This Applications booklet summarizes the features and benefits of individual Clarus™ Arnel GC analyzers that are available for a
variety of petrochemical laboratory and process applications, from hydrogen to asphalt analysis.
More details and specifications are available for each model by obtaining the referenced companion literature for each product. Almost
100 turnkey analyzers are presented, ensuring that consultation with a PerkinElmer representative will deliver the analyzer that will
meet your needs.
Tables 1, 2 and 3 below show industry-standard methods compliance for standard model Refinery Gas Analyzers (RGA), Natural
Gas Analyzers (NGA) and Trace Sulfur Analyzer Systems. Other standard model analyzers for light gas, chlorine, transformer oil
gas, LPG and simulated distillation through detailed hydrocarbon analyses that are compliant with ASTM and various industry standard
methods are also summarized in this applications booklet.
PerkinElmer
ASTM Methods GPA Methods ISO Methods IP Methods
Arnel NGA Model
2001, 2101 D1945, D1946, D2597 2261 6568, 6974-5 None
2201, 2301 D1945 2177, 2261 6568, 6974-5 None
2002, 2102, 2202,
D1945 2261 6974-5 None
2302
2003, 2103, 2203,
D1945, D1946 2261 None None
2303
2206, 2306 None 2165, 2177, 2186 None None
2008 D1946 2286 6975 345
2108 D1946 2286 6974E, 6974-3, 6975 345
2208 D1946 2165, 2186, 2286 6974E, 6974-3, 6975 345
2009, 2109, 2209 D1946 2286 6975 345
2000, 2100, 2200,
D1945, D1946 None None None
2300,
2006, 2106 D1945, D1946 None None None
2010, 2110, 2210 D1945 2261 6974-5 None
1
Table of Contents
Refinery Gas Analyzers
Model 1001…………………………………………………………………………………….….5
Model 1015………………………………………………………………………………….…….6
Model 1016………………………………………………………………………………….…….7
Model 1101……………………………………………………………………………….……….8
Model 1115……………………………………………………………………………….……….9
Model 1116………………………………………………………………………………………10
Model 1201……………………………………………………………………………………....11
Model 1215.......................................................................................................................12
Model 1216.......................................................................................................................13
Model 1301.......................................................................................................................14
Model 1315.......................................................................................................................15
Model 1316.......................................................................................................................16
2
Trace Sulfur Analyzers
Model 4025.......................................................................................................................53
Model 4026.......................................................................................................................54
Model 4027.......................................................................................................................55
Model 4028.......................................................................................................................56
Model 4029.......................................................................................................................57
Model 4227.......................................................................................................................58
Model 4229.......................................................................................................................59
Model 4425.......................................................................................................................60
Model 4428.......................................................................................................................60
Model 4429.......................................................................................................................60
Model 4629.......................................................................................................................60
Other Analyzers
(For light gas, chlorine, transformer oil gas, LPG and simulated distillation through detailed hydrocarbon analyses)
Model 4001……………………………………………………………………………………….62
Model 4002……………………………………………………………………………………….62
Model 4003……………………………………………………………………………………….63
Model 4004……………………………………………………………………………………….64
Model 4005……………………………………………………………………………………….65
Model 4006……………………………………………………………………………………….66
Model 4007……………………………………………………………………………………….67
Model 4008……………………………………………………………………………………….67
Model 4009……………………………………………………………………………………….68
Model 4010……………………………………………………………………………………….68
Model 4011……………………………………………………………………………………….69
Model 4012……………………………………………………………………………………….69
Model 4013……………………………………………………………………………………….70
Model 4014……………………………………………………………………………………….71
Model 4015……………………………………………………………………………………….72
Model 4016……………………………………………………………………………………….73
Model 4017……………………………………………………………………………………….74
Model 4018……………………………………………………………………………………….75
Model 4019……………………………………………………………………………………….76
Model 4020……………………………………………………………………………………….77
Model 4021……………………………………………………………………………………….78
Model 4022……………………………………………………………………………………….79
Model 4023……………………………………………………………………………………….80
Model 4024……………………………………………………………………………………….81
Model 4032……………………………………………………………………………………….82
Model 4033……………………………………………………………………………………….82
Model 4034……………………………………………………………………………………….82
Model 4035……………………………………………………………………………………….83
Model 4036……………………………………………………………………………………….84
Model 4037……………………………………………………………………………………….85
Model 4080……………………………………………………………………………………….86
Model 4081……………………………………………………………………………………….63
Model 4082……………………………………………………………………………………….87
Model 4083……………………………………………………………………………………….88
Model 4086……………………………………………………………………………………….89
C5=: Pentenes
C6+: C6 hydrocarbons and higher
3
Refinery Gas Analyzers
Refine
Refinery Gas Analyzers
Model 1001
Model 1015
Model 1016
Model 1101
Model 1115
Model 1116
Model 1201
Model 1215
Model 1216
Model 1301
Model 1315
Model 1316
Refinery Gas Analyzer Model 1001
Refinery Gas Analyzers
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, ASTM D2597, UOP 539-73, DIN 51872-4
Types of sample analyzed: Gas
Compounds separated: H2; O2; N2; CO2; CO; H2S; C1 through C5; C5=, C6+, H2O, NH3
composite peak
Guaranteed detection ranges/concentration levels: H2 - 1 to 10%; H2S - 0.1 to 25%; all other
components - 0.01 to 100% except the C5=, C6+, H2O, NH3 composite peak - 0.01 to 40%
Application note reference number with technical details: D-6620
Typical Chromatogram
5
Refinery Gas Analyzer Model 1015
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, ASTM D2597, UOP 539-73, UOP 709, DIN
51872-4
Types of sample analyzed: Gas
Compounds separated: H2; O2; N2; CO2; CO; H2S; C1 through C5; C5=, C6+ composite peak
Guaranteed detection ranges/concentration levels: H2 - 1 to 10%; H2S - 0.02 to 100%; C1 through
C5 - 0.001 to 100% on FID; C5=, C6+ composite peak - 0.002 to 40%; O2, N2, CO2, CO, C1 and
C2 hydrocarbons on TCD - 0.01 to 100%
Application note reference number with technical details: D-6621
Typical Chromatograms
Propane
Ethane,Ethylene
i-Butane
Methane
Propylene
n-Butane
1,3-Butadiene
C5=,C6+
t-Butene-2
c-Butene-2
1-Butene
Propadiene
15 Minutes
n-Pentane
Acetylene
i-Pentane
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, ASTM D2597, UOP 539-73, UOP 709, DIN
51872-4
Types of sample analyzed: Gas
Compounds separated: H2; O2; N2; CO2; CO; H2S; C1 through C5; C5=, C6+, H2O, NH3
composite peak
Guaranteed detection ranges/concentration levels: H2 - 1 to 10%; H2S - 0.02 to 100%; C5=, C6+,
H2O, NH3 composite peak - 0.2 to 40%; all other components - 0.01 to 100%
Application note reference number with technical details: D-6621
Typical Chromatograms
Methane, Air
Ethane, Ethylene
Propane
C5=, C6+, H2O, NH3
i-Butane
Propylene
1,3-Butadiene
n-Butane
Acetylene
t-Butene-2
15 Minutes
Butene-1
CO2
c-Butene-2
Propadiene
i-Butylene
n-Pentane
i-Pentane
H2S
7
Refinery Gas Analyzer Model 1101
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, ASTM D2597, UOP 539-73, UOP 539-86,
DIN 51872-4
Types of sample analyzed: Gas
Compounds separated: He; H2; O2; N2; CO2; CO; H2S; C1 through C5; C5=, C6+, H2O, NH3
composite peak
Guaranteed detection ranges/concentration levels: H2S - 0.1 to 25%; all other components - 0.01
to 100% except the C5=, C6+, H2O, NH3 composite peak - 0.01 to 40%
Application note reference number with technical details: D-6620, D-6651, D-6653
• Full-range hydrogen detection using separate channel with nitrogen carrier gas
• Dual channel/dual thermal conductivity detector (TCD/TCD)
• Includes Model 3062 Software, which contains a merge program to combine
multichannel data into one report
• Rugged packed columns, no degradation from H2O, NH3 or other common
contaminants
• 35 - 40 minute analysis time (see Model 1015 for faster analysis time)
• Plumbed in sulfur-resistant materials to enhance H2S detection
Typical Chromatogram
8
Refinery Gas Analyzer Model 1115
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, ASTM D2597, UOP 539-73, UOP 539-86,
UOP 709, DIN 51872-4
Types of sample analyzed: Gas
Compounds separated: He; H2; O2; N2; CO2; CO; H2S; C1 through C5; C5=, C6+ composite peak
Guaranteed detection ranges/concentration levels: H2S - 0.02 to 100%; C1 through C5 - 0.001 to
100% on FID; C5=, C6+ composite peak - 0.002 to 40%; He, H2, O2, N2, CO2, CO, C1, C2
hydrocarbons on TCD - 0.01 to 100%
Application note reference number with technical details: D-6621, D-6651, D-6653
Typical Chromatograms
Propane
Nitrogen
Hydrogen
Ethane,Ethylene
i-Butane
Methane
Propylene
n-Butane
1,3-Butadiene
C5=,C6+
t-Butene-2
c-Butene-2
1-Butene
Methane
Propadiene
15 Minutes
n-Pentane
Ethane
Acetylene
CO2
Ethylene
Acetylene
15 Minutes
i-Pentane
Oxygen
H2S
CO
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, ASTM D2597, UOP 539-73, UOP 539-86,
UOP 709, DIN 51872-4
Types of sample analyzed: Gas
Compounds separated: He; H2; O2; N2; CO2; CO; H2S; C1 through C5; C5=, C6+, H2O, NH3
composite peak
Guaranteed detection ranges/concentration levels: H2S - 0.02 to 100%; C5=, C6+, H2O, NH3
composite peak - 0.2 to 40%; all other components - 0.01 to 100%
Application note reference number with technical details: D-6621, D-6651, D-6653
Typical Chromatograms
Methane, Air
Nitrogen
Hydrogen
Ethane, Ethylene
Propane
C5=, C6+, H2O, NH3
i-Butane
Propylene
1,3-Butadiene
n-Butane
Methane
Acetylene
Ethane
t-Butene-2
15 Minutes
CO2
Butene-1
CO2
c-Butene-2
Ethylene
Acetylene
Propadiene
i-Butylene
15 Minutes
n-Pentane
i-Pentane
Oxygen
H2S
H2S
CO
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, ASTM D2597, UOP 539-73, DIN 51872-4
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: H2; O2; N2; CO2; CO; H2S; C1 through C5; C5=, C6+, H2O, NH3
composite peak
Guaranteed detection ranges/concentration levels: H2 - 1 to 10%; H2S - 0.1 to 25%; all other
components - 0.01 to 100% except the C5=, C6+, H2O, NH3 composite peak - 0.01 to 40%
Application note reference number with technical details: D-6620, D-6652, D-6653
Typical Chromatogram
11
Refinery Gas Analyzer Model 1215
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, ASTM D2597, UOP 539-73, UOP 709, DIN
51872-4
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: H2; O2; N2; CO2; CO; H2S; C1 through C5; C5=, C6+ composite peak
Guaranteed detection ranges/concentration levels: H2 - 1 to 10%; H2S - 0.02 to 100%; C1 through
C5 - 0.001 to 100% on FID; C5=, C6+ composite peak - 0.002 to 40%; O2, N2, CO2, CO, C1, C2s
on TCD - 0.01 to 100%
Application note reference number with technical details: D-6621, D-6652, D-6653
Typical Chromatograms
Propane
Ethane,Ethylene
i-Butane
Methane
Propylene
n-Butane
1,3-Butadiene
C5=,C6+
t-Butene-2
c-Butene-2
1-Butene
Propadiene
15 Minutes
n-Pentane
Acetylene
i-Pentane
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, ASTM D2597, UOP 539-73, UOP 709, DIN
51872-4
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: H2; O2; N2; CO2; CO; H2S; C1 through C5; C5=, C6+, H2O, NH3
composite peak
Guaranteed detection ranges/concentration levels: H2 - 1 to 10%; H2S - 0.02 to 100%; C5=, C6+,
H2O, NH3 composite peak - 0.2 to 40%; all other components - 0.01 to 100%
Application note reference number with technical details: D-6621, D-6652, D-6653
Typical Chromatograms
Methane, Air
Ethane, Ethylene
Propane
C5=, C6+, H2O, NH3
i-Butane
Propylene
1,3-Butadiene
n-Butane
Acetylene
t-Butene-2
15 Minutes
Butene-1
CO2
c-Butene-2
Propadiene
i-Butylene
n-Pentane
i-Pentane
H2S
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, ASTM D2597, UOP 539-73, UOP 539-86,
DIN 51872-4
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: He; H2; O2; N2; CO2; CO; H2S; C1 through C5; C5=, C6+, H2O, NH3
composite peak
Guaranteed detection ranges/concentration levels: H2S - 0.1 to 25%; all other components - 0.01
to 100% except the C5=, C6+, H2O, NH3 composite peak - 0.01 to 40%
Application note reference number with technical details: D-6620, D-6651, D-6652, D-6653
• Full-range hydrogen detection using separate channel with nitrogen carrier gas
• Added liquid-sampling capability to Model 1101
• Includes sample-loop purge feature that prevents contamination when changing
sample types, a secondary filter and back pressure regulation for proper liquid
sample introduction, plus a vaporization zone for proper liquid sample injection
• Dual channel/dual thermal conductivity detector (TCD/TCD)
• Includes Model 3062 Software, which contains a merge program to combine
multichannel data into one report
• Rugged packed columns, no degradation from H2O, NH3 or other common
contaminants
• 35 - 40 minute analysis time (see Model 1015 for faster analysis time)
• Plumbed in sulfur-resistant materials to enhance H2S detection
Typical Chromatogram
14
Refinery Gas Analyzer Model 1315
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, ASTM D2597, UOP 539-73, UOP 539-86,
UOP 709, DIN 51872-4
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: He; H2; O2; N2; CO2; CO; H2S; C1 through C5; C5=, C6+ composite peak
Guaranteed detection ranges/concentration levels: H2S - 0.02 to 100%; C1 through C5 - 0.001 to
100% on FID; C5=, C6+ composite peak - 0.002 to 40%; He, H2, O2, N2, CO2, CO, C1 and C2
hydrocarbons on TCD - 0.01 to 100%
Application note reference number with technical details: D-6621, D-6651, D-6652, D-6653
Typical Chromatograms
Propane
Nitrogen
Hydrogen
Ethane,Ethylene
i-Butane
Methane
Propylene
n-Butane
1,3-Butadiene
C5=,C6+
t-Butene-2
c-Butene-2
1-Butene
Methane
Propadiene
15 Minutes
n-Pentane
Ethane
Acetylene
CO2
Ethylene
Acetylene
15 Minutes
i-Pentane
Oxygen
H2S
CO
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, ASTM D2597, UOP 539-73, UOP 539-86,
UOP 709, DIN 51872-4
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: He; H2; O2; N2; CO2; CO; H2S; C1 through C5; C5=, C6+, H2O, NH3
composite peak
Guaranteed detection ranges/concentration levels: H2S - 0.02 to 100%; C5=, C6+, H2O, NH3
composite peak - 0.2 to 40%; all other components - 0.01 to 100%
Application note reference number with technical details: D-6621, D-6651, D-6652, D-6653
Typical Chromatograms
Methane, Air
Nitrogen
Hydrogen
Ethane, Ethylene
Propane
C5=, C6+, H2O, NH3
i-Butane
Propylene
1,3-Butadiene
n-Butane
Methane
Acetylene
Ethane
t-Butene-2
15 Minutes
CO2
Butene-1
CO2
c-Butene-2
Ethylene
Propadiene
Acetylene
i-Butylene
15 Minutes
n-Pentane
i-Pentane
Oxygen
H2S
H2S
CO
Model 2000
Model 2001
Model 2002
Model 2003
Model 2006
Model 2008
Model 2009
Model 2010
Model 2100
Model 2101
Model 2102
Model 2103
Model 2106
Model 2108
Model 2109
Model 2110
Model 2200
Model 2201
Model 2202
Model 2203
Model 2206
Model 2208
Model 2209
Model 2210
Model 2300
Model 2301
Model 2302
Model 2303
Model 2306
Model 2310
Natural Gas Analyzer Model 2000
Analyzer Description
Typical Chromatogram
Methane
Ethane
Propane
n - Butane
i - Butane
n - Pentane
i - Pentane
C6+
2 4 6 8 10 12 14
Time (min)
Analyzer Description
Published methods met: ASTM D1945, GPA 2261, ISO 6568, ISO 6974-5, DIN 51872-5
Types of sample analyzed: Gas
Compounds separated: O2, N2 composite; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: O2, N2 composite and CO2 – 0.01 to 20%, C1
through C5 - 0.01 to 100%, H2S - 0.1 to 25%, C6+ composite - 0.01 to 10%
Application note reference number with technical details: D-6622, D-6624
Typical Chromatogram
Methane
O2/N2
Ethane
Propane
C6+
n - Butane
i - Butane
n - Pentane
i - Pentane
H2S
CO2
0 2 4 6 8 10 12 14
Time (min)
19
Natural Gas Analyzer Model 2002
Analyzer Description
Published methods met: ASTM D1945, GPA 2261, ISO 6974-5, DIN 51872-5
Types of sample analyzed: Gas
Compounds separated: Ar; O2, N2 composite; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: Ar – 0.01 to 4%, H2S - 0.1 to 5%, C6+
composite - 0.01 to 10%, all other components - 0.01 to 100%,
Application note reference number with technical details: D-6622, D-6625
Typical Chromatogram
Methane
O2 / N 2
Propane
Ethane
n - Butane
Argon
i - Butane
n - Pentane
i - Pentane
H2S
C6+
CO2
10 20
Time (min)
20
Natural Gas Analyzer Model 2003
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, GPA 2261, DIN 51872-5
Types of sample analyzed: Gas
Compounds separated: O2; N2; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: O2 and C6+ composite - 0.01 to 10%, N2 –
0.01 to 20%, H2S - 0.1 to 5%, CO2 and C1 through C5 - 0.01 to 100%,
Application note reference number with technical details: D-6622, D-6626
Typical Chromatogram
Methane
Nitrogen
Ethane
Propane
C6+
n-Butane
Oxygen
i-Butane
n-Pentane
i-Pentane
H2S
CO2
2 4 6 8 10 12 14
Time (min)
21
Natural Gas Analyzer Model 2006
Analyzer Description
Published methods met: None
Types of sample analyzed: Gas
Compounds separated: O2, N2 composite; CO2; H2S; C1 through C10; C6+ composite
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2, N2 composite and
CO2 – 0.01 to 20%, C1 through C5 - 0.01 to 100%, H2S - 0.1 to 25%, C6+ composite - 0.01 to
10%, Channel two (FID): C4 to C10 - 0.005 to 100% (depends upon split ratio)
Application note reference number with technical details: D-6622, D-6627
Typical Chromatograms
Methane
Hexane
Heptane
p-Xylene
Pentanes
O2/N2
Ethane
Octane
n-Propylbenzene
n-Butylbenzene
Propane
Butanes
C6+
n - Butane
i - Butane
Decane
n - Pentane
i - Pentane
H2S
CO2
0 2 4 6 8 10 12 14 10 20
Time (min) Time (min)
Analyzer Description
Published methods met: ASTM D1946, GPA 2286, IP 345, ISO 6975
Types of sample analyzed: Gas
Compounds separated: O2; N2; CO2; H2S; C1 through C10; C6+ composite
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2 and C6+ composite -
0.01 to 10%, N2 – 0.01 to 20%, H2S - 0.1 to 5%, CO2 and C1 through C5 - 0.01 to 100%
Channel two (FID): C4 to C10 - 0.005 to 100% (depends upon split ratio)
Application note reference number with technical details: D-6622, D-6628
Typical Chromatograms
Methane
Nitrogen
Hexane
Heptane
p-Xylene
Pentanes
Octane
n-Propylbenzene
Ethane
Propane
n-Butylbenzene
C6+
Butanes
n-Butane
Decane
Oxygen
i-Butane
n-Pentane
i-Pentane
H2S
CO2
10 20
2 4 6 8 10 12 14 Time (min)
Time (min)
23
Natural Gas Analyzer Model 2009
Analyzer Description
Published methods met: ASTM D1946, GPA 2286, IP 345, ISO 6975
Types of sample analyzed: Gas
Compounds separated: O2; N2; CO2; H2S; C1 through C10
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2 - 0.01 to 10%,
N2 – 0.01 to 20%, H2S - 0.1 to 5%, CO2 and C1 through C4 - 0.01 to 100% Channel two (FID):
C4 to C10 - 0.005 to 100% (depends upon split ratio)
Application note reference number with technical details: D-6622, D-6629
Typical Chromatograms
Hexane
Heptane
p-Xylene
Pentanes
Octane
n-Propylbenzene
n-Butylbenzene
Butanes
Decane
10 20
Time (min)
24
Natural Gas Analyzer Model 2010
Analyzer Description
Published methods met: ASTM D1945, GPA 2261, ISO 6974-5
Types of sample analyzed: Gas
Compounds separated: O2, N2 composite; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2, N2 composite and
CO2 – 0.01 to 20%; He, H2, C1 through C5 - 0.01 to 100%; H2S - 0.1 to 25% Channel two
(FID): C6+ composite - 0.001 to 100%
Application note reference number with technical details: D-6622, D-6630
Ethane
Propane
Carbon Dioxide
N-Butane
i-Butane
n-Pentane
I-Pentane
H2S
25
Natural Gas Analyzer Model 2100
Analyzer Description
Typical Chromatogram
26
Natural Gas Analyzer Model 2101
Analyzer Description
Published methods met: ASTM D1945, GPA 2261, ISO 6568, ISO 6974-5, DIN 51872-5
Types of sample analyzed: Gas
Compounds separated: He; H2; O2, N2 composite; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: He, H2, C1 through C5 – 0.01 to 100%; O2, N2
composite and CO2 – 0.01 to 20%, H2S - 0.1 to 25%, C6+ composite - 0.01 to 10%
Application note reference number with technical details: D-6622, D-6624, D-6651
Typical Chromatogram
27
Natural Gas Analyzer Model 2102
Analyzer Description
Published methods met: ASTM D1945, GPA 2261, ISO 6974-5, DIN 51872-5
Types of sample analyzed: Gas
Compounds separated: He; H2; Ar; O2, N2 composite; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: Ar – 0.01 to 4%, H2S - 0.1 to 5%, C6+
composite - 0.01 to 10%, all other components - 0.01 to 100%,
Application note reference number with technical details: D-6622, D-6625, D-6651
Typical Chromatogram
H2 measured in one TCD utilizing a nitrogen carrier gas and the remaining
compounds measured in a second TCD utilizing a helium carrier gas.
28
Natural Gas Analyzer Model 2103
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, GPA 2261, DIN 51872-5
Types of sample analyzed: Gas
Compounds separated: He; H2; O2; N2; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: O2 and C6+ composite - 0.01 to 10%; N2 –
0.01 to 20%; H2S - 0.1 to 5%; He, H2, CO2 and C1 through C5 - 0.01 to 100%,
Application note reference number with technical details: D-6622, D-6626, D-6651
Typical Chromatogram
Analyzer Description
Published methods met: None
Types of sample analyzed: Gas
Compounds separated: He; H2; O2, N2 composite; CO2; H2S; C1 through C10; C6+ composite
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2, N2 composite and
CO2 – 0.01 to 20%; He, H2, C1 through C5 - 0.01 to 100%; H2S - 0.1 to 25%; C6+ composite -
0.01 to 10%; Channel two (FID): C4 to C10 - 0.005 to 100% (depends upon split ratio)
Application note reference number with technical details: D-6622, D-6627, D-6651
Typical Chromatograms
p-Xylene
Hexane
Heptane
Pentanes
Octane
n-Propylbenzene
n-Butylbenzene
Butanes
Decane
10 20
Time (min)
Analyzer Description
Published methods met: ASTM D1946, GPA 2286, IP 345, ISO 6974E, ISO 6974-3, ISO 6975
Types of sample analyzed: Gas
Compounds separated: He; H2; O2; N2; CO2; H2S; C1 through C10; C6+ composite
Guaranteed detection ranges/concentration levels: Channel one (TCD): He, H2, O2 and C6+
composite - 0.01 to 10%; N2 – 0.01 to 20%; H2S - 0.1 to 5%; CO2 and C1 through C5 - 0.01 to
100% Channel two (FID): C4 to C10 - 0.005 to 100% (depends upon split ratio)
Application note reference number with technical details: D-6622, D-6628, D-6651
Typical Chromatograms
Hexane
Heptane
p-Xylene
Pentanes
Octane
n-Propylbenzene
n-Butylbenzene
Butanes
Decane
10 20
Time (min)
31
Natural Gas Analyzer Model 2109
Analyzer Description
Published methods met: ASTM D1946, GPA 2286, IP 345, ISO 6975
Types of sample analyzed: Gas
Compounds separated: He; H2; O2; N2; CO2; H2S; C1 through C10
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2 - 0.01 to 10%;
N2 – 0.01 to 20%; H2S - 0.1 to 5%; He, H2, CO2 and C1 through C4 - 0.01 to 100% Channel two
(FID): C4 to C10 - 0.005 to 100% (depends upon split ratio)
Application note reference number with technical details: D-6622, D-6629, D-6651
Typical Chromatograms
Hexane
Heptane
p-Xylene
Pentanes
Octane
n-Propylbenzene
n-Butylbenzene
Butanes
Decane
10 20
Time (min)
32
Natural Gas Analyzer Model 2110
Analyzer Description
Published methods met: ASTM D1945, GPA 2261, ISO 6974-5
Types of sample analyzed: Gas
Compounds separated: He; H2; O2, N2 composite; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2, N2 composite and
CO2 – 0.01 to 20%; He, H2, C1 through C5 - 0.01 to 100%; H2S - 0.1 to 25% Channel two
(FID): C6+ composite - 0.001 to 100%
Application note reference number with technical details: D-6622, D-6630, D-6651
Typical Chromatogram
Methane
Air
Ethane
Propane
Carbon Dioxide
N-Butane
i-Butane
n-Pentane
I-Pentane
H2S
Analyzer Description
Typical Chromatogram
Methane
Ethane
Propane
n - Butane
i - Butane
n - Pentane
i - Pentane
C6+
2 4 6 8 10 12 14
Time (min)
34
Natural Gas Analyzer Model 2201
Analyzer Description
Published methods met: ASTM D1945, GPA 2177, GPA 2261, ISO 6568, ISO 6974-5,
DIN 51872-5
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: O2, N2 composite; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: O2, N2 composite and CO2 – 0.01 to 20%, C1
through C5 - 0.01 to 100%, H2S - 0.1 to 25%, C6+ composite - 0.01 to 10%
Application note reference number with technical details: D-6622, D-6624, D-6652
Typical Chromatogram
Methane
O2/N2
Ethane
Propane
C6+
n - Butane
i - Butane
n - Pentane
i - Pentane
H2S
CO2
0 2 4 6 8 10 12 14
Time (min)
35
Natural Gas Analyzer Model 2202
Analyzer Description
Published methods met: ASTM D1945, GPA 2261, ISO 6974-5, DIN 51872-5
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: Ar; O2, N2 composite; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: Ar – 0.01 to 4%, H2S - 0.1 to 5%, C6+
composite - 0.01 to 10%, all other components - 0.01 to 100%,
Application note reference number with technical details: D-6622, D-6625, D-6652
Typical Chromatogram
Methane
O2 / N2
Propane
Ethane
n - Butane
Argon
i - Butane
n - Pentane
i - Pentane
H2S
C6+
CO2
10 20
Time (min)
36
Natural Gas Analyzer Model 2203
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, GPA 2261, DIN 51872-5
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: O2; N2; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: O2 and C6+ composite - 0.01 to 10%, N2 –
0.01 to 20%, H2S - 0.1 to 5%, CO2 and C1 through C5 - 0.01 to 100%,
Application note reference number with technical details: D-6622, D-6626, D-6652
Typical Chromatogram
Methane
Nitrogen
Ethane
Propane
C6+
n-Butane
Oxygen
i-Butane
n-Pentane
i-Pentane
H 2S
CO2
2 4 6 8 10 12 14
Time (min)
37
Natural Gas Analyzer Model 2206
Analyzer Description
Published methods met: None
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: O2, N2 composite; CO2; H2S; C1 through C10; C6+ composite
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2, N2 composite and
CO2 – 0.01 to 20%, C1 through C5 - 0.01 to 100%, H2S - 0.1 to 25%, C6+ composite - 0.01 to
10%, Channel two (FID): C4 to C10 - 0.005 to 100% (depends upon split ratio)
Application note reference number with technical details: D-6622, D-6627, D-6652
Typical Chromatograms
Methane
Pentanes
Hexane
Heptane
p-Xylene
O2/N2
Ethane
Octane
n-Propylbenzene
n-Butylbenzene
Propane
C6+
Butanes
n - Butane
i - Butane
Decane
n - Pentane
i - Pentane
H2S
CO2
0 2 4 6 8 10 12 14 10 20
Time (min) Time (min)
Analyzer Description
Published methods met: ASTM D1946, GPA 2165, GPA 2186, GPA 2286, IP 345, ISO 6975
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: O2; N2; CO2; H2S; C1 through C10; C6+ composite
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2 and C6+ composite -
0.01 to 10%, N2 – 0.01 to 20%, H2S - 0.1 to 5%, CO2 and C1 through C5 - 0.01 to 100%
Channel two (FID): C4 to C10 - 0.005 to 100% (depends upon split ratio)
Application note reference number with technical details: D-6622, D-6628, D-6652
Typical Chromatograms
Methane
Nitrogen
Hexane
Heptane
p-Xylene
Pentanes
Octane
n-Propylbenzene
Ethane
Propane
n-Butylbenzene
C6+
Butanes
n-Butane
Oxygen
Decane
i-Butane
n-Pentane
i-Pentane
H2S
CO2
10 20
2 4 6 8 10 12 14 Time (min)
Time (min)
Analyzer Description
Published methods met: ASTM D1946, GPA 2286, IP 345, ISO 6975
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: O2; N2; CO2; H2S; C1 through C10
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2 - 0.01 to 10%;
N2 – 0.01 to 20%; H2S - 0.1 to 5%; CO2 and C1 through C4 - 0.01 to 100% Channel two (FID):
C4 to C10 - 0.005 to 100% (depends upon split ratio)
Application note reference number with technical details: D-6622, D-6629, D-6652
Typical Chromatograms
Hexane
Heptane
p-Xylene
Pentanes
Octane
n-Propylbenzene
n-Butylbenzene
Butanes
Decane
10 20
Time (min)
40
Natural Gas Analyzer Model 2210
Analyzer Description
Published methods met: ASTM D1945, GPA 2261, ISO 6974-5
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: O2, N2 composite; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2, N2 composite and
CO2 – 0.01 to 20%, C1 through C5 - 0.01 to 100%, H2S - 0.1 to 25%, Channel two (FID): C6+
composite - 0.001 to 100%
Application note reference number with technical details: D-6622, D-6630, D-6652
Typical Chromatogram
Analyzer Description
Published methods met: ASTM D1945 (precision statement only)
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: He; H2; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: He, H2 – 0.01 to 100%; all other components
except C6+ - 0.005 to 100%; C6+ - 0.005 to 10%
Application note reference number with technical details: D-6622, D-6623, D-6651, D-6652
Typical Chromatogram
Analyzer Description
Published methods met: ASTM D1945, GPA 2177, GPA 2261, ISO 6568, ISO 6974-5,
DIN 51872-5
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: He; H2; O2, N2 composite; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: He, H2, C1 through C5 - 0.01 to 100%, O2, N2
composite and CO2 – 0.01 to 20%, H2S - 0.1 to 25%, C6+ composite - 0.01 to 10%
Application note reference number with technical details: D-6622, D-6624, D-6651, D-6652
Typical Chromatogram
H2 measured in one TCD utilizing a nitrogen carrier gas and the remaining
compounds measured in a second TCD utilizing a helium carrier gas.
43
Natural Gas Analyzer Model 2302
Analyzer Description
Published methods met: ASTM D1945, GPA 2261, ISO 6974-5, DIN 51872-5
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: He; H2; Ar; O2, N2 composite; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: Ar – 0.01 to 4%, H2S - 0.1 to 5%, C6+
composite - 0.01 to 10%, all other components - 0.01 to 100%,
Application note reference number with technical details: D-6622, D-6625, D-6651, D-6652
Typical Chromatogram
H2 measured in one TCD utilizing a nitrogen carrier gas and the remaining
compounds measured in a second TCD utilizing a helium carrier gas.
44
Natural Gas Analyzer Model 2303
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946, GPA 2261, DIN 51872-5
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: He; H2; O2; N2; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: O2 and C6+ composite - 0.01 to 10%; N2 –
0.01 to 20%; H2S - 0.1 to 5%; He, H2, CO2 and C1 through C5 - 0.01 to 100%,
Application note reference number with technical details: D-6622, D-6626, D-6651, D-6652
Typical Chromatogram
H2 measured in one TCD utilizing a nitrogen carrier gas and the remaining
compounds measured in a second TCD utilizing a helium carrier gas.
45
Natural Gas Analyzer Model 2306
Analyzer Description
Published methods met: None
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: He; H2; O2, N2 composite; CO2; H2S; C1 through C10; C6+ composite
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2, N2 composite and
CO2 – 0.01 to 20%; He, H2, C1 through C5 - 0.01 to 100%; H2S - 0.1 to 25%; C6+ composite -
0.01 to 10%; Channel two (FID): C4 to C10 - 0.005 to 100% (depends upon split ratio)
Application note reference number with technical details: D-6622, D-6627, D-6651, D-6652
Typical Chromatograms
Hexane
Heptane
p-Xylene
Pentanes
Octane
n-Propylbenzene
n-Butylbenzene
Butanes
Decane
10 20
Time (min)
Analyzer Description
Published methods met: ASTM D1945, GPA 2261, ISO 6974-5
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: He; H2; O2, N2 composite; CO2; H2S; C1 through C5; C6+ composite
Guaranteed detection ranges/concentration levels: Channel one (TCD): O2, N2 composite and
CO2 – 0.01 to 20%; He, H2, C1 through C5 - 0.01 to 100%; H2S - 0.1 to 25% Channel two
(FID): C6+ composite - 0.001 to 100%
Application note reference number with technical details: D-6622, D-6630, D-6651, D-6652
Typical Chromatogram
Model 3012
Model 3013
Model 3062
Simulated Distillation Software Model 3012
Software Description
Published methods met: ASTM D2887, ASTM D3710, ASTM D5307
Types of sample analyzed: Liquid
Compounds analyzed: C4 through C110, depending upon method
Application note reference number with technical details: D-6631
130
120
110
100
90
80
70
60
50
40
30
20
10
49
Simulated Distillation Software Model 3013
Analyzer Description
Published methods met: ASTM D2887, ASTM D3710, ASTM D5307
Types of sample analyzed: Liquid
Compounds analyzed: C4 through C110, depending upon method
Application note reference number with technical details: D-6631
130
120
110
100
90
80
70
60
50
40
30
20
10
50
Software Enhancements Model 3062
Software Description
Published methods met: ASTM D3588
Types of sample analyzed: All
Application note reference number with technical details: D-6632
Compressibility: 0.9977
Heat (Saturated): 1045.63
Heat (Dry): 1064.15
Specific Gravity (R 0.6484
Channel:B
Raw Data File: c:\userprog\tc6demo\nga\pack.raw
Result File: c:\userprog\tc6demo\nga\pack.rst
Proc Method: C:\USERPROG\tc6DEMO\NGA\PACKED.mth
Calib Method: C:\USERPROG\tc6DEMO\NGA\PACKED.mth
Channel:A
Raw Data File: c:\userprog\tc6demo\nga\cap.raw
Result File: c:\userprog\tc6demo\nga\cap.rst
Proc Method: C:\USERPROG\tc6DEMO\NGA\CAP.mth
Calib Method: C:\USERPROG\tc6DEMO\NGA\CAP.mth
A/B Component Mol % Mol Fraction Mol FractioSpecific GrCalorific Compressibilty Rleative Density Calorific Value Compressibility
Name Bridged Normalizedfraction Fraction Fraction Btu/ft3 sqr rt (b)
Model 4025
Model 4026
Model 4027
Model 4028
Model 4029
Model 4227
Model 4229
Model 4425
Model 4428
Model 4429
Model 4629
Trace Sulfur Components Analyzer
Model 4025
Analyzer Description
Published methods met: ASTM D5504, ASTM D5623, ASTM D6021
Types of sample analyzed: Gas
Compounds separated: H2S, SO2, COS, mercaptans, aromatic sulfur compounds, sulfides
Guaranteed detection levels: less than 10 ppb
Application note reference number with technical details: D-6643
• Can be modified to analyze liquid and pressurized liquid samples (Model 4028)
Typical Chromatogram
COS
SO2
H2 S
Propylmercaptan
Methylmercaptan
t-Butylmercaptan
Ethylmercaptan
Dimethylsulfide
53
Trace Sulfur Components Analyzer
Model 4026
Analyzer Description
Published methods met: ASTM D5504, ASTM D5623, ASTM D6021
Types of sample analyzed: Liquid, pressurized liquid (valve injection)
Compounds separated: H2S, SO2, COS, mercaptans, aromatic sulfur compounds, sulfides
Guaranteed detection levels: less than 10 ppb
Application note reference number with technical details: D-6643
• The SCD has an equimolar response to sulfur compounds, lack of carbon interference
and requires simple calibration
Typical Chromatogram
COS
SO2
H2 S
Propylmercaptan
Methylmercaptan
t-Butylmercaptan
Ethylmercaptan
Dimethylsulfide
54
Trace Sulfur Components Analyzer
Model 4027
Analyzer Description
Published methods met: ASTM D5504, ASTM D5623, ASTM D6021
Types of sample analyzed: Liquid (syringe injection)
Compounds separated: H2S, SO2, COS, mercaptans, aromatic sulfur compounds, sulfides
Guaranteed detection levels: less than 10 ppb
Application note reference number with technical details: D-6643
Typical Chromatogram
COS
SO2
H2 S
Propylmercaptan
Methylmercaptan
t-Butylmercaptan
Ethylmercaptan
Dimethylsulfide
55
Trace Sulfur Components Analyzer
Model 4028
Analyzer Description
Published methods met: ASTM D5504, ASTM D5623, ASTM D6021
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: H2S, SO2, COS, mercaptans, aromatic sulfur compounds, sulfides
Guaranteed detection levels: less than 10 ppb
Application note reference number with technical details: D-6643
Typical Chromatogram
COS
SO2
H2 S
Propylmercaptan
Methylmercaptan
t-Butylmercaptan
Ethylmercaptan
Dimethylsulfide
56
Trace Sulfur Components Analyzer
Model 4029
Analyzer Description
Published methods met: ASTM D5504, ASTM D5623, ASTM D6021
Types of sample analyzed: Gas, liquid
Compounds separated: H2S, SO2, COS, mercaptans, aromatic sulfur compounds, sulfides
Guaranteed detection levels: less than 10 ppb
Application note reference number with technical details: D-6643
Propylmercaptan
Methylmercaptan
t-Butylmercaptan
Ethylmercaptan
Dimethylsulfide
57
Trace Sulfur Components Analyzer
Model 4227
Analyzer Description
Published methods met: ASTM D5504, ASTM D5623, ASTM D6021
Types of sample analyzed: Liquid (syringe injection), liquid and pressurized liquid (valve
injection
Compounds separated: H2S, SO2, COS, mercaptans, aromatic sulfur compounds, sulfides
Guaranteed detection levels: less than 10 ppb
Application note reference number with technical details: D-6643
Typical Chromatogram
COS
SO2
H2S
Propylmercaptan
Methylmercaptan
t-Butylmercaptan
Ethylmercaptan
Dimethylsulfide
58
Trace Sulfur Components Analyzer
Model 4229
Analyzer Description
Published methods met: ASTM D5504, ASTM D5623, ASTM D6021
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: H2S, SO2, COS, mercaptans, aromatic sulfur compounds, sulfides
Guaranteed detection levels: less than 10 ppb
Application note reference number with technical details: D-6643, D-6652
Typical Chromatogram
COS
SO2
H2S
Propylmercaptan
Methylmercaptan
t-Butylmercaptan
Ethylmercaptan
Dimethylsulfide
59
Trace Sulfur Components Analyzer
Models 4425, 4428, 4429 and 4629
Analyzer Description
Published methods met: ASTM D5504, ASTM D5623, ASTM D6021
Types of sample analyzed: Gas
Compounds separated: H2S, SO2, COS, mercaptans, aromatic sulfur compounds, sulfides
Guaranteed detection levels: less than 10 ppb
Application note reference number with technical details: D-6643
• Manual valve is used to select the calibration chamber or the sample input
• Permeation tube shipped with the product is tested with the analyzer, against an NBS
traceable calibrator for accuracy
Chamber Schematic
60
Other Analyzers
Other Analyzers
Analyzers for Light Gas, Chlorine, Transformer Oil Gas,
LPG and Simulated Distillation through Detailed
Hydrocarbon Analyses
• Fully automated
• Available with thermal conductivity detector (Model 4001) or flame ionization detector
(Model 4002)
Typical Chromatogram
62
Transformer Oil Gas Analyzer
Model 4003/4081
Analyzer Description
Published methods met: ASTM D3612-01
Types of sample analyzed: Gas
Compounds separated: H2, O2, N2, CO2, CO, CH4, C2H6, C2H4, C2H2, C3 and C4 hydrocarbons
Guaranteed detection ranges/concentration levels: CO2, CO, CH4, C2H6, C2H4 – 1.0 ppm; C2H2 -
2.0 ppm; C3s, C4s – 3.0 ppm; H2 – 50 ppm; O2, N2 – 500 ppm
Application note reference number with technical details: D-6634
• Injection port and gas sampling valve for various sample introductions
Methane
Carbon Monoxide
Acetylene
Oxygen
Hydrogen
63
Standard Method (ASTM D4815/D5580)
Analyzer Model 4004
Analyzer Description
Published methods met: ASTM D4815, ASTM D5580
Types of sample analyzed: Finished gasoline
Compounds separated: (ASTM D4815): Methanol, ethanol, isopropanol, tert-butanol, n-
propanol, MTBE, sec-butanol, DIPE, isobutanol, ETBE, tert-pentanol, n-butanol, benzene,
TAME, heavy hydrocarbons; (ASTM D5580): Benzene, toluene, ethylbenzene, p/m-xylene,
o-xylene, C9 and heavier aromatics, total aromatics
Guaranteed detection ranges/concentration levels: (ASTM D4815): MTBE, DIPE, ETBE, TAME –
0.1 to 20%; all other components – 0.1 to 12%; (ASTM D5580): (all v%) Benzene – 0.1 to 5%,
toluene – 1 to 15%, individual C8 aromatics – 0.5 to 10%, total C9 and heavier aromatics – 5
to 30%, total aromatics – 10 to 80%
Application note reference number with technical details: D-6633, D-6635
• ASTM D5580 requires a flame ionization detector (FID). For this reason and since the
same column system is utilized for both methods, only the FID performing ASTM
D4815 is appropriate for this analyzer
Typical Chromatograms
64
Standard Method (ASTM D5580) Analyzer
Model 4005
Analyzer Description
Published methods met: ASTM D5580
Types of sample analyzed: Finished Gasoline
Compounds separated: Benzene, toluene, ethylbenzene, p/m-xylene, o-xylene, C9 and heavier
aromatics, total aromatics
Guaranteed detection ranges/concentration levels: (all v%) Benzene – 0.1 to 5%, toluene – 1 to
15%, individual C8 aromatics – 0.5 to 10%, total C9 and heavier aromatics – 5 to 30%, total
aromatics – 10 to 80%
Application note reference number with technical details: D-6635
Typical Chromatogram
Analysis 1 Analysis 2
65
Standard Method (ASTM D4420) Analyzer
Model 4006
Analyzer Description
Published methods met: ASTM D4420
Types of sample analyzed: Finished Gasoline
Compounds separated: Benzene; toluene; C8, C9 and heavier aromatics, total aromatics
Guaranteed detection ranges/concentration levels: Benzene - 0.1 to 5V%, toluene – 1 to 20V%,
C8 aromatics – 3 to 25V%, C9 and heavier aromatics – 5 to 30V%, total aromatics – 10 to
80V%
Application note reference number with technical details: D-6636
• This single channel analyzer can have a second channel added, performing another
method or application. The column oven temperature requirement of the second
channel cannot operate above the maximum temperature allowed by Model 4006
(200 oC).
Typical Chromatogram
66
Standard Method (ASTM D4815/D4420)
Analyzer Model 4007/4008
Analyzer Description
Published methods met: ASTM D4815, ASTM D4420
Types of sample analyzed: Finished gasoline
Compounds separated: (ASTM D4815): Methanol, ethanol, isopropanol, tert-butanol, n-
propanol, MTBE, sec-butanol, DIPE, isobutanol, ETBE, tert-pentanol, n-butanol, benzene,
TAME, heavy hydrocarbons; (ASTM D4420): Benzene; toluene; C8, C9 and heavier
aromatics, total aromatics
Guaranteed detection ranges/concentration levels: (ASTM D4815): MTBE, DIPE, ETBE, TAME –
0.1 to 20%; all other components – 0.1 to 12%; (ASTM D4420): Benzene - 0.1 to 5V%, toluene
– 1 to 20V%, C8 aromatics – 3 to 25V%, C9 and heavier aromatics – 5 to 30V%, total
aromatics – 10 to 80V%
Application note reference number with technical details: D-6633, D-6636
• Analyzer has the capability to determine the ether and alcohol additives (see Model
4001/4002), as well as the benzene, toluene, ethylbenzene, C8 and other heavier
aromatics, as required, (see Model 4006) in gasoline
Typical Chromatograms
67
Standard Method (ASTM D4815/D3606)
Analyzer Model 4009/4010
Analyzer Description
• Analyzer has the capability to determine the ether and alcohol additives (see Model
4001/4002), as well as the benzene and toluene, as required, (see Model 4013) in
gasoline
Typical Chromatograms
68
Standard Method (ASTM D4815/D3606-
EPA modified) Analyzer Model 4011/4012
Analyzer Description
• EPA modification of ASTM D3606 adds additional column for enhanced benzene –
ethanol (if present) separation. All other specifications and method intent remain
unchanged.
Typical Chromatograms
69
Standard Method (ASTM D3606) Analyzer
Model 4013
Analyzer Description
Published methods met: ASTM D3606
Types of sample analyzed: Finished Motor and Aviation Gasoline
Compounds separated: Benzene, toluene
Guaranteed detection ranges/concentration levels: Benzene – 0.1 to 5V%, toluene – 2 to 20V%
Application note reference number with technical details: D-6637
• This single-channel analyzer can have a second channel added, performing another
method or application. The column oven temperature requirement of the second
channel cannot operate above the maximum temperature allowed by Model 4013
(120 oC).
Typical Chromatogram
70
Standard Method (ASTM D3606-EPA
modified) Analyzer Model 4014
Analyzer Description
Published methods met: ASTM D3606 (EPA modification)
Types of sample analyzed: Finished Motor and Aviation Gasoline
Compounds separated: Benzene, toluene
Guaranteed detection ranges/concentration levels: Benzene – 0.1 to 5V%, toluene – 2 to 20V%
Application note reference number with technical details: D-6637
• EPA modification of ASTM D3606 adds additional column for enhanced benzene –
ethanol (if present) separation
• All other specifications and method intent remain unchanged. See Model 4013 for
details
• This single channel analyzer can have a second channel added, performing another
method or application. The column oven temperature requirement of the second
channel cannot operate above the maximum temperature allowed by Model 4014
(120 oC).
Typical Chromatogram
71
Detailed Hydrocarbon Analyzer
Model 4015
Analyzer Description
Published methods met: ASTM D5443, ASTM D6623, ASTM D6730, ASTM D6733
Types of sample analyzed: liquid
Compounds separated: Paraffins, Isoparaffins, Naphthenes, Olefins, Aromatics and
Oxygenates (reported by carbon number) through C13
Guaranteed detection ranges/concentration levels: All components – 10 ppm
Application note reference number with technical details: D-6638
Summary Report
Oxygenates:
Total: 15.471 (mass%) 15.385 (vol%)
Total Oxygen Content: 2.808 (mass%)
• Typical analyses are light gases, furnace gases, stack gases, combustion gases,
decoking gases, mine gases
Typical Chromatogram
73
Light Gas Analyzer Model 4017
Analyzer Description
Published methods met: ASTM D1946
Types of sample analyzed: Gas
Compounds separated: He; H2; Ar - O2 composite; N2; CH4; CO; CO2; Ethylene; Ethane;
Acetylene; H2S
Guaranteed detection ranges/concentration levels: H2S – 0.02 to 100%, all other components –
0.01 to 100%
• Typical analyses are light gases, furnace gases, stack gases, combustion gases,
decoking gases, mine gases
• Full-range hydrogen detection using separate channel with nitrogen carrier gas added
to Model 4016
Typical Chromatogram
Nitrogen
Hydrogen
Oxygen
H2S
Ethylene
Acetylene
CO2
Ethane
Methane
CO
74
Light Gas Analyzer Model 4018
Analyzer Description
Published methods met: ASTM D1946
Types of sample analyzed: Gas
Compounds separated: Ar; O2; N2; CH4; CO; CO2; Ethylene; Ethane; Acetylene; H2S
Guaranteed detection ranges/concentration levels: H2S – 0.02 to 100%, all other components –
0.01 to 100%
Application note reference number with technical details: D-6639
• Typical analyses are light gases, furnace gases, stack gases, combustion gases,
decoking gases, mine gases
Typical Chromatograms
Nitrogen
Oxygen
13 Minutes
Argon
75
Light Gas Analyzer Model 4019
Analyzer Description
Published methods met: ASTM D1946
Types of sample analyzed: Gas
Compounds separated: He; H2; Ar; O2; N2; CH4; CO; CO2; Ethylene; Ethane; Acetylene; H2S
Guaranteed detection ranges/concentration levels: H2S – 0.02 to 100%, all other components –
0.01 to 100%
Application note reference number with technical details: D-6639
• Typical analyses are light gases, furnace gases, stack gases, combustion gases,
decoking gases, mine gases
Typical Chromatograms
Nitrogen
Nitrogen
Hydrogen
Oxygen
Oxygen
13 Minutes
H2S
Ethylene
Acetylene
CO2
Ethane
Argon
Methane
CO
76
Light Gases in Chlorine Model 4020
Analyzer Description
Published methods met: ASTM E1746
Types of sample analyzed: Gas
Compounds separated: He, H2, O2, N2, CH4, CO2, CO
Guaranteed detection ranges/concentration levels: He, H2 - 0.001 to 100%; all other components
- .01 to 100%
Application note reference number with technical details: D-6640
• An important feature is the use of proprietary packing materials as the first column in
both channels
• Samples containing water do not interfere with the analysis or create hardware
problems
Typical Chromatogram
77
Trace CO, CH4 and CO2 Analyzer
Model 4021
Analyzer Description
Published methods met: ASTM D3416, ISO 6381
Types of sample analyzed: Gas
Compounds separated: CO, CO2, CH4, ethylene – ethane composite, acetylene
Guaranteed detection ranges/concentration levels: All components – 0.1 ppm
Application note reference number with technical details: D-6641
• Acetylene elutes after ethylene – trace acetylene can not be analyzed in large
quantities of ethylene
• Purged valve housings and passivated tubing are used to enhance sensitivity
Typical Chromatogram
Carbon Monoxide
Methane
Carbon Dioxide
8 Minutes
78
Trace CO, CH4 and CO2 Analyzer
Model 4022
Analyzer Description
Published methods met: ASTM D3416, ISO 6381
Types of sample analyzed: Gas
Compounds separated: CO, CO2, CH4, ethylene, ethane, acetylene
Guaranteed detection ranges/concentration levels: All components – 0.1 ppm
Application note reference number with technical details: D-6641
• Acetylene elutes before ethylene – trace acetylene, CO, CO2, CH4 can be analyzed in
large quantities of ethylene
• Purged valve housings and passivated tubing are used to enhance sensitivity
Typical Chromatogram
Carbon Dioxide
Carbon Monoxide
Methane
Ethylene
Ethane
Acetylene
4 Minutes
79
Liquid Sample Expansion Apparatus
Model 4023
Analyzer Description
Published methods met: ASTM D2712
Types of sample analyzed: Pressurized liquid
Compounds separated: N/A
Guaranteed detection ranges/concentration levels: N/A
Application note reference number with technical details: D-6642
Product Schematic
80
Trace CO, CH4 and CO2 Analyzer
Model 4024
Analyzer Description
Published methods met: UOP 603
Types of sample analyzed: Gas
Compounds separated: CO, CO2, CH4
Guaranteed detection ranges/concentration levels: All components – 0.5 ppm
Application note reference number with technical details: D-6641
• Does not include any C2 analysis (Acetylene and ethylene coelute off of the primary
column)
• Purged valve housings and passivated tubing are used to enhance sensitivity
Typical Chromatogram
Carbon Monoxide
Carbon Dinoxide
Methane
81
Full Range H2, He, O2 and N2 Analysis
Model 4032/4033/4034
Analyzer Description
Published methods met: ASTM D1945, ASTM D1946
Types of sample analyzed: Gas
Compounds separated: Model 4032- He, H2; Model 4033- O2, N2; Model 4034- He, H2, O2, N2
Guaranteed detection ranges/concentration levels: All components – 10 ppm to 100%
Application note reference number with technical details: D-6645
• Models 4032 and 4033 are single channel, thermal conductivity detector analyzers.
Model 4034 is a dual-channel, dual-thermal conductivity detector analyzer.
• Better detection and measurement than 10 ppm can be achieved, with modification, in
some cases. Some limitations are present, however, with certain combinations of
compound levels present. Information on the complete sample matrix is required to
determine true detection and measurement ranges.
• Model 4032 and 4033 have the ability to be combined with other application channels
for cost savings
Typical Chromatogram
Model 4032 and Model 4034 analysis one Model 4033 and Model 4034 analysis two
82
Light Hydrocarbon Analyzer Model 4035
Analyzer Description
Published methods met: ASTM D2163, ASTM D2593, ASTM D2712, ASTM D2820, ASTM
D4424, ISO 7941
Types of sample analyzed: Gas
Compounds separated: All hydrocarbons from methane through hexane
Guaranteed detection ranges/concentration levels: All components - 0.1 ppm
Application note reference number with technical details: D-6663
Typical Chromatogram
83
Light Hydrocarbon Analyzer Model 4036
Analyzer Description
Published methods met: ASTM D2163, ASTM D2593, ASTM D2712, ASTM D2820, ASTM
D4424, ISO 7941
Types of sample analyzed: Liquid, pressurized liquid
Compounds separated: All hydrocarbons from methane through hexane
Guaranteed detection ranges/concentration levels: All components - 0.1 ppm
Application note reference number with technical details: D-6663
Typical Chromatogram
84
Light Hydrocarbon Analyzer Model 4037
Analyzer Description
Published methods met: ASTM D2163, ASTM D2593, ASTM D2712, ASTM D2820, ASTM
D4424, ISO 7941
Types of sample analyzed: Gas, liquid, pressurized liquid
Compounds separated: All hydrocarbons from methane through hexane
Guaranteed detection ranges/concentration levels: All components - 0.1 ppm
Application note reference number with technical details: D-6663
Typical Chromatogram
85
Benzene and Oxygenates Analyzer
Model 4080
Analyzer Description
Published methods met: EN 12177, EN 13132, IP 425/96
Types of sample analyzed: Unleaded petrol (gasoline) with a boiling point not to exceed 220 oC
Compounds separated: Benzene, aromatics, MTBE, ethanol, oxygenates
Guaranteed detection ranges/concentration levels: Benzene - 0.05 to 6 % (V/V), individual organic
oxygenate compounds - 0.17% to 15 % (m/m), total organically bound oxygen - up to 3.7%
(m/m)
Application note reference number with technical details: D-6647
• The analyzer features a capillary injection port, PreVent and three capillary columns (a
liquid-sampling valve can also be used)
• Total flexibility - Choosing the correct internal standard and varying the “heartcut”
times, it is possible to determine any oxygenate or aromatic found in petrol.
• The same sample can be run twice with different “heartcut” times. By spreading the
components into two analyses, the interferences found with a large number of
components are eliminated.
Typical Chromatogram
Toluene
Measured MTBE,
Benzene
benzene
86
Liquid Sampling Apparatus Model 4082
Analyzer Description
Published methods met: UOP 911
Types of sample analyzed: Pressurized liquid
Compounds separated: N/A
Guaranteed detection ranges/concentration levels: N/A
Application note reference number with technical details: None
• This product is normally sold as an “add on” to liquid sampling valved analyzers
• OPERATION OF THE MODEL 4082: Connect the filtered sample to the sample inlet and
start the liquid flow. There may be bubbles passing through the sight glass. Gently
close the valve at the top of the sight glass until the bubbles stop. This will provide a
slight backpressure in order to prevent the vaporizing of the liquid sample. Upon
minimizing the vaporization (bubbles) of the sample and obtaining a desired liquid flow
rate, close the toggle valve to stop the flow and make your injection. Once the run has
commenced, open the toggle valve and the backpressure valve to vent the remaining
liquid.
Schematic of Mounted Model 4082
87
Trace H2, O2, Ar, N2, CH4 and CO in Light
Hydrocarbons Analyzer Model 4083
Analyzer Description
Published methods met: ASTM D2504
Types of sample analyzed: Gas
Compounds separated: Hydrogen, oxygen /argon coelution, nitrogen, methane, carbon
monoxide
Guaranteed detection ranges/concentration levels: CO – 100 ppb, all other components – 50
ppb
Application note reference number with technical details: D-6648
Typical Chromatogram
88
Trace Methanol and MTBE in Liquid
Hydrocarbons Analyzer Model 4086
Analyzer Description
Published methods met: None
Types of sample analyzed: Liquid
Compounds separated: Acetaldehyde, MTBE, methanol, acetone, 2-propanol, others
Guaranteed detection ranges/concentration levels: Acetaldehyde, MTBE, methanol – 0.5 V ppm
to 1V%; all other components – 1 V ppm to 1V%
Application note reference number with technical details: None
• Split/splitless injector, dual capillary column, PreVent and flame ionization detector
(FID)
• An autosampler should be included for precision
• Optional gas-sampling or liquid-sampling valves can be added
• Capillary non-polar precolumn for separation of light non-polar compounds and
backflushing heavies. The use of a precolumn and backflushing keeps high boiling
material from contaminating the main column. Backflushing of the precolumn is
provided by PreVent hardware and a mid-point pressure switch time programmed from
the Clarus Interface or TotalChrom.
• Capillary PLOT silica main column for separation of oxygenated compounds
Typical Chromatogram
89
Related Industries Also Benefit
The dedicated and guaranteed PerkinElmer Arnel Analyzers and Systems in this booklet specifically address
the petrochemical laboratory and process industry. Many other standard and custom Analyzers, Systems and
Accessories are available for other industries such as pharmaceutical, food & beverage, and specialty gases
for QC, general testing and environmental - occupational safety monitoring applications.
Explore both the PerkinElmer web site (www.perkinelmer.com) and the Arnel Inc. web site (www.arnelinc.com)
for other products, plus contact your local PerkinElmer sales engineer or email Arnel at info@arnelinc.com for
more details, literature and a quotation.
90
PerkinElmer Life and
Analytical Sciences
710 Bridgeport Avenue
Shelton, CT 06484-4794 USA
Phone: 800-762-4000 or
(+1) 203-925-4602
www.perkinelmer.com
© 2003 PerkinElmer, Inc. All rights reserved. PerkinElmer is a registered trademark of PerkinElmer, Inc. Arnel is a trademark of Arnel Inc.
PerkinElmer reserves the right to change this document at any time and disclaims liability for editorial, pictorial or typographical errors.