Académique Documents
Professionnel Documents
Culture Documents
The new configurable system EMAI, model MX-300 presents the ultimate innovations in cardiologic sector. With digital technology and powerful microprocessors that provide high quality and optimal resolution parameters, signs and records, it offers the better relationship cost/benefit of the market.
ECG
Gain automatic control. Wave detection by smart software. Recognition and validation of pacemaker's pulse. Allows the use of patient's cable of 3 and 5 ducts. Visualization of seven derivations Allows the interconnection with central and defibrillator. Operation by the mode Filter and Diagnostic. Measure ranges 0 to 300 bpm. Accuracy 1 bpm or 2% (wichever is greater).
GENERAL SPECIFICATIONS
Wave forms freezing of ECG and SpO2. Monitoring in modes adult and neonatal. Digital adjustment of display's contrast . Digital control of bip's tone of pulse and alarm. Speed adjustments of tracing to 25 or 50 mm/s. On-line helps in screen with 41 pages. Battery charging indication, low battery, energy network or only battery. Automatic turning-off in case of minimum limit of battery charge.
SpO2 (Optional)
Spectrophotometry and plethysmographic techniques. Shows the measure values of SpO2 and pulse. Plethysmographic curve. Indicates the quality of sign, searching pulse and weak pulse Measures range SpO: 0 to100% 2 Range SpO2: 2% 70 to 100%. Measures range Pulse: 30 to 240 bpm. Accuracy Pulse: 3% or 2% (wichever is greater).
ALARMS
Situation alarms for oximetry probe absences, of patient cable, asystolia, of maximum and minimum limits for all the parameters. Sounding indication in three different tones and digital adjustable tones. Visual indication. Differentiated indication to asystolia.
CHART OF TENDENCIES
Tendency for variation of heart frequency and SpO2 in the last 72 hours. Visualization in adjustable intervals of 30 min., 1 h, 2 h, 3 h, 6 h, 12 h, 24 h, 48 h, and 72 h. Automatic scale of 0-max. or min-max.
PNI (Optional)
Non-invasive pressure measures by oscillometric method. Systolic, diastolic and medium pressure. It captures and tabulates the last valid measurements, with date and hour registration. Accomplishes measurements in manual or automatic mode, and this with adjustable intervals between 2 and 99 minutes. Range measures: Systolic: 60 to 250 mmHg Diastolic: 45 to 235 mmHg Media: 40 to 220 mmHg Accuracy: equal or superior to Norm SP10-1992 AAMI
PRINTER OUTPUT
Specifications subjected to changes without previous notice.
To values printing of ECG, SpO2, PIN and alarm conditions. Two printing modes, report and continuous. Compatible with matrix printer.
POWER REQUIREMENTS
Energy network 110 to 220 Vac 50/60Hz. External font: 12 Vdc (stabilized). Battery: plumb-acid, 12 Vdc, internal and automatic recharge, use near 8h to ECG, when totally charged. Recharge time: 16 hours when totally discharged.
MONITOR'S CONFIGURATION
* MX-300-A: ECG * MX-300-B: ECG + SpO2 * MX-300-C: ECG+PNI * MX-300-D: ECG + SpO2 + PNI
DIMENSIONS
Height ................................................................. 14,5 cm Width .................................................................. 22,8 cm Depth .................................................................. 21,3 cm Weight .................................................................... 4,0 kg
Av. Maria Estela, 33 - Jardim Maria Estela Phone: (55-011) 6335-1000 Fax: extension line 210 ZIP 04180-010 So Paulo Brazil www.transmai.com.br - e-mail: transmai@transmai.com.br
Registered in ANVISA AM 80052640004