Académique Documents
Professionnel Documents
Culture Documents
Radiation Guideline
Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging
Part 6
Test protocols for parts 25
These are the prescribed Test Protocols for Parts 2 5 of Radiation Guideline 6, defined in clause 3 of the Radiation Control Regulation 2003 (except the Mammography Radiation Guideline). This edition supersedes the Test Protocols published in June 2000. From 24 September 2003 the Department of Environment and Conservation (DEC) incorporates the Environment Protection Authority (EPA), which is defined in section 4 of the Radiation Control Act 1990 as the Authority responsible for administering the Act and Regulation. Statutory functions and powers in the Radiation Control Act 1990 continue to be exercised in the name of the EPA.
For technical information about this Guideline contact the Radiation Control Section of the DEC on (02) 9995 5959.
Disclaimer: The EPA and the DEC have prepared this Guideline in good faith exercising all due care and attention, but no representation or warranty, express or implied, is made as to the relevance, accuracy, completeness or fitness for purpose of this Guideline in respect of any particular users circumstances. The owner of apparatus should rely on their own inquiries and, where appropriate, seek expert advice as to the suitability of the application of this Guideline in particular cases. Comments on the Guideline should be made to the Director Radiation Control, DEC, so that changes can be considered. The EPA and the DEC accept no responsibility for any loss or damage resulting from the application of the Guideline. This document is subject to revision without notice. It is the responsibility of the reader to ensure that the latest version is being used.
Published by: NSW Environment Protection Authority 5961 Goulburn Street PO Box A290 Sydney South 1232 Ph: (02) 9995 5000 (switchboard) Ph: 131 555 (information & publications requests) Fax: (02) 9995 5999 (general) TTY: (02) 9211 4723 Email: info@epa.nsw.gov.au Web: www.environment.nsw.gov.au ISBN 1 74137 049 3 DEC 2004/30 First published June 2000 Second edition March 2004 This material may be reproduced in whole or in part, provided the meaning is unchanged and the source is acknowledged. Printed on recycled paper
1 2
2 2
Section 2General
2.1 2.2 2.3 2.4 2.5 Protection of the x-ray tube Line voltage Radiation units Certificate of Compliance Minor non-compliance issues
3
3 3 3 3 4
5
5 5 6 7 8
Section 4Kilovoltage
4.1 4.2 Accuracy Reproducibility
10
10 10
12
12 13 13
Section 6Collimation
6.1 6.2 6.3 6.4 6.5 6.6 6.7 Radiographic apparatus Fluoroscopic apparatus Intra-oral dental apparatus Panoramic apparatus Cephalometric apparatus Intensity of collimator light source Dark shutter (condenser discharge apparatus)
15
15 16 17 17 18 18 19
20
20
22
22 22 24
26
26 27
29
29
30
30 31 32 34
35
35 35
37 38
INTRODUCTION
This Guideline is for the information of owners, licensed users of ionising radiation apparatus and persons accredited under section 9 of the Radiation Control Act 1990 (the Act) as Consulting Radiation Experts (CREs). The Guideline sets out the basic test protocols that must be followed by CREs in the assessment of radiation apparatus for compliance with the mandatory requirements of Parts 25 of Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging. Compliance testing for the mandatory requirements of the Mammography Radiation Guideline (Radiation Guideline 6: Part 1) must be conducted by following the Royal Australian and New Zealand College of Radiologists Mammography Quality Control Manual (2002). This Guideline may also be used by persons performing quality control testing of diagnostic imaging apparatus to standardise their test procedures. Quality assurance (including quality control) is widely recognised as an essential element of good radiological practice. The primary objective in performing quality control of radiation apparatus is to ensure that it is functioning to a prescribed minimum standard. In the event of amendment to the Act or the Radiation Control Regulation 2003, references to the legislation in this document shall be deemed to refer to the current legislation. In the event of an inconsistency between the Guideline and the current legislation, the requirements of the legislation prevail to the extent of the inconsistency. From 24 September 2003 the Department of Environment and Conservation (NSW) incorporates the Environment Protection Authority (EPA). The EPA is defined in section 4 of the Radiation Control Act 1990 as the Authority responsible for administering the Act. Therefore, statutory functions and powers in the Act and the Radiation Control Regulation 2003 continue to be exercised in the name of the EPA. The Guideline was developed by the Radiation Control Section of the Department of Environment and Conservation (NSW) in consultation with the Radiation Advisory Council. The Department of Environment and Conservation acknowledges the assistance of Mr Lee Collins, Dr Donald McLean and Mr John Robinson, and the input received from stakeholders, in preparing this edition.
1.2
Test equipment
The following items are required to carry out compliance testing: a dosimeter with appropriate ionisation chambers a non-invasive beam analyser or kVp meter and electronic timer aluminium filters (Grade 1100 or equivalent) masking tape a collimator alignment test grid or lead markers/paper clips a perpendicularity test tool a light meter lead sheets a tape measure radiographic cassettes & film/fluorescent screen non-screen film a calculator with statistical functions a resolution gauge a Westmead test object (or equivalent) a 20 cm water phantom 6 sheets of perspex, each 24 cm x 30 cm, 1cm thick 2 mm copper sheet a pinhole camera/star test pattern (0.5 or 1) appropriate phantoms for CT and bone mineral densitometry a densitometer.
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 2
2.2
Line voltage
Variations in line voltage from 240 V may cause equipment to fail the kVp requirements specified in this Guideline. Compliance testing should be carried out at 240 V, which is the optimal line voltage at which diagnostic imaging apparatus should be used. If equipment has failed kVp requirements the owner should have a qualified person monitor the line voltage.
2.3
Radiation units
Quantity Exposure SI Unit coulombs per kilogram (C/kg) Gray (Gy) sievert (Sv) Old Unit roentgen (R) Conversion 1 C/kg = 3876 R
Note: Equivalent Dose = absorbed dose x radiation weighting factor (weighting factor for x-rays = 1)
All measurements should be in SI units. All radiation output measurements should be recorded as absorbed dose in air [1 R = 8.73 mGy in air].
(Source: ICRP60)
2.4
Certificate of Compliance
A Certificate of Compliance must be completed by the CRE and provided to the owner of the apparatus within 21 days of the date of compliance testing for the purpose of certification for registration, regardless of whether the apparatus has passed or failed. In addition, the CRE must, within a reasonable period after the inspection, issue the owner with a report. The report must include readings and calculations, details of non-compliance with mandatory requirements of Radiation Guideline 6 and may include recommendations relating to matters outside mandatory requirements in the Guideline (for example, recommended best practice). The report should note any mandatory requirements that are not applicable to the apparatus.
2.5
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 4
Equipment:
2 mm lead sheet Electronic timer or non-invasive beam analyser
Factors:
50 kVp
Method:
Place the lead sheet between timer and AEC chamber. Set automatic exposure control density to 0. Expose and note the backup time from the electronic timer. Use low mA setting to test time cutoff. Use high mA setting to test for mAs cutoff.
Analysis:
Exposure should terminate before 6 seconds or 600 mAs, whichever occurs first.
3.2
Reproducibility
Aim:
To assess the variation in radiation output, average kVp and exposure time for a number of exposures of the same object in AEC mode.
Equipment:
20 cm water phantom or 2 mm copper Dosimeter (small volume chamber) and electronic timer or non-invasive beam analyser Three loaded radiographic cassettes Densitometer
Factors:
80 kVp, 200 mA
Method:
Position phantom on film receptor. Set AEC density to 0. Set SID and focal spot to clinical conditions. Place dosimeter and electronic timer on tube side of phantom. Expose cassette and process film. Repeat twice.
Analysis:
Record density of each radiograph. Calculate average and standard deviations for optical density, input exposure and exposure time. Coefficient of variation must not exceed 0.05. Variation in optical density must not exceed 0.1 at a density of 1.2.
Note:
Taping a metal washer, or similar, to the cassette provides a consistent point for density measurements.
3.3
Response time
Aim:
To measure the minimum response time of the AEC.
Equipment:
Dosimeter (small volume chamber) and electronic timer or non-invasive beam analyser Six sheets of perspex, each 24 cm x 30 cm x 1cm thick
Factors:
High kVp, high mA (e.g. 120 kVp, 200 mA)
Method:
Position dosimeter/timer or beam analyser on image receptor. Place perspex sheets in front of detector. Note exposure time and output. Repeat exposures, decreasing perspex thickness each time. Keep decreasing perspex thickness until minimum time is reached, i.e. when the time no longer changes as perspex is removed.
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 6
Analysis:
Minimum response time must not exceed 0.02 seconds.
Note:
Ensure AEC chamber is not obscured by beam analyser.
3.4
Exposure timer
(a) Accuracy Aim:
To determine how the exposure time compares with the selected time.
Equipment:
Digital timer or non-invasive beam analyser
Factors:
Fixed kVp, fixed mA, (e.g. 70 kVp, 200 mA) variable time
Method:
Position digital timer or beam analyser at the distance recommended by the manufacturer. Collimate to size of detector. Make a series of exposures commencing at the shortest exposure time, then across the range of the timer at commonly used settings up to 1 second.
Analysis:
For intra-oral dental apparatus, measured times must be within 10% or 20 milliseconds of the indicated value, whichever is the greater, for times within normal working range. For all other apparatus measured times must be within 5% of nominal values.
Notes:
For half-wave rectified dental apparatus the exposure time must be determined from the time the kV waveform first rises to 75% of the kVp until it falls below this value. This test is not required for apparatus where mAs is selected as a single component.
Equipment:
Digital timer Lead backing sheet
Factors:
70 kVp, 20 mAs
Method:
Position digital timer in primary beam at the distance from the focal spot specified by the manufacturer. Place lead sheet under digital timer to absorb backscatter. Collimate beam to size of detector. Make a minimum of five exposures. Check commonly used times.
Analysis:
Use statistical function on calculator to calculate average and standard deviation. Calculate coefficient of variation (C): C = standard deviation average
Notes:
If a unit fails output reproducibility other measurements may be meaningless. If an appropriate calculator is not available an approximation may be obtained using the formula: C = (max. value min. value) average 2
For older apparatus with mechanical switching, turn controls from set positions between exposures and then return to original settings. This test is not required for apparatus where mAs is selected as a single component. This test may not be necessary if reproducibility of radiation output is acceptable.
3.5
Equipment:
Electronic timer or non-invasive beam analyser
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 8
Factors:
Low kV, mA, long exposure time (e.g. 1 second)
Method:
Position timer in the primary beam at 50 cm from focus. Initiate exposure and release switch before exposure is terminated.
Analysis:
Radiation emission must cease when switch is released. Measuring instrument will indicate time when exposure is terminated.
Equipment:
Digital kV meter
Factors:
Variable kVp, fixed mA, fixed time
Method:
Position kV meter at the distance recommended by the manufacturer. Collimate to size of detector. Observe radiation waveform at 70 kVp and obtain hardcopy if possible. Make a series of exposures across the clinically used range.
Analysis:
Check radiation waveform for any unusual characteristics that may affect subsequent measurements. Measured kVpavg must be within 5% of the nominal value. For capacitor discharge apparatus kVmax must be within 5% of nominal value. For dental apparatus the measured KVpavg must be within 5KVp of nominal value.
Notes:
Do not use times below 0.1 seconds. If using a non-invasive beam analyser use low sensitivity unless results dictate otherwise.
4.2
Reproducibility
Aim:
To determine the variation in average kVp over a number of exposures at the same generator setting.
Equipment:
Digital kV meter Lead backing sheet
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 10
Factors:
Clinical setting, for example 70 kVp, 20 mAs
Method:
Position kV meter in primary beam at the distance from the focal spot specified by the manufacturer. Place lead sheet under meter to absorb backscatter. Collimate beam to size of detector. Make a minimum of five exposures.
Analysis:
Use statistical function on calculator to calculate average and standard deviation. Calculate coefficient of variation (C): C = standard deviation average
Notes:
If a unit fails output reproducibility other measurements may be meaningless. If an appropriate calculator is not available an approximation may be obtained using the formula: C = (max. value min. value) average 2
For older apparatus with mechanical switching, turn controls from set positions between exposures and then return to original settings. This test may not be necessary if reproducibility of radiation output is acceptable.
Equipment:
Dosimeter (small volume chamber) or non-invasive beam analyser Lead backing sheet
Factors:
70 kVp, variable mA, 0.1 s or 70 kVp, variable mAs
Method:
Position ion chamber in primary beam at 75 cm from focal spot (record actual distance) or beam analyser at distance specified by manufacturer. Set dosimeter to Exposure mode (Total or Integrate). Place lead sheet under chamber to absorb backscatter. Collimate beam to size of chamber. Measurements should be made at as many mA or mAs stations as practicable.
Analysis:
Calculate Gy/mAs (X) by dividing output by the nominal mAs. Determine Xmax and Xmin. Calculate linearity coefficient: linearity coefficient = Xmax Xmin Xmin + Xmax
Notes:
kVp should be measured at each mA station to assess kVp compensation. Linearity should be measured for both/all focal spot(s) sizes as Gy/mAs may vary. This test does not directly check if mA stations have been correctly calibrated. For mammography, appropriate kVp and distance factors should be used.
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 12
5.2
Programmed timer
Aim:
To determine the linearity of the radiation output over a range of clinically used timer settings.
Equipment:
Dosimeter (small volume chamber) or non-invasive beam analyser
Factors:
Full range of clinically used timer settings
Method:
Make a minimum of five exposures. Calculate coefficient of linearity using the formula in 5.1 above (if time is not displayed on control, use time chart supplied by manufacturer).
Analysis:
Coefficient of linearity must not exceed 0.1.
Notes:
Readings should be taken on the different patient type settings with all other factors constant, to ensure satisfactory operation. Readings should be taken on each film type setting with all other factors constant, to ensure satisfactory operation.
5.3
Reproducibility
Aim:
To determine the variation in radiation output over a number of exposures at the same generator setting.
Equipment:
Dosimeter and appropriate ionisation chamber or non-invasive beam analyser Lead backing sheet
Factors:
70 kVp, 20 mAs for intra-oral dental: fixed kVp, fixed mA, 0.1 seconds.
Method:
Position the ion chamber in the primary beam at 75 cm from focal spot (record actual distance) or beam analyser at the distance specified by the manufacturer. Set dosimeter to Exposure mode (Total or Integrate). Place lead sheet under chamber to absorb backscatter. Collimate beam to size of chamber. Make a minimum of five exposures.
Analysis:
Use statistical function on calculator to calculate average and standard deviation. Calculate coefficient of variation (C): C = standard deviation average
Notes:
If a unit fails output reproducibility other measurements may be meaningless. For older apparatus with mechanical switching turn controls away from set positions between exposures and return to original settings. If an appropriate calculator is not available an approximation may be obtained using the formula: C = (max. value min. value) average 2
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 14
Equipment:
Loaded cassette Collimator alignment test tool or metal markers to delineate edges of light field Perpendicularity test tool
Factors:
50 kVp, 10 mAs or similar low factors depending on film/screen combination
Method:
Position cassette at 1 m SID, adjust light field to alignment marks on test grid, or collimate to approximately two-thirds of cassette size and mark the edges of the light field. Mark cathode end of tube for orientation. Place perpendicularity test tool in centre of the light field.
Analysis:
The irradiated area should be within 1% of the distance from the focus-to-image receptor in all directions. The image of the upper bead in the perpendicularity test tool should fall within 5 mm of the lower bead or ring.
Notes:
Repeat with collimator rotated 90. Repeat for each focus. This method can also be used to assess alignment of primary beam with bucky tray by placing a second cassette in the bucky tray and exposing simultaneously using an exposure time greater than 0.1 seconds. X-ray assembly and collimator should be visually inspected to assess perpendicularity before starting alignment test.
6.2
Fluoroscopic apparatus
Aim:
To ensure that the primary beam is confined to the image receptor area.
Equipment:
Loaded radiographic cassette or non-screen film Masking tape Tape measure
Factors:
Low or ABC
Method:
Maximise SID. Ensure collimators are fully open. Place cassette or non-screen film as close as possible to image intensifier (II) input surface. Expose cassette under ABC for 12 seconds to produce an optical density of approximately 1.0. Process film and measure dimensions of the x-ray field. Record dimensions of image on TV monitor. Repeat for all II field sizes. Repeat with minimum SID. Compare area exposed on film with area viewed on the TV monitor at both maximum and minimum SID for all II field sizes.
Analysis:
The primary beam must not fall outside the image receptor under any circumstances. The collimator must automatically limit the primary beam to the field size selected regardless of SID. The field size viewed on the TV monitor should be within 10% of the selected field size. The primary beam must be centred to the II field; the centre of the II field must appear as the centre of the image on the TV monitor.
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 16
6.3
Equipment:
Radiographic film of appropriate size
Factors:
Low
Method:
Place cone in direct contact with film surface and expose.
Analysis:
Beam diameter must not exceed 60 mm. If beam is not circular, maximum diagonal dimension must not exceed 60 mm.
Notes:
Where processing facilities are not available a fluorescent screen may be used to assess beam size.
6.4
Panoramic apparatus
Aim:
To ensure that the dimensions of the primary beam at the secondary collimator do not exceed, yet adequately cover, the dimensions of the slot of the secondary collimator.
Equipment:
Two pieces of radiographic film of a size large enough to cover the slot in the secondary collimator
Factors:
Low
Method:
Place one film behind the slot of the secondary collimator and the second film in front, ensuring the entire slot is covered and the film cannot move during exposure; mark each film for identification. Expose film (a full scan is not necessary).
Analysis:
Superimpose the processed films, aligning the slit images; the dimensions of the slit on both films should be equivalent.
Notes:
Where it is difficult to perform this test on an apparatus, it is sufficient to measure the radiation field with the front film only, provided the film can be directly related to the slit, via the use of orientation markers.
6.5
Cephalometric apparatus
Aim:
To ensure that the dimensions of the primary beam at the image receptor plane do not exceed the prescribed dimensions.
Equipment:
350 mm x 350 mm film or larger
Factors:
Low
Method:
Place film at receptor plane and expose.
Analysis:
Where possible the dimensions of the image should not exceed 180 mm x 240 mm and in all cases must not exceed 240 mm x 300 mm. In certain circumstances use of a larger field may be approved after the apparatus has been assessed by a CRE and the details of that assessment have been provided to the Authority. The Authority must be satisfied that such a practice does not cause a safety risk to any person.
6.6
Equipment:
Light meter
Method:
With collimator fully open, turn on light. Position light meter detector at 1 m from focal spot. Take reading in each quadrant of light field to determine uniformity.
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 18
Analysis:
The illuminance of the light field must be not less than 100 lux at 1 m.
Notes:
Ensure detector is parallel to anode-cathode axis. Ambient room light should be as low as possible.
6.7
Equipment:
Dosimeter (large volume chamber)
Factors:
Maximum kVp, minimum mAs
Method:
Position chamber 5 cm below face of collimator with shutters fully open. Take one reading immediately on charging the unit. Take one reading immediately after discharge.
Analysis:
Neither reading must exceed 0.02 mGy/hr.
Equipment:
Dosimeter (small volume chamber) or non-invasive beam analyser Lead backing sheet Aluminium filters (Grade 1100) Tape for positioning filters
Factors:
70100 kVp, fixed mAs
Method:
Remove all optional or easily removable filtration. Position ion chamber in primary beam at least 75 cm from focal spot or beam analyser at distance specified by manufacturer. Set dosimeter to Exposure mode (Total or Integrate). Place the lead sheet under the chamber to absorb backscatter. Collimate the beam to the size of the chamber. Make three exposures with no filters added (free in air), then take the average. Tape 1 mm of the aluminium filter on the face of the collimating device. Repeat exposures with 2, 3 and 4 mm aluminium filters.
Analysis:
Plot exposure against thickness of filter. Halve the average free in air exposure and determine corresponding thickness of aluminium from graph. The HVL must not be less than the values indicated in Table 7.1.
Notes:
kVp should be checked before HVL assessment. If using a non-invasive beam analyser only two measurements are usually required. The HVL will be calculated following the second measurement, provided the tube voltage and exposure time have not varied significantly. Any removable filters should be removed prior to HVL determination.
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 20
For undertable fluoroscopic tubes, position the chamber midway between table top and image receptor and place aluminium filters on the table top under the chamber. Ensure entire beam is intercepted by filters. For intra-oral dental apparatus, aluminium filters should be taped over the open end of the cone.
TABLE 7.1 HALF-VALUE LAYERS X-ray tube voltage (kVp) <50 50 60 70 80 90 100 110 120 130 140 150 >150 Minimum HVL (mmAl) * 1.5 1.8 2.1 2.3 2.5 2.7 3.0 3.2 3.5 3.8 4.1 *
Equipment:
Dosimeter (large volume chamber) Fluorescent screen or non-screen film
Factors:
Maximum kVp, 100 mAs (time should not exceed 1 second). Ensure tube rating is not exceeded.
Method:
Collimator should be fully closed or covered with 20 HVL of lead. Position chamber 1 m from focal spot and make exposures at positions, including cathode, anode and front of tube assembly. Calculate time averaged leakage using tube duty cycle or cooling curve data.
Analysis:
For dental apparatus used with intra-oral image receptors, leakage radiation must not exceed 0.25 mGy in an hour at a distance of 1 m from the focus. For all other apparatus, leakage radiation must not exceed 1.0 mGy in an hour at a distance of 1 m from the focus, taking into account the tube duty cycle.
Notes:
An incorrectly positioned x-ray tube insert or flaws in the lead shielding in a housing may give rise to narrow but intense beams of leakage radiation which fail to ionise the entire chamber and therefore appear not to exceed the specified limit; such beams are highly undesirable and the cause should be remedied. Pinhole leaks or hotspots can be detected by the use of a fluorescent screen or nonscreen film wrapped around the x-ray tube assembly.
8.2
Equipment:
Dosimeter (small volume chamber) or beam analyser
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 22
Factors:
Maximum kVp and mA
Method:
Minimise SID. Place chamber at the position appropriate for that apparatus (see Table 8.1). Set dosimeter to Rate mode. Ensure chamber is central to the x-ray field (use ABC to check). Protect image intensifier by covering with at least 2 mm of lead. Irradiate chamber at maximum kVp and mA until dosimeter reading stabilises.
Analysis:
For all systems where Automatic exposure control (AEC) is not provided, and for systems used exclusively for paediatric aims, the absorbed dose in air must not exceed 50 mGy/min. Where AEC is provided the absorbed dose in air must not exceed 100 mGy/min. Where a boost or high dose facility is activated the absorbed dose in air must not exceed 150 mGy/min. Where a boost or high dose facility is activated it must be restricted to a maximum of 20 seconds, after which it must return to normal fluoroscopic mode
Notes:
Ensure that the II is completely covered with lead to avoid damage.
TABLE 8.1 DETECTOR POSITIONS FOR MEASUREMENT OF MAXIMUM ABSORBED DOSE IN AIR Condition 1 2 X-ray tube permanently UNDER table Image Intensifier permanently UNDER table C or U arm apparatus (no permanent table) X-ray tube and image intensifier moveable around table Detector Position ON the table 300 mm ABOVE table top
300 mm from image intensifier but not less than 400 mm from focal spot 400 mm from focal spot
8.3
Scatter radiation
(a) Radiographic/fluoroscopic apparatus Aim:
To determine that sufficient shielding is provided for all occupied areas.
Equipment:
Dosimeter (large volume chamber) 20 cm water phantom
Factors:
100 kVp, 100 mAs (or similar factors)
Method:
Collimate primary beam to phantom size. Position chamber at 1 m horizontally from centre of phantom and take measurement (use this as a reference). Take a measurement immediately outside control area. Take a measurement inside control area. Repeat for other areas where adequacy of shielding is to be determined.
Analysis:
Determine weekly dose based on workload. Shielding should ensure that radiation dose does not exceed the limits prescribed in Schedule 2 of the Radiation Control Regulation 2003.
Notes:
Beware of backscatter. Ensure that the dimensions of the protective shield are adequate to fully intercept leakage radiation from the x-ray tube assembly and scattered radiation from the patient.
Equipment:
Dosimeter (large volume chamber) Appropriate body phantom
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 24
Factors:
Maximum kVp, mA and scan time 10 mm CT slice thickness
Method:
Position phantom centrally within scan field. Position chamber 1 m from centre of field at 45 to phantom and take measurement. Take a measurement at the end of the couch. Take a measurement immediately outside control area. Take a measurement inside control area. Repeat for other areas where adequacy of shielding is to be determined.
Analysis:
For measurement of scatter radiation in CT facilities a phantom with minimum diameter of 360 mm must be used; clinical abdominal scanning parameters should be employed. Determine weekly dose based on workload and average scans per patient. Shielding should ensure that radiation dose does not exceed the limits prescribed in Schedule 2 of the Radiation Control Regulation 2003. Measurements should approximate isodose curves supplied by manufacturer.
Equipment:
Line pair gauge 20 cm water phantom
Factors:
ABC or 70 kVp, 1 mA
Method:
Place line pair gauge directly onto centre of image receptor surface at 45 to the grid and to the raster lines. Collimate to gauge, place the water phantom in the beam and note the number of line pairs visible on the monitor. Monitor should be viewed from a distance of 4 times the screen diameter, in ambient light.
Analysis:
The high contrast resolution of the live image must not be less than the following values:
TABLE 9.1 HIGH-CONTRAST RESOLUTION Apparatus New Field size (cm) 11 to < 18 18 to < 26 26 to <30 30 to 36 >36 Existing 25 > 25 Resolution (line pairs/cm) 18 16 14 12 10 12 10
Notes:
This is a subjective test that can be affected by room lighting, monitor contrast and brightness settings and orientation of test gauge. This test must be performed without any time integration or image enhancement.
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 26
9.2
Equipment:
Westmead test object (or equivalent) 20 cm water for acceptance testing of new apparatus 2 mm copper for routine QA
Factors:
ABC or 70 kVp, 1 mA
Method:
Set SID at normal operating distance or 100 cm. Place test object directly onto centre of image receptor at 45 to the raster lines. Place water or copper in the beam on the table top or on the test object. Monitors should be viewed from a distance of four times the screen diameter in ambient light. Adjust monitor brightness and contrast settings to optimum (see Notes below). View all field sizes and record the following: (a) field size (b) distortion note any S or pin cushion distortion (c) contrast threshold: the number of large circles detectable on the live image (d) contrast detail the number of hole sizes visible from the contrast detail portion of the test object on the live image.
Analysis:
Field size should be within 1 cm of nominal size. The low contrast threshold must not be less than 4%. Minimum contrast detail for the clinical field size used should not be less than the following: (a) general fluoro 6 holes (1.5 mm) (b) interventional application 7 holes (1.0 mm).
Notes:
These are subjective tests which can be affected by room lighting, monitor contrast and brightness settings and orientation of test object. This test must be performed without any time integration or image enhancement. If possible, hard copy images should be taken. Optimal adjustment of TV monitor is best done using the following method: (a) turn contrast to zero (b) turn brightness until image is just not visible (c) adjust contrast until both circles are visible. As most image intensifiers do not display an area as large as quoted, the manufacturer should be consulted for the exact size.
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 28
SECTION 10MEAN CT NUMBER, NOISE AND UNIFORMITY 10.1 Mean CT number, noise and uniformity
Aim:
To determine the mean CT number, noise and uniformity of an image.
Equipment:
CT phantom
Factors:
Appropriate clinical factors
Method:
Position phantom in centre of gantry within scan field. Scan phantom. Select a region of interest (ROI) at the centre of the image and determine the mean CT number and standard deviation of pixels in ROI (noise). Compare the mean CT numbers of the ROI in the central position with those in the outer ROI. Repeat at three other positions within ROI.
Analysis:
Noise should not deviate from baseline values by more than 10% or 0.2 Hounsfield unit (HU) whichever is larger. Mean CT number in the central region of interest should fall within " 4 HU of the baseline value. The difference between the mean CT numbers of the central ROI and the outer ROI should not vary by more than 2 HU from those of the baseline values.
Notes:
All significant settings of radiation apparatus and accessories should be recorded to ensure duplication in future testing. Select positions in the ROI corresponding to 3, 6, 9 and 12 oclock to allow for easy duplication. A decrease in noise may be the result of an unnoticed increase in dose. Most CT systems have in-built software which performs the above tests using a special calibration phantom.
Equipment:
Dosimeter (small volume chamber) Lead backing sheet
Factors:
Nominal factors for standard size patient (70 kg)
Method:
Position chamber in centre of beam at appropriate focus-to-skin distance. Place lead backing sheet under chamber. Use standard collimation. Initiate exposure and record results.
Analysis:
The skin entrance dose should not exceed the reference values in Table 11.1.
Notes:
The skin surface entrance doses in Table 11.1 are in air with backscatter for conventional film-screen combinations up to 200. For faster film-screen combinations (400600) the values should be reduced by a factor of 23. Doses in excess of the guidance levels should be based on sound clinical judgement.
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 30
TABLE 11.1 DIAGNOSTIC REFERENCE LEVELS FOR RADIOGRAPHIC EXAMINATIONS FOR A STANDARD SIZE PATIENT (70 KG) Surface entrance dose per radiograph with backscatter (mGy) AP Lat LSJ Chest PA Lat Pelvis Abdomen, IV urography & cholecystography Skull AP AP 10 25 30 0.4 1 8 8
AP PA Lat
5 5 3
PA AP Lat LSJ
11.2
Fluoroscopic examinations
Aim:
To evaluate the patient dose for typical fluoroscopic examinations.
Equipment:
Dosimeter Standard patient equivalent phantom (20 cm water)
Factors:
ABC or manual factors commonly used
Method:
Position chamber on table in centre of field. Using appropriate supports, place phantom above chamber. Adjust collimator so edges are just visible on monitor. Set dosimeter to Rate mode. Activate image intensifier long enough for dosimeter reading to stabilise. Record exposure rate in all modes.
Analysis:
Dose rates should not exceed the guidance values in Table 11.2.
Notes:
Values in Table 11.2 include backscatter and relate to a 23 cm field size. Record indicated mA and kVp, collimation, carriage height, etc. for future reference. For overtable tubes or C-arm systems the chamber should be positioned as per Table 8.1 and source-to-chamber distance should be noted.
TABLE 11.2 DOSE GUIDANCE LEVELS FOR FLUOROSCOPIC EXAMINATIONS Surface entrance dose rate* (mGy/min) 17
Mode of operation Normal II field size >17cm Normal II field size 17cm High Level * in air with backscatter
25
100
11.3
CT examinations
Aim:
To evaluate the scanner output for typical CT examinations by measuring the CT Dose Index (CTDI)1
EC, Report EUR 16262, European Guidelines on Quality Criteria for Computed Tomography, 1999.
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 32
Equipment:
Pencil ion chamber with at least 10 cm length of active volume CT dosimeter phantom (see Notes)
Factors:
Appropriate clinical factors for the procedures to be tested, including pitch
Method:
Centre and align the appropriate CTDI phantom in the gantry with long axis within 2 mm of scanner centre line in both horizontal and vertical planes and CT scan plane in the centre of phantom. Scan to check alignment. Insert the ion chamber in the centre of phantom and repeat at peripheral locations. Scan a single axial CT rotation without movement of the table. Calculate the CTDI, CTDIw and CTDIvol
CTDI =
c.L.D T
where c = ion chamber calibration factor L = length of chamber active volume T = CT slice thickness D = measured dose from single scan (mGy).
Analysis:
The CTDI should not exceed the guidance values in Table 11.3.
TABLE 11.3 SOME DIAGNOSTIC GUIDANCE LEVELS FOR CT PROCEDURES2 Examination Routine Head Routine Chest Routine Abdomen Routine Pelvis CT Dose Indexw (mGy) Dose Length Product (mGy cm) 60 30 35 35 1050 650 780 570
Nagel HD, ed. Radiation Exposure in Computed Tomography: Fundamentals, Influencing Parameters, Dose Assessment, Optimisation, Scanner Data, Terminology. 4th ed. Hamburg: CTB Publications, 2002.
Notes:
A CTDI phantom is a 1416 cm long acrylic (polymethyl methacrylate) cylinder with diameter 16 cm (head), 32 cm (body) or 8 cm (paediatric). Holes are drilled for placement of the pencil ion chamber. One hole is at the centre and four others located at a depth of 1 cm below the surface of the phantom at the 3, 6, 9 and 12 oclock positions. Acrylic rods are inserted into all holes except at the position where the dose is to be measured. CTDIc is the CTDI measured at the central location while CTDIp is the average CTDI measured at the peripheral locations. CTDIw = 1/3 CTDIc + 2/3 CTDIp. CTDIvol = CTDIw/p where p = pitch. Dose length product (DLP) = CTDI x scan length (mGy.cm) for an examination. An alternate measurement, the CTDI in air, does not require a phantom as the dose is measured using the pencil ion chamber placed in-air at the gantry centre of rotation. This method may be used in conjunction with software to determine the effective dose, provided the scan regions of the patient are accurately known. Because of the expense of the CTDI phantom, CTDI air should be adequate for baseline constancy tests.
11.4
Equipment:
Dosimeter (small volume chamber) or beam analyser
Factors:
Bite-wing exposure
Method:
Position chamber in primary beam at 10 mm from end of cone and expose.
Analysis:
Skin entrance exposure for a bite-wing should be in the range of 23 mGy and must not exceed 5 mGy for any intra-oral radiograph.
Notes:
Ensure entire active volume of chamber is irradiated.
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 34
Equipment:
Pin-hole camera (focal spot > 0.3 mm) or star test pattern (focal spot 0.3 mm)
Factors:
75 kVp, 10 mAs
Method:
As per test tool manufacturers instructions.
Analysis:
Determine apparent focal spot size as per instructions. Tolerance limits are given in Table 12.1.
Notes:
Measurement should be performed for each focus.
TABLE 12.1 FOCAL SPOT TOLERANCES Nominal size of focal spot (mm) <0.8 0.8 1.5 >1.5 Tolerance minus 0 0 0 Tolerance plus 50% 40% 30%
12.2
Image viewing
Aim:
To ensure the level of light intensity emitted by the viewing box is adequate and sufficiently uniform.
Equipment:
Light meter with luminance detector
Method:
Visually check for any inconsistencies in colour or brightness. Place detector over centre of viewing box and record luminance. Repeat in each quadrant of viewing box.
Analysis:
The minimum luminance in the centre and each quadrant of the viewing box should be >1000 nit (candela.m-2). The five luminance levels should be within 10% of mean value. The colour of the viewing box should be white or blue and should be consistent throughout a complete set.
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 36
DEFINITIONS
In this Guideline: ABC means automatic brightness control Act means the Radiation Control Act 1990 AEC means automatic exposure control Absorbed dose means energy delivered from radiation per unit mass of absorbing material, measured in Gray (Gy) or mGy. One Gray equals one joule per kilogram Added filtration means quantity indicating the filtration affected by added filters in the useful beam, but excluding inherent filtration Air kerma means kerma measured in a mass of air Authority means the NSW Environment Protection Authority Coefficient of variation means the standard deviation divided by the mean of a set of numbers Coefficient of linearity equals ((Xmax-Xmin)/(Xmax +Xmin)) where Xi = Gy/mAs for exposure setting 'i' CRE means consulting radiation expert CT means computed tomography CTDI means CT dose index, the integral of the dose profile along a line perpendicular to the tomographic plane from 7T to +7T (where T is the nominal slice thickness), divided by the product of the nominal slice thickness and the number of tomograms (N) produced in a single scan Filtration means modification of the spectral distribution of an x-ray beam as it passes through matter by the differential absorption of poly-energetic photons Fluoroscopic apparatus means radiation apparatus that emits ionising radiation, as defined in the Act, used for the purpose of fluoroscopy or radioscopy. (It should be noted that Standards Australia and the International Electrotechnical Commission have adopted the term radioscopic, but for the purposes of this document the term fluoroscopic is used.) Focal spot means the area of the target from which x-rays are emitted Half-value layer (HVL) means the thickness of a specified material that reduces the exposure rate of a given x-ray beam to half its original value Inherent filtration means the filtration affected by the irremovable materials of an x-ray tube assembly (i.e. glass, oil and port seal), through which the radiation beam passes before emerging from the x-ray tube assembly. It is expressed in terms of thickness of a reference material that, at a specified potential difference and waveform, gives the same radiation quality in terms of half-value layer Kerma (K) means Kinetic Energy Released in a material by ionising radiation and is determined as the quotient of dEtr by dm, where dEtr is the sum of the initial kinetic energies of all the charged ionising particles liberated by uncharged ionising particles in a material of mass dm (K = dEtr/dm). The unit of kerma is the gray (Gy), or joule per kilogram
Radiation Guideline 6: Registration requirements & industry best practice for ionising radiation apparatus used in diagnostic imaging 38
Kerma rate means kerma per unit time and is determined as the quotient of dK by dt, where dK is the increment of kerma in the time interval dt Optical density (OD) means the degree of film blackening produced during development, where optical density is the log of the reciprocal of the fraction of light transmitted through the blackened film Operator means a person licensed under Section 6 of the Act to use ionising radiation apparatus owner means the owner of the radiation apparatus to which Section 7 of the Act applies Phantom means the test object that simulates the average composition of various structures Primary beam means all ionising radiation that emerges through the specified aperture of the protective shielding of the x-ray tube and the collimating device Radiographic apparatus means ionising radiation apparatus, which emits ionising radiation, used for the purpose of radiography Radiation leakage means ionising radiation transmitted through the protective shielding of a radiation source other than the primary beam Radiation quality means a characteristic of radiation determined by its spectral distribution that controls its penetration. It depends on the magnitude and waveform of the tubes potential difference and the total filtration, and is described in part by its half-value layer. Regulation means the Radiation Control Regulation 2003 Scattered radiation means ionising radiation produced from the interaction of electromagnetic ionising radiation with matter. It has a lower energy than or a different direction from that of the original incident ionising radiation. SID means source-to-image receptor distance. This is the same as focus-to-image receptor distance and focus-to-film distance. Target means the area of the anode which is struck by the electrons from the cathode Total filtration means the sum of inherent filtration and added filtration between the radiation source and the patient or other defined plane X-ray tube assembly means the x-ray tube housing with an x-ray tube insert, but not including a collimating device X-ray tube housing means a container in which an x-ray tube is mounted for normal use, providing protection against electric shock and against ionising radiation except for an aperture for the useful beam. It may contain other components. X-ray tube insert means a highly evacuated vessel for the production of x-radiation by the bombardment of a target, usually contained in an anode, with a beam of electrons accelerated by a potential difference X-ray tube potential difference means the peak value of the potential difference applied to the x-ray tube, expressed as kilovolts peak (kVp) Unless otherwise defined, all words in this Guideline have the same meaning as in the Act and the Regulation