Vous êtes sur la page 1sur 3

SIDERMES Measuring Equipments

- SIDERTOP DATA TRANSMISSION WITH SIDERMES PROCEDURE (EPROM SD 148)

DATA TRANSMISSION CHARACTERISTICS: - ACTIVE CURRENT LOOP: 20mA (half duplex - 4 wires) active transmission, passive receipt. With proper selection on the related PC board, it is possible to obtain transmission and receipt both passive. - TRANSMISSION SPEED: 600 Baud - ASCII CHARACTERS: 7 bits + 2 stop bits + parity even - PROTOCOL: Polling type - HANDSHAKE obtained in the following way: The host computer sends the inquiry character ENQ. The equipment answers with NAK if no transmitting characters are present, or with the sequence of characters indicated below. The host computer, after receiving the Message and the validity check, sends back the character ACK. The equipment receives the character ACK and realizes the transmission was OK. If ACK is not received the Message will be re-transmitted after the following ENQ. CONTROL CHARACTERS USED: ASCII ENQ ACK NAK EOT MESSAGE SEQUENCE: ASCII (1) - Valid temperature measurement - Character before ID digits - 6 (six) ID digits - 4 (four) measure digits - Celsius degrees - End of transmission or: (2) - Not valid temperature measurement - Character before ID digits - 6 (six) ID digits - 4 (four) measure digits - Celsius degrees - End of transmission T Y C EOT t Y C EOT HEXADECIMAL CODE 54 59 43 04 74 59 43 04 HEXADECIMAL CODE 05 06 15 04

or: (3) - Valid temperature + Oxygen measurement - Character before ID digits - 6 (six) ID digits - 4 (four) measure digits for temperature - Celsius degrees - Sign + or Sign - 3 (three) measure digits for mV - 4 (four) measure digits for ppmO2 (see below**) - End of transmission or: (4) - Not valid temperature + Oxygen measurement - Character before ID digits - 6 (six) ID digits - 4 (four) measure digits for temperature - Celsius degrees - Sign + or Sign - 3 (three) measure digits for mV - 4 (four) measure digits for ppmO2 (see below**) - End of transmission or: (5) - Valid temperature + Oxygen measurement, with calculated Aluminium - Character before ID digits - 6 (six) ID digits - 4 (four) measure digits for temperature - Celsius degrees - Sign + or Sign - 3 (three) measure digits for mV - 4 (four) measure digits for ppmO2 (see below**) - 4 (four) digits for calculated Aluminium - End of transmission or: (6) - Not valid temperature + Oxygen measurement, with calculated Aluminium - Character before ID digits - 6 (six) ID digits - 4 (four) measure digits for temperature - Celsius degrees - Sign + or Sign - 3 (three) measure digits for mV - 4 (four) measure digits for ppmO2 (see below**) - 4 (four) digits for calculated Aluminium - End of transmission or: (7) - Valid temperature + Oxygen measurement, with calculated Carbon - Character before ID digits

X Y C SP EOT x Y C SP EOT

58 59 43 20 2d 04 78 59 43 20 2d 04

J Y C SP -

4a 59 43 20 2d

EOT

04

j Y C SP -

6a 59 43 20 2d

EOT

04

K Y

4b 59

- 6 (six) ID digits - 4 (four) measure digits for temperature - Celsius degrees - Sign + or Sign - 3 (three) measure digits for mV - 4 (four) measure digits for ppmO2 (see below**) - 4 (four) digits for calculated Carbon + the decimal point - End of transmission or: (8) - Not valid temperature + Oxygen measurement, with calculated Carbon - Character before ID digits - 6 (six) ID digits - 4 (four) measure digits for temperature - Celsius degrees - Sign + or Sign - 3 (three) measure digits for mV - 4 (four) measure digits for ppmO2 (see below**) - 4 (four) digits for calculated Carbon + the decimal point - End of transmission INTERNAL TESTS: TEST 1 - Valid temperature + Oxygen measurement - Character before ID digits - 6 (six) ID digits - 4 (four) TEST 1 digits for temperature - Celsius degrees - Sign + or Sign - 3 (three) TEST 1 digits for mV - 4 (four) TEST 1 digits for ppmO2 (see below**) - Identification of TEST - End of transmission TEST 2 - Valid temperature + Oxygen measurement - Character before ID digits - 6 (six) ID digits - 4 (four) TEST 2 digits for temperature - Celsius degrees - Sign + or Sign - 3 (three) TEST 2 digits for mV - 4 (four) TEST 2 digits for ppmO2 (see below**) - Identification of TEST - End of transmission

C SP -

43 20 2d

EOT

04

k Y C SP -

6b 59 43 20 2d

EOT

04

X Y C SP T EOT X Y C SP T EOT

58 59 43 20 2d 54 04 58 59 43 20 2d 54 04

** = In case Oxygen measurement is lower than 20 ppm, last transmitted digit is a decimal and so, just before it, will be transmitted the decimal point (HEX code 2e).