Isotope Dilution in Chemical Analysis Introduction are not well behaved during chromatography or sample extraction. Variability caused by such problems is usually compensated for in part or monitored by using internal standards and surrogate analytes. An isotope dilution standard is the "perfect" internal standard or surrogate.

The technique of isotope dilution is


being used more and more in our lab to improve precision and accuracy by reducing the problems with calibration and sample preparation matrix effects. Some analytes

Isotope Dilution Standards An internal standard or surrogate is a compound similar to the analyte of interest. An isotope dilution standard is an isotope of an element or a molecular compound labeled with an isotope. Good examples of both of these are 204Pb and benzene-d6. 204Pb is a minor isotope of lead (Pb). The Pb isotopes which are stable and occur naturally have atomic masses of 204

Pb was in the sample. There are a couple of catches, because nothing is ever "perfect". One catch is that you need to use a mass spectrometer to separate analyte masses. A second catch is that enriched isotopic standards are expensive. The other catches are beyond the scope of this article. In the example benzene-d6, the hydrogens (H) normally attached to the

(1.4%), 206 (24.1%), 207 (22.1%), and 208 (52.4%) where the natural abundances appear in parenthesis. By adding a known amount of 204Pb to a sample before testing for total Pb and by testing for each of the lead isotopes, one can accurately determine how much total

benzene ring are replaced by deuteriums (D), an isotope of hydrogen (2H), thus the -d6. In both these cases, the isotopes have almost exactly the same chemical properties as the original or naturally occurring isotopes. They just differ in mass. This is as close as it gets to a "perfect" internal standard or surrogate.

Minimizing Errors To give you an idea of how much the error can be reduced in a chemical analysis, consider the following. In a gas chromatography analysis, the error from injection can be approximately 5%. By adding an internal standard, the error can be reduced by half to approximately 2-3%. With isotope dilution, the error should be half of that again, approximately 1%. In the analysis of sub-ppm levels of lead (Pb) in supplements, errors from traditional approaches using ICPMS and GFAA can be 25% or more, characteristic of a trace analysis in a solid matrix. With isotope dilution, the error is generally less than 5%. Where ever a mass spectrographic technique can be used for analysis, isotope dilution should

Because of this, NIST commonly uses isotope dilution mass spectrometry for certifying reference samples. Below are links to other articles on isotope dilution techniques using either gas chromatography-mass spectrometry (GCMS) or inductively coupled plasma-mass spectrometry (ICPMS).

Lead in Supplements by ICPMS Boron by ICPMS 1,4-Dioxane by GCMS Ethylene Glycol and Ethylene Chlorohydrin by GCMS Volatile Organics Analysis by EPA 1666 (pdf file) For a quotation...

minimize the error.


For a full listing of Exova services, please visit our new Exova web site, Exova Pharmaceutical Solutions web site and our new Santa Fe Springs laboratory home page.

9240 Santa Fe Springs Rd Santa Fe Springs, CA 90670 562.948.2225 Fax 562.948.5850 info400@exova.com

Formerly West Coast Analytical Service (WCAS) and Bodycote Testing Group