Introduction
12
absorbed
dose points
>> 2
absorbed
dose points
DQA x 2 ?
< 8 Gy
Material
ArcCHECK (detector)
21 cm
L 21 cm
No warm-up
No dose or acquisition limit
Real time updates (50 ms)
ArcCHECK (software)
BEV display
Method
ArcCHECK calibration
Clinical cases
Analyze
measured
abs(2%, 2mm)
92.0%
Hot points
Cold points
Compare
planned
Breast
(2 Gy x 28)
Results
> 92%
> 92%
Results
> 97%
Results: Comparison
Eye Orbit
(2 Gy x 33)
Absorbed dose
sampling
ArcCHECK
rel(3, 3)
99.3%
A1SL
-3.1%
-3.4%
EBT2
rel(3, 3)
95.8%
Results: Comparison
+0.4%
-1.9%
2%
>
Conclusions
ArcCHECK vs TPS
abs (3% 3mm) > 92%
rel (3% 3mm) > 97%
Small volumes: abs (3% 3mm)>99%
2%
Conclusions
ArcCHECK is suitable to measure IMRT TomoTherapy