The Assessment of Four Different Methods to Verify Tracheal Tube Placement in the Critical Care Setting
Sylvia Knapp, MD, Julia Kofler, MD, Brigitte Stoiser, MD, Florian Thalhammer, MD, Heinz Burgmann, MD, Martin Posch, PhD, Roland Hofbauer, MD, Margit Stanzel, MD, and Michael Frass, MD
Department of Internal Medicine I and Department of Medical Statistics, University of Vienna, Vienna, Austria

One of the most serious complications of conventional endotracheal intubation is unidentified placement of the tube in the esophagus. The aim of our study was to evaluate four different methods for immediate detection of the tube position: auscultation, capnographic determination of ETco2, esophageal detection method (EDM) using a self-inflating bulb, and the transillumination method using a lighted stylet (Trachlight; Laerdal, Armonk, NY). Thirty-eight endotracheally intubated patients admitted to our medical intensive care unit were enrolled in the study. A second identical tube was inserted into the esophagus under laryngoscopic control. The endotracheal tube was then disconnected from the ventilator. Two blinded examiners, one experienced, the other inexperienced, determined the tube position within 30 s using one of the four methods. The order of the tubes tested and the methods used were randomized. In 130 of 152 examinations, both examiners correctly diagnosed the position of the tube. The wrong result was obtained by both examiners 4 times; only the experienced examiner was wrong 4 times, and only the inexperienced examiner was wrong 14 times. Using ETco2, both examiners were correct in all cases. Auscultation showed an obvious

relation to the examiners experience: the experienced examiner was correct in all cases, the inexperienced examiner was correct in only 68% of cases. Using the selfinflating bulb, there were two wrong results of the experienced examiner and one wrong result of the inexperienced examiner. The transillumination technique was associated with a high error rate by both examiners (16% and 13%, respectively). Comparing all four methods showed that capnography is superior to auscultation (P 0.0005) and to the Trachlight detection method (P 0.0078). EDM was not statistically superior to auscultation and transillumination. Capnography was the most reliable method for rapid evaluation of tube position, followed by EDM, whereas auscultation and Trachlight did not seem to be of comparable value. Experience was a determining factor for auscultation. Implications: To prevent unidentified esophageal intubation, a serious complication in the critical care setting, four methods for detecting tube position were tested by two examiners (one experienced, the other inexperienced) in endotracheally intubated patients after insertion of a second tube into the esophagus. (Anesth Analg 1999;88:766 70)

utcome studies have repeatedly shown that unrecognized esophageal intubation is a leading cause of death or brain damage in anesthetic practice (1 4). Esophageal intubation as a complication in emergency airway management in critically ill patients occurs in 8% of the attempts (5). Contributing factors to this high incidence include intubation under unfavorable conditions, unavailability of monitoring equipment, violation of the standard technique of auscultation of lung fields and epigastrium, and intubations performed by inexperienced personnel. Thus,

Accepted for publication December 21, 1998. Address correspondence and reprint requests to Sylvia Knapp, MD, Department of Internal Medicine I, Waehringer Guertel 1820, A-1090 Vienna, Austria. Address e-mail to sylvia.knapp@akh-wien.ac.at.

reliable methods to confirm correct tube position are necessary. The methods most commonly used for confirming tube position are clinical evaluation by auscultation of the lungs and the epigastrium; capnographic determination of ETco2; the esophageal detection method (EDM) using a self-inflating bulb (SIB) or syringe; and the tracheal transillumination technique using a lighted stylet. Although various studies have demonstrated the reliability of each of these methods, no study has directly compared them. In addition, the influence of the examiners experience on the use and interpretation of these various detection methods has not been assessed. We therefore conducted a study to compare the reliability of four different methods for immediate detection of the tube position and to assess the influence of the examiner experience.
1999 by the International Anesthesia Research Society 0003-2999/99/$5.00

766

Anesth Analg 1999;88:76670

ANESTH ANALG 1999;88:766 70

CRITICAL CARE AND TRAUMA KNAPP ET AL. ASSESSMENT OF TRACHEAL TUBE POSITION

767

Methods
The study was performed at the Medical Intensive Care Unit of the University Hospital of Vienna after approval of the local ethical review board and in conformance with Austrian law. Thirty-eight consecutive, orotracheally intubated, adult patients requiring prolonged controlled mechanical ventilation were eligible for study inclusion. The correct tube position was verified by a chest radiograph. Hemodynamic and respiratory stability, as well as adequate sedation and analgesia with midazolam and sufentanil, were prerequisites for study entry. Exclusion criteria were severe respiratory failure (fraction of inspired oxygen [Fio2] 0.5, positive endexpiratory pressure 10 cm H2O, and/or inspiratory to expiratory ratio 0.5), need for vasopressors, gastric reflux via the feeding tube of 200 mL 4 h before inclusion, pregnancy, hemorrhagic diathesis, gastroesophageal diseases, and tracheostomy. A second tube (7.59.0 mm inner diameter) was inserted into the esophagus under laryngoscopic control. This tube was identical to the tracheal airway and was always inserted by the same investigator. The cuffs of both tubes were inflated with 8 3 mL of air. The patients head was covered with a sheet so that only the ends of both tubes were visible. The tracheal tube was then disconnected from the ventilator, and the first of two examiners was allowed to enter the room. Examiner 1 was one of two physicians with at least 4 yr experience in critical care; Examiner 2 was one of two medical students in the last year of training with little or no experience in critical care. Thirty seconds was allowed for determination of the tube position using one of the four methods described below. During the study period, patients were ventilated manually by the investigator with an approximate tidal volume of 10 mL/kg body weight at a frequency of 12 breaths/min. In cases in which the esophageal tube was ventilated, the stomach was decompressed afterward. The selection of the tube (tracheal or esophageal) to be tested and the detection method was chosen according to a randomization list. After the experienced examiner left the room, the inexperienced examiner evaluated the same tube using the same method of evaluation. This procedure was repeated until all detection methods were tested by both examiners in a particular patient. During the study, heart rate, blood pressure, and oxygen saturation were monitored continuously. Before and immediately after the last examination, arterial blood gases were checked. In case of respiratory or hemodynamic deterioration, the study was discontinued. Auscultation was performed with patients in the supine position over both lungs in the infraclavicular fossa and the fifth intercostal space in the midaxillary line and over the epigastrium.

ETco2 was determined using a spectrographic/ infrared capnograph (Capnosat; Dra ger, Lu beck, Germany). The examiners looked for the characteristic CO2 waveform. EDM was performed by using the SIB (Tube Chek; Ambu, Glostrup, Denmark; capacity 60 mL), which was compressed before connection to the tube. Rapid reinflation (4 s) of the SIB was indicative of tracheal positioning, whereas delayed (4 s) or no reinflation was considered indicative of esophageal placement. For transillumination, a lighted stylet (Trachlight; Laerdal, Armonk, NY) was inserted into the tube. If a bright, circumscribed glow was seen below the cricothyroid membrane, the tube position was identified as tracheal. If no light or only a very weak light was detectable, the tube was considered to be in an esophageal position. The room was darkened before examination. Continuous variables are presented as means sd, nominal variables are presented as percentages. For every detection technique, an indicator variable was created (1 the examiner was right in the particular patient, 0 the examiner was wrong). Differences in the indicator variable between the two examiners per detection technique were analyzed by using the sign test. For an overall comparison between the examiners, the differences in the indicator variables were summed over the detection techniques. The Wilcoxon signed-rank test was used to compare the examiners. To compare the four methods, the indicator variables were summed over the examiners. All pairwise differences of these aggregated indicators among the four methods were analyzed again using the Wilcoxon signed-rank test. All confidence limits were calculated according to the method of Pearson and Clopper (6). The significance level was set to 0.05. A Bonferroni-Shaffer correction for multiplicity of testing for the differences among the detection methods was performed (7). Sensitivity and specificity were calculated for each method pooled over the examiners.

Results
Thirty-eight patients (62 11 yr; 22 male, 16 female; body mass index 27 5 kg/m2) were included in the study. The main reasons for admittance to the intensive care unit were cardiopulmonary resuscitation (n 16), heart surgery (n 6), and respiratory insufficiency due to various disorders (e.g., pulmonary embolism, pneumonia, pulmonary edema) (n 16). All patients were in a stable respiratory and hemodynamic state at study entry (Fio2 0.4 0.1, positive end-expiratory pressure 5 2 cm H2O, oxygen saturation 96% 3%, heart rate 89 27 bpm, mean

768

CRITICAL CARE AND TRAUMA KNAPP ET AL. ASSESSMENT OF TRACHEAL TUBE POSITION

ANESTH ANALG 1999;88:766 70

Figure 1. Error rates of experienced and inexperienced examiners obtained by evaluating tube placement using one of four methods.

arterial blood pressure 84 17 mm Hg). Blood pressure, oxygen saturation, and heart rate remained stable in all study patients. Arterial blood gas checks did not show major changes before and after the study. In none of the cases did the study have to be discontinued prematurely because of complications. The two examiners performed 152 examinations. Of the tubes tested, 74 were in the tracheal position, and 78 were in the esophageal position. In 130 of 152 examinations, both examiners correctly diagnosed the position of the tube. Four times a wrong result was obtained by both. Four times only the experienced examiner was wrong, and 14 times only the inexperienced examiner was wrong. The indicator variables aggregated over techniques in the 38 patients were significantly different between the examiners (P 0.0129). Using auscultation, the experienced examiner diagnosed the proper tube position in all cases, whereas the inexperienced examiner was correct in only 68% of cases (P 0.0005) (Fig 1). Ten (83%) of these incorrect results were false-positive (esophageal tube identified as tracheal), and only two (17%) were false-negative (tracheal tube identified as esophageal Using ETco2, both examiners were correct in all cases (Fig. 1). Using EDM, in one case, the experienced examiner determined a wrong tube position; in another case, both examiners determined a wrong tube position (Fig 1). All incorrect results were false-negatives (tube positioned in trachea, but SIB failed to reinflate). Using transillumination, the experienced examiner was incorrect in six cases (16%; twice false-positive, four times false-negative), and the inexperienced examiner was wrong in five cases (13%; twice falsepositive, three times false-negative) (Fig. 1). Three times both examiners were incorrect in the same patients.

Comparing the four methods for both examiners, a statistical difference was found between ETco2 and auscultation (P 0.0005) and ETco2 and Trachlight (P 0.0078). The difference between EDM and auscultation (P 0.0225) failed to reach significance when applying the critical limit 0.05/3 after adjustment for multiple testing. The calculated differences between ETco2 and EDM, EDM and Trachlight, and auscultation and Trachlight were not statistically significant. Sensitivity and specificity of the four methods pooled for the two examiners are shown in Table 1.

Discussion
Many methods have been used to distinguish tracheal from esophageal tube placement, and almost all can fail (8,9). One of the most reliable signs of correct tracheal placement is visualization of the tube entering the larynx (8,10). However, complete visualization is sometimes impossible, particularly under emergency conditions. Furthermore, the tube may slip out of the larynx after initial successful tracheal placement. In our study, the use of capnography led to correct diagnoses of tube positioning in all patients tested, once more confirming its reliability in noncardiac arrest patients. However, during cardiac arrest and cardiopulmonary resuscitation, when low cardiac output leads to low expired carbon dioxide concentrations, false-negative results have often been observed. Bozeman et al. (11) reported an error rate of approximately 30% in ETco2 determination in cardiac arrest victims, whereas the EDM correctly indicated tracheal tube placement in all of these patients. The EDM relies on the anatomic differences between the trachea and the esophagus. Application of negative pressure leads to collapse of the esophagus because of its fibromuscular

ANESTH ANALG 1999;88:766 70

CRITICAL CARE AND TRAUMA KNAPP ET AL. ASSESSMENT OF TRACHEAL TUBE POSITION

769

Table 1. Sensitivity and Specificity of the Four Methods Auscultation Sensitivity Specificity 0.74 0.95 ETco2 1.0 1.0 EDM 1.0 0.91 Trachlight 0.9 0.81

EDM esophageal detection method.

structure, whereas the trachea is held open by its rigid cartilaginous rings. Because this anatomic difference persists regardless of the metabolic and perfusion state of the patient, the SIB technique seems superior to capnography in cardiac arrest patients (11,12). Therefore, some authors recommend the EDM for initial assessment of tube position in the preclinical setting, eventually followed by capnometry (13,14). Nevertheless, both methods may lead to false-negative results in the case of tracheobronchial obstruction or pulmonary edema, which may be encountered during emergency intubations (9,15).1,2 In the current study, the SIB failed three times, and a false-negative result was obtained by both examiners in one patient. This might be explained by the patients body mass index of 34 kg/m2, because morbid obesity has been described as the most common factor causing false-negative results (16).2 Lang et al. (16) explained this phenomenon to be a consequence of the decreased functional residual capacity in morbid obesity and the collapse of large airways due to invagination of the posterior wall when subatmospheric pressure is generated by the SIB. An improvement of their results could be achieved when the SIB was compressed after connection to the tube, thereby maintaining the patency of the tracheobronchial tree and allowing the SIB to refill. However, in one of our patients who was not obese and who showed no signs of tracheobronchial obstruction, bronchospasm, or pulmonary edema, a wrong result was obtained. We hypothesize that the tip of the tube was impinging on the carina. Possibly, reinflation would have occurred if more time had been allowed (i.e., delayed reinflation). However, allowing 30 s for reinflation, as recommended by Zaleski et al. (17), seems inappropriate, especially in emergency situations. Tracheal transillumination using a lighted stylet has been used for facilitation of endotracheal intubation, as well as for confirmation of tracheal tube placement (18). An intense, circumscribed midline glow in the region between the laryngeal prominence and the sternal notch is visible when the tube is correctly placed. In the case of esophageal intubation, the light is either absent or dull and diffuse. The potential advantage of
Czinn EA, Wafai Y, Salem MR, et al. Efficacy of the self-inflating bulb for confirmation of emergency tracheal intubation in critically ill patients [abstract]. Anesthesiology 1995;83:A266. 2 Wafai Y, Salem MR, Joseph NJ, et al. The self-inflating bulb for confirmation of tracheal intubation: incidence and demography of false negatives [abstract]. Anesthesiology 1994;81:A1304.
1

this method lies in its independence from hemodynamic and/or pulmonary impairment. Besides its dependence on ambient light conditions, it may fail in the presence of neck tumors or swellings and in obese patients with short and thick necks (19,20). In our study, rather high error rates (13% and 16%) were observed with transillumination, independent of the users experience. Patients in whom incorrect determinations of tube position were made had a higher mean body weight, but the differences were not statistically significant. Regarding our results, this technique seems unreliable for the assessment of tube position, even under favorable light conditions. The current study further shows that the assessment of tracheal tube position by auscultation was accurate when performed by experienced examiners. Falsepositive results by auscultation have been described by several authors (8,2125). In these cases, air movement through the esophagus was mistakenly identified as breath sounds. Andersen et al. (25) showed that auscultation of the epigastrium and the right and left axilla is more reliable for detecting tube position than chest auscultation. Our data suggest that capnography is the most reliable method for detecting tube position, independent of the examiners experience. Investigator experience, however, seems to be an important factor influencing the reliability of auscultation. Using the SIB led to correct results in 96% of cases, whereas transillumination was not trustworthy, even under optimal conditions.
The authors thank Ramez Salem, MD, Department of Anesthesiology, Illinois Masonic Medical Center, University of Illinois College of Medicine, Chicago, IL, for carefully reviewing our manuscript and for helpful comments.

References
1. Keenan RL, Boyan CP. Cardiac arrest due to anesthesia: a study of incidence and causes. JAMA 1985;253:23737. 2. Caplan RA, Posner KL, Ward RJ, Cheney FW. Adverse respiratory events in anesthesia: a closed claims analysis. Anesthesiology 1990;72:828 33. 3. Utting JE, Gray TC, Shelley FC. Human misadventure in anaesthesia. Can Anaesth Soc J 1979;26:472 8. 4. Holland R, Webb RK, Runciman WB. Oesophageal intubation: an analysis of 2000 incident reports. Anaesth Intensive Care 1993;21:608 10. 5. Schwartz DE, Matthay MA, Cohen NH. Death and other complications of emergency airway management in critically ill adults: a prospective investigation of 297 tracheal intubations. Anesthesiology 1995;82:36776. 6. Hartung J, Elpelt B, Klo sener K. Statistik: Lehr und Handbuch der angewandten Statistik. Mu nchen: Oldenbourg Verlag, 1993:9. 7. Shaffer JP. Modified sequentially rejective multiple test procedures. J Am Stat Assoc 1986;81:826 31. 8. Birmingham PK, Cheney FW, Ward RJ. Esophageal intubation: a review of detection techniques. Anesth Analg 1986;65:886 91.

770

CRITICAL CARE AND TRAUMA KNAPP ET AL. ASSESSMENT OF TRACHEAL TUBE POSITION

ANESTH ANALG 1999;88:766 70

9. Salem MR, Baraka A. Confirmation of tracheal intubation. In: Benumof JL, ed. Airway management: principles and practice. New York: Mosby, 1996:531 60. 10. Joshi GP, Smith I, White PF. Laryngeal mask airway. In: Benumof JL, ed. Airway management: principles and practice. New York: Mosby, 1996:35373. 11. Bozeman WP, Hexter D, Liang HK, Kelen GD. Esophageal detector device versus detection of end-tidal carbon dioxide level in emergency intubation. Ann Emerg Med 1996;27:5959. 12. White SJ, Slovis CM. Inadvertent esophageal intubation in the field: reliance on a fools gold standard. Acad Emerg Med 1997;4:89 91. 13. Jenkins WA, Verdile VP, Paris PM. The syringe aspiration technique to verify endotracheal tube misplacement. Am J Emerg Med 1994;12:413 6. 14. Petroianu GA, Maleck WH. Detection of an oesophageal intubation: state of the art. Anaesth Intensive Care 1994;22: 744 6. 15. Baraka A. The oesophageal detector device [letter]. Anaesthesia 1991;46:697. 16. Lang DJ, Wafai Y, Salem MR, et al. Efficacy of the self-inflating bulb in confirming tracheal intubation in the morbidly obese. Anesthesiology 1996;85:246 53.

17. Zaleski L, Abello D, Gold MI. The esophageal detector device: does it work? Anesthesiology 1993;79:244 7. 18. Hung OR, Pytka S, Morris I, et al. Clinical trial of a new lightwand device (Trachlight) to intubate the trachea. Anesthesiology 1995;83:509 14. 19. Stewart RD, LaRosee A, Stoy WA, Heller MB. Use of a lighted stylet to confirm correct endotracheal tube placement. Chest 1987;92:900 3. 20. Mehta S. Transtracheal illumination for optimal tracheal tube placement. Anaesthesia 1989;44:970 2. 21. Szekely SM, Webb RK, Williamson JA, Russell WJ. Problems related to the endotracheal tube: an analysis of 2000 incident reports. Anaesth Intensive Care 1993;21:611 6. 22. Pollard BJ, Junius F. Accidental intubation of the oesophagus. Anaesth Intensive Care. 1980;8:183 6. 23. Howells TH, Riethmuller RJ. Signs of endotracheal intubation. Anaesthesia 1980;35:984 6. 24. Stirt JA. Endotracheal tube misplacement. Anaesth Intensive Care 1982;10:274 6. 25. Andersen KH, Schultz-Lebahn T. Oesophageal intubation can be undetected by auscultation of the chest. Acta Anaesthesiol Scand 1994;38:580 2.