HP-PLOT Q Columns
HP-PLOT Q columns feature baseline separation starting at 60_C for alkane hydrocarbons up to C6; all C1-C3 isomers; air/CO, water, CO2; and other polar compounds resulting in up to a 30% reduction in cycle time. These columns also pro vide excellent stability with little or no detector spiking, low bleed for decreased condi tioning times and improved sensitivity, outstanding repro ducibility. These columns also display long life: they have been demonstrated to with stand 1800 injections of alco hol samples. Because selectivity is main tained with HP-PLOT Q versus other commercially available DBV phases, transferring methods to HP-PLOT Q is easy. HP-PLOT Q columns have a maximum isothermal operating temperature of 270_C and are excellent for valved and GC/MS analyses.
Resolution of air/CO, CO2, and methane in a natural gas sample (C7 concentration is approximately 200 ppm)
Column: Methane N2-Air CO2 Ethane Carrier: Oven: Detector: Injection: Sample: H2O 0.53 mm x 30 m, HP-PLOT Q (Agilent part no. 19095P-QO4) Helium, 60 cm/sec, constant flow 60C (2 min), 30C/min to 240C (1 min) TCD 250C 250C Split mode (18:1) 0.25 cc natural gas sample, Methane, 80% C3
i-C4/n-C4
H2S
n-C5 C6 C7
8 minutes