Vous êtes sur la page 1sur 2

User’s Manual for Xentaur Dewpoint Generator

Blended Gas Fill port with Optional Saturator Gas Dry Gas
Outlet Vent Valve Non-Dry Gas source Inlet
Inlet selection valve

Saturated gas Dry Gas Coalescing Saturator Dry Gas Particulate


flow meters flow meter Filter Regulator Filter
Specifications for XENTAUR Gas Dilution Dewpoint Generator

Dewpoint Range: Low: within 1% of inlet gas water concentration


(e.g. 10ppm inlet gas produces 11ppm outlet gas)
High: typically 10°C below ambient temperature, can be placed in a
heated enclosure for higher dewpoint measurements.

Repeatability: +/- 2°C dp, in the range of -65°C to 0°C dp

Inlet Gas: Dewpoint: not to exceed 1% of desired low water concentration


Pressure: Regulate to 50 psig maximum
Flow rate: 20 LPM maximum

Inlet & Outlet connections: ¼” Swagelok tube fittings.

Materials: All materials exposed to dry gas are: stainless steel, borosilicate glass tubes,
Viton ‘O’ rings. Materials exposed to non-dry gas have in addition: brass, aluminum
alloy, polycarbonate.

Principle of Operation: The flow meters are adjusted to produce a blend of dry and
saturated gas, to achieve the desired dewpoint.

Check
Regulator Valve
Optional Flow Meters w/Needle Valves
Non-Dried
DRY Gas Inlet BLENDED
3-Way
GAS GAS
Valve
INLET OUTLET
500 ml/min
50 ml/min

10 L/min

15 L/min

Particulate
Filter
Coiled
Tubing
~3.5m Vent Valve
W1 W2 W3 DRY
Pressure
Gauge
Fill
Line

Coalescing
Filter
Saturator

Flow Diagram of XENTAUR Gas Dilution Dewpoint Generator