Vous êtes sur la page 1sur 3

Items to be purchased in Scientific Metrology Division, NBSM for FY-2069/70 Prepared on March 4, 2013 Item I, Laptops Hard Drive

: >500GB SATA 5400rpm Display: > 15.6 inch, Led Processor: > Intel 2nd Gen Core i5 , Speed - >2.3GhZ, Bus Speed : >1333 MHz Graphics Memory: 1GB Memory: > 4GB DDR3 Speed: >1333 MHz Hard Drive Capacity: > 500GB Serial ATA - 5400.0 rpm Operating System: Microsoft Windows 8 64 bit , Microsoft Windows 7 Professional 64 bit Network & Communication : WAN, Bluetooth, Ethernet, USB, HDMI Power: >6 hours

Item 2, Low temperature liquid baths Temperature ranges from 30C to 150C Stability to 0.02 C Uniformity: 0.05 C Accuracy: 0.3C Calibration Depth: not less than 300 mm Computer communications: USB /RS-232/LAN/WAN

Item 3, Thermometric Bridge (For calibration of SPRTs, Thermistors etc) Measuring Range Temperature: -50 to 600 C ; Resistance: 0 to 400 Resolution Temperature: 0.1 mK (Pt 100) Resistance: 50 m Measurement Uncertainty: Within the values of its resolution (95% Confidence Level & at (202)C Sensor Connection: 4 wire Number of sensor inputs: 2 Computer communications: USB/RS-232/LAN/WAN (Calibration Certificates from laboratories accredited as per ISO/IEC 17025 required) Assumed Cost: $8000 Item 4: Laboratory Ice Cube makers Required No. : 1 Production Capacity >30 Kg/24 Hours Storage Capacity > 1 Year Accessories Ice Cube Crusher (Mechanical) Assumed Cost: $1000

Item 5: Data Logger with humidity and temperature sensor : 1 set


1/3

Items to be purchased in Scientific Metrology Division, NBSM for FY-2069/70 Prepared on March 4, 2013 Temperature: Thermistor Probe Measurement Range: 40 C to 125 C (40 F to 257 F) Internal Resolution: 0.1 C typ Thermistor Probe Temperature Accuracy (Overall Error): 0.5 C typ, 2.0 C max Alarm Threshold Range (Software-Selectable): 40 C to 125 C (40 F to 257 F) range for both high alarms and low alarms Data Sampling Sampling Rate (Software-Selectable): 10 s, 30 s, 1 min, 5 min, 30 min, 1 hr, 6 hr, 12 hr Temperature Samples: More than 1,000,000 max Temperature Units: C or F Relative Humidity Measurement Range: 0% RH to 100% RH Accuracy (Overall Error Between 20% RH and 80 %RH): 3.0% RH typ Alarm Threshold Range: 0% RH to 100% RH (high and low alarms) Data Sampling Sampling Rate (Software-Selectable): 10 s, 30 s, 1 min, 5 min, 30 min, 1 hr, 6 hr, 12 hr Combined Temperature and Humidity Samples: More than 500,000 max Temperature Units: C or F Other requirements Rechargeable battery-powered: Internal lithium polymer battery (rechargeable using included microUSB cable) Internal Storage: Logs more than 1 million temperature samples and more than 500,000 temperature/humidity samples LCD Display: Integrated LCD screen displays current readings, min/max readings, high and low alarm occurrences, and battery charge status Connectivity: USB, WiFi Software

Item 6: Force Transducer for Calibration of Universal testing machine/ Compressive testing machine (1 Set)
Capacity: 2 MN Class : 0.5 (as per ISO 376) Direction of force: Tensile and Compression Protection: IP67 Normal Sensitivity: 2mV/V Calibration Certificate: from ISO 17025 accredited Laboratory Other requirements: Mounting for tensile and compressive mode, display unit

Item 6: Force Transducer for Calibration of Universal testing machine/ Compressive testing machine (1 Set)
Capacity: 500kN Class : 0.5 (as per ISO 376) Direction of force: Tensile and Compression Protection: IP67 2/3

Items to be purchased in Scientific Metrology Division, NBSM for FY-2069/70 Prepared on March 4, 2013 Normal Sensitivity: 2mV/V Calibration Certificate: from ISO 17025 accredited Laboratory Other requirements: Mounting for tensile and compressive mode, display unit