the DiCOM box

INSTALLATION MANUAL
Manual Part No. 005-00148-00

Date of manufacture:_______________

Note: This manual includes instructions for

version software

NAI Tech Products PN 105-00504-00

This manual presents installation instructions for the NAI Tech Products Acquisition Module. Read it in its
entirety prior to operation to ensure that the NAI Tech Products Acquisition Module is properly set up and
keep it available for future reference.

This document is provided as a reference only regarding the NAI Tech Products Acquisition Module. All
questions must be referred to NAI Tech Products
Uncontrolled copies are for reference only, and users are responsible to obtain the latest information.

© 2007 NAI Tech Products


the Dicom box is a registered trademark of
NAI Tech Products

NAI Tech Products MediTech Strategic Consultants B.V.

M 12919 Earhart Ave.


Auburn, Ca. 95602
Maastrichterlaan 127
NL - 6291 EN Vaals
USA

tel: (866) DICOMBX (342-6629) (toll free)


(530) 887-1008 (direct)
fax: (530) 887-1108
web: www.NAITechProducts.com

005-00148-00
Rev. D
PROPRIETARY INFORMATION
AND SOFTWARE LICENSE
The Customer shall keep confidential all proprietary information furnished or disclosed to the Customer by NAI Tech
Products, unless such information has become part of the public domain through no fault of the Customer. The Customer
shall not use such proprietary information for any purpose other than the maintenance, repair or operation of the goods,
without the prior written consent of NAI Tech Products.
NAI Tech Products systems contain NAI Tech Products and others proprietary software in machine-readable form. NAI
Tech Products and others retain all their rights, title and interest in the software; except that purchase of this product
includes a license to use the machine-readable software contained in it. The Customer shall not copy, trace, disassemble
or modify the software, nor cause or permit the copying, tracing, disassembly or modification of such software. Transfer of
this product by the Customer shall constitute a transfer of this license which shall not be otherwise transferable. Upon
cancellation or termination of this contract or return of the goods for reasons other than repair or modification, the Customer
shall return to NAI Tech Products all such proprietary information.

i 005-00148-00
Rev. D
5 IMPORTANT SAFETY INFORMATION
• REFER SERVICING TO QUALIFIED SERVICE PERSONNEL. THIS INCLUDES FUSE AND BATTERY
REPLACEMENT. IMPROPER FUSE REPLACEMENT CAN CAUSE SHOCK AND FIRE HAZARD.
DANGEROUS VOLTAGE
Shock hazards may exist if this system is not properly grounded. Protection against electrical shock is
provided by grounding the chassis with a 3-wire cable and plug. The DiCOM box must be plugged into a
hospital-grade three-contact outlet. The grounding wire must not be removed or defeated.
• Risk current of this device does not exceed 200 micro amps.
• CAUTION: Do not remove the protective covers on the DiCOM box; hazardous voltages are present inside.
Cabinet covers must be in place while the unit is in use. All internal adjustments and replacements must be
made by a qualified electronic technician.
WARNING:
• Do not operate the DiCOM box in the presence of flammable gasses or anesthetics. Explosion can result.
• The appropriate IEC cord set must be used to ensure that protective earthing (grounding) is secure prior to
application of power.

WARNING:
• Always disconnect the DiCOM box from the power source prior to cleaning the unit.
• Accessory equipment connected to the video and digital interfaces must be certified according to the
respective IEC standards (i.e. IEC 950 for data processing equipment and IEC 601 for medical equipment). All
configurations shall comply with the system standards IEC 601-1-1. Additional equipment connected to the
signal input or output shall meet the system standard IEC 601-1-1. If in doubt, consult the technical service
department or your local representative.
• The operator of this equipment must never physically contact the patient and the DiCOM box input and output
connectors.
• There are no user serviceable components inside the instrument. Service, other than cleaning the fan filter, will
be performed by returning the instrument to NAI Tech Products, 12919 Earhart Ave, Auburn, CA 95602.

5 EMC INTERFERENCE
• This equipment may be affected by electro-magnetic interference. Should interference occur determine which
equipment is causing the interference; this can be done by turning the suspected equipment off. Once the
source of EMC interference has been found try correcting the interference by one or more of the following
measures:
• Increase the separation between the equipment and that source of interference
• Connect the equipment into an outlet on a circuit different from that of the interference source
• Consult the service department or qualified technician.

STATIC SENSITIVE DEVICE


• This product meets the requirements of IEC 601-1-2 for electro-static discharge. Care should be taken when
operating this product in very low humidity environments. Common ESD precautions apply. Some of these
precautions are:
1. Avoid carpeted areas.
2. Discharge the body by touching the DiCOM box chassis prior to handling the hand controller.
If ESD problems persist, use of wrist straps, ionizing blowers and conductive floor wax may be indicated.

5 DISPOSAL
Disposal of this product shall be in accordance with all applicable laws and regulations in effect at the locality,
at the time of disposal.

005-00148-00 ii
Rev. D
SYSTEM SAFETY SYMBOLOGY
The following table defines the symbols used in the product labeling
Title/Symbol English / Identifies
Note An important feature or special information. A note may apply only in special cases.

Important 5 An important note in procedures that provides information necessary to complete the task.

A helpful tip to apply a technique to your needs.

CAUTION Critical information that you must follow to prevent data loss or damage.
General warning

Attention

M Manufactured by

N Date of Manufacture

f Serial Number

Authorized Representative

C CE Mark

i Consult Instructions for Use

Static sensitive

Do not touch

Dangerous voltage

Represents AC power is ON.

O Represents AC power is OFF.

~ Alternating current (A.C.)

Protective earth (ground)

Equipotentiality. Indicates the connection point of the potential equalization conductor


should it be required by application of the installation.

Type 1 Logo - Pollution Control Logo for Electronic Information Products

iii 005-00148-00
Rev. D
Table 1:
Marking Styles for Names and Contents of Toxic or Hazardous Substances or Elements

Part Name Toxic or Hazardous Substances and Elements

Hexavalent Polybrominated
Lead Mercury Cadmium Polybrominated
Chromium diphenyl ethers
(Pb) (Hg) (Cd) biphenyls (PBB)
(Cr (VI) (PBDE)

Remote O O O O O O
Control

DVI O O O O O O

CA+ O O O O O O

CA O O O O O O

MDR O O O O O O

MDR Video O O O O O O

Legend:
O: Indicates that this toxic or hazardous substance contained in all of the homogeneous materials for this part is below the
limit requirement in SJ/T11363-2006.
X: Indicates that this toxic or hazardous substance contained in at least one of the homogeneous materials for this part is
above the limit requirement in SJ/T11363-2006.
General Upkeep

The outside of the DiCOM box module and control panel may be cleaned occasionally as required. First, turn off
power to the DiCOM box. Gently wipe the outside of the unit with a soft, lint-free cloth. The cloth may be moistened
with a mild soap solution. Be certain to avoid solvent type cleaners, as these chemicals can cause deterioration
which may cause a safety hazard.
Clean and vacuum the air intake openings in the front of the module. Remove the fan filter by prying out it’s retaining clip,
clean it thoroughly once a month or on an as need interval. (See Figure 3 for the location of the fan filter.) Failure to keep
the fan filter clean may result in system lock-ups and/or failure.
SERVICE DIRECTIONS
Other than the fan filter, the DiCOM box contains no user serviceable components. The unit must only be serviced by NAI
Tech Products personnel. For service, call (530) 887-1008. Refer to Installation section 1.0 for complete return information.
MANUAL SUPPLEMENTS
Manual Supplements, if present, will provide updated operator information. Supplements should be added to the manual as
the DICOM box software is updated.

005-00148-00 iv
Rev. D
Table Of Contents

1 Unpack DiCOM box From Carton . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1


2 Installation Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
2.1 Preparation for Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
2.2 Equipment Required for Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
3 System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
3.1 Host Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
3.2 Network Store/Print Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
3.3 Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
3.4 Optional Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
3.5 Optional Flat Panel Touchscreen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
4 SPECIFICATIONS: the DiCOM box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
5 SPECIFICATIONS: the DiCOM box CA+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
6 Installation Flowcharts and Video Input Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12
7 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-26
7.1 Verify DiCOM box Software is Current Revision. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
7.1.1 Installation Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-14
7.2 Select/Setup Host . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-23
7.2.1 Calibrate Phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
7.2.2 Calibrate ADC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
7.2.3 Modify Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15-17
7.2.4 Find Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-19
7.2.5 ACR Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19-23
7.3 DiCOM box Option Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.4 Setup Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23-24
7.4.1 Modify DiCOM box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.4.2 Ping Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.4.3 Add Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.4.4 Modify Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.4.5 Delete Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.4.6 Ping Worklist Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.4.7 Add Worklist Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.4.8 Modify Worklist Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.4.9 Delete Worklist Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.4.10 Ping MPPS Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
7.4.11 Add MPPS Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.4.12 Modify MPPS Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.4.13 Delete MPPS Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.4.14 Ping Commit Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.4.15 Add Commit Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.4.16 Modify Commit Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.4.17 Delete Commit Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.4.18 Ping Archive Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.4.19 Add Archive Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.4.20 Modify Archive Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.4.21 Delete Archive Server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.4.22 Archive Storage Class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.5 Exam ID Config . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
7.6 Exam Transfer Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.7 Product Name Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.8 Institution Name Entry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.9 Station Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.10 Host Control Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.11 Frame Grabber Configuration (CA+ Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.12 Temperature Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.13 Clean Fan Filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
7.14 Load Backup Files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

v 005-00148-00
Rev. D
7.15 Saving Backup Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7.16 Save Log File to a Floppy Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
7.17 Final Installation Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
APPENDIX A RS-422 Host Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27-28
APPENDIX B REMOTE CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28-31
B.1 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
B.1.1 Control Keypad Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
B.1.2 Ethernet Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
B.1.3 Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
B.2 Environmental . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
B.3 Remote Control Installation Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29-31
APPENDIX C Flat Panel Touch Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32-34
C.1 Operator Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32-33
C.2 Adjusting the Flat Panel Touch Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
C.3 Connecting the Flat Panel Touch Screen to the DiCOM box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

005-00148-00
vi
Rev. D
1 Unpack DiCOM box From Carton
A Ensure shipping carton is not damaged. Unpack the carton and inspect contents and verify all of the following are
present:
• Acquisition Module (DiCOM box)
• DiCOM box Control Panel
• Input Video Cabling (DiCOM box analog input only)
• Power Cord (hospital grade)
• Operator Manual and CD containing Installation Manual
B Connect the control panel cable to the connector on the DiCOM box rear panel marked “Control Panel”, after
removing the connector cover plate.
C Connect the power cord to the DiCOM box rear panel and to the wall plug power receptacle.
D Remove floppy drive access cover (2 screws). (For system backup after installation) See Figure 1.
Figure 1 Floppy Disk Access Cover Diagram

2 Installation Information
Other than the fan filter, the DiCOM box contains no user serviceable components. The unit must only be serviced by NAI
personnel. To return a unit to NAI, call NAI Customer Service at (530) 887-1008 and request an RMA (Return Material
Authorization) number. Units returned to NAI without an RMA number will not be accepted.
2.1 Preparation for Installation
Prior to leaving for installation, it is necessary to acquire the following information regarding the host system and network
configuration of the installation site.
• Location of DiCOM box - Relative to system, network connection. Ensure cable lengths and types are
adequate.
• Power - Availability and access
- Type
- Connector
• Host System - Manufacturer
- Model Number
- Output video and connector types.
• Network Characteristics - Ethernet type, connector type and physical access.
- Network connection cable requirements.

1 005-00148-00
Rev. D
- Ensure network is functioning and meets safety requirements.
- Printer identification.
- DiCOM box identification.
2.2 Equipment Required for Installation
• This Manual (supplied with DiCOM box)
• Current revision of system software and any associated operator/installation manual supplements.
• Formatted 3½ inch floppy disks
• Hand Tools
• Electrostatic Control Grounding Equipment
• PC or Laptop with NIC
3 System Description
The Acquisition Module (DiCOM box) is an image capture, and transfer system intended for use with medical imaging
equipment. The DiCOM box base unit will capture single image frames of color (RGB) or monochrome (B&W) video.
Optionally, the DiCOM box may be upgraded, via software key, to any combination of the following configurations:
• DICOM Print Class Enabled (allows transfer of images to a network printer)
• DICOM Storage Class Enabled (allows transfer of images to a network archive or workstation)
• DICOM Worklist Retrieval Enabled (allows Worklist Retrieval)
• Host Control Enabled (standard on digital image input versions)
• DICOM Modality Performed Procedure Step, (MPPS), may be used to notify the Hospital Information System when
a procedure for a specific patient has started and when it has completed.
• DICOM Storage Commitment may be used to notify an archive device that a study has been completed and to
request that the archive device will notify the DiCOM box that all of the images have been successfully stored.
The DiCOM box CA+ will capture single frame or multiframe images of color (RGB) or monochrome (B&W) video.
3.1 Host Interface
The DiCOM box will capture images from the host via an analog framegrabber. The analog framegrabber will capture
single frames of a color or monochrome video signal. Sheet formatting and image capture control will be done via the
DiCOM box control panel.
3.2 Network Store/Print Interface
By connecting to the Ethernet/DICOM interface, the DiCOM box will function as a client device in a networked environment
conforming to the Digital Imaging and Communications in Medicine (DICOM) standard. The DiCOM box has Ethernet
network interface capability that allows connection to the networked environment via an IEEE 802.3, 10/100 BASE-T
network interface. The DiCOM box is capable of sending data to one or more server devices in the networked environment.
3.3 Control Panel
The control panel keypad is composed of twenty keys. The keypad display is a 4-line by 20-character liquid crystal display
(LCD) with back lighting. The control panel is connected to the DiCOM box via a 15-foot cord to allow for remote use. The
control panel provides the user with prompting, status messages, error messages, and other operational information. Note
that all of the screens presented in this manual are examples. The actual screens displayed on the control panel may vary.
3.4 Optional Keyboard
A keyboard may be attached to the DiCOM box to allow for alphanumeric data entry. The control panel keypad is still
required to view the entered data. The keyboard is connected to the DiCOM box via a 6-foot cord to allow for remote use.
3.5 Optional Flat Panel Touchscreen
Replaces the control panel with a flat panel touchscreen display. The display allows viewing patient worklists, returned from
the hospital server, four patients at a time. Also allows for alphanumeric data entry without a separate keyboard.

005-00148-00 2
Rev. D
4 SPECIFICATIONS: the DiCOM box
Video Input (CA only) - Red, Green, Blue (Terminated, 75 Ohms).
- Composite monochrome input (Terminated, 75 Ohms).
- Composite monochrome into Green input (Terminated, 75 Ohms).
- Composite sync (Termination selectable for 75 Ohms/1K Ohms).
- Sync on Green (Terminated, 75 Ohms).
- Separate horizontal sync (Termination selectable for 75 Ohms/1K Ohms).
- Separate vertical sync (Termination set to 1K Ohms).
Video Format (CA only) - Nominal 525 lines 30 frames per second, interlaced or non-interlaced.
- Nominal 625 lines 25 frames per second, interlaced or non-interlaced.
Pixel Clock Min: 7 MHz.
Pixel Clock Max: 166 MHz
Image Width Max: 2048
Image Height Max: 1024
Digital Video Input (DVI only) - Medical Digital Image Interface Emulation. (terminated, 120 ohms)
Digital Image Format (DVI only) - Up to 4K x 5K
Interface Clock Speed (DVI only) - Up to 2 MHz
Network Output - Ethernet 10/100 BASE-T.
Connectors - Digital Video Input: DB37, Female pins, (DVI only).
- Video Input: High Density DB15, Female pins (CA only).
- Video Loop Through: High Density DB15, Female pins (CA only).
- Network: 8 pin modular jack (RJ45).
- Control Panel Keypad: DB9, Male pins.
- Host Control Input: DB9, Male pins.
Power Input - 115-230 Volts, +15% 47-63 Hz, 120 VA.
Dimensions - Module; Height: 3.25 inches, (8.25 cm).
Width: 10.125 inches, (25.72 cm).
Depth: 14.125 inches, (35.88 cm).
- Remote Control Panel; (not including cable);
Height: 7.15 inches (18.16 cm).
Width: 4.10 inches (10.41 cm).
Depth: 0.95 inches (2.41 cm).
Weight - Module, 8 pounds, (3.68 kg).
- Remote Control Panel, 8 ounces, (227g).
Environmental - Operating
Temperature: +50ºF to +104ºF, (+10ºC to +40ºC).
Barometric pressure: 700 hPa to 1060 hPa
Relative Humidity: 5% to 80% non condensing.
- Transportation and Storage
Temperature: +14ºF to +122ºF (-10ºC to +50ºC).
Barometric pressure: 500 hPa to 1060 hPa
Relative Humidity: 5% to 85% non condensing.
Clearance Requirements - Approximately 3 inches for rear panel cable connection.
- Do not obstruct ventilation air flow at front and rear of module.
IEC 601 Classification - Class I, type B ordinary equipment, continuous operation.

3 005-00148-00
Rev. D
5 SPECIFICATIONS: the DiCOM box CA+
Video Input - Red, Green, Blue (Terminated, 75 Ohms).
- Composite monochrome into Green input (Terminated, 75 Ohms).
- Composite sync (Terminated, 75 Ohms).
- Sync on Green (Terminated, 75 Ohms).
- Separate horizontal sync (Terminated, 1K Ohms).
- Separate vertical sync (Terminated, 1K Ohms).
Video Format - Nominal 525 lines 30 frames per second, interlaced or non-interlaced.
- Nominal 625 lines 25 frames per second, interlaced or non-interlaced.
Pixel Clock Min: 7 MHz.
Pixel Clock Max: 40 MHz / Cine RGB with Recall
120 MHz / Cine Monochrome with Recall
Pixel Clock Max*: 80 MHz / Cine RGB without Recall
160 MHz / Cine Monochrome without Recall
Image Size Max: 2048 x 1024 Single Frame
800 x 600 Cine RGB
1600 x 1024 Cine Monochrome
*Special cine mode (Contact NAI product support)
Network Output - Ethernet 10/100 BASE-T.
Connectors - Video Input: High Density DB15, Female pins.
- Video Loop Through: High Density DB15, Female pins.
- Network: 8 pin modular jack (RJ45).
- Control Panel Keypad: DB9, Male pins.
- Host Control Input: DB9, Male pins.
Power Input - 115-230 Volts, +15% 47-63 Hz, 120 VA.
Dimensions - Module; Height: 3.25 inches, (8.25 cm).
Width: 10.125 inches, (25.72 cm).
Depth: 14.125 inches, (35.88 cm).
- Remote Control Panel; (not including cable);
Height: 7.15 inches (18.16 cm).
Width: 4.10 inches (10.41 cm).
Depth: 0.95 inches (2.41 cm).
Weight - Module, 8 pounds, (3.68 kg).
- Remote Control Panel, 8 ounces, (227g).
Environmental - Operating
Temperature: +50ºF to +104ºF, (+10ºC to +40ºC).
Barometric pressure: 700 hPa to 1060 hPa
Relative Humidity: 5% to 80% non condensing.
- Transportation and Storage
Temperature: +14ºF to +122ºF (-10ºC to +50ºC).
Barometric pressure: 500 hPa to 1060 hPa
Relative Humidity: 5% to 85% non condensing.
Clearance Requirements - Approximately 3 inches for rear panel cable connection.
- Do not obstruct ventilation air flow at front and rear of module.
IEC 601 Classification - Class I, type B ordinary equipment, continuous operation.

005-00148-00 4
Rev. D
Figure 2 Front Panel

1. Power on indicator
2. Hard disk busy indicator
3. Floppy drive access

5 005-00148-00
Rev. D
Figure 3 Rear Panel (the DiCOM box CA)

Figure 4 Rear Panel (the DiCOM box CA+)

Figure 5 Rear Panel (the DiCOM box DVI)

1. Hand Control Input


2. Power Input Module
3. Remote Capture Input (Color Remote Capture not used on digital systems)
4. Potential Equalization Conductor
5. Ethernet Output
6. Video Input
7. Video Loop Through (Video Out for the DiCOM box CA+) NOTE: Paint and marking per NAI
8. Fan Filter Tech Products Specification
9. Digital Image Input unless other wise specified.
10. Host Control Input
11. Optional Keyboard Input
12. VGA out for optional touch screen display

005-00148-00 6
Rev. D
Figure 6 Typical Connection Diagram (Analog Video Input)

HOST CONTROL
PINOUT RS-232
PIN SIGNAL
2 RECEIVE
3 TRANSMIT
5 GROUND

VIDEO CONNECTION
NOTES:
CA+ FOR TTL LEVEL
SYNC USE THE
WHITE LEAD.
FOR VIDEO LEVEL
SYNC USE THE
GRAY LEAD.
CA FOR TTL LEVEL
SYNC SET C-SYNC
TERM TO NO.
FOR VIDEO LEVEL
SYNC SET C-SYNC
TERM TO YES.

CONNECTOR STYLE QUANTITY


Digital Video Input DB 37 Female Pins 1
Loop Through DB 37 Female Pins 1
Remote Capture BNC 2
Ethernet Output RJ45 8 Pin Modular 1
Hand Control input DB 9 Male Pins 1
Host Control Input DB 9 Male Pins 1
Power Input module IEC 320/C14 1
Keyboard Input PS 2 1

HOST CONTROL
PINOUT RS-422

PIN SIGNAL NOTE: See Appendix A for the procedure


1 TXD- for setting jumpers on the single board
computer for RS-422 host control.
2 TXD+
3 RXD+
4 RXD-

7 005-00148-00
Rev. D
Figure 7 Typical Connection Diagram (Digital Video Input)

HOST CONTROL
PINOUT RS-232
PIN SIGNAL
2 RECEIVE
3 TRANSMIT
5 GROUND

CONNECTOR STYLE QUANTITY


Digital Video Input DB 37 Female Pins 1
Remote Capture BNC 1
Ethernet Output RJ45 8 Pin Modular 1
Hand Control input DB 9 Male Pins 1
Host Control Input DB 9 Male Pins 1
Power Input module IEC 320/C14 1
Keyboard Input PS 2 1

HOST CONTROL PINOUT


RS-422
PIN SIGNAL
1 TXD-
NOTE: See Appendix A for the procedure
2 TXD+
for setting jumpers on the single board
3 RXD+ computer for RS-422 host control.
4 RXD-

005-00148-00 8
Rev. D
6 Installation Flowcharts and Video Input Configurations
The Acquisition Module (DiCOM box) must be configured for a particular installation by setting up interface parameters to
be compatible with the resident scanner and the specific network applications. The following flowcharts present the
recommended installation sequence. Refer to the Operators Manual for information regarding menu item selection.
Installation Flowchart (number 1)

9 005-00148-00
Rev. D
Installation Flowchart (number 2)

005-00148-00 10
Rev. D
Installation Flowchart (number 3)

Framegrabber Configurations

Table 1:

Info Screen SN Cutoff Single Frame Cine Model

DFG NA Yes NA DVI

EFG 8478 Yes NA CA

MFG/8 NA Yes 40/120 MHz Cine + Recall CA+

MFG2 NA 166 Single Frame NA CA

MFG3/8 NA Yes 80/160 MHz Cine - Recall CA+

MFG4/8 NA Yes 121 MHz Cine - Recall 4 field interlace

MFG5 NA 166 Single Frame NA i-Print

Note: MFG framegrabbers have either 384 Mb or 768 Mb memory size. MFG2/4 designates a single frame configuration
with 384 Mb of memory, MFG/8 designates Cine + Recall with 768 Mb of memory.

11 005-00148-00
Rev. D
Installation Flowchart (number 4)

*NOTE: If the Find Parameters/Automatic menu item fails, proceed to Section 7.2 to complete video setup.

005-00148-00 12
Rev. D
7 Installation
Before beginning, ensure that the prior site information acquired is correct regarding the configuration of the system and
the network. For user interface guidelines on using the installation menus, please refer to the Operators Manual.
7.1 Verify DiCOM box Software is Current Revision.
Follow instructions on disk label when loading software.

INSERT AAM-CA INSTALL DISK 1


INTO THE FLOPPY DISK DRIVE
AND APPLY POWER

CANCEL
INSERT DISK 2 PUSH OK SOFTWARE FULL EJECT DISK AND
INSTALL? END
WHEN READY UPGRADE? PRESS ANY KEY

OK

LOADING FILES OK
WAIT UNTIL
BEEP

ERASES OK FORMATTING HARD LOADING FILES


DATA AND DRIVE WAIT UNTIL
INSERT DISK 3 PUSH OK SETUP PLEASE WAIT.... BEEP
WHEN READY
CANCEL
OK
INSERT DISK 2 PUSH OK
INSTALLATION COMPLETE OK LOADING FILES WHEN READY
REMOVE FLOPPY PRESS ENTER THE AAM-CA SERIAL WAIT UNTIL
ANY KEY TO RESTART AAM # PUSH OK WHEN DONE BEEP OK

OK LOADING FILES
WAIT UNTIL
INSERT DISK 3 PUSH OK BEEP
END WHEN READY

7.1.1 Installation Menu


This operation pertains to installation of the DiCOM box: The Installation sub-menu will be activated by pressing the
CANCEL key four times followed by pressing the Color Capture A key. The module must be idle, i.e., ready to begin an
exam, to activate the Installation menu. Once the Installation menu has been activated, press the Menu key to access it.
I N S T A L L A T I O N
 S e l e c t / s e t u p H o s t
O p t i o n C o n f i g
S e t u p N e t w o r k
E x a m I D C o n f i g
E x a m T r a n s f e r M o d e
P r o d u c t N a m e
I n s t r u c t i o n N a m e
S t a t i o n N a m e
H o s t C o n t r o l C o n f i g
F r a m e G r a b b e r C o n f i g
T e m p e r a t u r e L i m i t
C l e a n F a n F i l t e r
L o a d B a c k u p F i l e s
S a v e B a c k u p F i l e s
S a v e L o g F i l e s
Some menu parameters are dependent on the model of the DICOM box. Use the Info key to determine the frame grabber
type of the DICOM box to be installed.

13 005-00148-00
Rev. D
MENU TREE (Figure IN1)

005-00148-00 14
Rev. D
7.2 Select/Setup Host
This operation allows the operator to select the modality and host that the DiCOM box is connected to. First select the
modality and then the specific host that the DiCOM box will be connected to. When a host is selected, the DiCOM box will
automatically be configured with the appropriate video parameters (for analog frame grabber equipped systems). Also,
when selecting a new host, a frame grabber calibration (for analog frame grabber equipped systems) must be performed.
See sections 7.2.1 and 7.2.2 for further instruction on frame grabber calibration. If the connected host does not appear in
the SELECT/SETUP HOST menu, a custom data file will need to be created by a qualified service person. Make certain
that the host system has reached its normal operating temperature prior to starting this procedure. Upon completion of this
procedure go to section 7.2.5 and perform the ACR setup procedure (if ACR is required).
NOTE: Use the method outlined in the installation flowchart (number 4) before proceeding with this procedure.
1. For video input to the DiCOM box select the UNCLASSIFIED listing in the SELECT/SETUP HOST menu.
2. Select GENERIC VIDEO but press CANCEL when prompted to calibrate video.
3. Reselect UNCLASSIFIED and GENERIC VIDEO to gain access to the SETUP HOST menu.
4. Refer to figure IN1 and select the appropriate video and sync source for the host system that is connected to the
DiCOM box.
5. Use the FIND PARAMETERS menu item. If pixel clock frequency is known, use the FIND LINE RATE sub-menu to
determine the line rate. Then calculate Htotal by the following formula.
Htotal = pixel clock frequency/line rate.
6. After the correct Htotal value has been determined use the AUTOMATIC SIZING menu item to detect the appropriate
image height and width, and the left and top margins.
7. Next, use the CALIBRATE PHASE and CALIBRATE ADC menu items to fine tune the pixel clock and ADC values.
8. Send an image to the printer or PACS to verify image quality.
NOTE: Use the MODIFY PARAMETERS menu item to correct any improperly set video parameter.
7.2.1 Calibrate Phase
Use this menu item to calibrate the pixel clock phase to that of the host video. The host system video must contain high
contrast black to white, and white to black transitions.
7.2.2 Calibrate ADC
Use this menu item to calibrate the ADC so that the captured image will contain the full dynamic range of the host video.
The host system video must contain peak white areas and black areas.
7.2.3 Modify Parameters
Use this menu item to adjust image height and width, image margins, video and sync source, and pixel aspect ratio.
7.2.3.1 Advanced Parameters
Video Source
Select the video source that is appropriate for the host system you are connecting to.
• RGB for color host systems with composite or separate sync signals.
• RGB&MONO-BW for host systems with color and monochrome video output with composite or separate sync
signals. (EFG Only)
• GREEN-BW for host systems with green channel video output with composite or separate sync signals.
• MONO-BW for host systems with monochrome video output with composite or separate sync signals.
Sync Source
Select the sync source that is appropriate for the host system you are connecting to.
• COMPOSITE (RGBS) for RGB host systems with a separate composite sync.
• SEPARATE/NEG HSYNC for host systems, color or monochrome, with separate negative horizontal sync.
• SEPARATE/POS HSYNC for host systems, color or monochrome, with separate positive horizontal sync.
• SYNC ON GREEN for color host systems with sync on the green channel.
• COMPOSITE (MONO) for monochrome host systems with composite sync.
• VIDEO LEVEL SYNC for host systems with non-TTL level sync signals. (MFG Only)
• TTL LEVEL SYNC for host systems with TTL sync signals. (MFG Only)
• SEPARATE H/V SYNC for host systems, color or monochrome, with separate sync. (MFG Only)

15 005-00148-00
Rev. D
Htotal
Htotal is the total horizontal line time or horizontal period expressed in pixels. Htotal is the sum of the following
terms: Front Porch + Sync Pulse + Back Porch + Unblanked Period = Htotal
If the pixel clock frequency is known, use the FIND LINE RATE sub-menu item to determine the line rate. Then
calculate Htotal by the following formula.
Htotal = Pixel Clock Frequency
Line Rate
Select the Htotal sub-menu item and enter the derived value.
Line Rate
The line rate of the video signal is the horizontal line frequency expressed in Hertz. If the line rate is not known, use
the FIND LINE RATE sub-menu item to determine the line rate.
Composite Sync Termination
For video with TTL level sync, select NO. For video with approximately 300 mV sync level select, YES.
Interlaced Video
This parameter indicates whether the video source is interlaced or non-interlaced. Enter the INTERLACED VIDEO
sub-menu item and select YES if interlaced or NO for non-interlaced. If the interlace/non-interlace is not known, the
FIND LINE RATE sub-menu item will find it automatically.
Rotate Image
If the host system display is rotated to portrait, use this menu item to send the acquired image in landscape.
Odd Start (MFG Only)
If the host system updates image information at the start of the odd field, set odd start to YES. Captured cine
images will contain motion artifacts if this item is set incorrectly.
PLL Pre Vertical Sync Coast (MFG Only)
PLL Pre Vertical Sync Coast is used to set the number of lines, prior to vertical sync, to be ignored.
PLL Post Vertical Sync Coast (MFG Only)
PLL Post Vertical Sync Coast is used to set the number of lines, after vertical sync, to be ignored.
Sync on Green Threshold (MFG Only)
Sync on Green Threshold is used to compensate for marginal sync signals.
Display Lock (MFG Only)
Display Lock (Yes) sets video output parameters to match video input parameters.
Output Parameters (MFG Only)
Output Parameters are used to configure the video output of the DiCOM box for an external display.
Display Vertical Reference
Display Vertical Reference allows configuration of the amplitude of the video output signal.
Display Setup
Display Setup is used to incorporate a setup or pedestal in the video output signal.
Display Sync Type
Display Sync Type allows configuration of the video output sync.
Display Interlace
Display Interlace sets the video output to interlaced or non-interlaced (Progressive scan).
Display Line Rate
Display Line Rate sets the frequency of the video output.
Display Htotal
Display Htotal allows configuration of the duration of the horizontal period. Htotal can be calculated by dividing
pixel clock frequency by the line rate.
Display Vtotal
Display Vtotal allows configuration of the duration of the vertical period. Vtotal can be calculated by dividing the
line rate by the frame rate.
Display H Sync Polarity
Display H Sync Polarity allows configuration of the polarity of the output horizontal sync signal.

005-00148-00 16
Rev. D
Display V Sync Polarity
Display V Sync allows configuration of the polarity of the output vertical sync signal.
Display H Sync
Display H Sync allows configuration of the duration of the output horizontal sync signal.
Display V Sync
Display V Sync allows configuration of the duration of the output vertical sync signal.
Display Pre Equalization
Display Pre Equalization allows configuration of the duration of the Pre Equalization period.
Display Post Equalization
Display Post Equalization allows configuration of the duration of the Post Equalization period.
RGB or Monochrome Phase Delay
Use this sub-menu item to create a phase shift relative to the current pixel clock value.
Horizontal Pixel Resolution
If the horizontal pixel aspect ratio of the captured image is not correct, use this sub-menu item to correct it. For
example, if a square object in the image is 320 pixels wide and 300 pixels high, change horizontal pixel resolution to
320 and vertical pixel resolution to 300.
Vertical Pixel Resolution
If the vertical pixel aspect ratio of the captured image is not correct, use this sub-menu item to correct it.
RGB or Monochrome Image Lightness
The ADC calibration process should result in image captures with maximum dynamic range. This sub-menu is
provided to allow manual adjustment of the ADC image lightness.
RGB or Monochrome Image Darkness
The ADC calibration process should result in image captures with maximum dynamic range. This sub-menu item is
provided to allow manual adjustment of the ADC image darkness value.
Set to Defaults
Use this sub-menu item to reset all video parameters to their original values.
Save Host to Disk (Floppy Disk)
When a listed host video setup has been modified, use this menu item to save the modified host video setup.
Save as a New Host
When a listed host video setup has been modified, (i.e., the installation required using unclassified/generic video for
the host system), use this menu item to name and save the modified host video setup.
7.2.3.2 Left Margin
If the left margin of the image is too large, i.e. too much black area to the left of the image, increase the left margin
pixel value. If the left margin of the image is too small, i.e., the image is cropped on the left, decrease the left margin
pixel value.
7.2.3.3 Width
If the width of the image is too large, i.e. too much black area to the right of the image, decrease the width pixel
value. If the width of the image is too small, i.e., the image is cropped on the right, increase the width pixel value.
7.2.3.4 Top Margin
If the top margin of the image is too large, i.e. too much black area above the image, increase the top margin line
value. If the top margin of the image is too small, i.e., the image is cropped on the top, decrease the top margin line
value.
7.2.3.5 Height
If the height of the image is too large, i.e. too much black area below the image, decrease the height line value. If
the height of the image is too small, i.e., the image is cropped at the bottom, increase the height line value.
7.2.3.6 Display Top Margin (MFG Only)
Display top margin allows adjustment of the Vertical Back Porch of the video output.
7.2.3.7 Display Left Margin (MFG Only)
Display left margin allows adjustment of the Horizontal Back Porch of the video output.

17 005-00148-00
Rev. D
7.2.4 Find Parameters
Use the find parameters menu item to determine the line rate and/or Htotal for host systems that were not listed in section
7.2. If you have a video specification for the host system you are connecting to, go to the MODIFY PARAMETERS menu
item and enter the video parameters.
7.2.4.1 Automatic
The FIND PARAMETERS AUTOMATIC sub-menu item attempts to determine the timing characteristics of the host
system video. Before proceeding, make sure that the image on the host scanner contains many high contrast,
single pixel wide features(i.e., a SMPTE Pattern will usually contain adjacent on, then off, then back on, single
pixels. This feature is also normally present in a multi-burst resolution pattern.). When the procedure is complete,
go to MODIFY PARAMETERS to review and/or change the results. NOTE: The “found” value for Htotal must be
greater than the width of the image (i.e., if the image size is 640 x 480, Htotal must be greater than 640).
7.2.4.2 Find Htotal Coarse (EFG Only)
The FIND HTOTAL COARSE sub-menu item will cause two four up sheets to be printed. Htotal will be incriminated
by 16 pixels in each frame. The image on the host system must contain as much high contrast information as
possible. In the illustration below, the captured image is filled with text. The printed images will contain node lines of
varying width and frequency. As the Htotal value approaches the correct number for the connected video, the width
of the node lines will increase and the frequency of node lines in the frame will decrease. Access the Htotal sub-
menu item to enter the rough value and proceed to Find Htotal Fine.
7.2.4.3 Find Htotal Fine (EFG Only)
The FIND HTOTAL FINE sub-menu item will cause two four up sheets to be printed. Htotal will be incriminated by 4
pixels in each frame. The image on the host system must contain as much high contrast information as possible. In
the illustration below, the captured image is filled with text. The printed images will contain node lines of varying
width and frequency. As the Htotal value approaches the correct number for the connected video, the width of the
node lines will increase and the frequency of node lines in the frame will decrease. Access the Htotal sub-menu
item to set Htotal to the value indicated in the best image of the two prints. If further testing is indicated, change the
Htotal value by two and repeat FIND HTOTAL FINE sub-menu item to finish determining the correct value for Htotal.

7.2.4.4 Find Vtotal (MFG Only)


The FIND VTOTAL sub-menu item determines the number of lines for the video signal.
7.2.4.5 Find Line Rate
The FIND LINE RATE sub-menu item determines the horizontal frequency, and whether the video is interlaced or
non-interlaced, on the host system video. If the pixel clock frequency is known, Htotal can be derived by dividing the
pixel clock frequency by the line rate.

005-00148-00 18
Rev. D
7.2.4.6 Automatic Sizing
Use the AUTOMATIC SIZING sub-menu item to find initial values for LEFT MARGIN, WIDTH, TOP MARGIN and
HEIGHT. Use the CALIBRATE PHASE and CALIBRATE ADC sub-menu items prior to performing an AUTOMATIC
SIZING. After using the AUTOMATIC SIZING sub-menu repeat the CALIBRATE PHASE and CALIBRATE ADC
menu items and then the AUTOMATIC SIZING sub-menu item. The found video may contain information not
intended to be displayed, such as hidden digital patient information. Proceed to the MODIFY PARAMETERS sub-
menu to eliminate any extraneous information present in the captured image.
7.2.5 ACR Setup
7.2.5.1 Introduction
This procedure describes how to modify the parameters that define the format of the patient name, identification,
(ID), and the accession number as they appear on the host system’s display screen. These parameters are used by
the Automatic Character Recognition (ACR) process which is exercised each time the operator starts a new exam
by pressing the NEW/END EXAM key on the DiCOM box Control Panel. Each model of host system may place the
patient name, ID, and accession number character strings in different locations in the image. The format of these
fields is varied as well as the character font. The ACR parameters include the location of the fields within the image,
the format of the name and ID fields, and information about the character font.
7.2.5.2 Procedure
The name and ID format parameters are changeable within the ACR Setup menu. The ACR Setup menu is part of
the Installation Menu. (See section 7.1.1 in this manual on how to enable access to the Installation Menu). The
procedure is divided into nine tasks.
Task 1. Name ID Field Order
The first task is to select the number of text lines in the image that are provided for the name and/or ID fields. The
control panel will display:
N A M E / I D F I E L D O R D E R
 * I n S e p a r a t e F i e l d s
I D b e f o r e N a m e
N a m e b e f o r e I D
Enter ID before name or name before ID if the name and ID character strings appear on a single line and there is no
set starting point for the two fields. Enter in separate fields if the name and ID character strings are on separate lines
or if there are distinct starting points for the two fields displayed on one line. Press the CANCEL key to keep the
previous setting and continue to the next task. If you have a video specification for the host system you are
connecting to, go to the MODIFY PARAMETERS menu item and enter the video parameters.
Task 2. Select Name Format
This task selects one of the following name formats:
FIRST LAST
LAST, FIRST MIDDLE
FIRST MIDDLE LAST
LAST, FIRST MI.
FIRST MI. LAST
LAST
LAST, FIRST
NOTE: FIRST = first name, MIDDLE = middle name, LAST = last name, and MI. = middle initial.
The Control Panel will display:
S E L E C T N A M E F O R M A T
 * L a s t F i r s t M i d d l e
F i r s t L a s t
F i r s t M i d d l e L a s t
L a s t F i r s t M i .
F i r s t M i . L a s t
L a s t
L a s t F i r s t
The default name format is LAST, FIRST MIDDLE. Check with the customer to obtain their preference. Select the
desired order for the name format and press the OK key. Press the CANCEL key to keep the previous value and
continue to the next task.

19 005-00148-00
Rev. D
Task 3. Setup Accession
This task defines the location of the accession field within the image. The Control Panel will display:
A C C E S S I O N F I E L D
 * E n a b l e A c c e s s i o n
D i s a b l e A c c e s s i o n
Select either to enable or disable the accession field. Press OK to proceed with ACR setup for name/ID/accession
number fields. If the accession number field has been disabled continue the ACR setup by proceeding to Task 4,
and disregard references to accession field setup.
Task 4. Font Spacing
If the space between characters is independent of the actual character, (i.e., two lower case L’s are not closer
together than other characters), then set font spacing to “EQUALLY”. Otherwise select “PROPORTIONATELY”.
Task 5. Active Clock Field (Automatic Setup Only)
If NAME, ID, and/or ACCESSION fields are found using the Automatic Define feature, this item must be set to YES.
If manual entry for CLOCK field, NAME field, ID field and ACCESSION field will be used, skip this task.
Task 6. Define Clock Field
* See note in Task 9 prior to proceeding. Manually Define Clock Field:
Send an image to a DICOM validation tool and open the image with image viewing software. Locate the Clock Field
and note the X and Y pixel coordinates for the upper left hand corner and the lower right hand corner of the field.
Use the upper left hand coordinates for X0 and Y0. Use the lower right hand coordinates for X1 and Y1.
Automatically Define Clock Field:
Empty the NAME, ID and ACCESSION fields and press OK.
Task 7. Define Name Field
* See note in Task 9 prior to proceeding.
It is necessary to fill the name field with characters to insure that the whole field is defined. For example. you may
need to pick the widest character at the beginning and end of the field to insure that the maximum field width is
represented. If the character font includes characters with true descending components, e.g. a lower-case “g”, then
those characters should be included near the end of the field to insure that the maximum field height is defined.
(See Illustration 1 below.) Another characteristic to be aware of is proportionally spaced characters that may be in
some character fonts. Proportionally spaced fonts are those in which the narrow characters occupy less horizontal
space than wider characters. Some host systems that have a proportionally spaced font have a limit to the number
of characters in a given line of text. In this case, if name and ID fields were filled with narrow characters, it would not
represent the maximum width of the field. Therefore, for the purposes of this task, it may be necessary to fill the field
with wide characters. Manually Define Name Field:
Send an image to a DICOM validation tool and open the image with image viewing software. Locate the Name Field
and note the X and Y pixel coordinates for the upper left hand corner and the lower right hand corner of the field.
Use the upper left hand coordinates for X0 and Y0. Use the lower right hand coordinates for X1 and Y1.
Automatically Define Name Field:
This task defines the location of the name field within the image. If the Name ID field order parameter entered in
Task 1 is Name before ID or ID before Name, then this task defines the location of the combined name and ID
fields. If there is an active clock in the image, its location must be identified prior to establishing the Name field (see
Task 6). Start an exam at the host system but leave the patient name and ID fields blank.
The Control Panel will display:
D E F I N E N A M E F I E L D
E m p t y n a m e , I D , a n d
a c c e s s i o n f i e l d s
P r e s s O K o r C A N C E L
Use the procedure provided by the host instrument to enter characters into the patient name field. Be certain that
the ID and Accession fields are empty. After you have filled the field with characters, press the OK key.
The Control Panel will display:
D E F I N E N A M E F I E L D
F i l l t h e n a m e f i e l d
w i t h c h a r a c t e r s
P r e s s O K o r C A N C E L

005-00148-00 20
Rev. D
If you want to skip over the task of defining the location of the name field, press the CANCEL key. If “# of lines”
equals “2”, then the procedure accepts the previously defined name field location and then skips to task 8. If “# of
lines” equals “1”, then the procedure accepts the previously defined name field location and skips to task 9.
WWWWWWWWWWWWWWWjW
Illustration 1. The patient name field has been filled with characters including lower case characters that extend
below the baseline.
Task 8. Define ID Field
* See note in Task 9 prior to proceeding.
Manually Define ID Field:
Send an image to a DICOM validation tool and open the image with image viewing software. Locate the ID Field
and note the X and Y pixel coordinates for the upper left hand corner and the lower right hand corner of the field.
Use the upper left hand coordinates for X0 and Y0. Use the lower right hand coordinates for X1 and Y1.
Automatically Define ID Field:
This task defines the location of the ID field within the image. If the Name ID field order parameter is Name before
ID or ID before Name, this task is not performed. See task 5 for instructions on preparing for this task. The Control
Panel will display:
D E F I N E I D F I E L D
F i l l t h e I D f i e l d
w i t h c h a r a c t e r s
P r e s s O K o r C A N C E L
Use the procedure provided by the host instrument to enter characters into the patient ID field (see illustration 2). Be
certain that the Name and Accession fields are empty. After you have filled the field with characters, press the OK
key.
WWWWWWWWWWWWWWWjW
Illustration 2. The patient ID field has been filled with characters including lower case characters that extend below
the baseline.
If you want to skip over the task of defining the location of the ID field, press the CANCEL key. The procedure
accepts the previously defined ID field location and continues on to the next task.
Task 9, Define ACCESSION Field
* See note below prior to proceeding.
Manually Define Accession Field:
Send an image to a DICOM validation tool and open the image with image viewing software. Locate the Accession
Field and note the X and Y pixel coordinates for the upper left hand corner and the lower right hand corner of the
field. Use the upper left hand coordinates for X0 and Y0. Use the lower right hand coordinates for X1 and Y1.
Automatically Define Accession Field:
This task defines the location of the ACCESSION field within the image. See task 4 for instructions on preparing for
this task. The Control Panel will display:
D E F I N E A C C F I E L D
F i l l t h e a c c f i e l d
w i t h c h a r a c t e r s
P r e s s O K o r C A N C E L
Use the procedure provided by the host instrument to enter characters into the accession field (see illustration 3).
Be certain that the Name and ID fields are empty. After you have filled the field with characters, press the OK key.
WWWWWWWWWWWWWWWjW
Illustration 3. The accession field has been filled with characters including lower case characters that extend below
the baseline.
If you want to skip over the task of defining the location of the ACCESSION field, press the CANCEL key. The
procedure accepts the previously defined ACCESSION field location and continues on to the next task.
*NOTE: If manual entry of CLOCK, NAME, ID or ACCESSION fields is attempted, the following hardware
and software are required:
1. PC with network interface card
2. ADVTTM (AGFA DICOM Validation tool) or equivalent
3. ADOBE PHOTOSHOPTM or equivalent
21 005-00148-00
Rev. D
ADVT if a registered trademark of AGFA Incorporated.
PHOTOSHOP is a registered trademark of ADOBE Systems Incorporated.
Task 10. ACR Font Setup
The purpose of these tasks is to obtain a bit map of each character in the host instrument’s character font. This is
accomplished by having the installer enter each requested character into the name field. The Control Panel will
display:
E n t e r c h a r a c t e r s
i n t o n a m e f i e l d
A B C D E F G H I J K L M
  O K o r C a n c e l
Use the procedure provided by the host instrument to enter the requested characters into the patient name field.
The first characters requested are “ABCDEFGHIJKLM”, (See Illustration 4). If the requested character isn’t
available on the host instrument’s keyboard, leave a space in place of the missing character. Care needs to be
taken to insure that the requested characters are the only characters in the field. For example, if there is a cursor
displayed when entering characters into the name field, you need to move the cursor to another field. Or, if the host
instrument automatically inserts a comma after the last name, leave the last name blank and insert the requested
character set into the first name area.
Name: ABCDEFGHIJKLM
Illustration 4. The requested characters, (‘ABCDEFGHIJKLM’ in this case), are the only characters in the name
field. See Appendix A for a list of the requested characters groups.
Press the OK key when the requested characters have been entered into the name field. The UP key will skip the
current requested characters. The last accepted character set bit map for that character set will be used. The
DOWN key will cause the previous requested character set to become the current requested character set. The
CANCEL key will continue to the next task without saving any changes in the font. If the CANCEL key is pressed,
the control panel will display:
A b o r t f o n t s e t u p ?

Press the OK key to continue to the next task without saving any changes to the font. Press the CANCEL key to
return to the previous control panel display.
Task 11. Define Space Minimum Width
The minimum space between characters is automatically found in task 10. This menu item allows adjustment of the
automatically found value.
Task 12. Auto Threshold
This menu item is used to automatically determine the pixel value differential between the character to be ACR’ed
and the background for that character.
Task 13. Set Threshold
This menu item is used to set the threshold value for the ACR function. Determine the pixel value differential
between the character to be ACR’ed and the background adjacent to the character using image viewing software.
Sum the two values and divide by two. Enter this value as the threshold.
Task 14. Set Polarity
This menu item is used to setup polarity of the ACR’ed text. If the text in the Patient NAME, ID, and ACCESSION
number, are black on a white background set polarity to “INVERSE.” Otherwise set polarity to “NORMAL.”
Allowable Character List
CAUTION: The AAM will only accommodate the following defined list of characters for use in patient ID and name
sensing.

005-00148-00 22
Rev. D
Any attempt to use characters not on this list may result in erroneous messages presented on the Control Panel
display. This list of characters does not apply to Control Panels with extended character support.
0 1 2 3 4 5 6 7 8 9

A B C D E F G H I J K L M N O P Q R S T U V W X Y Z

a b c d e f g h i j k l m n o p q r s t u v w x y z

SP ! “ # $ % & ‘ ( ) * + ‘ - . /

: ; < = > ? [ ] ^ _ @ ~ { }

7.3 DiCOM box Option Configurations


Allows enabling of Color printer support, host control support, storage class support, print class support, high line rate video
capture, worklist retrieval, frame grabber configuration (MFG Only) and or 12 bit image acquisition (DFG only).
7.4 Setup Network
This sub-menu is used to configure the DICOM and TCP/IP network parameters for the DiCOM box and the destination
printer, Archive Server or Worklist Server. IP Address is the IP address of the DiCOM box or destination server. Subnet is
the subnet mask for the subnet on which the DiCOM box or destination server will be placed. Gateway is the default
gateway address for dynamic routing to other subnets. Please consult your network administrator for information on how
these parameters should be configured.
NOTE: Always disconnect from the network prior to making changes in setup network.
7.4.1 Modify the DiCOM box
M O D I F Y t h e D I C O M b o x
 I P A d d r e s s
S u b n e t M a s k
G a t e w a y
A E T i t l e
N e t w o r k S p e e d
Selection of any of these menu items will invoke a input field such as the one shown below:
I P A D D R E S S
1 2 3 . 4 5 6 . 7 8 9 . 0 1 1
The current address will be displayed on the hand controller. Use the network speed menu item to set the speed, 100baseT
or 10baseT, and duplex, full or half. The default setting is auto which allows the DiCOM box to negotiate the network speed.
7.4.2 Ping Printer
Use this sub-menu to ping a defined print destination on the LAN.
7.4.3 Add Printer
Use this sub-menu to add a second print destination on the LAN.
7.4.4 Modify Printer
Use this sub-menu to define a print destination specifically for the LAN connected to the DiCOM box.
7.4.5 Delete Printer
Use this sub-menu to remove a print destination on the LAN.
7.4.6 Ping Worklist Server
Use this sub-menu to ping a defined Worklist Server destination on the LAN.
7.4.7 Add Worklist Server
Use this sub-menu to add a second Worklist Server destination on the LAN.
7.4.8 Modify Worklist Server
Use this sub-menu to define a Worklist Server destination specifically for the LAN connected to the DiCOM box.
7.4.9 Delete Worklist Server
Use this sub-menu to remove a Worklist Server destination on the LAN.
7.4.10 Ping MPPS Server
Use this sub-menu to ping a defined MPPS Server destination on the LAN.

23 005-00148-00
Rev. D
7.4.11 Add MPPS Server
Use this sub-menu to add a second MPPS Server destination on the LAN.
7.4.12 Modify MPPS Server
Use this sub-menu to define a MPPS Server destination specifically for the LAN connected to the DiCOM box.
7.4.13 Delete MPPS Server
Use this sub-menu to remove a MPPS Server destination on the LAN.
7.4.14 Ping Commit Server
Use this sub-menu to ping a defined Commit Server destination on the LAN.
7.4.15 Add Commit Server
Use this sub-menu to add a second Commit Server destination on the LAN.
7.4.16 Modify Commit Server
Use this sub-menu to define an Commit Server destination specifically for the LAN connected to the DiCOM box.
7.4.17 Delete Commit Server
Use this sub-menu to remove an Commit Server destination on the LAN.
7.4.18 Ping Archive Server
Use this sub-menu to ping a defined Archive Server destination on the LAN.
7.4.19 Add Archive Server
Use this sub-menu to add a second Archive Server destination on the LAN.
7.4.20 Modify Archive Server
Use this sub-menu to define an Archive Server destination specifically for the LAN connected to the DiCOM box.
7.4.21 Delete Archive Server
Use this sub-menu to remove an Archive Server destination on the LAN.
7.4.22 Archive Storage Class
Use this sub-menu to select the modality to which the DiCOM box is connected. Store aspect ratio allows changing the
pixel aspect ratio of the stored image. Store as 12 bit enables sending an 8 bit image to an archive as a 12 bit image. The
send cropped image function allows elimination of non-diagnostic portions of the captured image. Coordinates X0 and Y0
correspond to the upper left corner of the cropped image, coordinates X1 and Y1 correspond to the lower right corner of the
cropped image.
7.5 Exam ID Config
This operation allows for selection of the method to be used for generating an exam ID. In Query mode, the user will be
queried for an exam ID whenever a new exam is created. In Query Alphanumeric mode, the user enters the exam
information from the optional keyboard. In ACR mode the DiCOM box detects the patient name and ID if present, from the
host video signal. Auto-open with ACR will perform the exam open function when any of the Capture keys are pressed.
Auto Date/Time is used when the DiCOM box is in Host Control Mode. In Auto Unique ID mode, the DiCOM box will
automatically generate a unique exam ID whenever a new exam is created (without prompting the user). In Worklist/
Accession mode, the operator is prompted to enter the accession number for the exam, when the New/End Exam key has
been pressed. In Worklist Search mode, the worklist server is queried when the New/End key has been pressed. The
operator is then presented with a list of patients to select from. Worklist with ACR captures the exam ID from the modality
display and queries the worklist server with the exam ID. Use this selection list to specify the type of exam ID to use. The
current setting will be preceded by an asterisk.
E X A M I D C O N F I G
 * Q u e r y
Q u e r y A l p h a n u m e r i c
A u t o U n i q u e I D
A u t o D a t e / T i m e
A C R
A u t o - o p e n w i t h A C R
W o r k l i s t S e a r c h
W o r k l i s t / A c c e s s i o n
W o r k l i s t w i t h A C R

005-00148-00 24
Rev. D
7.6 Exam Transfer Mode
Allows setting the exam transfer to archive (PACS) to be during the exam or after the exam is closed.
7.7 Product Name Entry
Allows configuration of the product name as presented in the INFO screen.
7.8 Institution Name Entry
Allows identifying to PACS the institution where the equipment is located that produced the digital image.
7.9 Station Name
Allows the user to define the name identifying the machine that produced the digital image to PACS.
7.10 Host Control Configuration
Allows for setup of the serial communication for host control. Serial communication is jumper selectable for RS232 or
RS422 operation (see appendix A for jumper location). Host control mode is selectable for M952, P831, KCL or SPCI. The
SPCI selection contains four sub-selections for specific protocol types and auto film on/off selection to accommodate SPCI
implementations that do not send an end sheet command.
7.11 Frame Grabber Configuration (CA+ Only)
Allows the frame grabber configuration to be changed. Option J must be enabled to access this menu item. If the frame
grabber configuration is changed, the following message is displayed:
Updating frame grabber microcode. This takes a minute. DO NOT INTERRUPT!
If power is lost during the updating process, the DICOM box may require service.
If the host video signal pixel clock frequency exceeds 40 MHz for RGB or 80 MHz for monochrome, the framegrabber
config menu item allows changing the framegrabber micro code to allow higher frequency video signals to be captured.
7.12 Temperature Limit
Allows setting a CPU temperature threshold that will alert the operator with the following message if the threshold is
exceeded:
Internal temperature limit exceeded. Check fan filter. Press OK to continue.
7.13 Clean Fan Filter
Allows setting a timer to remind the operator to clean the fan filter.
7.14 Load Backup Files
If pre-installation site survey has determined that site-specific module data files are necessary, and dynamic, i.e. keypad,
entry of such data will not be completed perform the following:
NOTE: Option keys are module specific. Do not load Option keys from another DiCOM box.
1. Remove the the DiCOM box cover, or floppy drive access cover. (See Figure 1 Floppy Disk Access Cover Diagram)
2. Insert the floppy disk containing the appropriate table or file into the floppy disk drive located inboard of the module
side panel.
3. Select and access the Load Backup Files item from the Installation menu.
4. After confirming the operation, wait for the control panel to indicate that the operation is complete.
5. Eject the floppy disk. Press the OK/Enter key to re-start the module.
6. Reinstall the the DiCOM box cover, or floppy drive access cover.
NOTE: Power must be cycled before the new files or tables take effect.

25 005-00148-00
Rev. D
7.15 Saving Backup Files
NOTE: One copy for site, one copy to take off site and retain.
After completing the installation of the DiCOM box CA or DVI a backup of the installation should be performed.
Selecting save backup files saves AAM Files, the host table, destination files, host control settings, current host selection
and user setups. Do not load AAM Files onto another AAM.
Use the Save Host to Disk function (in Advanced Parameters) to save an Unclassified video file.
1. Remove the the DiCOM box cover, or floppy drive access cover.
2. Insert a formatted, blank floppy disk into the floppy disk drive on the DiCOM box.
3. Select Save Backup Files from the Installation menu.
4. After confirming the operation, wait for the control panel to indicate that the operation is complete.
5. Eject the floppy disk.
6. Reinstall the the DiCOM box cover or floppy drive access cover.
7.16 Save Log File to a Floppy Disk
This menu item is used to copy the network error log file from the DiCOM box’s hard disk to a floppy disk.
1. Insert a formatted, empty floppy disk into the floppy disk drive on the DiCOM box front panel. Select and
access the Save Log Files item from the Installation menu. The control panel will display the following
message to confirm the operation:
S a v e e r r o r l o g f i l e ?

2. Press the OK/ENTER key to save the error log to disk. Press the CANCEL key to return to the Installation
Menu.
3. Eject the floppy disk.
4. Reinstall the DiCOM box cover.
7.17 Final Installation Steps
Insert Supplement into Operator Manual and Reinstall DiCOM box Cover or floppy drive access plate.
If operator manual supplement(s) are associated with the current DiCOM box software installed, insert the supplement
page(s) into the Operator Manual. If the cover has not been installed, do so at this time.

005-00148-00 26
Rev. D
APPENDIX A RS-422 Host Control

Configure for RS-422 host control as follows:


NOTE: Always follow ESD precautions when handling static sensitive devices.
• Remove the DiCOM box cover (8 screws).
• See the diagram below for jumper location.
• At SC2T1 move the jumper from pins 1-2 to pins 3-4.
• At SC2T2 move the jumpers from 1-5, 2-6, 3-7 and 4-8 to 5-9, 6-10, 7-11 and 8-12.
• Replace the DiCOM box cover (8 screws).

27 005-00148-00
Rev. D
New Single Board Computer has revised jumper locations, pinout for RS232 and RS422 is identical to old SBC.

APPENDIX B REMOTE CONTROL

If the DiCOM box must be located away from the modality control console, the REMOTE CONTROL may be installed
adjacent to the control console for remote access to the DiCOM box system menu. The REMOTE CONTROL must be
configured prior to use, by setting it's IP address and entering the IP address of the DiCOM box it is intended to control.
Once configured, all key presses and display messages traffic is routed to and from the DiCOM box via the network
connection.
B.1 Features
B.1.1 Control Keypad Connection
The keypad is connected using a 9 Pin D connector.
B.1.2 Ethernet Connection
The Ethernet port is 10Base-T. Configure network node for 10 half UDP.
B.1.3 Power Connection
The power supply has an IEC320 type power cord connector. The power cord is appropriate for use in North America.

005-00148-00 28
Rev. D
B.2 Environmental

Operating
Temperature: +10(C to +40(C (+50(F to +104(F)
Barometric Pressure: 700hPa to 1060hPa
Relative Humidity: 5% to 80% (Non Condensing)
Transportation/Storage
Temperature: -10(C to +50(C (+14(F to +122(F)
Barometric Pressure: 500hPa to 1060hPa
Relative Humidity: 5% to 85% (Non Condensing)

Power On Indicator Network Activity Indicator

Network Link Indicator

KEYPAD

PWR ON

ACT LNK

NAI TECHNOLOGY PRODUCTS

MODEL NO: NAI PART NO: 105-00530-00

SERIAL NO: CUSTOMER PART NO:

MFR DATE:

NETWORK

POWER

B.3 Remote Control Installation Instructions


NOTE: Obtain an IP address for the REMOTE CONTROL prior to proceeding with installation.
1. Note or enter the network parameters for the DiCOM box to be accessed via the REMOTE CONTROL.
the DICOM box network parameters

IP Address

Subnet Mask

Gateway Address
2. Power down the DiCOM box and remove the keypad. Install the keypad on the REMOTE CONTROL, then apply
power to the REMOTE CONTROL.
NOTE: The keypad must be removed from the DiCOM box to allow the REMOTE CONTROL to establish a
connection.
Warning: Do not apply power to the REMOTE CONTROL prior to attaching the keypad.
29 005-00148-00
Rev. D
3. As the REMOTE CONTROL initializes, the following banner will be displayed for a short period of time:
REMOTE CONTROL V2 . 0 2 Contacting Host . . . . /
C o p y r i g h t (c) 2004 followed by:
NAI Technology
Products
Press the Cancel key four times, then press the Color Capture A key once, to access the REMOTE CONTROL
network configuration mode. This key sequence must be performed while the REMOTE CONTROL software version
is displayed.
4. Enter the IP address assigned to the REMOTE CONTROL when the following screen is displayed:

KeyPad ID Address

192.168.001.110

After entering the IP address press the OK key to proceed to the next screen.
5. Enter the Subnet Mask when the following screen is displayed:

Subnet Mask

255.255.255.000

After entering the Subnet Mask press the OK key to proceed to the next screen.
6. Enter the Gateway Address when the following screen is displayed:

Gateway Address

000.000.000.000

After entering the Gateway Address press the OK key to proceed to the next screen.
7. Enter the DiCOM box (Host) address when the following screen is displayed:

Host IP Address

192.168.001.100

After entering the IP address press the OK key to proceed to the next screen.
8. Connect the Patch cord to the REMOTE CONTROL, the following screen will displayed until the REMOTE
CONTROL establishes a connection with the DiCOM box.

Contacting Host..../

9. Once A connection is made, the following screen will be displayed.

Connected to Host

<host AE title>

Press OK to Continue

005-00148-00 30
Rev. D
REMOTE CONTROL Typical Connection Diagram

KEYPAD DISPLAY

the DiCOM box


Keypad
5

KEYPAD

TO NETWORK
DROP

POWER NETWORK

REMOTE CONTROL POWER SUPPLY


TO POWER
OUTLET

31 005-00148-00
Rev. D
APPENDIX C Flat Panel Touch Screen

The DiCOM box keypad may be replaced with a Flat Panel Touch Screen. This feature allows viewing Worklist search
results a page at a time. It also enables data entry for single patient Worklist queries.
C.1 Operater Interface
The flat panel is a keypad replacement, all functions supported by the keypad buttons are present in the flat panel touch
screen.

If exam ID config is set to Worklist Search, pressing the New/End Exam key will bring up the following screen:

Press the OK/Enter key to use the last list that was obtained from the Worklist server or select "Find New Worklist" to obtain
a new worklist. The worklist will be organized by the sort method that is specified in the Worklist Manager / Select Sort
Method menu item.

005-00148-00 32
Rev. D
Use the "Scroll Up" and "Scroll Down" keys to view the list of exams. Use the "Select Patient" key or touch the desired
exam to select an exam from the list.
C.2 Adjusting the Flat Panel Touch Screen
Use the buttons on the top of the monitor to apply power and make adjustments to the display. Press the Up or Down
button to activate the On Screen Display. Press the Up or Down button to select the screen function to be adjusted then
press the + or - button set the value for the function. Press the Up or Down button to exit the function adjustment screen.
See the touchscreen user's manual for detailed instructions on image quality adjustment.

33 005-00148-00
Rev. D
C.3 Connecting the Flat Panel Touch Screen to the DiCOM box
Make all connections between the DiCOM box and the Touch Screen prior to appling power to the DiCOM box and Touch
Screen.