Type A and B uncertainty analysis For ASTM E74 Calibrations

COMPANY TRANSDUCER MANUFACTURER TRANSDUCER S/N CAPACITY ASTM E74 LLF for K=2.4
THE LOWEST FORCE OR TORQUE POINT USED IN THE LOADING RANGE

PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K=1 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) CAL DATE

MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 200 LBF 0.001% 9/26/13

ENTER YOUR CALIBRATION INFORMATION IN HIGHLIGHTED COLUMNS

CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1 TRANSDUCER UNCERTAINTY IN % FOR LOWEST FORCE APPLIED K =2.4 TRANSDUCERUNCERTAINTY IN % FOR LOWEST FORCE APPLIED K =1

0.00500% 0.00208% 0.25000% 0.10417%

Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

ASTM E74 OR E2428 LLF % at the lowest calibration force or torque to be used Combined Type A Uncertainty

0.10417%

normal

32

1.04E-3 1.04E-3

1.09E-6 1.09E-6

99.91% 999.07E-3

36.79E-15 36.79E-15

Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

0.001% 0.000% 0.004% 0.0015% 0.003% 0.000%

normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty

10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 1.04E-3 32 2.04 2.12E-3

100.00E-12 000.00E+0 533.36E-12 75.00E-12 300.02E-12 000.00E+0 1.09E-6

0.01% 0.00% 0.05% 0.01% 0.03% 0.00% 100.00%

100.00E-24 000.00E+0 2.84E-21 56.26E-24 900.11E-24 000.00E+0 36.79E-15

FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % AT THE LOWEST FORCE TO BE APPLIED) 0.10422%

0.208% 0.417 LBF

Label 1 2 3 4 5 6 7 8 9 10 Average Std. Dev Variance Reproducibility Repeatability

1 0.00201 0.00201 0.00200 0.00200 0.00202 0.00200 0.00199 0.00200 0.00200 0.00199 0.002002 9.19E-6 84.44E-12 11.6762E-6 18.827E-6

Repeatability and Reproducibility PLUS ANOVA 2 3 4 5 6 0.00201 0.00200 0.00199 0.00199 0.00195 0.00203 0.00201 0.00200 0.00200 0.00201 0.00195 0.00200 0.00200 0.00201 0.00196 0.00199 0.00198 0.00203 0.00202 0.00204 0.001987 0.00201 #DIV/0! #DIV/0! #DIV/0! 25.84E-6 17.64E-6 #DIV/0! #DIV/0! #DIV/0! 667.78E-12 311.11E-12 #DIV/0! #DIV/0! #DIV/0! F calc Maximum Std. Dev. 25.84E-6 3.8464 F crit Minimum Std. Dev. 9.19E-6 3.3541

If F calc > F crit , there is significance of Reproducibility data Include Reproducibility

PLUS ANOVA 7 8 9 10

#DIV/0! #DIV/0! #DIV/0!

#DIV/0! #DIV/0! #DIV/0!

Between Groups MS Within Groups MS

#DIV/0! #DIV/0! #DIV/0! 1.36E-9 354.44E-12 P-Value 33.90E-3

#DIV/0! #DIV/0! #DIV/0! df Numerator 2 df Denominator 27

ity data

Type A and B uncertainty analysis


COMPANY TRANSDUCER MANUFACTURER TRANSDUCER S/N CAPACITY ASTM E74 LLF for K=2.4 THE FORCE OR TORQUE POINT YOU WISH TO CALCULATE COMBINED UNCERTAINY FOR PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K=1 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) CAL DATE

MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 10000 LBF 0.001% 0.000% 9/26/2013

ENTER YOUR CALIBRATION INFORMATION IN HIGHLIGHTED COLUMNS

CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1

0.00500% 0.00208% 0.00500% 0.00208%

Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

ASTM E74 OR E2428 LLF % Combined Type A Uncertainty

0.00208%

normal

32

20.83E-6 20.83E-6

434.03E-12 434.03E-12

30.09% 300.90E-3

5.89E-21 5.89E-21

Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

0.001% 0.000% 0.004% 0.002% 0.003% 0.000%

normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty

10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 37.98E-6 212 1.97 74.87E-6

100.00E-12 000.00E+0 533.36E-12 75.00E-12 300.02E-12 000.00E+0 1.44E-9

6.93% 0.00% 36.98% 5.20% 20.80% 0.00% 100.00%

100.00E-24 000.00E+0 2.84E-21 56.26E-24 900.11E-24 000.00E+0 9.79E-21

FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00380%

0.008% 0.760 LBF

Type A and B uncertainty analysis


COMPANY TRANSDUCER MANUFACTURER TRANSDUCER S/N CAPACITY ASTM E74 LLF for K=2.4 THE FORCE OR TORQUE POINT YOU WISH TO CALCULATE COMBINED UNCERTAINY FOR PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K=1 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) CAL DATE

MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 9000 LBF 0.001% 0.000% 9/26/2013

ENTER YOUR CALIBRATION INFORMATION IN HIGHLIGHTED COLUMNS

CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1

0.00500% 0.00208% 0.00556% 0.00231%

Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

ASTM E74 OR E2428 LLF % Combined Type A Uncertainty

0.00231%

normal

32

23.15E-6 23.15E-6

535.84E-12 535.84E-12

34.70% 346.99E-3

8.97E-21 8.97E-21

Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

0.001% 0.000% 0.004% 0.002% 0.003% 0.000%

normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty

10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 39.30E-6 185 1.97 77.53E-6

100.00E-12 000.00E+0 533.36E-12 75.00E-12 300.02E-12 000.00E+0 1.54E-9

6.48% 0.00% 34.54% 4.86% 19.43% 0.00% 100.00%

100.00E-24 000.00E+0 2.84E-21 56.26E-24 900.11E-24 000.00E+0 12.87E-21

FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00393%

0.008% 0.707 LBF

Type A and B uncertainty analysis


COMPANY TRANSDUCER MANUFACTURER TRANSDUCER S/N CAPACITY ASTM E74 LLF for K=2.4 THE FORCE OR TORQUE POINT YOU WISH TO CALCULATE COMBINED UNCERTAINY FOR PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K=1 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) CAL DATE

MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 8000 LBF 0.001% 0.000% 9/26/2013

ENTER YOUR CALIBRATION INFORMATION IN HIGHLIGHTED COLUMNS

CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1

0.00500% 0.00208% 0.00625% 0.00260%

Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

ASTM E74 OR E2428 LLF % Combined Type A Uncertainty

0.00260%

normal

32

26.04E-6 26.04E-6

678.17E-12 678.17E-12

40.21% 402.10E-3

14.37E-21 14.37E-21

Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

0.001% 0.000% 0.004% 0.002% 0.003% 0.000%

normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty

10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 41.07E-6 155 1.98 81.12E-6

100.00E-12 000.00E+0 533.36E-12 75.00E-12 300.02E-12 000.00E+0 1.69E-9

5.93% 0.00% 31.62% 4.45% 17.79% 0.00% 100.00%

100.00E-24 000.00E+0 2.84E-21 56.26E-24 900.11E-24 000.00E+0 18.27E-21

FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00411%

0.008% 0.657 LBF

Type A and B uncertainty analysis


COMPANY TRANSDUCER MANUFACTURER TRANSDUCER S/N CAPACITY ASTM E74 LLF for K=2.4 THE FORCE OR TORQUE POINT YOU WISH TO CALCULATE COMBINED UNCERTAINY FOR PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K=1 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) CAL DATE

MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 7000 LBF 0.001% 0.000% 9/26/2013

ENTER YOUR CALIBRATION INFORMATION IN HIGHLIGHTED COLUMNS

CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1

0.00500% 0.00208% 0.00714% 0.00298%

Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

ASTM E74 OR E2428 LLF % Combined Type A Uncertainty

0.00298%

normal

32 9

29.76E-6 29.76E-6

885.77E-12 885.77E-12

46.76% 467.63E-3

24.52E-21 24.52E-21

Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

0.001% 0.000% 0.004% 0.002% 0.003% 0.000%

normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty

10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 43.52E-6 126 1.98 86.13E-6

100.00E-12 000.00E+0 533.36E-12 75.00E-12 300.02E-12 000.00E+0 1.89E-9

5.28% 0.00% 28.16% 3.96% 15.84% 0.00% 100.00%

100.00E-24 000.00E+0 2.84E-21 56.26E-24 900.11E-24 000.00E+0 28.42E-21

FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00435%

0.009% 0.609 LBF

Type A and B uncertainty analysis


COMPANY TRANSDUCER MANUFACTURER TRANSDUCER S/N CAPACITY ASTM E74 LLF for K=2.4 THE FORCE OR TORQUE POINT YOU WISH TO CALCULATE COMBINED UNCERTAINY FOR PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K=1 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) CAL DATE

MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 6000 LBF 0.001% 0.000% 9/26/2013

ENTER YOUR CALIBRATION INFORMATION IN HIGHLIGHTED COLUMNS

CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1

0.00500% 0.00208% 0.00833% 0.00347%

Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

ASTM E74 OR E2428 LLF % Combined Type A Uncertainty

0.00347%

normal

32 9

34.72E-6 34.72E-6

1.21E-9 1.21E-9

54.45% 544.54E-3

45.42E-21 #DIV/0! 45.42E-21

Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

0.001% 0.000% 0.004% 0.002% 0.003% 0.000%

normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty

10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 47.05E-6 99 1.98 93.36E-6

100.00E-12 000.00E+0 533.36E-12 75.00E-12 300.02E-12 000.00E+0 2.21E-9

4.52% 0.00% 24.09% 3.39% 13.55% 0.00% 100.00%

100.00E-24 000.00E+0 2.84E-21 56.26E-24 900.11E-24 000.00E+0 49.32E-21

FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00471%

0.009% 0.565 LBF

Type A and B uncertainty analysis


COMPANY TRANSDUCER MANUFACTURER TRANSDUCER S/N CAPACITY ASTM E74 LLF for K=2.4 THE FORCE OR TORQUE POINT YOU WISH TO CALCULATE COMBINED UNCERTAINY FOR PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K=1 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) CAL DATE

MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 5000 LBF 0.001% 0.000% 9/26/2013

ENTER YOUR CALIBRATION INFORMATION IN HIGHLIGHTED COLUMNS

CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1

0.00500% 0.00208% 0.01000% 0.00417%

Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

ASTM E74 OR E2428 LLF % Combined Type A Uncertainty

0.00417%

normal

32 9

41.67E-6 41.67E-6

1.74E-9 1.74E-9

63.26% 632.58E-3

94.19E-21 #DIV/0! 94.19E-21

Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

0.001% 0.000% 0.004% 0.002% 0.003% 0.000%

normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty

10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 52.39E-6 76 1.99 104.34E-6

100.00E-12 000.00E+0 533.36E-12 75.00E-12 300.02E-12 000.00E+0 2.74E-9

3.64% 0.00% 19.43% 2.73% 10.93% 0.00% 100.00%

100.00E-24 000.00E+0 2.84E-21 56.26E-24 900.11E-24 000.00E+0 98.09E-21

FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00524%

0.010% 0.524 LBF

Type A and B uncertainty analysis


COMPANY TRANSDUCER MANUFACTURER TRANSDUCER S/N CAPACITY ASTM E74 LLF for K=2.4 THE FORCE OR TORQUE POINT YOU WISH TO CALCULATE COMBINED UNCERTAINY FOR PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K=1 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) CAL DATE

MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 4000 LBF 0.001% 0.000% 9/26/2013

ENTER YOUR CALIBRATION INFORMATION IN HIGHLIGHTED COLUMNS

CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1

0.00500% 0.00208% 0.01250% 0.00521%

Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

ASTM E74 OR E2428 LLF % Combined Type A Uncertainty

0.00521%

normal

32 9

52.08E-6 52.08E-6

2.71E-9 2.71E-9

72.90% 729.01E-3

229.96E-21 #DIV/0! 229.96E-21

Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

0.001% 0.000% 0.004% 0.002% 0.003% 0.000%

normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty

10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 61.00E-6 59 2.00 122.06E-6

100.00E-12 000.00E+0 533.36E-12 75.00E-12 300.02E-12 000.00E+0 3.72E-9

2.69% 0.00% 14.33% 2.02% 8.06% 0.00% 100.00%

100.00E-24 000.00E+0 2.84E-21 56.26E-24 900.11E-24 000.00E+0 233.86E-21

FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00610%

0.012% 0.488 LBF

Type A and B uncertainty analysis


COMPANY TRANSDUCER MANUFACTURER TRANSDUCER S/N CAPACITY ASTM E74 LLF for K=2.4 THE FORCE OR TORQUE POINT YOU WISH TO CALCULATE COMBINED UNCERTAINY FOR PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K=1 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) CAL DATE

MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 3000 LBF 0.001% 0.000% 9/26/2013

ENTER YOUR CALIBRATION INFORMATION IN HIGHLIGHTED COLUMNS

CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1

0.00500% 0.00208% 0.01667% 0.00694%

Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

ASTM E74 OR E2428 LLF % Combined Type A Uncertainty

0.00694%

normal

32 9

69.44E-6 69.44E-6

4.82E-9 4.82E-9

82.71% 827.06E-3

726.78E-21 #DIV/0! 726.78E-21

Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

0.001% 0.000% 0.004% 0.002% 0.003% 0.000%

normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty

10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 76.36E-6 46 2.01 153.71E-6

100.00E-12 000.00E+0 533.36E-12 75.00E-12 300.02E-12 000.00E+0 5.83E-9

1.71% 0.00% 9.15% 1.29% 5.15% 0.00% 100.00%

100.00E-24 000.00E+0 2.84E-21 56.26E-24 900.11E-24 000.00E+0 730.68E-21

FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.00764%

0.015% 0.458 LBF

Type A and B uncertainty analysis


COMPANY TRANSDUCER MANUFACTURER TRANSDUCER S/N CAPACITY ASTM E74 LLF for K=2.4 THE FORCE OR TORQUE POINT YOU WISH TO CALCULATE COMBINED UNCERTAINY FOR PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K=1 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) CAL DATE

MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 2000 LBF 0.001% 0.000% 9/26/2013

ENTER YOUR CALIBRATION INFORMATION IN HIGHLIGHTED COLUMNS

CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1

0.00500% 0.00208% 0.02500% 0.01042%

Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

ASTM E74 OR E2428 LLF % Combined Type A Uncertainty

0.01042%

normal

32 9

104.17E-6 104.17E-6

10.85E-9 10.85E-9

91.50% 914.97E-3

3.68E-18 #DIV/0! 3.68E-18

Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

0.001% 0.000% 0.004% 0.002% 0.003% 0.000%

normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty

10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 108.90E-6 38 2.02 220.46E-6

100.00E-12 000.00E+0 533.36E-12 75.00E-12 300.02E-12 000.00E+0 11.86E-9

0.84% 0.00% 4.50% 0.63% 2.53% 0.00% 100.00%

100.00E-24 000.00E+0 2.84E-21 56.26E-24 900.11E-24 000.00E+0 3.68E-18

FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.01089%

0.022% 0.436 LBF

Type A and B uncertainty analysis


COMPANY TRANSDUCER MANUFACTURER TRANSDUCER S/N CAPACITY ASTM E74 LLF for K=2.4 THE FORCE OR TORQUE POINT YOU WISH TO CALCULATE COMBINED UNCERTAINY FOR PRIMARY FORCE CALIBRATION STANDARD UNCERTAINTY K=1 PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY K=1 (IF APPLICABLE) CAL DATE

MOREHOUSE MOREOUSE U-7643 10000 LBF 0.5 LBF 1000 LBF 0.001% 0.000% 9/26/2013

ENTER YOUR CALIBRATION INFORMATION IN HIGHLIGHTED COLUMNS

CALCULATED VALUES TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=2.4 TRANSDUCER UNCERTAINTY IN % FOR FULL SCALE K=1
TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =2 TRANSDUCER UNCERTAINTY IN % FOR THE FORCE UNC IS TO BE CALCULATED FOR K =1

0.00500% 0.00208% 0.05000% 0.02083%

Type A Uncertainty %
Uncertainty Description Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

ASTM E74 OR E2428 LLF % Combined Type A Uncertainty

0.02083%

normal

32

208.33E-6 208.33E-6

43.40E-9 43.40E-9

97.73% 977.29E-3

58.87E-18 #DIV/0! 58.87E-18

Type B Uncertainty %
Uncertainty Description PRIMARY OR SECONDARY CALIBRATION STANDARD UNCERTAINTY PRIMARY ELECTRICAL CALIBRATION STANDARD UNCERTAINTY (IF APPLICABLE) STABILITY OF THE SECONDARY STANDARD OVER TIME TEMPERATURE ERROR +/- FROM INSTRUMENT SPEC SHEET POTENTIAL ERROR SOURCES IF KNOWN DISSEMINATION ERROR (FOR CALIBRATION LABORATORIES) Uncertainty Distribution Divisor Degrees of Freedom Standard Uncertainty Squared % Contribution u^4/df

0.001% 0.000% 0.004% 0.002% 0.003% 0.000%

normal 1 100 normal 1 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 rectangular 1.732 100 Combined Uncertainty Effective Degrees of Freedom Coverage Factor (k) = Expanded Uncertainty

10.00E-6 000.00E+0 23.09E-6 8.66E-6 17.32E-6 000.00E+0 210.74E-6 33 2.03 428.75E-6

100.00E-12 000.00E+0 533.36E-12 75.00E-12 300.02E-12 000.00E+0 44.41E-9

0.23% 0.00% 1.20% 0.17% 0.68% 0.00% 100.00%

100.00E-24 000.00E+0 2.84E-21 56.26E-24 900.11E-24 000.00E+0 58.87E-18

FOR K= 1 UC= SQUARE ROOT OF TOTAL COMBINED TYPE A AND B FOR K=2 U = K * UC (UNCERTAINTY % ) 0.02107%

0.042% 0.421 LBF