Académique Documents
Professionnel Documents
Culture Documents
Table of Contents
TERMS OF WARRANTY................................................................3 CAUTION.........................................................................................3 HOW TO CONTACT US .............................................................4 DEFINITION OF WARNING, CAUTION, NOTE.............................6 GENERAL PRECAUTION ON ENVIRONMENT............................7 GENERAL PRECAUTION ON ELECTRIC SAFETY.....................8 SECTION 1. GENERAL INFORMATION .................9
1.1 Product Overview....................................................................9 1.2 Product Features...................................................................10 1.3 Product Configuration..........................................................10 SECTION 2. TECHNICAL DESCRIPTION...................................11 2.2 Main Unit.................................................................................12 1.2. Composition.........................................................................12 2.2. Main Board..........................................................................13 2.2.3Key Board.............................................................................25 2.3Doppler Probe.........................................................................26 2.3.1 Composition.......................................................................27 2.4UC Probe................................................................................27 2.4.1 Composition.......................................................................27 2.4.2 Technical Description.........................................................28 SECTION 3. TROUBLESHOOTING.............................................30
BT-300 Service Manual V2.0 3.1 Warning Sign Occasion and Status Checking...................30 Alarm Occasion and Status Checking Up.................................31 3.2 Troubleshooting on the Other Situation.............................31 SECTION 4. MAINTENANCE AND CLEANING..........................33 4.1 Maintenance and Cleaning...................................................33 Caution....................................................................................34 4.2 Regular Inspection................................................................34 SECTION 5. SPECIFICATIONS....................................................35
Terms of Warranty
This product is manufactured and examined by strict quality control and inspection system.
Compensation standard concerning repair, replacement and refund of the product complies with Consumers protection law issued by Economic Planning Dept.
BT-300 is guaranteed by BISTOS Co., Ltd. And its warranty period is two years from the date of purchase.
Warranty repair or replacement will be made by BISTOS Service Center with free of charge during warranty period if properly used under normal condition in accordance with the instructions for use.
In the event of a malfunction or failure during warranty period, customer should inform BISTOS Co., Ltd. of the model name, serial number, date of purchase and explanation of failure about the defective equipment.
Caution Federal law restricts this device to sale by or on the order of a physician
How to contact us
You can contact us the following address and telephone numbers for services and supplies.
Oversea Sales Team BISTOS Co., Ltd. Product and Order Inquiries 201, 2F, 239-15, Gasan-Dong, Geumcheon-Gu, Seoul, Korea Tel : +82-2-862-0642 Fax: +82-2-862-0644
Customer Service Team Service Request and Technical Support BISTOS Co., Ltd. 201, 2F, 239-15, Gasan-Dong, Geumcheon-Gu, Seoul, Korea Tel : +82-2-862-0642 Fax: +82-2-862-0644
Warning A Warning indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Caution A Caution indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
Note A Note indicates a particular point of information; something on which to focus your attention.
Avoid placing in an area exposed to moisture. Do not touch the equipment with wet hand. Avoid placing in an area where there is a high variation of temperature. Operating temperature ranges from 10C to 40C. Operating humidity ranges from 30% to 85%. Avoid placing in an area where there is an excessive humidity rise or ventilation problem. Avoid placing in an area where chemicals are stored or where there is in danger of gas leakage.
Avoid placing in an area where there is an excessive shock or vibration. Avoid dust and especially metal material into the equipment. Power off when the equipment is not fully installed. Otherwise, the equipment could be
Do not disjoint or disassemble the equipment. BISTOS Co., Ltd. takes no responsibility of it
(Power Adaptor Input: 100 ~ 250V AC, Power Adaptor Output: 16V, 2.8A). Be sure that the entire connection cable of the system is properly
BT-300 system consists of the followings. Unpack the package and check out the following items. Also, be sure to check to see any damage of main body and accessories BT-300 Main Body Ultrasound Transducer(Doppler Probe) - 2 EA(1 EA for BT-300S) Tocotransducer(UC Probe) - 1EA Event Marker -1EA Thermal Print Paper - 2EA Power Adaptor -1EA Power Cord -1EA Ultrasound Gel -1EA Abdominal Transducer(Probe) Belt -3 EA(2 EA for BT-300S) Users Manual -1EA
Contents
Name
Description
10
Main body of BT-300 (2-year Warranty) Ultrasound Transducer for Measuring FHR Pressure Sensor(Tocotransducer) for Measuring Uterine contraction Used for a specific event(for example, fetal Movement) Z-folder type thermal Paper Used for Holding Doppler Probe and/or UC Probe Adaptor for transform AC Power to DC 16V AC Power cord Ultrasonic coupling gel
Event Marker
BT-300 is capable of monitoring fetal heart rate and maternal uterine contractions. Simultaneous trends of fetal heart rates (FHR) and uterine activity (UA) are plotted by the built-in chart recorder. FHR and UA are displayed continuously at numeric display on the front panel. Measuring fetal heart rate is using 1MHz ultrasound wave which transmits the burst pulse to the 7 ultrasound sensors about 3.47KHz periodically. This wave is reflected to the part of different medium.
11
1.2.
12
2.2.
Main Board
13
J1
R 12 C 10 11 TL 06 4 1 + U 3A
N1 N2 N3
L M B 0 5 0 9 -6 P D O P 1 _ ID
13 V+ 4 V- 12 VC C 5 GND
4 U 1C TL064 R58 IV P 1 6 + U 4A 11 1 C 40 8
R 23
C 170
C 24
R 27
D2
AVP5 D4 R 33 C 29 R 34 R 35 R36 R 37 R 38 R39 C 30 R 42 R 43 6 5 C 34 AVP5 AVP12 C36 4 C 35 4 11 U 4B TL06 4 7 AVP5 + R 44 9 10 11 TL064 8 D O P1_H R + U 4C
D6 2 3 D7 4
TL0 64 +
C 37
R 48
R 49
R 50
R 51 R 54
R 53
U 6D 11 13 AVP5 12 TL064 14 +
D8 11 2 3 D9
U 6A TL 064 1 + R57 C 39 R 61
R 65
C 38 11 6
U6B TL064 7 +
R 56 R 60 R 62 R 55
R 59 C 41 C 42
5 D 1_R EF
4 AVP12 R 66 R 64 C 45
R 67 R 68 C46
J2
N1 N2 N3
L M B 0 5 0 9 -6 P
TR AN S D U C ER
D O P1_S
1 3 5 7
2 4 6 8
J5 A N A L_ S IG J7 1 2 3 5 6 8 R 79
AVP5
N1 N2 N3
L M B 0 5 0 9 -6 P
C 62
C 63
78L05
(MARK JACK)
14
1 1 1 1 1 1 1
+ G7 G6 G1 G2 G3 G4 G5
TD A 2030
14
U 2 2A 2
74H C 04
U 22B 4
7 4H C 0 4 TPH _C K
14 3
7 14
7 14
9 U 22D
8 74H C 04
TPH _D I
C LK VC C OE GND
U 2 2C
74H C 04
74H C 574 U 13
C 7 5 C 7 6 C 7 7 C 7 8 C 7 9 C 8 0 C 81 C 8 2 C 8 3 C 8 4 C 8 5 C 8 6 D VP5 U 43 R 96 D V P 1 .8 R TC _/C S R TC _AS R TC _D S 19 R C 0 / T 1 O S O /T 1 C L K 2 0 R A 0 /A N 0 R C 1 / T 1 O S I/ C C P 2 2 1 R A 1 /A N 1 R C 2/C C P 1 2 2 R A 2 /A N 2 R C 3 /S C K / S C L 2 3 R A 3 /A N 3 / R E F R C 4 / S D I/ S D A 2 4 R A 4 /T 0 C L K R A 5 /A N 4 / S S R C 5/S D O R C 6 /TX/C K 8 R B 0 /I N T R C 7 /R X/ D T 9 10 R B1 R D 0 /P S P 0 11 R B2 R D 1 /P S P 1 14 R B3 R D 2 /P S P 2 15 R B4 R B 5 R D 3 /P S P 3 16 R D 4 /P S P 4 17 R B6 R B7 R D 5 /P S P 5 R D 6 /P S P 6 30 R D 7 /P S P 7 3 1 O S C 1 /C L K O S C 2 /C L K O U T R E 0 /R D / A N 5 18 M C L R /V P P R E 1 /W R /A N 6 R E 2 /C S / A N 7 7 28 V D D VDD P IC 1 6 C 7 7 /T Q F P C 98 DVP5 D VP5 R 10 4 PD0 PD1 PD2 PD3 PD4 PD5 PD6 PD7 /LE D D _ W R IN I _ E N P D [ 0 .. 7 ] P IC D / L PD0 PD1 PD2 PD3 PD4 PD5 PD6 PD7 U 1 2D U 26 1 C 13 5 C 122 104 U 23 R 128 R 129 DVP5 4 VC C NC 1 P1 D -S U B 9 P 1 6 2 7 3 8 4 9 5 C 13 9 5 R 148 R 149 C 145 C 146 C2GND 15 TX RX 3 4 C1C2+ V+ V2 6 C 14 0 7 5 14 C1+ VDD 16 C 137 C 13 6 D VP5 74H C 04 9 8 IN I _ E N /K E Y D _R D 7 2 3 4 5 6 7 8 9 2 3 4 5 6 7 8 9 11 1 32 35 36 37 42 43 44 1 38 39 40 41 2 3 4 5 25 26 27 R T C _ R /W R S _D S P IF _S C K IF _S D I IF _S D O TX RX /S E G D _ W R /S E G E _ W R /L E D D _ W R /K E Y D _ R D /V O L D _ W R /P R N C _ W R /I D D _ R D /T P H _ L E /S E G V _ W R J1 8 1 R 164 2 /S E G E _W R D U A L / S IN G L E DVP5 PW M /S E G D _ W R PD0 PD1 PD2 PD3 PD4 PD5 PD6 PD7 2 3 4 5 6 7 8 9 11 1
D1 D2 D3 D4 D5 D6 D7 D8
C LK VC C OE GND
D V P 3 .3
U 16 89 88 87 86 85 88 43 11 24 25 91 68 15 26 11 13 02 75 53 63 4 128 111 1 06 76 70 57 45 00 34 31 4 TMS320VC 5402 TM S TCK T R S TT D I TDO E M U 1 /O F FE M U 0 Vss Vss VV ss ss Vss Vss VV ss ss Vss VV ss ss CC VV DD DD CVD D CVDD CVDD CVD D DVDD D VDD DVDD DVDD DD VV DD DD U 17 HD0 HD1 HD2 HD3 HD4 HD5 HD6 HD7 H C N TL0 H C N TL1 BC LKR 0 BFSR0 BD R0 B C LK X0 B F S X0 B D X0 BC LKR 1 BFSR1 BD R1 B C LK X1 B F S X1 B D X1 H IN T / T O U T 1 C LK M D 1 C LK M D 2 C LK M D 3 TO U T0 Vss C LK O U T X1 C L K IN / X 2 RS 58 69 81 95 1 20 1 24 1 35 6 39 46 41 43 45 48 53 59 42 44 47 49 54 60 51 77 78 79 82 93 94 96 97 98 D V P 3.3 C 1 03 J 11 U 12A 2 14 D VP5 R 111 IF _ S C K IF _ S D I 74 H C 0 4 7 1 R 112 IF _ S D O 1 2 3 4 5 6 7 VPP D VP5 PD7 PD6 /K E Y D _ R D /I D D _ R D U 1 8 ,24 P C 1 11 1 8 2 TC R ESET 7 3 RESET Vs 6 4 CK RCT 5 GND VCC M M 1 07 5 C 100 VPP 1 3 P IC D /L R S T R 1 01 J10 2 4 DVP5
D0 D1 D2 D3 D4 D5 D6 D7
D [ 0 .. 7 ]
99 10 0 10 1 10 2 10 3 10 4 11 3 11 4 11 5 11 6 11 7 11 8 11 9 12 1 12 2 12 3 13 1 13 2 13 3 13 4 13 6 13 7 13 8 13 9 14 0 14 1 5 7 8 9 10 11 10 5 10 7 10 8 10 9
D0 D1 D2 D3 D4 D5 D6 D7 D8 D9 D 10 D 11 D 12 D 13 D 14 D 15 A0 A1 A2 A3 A4 A5 A6 A7 A8 A9 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19
C V D D : 1 .8 V
D V D D : 3 .3 V
C95
C 96
DVP5
C 99
C LK VC C OE GND
74H C 574 U18 PD0 PD1 PD2 PD3 PD4 PD5 PD6 PD7 4 5 6 7 8 9 10 11 12 1 18 AD0 19 A D 1 IR Q 23 AD 2 SQ W 24 AD 3 24 AD4 AD5 13 AD 6 C S 14 AD 7 AS 15 G N D R /W 17 MOT DS RST D S 1 2C 88 7 U 19 D1 D2 D3 D4 D5 D6 D7 D8
TM S 3 2 0 V C5 4 0 2
R T C _ /C S R TC _AS R T C _ R /W R TC _D S
DVP5
C LK VC C OE GND
D V P 3 .3
74H C 574 U 21 A0 A1 A2 A3 A4 A5 A6 A7
C 114
U 12B 4
14
S S T3 9 V F 51 2
129 19 22 7 62 15 85 17 13 25 24 22 23 21 12 90 67 66 65 64 63 61 32 31 23 09 28 22 76
1 20 1 9 D IR V C C 1 0 OE GND
R 16 2 1 5 .9 7 4 4 M D V P 3.3 D V P 3 .3 R 139 D O P1_H R R 141 D O P2_H R R 142 UC 1 2 3 4 5 6 7 U 25 CH0 VDD CH1 VREF CH2 AGND CH3 C LK NC DOUT NC D IN D G N D C S /S H D N M C P 3 2 0 4 -B I/ S L 14 13 12 11 10 9 8 C 13 2 R 14 0 C 133 C 1 34 + D V P 3.3 3 OUT GND 2 N1 N2
1 5 .7 6 M H z O S C (5 V )
14 1 3 T1 O U T 7 R 1 IN 8 T2 O U T R 2 IN
C LK VC C OE GND
M A X 2 3 2 C S E ( S MD )
74H C 574 U 27 PD0 PD1 PD2 PD3 PD4 PD5 PD6 PD7 /P R N C _W R IN I _ E N C 163 2 3 4 5 6 7 8 9 11 1 D1 D2 D3 D4 D5 D6 D7 D8
(FEMALE)
U 39 A
R 14 5 R 1 46 R 1 4 7
AD C _C K AD C _R AD C _X AD C _C S 3
4K DVP5
74H C 04 U 39 B 4 3 2K
U 3 9 ,1 4 P
C LK VC C OE GND
U 4 0B R X_I 6
74H C 574 74H C 04 U 30 PD0 PD1 PD2 PD3 PD4 PD5 PD6 PD7 C 165 /I D D _ R D 2 3 4 5 6 7 8 9 A0 A1 A2 A3 A4 A5 A6 A7
P V P 24 2
D10 1
L3
U 41 B 5 4 74H C 08
16 K 4K U 4 1 ,1 4 P
DVP5
C 143 +
R 150 1 2
U 29 PQ 2C F1 R 151 4 D O P1_P 3 1 U 3 2 ,14 P 74H C 08 U 32B D O P1_I 6 5 4 74H C 08 U 32C 10 D O P1_Q J 16 74H C 08 5 2 6 7-2 U35 L M 2 5 7 6 S -5 . 0 4 L5 2 2 FB V IN VOUT GND SW D O P2_I 5 3 6 4 74H C 08 U 37C 10 D O P2_Q 8 9 74 H C 08 74H C 08 11 12 3 1 U 3 7 ,1 4 P 1 74H C 08 U 37B 5 C 157 U 37A 2 D O P2_P DVP5 8 9 R X_Q 74 H C 08 11 12 R X_I C 166 C 1 54 U 4 2 ,1 4 P D VP5 + C 147 U 32A 2 R X_Q 8
U 4 0C 10 9
1 C LR 4 PRE 74H C 74
SW
5 C 144 R 152
FB GND
V IN
Vc
U 39 D U 41C U 41D 11 10 9 7 4H C 0 8 U 41A 1 U 42B 3 9 Q 8 Q 2 1M 74 H C 0 8 11 U 39E 74H C 04 10 7 AN A_R S 12 16K 74H C 04 74 H C 0 8 1 1 5 .97 4 4 M 8 9 8K
7 4 H C 08 DVP5
13 12 2K
A GND VCC C LR
U 31 D V P 3.3 C 1 53 10 4 2 4 OUT 4 + C 15 0 G N D /A D J IN 3
IV P 1 6
L4
J15 H T J - 0 2 0 -0 4
U 3 2D 13 G ATE1 1M DVP5
D C LK
11
L M 1 1 1 7 M P X- 3 .3
2 1
C 151
C 1 52
23
13 10 C LR PRE 74H C 74
/A N A _R S 7 U 34A 5 6 Q Q DVP5 U 3 4,1 4 P C 161 U 3 7D 13 G ATE2 U 34B 9 8 Q Q D C LK 12 11 1M DVP5 DUAL 7 2K 7 4H C 7 4 DVP5 D C LK 14 2 3 DVP5 DUAL 16K 8K U 38A 74H C 393 3 4 QA 5 QB 6 QC QD
DVP5
A VCC C LR GND
1 2
C 162
C 159
D 11
1 C LR 4 PRE
L M 1 1 1 7 M P X- 1 .8 1
A VCC C LR GND
13 12 AN A_R S
V 1 V2 1 V2 12 V2 31 V2 14 V2 15 V2 16 V2 71 V2 81 V2 19 V3 01 3 1 DVP5 1 11 1 11 1 11 1 1
13 C LR 10 PRE
Technical Description Doppler signal generation circuit The burst signal drives the ultrasound sensor using by Q1 ~
15
1 1 11 1 11 1 1 11 1 11 1 11 1 1 11 1 11 1 11 1 1 1 V 9 V1 9 V2 9 V3 9 V4 9 V5 9 V6 9 V7 9 V8 9 V9 1 V0 10 V0 1 V0 21 V0 13 V0 41 V0 51 V0 61 V0 71 V0 81 V0 91 V1 10 V1 1 V1 21 V1 31 V1 41 V1 51 V1 16 V1 71 V1 18 V1 91 2 0
1 1 11 1 11 1 1 11 1 11 1 11 1 1 11 1 11 1 11 1 1 1 V 6 V1 6 V2 6 V3 6 V4 6 V5 6 V6 6 V7 6 V8 6 V9 7 V0 7 V1 7 V2 7 V3 7 V4 7 V5 7 V6 7 V7 7 V8 7 V9 8 V0 8 V1 8 V2 8 V3 8 V4 8 V5 8 V6 8 V7 8 V8 8 V9 9 0
A/D converter & DSP circuit MCP3204(U28) A/D converter is possible to convert to 4 channels and converts each signal DOP_1HR, DOP2_HR, UC to 12 bits digital signal by 200Hz periodically. The range of analog input signal of this A/D converter must be in 0 ~ +3.3V. The converted digital signal is transmitted to DSP(U16) by serial and calculated an useful value using many signal processing techniques. DSP is TMS320VC5402 by TI company and the fetal heart rate is calculated at 0.25 second using autocorrelation algorithm. The output is transmitted to CPU to display and record. TX burst pulse generation circuit U33, U34, U37~42 generate the burst signal for driving the ultrasound sensor and control signal for demodulation. The timing diagram is below. < Timing Diagram > 3.47KHz
16
CPU circuit CPU(U43) is PIC16F77, when it has malfunction, CPU resets itself using MM1075(U17) watch-dog IC. FHR and UA are displayed on the 7-segment and signal quality is displayed on the LED. Thermal print is operated +24V. CPU controls the stepping motor and TPH to record the calculated data on the thermal paper. Door open or paper out is recognized by sensor. If that situation happens during recording, then CPU stops recording and displays See Prn on the display panel. Time and date are calculated automatically at the DS12C887(U18). CPU reads the value and records on the recording paper. Each key is recognized by CPU which performs each operation.
17
J1
J2 S IG N A L N 1 N 2 N 3
S IG N A L
N 1 N 2 N 3
L M B 0 5 0 9 -6 P D O P 1 _ ID
L M B 0 5 0 9 -6 P
D O P 2 _ ID
Pin 1 2 3 5 6 8
I/O I O I/O I
Description Acoustic stimulator enable signal Ground +12V power Ground Transmitting and signal Identity signal
receiving
ultrasound
B. UC Probe Connector
J7 1 2 3 5 6 8
AVP12
S IG N A L U C _ ID
N 1 N 2 N 3
L M B 0 5 0 9 -6 P
Pin 1 2 3 5 6 8
Name I/O NC GND Ground AVP12 O +12V power Signal I UC out signal NC UC_ID I Identity signal C. Mark Jack Connection
Description
18
5 2 6 7 -2
Pin 1 2
I/O I -
AVP5
J3 AN AL_C O N 1 3 5 7 1 3 5 7
J5 A N A L _ S IG
AVP12
Name Signal DOP2_P VOL2_0 VOL2_1 VOL2_2 DOP2_Q DOP2_I DOP2_S DOP2_HR AVP5 AVP12
Description Transmitting and receiving ultrasound signal Digital pulse for transmitting Volume control signal 0 Volume control signal 1 Volume control signal 2 180 control signal for demodulation 0 control signal for demodulation Doppler sound signal Doppler heart rate signal +5V power +12V power
E. Speaker Connector
19
5 2 6 7 -2
Pin 1 2 F.
I/O O -
Pin 1 2
I/O I I
Description +16V input power +16V input power through power switch
Pin 1 2 3 4
I/O I O O
Description Power for program writing to CPU CPU reset signal CPU reset signal
20
S IG N A L
J18 1
D U A L /S IN G L E
Pin 1 2 I.
I/O O I
P IC D /L
This connector is used only to download the CPU program. J. Central Monitor Interface Connector (RS-232C)
P1 D -S U B 9 P 1 6 2 7 3 8 4 9 5
R X TX
N 1 N 2
(FEMALE)
Pin 2 3 5
Name RX TX GND
I/O I O -
Description Receiving signal for central monitor Transmitting signal for central monitor Ground
21
1 1 1 1 1 2 2 2 2 2 3 3 3 3 3
0 2 4 6 8 0 2 4 6 8 0 2 4 6 8 0
H E A D E R 20X2
Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25
Name LED_H1G LED_H1R LED_H2G LED_H2R LED_REC/ZOOM LED_ALM/REC LED_INFO LED_ALM SEG_V1 SEG_V2 SEG_V3 SEG_V4 SEG_V5 SEG_V6 SEG_V7 VOL_UP1/ALM VOL_DN1/VOL_UP VOL_UP2 VOL_DN2/VOL_D N REC/ZOOM UC_REF ALM/REC SEG0 SEG1 SEG2
I/O O O O O O O O O O O O O O O O I I I I I I I O O O
Description FHR1 green rhythm LED signal FHR1 red rhythm LED signal FHR2 green rhythm LED signal FHR2 red rhythm LED signal Record(D) or zoom(S) LED signal Alarm(D) or record(S) LED signal Information LED signal Alarm LED signal Volume 1 step display signal Volume 2 step display signal Volume 3 step display signal Volume 4 step display signal Volume 5 step display signal Volume 6 step display signal Volume 7 step display signal DOP1 volume up(D) or alarm(S) key DOP1 volume down(D) or volume up(S) key DOP2 volume up key DOP2 volume down(D) or volume down(S) key Record(D) or zoom(S) key UC reference key Alarm(D) or record(S) key 7-segment 0 enable signal 7-segment 1 enable signal 7-segment 2 enable signal
22
L. TPH Connector
PVP24 J9 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 15P
D VP5 S TR O B E 0 S TR O B E 1 TP H _C K /T P H _ L E TP H _D I
Pin 1 2 3 4 5 6 7 8
I/O O O O O O
Description +24V power +24V power Ground Ground +5V power TPH strobe 0 signal TPH strobe 1 signal
23
M. Motor Connector
J12 3 6 11 14 U 28 B B B B 1 2 3 4 C C C C 1 2 3 4 2 7 9 16 1 8 M M M M C C C C K K K K 0 1 2 3 1 2 3 4 5 6
PVP24 IV P 1 6
G 2 G 1
K1&4 K2&3
6P
Pin 1 2 3 4 5 6
I/O O O O O O O
Motor clock 0 Motor clock 1 Motor clock 2 Motor clock 3 +24V power +24V power
5 2 6 7 -2
Pin 1 2
I/O I -
24
74H C 04
14
R 161
R 159
Pin 1 2 3
I/O O O I
25
G C D P D E
5 4 3 2 1
G C D P D E
5 4 3 2 1
D 1 SEG _V7 U 3 S EG _E S EG _D S EG _C S S S S E E E E G G G G _E _D _G _C 1 2 3 4 5 6 7 8 9 1 1 1 1 2 2 2 2 2 E D C D E D G C D 1F 1G 1A P 1B 1C O M 2C O M 2F 2A P 2B 18 17 16 15 14 13 12 11 10 S S S S S S S S S E E E E E E E E E G G G G G G G G G _F _G _A _B E6 E7 _F _A _B SEG _V6 SEG _V5 SEG _V4 SEG _V3 SEG _V2 SEG _V1 1 2 3 4 5 6 7 8 D P G F E D C B A 16 15 14 13 12 11 10 9 16 15 14 13 12 11 10 9
H E A D E R 20X2
SW 1 V O L _ U P 1 /A L M 1 2 V O L _ D N 1 /V O L _ U P 1
SW 2 2 VO L_U P2 1
SW 3 2 V O L _ D N 2 /V O L _ D N 1
SW 4 2 SW 6
D 3
R E C /Z O O M
D 5 D 6 L E D _ R E C /Z O O M
D 4
Technical Description FHR and UA are displayed on the 7-segment U1, U2, U3 and signal quality is displayed on the two colors LED D2, D4. D3, D5, D6, D7 LED display the state of alarm, recording, zoom and information operation respectively. SW1 ~ SW7 are key switches. D1 is displayed the volume step. Connection This connection is same as the key board connection in main board.
2.3
Doppler Probe
26
2.4
UC Probe
2.4.1 Composition
27
28
JP1 1
VR 1 2 R 3 AV P12 R 6 R 7 1 R 4
VR 2
R 9
R 10 4
U 1A R 11 AVP12 C 1 U C _R EF
R 12
U C _R EF
U 2 U C _R EF 1 O U T IN 3
AVP12
C 3 G N D C 2
U 1B
G 1 G 2 G 3G 4
V 1 V5 1 V6 1 V7 1 V 2 V 3 V 4 V 5 V 6 V 7 V 8 V 9 V 1 V0 1 V1 1 V 2 1 V3 1 4
78L05
1 1 1 1
T i t le S iz e A D a te :
Technical Description
The uterine contraction is converted to electrical signal by using pressure sensor strain gauge. This signal is compensated the difference of production by adjusting the offset and gain. And it is transmitted to main board. Pin 1 2 3 4 Connection Name UC_OUT 1 GND1 AVP12 SGND1 I/O O I Description UC output Ground +12V power Shield ground
29
1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1
R 8
TL064
10
11
TL064
Section 3. Troubleshooting
3.1Warning Sign Occasion and Status Checking 1 2 After the monitor power on, Information sound rings once. If the ultrasound transducer pulls out of the connector during monitoring(error #4), it rings once every 3 seconds until the transducer is connected to the connector or pressing the ALM(alarm) pushbutton causes the alarm function disable. 3 When the REC(record) pushbutton is pressed for recording in case of print door open or under recording, when the door is opened, it rings once every 3 seconds and ERR 1 - REC ERR 1 in BT-300D -appears in the FHR and UC display as shown below until the door is closed or pressing the ALM(alarm) pushbutton causes the alarm function to be disabled and return to the normal monitoring mode.
[Figure 4.2 Front Panel for BT-300D] 4 When the REC(record) pushbutton is pressed for recording in case of paper empty or printer door open, or when the paper runs out under recording, it rings once every 3 seconds and ERR 2 - REC ERR 2 in BT-300D - appears in the FHR and the UC display until the paper is installed or pressing the ALM(alarm) pushbutton causes the alarm function to be disabled and go back to the normal monitoring mode.
30
Error Number 1 2 3 4 5
Printer door open & Paper Close the door & Insert the empty paper Dop1 probe open under normal operation Dop2 probe open under normal operation Connect the Dop1 probe Connect the Dop2 probe
When the measured FHR is beyond the alarm setting range for more than 20 seconds during monitoring, it rings every 3 second until the FHR is within the alarm setting range at least once or pressing the ALM(alarm) pushbutton causes the alarm function to be disabled on purpose.
31
3.2.3 UC operation
Not display the UC value 1 Check UC signal in Main Board U7C - 8pin 2 Refer to display operation 3.2.4 Sound operation
1 2
Not hear the Alarm / Information sound Check the PWM U43 - 36pin Check the Audio Amp U8
32
33
34
Section 5. Specifications
Ultrasound Mode Technique: Pulse Repetition Frequency: Pulse Duration: Center Frequency: Intensity: Heart Rate Counting Range: FHR Accuracy:
Leakage Current: Dual Fetus
Pulsed Doppler Wave / Autocorrelation Processing 3.5 KHz 128 1MHz <10mW/cm2 50 ~ 240 BPM 2% of range
<10
Uterine Activity Mode Range, Tocotransducer: Excitation Voltage: Resolution: Bandwidth, Tocotransducer: 0 ~ 99 Relative Units +5.0Vdc 1 Relative Unit DC ~ 0.5Hz
Power Requirements Line Voltage: Line Frequency: Power Consumption: 100 ~ 250VAC 50 or 60Hz 80VA max
Physical Characteristic Height: Width: Depth: Weight: 19.1cm (7.5 in) 18.7cm(7.4 in) 20.1cm(7.9 in) 6Kg(13.2 lbs)
Environmental Characteristics Operating Temperature: Operating Humidity: Storage Temperature: 60(131) Storage Humidity: Operating Atmospheric Pressure: Storage Atmospheric Pressure: 10(50) to 40(104) 30% ~ 85% non-condensing -10(14) to 20% ~ 95% non-condensing 70 ~ 106 70 ~ 106
35
Printer Thermal Array Type Print Speed: 10, 20, 30 mm/min Paper Feeding Function Print Contrast: 1, 2, 3 steps Auto Print Period: 0(Off), 10, 20, 30, 40, 50, 60min Display 7-Segment LED 2 Channels (HR, UC)-BT-300S/3 Channels (HR1, HR2, UC)-BT-300D Indicators Heart Rhythm (Green: Stable, Red: Unstable) Alarm On/Off State Print On/Off State AC Power (Green LED) Sound Doppler Sound with Volume Control (8 steps) Alarms Sound: exceed FHR Range Information Sound: Doppler Probe Off, Paper Off, Door Open, Watch Dog Set-up Alarm Upper/Lower Limit Value Print Speed Print Contrast Auto Print Period Function Event Mark Function Zoom in Function Doppler2 Offset Function
36
Service Telephone and Fax. Numbers Telephone: 82-2-8620642, 0643 Fax: 82-2-8620644 BISTOS Co., Ltd 201, 2F, 239-15, Gasan-Dong Geumcheon-Gu, Seoul, Korea Model Name: BT-300 www.bistos.co.kr bistos@bistos.co.kr EC Representative: Econet Medical Gmbh High-t'Park Mainstr. 6c-6d D-45768 Marl/Germany Telephone: +49 2365 92 437 - 0 Fax: +49 2365 92 437 55
37