1) Co-current flow

FT1(L/min)
10
10
10
10
10

FT2(L/min)
2
4
6
8
10

TT1( )
41.2
37.1
34.5
33.6
33.5

TT2( )
28.1
28.4
28.7
29.0
29.3

Run 1
CW

Nominal
flow

TT3( )
47.5
46.3
45.4
45.0
44.7

Run 2
HW

10

TT4( )
49.4
49.2
49.5
48.9
48.7

Run 3

DPT1( )
54.0
53.0
117.0
178.0
265.0

Run 4

DPT2( )
118.0
127.0
121.0
120.0
121.0

Run 5

CW

HW

CW

HW

CW

HW

CW

HW

10

10

10

10

10

(L/min)
2

10

10

10

10

10

10

Temp Inlet
(C)

T2: 28.1

T4: 49.4

28.4

49.4

28.7

49.5

29.0

48.9

29.3

48.7

Temp
Outlet (C)

T3: 47.5

T1: 41.2

46.3

37.1

45.4

34.5

45.0

33.6

44.7

33.5

Pressure
Drop
(kPa)

DP :
54.0

DP :
118.0

53.0

127.0

117.0

121.0

178.0

120.0

265.0

121.0

(Shell)

(Tube)

T3 T2

T4 T1

= 19.4

=8.2

Flow rate
(L/min)

Temp
Change,
(C)

(Shell)

=17.9

(Shell)

=12.1

=16.7

(Shell)

=15.0

=16.0

(Shell)

=15.3

=15.4

=15.2

Average
Temp
(C)

=37.4
=37.8

Q, Head
Load,
BTU/HR

Compute
ratio
LMTD
REYNOLDS
U

QC:

=43.2

=37.1

=42.0

=37.0

=41.3

=45.3
QH:

=37.0

=41.1

1) Constant FT2

FT1(L/min)
2
4
6
8
10

FT2(L/min)
10
10
10
10
10

TT1( )
30.7
31.9
32.6
33.4
33.8

TT2( )
29.5
29.6
29.7
29.8
29.9

Run 1

CW

TT3( )
39.2
43.5
44.1
44.5
45.0

Run 2

HW

CW

TT4( )
49.5
49.4
48.8
48.8
49.2

Run 3

HW

CW

DPT1( )
207.0
208.0
206.0
209.0
207.0

Run 4

HW

CW

DPT2( )
-5.0
-5.0
-5.0
61.0
125.0

Run 5

HW

CW

HW

Nominal
Flow,
USGPM

FC: 10

FH: 2

10

10

10

10

10

Actual Flow,
USGPM

FC: 10

FH: 2

10

10

10

10

10

Temp, C/F,
Inlet

T2: 29.5

T4: 49.5 29.6

49.4

29.7

48.8

29.8

48.8

29.9

49.2

T3: 39.2

T1: 30.7 43.5

31.9

44.1

32.6

44.5

33.4

44.0

33.8

Pressure,
psig, Inlet

PG-C:
______

PG-H:
______

Pressure
Drop, mm
H2O

DP :
207.0

DP : 5.0

208

(Shell)

(Tube)

(Shell)

Temp, C/F
Outlet

-5.0

206.0
(Shell)

-5.0

209.0
(Shell)

61.0

207.0
(Shell)

125.0

CALCULATE
FOLLOWING:
Temp
Change,
C/F

T3 T2

T4 T1

= 9.7

=18.8

=13.9

=17.5

=14.4

=16.2

=14.7

=15.4

=15.1

=15.4

=36.5

=40.7

=36.9

=40.7

=37.2

=41.1

=37.5

=41.5

Average
Temp, C/F
=34.3

Q, Head
Load,
BTU/HR

Compute
ratio
0.5 (QC +
QH), BTU/HR
LMTD
REYNOLDS
U

QC:

=40.1

QH:

CALCULATION

Temperature change
i.

Temperature change of cold water (CW);


T = T T

ii.

= 19.4 C

Temperature change of hot water (HW):


T = T T

= 8.2 C

Average Temperature
i.

Average temperature of cold water;


3+2
2

ii.

= 37.8C

Average temperature of hot water;


4+1
2

= 45.3 C

Amount of heat

QC =

ii.

QH = 25 USG x 8.19 lbm x 1.0 BTU x 9.9 x 1.8F x 60 min = 218,919 BTU

i.

min

USG

x 1.0 BTU x 19.4 x 1.8F x 60 min = 256,474 BTU

lbmF

1 hr

HR

QC = 10.1 USG x 8.27 lbm x 1.0 BTU x 12.8 x 1.8F x 60 min = 115,468 BTU
min

USG

bmF

1 hr

HR