WV.
A
TROUBLESHOOTING
General
The "Troubleshooting" section of this manual is written to
facilitate maintenance using simple hand tools and a multimeter,
such as a Simpson Model 260 or Triplet Model 360. Complete
adjustment. requires the use of an oscilloscope. The equipment does
come from the factory with all adjustments preset and sealed. Most
maintenance problems require only the use of a multimeter, A few
simple multimeter checks and use of operator panel meters or tell-
tale lamps will generally locate the problem. Many times a mere
visual inspection will locate a source of trouble, but an ohnmeter
check should thereafter be made to verify all parts are good before
reapplying power.
ADVICE Do not break the seal on pots of the control board and
adjust in troubleshooting. The control can be misaligned such that
further damage to the system will result and an oscilloscope setup
be required. Avoid the temptation to turn pots in diagnosing a
problen. AS previously stated, the majority of problems can be
diagnosed by use of a multimeter, panel meters, and indicator
amps.
How to Use an Ohmneter
A simple ohnmeter is a very useful instrument that may be used for
the location of circuits. It allows one to make checks with no
power on the equipment, thus is the safest instrument for
inexperienced personnel. Usually, VOM (volt-chm meter) is used for
this purpose as it requires no external power to operate the
instrument, making it portable. A VOM of at least 20,000 ohms per
volt should be used. A meter like a Simpson 260 VOM is recommended.
Other inexpensive meters may be found with the mentioned ohms per
volt. Some meter manufacturers may design their meters to read
polarity in the reverse direction and must be recognized for this
fact when making polarity measurements; for example, diodes read in
one direction only when good.
Semiconductors are to be checked on the Rx 1 scale of an ohmmeter.
Usually a selector switch is provided for selecting this range.
With one end of the leads inserted into the meter (see meter
instructions) and the.other ends open, the meter will see an
infinite resistance and remain at infinity. (Infinity symbol = 00)
By touching the leads together, a short or zero ohms is seen by the
meter. A control knob is provided on the meter for calibrating the
meter to zero ohms. This must be adjustéd before using the meter
and checked for zero if moved to a different scale. If the meter
cannot be zeroed, it is generally a sign that the battery is weak
and needs replacing. See the following "Ohnueter Tests for
Semiconductors" and the troubleshooting tests section for other
tests that may be performed with an ohmmeterAs mentioned earlier, the ohmmeter may be used to check continuity
The leads may be placed across a conductor and measure the
resistance between the leads. A copper wire or bus will have a very
low resistance and appear to be a short when looking across it. A
resistance in the line between the leads will cause the meter to
read the value if properly calibrated.
Ohaweter Tests for Semiconductors (Out of Circuit
Set the multimeter to R x 1 scate. The meter has an internal
battery which supplies power when used to measure resistance
Determine which lead of the meter is connected to the positive end
of this battery. This can be done by checking a diode which is
known to be good.
Semiconductors almost always fail short. A meter reading of
0 ohms indicates a failed device
‘i 1. Place plus lead on anode and
vier minus lead on cathode. Read
ANODE >. CATHODE approximately 5 to 10 ohms
sj CATHODE
: 2. Place plus lead on cathode
and minus lead on anode. Read
more than 10,000 ohms.
Mote: This test is good for both small signal diodes as well as
large power diodes. Power diodes are made in both forward polarity
and reverse polarity. The stud is the cathode on forward polarity
diodes. The stud is the anode on reverse polarity diodes
1. Place plus lead on anode and
minus lead on cathode. Read
more than 10,000 ohms.
ANODE PS CATHOPE 2, Place plus lead on cathode and
“GATE. minus lead on anode. Read more
than 10,000 ohms.
3. Place plus lead on gate and minus
lead on cathode. Read approx-
imately 15 to 30 ohms.
THYRISTOR (SCR)Ohmmeter test for semiconductors (In Circuit)
1. Remove al) power from cabinet.
2. Set ohmmeter on R x 1 scale
3. Check module diagram for device location.
4. Take ohmmeter readings across each device and check
per Table I.
5. If O ohms is read across a device, isolate device and
check again
6. Remove failed device
7. Install new device using recommended procedure
See Procedure for Removal and Replacement of Power
Semiconductors
8. Before discarding failed device, always ohmmeter check it one
more time
TABLE 1
OHNMETER READINGS
DEVICE
FORWARD, REVERSE
RECTIFIER DIODES 5 to 10 OHMS MORE THAN
D1,02,03,04, 05,06 10,000 OHMS
BLOCK DIODES 5 TO 10 OHMS MORE THAN
801, B02 10,000 OHMS
CLAMP DIODES 5 TO 10 OHMS MORE THAN
01, coz 10,000 OHMS
DISCHARGE O100ES 5 TO 10 OMS MORE THAN
07,08 - 10,000 OHMS
THYRISITORS 15 TO 30 OHMS MORE THAN
SCRI sSCR2 10,000 OHMS
SCR3,SCREBasic Power Circuit Check with Ohmmeter
These tests should be performed before applying power upon unit
installation or after a failure has occurred. If fuses are found to be
bad, a power semiconductor check should be made. Making these checks
will be time well spent, rather than gambling that everything is proper
in applying power to the system.
Caution: Ohnmeter checks are to be performed with the power
supply de-energized. One should insure that no charge remains on
capacitor studs or bus bars before handling.
1, Fuses - Check the power fuses with the miltimeter conditioned for
an ohmmeter test on the R x 1 scale. These are located at the
‘output of the disconnect switch for incoming three phase Tine
voltage. A good fuse will read zero ohms. A defective fuse will
read more than 1000 ohms. Because of sneak paths through other
components in the system, a reading other than zero ohms should be
interpreted as defective.
The various control fuses will usually be good when checked after
most fault conditions. If necessary, they can easily be checked by
removing them fron their respective Sockets. If removed, insure
that they are returned properly. Contro] fuses will usually relate
to obvious problens, such as no control lights, no voltmeter
reading (potential transformer fuses), or Tack of operator control
with the power dial
2. Power Semiconductors - The power devices (diodes and SCR’s) can be
checked without removing them frofl their heat sinks. This is
advised in all cases where possible. fach should be checked in
the forward and reverse direction as indicated in Table 1.
Parallel circuit paths will cause the meter to read an ohm value in
one direction on good devices in some instances. A zero ohm
reading (short) across a device in both directions is an improper
reading. Power devices of this nature usually fail short circuit.
Caution: Where devices are in parallel, it may be another part in
the system that is bad, which will reflect a short on the device
being measured. If a short is observed, the branch circuits should
be isolated until a bad device is located. Another Word of
Caution: When a semiconductor is removed from the circuit, it may
appear open (high reading in both directions) which is caused by
no pressure .applied to the pole faces of the device.
3. Capacitors - If a capacitor develops a short, it will usually read
near or at zero ohms on the ohmmeter check. - Again, sneak paths of
parallel circuit should be watched for when making checks of this
nature, If switching leads or polarity 6f the multimeter indicates
a different reading on the multimeter (one high - one low), the
capacitor is most likely a good capacitor will slowly charge through
the batteries of the mltimeter. If a short is observed on themultimeter and the capacitors are in parallel, it will be necessary
to isolate the individual section for the short. This should be
verified before removing the part. A swollen capacitor in a group
would be an obvious candidate for failure.
F. General Troubleshooting Outline
1. Visual Cheek
Check for bulged, or leaky capacitors
Check buswork for any loose connections
Examine KS contactor and RI resistor
Check for any loose or burnt wiring
Check for water leaks or condensation problems
Check a1] water-cooled resistor assemblies
2. Ohmmeter Check
Line diodes
Clamp and block diodes
SCR’s (Thyristors)
CP capacitors
___ OF capacitors
3. DC Bus Voltage Balance (Breaker on, Unit off)
Check positive bus to neutral bus (+ 325 VDC) for 5OKW and
up. +175 for 10,20,and 30KW.
Check negative bus to neutral bus (- 325 VOC) for SOKW and
up. -175 for 10,20, and 30KW.
OC bus voltage must be balanced within 20 volts.
If not, check the following:
Any accumulation of dirt or condensation
that may cause leakage.
Line voltage. Check phase to phase and
phase to ground. One phase may be
grounded. (480 .VAC nominal)
CF capacitor bleeder resistors, R2 and R3Isolate one CF capacitor at a time on low
side, (Must remove one CF capacitor on
high side to maintain balance).
Isolate or replace on SCR at a time on
low side.
Isolate other components on low side one
at a time (i.e., CP, RC snubbers).
Check if any water hose has been replaced.
May be conductive.
V. CORRECTIVE ACTION
AL
Circuit breaker will not energize.
Check that 211 doors are closed.
Check that door switches are operating properly.
Check for 24 VOC at breaker undervoltage coil. If
voltage is low, replace batteries on 24 volt supply
board.
Check circuit breaker linkage. May have to reset
breaker manually.
Power Supply will not start.
lL.
2
Check that al] fault lamps are out.
Check that “power ready" lamp is on. If not, check
that any customer connected interlocks are operating
properly.
If unit has auto/manual switch and the switch is in the
auto position, you mist push the start button to get a
power ready Tight.
Check control board Pin 2. Pin 2 must go to 420 volts
to start. If Pin 3 goes to +20 and unit does not start,
replace or troubleshoot control board (single control
board units). Check oscillator board pin 3 for +20 volts
to start.
Fuse Failure.
1
2.
3.
Perform visual check of unit
Ohmmeter check unit for failed component.
Replace failed component. Be careful to observe
proper installation procedureD. Fuse Failure with No Bad Parts
If failure of fuses occurs with no bad parts found, a number
of possibilities arise. On a rare occasion, a capacitor will
are or break down internally while operating and cause fuses to
fail, then appear good when making tests. This type of failure
is difficult to locate, but usually it will not occur too many
times before total failure of the capacitor occurs. A visual
inspection for a swelled capacitor or listening for the arc
while running will sometimes aid in locating the source of
failure.
A visual inspection may locate where a breakdown has occurred
external to components. Such things as coolant leaks, loose
connections, foreign material (i.e., dropped pieces of wire,
loose hardware, metal filings, discolored or bent bus, tools
left in enclosure, etc.) may be the result of intermittent fuse
failures
The cooling system may relate to fuse failure if it is improper.
A minimum of 30 PSI differential pressure is to be maintained
across’ the inlet and outlet water manifolds to insure adequate
flow in all lines. The inlet water temperature cannot be above
a maximum rating of 95°F. Switches in the system monitor the
inlet pressure and temperature for safe operation. Caution
Tampering with these devices can only lead to trouble and will
void the warranty. Outlet temperature switches are utilized to
insure safe operating temperatures of individual lines. Any
improper performance of water Tines should be checked out and
corrected. The cooling system is important for protection of
the water-cooled components. Kinked hoses or restrictions from
deposits will lead to eventual failure. Targets as described
earlier in the maintenance section are used for electrolysis
protection and should be watched until the rate of decay is
determined.
A more elusive and complex problem may result from radio
frequency "noise." This can result from improper or poor wiring
practices where radiated or conducted electrical transients are
fed into the control and disturb its normal operation. This can
usually be completely eliminated or avoided if a few simple
rules are followed. If none of the previous usual problems have
been found, then it is advised that the following questions be
answered.
‘Are control wires and power cables separated
between the load station and power supply?
ii Are wires twisted or shielded as indicated on
drawings, particularly where Tong runs are used?
Verify that no cut or frayed wires exist.iii Oo all relays associated with control or
connected to the power supply 115 VAC line have
snubber protection around their solenoid coils?
iv If the equipment has been running up to now, has
any new circuitry been added to the control about
he time trouble started?
v Is there proper clearance and insulation at the work
station for bus bars and coil? If very severe arcing has
occurred, disturbances may be getting back to the power
supply even though there is isolation between the power
supply and load. The power supply is capable of operating
into shorts or open circuits at the work station
Protective circuitry in the power supply will stop the unit
if a continuous short exists at the Toad. An open will
usually result in the power supply operating in voltage
Vimit
Intermittent Iripping
Intermittent tripping can be caused by a part failing under oad, noise
problems, a control board problem (false tripping), or a malfunciion in
the load circuit
1
Part failing under load
The most common part to fail under load would be an SCR. Next would
be a capacitor. The only way to find a part that is failing under
load is through a process of elimination.
A
Monitor OC bus voltage with respect to neutral while running
unit. If a part is failing under load, either the positive or
negative bus voltage will collapse to 0 volts and then recover
to normal levels after the unit has tripped
Turn unit on. Turn power up slowly while watching voltmeter.
If positive bus voltage collapses to zero, then part failure is
in the positive section. If negative bus voltage collapses Lo
zero, then part failure is in the negative section.
If bus voltage collapse is only partial (100-150 VDC), then it
is probably a control board problem (misfiring). Change or
troubleshoot control board. If bus voltage stays constant at
point of trip, then problem is either load related (i.e., tank
capacitor breaking down, severe arching at work coil) or a
control board problem (false tripping)0. If bus voltage collapses to zero then:
1. Replace SCR's one at a time on side that
collapses
2. Swap CP capacitors. The idea is then to try
to get the bus collapse to move to the opposite
side.
3. Swap other inverter components untiT bus collapse
moves to the opposite side
>
CAUTION: When using the oscilloscope to monitor waveshapes in the
power section, high voltages will exist on the chassis if the ground
Vead of the probe is connected into the power circuit. Care must be
used in handling the oscilloscope. A good insulating material
should be placed between the operator of the oscilloscope and earth
or frame ground.
The oscilloscope must always be isolated from the line! This may be
accomplished by using an isolation transformer between the 115 VAC
Vine and oscilloscope or by not connecting the ground of the three
prong plug. Failure to operate the oscilloscope in this manner will
result in damage to the instrument or probe.
Before looking at any waveshapes of the equipment, it is always
necessary to compensate the probe. Without doing this first, some
distortion of the observed wave will exist. The oscilloscope manual
will explain how to perform this function, but generally a square
wave should be observed at the calibrator terminal of the
oscilloscope.
The x 10 probe would be the most, convenient for monitoring
waveshapes. A x 1 probe should be avoided. A x 100 probe will be
useful for some of the waveshapes in the power circuit, however, the
x 10 probe will suffice for most measurements.
As mentioned previously, the scope chassis may have high potentials
on it when monitoring the power circuit with the ground lead
attached to the high frequency bus. This may be overcome by looking
at a signal using both channels of a dual channel -oscilloscope to
monitor a waveshape. 8y setting the gain for each channel equal
{both probes need to be the same "X" value), ihverting one channel,
establishing a common reference for each channel, and monitoring the
signal with one probe acting as a ground, a waveshape may be
observed using the scope amplifiers for isolation_WrvE FORMS
OsciLtLoscor
MONITOR AY
FIRING CONTROL BOARD
PIN IB REE: PIN IT
POSITIVE WAVEFORM SHOW”
NEGATIVE WAVEFORM
INNERTED
orm
SCR Wav
PEAK REVERSE
VOLTAGE -S™
+600
PIN IS REED PIN IT
Wes aye BaaS
G. (teas
fES> 08 DI-daS
tous
| ON
TIME
280 (
NOTE
1. ON-TIME - 20 To 25 ~ SEC. NOMINAL
2. TURM-OFF-TIME — 10 USEC. MINIMUM
3B. PEAK REVERSE VOLTAGE - S00 VouTS Maximum
CAUTION! OSCILLOSCOPE MuST BE UNGROUNDED