Académique Documents
Professionnel Documents
Culture Documents
Agenda 1. Qualification of stability chambers 2. Calibration of equipment of stability chambers 3. Maintenance of stability chambers
Qualification
1. Qualification of stability chambers 1.1 General aspects 1.2 URS, analysis of local conditions 1.3 Qualification: DQ, IQ, OQ, PQ 1.4 periodic Re-Qualification
define good Team-Members, experts where needed QA, Engineer, technician, system owner (teamleader) the team determines your success
approach: planning doing reporting using professional project management Tool (MS-Project) define milestones arrange periodical status- meetings
Qualification
1. Qualification of stability chambers 1.1 General aspects 1.2 URS, analysis of local conditions 1.3 Qualification: DQ, IQ, OQ, PQ 1.4 periodic Re-Qualification
define usage very precisely (which kind of samples will be stored?) define climatic condition -nominal value, according ICH Q1A, WHO (e.g. accelerated conditions: 40C 2C / 75 % r.H. 5 % r.H.) - limits, according ICH Q1A, - additional conditions (e.g. Humidity in 5C-Cells is not recommended, but very important)
define location, shape and area needed define equipment (shelves, drawer needed?) define safety and enviromental aspects (fire detector) define security requirements (on-line monitoring, locking system) define budget define basis and scale of qualification
Find the best location (building) for stability chambers consider flow line (Short way, minimal horizontal moving) infrastructure (water, air condition, additional deconstruction) safety and enviromental aspects (energy saving) security requirements (available access control-systems)
Qualification
1. Qualification of stability chambers: 1.1 General aspects 1.2 URS, analysis of local conditions 1.3 Qualification: DQ, IQ, OQ, PQ 1.4 periodic Re-Qualification
Qualification overview
1.Phase MQP DQ
2.Phase IQ
3.Phase OQ
4.Phase PQ MQR
DQ: Design Qualification MQP: Master-Qualification Plan MQR: Master-Qualification Report IQ: Installation Qualification OQ: Operational Qualification PQ: Performance Qualification
write and approve DQ-Plan getting different tenders Check technical manuals URS available write and approve DQ-Report (with reported open issues) before starting IQ End Order of the designed chamber
write and approve IQ-Plan compare ordered and delivered material Installation of the climatic chamber according local requirements / standards first function check prepare a (technical) logbook Installation of additional equipment (monitoring sensor or flash-light) write and approve IQ-Report (with reported open issues) before starting OQ End installed climatic chamber
write and approve OQ-Plan OQ-tests with empty chamber: running of the climatic cell: does it work as expected perform extensive Temperature- and Humidity-Room-Mapping testing any influence of the climatic conditions (light, open doors) simulate a power failure and protocol the behavior of the chamber to make conclusions find a representative position for the monitoring sensor write and approve OQ-Report (with reported open issues) before starting PQ End Climatic chamber ready for operation
write and approve PQ-Plan repeat the OQ-tests with filled chamber: running of the climatic cell: does it work as expected perform extensive Temperature- and Humidity-Room-Mapping testing any influence of the climatic conditions (light, open doors) simulate a power failure and protocol the behavior of the chamber to make conclusions Go live with the on-line monitoring, if available Calibrate the sensors write + approve PQ-Report (without any open issues) End completely qualified climatic chamber
Qualification
Qualification
1. Qualification of stability chambers: 1.1 General aspects 1.2 URS, analysis of local conditions 1.3 Qualification: DQ, IQ, OQ, PQ 1.4 periodic Re-Qualification 2. Calibration of equipment: 3. Maintenance of stability chambers:
Start
Strategy: Review of qualification status after a pre-defined time period write and approve Re-Qualification-Plan check all relevant parameters (i.e. MSR) is the change documentation complete? Repeat the Temperatur- and Humidity- Room-Mapping write and approve PQ-Report before writing the Final Report End re-qualified climatic chamber
Calibration
1. Qualification of stability chambers 2. Calibration of equipment of stability chambers 3. Maintenance of stability chambers
Calibration concepts
Calibration of equipment (e.g. sensors for on-line monitoring) in place Calibration including whole measured distance one point calibration at set point
Calibration of equipment (e.g. data-logger for temperature mapping) with a certified, adjusted master tool (Certified Institution) Calibration without measured distance Multi point calibration all set points must lie within calibration range Data logger must be suitable for intended conditions (e.g. 40C / 75%r.H.)
Maintenance
3. Maintenance of stability chambers 3.1 periodical maintenance, technical support 3.2 microbiological monitoring, cleaning 3.3 on-line monitoring 3.4 Alarm-System 3.5 Change Control 3.6 Temperature- or Humidity-Deviations
Defined maintenance according service-protocol: checking pressure, air flow, valves on-going maintenance including auxiliary media (steam, cooling water) support in any case of emergency, 24h a day
Maintenance
Maintenance
3. Maintenance of stability chambers: 3.1 periodical maintenance, technical support 3.2 microbiological monitoring, cleaning 3.3 on-line monitoring 3.4 Alarm-System 3.5 Change Control 3.6 Temperature- or Humidity-Deviations
periodical microbiological monitoring according written procedure periodical cleaning of stability chambers including according written procedure, e.g. every 6 months define acceptance criteria
Maintenance
3. Maintenance of stability chambers 3.1 periodical maintenance, technical support 3.2 microbiological monitoring, cleaning 3.3 on-line monitoring 3.4 Alarm-System 3.5 Change Control 3.6 Temperature- or Humidity-Deviations
confirm the defined position for any monitoring sensor which is used to visualize the climatic conditions, generating GMP-Data to document any deviations Date could be requested by inspectors monitoring sensors are calibrated annually
Maintenance
3. Maintenance of stability chambers 3.1 periodical maintenance, technical support 3.2 microbiological monitoring, cleaning 3.3 on-line monitoring 3.4 Alarm-System 3.5 Change Control 3.6 Temperature- or Humidity-Deviations
double- control of the climatic limits using another, different, independent alarm system
Maintenance
3. Maintenance of stability chambers 3.1 periodical maintenance, technical support 3.2 microbiological monitoring, cleaning 3.3 on-line monitoring 3.4 Alarm-System 3.5 Change Control 3.6 Temperature- or Humidity-Deviations
Approved changes Planned change has to be assessed Relevant changes have to be approved by Quality Assurance before realization
Maintenance
3. Maintenance of stability chambers 3.1 periodical maintenance, technical support 3.2 microbiological monitoring, cleaning 3.3 on-line monitoring 3.4 Alarm-System 3.5 Change Control 3.6 Temperature- or Humidity-Deviations
Short time deviations are allowed by ICH (short term spikes due to opening of doorsare accepted as unavoidable) important deviations have to be documented (i.e. in a Deviation Report) according ICH (Excursions that exceed the defined tolerances for more than 24 hours should be describes in the study report)
Chamber
drawer
access controlsystems
escape route
ramp
We Innovate Healthcare