Alarm 4

Picture

Input: E1.0
Type: internal hardware
Normal status: high
Bypass: 4
Activation: Nointerlock = 0 or 2 or 10 or 12
Existing on: FB40, FB60, FB80, FB120, FB160

Message
Interlock open. Close the front machine panel or check the key selector.

Problem and solution


The main door of the FB machine is open. This Alarm is on available only on old
machines. On new machines the Alarm number related to this problem is Alarm 0. If the
door is already closed, check the proximity switches. If the switches are OK, check the
electrical cables. On new machines this alarm is related to the key switch used to enable
or disable x-rays.