Planning Guide
COROSKOP Plus hier eintragen Titel der Unterlage (+ KOORDINAT M)
0-2
Chapter all Page all Revision 01
Revision
COROSKOP Plus
Page 2 of 4 TD SD 31
Contents
0-3
Page
3 _______Installation Preparation__________________________________________ 3 - 1
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . COROSKOP Hi-P . . . . . . . . . . . . . . . . . . . . . . . KOORDINAT M . . . . . . . . . . . . . . . . . . . . . . . . MTS-A3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . Control Room Distributor . . . . . . . . . . . . . . . . . . . VideoRec or Video Recorder + Monitor M44-2M (Option). . . M102 with Video Distributor Amplifier (Option) . . . . . . . . Reference Monitor (Option) . . . . . . . . . . . . . . . . . . Control Cabinets . . . . . . . . . . . . . . . . . . . . . . . . Cooling System . . . . . . . . . . . . . . . . . . . . . . . . Upper Body Radiation Shield, Surgery Light, Injector (Option) Lower Body Radiation Shield (Option). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1 .3-1 .3-1 .3-2 .3-2 .3-2 .3-2 .3-2 .3-2 .3-2 .3-3 .3-3
Page 3 of 6 TD SD 31
COROSKOP Plus
0-4
Contents
Page Second Display (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3 On-Floor-mounting with Small Mounting Plate (not for USA) . . . . . . . . . . . . . 3 - 4 COROSKOP Hi-P/KOORDINAT M . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4 COROSKOP Hi-P: Dimensional Drawing, "Small" Mounting Plate (not USA) . . . 3 - 5 KOORDINAT M: Table Base Dimensional Drawing . . . . . . . . . . . . . . . . 3 - 6 On-Floor-mounting with Large Mounting Plate (for USA only) . . . . . . . . . . . . . 3 - 7 COROSKOP Hi-P/KOORDINAT M . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 7 COROSKOP Hi-P: Dimensional Drawing "Large" Mounting Plate (USA only) . . . 3 - 8 Floor-mounting in Self-leveling Grout with Mounting Frame . . . . . . . . . . . . . . 3 - 9 COROSKOP Hi-P/KOORDINAT M . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 9 COROSKOP Hi-P: Dimensional Drawing, Mounting Frame . . . . . . . . . . . .3 - 10 KOORDINAT M: Mounting Frame Dimensional Drawing . . . . . . . . . . . . . .3 - 12 System Ceiling Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 14 MTS Version 1 with Fixed-position Upper Body Radiation Shield (Option). . . . .3 - 14 MTS Version 1 with Moveable Upper Body Radiation Shield (Option) . . . . . . .3 - 15 In-house Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 16 Remarks regarding the recommendation on Page 3-17 for the room electrical installation per DIN VDE 0107 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 16 Recommendation for Application Group 2 . . . . . . . . . . . . . . . . . . . . .3 - 17 Power Cables by Customers Contractor . . . . . . . . . . . . . . . . . . . . . .3 - 18
COROSKOP Plus
Page 4 of 6 TD SD 31
Contents
0-5
Page
8 _______Detail Planning_________________________________________________ 8 - 1
Power Unit 1 (POLYDOROS IS-C) . . . . . . . . . . . . . . . . . . System Control Cabinets (Double Cabinet) . . . . . . . . . . . . . . HICOR T.O.P. Cabinet. . . . . . . . . . . . . . . . . . . . . . . . . Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . Recommendation for Cabinet Setup. . . . . . . . . . . . . . . . COROSKOP Hi-P Ceiling Outlet . . . . . . . . . . . . . . . . . . . MTS-A3 Positioning per Version 2 . . . . . . . . . . . . . . . . . . System: Top-down View . . . . . . . . . . . . . . . . . . . . . . Ceiling Installation . . . . . . . . . . . . . . . . . . . . . . . . . Installation Recommendation for MTS-A3 Longitudinal Rails . . . N465 System Control Console . . . . . . . . . . . . . . . . . . . . Table Installation. . . . . . . . . . . . . . . . . . . . . . . . . . Wall Installation . . . . . . . . . . . . . . . . . . . . . . . . . . Control Room Distributor . . . . . . . . . . . . . . . . . . . . . . . Control Console (Option)/Container (Option) . . . . . . . . . . . . . 110 V Power Pack (Option) . . . . . . . . . . . . . . . . . . . . . . M102 with Video Distributor Amplifier (Option) . . . . . . . . . . . . Fixed Mounted Upper Body Radiation Shield, Surgery Light (Option) Moveable Upper Body Radiation Shield, Surgery Light (Option) . . . Injector Wall Terminal Box (Option) . . . . . . . . . . . . . . . . . . Shown without Injector Panel . . . . . . . . . . . . . . . . . . . Planning Recommendation for Injector Wall Connection . . . . . Injector Rack Mount Unit (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-1 .8-3 .8-5 .8-5 .8-6 .8-7 .8-8 .8-8 .8-9 8 - 10 8 - 11 8 - 11 8 - 11 8 - 12 8 - 13 8 - 14 8 - 15 8 - 18 8 - 19 8 - 20 8 - 20 8 - 21 8 - 22
Page 5 of 6 TD SD 31
COROSKOP Plus
0-6
Contents
COROSKOP Plus
Page 6 of 6 TD SD 31
General Remarks
System Configuration
1-1
1
This Planning Guide (PG) has been prepared for the COROSKOP Plus System. The COROSKOP Plus System is comprised of the following system components: COROSKOP Hi-P + KOORDINAT M + MTS-A3 + HICOR T.O.P.
Remarks A note must be made on all room plans that are created by the Planning Departments
which makes reference to the General Installation and Delivery Conditions of Siemens Medical Engineering. A copy of the General Installation and Delivery Conditions must accompany the plans when they are turned over to the customer.
For required spaces, the documents of the application fire safety regulations must be
observed.
The system was developed in accordance with IEC 601-1 (corresponds to DIN VDE
0750, Part 1 and EN 60601-1).
Page 1 of 4 TD SD 31
COROSKOP Plus
1-2
Definition of Terms
General Remarks
1
Technically feasible area, Orientation points and so on Technically feasible area, minimum length, minimum width
The technically feasible area is the rectangle that is formed from the minimum length and minimum width as the space requirement for a system component. The following conventions apply for the technically feasible area: 1. It is the minimum area that is possible without making restrictions to unit movements necessary. 2. Unless otherwise noted, it may not be less than this area! 3. The required safety distances are included (caution: it is possible that national regulations differ from these); 4. The service areas for the unit are included; 5. The required areas for unhindered examination and operation, as well as the required installation areas for control cabinets are not taken into consideration; 6. Illustration: the technically feasible area are illustrated in the top-down views in this PG as a broken line.
Orientation points
Points specific to system components to which reference is made when positioning system components to each other or in the room. The isocenter of a radiographic system is always illustrated as the orientation point. Illustration in the drawings:
Fixpoints
Clearly marked points on system components, installation ceiling, walls or floor on which cable outlets are located. Illustration in the drawings: circle with letter/number- combination. Example: PU
The cable lengths establish the maximum fixpoint distances and thus the maximum distances between the individual system components.
Room height
The room height is the distance measured from the top surface of the finished floor to the bottom surface of the ceiling structural elements (Unistrut rails) (bottom surface of drop ceiling).
COROSKOP Plus
Page 2 of 4 TD SD 31
General Remarks
Lists of Fixpoints Used
Fixpoint CRX Subsystem Control Room X-ray (console) Remark
1-3
1
Fixpoint on the bottom of the control room distributor Floor fixpoint Floor fixpoint Ceiling fixpoint Wall fixpoint Ceiling fixpoint Ceiling fixpoint Floor fixpoint Wall fixpoint Ceiling fixpoint Floor fixpoint Floor fixpoint Floor fixpoint Ceiling fixpoint Floor fixpoint
1
Control Room E-Medicine (measuring station) (Option) Cooling Unit 1 (cooling unit for cooling X-ray tube unit) Display (MTS-A3) Second display (display of unit data + radiation indicator) Injector head mounted on injector ceiling column in radiation shield support arm system (option) Injector head permanently mounted on ceiling (option) Image system (HICOR T.O.P.) Injector wall outlet in the examination room Plane 1 (COROSKOP Hi-P) System control cabinets (double cabinet) PD: Power Distributor SD1: System Distributor 1 Power Unit 1 (POLYDOROS IS-C) ACOM M (option) Surgery Light (option) Table (KOORDINAT M)
PU1 R1 SU T
Page 3 of 4 TD SD 31
COROSKOP Plus
1-4
EPS (Emergency Power Supply) (Option)
The EPS is primarily comprised of:
1.
General Remarks
1
a UPS (Uninterruptable Power Supply) The power distributor (Fixpoint PD) is connected to the UPS. This way, if there is a failure of the line power supply, with the exception of the generator all system components (unit, imaging system, ...) will still be supplied with power. an ATS (Automatic Transfer Switch) The ATS is intended to be an expansion to the UPS. If line power fails, the generator is connected to the UPS via the ATS. The normal exposure mode will be blocked and only continuous fluoroscopy will be possible (with fully charged batteries, min. 30 minutes).
2.
Detailed information about the EPS is available from the headquarters of the Exide
Company in the particular country, e.g. from: Country USA Address Exide World Headquarters 8609 Six Forks Road Raleigh, North Carolina 27615 USA Exide Electronics Mrekestr. 7 D-63814 Mainaschaff Telephone/Fax Tel.: 001 919 870-3484 Fax: -3031
Germany
COROSKOP Plus
Page 4 of 4 TD SD 31
Room Planning
Room Plan
Recommendation
Cable cabinet System controller
2
2-1
2
Examination room
MTS-A3
Control room
Hardcopy System control console
POLYDOROS IS
Cooling unit
HICOR T.O.P.
Equipment room
ACOM T.O.P.
KOORDINAT M
Plan the control room at the foot-end of the KOORDINAT M (optimum visual contact to
the examination room).
Plan the radiation shield window in terms of size and position so that the examination
room can be clearly viewed from the control room.
Plan the entrance area as illustrated. Plan the equipment room to the left of the examination room. Air condition the equipment room. If at all possible, do not plan the POLYDOROS IS and cooling unit in the examination
room (generation of noise and heat).
Page 1 of 18 TD SD 31
COROSKOP Plus
Orientation point /Isocenter COROSKOP Hi-P in normal position Isocenter COROSKOP Hi-P in oblique position
*2
2-2
1:50
Room Planning
*1 *1 Including 500 mm safety distance between COROSKOP Hi-P and wall. *2 Dimensions not in parenthesis: Table longitudinal movement not limited (standard) Dimensions in parenthesis: Table longitudinal movement limited by the room width.
2
Room Planning
System: Position of the MTS-A3
Position the MTS-A3 different from the standard position (Version 1) only in exceptional cases.
2-3
2
Version 1 (standard position) Room plan, see Page 2-2 Ceiling installation, see Page 3-14/15 Fixed-mounted or moveable upper body radiation shield (option), see Page 8-18/19
Version 2 Room plan, see Page 8-8 Ceiling installation, see Page 8-9 Only moveable upper body radiation shield (option) allowed, see Page 8-19
1:100
Versions 3, 4 Fixed-mounted or moveable upper body radiation shield (option), see Page 8-18/19
Page 3 of 18 TD SD 31
COROSKOP Plus
2-4
System: Side View
Room Planning
2
*1 1:50 *2 *1 *2 D P1 With room heights lower/equal to 2830 mm, there is risk of collision between MTS-A3 and KOORDINAT M; risk of injury between MTS-A3 and patient! Orientation point/isocenter COROSKOP Hi-P in normal position MTS-A3 (ceiling fixpoint) COROSKOP Hi-P (ceiling fixpoint) T KOORDINAT M (floor fixpoint)
COROSKOP Plus RAA2-315.021.14 Rev. 01 10.97 Page 4 of 18 TD SD 31 Siemens AG Medical Engineering
Room Planning
Planning Information
Component Movement Ranges
2-5
2
When planning external components in the examination room, pay attention to the movement ranges of the system components (risk of collision!). MTS-A3 movement range
Page 5 of 18 TD SD 31
COROSKOP Plus
2-6
Max. Admissible Floor Unevenness
Room Planning
2
Max. admissible floor unevenness of the subfloor in the cross-hatched area is 8 mm. The cross-hatched area corresponds to the rotation area (normal/oblique position) of the stand, as well as to the C-Arm movement range.
1:20
COROSKOP Plus
Page 6 of 18 TD SD 31
Room Planning
COROSKOP Hi-P: Side View
2-7
2
1:50
Page 7 of 18 TD SD 31
COROSKOP Plus
2-8
COROSKOP Hi-P: Top-down View
Room Planning
2
Axis of rotation
1:50
COROSKOP Plus
Page 8 of 18 TD SD 31
Room Planning
KOORDINAT M: Views
View X *1 *2 *3 *4
2-9
2
*5 *6 *7
1:50
*1 *2 *3 *4
Accessory rail, e.g. used to hook on the ECG preamplifier (with CATHCOR/RECOR) Blank panel, 135x95 mm for installation of an electrical outlet (page 3-20) Panel with a nipple for compressed air connection Panel with injector connector
*5 *6 *7 T
Page 9 of 18 TD SD 31
COROSKOP Plus
2 - 10
Arrangement of Cabinets and Components
Recommendation 1 Recommendation 2 2
Room Planning
2 2
POLYDOROS IS Cabinet
Recommendation 1:
POLYDOROS IS (PU1) and cooling unit (CU1) located in equipment room (due to generation of noise and heat).
Recommendation 2:
Control cabinets located in the examination room, cooling unit (CU1) and HICOR T.O.P. (IS) in equipment room. Caution: Take note of noise and heat generated by the POLYDOROS IS (PU1)!
Air-condition the equipment room. The cabinets can be set up as illustrated. Take note of the following:
- The cabinets can be set up separated from each other or corner-to-corner only at the places marked with arrows. When this is done, take note of the cable list on Page 4-3. - If set up is corner-to-corner, maintain at least 40 mm space to the next cabinet (needed to open the cabinet doors). - PU1 and cable cabinet can also be positioned to the right of the SD1. - For additional information, see Pages 8-1 to 8-6.
COROSKOP Plus
Cable cabinet
HICOR T.O.P.
Page 10 of 18 TD SD 31
Cooling unit
Siemens AG Medical Engineering
Room Planning
Control Room
2 - 11
2
Measuring station
Plan the control room so it is at the foot-end to the examination room. The control console, container, second display (not shown; for details, see Page 8-16)
and ACOM T.O.P. can be ordered as options.
Position the components in the vicinity of the control room distributor (see Pages
8-11/12/13/14).
Page 11 of 18 TD SD 31
COROSKOP Plus
2 - 12
Room Planning
2nd monitor (44 cm or 54 cm) *1*3 N465 System Control Console 2nd unit control console *1 ECC2*1 Reference monitor (44 cm or 54 cm) *1*4
Video recorder1*2
*1 *2
Option VideoRec (= video recorder + VCR interface + scan converter) or video recorder SONY SVO 9500... + M44-2M N-Monitor (not shown). As an alternative to the SONY video recorder every commercially available VHS recorder can be used. The Power Pack is required for the video recorder and 110V/60 Hz power systems. For details, see Page 8-14.
*3 *4
The M102 is required for the connection of additional monitors (see Page 8-15). Positioned in the MTS (standard) or in the control room.
For details, as well as for additional components that can be connected to the control room distributor, see Pages 8-12.
COROSKOP Plus
Page 12 of 18 TD SD 31
Room Planning
Static Requirements
2 - 13
2
COROSKOP Hi-P, Floor-mounted with Small Mounting Plate (not for USA)
Tension occurs only at the following mounting points: F5max F6max F7max + 2.0 kN + 4.0 kN + 3.0 kN
These maximum values do not occur simultaneously. The tension depends on the operating position and on the unit movement.
Mounting holes 1-7 for installation directly on the floor Small mounting plate (722 mm dia.) Spare mounting holes 1:10
Center of gravity, unit in oblique position Orientation point/isocenter, unit in normal position Center of gravity, unit in normal position 1:20
*1 Load force (unit + small mounting plate) = 5.9 kN Range of shifting load force caused by the park position, the two operating positions, as well as the unit C-Arm movement.
Page 13 of 18 TD SD 31
COROSKOP Plus
2 - 14
Room Planning
COROSKOP Hi-P, Floor-mounted with Large Mounting Plate (for USA only) 2
Large mounting plate (860 mm dia.) Tension occurs only at the following mounting points: F5max F6max F7max + 3.5 kN + 3.5 kN + 3.5 kN
These maximum values do not occur simultaneously. The tension depends on the operating position and on the unit movement.
Mounting holes 1-7 for installation directly on the floor 1:10 Spare mounting holes
1:20 *1 Load force (unit + large mounting plate) = 6.0 kN Range of shifting load force caused by the park position, the two operating positions, as well as the unit C-Arm movement.
COROSKOP Plus
Page 14 of 18 TD SD 31
Room Planning
2 - 15
1:20
*1 Load force (unit + small mounting plate + mounting frame) = 6.7 kN Range of shifting load force caused by the park position, the two operating positions, as well as the unit C-Arm movement.
Page 15 of 18 TD SD 31
COROSKOP Plus
2 - 16
MTS-A3
Room Planning
2
Vertical ceiling load Fxmax: 4.0 kN (the dynamic load is exerted on each mounting point, because this is a floating single load). For further information, see PG RA53-020.021.02...
KOORDINAT M The max. tension is calculated with the table turned (+90o or -90o) and with the tabletop
extended completely in the head end position. The max. tension is determined with the weight of the patient (200 kg) and accessories (40 kg) factored in.
Mounting frame:
Tension at head-end and/or foot-end: per expansion bolt, max. tension 4.0 kN
COROSKOP Plus
Page 16 of 18 TD SD 31
Room Planning
KOORDINAT M Center of Gravity The center of gravity when the table is turned was determined under the following conditions: Tabletop extended completely in the head position. Tabletop loaded with a patient weight of 200 kg. Table accessories (40 kg) installed on the accessory rail on the left.
2 - 17
Page 17 of 18 TD SD 31
COROSKOP Plus
2 - 18
This page intentionally left blank.
Room Planning
COROSKOP Plus
Page 18 of 18 TD SD 31
Installation Preparation
General Remarks
COROSKOP Hi-P
3-1
3
The COROSKOP Hi-P preinstallation parts are part of the preinstallation shipment. The preinstallation kit contains the mounting frame including the small mounting plate
respectively the small mounting plate or the large mounting plate, as well as the ceiling outlet (for details, see Page 8-7).
KOORDINAT M The KOORDINAT M preinstallation parts are part of the preinstallation shipment. Preinstallation on the floor:
Installation Kit 50; contains base with installation hardware.
With installation in the basement, use the mounting frame. Panels are located at the foot end of the table (see Page 2-9) for:
- Injector connection - Compressed air connection - Blank panel for installation of an electrical outlet. For details, see Page 3-20.
Page 1 of 20 TD SD 31
COROSKOP Plus
3-2
MTS-A3
Installation Preparation
3
The MTS-A3 preinstallation kit is part of the preinstallation shipment. Transversely installed Unistrut rails are prerequisite for the MTS-A3 longitudinal rails.
Grid spacing for ceiling installation is max. 1350 mm.
Additional Unistrut rails for the MTS-A3 are needed (customers contractor) to position
the MTS-A3 longitudinal rails per Version 2. These must be installed turned 90 o to the exiting Unistrut rails (see Pages 8-8/9/10). Page 2-3).
The longitudinal rails can be shortened (see MTS-A3 Positioning according to Version 4,
Control Room Distributor Place the control room distributor in the control room either under the control console
(option) or mount it on the wall (obtain mounting hardware locally). For details, see Pages 8-12/13.
If configured, plan the second footswitch and/or second unit control console options in
the control room so that when they are operated, there is visual contact to operating personnel, to the patient and to the injector in the examination room.
VideoRec or Video Recorder + Monitor M44-2M (Option) When setting up the video recorder, make sure there is sufficient air circulation to
dissipate heat. The clearance to the walls of built-in fixtures or other components must be at least 3 cm.
M102 with Video Distributor Amplifier (Option) Position the M102 in the vicinity of the control room distributor on the wall. For details, see Page 8-15. Reference Monitor (Option) Positioned in the MTS (standard) or in the control room (see Page 2-12). Control Cabinets For floor and wall mounting, as well as for cable lead-in, see the chapter "Detail
Planning", Pages 8-1 to 8-4.
Store excess cable lengths in the cable cabinet, under computer flooring or in cable
conduits, etc. (see Page 4-6).
Cooling System For further information, see the MEGALIX Cooling System PG, R71-020.021.06.
COROSKOP Plus
Page 2 of 20 TD SD 31
Installation Preparation
Upper Body Radiation Shield, Surgery Light, Injector (Option)
1. Overview Approved Attachments
Ceiling column, fixed mounting (for Dimensional Drawing, see Page 8-18; for positioning, see Page 3-14) or ceiling column mounted on the carriage (for Dimensional Drawing, see Page 8-19; for positioning, see Pages 3-15, 8-9) Upper body radiation shield or injector stand or upper body radiation shield and surgery light Upper body radiation shield and/or surgery light Second carriage for injector stand possible
3-3
3
MTS-A3 longitudinal and transverse arrangement (MTS versions 1-4; see Page 2-3)
All mounting hardware is included in the shipment. For ordering information, see the
particular SIR-PM. 3. Surgery Light (Option)
For dimensional drawing, see Page 8-18/19. For positioning, see Pages 3-14/15, 8-9. For electrical connection, see Page 3-19.
4. Injector (Option)
There are 3 different ways in which the Injector Rack Mount Unit can be installed. For
details, see Page 8-22.
For cable lengths, see Page 8-22. Lower Body Radiation Shield (Option) The lower body radiation shield is mounted on the left or right on the table. Second Display (Option) Positioned in the control room or in the examination room on the wall or on a console
(installation materials must be ordered locally).
3 3
Page 3 of 20 TD SD 31
COROSKOP Plus
3-4
Installation Preparation
On-Floor-mounting with Small Mounting Plate (not for USA)
COROSKOP Hi-P/KOORDINAT M
The small COROSKOP Hi-P mounting plate and the table base are installed directly on solid concrete or load-bearing, self-leveling grout (cut away existing floor covering).
3
COROSKOP Hi-P small mounting plate (722 mm dia.) KOORDINAT M table base
1:50
*1
Guide cables to the table base through the opening in the ceiling of the room below (see Page 3-6). With installation in the basement, use the mounting frame.
COROSKOP Plus
Page 4 of 20 TD SD 31
Installation Preparation
3-5
3
Small mounting plate with 13 holes (25 mm dia.): - for installation directly on concrete with Liebig expansion bolts (15 mm dia.); 7 holes, evenly spaced, must be used (see Page 2-13). - for installation directly on the mounting frame
Orientation point/isocenter, COROSKOP Hi-P in normal position Mounting points for COROSKOP Hi-P
Insulation parts
*1, *2
Solid concrete
*1
S15/70 Liebig expansion bolts (7 pcs. included in shipment) Floor mounting using studs through the ceiling of room below (obtain studs locally)
*2
1:2
Page 5 of 20 TD SD 31
COROSKOP Plus
3-6
Installation Preparation
KOORDINAT M: Table Base Dimensional Drawing
View A-A:
3
Table column
Solid concrete
*1
Orientation point
1:5 *2 *1 *2 Shaded area for cable lead-in through the solid ceiling (a 75 mm dia. hole is sufficient for the standard cables) 6 holes for threaded studs
COROSKOP Plus
Page 6 of 20 TD SD 31
Installation Preparation
3-7
3
*1
Orientation point, KOORDINAT M Orientation point/isocenter, COROSKOP Hi-P in normal position 1:50
*1
Guide cables into the table base through openings made in the ceiling of the rooms below (see Page 3-6). If installation is in the basement, use the mounting frame.
Page 7 of 20 TD SD 31
COROSKOP Plus
3-8
Installation Preparation
COROSKOP Hi-P: Dimensional Drawing "Large" Mounting Plate (USA only) 3
Dimensions for installation on concrete
Large mounting plate with 13 mounting holes (25 mm dia.): - for installation directly on concrete with Liebig expansion bolts (15 mm dia.); 7 evenly spaced mounting holes must be used (see Page 2-14).
Insulation parts
*1, *2
1:10
Solid concrete
1:2
COROSKOP Plus
Page 8 of 20 TD SD 31
Installation Preparation
Floor-mounting in Self-leveling Grout with Mounting Frame
COROSKOP Hi-P/KOORDINAT M
3-9
3
1:50
Page 9 of 20 TD SD 31
COROSKOP Plus
3 - 10
Installation Preparation
COROSKOP Hi-P: Dimensional Drawing, Mounting Frame
Dimensions for installation in self-leveling grout
3
4 x floor mounting *1
13xM10 thread for installing the small COROSKOP Hi-P mounting plate on the mounting frame
*1
COROSKOP Plus
Page 10 of 20 TD SD 31
Installation Preparation
3 - 11
Insulation parts COROSKOP Hi-P: small mounting plate Floor covering (max. 10 mm thick)
Mounting frame
1:2 *1 Mounting frame installation to allow adjustment to floor level With grout thickness of 62 mm to 82 mm: adjust using shims Solid concrete
Fill in space around floor frame with concrete after installation. Solid concrete 1:5
*1 *2
B15/45 + B200 Liebig expansion bolts (4 pcs. included in shipment). After installation, cut off excessive length. Special lengths available, see Liebig catalog
Page 11 of 20 TD SD 31
*2
COROSKOP Plus
3 - 12
Installation Preparation
KOORDINAT M: Mounting Frame Dimensional Drawing
Dimensions for installation in self-leveling grout Front view
3
Orientation point
Table base
1:5
COROSKOP Plus
Page 12 of 20 TD SD 31
Installation Preparation
3 - 13
Mounting frame
Mounting frame installation to allow adjustment to floor level Self-leveling grout: minimum depth 62 mm *1
*1
Solid concrete
Solid concrete
*2 1:5 *1 *2 B15/45 + B200 Liebig expansion bolts (4 pcs. included in shipment) Cut off excess length after installation Special lengths available, see Liebig catalog
Siemens AG Medical Engineering RAA2-315.021.14 Rev. 01 10.97 Page 13 of 20 TD SD 31 COROSKOP Plus
3 - 14
System Ceiling Installation
Installation Preparation
3 3
Grid spacing *3 *5
*2
*1
*6
*7 1:50
*1 *2 *3 *4 *5 *6
MTS-A3 Do not plan any lighting fixtures, ventilation inlets, ventilation outlets, sprinkler elements, etc. between the rails.
*7 IN2 SU
Orientation point/isocenter, COROSKOP Hi-P in normal position COROSKOP Hi-P ceiling outlet (see Page 8-7) MTS-A3 ceiling outlet Mounting hole spacing Observe max. admissible rail overhang Mounting plate (option) for - radiation shield and surgery light (see Page 8-18) - or injector stand (see Page 8-18) With a smaller ceiling grid, shift the position of the installation plate to the side until it is centered to the ceiling grid. Injector head fixed installation on the ceiling (option) Surgery light (option) (see Pages 3-19, 8-18)
COROSKOP Plus
Page 14 of 20 TD SD 31
*4
Siemens AG Medical Engineering
Installation Preparation
MTS Version 1 with Moveable Upper Body Radiation Shield (Option)
Recommendation: ceiling grid: 67.5 cm
3 - 15
Grid spacing *3 *5
*2
*1
*7
*5
*6
*4
1:50
Do not plan any lighting fixtures, ventilation inlets, ventilation outlets, sprinkler elements, etc. between the rails.
*1 *2 *3 *4 *5 *6
MTS-A3
*7
Orientation point/isocenter, COROSKOP Hi-P in normal position COROSKOP Hi-P ceiling outlet (see Page 8-7) MTS-A3 ceiling outlet Mounting hole spacing Observe max. admissible rail overhang. longitudinal rails (option) for radiation shield, surgery light, injector stand (option) (see Page 8-19); length when shipped: 4m, can be shortened. With a smaller ceiling grid, shift position of longitudinal rails to the side if needed so that the max. rail overhang is not exceeded. Injector head mounted on injector ceiling column in the radiation shield rails (option) Surgery light (option) (see Pages 3-19, 8-19)
Page 15 of 20 TD SD 31
COROSKOP Plus
3 - 16
In-house Electrical Installation
Installation Preparation
3
For power line connection voltages, see Chapter 5. Remarks regarding the recommendation on Page 3-17 for the room electrical installation per DIN VDE 0107
Ground Fault Interrupters
Use only ground fault interrupters for AC and pulsed DC fault currents, surge-protected,
with I!N=30 mA.
Ground fault interrupters cannot be used in power systems without an N-conductor. Ground Fault Interrupter for System Power Distributor (PD):
IN = 40 A / I!N = 30 mA
ON-OFF Switch (with Control Lamp) Install one ON-OFF switch in the control room for the COROSKOP Plus system. This ONOFF switch should be installed at a height of approx. 1.8 m in the vicinity of the control console. EMERGENCY OFF Switch (with Locking Mechanism) At least one Emergency OFF switch (with locking mechanism) must be installed in the examination room, control room and equipment room to enable quick switch-off of the COROSKOP Plus system in emergency situations. Position these at a height of approx. 1.8 m so that they are easily reached, but cannot be pressed unintentionally.
COROSKOP Plus
Page 16 of 20 TD SD 31
Installation Preparation
Recommendation for Application Group 2
Electrical Installation on Site per DIN VDE 0107
Power supply Backup power supply SV Power lead-in for X-ray
3 - 17
3
Second cable
First cable
5 x per calculation
Legend: 1 Voltage monitoring and switching device per VDE 0107 2 Isolating transformer with overload monitoring 3 Isolation monitor 4 Response and test combination 5 Potential equalization leads to external, conductible parts 6 Power line to permanent or mobile operating devices 7 Power line to permanent or mobile operating devices, e.g. developer machine 8 GFI (ground fault interrupter) 8a GFI for system power distributor, see notice on Page 3-16 8b GFI for generator, see notice Page 3-16 9 System breaker EAT On-Off switch with control lamp for the X-ray system AT1 Emergency OFF switch in examination room AT2 Emergency OFF switch in control room AT3 Emergency OFF switch in equipment room SC 4-part connector socket for CATHCOR (option) PIN Gnd pin socket [] Figures are free cable lengths in m Note: The ratings of the short-circuit prefuses apply only for SIEMENS GFIs
Page 17 of 20 TD SD 31
COROSKOP Plus
3 - 18
Power Cables by Customers Contractor
Installation Preparation
3
- Power cable, 5x16 mm2. - The connector terminals are set up for max. 16 mm2. - Power line cabling 5 x per the calculation. Take note of the internal line resistance: 1. 2. Measure the internal line resistance, Ri, at the subdistributor using the line resistance meter. Size the power lead-in cable so that the max. admissible Rimax (see table) is not exceeded. Measured internal line resistance plus the resistance of the power lead-in cable " Rimax. UN P 100 kW 80 kW
" 0.10 Ohm " 0.14 Ohm " 0.16 Ohm " 0.20 Ohm
" 0.14 Ohm " 0.17 Ohm " 0.21 Ohm " 0.27 Ohm
*Resistance values in Ohms with UN -15% ** Resistance values in Ohms with UN -10% - The POLYDOROS IS is a 100 kW generator. If there is a greater internal line resistance than listed in the above table, generator power is automatically reduced. A max. generator power of 80 kW is sufficient for the X-ray tube unit assemblies currently being shipped. - Minimum Ri: 0.2 Ohm. - Max. cross section of the cable that can be clamped: 70 mm2.
COROSKOP Plus
Page 18 of 20 TD SD 31
Installation Preparation
Connection for Surgery Light (Option)
Surgery Light: - Power consumption: 50 W - Can be connected to 115 V~ or 230 V~
3 - 19
*2a
*2b
*1 *2
For connection to line power, see Page 3-17, *1 The brightness of the surgery lamp is adjusted using an internal dimmer. The light can be connected to line power: - directly, or - a wall switch (*2a), - a second footswitch, Part No. 17 41 313 (*2b), - a relay (*2c) Part No. 10 13 085 When connected over a relay, the surgery lamp is switched on and off when fluoro/exposure is triggered.
Page 19 of 20 TD SD 31
COROSKOP Plus
3 - 20
Installation Preparation
A table electrical outlet can be installed in the KOORDINAT M:
- The electrical outlet and the power lead-in cable must be obtained from the customers contractor. - A 135x95 mm blank panel is provided at the foot end of the table for installation of the electrical outlet. - Use only a fully-enclosed electrical outlet (insulated and splash-water protected). - Labeling of the max. admissible consumption at the power outlet is required. - Observe the country-specific regulations. - For connection to the room distributor, see *2 Page 3-17. - The power cable must be led directly into the table and connected to the electrical outlet, i.e. intermediate connection in the table is not allowed.
COROSKOP Plus
Page 20 of 20 TD SD 31
System Connections
General Remarks
All system-specific cables are included with the shipment.
4-1
4
The system cabling is comprised of standard cables and standard wiring harnesses with
a fixed length. These may not be shortened.
Store excess cable lengths in the cable cabinet, in cable conduits, under computer
flooring, etc.
Standard cabling has already been prepared for the following options:
Second ECC Second monitor ACOM T.O.P. (ACOM M) Hardcopy Video printer Second footswitch Second unit control console VideoRec (Video-Recorder + VCR-Interface + scan converter) 110 V Power Pack
For remarks regarding the room electrical installation, see Page 3-16.
Page 1 of 6 TD SD 31
COROSKOP Plus
4-2
Overview of System Connections
System Connections
4
RD
SU
RD
RD
RD
Remarks:
RD = room distributor Room distributor and power cables are not included in the shipment. The illustration includes the cable cabinet.
The wiring harnesses can be laid directly to the fixpoints, i.e. without first going through the cable cabinet. Excess cable lengths must then be stored in some other manner (in cable conduits, under computer flooring, etc.).
COROSKOP Plus
Page 2 of 6 TD SD 31
System Connections
4-3
4
to Fixpoint
Cable conduit cross section in mm2 780 500 500 3125 3125 2000 780 3125 3125 780 780 3125 3125 220 220 120 150 100 120 900 290 160
Conduit, Minimum Maximum size Opening in Fixpointin inches mm Distance in m 2" 3" 3" 3" 3" 2.5" 3" 3" 3" 2" 2" 3" 3" 3" 2" 2" 2.5" 3" 2" 2.5" 2" 2" 32 dia.
42 42
Remarks
P1 P1 P1 P1 P1 P1 SD1 SD1 SD1 SD1 SD1 SD1 IS IS CRX IS CRE CRE CRE SD1 SD1 R1 PU1 PD SU
T
PU1 PU1 PU1 SD1 SD1 CU1 D CRX T CU1 PU1 IS CRX T D CRE D CRX T IWO D1 CRX
15 15 15 15 15 30 19 19 14 48 5 34 27 29 19 27 27 27 27 18 25 20* Meas. station (opt.) Meas. station (opt.) with Cathcor (option) only Meas. station (opt.) Injector wall connection (option) 2nd display (option) Fiber optic cable for ACOM M ** (opt.) connection Contractor cable Contractor cable Contractor cable
Contractor cable
80 x 55 80 x 55 50 dia. 60 x 35 55 x 55 80 x 55 35 dia. 35 x 35 80 x 55 73 x 35
55 x 30
20 dia. 53x30 40 x 40 70 x 17 45 x 15 60 x 30 40 x 16 16 x 8
Pwr. cable, Power Unit 1 (POLYDOROS IS) -> P.3-16/ 17/18 Pwr. cable, Power Distributor -> P. 3-16/17/18 Pwr. cable, surgery lamp (option) -> P. 3-16/17/18
Pwr. cable, table electrical outlet (option) -> P. 3-17/20
* This fiber optic cable is always configured in the HICOR T.O.P. - control room distributor wiring harness. If the free cable length (20 m) from control room distributor to ACOM M is not sufficient, a 300 m fiber optic cable can be laid directly from the ACOM to the HICOR. ** If needed, plan in a wall power outlet (by customers contractor) for the ACOM B/M power line connection. Because of the relatively high power consumption, the power connection may not be made in the control room distributor.
Page 3 of 6 TD SD 31
COROSKOP Plus
4-4
System Connections
Room Light Connection, Radiation Display, Door Contact, EPS
*3
24 VAC / 1A
internal, 2 m
SD1
*3
24 VAC / 1A
internal, 2 m
SD1
2 (shielded)
internal, 2 m
SD1
2 (shielded)
Potential-free door contact (24 VDC / 100mA)*1; Door closed = contact closed
internal, 2 m
SD1
2 (shielded)
Potential-free contact (24 VDC / 100mA) of EPS*2; emergency power "on"= contact closed
*1
If needed, a switch for radiation blocking can be installed in series with the door contact: - The switch (24 VDC / 100 mA) must be provided by customers contractor. - Control is +24 V, internal or external. - Plan in the switch so that unintentional operation is not possible. - The switch must be provided with an appropritate symbol (or label) so that its function is easily recognizable. EPS = Emergency Power Supply (option) (see Page 1-4) If a relay is controlled over this contact with DC voltage, a free-running diode must be wired over the relay coil.
*2 *3
COROSKOP Plus
Page 4 of 6 TD SD 31
System Connections
Laying Cables
Example of Cables Laid in Cable Conduits
4-5
4 4
Remarks:
Lay cables / wiring harnesses separately in pipes or in enclosed cable conduits. If cables are laid in open cable conduits, separate cables using metal spacer plates or a
similar method.
The depth of the cable conduit should be at least 70 mm; if there are intersections, depth
plan a greater depth.
Page 5 of 6 TD SD 31
COROSKOP Plus
4-6
Remarks Regarding Laying of Cables
System Connections
4
Storage of Excess Cable Lengths: Cable stored in snake-like fashion in the (included in shipment) in the cable cabinet (service-friendly) in cable conduits (take note of accessibility) in ceiling/wall boxes (take note of accessibility) under computer flooring (service-friendly)
COROSKOP Plus
Page 6 of 6 TD SD 31
Technical Data
Electrical Data
Nom. connection voltage* [V] +/- 10% 3 ph. 400 POLYDOROS IS (Power Unit PU1) 3 ph. 415 3 ph. 440 3 ph. 480 3 ph. 400 3 ph. 415 Power Distributor PD 3 ph. 440 3 ph. 480 50/60 60 Frequency [Hz] +/- 1Hz 50/60 50 63 50/60 60 50/60 50 35 50 A/C 14 80 A/C 18 with fluoro 160 with exposure Fuse, internal [A] Fuse, external [A]
5-1
5
* **
Max. admissible voltage between phase (L1,L2,L3) and Gnd: 300V The connection value was calculated per the following formula: ConnectionxValue = 3 # ConnectionxVoltage # intern alxFu sin g
P1
Small mounting plate Large mounting plate Mounting frame KOORDINAT M (incl. base 16 kg, tabletop 12 kg, accessories 40 kg)
n. a. n. a. SU SU IWO
Examination control console Unit control console Fixed upper body radiation shield, surgery light (option) Moveable upper body radiation shield, surgery light (option) Injector wall terminal box (option)
Page 1 of 4 TD SD 31
COROSKOP Plus
5-2
Equipment Room
Fixpoint CU1 IS Cooling Unit 1 Image System (HICOR T.O.P.) Subsystem Dimensions (W x Dx H) [mm] 440 x 313 x 485 629 x 1006 x 2008 1200 x 434 x 2209 800 x 434 x 2209 800 x 434 x 2209
Technical Data
5
PD/SD1 System control cabinet (double cabinet) PU1 n. a. Power Unit 1 (POLYDOROS IS) Cable cabinet
Control Room
Fixpoint CRX n.a. n. a. Subsystem Control room distributor N465 System Control Console VIdeoRec ( option) comprised of: Video recorder VCR Interface + scan converter Video recorder SONY SVO 9500 ... (option) Monitor M44-2M (option) 110 V Power Pack (option) 2nd Examination Control Console (option) 2nd unit control console (option) 2nd 44 cm SIMOMED HM monitor 2nd 54 cm SIMOMED HM monitor M102 with video distributor amplifier (option) Container (Option) Control console 2nd display (option) Dimensions (W x D x H) [mm] 770 x 220 x 530 400 x 226 x 154 270 x 370 x 144 445 x 210 x 94 270 x 370 x 144 405 x 294 x 390 160 x 155 x 270 360 x 180 x 70/120 240 x 180 x 70/110 407 x 456 x 407 500 x 520 x 480 770 x 220 x 530 570 x 800 x 730 1400 x 876 x 730 see Page 8-15/16 Weight [kg] 15 2 12 7 12 approx. 15 5 2 2 22 31 approx. 15 approx. 30 approx. 50 approx. 9 Heat dissipation [kW] 0.1 0.1
0.1 0.1 0.1 0.1 n.a. n. a. n.a. 0.15 0.15 0.1 n.a. n.a. n.a.
COROSKOP Plus
Page 2 of 4 TD SD 31
Technical Data
Environmental Conditions
Examination and control Temperature range: room Image System HICOR T.O.P. Relative humidity: Temperature range: Relative humidity: max. temperature gradient: Air flow volume: max. noise generation: POLYDOROS IS Temperature range: Relative humidity: max. temperature gradient: Air flow volume: max. noise generation: Cooling unit (for Megalix X-ray tubes) Cooling air: Air flow volume: max. noise generation:
5-3
5
+150 C ... +350 C (recomm. temperature 220 C) 20 - 75 % non-condensing +150 C ... +280 C 20- 80 % non-condensing 15 K/h 1680 m3/h 57 dB(A) +100 C ... +400 C 20 - 75 % non-condensing 5 K/h 800 m3/h 55 dB(A) +40 C ... +300 C (frost-free room) 1000 m3/h 60 dB(A)
Paint Surfaces
Primary color Combination color White textured, Med-No. 4146, similar to RAL gray-white 9002 Neutral gray textured, Med-No. 4426, similar to RAL dusty-gray 7037 anthracite textured, Med-No. 4079
Page 3 of 4 TD SD 31
COROSKOP Plus
5-4
This page intentionally left blank.
Technical Data
COROSKOP Plus
Page 4 of 4 TD SD 31
Transport Conditions
Packaging and Transport Paths
Largest crate with packaging - weight
6-1
6
Largest single part with transport carriage W2450 x D830 x H1800 mm - 660 kg (min. door opening) - weight
Page 1 of 2 TD SD 31
COROSKOP Plus
6-2
This page intentionally left blank.
Transport Conditions
COROSKOP Plus
Page 2 of 2 TD SD 31
Project Management
n. a.
This is an empty chapter.
7-1
7
It was added only so that the standardized chapter sequence could be maintained.
Page 1 of 2 TD SD 31
COROSKOP Plus
7-2
This page intentionally left blank.
Project Management
COROSKOP Plus
Page 2 of 2 TD SD 31
Detail Planning
Power Unit 1 (POLYDOROS IS-C)
Front View *4 Exhaust air *4 Side View
8-1
8
*2 *3 *3
Air intake
*1 1:20
*1 *2 *3
Base panels removeable on all sides. No spacing to wall required on sides. When set up corner-to-corner, min. spacing to cable cabinet is 40 mm (needed to open cabinet doors). Wall anchoring needed because some assemblies can be swung out. Bolt cabinets to wall using supplied brackets and screw anchors. Obtain screw anchors locally. If needed, cabinet can also be bolted to floor (see next page).
*4 for Service
Page 1 of 22 TD SD 31
COROSKOP Plus
8-2
Detail Planning
Exhaust air
*2
*1
*2
*2
1:20
*1 *2
Holes 2 x 14 mm dia. to bolt to the floor. Obtain screw anchors locally. Cable lead-in. In order to provide strain relief for the power cable, lead the power cable into the cabinet base as far to the right as possible (in the direction of the arrows). The power cable cannot be led in from the left side.
COROSKOP Plus
Page 2 of 22 TD SD 31
Detail Planning
System Control Cabinets (Double Cabinet)
Front View *4 Side View
8-3
8
*4
Exhaust air
*3
*2
Air intake
1:20 *1 *1 *1 Cabinet base Top-down view *1 *2 *3 Base panels removable on all sides. No spacing to wall on side required. When set up corner-to-corner, 40 mm min. spacing to cable cabinet (needed to open cabinet doors). Wall mounting required because assemblies can be pivoted out. Bolt cabinet to wall using supplied brackets. Obtain screw anchors locally. If needed, cabinet can also be bolted to floor (see next page). Cabinet doors removable
*4 for Service
Siemens AG Medical Engineering
Page 3 of 22 TD SD 31
COROSKOP Plus
8-4
Detail Planning
Exhaust air
Exhaust air
*2
*2
*1
*3
*3 1:20
*1 *2 *3
14 mm dia. holes for floor mounting. Obtain screw anchors locally. Area (area outlined with heavy line) to guide in cables from the floor (cable conduit). Cables can be led into cabinet base from all sides, except from front, over the floor after removing the base cover panels.
COROSKOP Plus
Page 4 of 22 TD SD 31
Detail Planning
HICOR T.O.P. Cabinet
Dimensions
Front View Side View
8
8-5
8
Rear View
Exhaust air
Air intake
Bypass switch
1:20
Page 5 of 22 TD SD 31
COROSKOP Plus
8-6
Recommendation for Cabinet Setup
Detail Planning
8
Service area, approx. 4006 x 1429 mm; These minimum spaces must be provided for service purposes.
*2
Service position
*1
1:50
*1 *2
Admissible position during operation; can be moved on castors. Cable loop from fixpoint IS, min. 1.8 m
COROSKOP Plus
Page 6 of 22 TD SD 31
Detail Planning
COROSKOP Hi-P Ceiling Outlet
8-7
8
*3
Unistrut ceiling
Cover plate
Cover housing
*1 *2 *3
Ceiling mount, tension per screw, less than 1000 N Threaded stud spacing Cable run on ceiling. Cable inlet to be cut into the cover housing. Cable duct inner dimensions, min. 60 x 160 mm
1:5
Page 7 of 22 TD SD 31
COROSKOP Plus
Orientation point/Isocenter, COROSKOP Hi-P in normal position Isocenter, COROSKOP Hi-P in oblique position 1:50
*2
8-8
Detail Planning
*1 *1 Including 500 mm safety distance between COROSKOP Hi-P and wall. *2 Dimensions not in parenthesis: Table longitudinal movement not limited (standard) Dimensions in parenthesis: Table longitudinal movement limited by the room width.
8
Detail Planning
Ceiling Installation
Recommendation: 67.5 cm ceiling grid
8-9
8
Grid spacing
*3 *2
*1
Do not plan lighting fixtures, ventilation outlets, sprinklers, etc. between the ceiling rails. *1 *2 *3 *4 *5 *6 Orientation point/isocenter, COROSKOP Hi-P in normal position COROSKOP Hi-P ceiling outlet (see Page 8-7) MTS-A3 ceiling outlet Mounting hole spacing Observe max. admissible rail overhang Ceiling rails (option) for radiation shield, surgery light, injector stand (see Page 8-18); length when shipped: 4m; can be shortened Injector stand mounted on injector ceiling column in the radiation shield rail (option) Surgery light (option)
MTS-A3
IN1 SU
Radiation shield
Page 9 of 22 TD SD 31
COROSKOP Plus
8 - 10
Detail Planning
Installation Recommendation for MTS-A3 Longitudinal Rails
8
Base ceiling
1:5
Unistrut nut
Longitudinal rail
1:2
COROSKOP Plus
Page 10 of 22 TD SD 31
Detail Planning
N465 System Control Console
Position the N465 in the vicinity of the control room distributor.
8 - 11
8
Table Installation
Wall Installation
Air intake slots in floor panel *1
*1
1:5
Page 11 of 22 TD SD 31
COROSKOP Plus
8 - 12
Control Room Distributor
Standard (max. distance):
N465 (System console) 2.0m 20.0m 2.0m 21m 2.7m 2.7m 2.7m *** 3.0m
Detail Planning
8
CRX
* ** ***
COROSKOP Plus
Obtain installation hardware locally. If needed, a fiber optic cable (max. 300 m) can be laid directly from the HICOR T.O.P. to the ACOM M. See Page 8-22
RAA2-315.021.14 Rev. 01 10.97 Page 12 of 22 TD SD 31
1:10
Detail Planning
Control Console (Option)/Container (Option)
8 - 13
8
Container
Control console
*1
Container for VideoRec (video recorder + VCR interface + scan converter), video printer
Height adjustable
Cable conduit
*1
Page 13 of 22 TD SD 31
COROSKOP Plus
8 - 14
110 V Power Pack (Option)
Detail Planning
8
Das 110 V Power Pack is required only for 110V/60 Hz power nets for connection of the
video recorder (option).
Set it up in the vicinity of the control room distributor (see also Page 8-12).
Front View
Side View
*1
*1
1:5
*1
COROSKOP Plus
Page 14 of 22 TD SD 31
Detail Planning
M102 with Video Distributor Amplifier (Option)
8 - 15
8
Without the M102, 1 monitor can be connected to the control room distributor. The
M102 makes it possible to connect additional monitors (max. 4 monitors).
Position the M102 in the vicinity of the control room distributor (see Page 8-12).
*1
*2
*2
1:10
*1 *2 *3
Obtain installation hardware locally. Cable lead-in in the direction of arrow possible from above or below 2 different length cable sets (video, power, ground) can be shipped: 9 m or 29 m
RAA2-315.021.14 Rev. 01 10.97 Page 15 of 22 TD SD 31 COROSKOP Plus
*3
*3
8 - 16
Display
Display mount
Display mount
*1
*1
*2
Detail Planning
1:5 *1 *2 *3 Lead in cables from above or below 2 x 8 mm dia. mounting holes for installation on customers console Mounting holes for wall installation
Detail Planning
Display
*1
*2
*1
*2
*1 *2
Opening to guide in cables 2 x 8 mm dia. mounting holes for installation on customer-provided console
1:5
8 - 17
8 - 18
Detail Planning
Fixed Mounted Upper Body Radiation Shield, Surgery Light (Option)
Not approved for MTS-A3 positioning per Version 2 (MTS perpendicular)
(see Page 2-3).
8
Surgery light
Ceiling column
Large pane
Mounting plate
23 kg
Ceiling column with support, spring- 28 kg loaded arm and large pane, or Maximum configuration (lamp and large pane) 32 kg
COROSKOP Plus
Page 18 of 22 TD SD 31
Detail Planning
Approved for MTS-A3 positioning per Version 1-4 (see Page 2-3). Second carriage for injector head also possible.
8 - 19
8
Surgery light
Large pane
30 kg 5 kg
1) Ceiling column with support, spring- 28 kg loaded arm and large pane or 2) Ceiling column with support, spring- 22 kg loaded arm and surgery light or 3) Maximum configuration (light and large pane) 32 kg
Page 19 of 22 TD SD 31
COROSKOP Plus
8 - 20
Injector Wall Terminal Box (Option)
Shown without Injector Panel
Front View
Detail Planning
8
Side View
Mounting holes
COROSKOP Plus
Page 20 of 22 TD SD 31
Detail Planning
Planning Recommendation for Injector Wall Connection
8 - 21
8
*1
1:50
IWO Injector wall terminal box (option, injector wall outlet) *1 Orientation point/isocenter, COROSKOP Hi-P in normal position
Page 21 of 22 TD SD 31
COROSKOP Plus
8 - 22
Injector Rack Mount Unit (Option)
Versions:
Detail Planning
8
- with table base or - with permanently mounted ceiling stand (Page 8-18) or - with moveable ceiling stand (installed in rail system for radiation shield, Page 8-19).
Approved: ANGIOMAT 6000 and MEDRAD Mark V Plus For installation location and approved MTS-Variants see paragraph "1. Overview", Page
3-3.
Place injector control unit on contractor-installed wall console etc.,if possible, in the
equipment room.
Cable lengths:
Z51 (in control room distributor or wall terminal box) l1 l2 Injector control unit l3
Injector head
Cable lengths (m) l1 10.0 3.0 l2 24.0 max. 30.5 l3 24.0 max. 30.5
COROSKOP Plus
Page 22 of 22 TD SD 31
9-1
9
The COROSKOP Plus system is the successor of the COROSKOP T.O.P system and the COROSKOP CLASSIC system.
Chapter 2
Dimensions corrected. New: KOORDINAT M, view X Dimensions corrected. Now VideoRec is an option. New: Video recorder + monitor M44-2M (option) New: Reference monitor (option) M102 for connection of additional monitors added.
Chapter 3
Page 2
Now VideoRec is an option. New: M102 with video distributor amplifier (option). New: Reference monitor (option). Illustration for laying the cables to the table base changed.
Page 6
Chapter 4
Connection of the table power outlet expanded. Wiring harness No. 19 (SD1 - IWO) changed from 19 to 18 m. Note *1 (optional radiation blocking switch) rewritten.
Chapter 5
Page 2
PG Speed Info 02/96 incorporated. New: Video recorder, monitor M44-2M, M102
Page 3
Paragraph "Environmental Conditions": Rel. humidity: lower tolerance changed to 20%. Examination room and control room, temperature range: lower tolerance changed to +15 degrees C.
Chapter 8
Page 12 Page 15
New: Max. distance added for CRX - M102 and CRX - video recorder. New: Dimensional drawing M102 with video distributor amplifier (option)
Changes to the COROSKOP T.O.P. (+KOORDINAT M) PG, RAA2-315.021.13.01.02 are labeled with
Page 1 of 2 TD SD 31
COROSKOP Plus
9-2
Contacts:
Tel: +49(0)9191/18-9515, Fax -9903 e-Mail: heinrich.armer@med.siemens.de Tel: +49(0)9191/18-9511, Fax -9903 e-Mail: hermann.brueckl@med.siemens.de Tel. +49(0)9191/18-8080, Fax -9944
COROSKOP Plus
Page 2 of 2 TD SD 31