Vous êtes sur la page 1sur 24

CASE STUDY ON ANALYTICAL INSTRUMENTS (ANALYSERS)

BY VINAY KUHIKAR SR. MANAGER, TSX, PSWR

BHARAT HEAVY ELECTRICALS LIMITED POWER SECTOR WESTERN REGION

CONTENTS

SL. NO. 1.0 2.0 3.0 4.0 5.0 6.0 7.0 COVER SHEET CONTENTS INTRODUCTION

DESCRIPTION

NO. OF PAGES 01 01 01 05 02 02 11

PAGE NO. 01 02 03 04 - 08 08 - 09 10 - 11 12 - 22

FLUE GAS ANALYSERS OXYGEN ANALYSERS


DUST / OPACITY MONITOR STEAM AND WATER ANALYSIS SYSTEM (SWAS)

3. INTRODUCTION "Air pollution" is the presence in the atmosphere of one or more contaminants in sufficient quantities and of such characteristics and duration as to be injurious to human, plant, or animal life, to health, or to property, or to unreasonably interfere with the enjoyment of life or property. The pollutants monitored in Power plants Flue gas are Sulphur dioxide (SO2), Nitrogen dioxide (NO2), Carbon Monoxide (CO) and Suspended Particulate Matter (SPM). POLLUTION CONTROL BOARD NORMS National Ambient Air Quality Standards: 1. The level of air quality necessary with an adequate margin of safety, to protect the public health, vegetation and property. 2. Whenever and wherever two consecutives values exceeds the limit specified below for the respective category, it would be considered adequate reason to institute regular/continuous monitoring and further investigations.
POLLUTANT Time Weighted Average Industrial Area Residential Rural and other Area Sensitive Area SULPHUR DIOXIDE (SO2) Annual 24 hours Average * average ** 80 ug/m3 120 ug/m3 60 ug/m3 80 ug/m3 OXIDES OF NITROGEN (NO2) Annual 24 hours Average * Average ** 80 ug/m3 120 ug/m3 60 ug/m3 80 ug/m3 SUSPENDED PARTICULATE MATTER (SPM) Annual 24 hours Average * Average ** 360 ug/m3 500 ug/m3 140 ug/m3 200 ug/m3

15 ug/m3

30 ug/m3

15 ug/m3

30 ug/m3

70 ug/m3

100 ug/m3

Method of Measurement

Improved West and Gaeke methods

Ultraviolet fluorescence

Jacab and Hocheiser modified (Na-Arsenite methods)

Gas phase Chemilumio ne scence

High volume sampling (Average flow rate not less than 1.1 m3/minute)

* Annual Arithmetic Mean of minimum 104 measurements in a year taken twice a week 24-hourly at uniform interval.
** 24-hourly/8-hourly values should be met 98% of the time in a year. However 2% of the time, it may exceeded but not two consecutive days.

For coal based power plants, emission standard for particulate matter is 350 & 150 mg/Nm for the power plants having capacity less than 200/210 MW and more than 200/210 MW respectively. Gaseous emissions are controlled through dispersion at adequate heights. The stack height requirements for different capacities of power plants are as follows: Capacity Stack Height Less than 200/210 MW H=14(Q) Where Q is SO2 emission Kg/Hr. 200/210 MW to 500 MW: 220 Mtrs. More than 500 MW: 275 Mtrs. Concentration based standards for gaseous pollutants are not in existence for coal based power plants. However as and when it is required, the monitoring of those pollutants may be conducted following the methodology as suggested in CPCB (Central Pollution Control Board) publication namely Emission Regulation Part III CPCB does not have any guidelines or suggest methodology for verifying Flue gas analyser systems. It is up to buyer or user to procure these equipments as per their requirement. However, for valuation of performance of these equipments generally Calibration gas is used to verify the performance of the equipment as per suppliers guarantee. In case of particulate matter, the performance of Opacity meter / continuous monitoring equipment may be verified by manual isokinetic monitoring.

Flue Gas Analysers are required for the monitoring of Air Quality of Flue gases coming out of Power Plant Boiler Chimney. Flue gas analyser measures flue gases - Its probes are placed in chimney at required height or appliance exhaust and level of Oxygen (O2), Carbon Monoxide (CO), Carbon Dioxide (CO2), etc and flue gas temperatures are recorded. If the CO level in the flue of a gas appliance is above 100ppm (parts per million) then an investigation must take place. For oil or coal fired appliances the CO should not be above 200ppm. Equally, Oxygen (O2) levels should be in the region of 3-5% for gas appliances and 5 - 8% for oil & solid fuel appliances - Any other levels suggests an inefficient use of energy, wasting fuel and money, or a potentially unsafe situation where too much CO is created through oxygen deficiency

4. Flue Gas Analyser


Main components of this system are: 1. Flue Gas Analyser for measuring the quantity of SO2, NO & CO. 2. Flue Gas Sample Probe for collecting Flue gas sample from Chimney inlet. 3. Heated Flue Gas Sample line for transporting Flue gas sample from probe to Analyser. 4. Sample Gas Cooler for cooling of Flue gas sample to remove moisture. 5. NO2/NO Converter converts NO2 in Flue gas sample to NO for Analyser analysis.
4

6. Diaphragm Pump for sucking the Flue gas sample from probe to Analyser. 7. Peristaltic Pump for removing condensed water form the cooler drain. 8. Sample handling system consisting of Solenoids, valves, filters, etc. 9. Programmable Logic Controller (PLC) for logic programming of Purge cycle. 10. Power supply source 24 VDC Power supply for PLC 11. Oxygen Analyser for measuring the quantity of O2 in Flue gas. 12. O2 Probe (Zirconium) for collecting & conditioning of Flue gas sample from Chimney inlet. 13. Power supply MCB and alarm JBs 14. Air Conditioner for keeping the ambient temperature inside the panel around 26C so that electronic equipments like PLC and Analysers can funtion properly. 15. Certified Calibration Gas cylinders of SO2, NO, CO, N2 & O2 (Span & Zero). OPERATING PRINCIPLES FLUE GAS ANALYSER Flue gas analyser is for continuous measurement of IR components, including H2O for cross sensitivity correction. It measures gas components with NDIR measuring system. It measures all components simultaneously. The time that analyser needs to determine all measuring values once depends on the number of measuring components and the physical measuring range. Typical time: 5 20 Sec. A wide band infrared radiation is emitted by a non-metallic, high stability source. For each measured gas this radiation is routed alternatively through an interferential filter, a cuvette filled with nitrogen and a filter, and a cuvette filled with high partial pressure of the gas to be analyzed. A suitable optical system drives the IR radiation inside the analysis chamber and then to the detector, which receives and amplifies alternatively the two signals: one is the measure and the other is the reference. The gas concentration is proportional to the difference of the two signals. Gases which may have a cross sensitivity with the measured gas, generate the same variation of the measure and reference signals. Therefore the measure is not affected. The wheel that holds the cuvettes and filters rotates at 3000 r.p.m. The sensor can acquire 50 measures each time one of the cuvettes is in front of the I.R. source. A high quantity of measures is available. They must be amplified and computed, having a very high precision and stability. Two microprocessors connected via a dual ram are used. With this approach no measures are lost and the instrument can manage all functions. Due to physical reasons, the NDIR measuring principle has a certain cross-sensitivity against H2O when measuring SO2 or NO. The measuring system cannot be optimized for both measuring components, as it is when only one component is measured. As a result, the H2O cross-sensitivity is higher when simultaneously measuring SO2 and NO. Therefore this analyser additionally determines the actual H2O concentration in the sample gas and compensates the cross-sensitivities arithmetically with this measured H2O value. Three different test gases can be used to calibrate the sensitivities (Spans). INSTALLATION DETAILS The Flue gas analyser panel is generally installed inside or nearby Chimney. Panel have 1.0 Ton AC installed on it for maintaining the ambient temperature inside the panel.
5

During Operation of Flue Gas Analyser, the allowable ambient temperature range must be kept (545C), otherwise the measuring accuracy might not be as high as specified. LOCATION DETAILS The installation location should be free of mechanical oscillation and vibration. Especially low frequency vibration for example, caused by road traffic or heavy machinery can disturb the measuring operation. Please avoid exposing the analyser to direct sunlight. It must be installed / mounted horizontally. An extreme case inclination would interfere with the accuracy of the flowmeter. RECOMMENDATION The gas analysis instrument is part of a measuring system. In order to ensure that the analysis is trouble-free, that it requires a minimum of maintenance and produces good analysis data, it is necessary to set up the whole measuring system logically. The correct choice of the point of gas removal, the devices for preparing the sample gas and careful installation are as crucial to the success of measurement as the analyser itself. The Sample gas probe is installed on the gas duct before Chimney CALIBRATION Why is calibration necessary for Gas Analysers? The characteristics of optical and electronic components slightly change in the course of operation. These changes affect a high-precision measuring system and results in small changes of the measuring results. This effect is known as Drift. To compensate drift, a gas analyser must be regularly calibrated. In a calibration, the measurement result of the analyser is checked with the Certified Calibration Test Gases, and then the offset from the correct NOMINAL value is adjusted to bring the analyser back to the true reading. The two important parameters in the measuring system are: 1. The Zero point: defined as the reaction of the instrument when the component of interest (measuring component) is not present. 2. The sensitivity or span: defined as the reaction of the instrument when the component of interest (measuring component) is present. There is a Zero point drift and sensitivity drift for each measuring component. Each must be determined and corrected independently. Calibration gas standards are invaluable in ensuring exact results. Zero gas: The Zero Gas is used to calibrate the zero points of all measuring components. The Zero Gas must NOT contain any of the measuring components. As a rule Nitrogen can be used for the Zero Gas, either Technical of Top grade depending on the application. Test Gases: The Test Gases are used to calibrate sensitivities. A test gas is a mixture of Zero gas and one or more measuring components. Nominal values of the test gases The nominal values are the actual concentration of the measuring components in the test gas. The nominal values should be within 60 to 100 % of the physical measuring range end value of the corresponding measuring component.
6

STANDARD LIST OF CALIBRATION GAS CYLINDERS


1. 2. 3. 4. 5. 6.

Zero Gas: 100% N2 in a 10 liters C.S. Cylinder 1600-1800 ppm of SO2 balance N2 in a 10 liters Al. Cylinder 200-225 ppm of CO balance N2 in a 10 liters Al. Cylinder 800-900 ppm of NO balance N2 in a 10 liters Al. Cylinder O2 Zero Gas: 2 % O2 balance N2 in a 10 liters C.S. Cylinder O2 Span Gas: 18 % O2 balance N2 in a 10 liters C.S. Cylinder

CERTIFICATES 1. Concentration and Stability Certificates for Calibration Gas Cylinders of SO2, NO, CO, N2 & O2 (Span & Zero) gas is required. The validity of these certificates is specified by Calibration gas filling agency. Gas Validity may be for 6 or 12 months depending upon the type of gas and its concentration. 2. Vendor provides Chief Controller of Explosives Certificate for each Calibration gas cylinder. These certificates are useful document for refilling of calibration gas cylinders.

PROBLEMS ENCOUNTERED AND RESOLVED Flue gas sample line choking is the biggest problem in Thermal Power Plant Flue Gas analyser system. Ceramic Flue gas sample probe filter also gets choked frequently due to accumulation of fly ash and gas condensation. Most of the time the Instrument air purging cycle is ineffective in clearing the sample line choking. Another reason for improper air purging is failure of PLC controls. 24 VDC Power supply is required for PLC operation. This is provided by 24 VDC Power supply module having input 230 VAC. Improper slope of Flue gas sample line also result in gas sample condensation in line. Some time Flue Gas Sample Probe internals (filter & holding spring) was also found eroded and broken. The insufficient heating capacity of Heat Tracing line is also responsible for moisture condensation in sample line. At temperature less SO2 dew point moisture condenses to form Aserol (sticky powder). Over the period of few weeks this substance blocks the sample line. We have observed the choking of Flue gas sample cooler due to Ice formation in sample line if setting and wiring of potentiometer wire of temperature control circuit is not done properly. SO2 / NOx / CO readings become unstable due to insufficient flue gas sample flow. The condensed water from cooler drain needs to be drained without allowing air ingress into the system. This is made possible by using Peristaltic pump. The low rating pump initially provided was not able to drain all the water at faster rate. This resulted in accumulation of water in cooler and sample line inside the Analyser cabinet. The outer covering of the Electrically heated hose, which was kept in rolled form, got melted at overlap point within one week of its operation. Poor response of Vendors for attending to system problems and insufficient supply of commissioning spares may leads to the total system failure. COMMISSIONING FEEDBACKS The Ceramic filter in the Flue gas sample probe is for fly ash removal. It should be checked for its specification and its Test Certificate must be requested from the
7

Vendor. Extra Filter needed to be installed in sample line for the removal of very fine fly ash. Some times Flue gas sample line needed to be cleaned to remove choking, it can be done by using high-pressure hand operated pump. The Flue Gas Sample Probe casing must have heat-tracing wound on it. This is to avoid the flue gas sample condensation in probe. Peristaltic Pump operation should be checked as per the actual requirement. This pump must remove all the condensed water in cooler. If required higher capacity pump can be installed. It is necessary to check the setting of Heat tracing Thermostat. By installing RTD in the sample line we can monitor the temperature of Flue gas sample. Flue gas sample line (with heat tracing) must be installed with proper slope to avoiding moisture condensation in sample line. Timing of Air purging cycle must be adjusted so as to avoid frequent choking of Flue Gas Probe filter. The installation of electrically heated Teflon hose (if provided) must be done as per vendor instruction. It should not be kept in coil form otherwise its outer sheath will melt at contact points. Reading of O2, SO2, NOx, CO and load should be configured in DCS Trends for continuous monitoring of Flue gas analyser system. Certified Calibration gas cylinders validity must be checked before using it for Calibration of Analysers.

SUGGESTIONS 1. Flue gas sample line between sample Probe and analyzer panel must be purged at regular interval with Instrument air along with sample probe purging. For doing this, one solenoid valve with instrument air supply to be installed in analyser panel. Sample probe purging timer can be used for the operation of this solenoid. 2. A catch pot should be placed between sample probe and flue gas sample line, just after sample probe (near duct) with air operated solenoid drain valve. This drain valve should open during gas sample line air purging cycle. 3. Diaphragm pump should be of high capacity for the suction of Flue gas sample. 4. The AMC offer shall separately indicate one year annual maintenance contract including all the spares, consumables required. 5. Scope Of Analyser Vendor Should Include a. Necessary rectification / modification in the Analyser system as per site conditions. b. Total Commissioning of the system. c. Field acceptance of the system by BHEL and Customer. d. Handing over the system to Customer O&M with necessary guidelines and spares for further Operation and Maintenance. e. Annual Maintenance Contract (AMC) for Analysers. (During BHEL one year Warranty period) f. Refilling of Certified calibration gas cylinders. DOCUMENTS REQUIRED FOR COMMISSIONING OF FLUE GAS ANALYSERS 1. Installation and commissioning procedure of Flue Gas Analyser.
8

2. 3. 4. 5. 6. 7. 8.

OGA drawing (Location Drawing) of Flue gas analyser panel in Power plant. Heat tracing route drawing. Heat tracing Instruction Manual. PLC O&M and Instruction Manual. Write-up for the Routine Maintenance List of Mandatory spares for Flue Gas Analyser system. Concentration and Stability Certificates Certificate and Chief Controller of Explosives Certificate of calibration gas cylinder of N2, SO2, NOX, O2 (Span gas), O2 (Zero gas) and CO.

MODIFICATION DONE FOR BETTER OPERATION OF SYSTEM It was found that the moisture content in the flue gas is high because of Soot blowing in boiler at the rate of 2 hours per shift. The slope of sample line was not sufficient to drain the excess moisture from the sample line. It was jointly decided by vendor, Customer and BHEL Site Engineer in consultation with EDn Bangalore Design Engineer to modify / re-route the Flue gas sample lines to avoid moisture condensation. After above modification it was observed that the moisture condensation is not taking place in sample line.

5. OXYGEN ANALYSER INTRODUCTION Oxygen Analyser measure oxygen concentrations in boiler with the help of probes located at various locations in boiler. The zirconia method eliminates the need for sampling equipment, is easy to install, provides quick response and stable display, and is easy to maintain. The probe is attached to a terminal block. The lead wire, which is directly connected to the internal electric pole, and cell heater are contained within the probe. Therefore there is no contact with the direct exhaust gases. The measurement range, output signal, and calibration gas concentration can easily be changed.

OPERATING PRINCIPLES OXYGEN ANALYSER The working element of the gas sensor is a closed-end tube made of a ceramic oxide, Zirconium Oxide (Zirconia). When it is hot (temperatures that exceed 600 C) it becomes a conductor of electricity because of the mobility of the Oxygen ions in its crystal structure. Electrodes of porous platinum are coated onto the inside and outside of the cell and connected to the measurement circuitry. When the sensing cell is hot, a voltage is produced that is logarithmically proportional to the ratio of the oxygen concentration of the gas on the reference side of the cell (usually ambient air) and the oxygen concentration of the sample. If the oxygen in one gas is known (the reference gas is normally air 20.9 % O2), the oxygen of the other gas is indicated directly by the voltage from the cell. The sensing cell does not produce a signal when air is on both sides, and the voltage increases as the oxygen concentration in the sample diminishes, relative to air. Because of the high operating temperature of the cell, combustible gases that are
9

present may burn, when this occurs, the cell will generate a higher than expected millivolts and cause the display to indicate less oxygen than is actually in the gas.
INSTALLATION LOCATION Probe locations of Oxygen Analysers in boiler are generally as follows: 1. Economiser Outlet- (Left): 01 No. 2. Economiser Outlet - (Middle): 01 No. 3. Economiser Outlet - (Right): 01 No. 4. Air PreHeater-A Outlet: 01 No. 5. Air PreHeater-B Outlet: 01 No. COMMISSIONING PROCEDURE, CALIBRATION Commissioning procedure will be as given in Instruction Manual of Oxygen Analyser Calibration can be carried out with O2 Span (8.0 %) and Zero (0.4 %) gas for all Oxygen Analysers. CERTIFICATE & CALIBRATION ACCESSORIES 1. Oxygen Gas concentration and stability Certificates 2. Chief Controller of Explosives Certificate 3. Certified calibration Oxygen gas cylinder for Unit 3 & 4.The details of Calibration Oxygen cylinders are as follows: Zero Gas: 0.4 % O2 balance N2 in a 10 ltrs C.S. Cylinder (SGC18375, SGC18397): 02 Nos. Span Gas: 8.0 % O2 balance N2 in a 10 ltrs C.S. Cylinder (SGC18387, SGC18371): 02 Nos. PROBLEMS ENCOUNTERED AND RESOLVED Electronics module of Oxygen analyser got defective. Vendor representative repaired it at site.

6. DUST / OPACITY MONITOR INTRODUCTION


10

The Opacity Monitor is used for continuous opacity or dust monitoring of flue gas in stacks and ductwork. A opacity monitors give vital analysis of pollutant particulate emission levels and provide a primary indicator of overall boiler efficiency. Opacity measures the percentage of absorbed light. An advantage of opacity is that the measuring devices need not be calibrated. A disadvantage is that the results vary with the location of the monitor itself. However, the results can be recalculated for the distance from the monitor to the stack emission outlet. The influences of the color, shape, and surface structure of the dust particles are not taken into account, shown in the non-linear relationship between dust load and opacity. For example, opacity increases only about 15% when the dust load is doubled in the upper measurement ranges. Measurements can be displayed as opacity or optical density. The extinction (optical density) has a linear relationship to the dust load because it is calculated logarithmically: doubling the dust load results in a doubled measurement. A calibration curve is calculated for the measuring location according to VDI 2066 or EPA CFR 40 Part 60 #5 with the extinction coefficient k. Under normal operating conditions, however, the same dust load can lead to different readings depending on grain size and flow distribution, so the uncertainty ranges are also calculated. OPERATING PRINCIPLE Operating Principle: Cross-stack double pass transmissometer The beam transmissometer measures the fraction of light, from a collimated light source, reaching a light detector a set distance away. Light, which is absorbed or scattered by the media, between the source and the detector does not reach the detector. The fraction of light received is converted to the beam attenuation coefficient (usually called "c") by the formula c = ln(T)/z where T is the fraction of light transmitted and z is the path length of the instrument. The exact relationship varies with the type of particles present. The measuring system is based on the single pass/dual patharchitecture, illustrated by the enclosed picture. Method - I The laser beam sent by the laser unit crosses the measuring section only once and the receiver unit measures and evaluates the laser beams weakening caused by dust content. Additionally the laser unit sends the reference signal around the stack via optical fibre to the receiver unit, 40 times per second. This system improves the longterm accuracy of the analyser, because in traditional analysers with a dual-pass design, the retro-reflector is excluded from the zero/span calibration. The optical reference path around the stack ensures that all affected surfaces are included in the calibration. The monitor automatically corrects for dirt that can accumulate on these surfaces. Method - II Opacity monitor operates on the auto-collimation (double-pass) principle. The light beam crosses the measurement path twice and the system measures the light attenuation from dust in the stack. A photoelement alternately reads the measurement beam and a comparison light beam 2 minutes to insure accuracy. Since one joint amplifier is used for both the measuring and comparison light beams, temperature fluctuations and long-term amplifier drift are automatically compensated. The system generates light internally with the Super wideband Light Source to prevent distortions from sunlight or other sources. The broad spectrum of emitted light means results are not distorted by temperature. This ensures a more stable measurement than a conventional narrow band LED.

11

INSTALLATION DETAILS, LOCATION, RECOMMENDATION System alignment is performed using a direct view alignment port built into the light source optics. The retro reflector also contains a built in alignment port, to align the retro reflector. Alignment of both opacity monitor light source and reflector is critical for accurate off stack zeroing of opacity monitors. COMMISSIONING PROCEDURE & CALIBRATION Zero and span calibrations are performed hourly and timed for continuous stack sampling with no monitor down time. Calibration drift is automatically compensated for and we report the information daily on user-selected outputs. The hourly system calibrations assure minimal erroneous data collection. Dirty window compensation is corrected hourly and we display a visual alarm on the operator panel if the correction limit is exceeded. TYIPICAL CONFIGURATION DATA OF DUST / OPACITY MONITOR (LAND Model 4500) UNIT TYPE = MKII AVG TIME = 01 MIN MODE = DUST DUST_CAL = BLK AVG DUST BKGD = 0.000 ALERT = 0.00 mg/m3 ALRT DELAY = 10 SEC ALRM DELAY = 15 SEC CAL. EVERY = 24 HRS OUTPUT 1 = I DUST RANGE 1 = 1000 mgm RANGE 2 = 1000 mgm OUT 2 = 4 - 40 mA LINE FREQ = 50 HZ CAL IN = 6:00 BAUD RATE = 9600 Passwords: 1. For changing TYPE and O.D: 417 2. For normal calibration check: 10 3. For Input Current calibration: 27 PROBLEMS ENCOUNTERED AND RESOLVED 1. Display on LCU of Opacity/Dust Monitor not visible on both the units. 2. Opacity/Dust Monitor Blower found burnt and needed rewinding. It was repaired twice. 3. Transceiver units found faulty and cannot be repaired at site and need to be taken at vendor Service center. Following items serviced by vendor at their works and reinstalled at site: Transceiver unit of Unit #3 (Location: Flue gas duct): 01 No LCU of Unit # 3 (Location: LCU JB): 01 No Micro P.C.B of Unit # 4 (Location: inside transceiver): 01 No Control P.C.B of Unit # 4 (Location: inside transceiver): 01 No 4. Constant Over range reading in Dust Monitor was due to loose current loop connector. 5. CRU Units Range and Dust gain could not be set due to program error. Hence, the CRU unit has been bypassed and mA output has directly been connected to Control Room Panel. 6. BHEL Ranipet did not give opacity versus Dust level curve. 7. Calibration Check was not operational because of faulty calibration motor. PLCP = 0.500 DUST GAIN = 110 ALARM = 800 mg/m3 O.D. = 3.00 OUTPUT 2 = I DUST OUT 1 = 4 - 20 mA S/W VER = 3.12 INPUT RANGE = 4 - 20 Ma

12

7. STEAM & WATER ANALYSIS SYSTEM (SWAS) INTRODUCTION We understand that within power stations the aim of water and steam analysis is to minimize contamination of the circuit, thereby reducing corrosion as well as the risk of the formation of harmful impurities and operation costs. Whether it's water or steam, we need accurate control of pH & conductivity in our process. Errors can lead to costly equipment damage or wasted time & materials. The purity of Boiler Feed Water and Steam is absolutely crucial in turbine applications. To prevent costly plant shutdowns due to scaling and corrosion, over two dozen steam and water samples have been taken to measure pH, Conductivity, Dissolved Oxygen, Silica, Sodium and Hydrazine. These samples can be as hot as 560 C, with pressures up to 250 bars. SWAS system is for Conditioning and Analysing steam / water sample at elevated temperatures and pressures. For each Analysers there are single line units, pre-piped to include all necessary elements required to reduce temperature and pressure of hot samples, so that they can be safely handled and analyzed. Each unit consist of a Sample cooler, Pressure regulator, Isolating and regulating valves for sample and coolant, indicators for temperature, pressure and flow, as well as necessary tubing and fittings. SWAS generally consists of following Analysers: 1. PH Analyser (example 9135 - ZELLWEGER ANALYTICS, Polymetron make ) 2. Conductivity Analyser (example 9125 - ZELLWEGER ANALYTICS, Polymetron make) 3. Silica Analyser (example 9097 - ZELLWEGER ANALYTICS, Polymetron make) 4. Dissolved Oxygen Analyser (example 9182 - ZELLWEGER ANALYTICS, Polymetron make) 5. Sodium Analyser (example SODIMAT 9073 - ZELLWEGER ANALYTICS, Polymetron make) 6. Phosphate Analyser (example PHOSPHAMAT 8892 - ZELLWEGER ANALYTICS, Polymetron make) 7. Chlorine Analyser (example CHLOROMAT 9184 - ZELLWEGER ANALYTICS, Polymetron make) 8. Hydrazine Analyser (example HYDRASTAT 9186 ZELLWEGER ANALYTICS, Polymetron make) 9. Chloride Analyser (example MODEL 1517 ORION make) OPERATING PRINCIPLE SWAS ANALYSER PANEL The sample is tapped from the process stream and fed to the system through a 1 OD line. A bulkhead connector is provided at the sample inlet point. The sample reaches the sample cooler after passing through an isolating valve, which may be high pressure or low-pressure quality depending upon the service. The Sample cooler is coil in shell type construction and can operate under severe operating conditions. There could be one or two sample coolers in series depending upon the total heat load in the sample stream. The coolant enters the sample cooler from the coolant header, fills the shell and goes back to the return header. The shell is protected from excessive pressure, which may arise due to accidental rupture of sample coil by means of a relief valve. The flow indicator is provided to check the coolant flow in the coolant lines. The sample than passes through a sample filter, which removes any suspended matter from the sample to the pressure regulator, which brings the outlet pressure down to a set level. A relief valve is also built-in with the pressure regulator, which may be set to blow off at preset pressure, thus protecting the down stream apparatus.
13

After pressure reduction, the conditioned sample is taken in a header where its temperature and pressure is monitored. Temperature and Pressure gauges are provided for this purpose. The Analysers that are fed by this system operate under temperature and pressure that are very low (Typically 2 to 3 Kg/sqcm & 30 to 40 C). Thus it is essential to protect them from over temperature or over pressure conditions, which would damage the instruments. The temperature and pressure switches are fitted into SWAS as safety precautions to prevent such system failure. The switches receive a signal from temperature and pressure sensor mounted on the header and operate solenoid valve, which blows the sample down in case of over temperature or over pressure condition. The sample than goes to respective Analysers through Rotameters. Adjusting the regulating valve provided with the Rotameter regulates the flow rate. PH ANALYSER pH is the negative logarithm of the hydrogen ion activity and measure of the acidity, or alkalinity of a solution. pH = - log10 [aH+] pH is normally measured using a glass electrode and a reference electrode. the glass electrode acts as a transducer, converting chemical energy (the hydrogen activity) into an electrical energy (measured in millivolts). The reaction is balanced and the electrical circuit is completed by the flow of ions from the reference solution to the solution under test. The electrode and reference solution together develop a voltage (emf) whose magnitude depends on the type of reference electrode, the internal construction of the glass electrode, the pH of the solution and the temperature of the solution. For every unit change in pH (or decade change in ion concentration), the emf of the electrode pair changes by 59.16 mV at 25 Deg C. This value is known as the Nernstian slope of the electrode. The pH electrode pair is calibrated using solutions of known and constant hydrogen ion concentration, called buffer solutions. The buffer solutions are used to calibrate both the electrode isopotential and slope. CONDUCTIVITY ANALYSER Measuring the conductivity of the liquids has become an increasingly important quality control tool for the process industries where it is often used to indicate a change in the process rather than an absolute analysis. In other words, unlike ion selective measurements, the total quantity of ions present in the solution is measured, rather than the type of ions itself. The conductivity of a liquid is somewhat similar to that of an electric current flowing though a metal when a voltage is applied. The main difference, however, is that in a metal conductor the current flows by transport of electrons, whereas in a liquid, it flows by transport of ions. It follows that the greater the numbers of ions present in the liquid, the higher is the transport of charges per unit time and hence the conductance of the liquid. In effect it is the inverse of electrical resistance and the unit of measurement is Mho. The electrolytic conductivity of a liquid (sometimes called its conductance per cm or m) is dependent on the concentration of the liquid, (i.e the number of available ions), as well as the valency and mobility of the ions and the degree of dissociation of the electrolyte. The mobility of the ions is itself dependent on the temperature of the solution, so that at constant temperature, the conductivity is directly related to concentration. Typical industrial applications for conductivity measurement include water purity, Boiler Feed water protection, check on water hardness and boiler blow down control. The conductivity measurement of pure water is a special case, since the specific conductivity of pure water relates to the sum of the specific conductivity values of all the salts present and the dissociated H+ and OH- ions. SILICA ANALYSER
14

Polymetron Silkostats Silica Analysers offer online monitoring of low-level silica in pure sample water and steam. Soluble silica reacts with ammonium molybdate to form a yellow-coloured silicomolybdate complex. The latter may be reduced, by means of a suitable reducing agent, to the strongly blue-coloured molybdenumblue complex: Photometric analysis of silica by measurement of molybdenumblue is substantially more sensitive than the assay via the yellow silicomolybdate complex. The formation of molybdenum blue is strongly dependent upon the solution pH, the type and amount of reducing agent, as well as various other factors. It is important, therefore, that the instructions for the preparation of the reagents and the method are strictly adhered to. Phosphates, analogous to silicates, are also capable of producing yellow phosphomolybdate complexes and, in the presence of reducing agents, molybdenumblue. It is possible, within limits, to suppress this interference by judiciously adding oxalic acid. In order to assure optimum turbine performance, continuous monitoring of silica in superheated steam, boiler water and feed water is of utmost importance. DISSOLVED OXYGEN ANALYSER The measurement of dissolved oxygen is based on the Clark cell principle. An oxygen-permeable membrane isolates the electrodes from the sample water, thus obviating the need for sample conditioning. Other reducible or oxidisable ions do not interfere, because they cannot pass through the gas-permeable membrane. A constant voltage supply powers two electrodes, maintaining each at a constant potential. A gold working electrode (cathode) reduces the dissolved oxygen to hydroxyl ions: O2 + 2H2O + 4e- => 4OHA large silver counter electrode (anode) provides the oxidation reaction, which occurs on its surface: 4Ag+ + 4Br- => 4AgBr + 4eThe reduction of oxygen is the current limiting reaction, thus making the cell current linearity proportional to the dissolved oxygen concentration. Electrochemical reactions and diffusion rates are temperature sensitive. The measuring cell, therefore, is equipped with a temperature sensor, which allows automatic temperature compensation. Technical characteristic Sample: Temperature: 0 to 45 Deg C. Working pressure: Atmospheric pressure Flow rate: 4 to 10 liter/hr Material: Working electrode: Cathode: gold Counter electrode: Anode: silver Membrane holder: Noryl Membrane: PFA Transmitter: Aluminium + polyester Probe body: Noryl SODIUM ANALYSER Modern high-pressure power plants require highly pure Feedwater. Polymetron Sodimats offer online monitoring of low-level sodium in ultra pure water and steam. The measurement is based on a direct potentiometric technique using a highly sensitive sodium glass electrode. The sodium glass electrode is placed in a sample which has been previously conditioned to a pH > 10. Sample conditioning is necessary because sodium glass electrodes are not perfectly specific sensors, but are subject to interference by ions, especially ions H+. Low level sodium measurement, therefore, require that the H+ of the sample water be adjusted to a level several orders of magnitude below the level of sodium. The difference of potential between the glass electrode and the reference electrode is directly proportional to the sodium concentration. Sodium cations and anions are always linked. Most cations have a corrosive influence in water and vapor cooling circuits. Because of this chemical link between sodium ions and anions, sodium measurement presents particularly important risks of corrosion and other effects.
15

PHOSPHATE ANALYSER It uses photometric analysis to monitor the levels of phosphates in water. Orthophosphate reacts with ammonium molybdate to form a yellow-coloured phosphomolybdate complex. The latter may be reduced, be means of a suitable reducing agent, to the strongly blue coloured molybdenum blue complex. Photometric analysis of phosphates by measurement of molybdenumblue is substantially more sensitive than the assay via. the yellow phosphamolybdate complex. The formation of molybdenumblue is strongly dependent upon the solution pH, the type and the amount of reducing agent, as well as various other factors. It is important, therefore the instructions for the preparation of the reagents and the method are strictly adhered Ammoniumolybdate reacts with orthophosphate. The rate of reaction of this complex formation is practically instantaneous. Direct photometric assay of the yellow complex is possible, in principle, but the sensitivity of the measurement is insufficient for the analysis of phosphate concentration in the low ppm range. for this reason, the phosphamolybdate is reduced with ferrous ions to the much more sensitive molybdenumblue. Note: 1. Ferrous iron solution is particularly suitable as a reducing agent because of its stability against oxidation by air, as well as its fast rate of reaction with phosphomolybdate. 2. A high sample-to-reagent ration minimizes errors arising as a result of inaccuracies of the reagent delivery rate. Its principal application is phosphate analysis in Boiler Feed water, Steam generation, etc. Boilers operating at medium to low pressures frequently utilize sodium phosphates for alkalization (pH 9 10), and prevent the formation of residual water hardness scale in the steam generating components. CHLORINE ANALYSER Polymetron Chloromat Analysers were designed for the on line monitoring of Chlorine levels in water circuits. The Chloromats feature controlled potential amperometry and a self-cleaning working electrode system. Active chlorine measure lies on the Clark cell principle. The Sensor amperometric type is composed of: one gold working electrode, where the main reaction occurs. one silver counter electrode, which is also used as reference electrode. a potassium chloride based electrode. one microporous membrane selective to HOCl The HOCl molecules contained in the sample diffuse through the membrane. HOCl is then located in a thin electrolyte layer, between the membrane and the cathode. A constant working potential is applied to the working electrode (cathode) where HOCl is reduced: HOCl + H + 2e- => Cl + H2O At silver electrode (anode), silver is oxidised into Ag+ ions which then percipitate with the chloride ions 2Cl + 2Ag => 2AgCl + 2eHOCl reduction at the cathode generates a current directly proportional to its concentration. The electrochemical reaction and the diffusion through the membrane depend on temperature. The measuring cell is therefore equipped with a temperature sensor which enables to operate an automatic compensation of the measurement. HYDRAZINE ANALYSER
16

Hydrazine is one of the most common water treatment chemicals used on modern boiler houses. But dosing too mush is an expensive waste, while too little leads to equipment damage. Polymetron Hydrastat analysers are designed for accurate on-line monitoring of hydrazine levels in water samples.

This analyser continuously measures the amount of dissolved hydrazine and onter oxygen reducers in water. The measuring principle is based on the electrochemical method of 3-electrode amperometry. A polarization voltage (+ 480 mV) is applied between a platinum anode (working electrode) and a stainless steel cathode (counter-electrode). Hydrazine is oxidized at the surface of the platinum electrode working electrode and the resulting current is directly proportional to the hydrazine concentration in the range of 0 to 500 ppb N2H4. The reaction is enhanced in the alkaline environment, sample is conditioned at pH = 10.2 adding diethylamine or disopropylamine through a Venturi tube, before the sample enters the measuring cell. Compensation of the temperature effect is achived through a semiconductor sensor integrated to the measuring cell. The chemical reaction is (1) N2H4 + 4OH => N2+4H2O + 4eThe anode-cathode torque potential is kept constant par rapport with respect to a third electrode (reference electrode, Ag/AgCl). The system avoids interference effects interference effects resulting from the variations of water composition that appear when using the 2-electrode system. At + 480 mV, the cell current is linearly proportional to the Hydrazine Concentration. CHLORIDE ANALYSER The sample stream enters the Low Chloride Monitor and passes through a sample bypass valve, which allows excess sample to be diverted to a waste drain. Bypassing excess sample to drain allows the analyser to respond more quickly the changes in sample chloride concentration, especially where there is a long sample line prior to the analyser. Sample is filtered to remove particulate matter to prevent the clogging of analyser. Enters a pressure regulator and a flow meter / needle valve assembly to ensure a constant pressure and flow rate in the Analyser. Passes through the three way solenoid valve and then either flows through or bypasses a deionization cartridge before being cooled down to 5 C. Passes through the reagent diffusion bottle where pH adjustment takes place. Reagent consumption is held at a constant level because sample temperature is maintained at 5 C. Re-cooled to 5 C 0.1 C, eliminating the need for temperature compensation due to varying sample temperatures. Sample is than passed from the chloride or measuring electrode, whose voltage output varies with changes in chloride concentration. It passes by the reference electrode. The reference electrode maintains the fixed voltage output when in contact with the sample. the combined chloride and reference electrode voltage outputs are amplified and displayed. Sample is than drained into atmosphere. Note: The Reference electrodes position is downstream of the chloride electrode to avoid chloride contamination of the sample from the reference electrode filling solution. INSTALLATION DETAILS, LOCATION, RECOMMENDATION ANALYSERS INSTALLED FOR: Boiler Drum Saturated steam Superheated Steam Feed water Condensate Water Deaerator Makeup DM Water Conductivity (Mho), pH, SiO2 (ppb), PO4 (ppm) Conductivity (Mho), pH Cation Conductivity (Mho), pH, Conductivity (Mho), SiO2 (ppb) DO2 (ppb), pH, Conductivity (Mho), N2H4(ppb) Sodium (ppb), pH, Conductivity (Mho), SiO2 (ppb), Cation Conductivity (Mho), Chloride (ppb) DO2, pH Conductivity (Mho), pH
17

Condensate Cooling Water Hotwell LOCATION

Chlorine (ppb) Conductivity (Mho) - Right side, Conductivity (Mho) - Left side

DRY PANEL: All SWAS Analysers are mounted in this panel. During Operation of SWAS Analysers in Dry panel, the allowable ambient temperature range must be kept (545 C), otherwise the measuring accuracy might not be as high as specified. In other words Air-conditioning required. The installation location should be free of mechanical oscillation and vibration. Please avoid exposing the analyser to direct sunlight. It must be installed /mounted horizontally. WET PANEL: This entire panel must be located near to Dry panel. All Sample coolers, Pressure regulator, Isolating and regulating valves for sample and coolant, indicators for temperature, pressure and flow, temperature switches, as well as necessary tubing and fittings are mounted in this panel. It does not require Air-conditioning but installation location should be free of mechanical oscillation and vibration. Typically following nos. of each type of Analysers may be required in each Unit PH Analysers: 07 Nos. Conductivity Analysers: 08 Nos. Silica Analysers: 03 Nos. Dissolved Oxygen Analysers: 03 Nos. Sodium Analyser: 01 No. Phosphate Analyser: 01 No. Chlorine Analyser: 01 No. Hydrazine Analyser: 01 No.

General Specification of Analyser SN Descriptions O. 1 2 3 4 5 6 7 8 9 Accuracy Sample Flow rate pH Transmitters 9135 0 - 14 pH 2 relay contacts +/- 0.01 10~15 pH LPH Conductivity 9125 0-100 uS/cm 2 relay contacts +/- 0.01% 10~15 Transmitter LPH Dissolved 9182 0-2000 ppb O2 2 relay contacts +/- 0.5 10~15 Oxygen Analyser ppb LPH Silica Analyser 9097 0-1000 ppb 2 relay contacts +/- 2 ppb 1~5 LPH Hydrazine Hydrastat 0-500 ppb Amperometric +/- 2 ppb 10~12 Analyser 9186 LPH Sodium Analyser Sodimat 0-01 ppb ~ Microprocessor <5% 3-5 9073 10000 ppm based Ion selective LPH Phosphate Phosphama 0-10 ppm Microprocessor 0.2 ppm 3 - 5 Analyser t 8892 based LPH Chlorine Chloromat 0-5 Mg/L Microprocessor ---3-5 Analyser 9184 based LPH Chloride 1517-A1 0-500 ppb Ion selective --50 Analyser (Orion) Ml/Min Applicable for all Analysers Power Supply Moniotor Enclosure Ambient Temperature Output signal 230 V AC, 50 Hz IP65 5 - 55 Deg C 2x4~20 mA
18

Model No. Range

Type

Typical Dimensions 144x144x150(D) 144x144x150(D) 144x144x150(D) 890x485x290(D) 300x1000x212(D) 485x890x285(D) 483x746x290 (D) --1180x300x130(D)

SWAS Dos And Donts Dos 1. Always keep the coolants flowing, even if sample is stopped. 1. Check the wiring for the possible loose connections. 2. Always flush all the Sample / Coolant lines before starting operation. 3. Ensure that sufficient differential pressure exists in coolant inlet and outlet header (Typical Minimum 2.5 Kg/sqcm) 4. Check for leaks in both sample and coolant lines. 5. Check whether the connections to valves / pressure regulator / Flow indicators etc are in right direction. 6. Always switch Off the mains power supply while carrying out any maintenance on the system Donts 1. Dont disturb the settings of Pressure regulator, temperature switch, pressure switch and safety valve without consulting service Engineer. 2. Dont stop the coolant supply, before isolating the sample supply. 3. Dont carry any maintenance function without isolating the sample supply. COMMISSIONING OF SWAS 1. All samples to be made available in Wet and dry panel as per requirement. 2. After commissioning and calibration of all Analysers, daily readings may be jointly recorded atleast for one month along with Customer chemist in the Observation sheets. Readings to be made available in DCS (MAX). 3. Alarms to be commissioned 4. Insulation of Chiller water lines to be done properly. 5. Operation of chiller to be ensured 6. BHEL site to handed over the required chemicals for 12 months for the operation of Silica, Phosphate, sodium, Hydrazine Analysers to customer Major highlights of SWAS MOM with Customer during Handing over of SWAS: 1. Date of handing over to be mentioned clearly. 2. List of Chemicals and other consumables with quantity to mentioned. 3. Handing over of all Manual and Procedures (including O&M Manuals of all Analysers). 4. All Normal / routine maintenance jobs involving Instrument re-calibration, fuse replacement, leakage found if any will be attended by vendor 5. Major defects like equipment failure due to design deficiency or manufacturing defect will be attended by BHEL during Warranty period. In such case BHEL will make all out effort to rectify the defect in shortest possible time. TYPICAL CALIBRATION ACCESSORIES OF SWAS Box No. 01: Chloride analyser calibrator set a. Calibrator with 230 V AC charger: b. Tube set with connector: c. Syringe: d. Electrode accessories with sand paper: Box No. 02: Chloride analyser calibrator set a. Calibrator with 230 V AC charger: b. Tube set with connector:

01 No. 01 No. 02 Nos. 01 No. 01 No. 01 No.


19

c. Syringe: 02 Nos. d. Electrode accessories with sand paper: 01 No. Box No. 03: Reagent bottles of Chloride with tubes (faulty reagents): 08 Nos. Box No. 04: Reagent bottles of Chloride with tubes (O.K reagents): 03 Nos. Box No. 05: a. DO2 kit: 06 Nos. b. Oxistat solution bottles: 06 Nos. c. Oxistat membrane box (total 14 nos. membrane): 06 nos. d. Packing circular SS plates: 06 Nos. e. Silica Analyser keys: 02 Nos. f. Fuses: 19 Nos. g. PVC connectors: 15 Nos. Box No. 06: a. Chlorine kit: 02 Nos. b. Chloromat refilling solution bottles: 03 Nos. c. Chlorine membrane box (total 06 membrane): 02 Nos. d. Plastic can cover: 03 Nos. Material given to Customer (MSEB) by M/S Forbes Marshall: Empty Plastic cans of 10 Liters capacity each for: 12 Nos. Storage / Maintenance of Analyser Reagents

DOCUMENTS & CERTIFICATES 1. TEST CERTIFICATES for the SWAS wet panel sample line Isolation valves (03 pages) 2. O&M Manuals of SWAS: a. Instruction Manual for PHOSPHAMAT - Model 8892: 03 Nos. b. Instruction Manual for SODIMAT Model 9073: 03 Nos. c. Instruction Manual for CHLOROMAT Model 9184: 03 Nos. d. Instruction Manual for Metering Pump (SWAS): 03 Nos. e. O & M Manual for Diamond Chiller (SWAS): 03 Nos. f. Instruction Manual for HYDRASTAT Model 9186: 03 Nos. g. Instruction Manual for DYNAMIC CALIBRATOR Model 15 DC: 03 Nos. h. Instruction Manual for CHLORIDE MONITOR: 10 Nos. i. Maintenance Procedure: 03 Set i. Sample Cooler Maintenance: ii. SWAS Maintenance iii. Sodium Analyser Maintenance iv. Hydrazine Analyser Maintenance v. Silica Analyser Maintenance vi. Phosphate Analyser Maintenance vii. DO2 Polishing procedure viii. Regeneration procedure of Cation Column ix. Instruction for changing over of compressors of chiller unit j. O&M Manuals of SOX/NOX/CO Analyser k. O&M Manuals of Oxygen Analyser PRECAUTIONS
20

General Precautions Replace the calibration solution when the empty alarm is flagged. Check the reagents level & replenish after 40 days. While making new reagents use de-ionised water which conductivity is < 0.2 s/cm. Use the all laboratory utensils of plastic (Not glass). If the plant is under shutdown for more than week, run the analyser with DM water for 2 to 3 hours. MAINTENANCE PROCEDURES SILICA ANALYSER Periodic: Check reagents solution level & replenish if low Monthly: Check for possible dirt deposits in the flow circuit. Quarterly: Change pumps tubes, each tube set will last 2x90 days by using the reversible feature. Annually: Clean level detector with a soft tissue. DISSOLVED OXYGEN ANALYSER Approximate membrane lifetime: 6 month depending on the sample Cleaning: Clean the instrument with a soft tissue and without any aggressive agent. Transmitter: The electronic unit amplifies the signal of the amperometric measuring cell and converts it into a direct digital readout in ppm, mg/l, g/l, Deg. C, Deg F. The transmitter comprises the following items: 1. Potentiostat which maintains the working electrode potential constant. 2. Amperometric measuring module. 3. Analog multiplexer 4. Microprocessor The analog multiplexer allows measurements to be acquired from the measuring cell, temperature sensor and internal checkpoints. Further, the microprocessor operates the relays, the RS485 interface (optional) and analog outputs (2 x 4 to 20 mA isolated from input signal, 800 ohms load maximum). The unit has a built-in concentration autoranging feature and a microprocessor operated calibration routine. The output of the Potentiostat is monitored for possible overdrive of the Potentiostat output stage. This condition can occur with the connections to the measuring cell open, inoperable electrodes or a defective reference electrode.

Oxygen Electrode Rejuvenation Procedure After some months of operation (3 to 12 depending on sample oxygen concentration, plant shut-down frequency, etc), a dark AgBr coating may cover part of the silver anode. This coating does not affect the measurement until more than 90 % of the surface is contaminated. When changing the electrolyte and membrane, visually check the silver anode. If more than 2/3 of the surface is covered then electrode rejuvenation according to the following procedure is needed: 1. Soak the anode in 10 % ammonia for about one hour then rinse it with demineralised water and wipe it with a soft cloth. 2. If the ammonia cleaning is not sufficient, rejuvenation of the silver electrode has to be done by repolishing softly (coating is an only few micron thick) the area covered with silver bromide with
21

soft abrasive (N400 to 600). After polishing, rinse the anode with demineralised water and wipe it with a soft cloth. Obviously, re-calibration is required. SODIUM ANALYSER Weekly: Check the level of Calibration solution. Check the level of conditioning reagents. Check the level of electrolyte (KCl) Monthly: Check visually the filter cartridge, change or clean Half Yearly: Check the flow meter calibration. Electrode Reactivation Procedure If the sodium selective electrode may loose its sensitivity because of clogging, reactivate electrode with following procedure: 1. Immerse half of the measuring electrode in a de-mineralised solution with 2 % of hydrochloric acid for 5 to 10 Seconds. 2. Rinse immediately & wash electrode with DM water. 3. Put the electrode in a KCl 0.1 m solution for a couple of hours before putting it back into the cell. PHOSPHATE ANALYSER Periodic: Check reagents solution level & replenish if low Monthly: Check for possible dirt deposits in the flow circuit. Quarterly: Change pump tubes, each tube set will last 2x90 days by using the reversible feature. Annually: Clean level detector with a soft tissue. HYDRAZINE ANALYSER Weekly: Check the level of sample conditioner. Replenish if low. Check the sample flow rate. Reset to 10 Lt/hr if it has drifted. Check the filter for dirt. Replace or clean if flow is impeded. Monthly: If the platinum electrode is deposited or if this electrode is dirty do not remove it for cleaning, but proceed as follows: 1. Turn off the sample flow needle valve. 2. Using a syringe, inject 5 % nitric acid into the measuring cell via the venturi-valve inlet tubing. 3. Wait 5 minutes for the acid react. 4. Turn the sample flow again and set it to 10 Lt/hr 5. Recalibrate the instrument. Half Yearly: Check the flow meter calibration. Regeneration Procedure Of Cation Column Remove the Cation column, which is exhausted. Take 25 ml (36%) HCL in beaker, dilute to 250 ml in water add 1000 ml DM water = 1250 ml. Take above solution in 5-liter bucket. Add cation resin into above solution; stir well about 30 to 40 minutes. Rinse the resin 3 to 4 times by using five-liter DM water every time. Fill the resin in to vessel along with DM water.
22

Always keep resin in wet condition by using low conductivity water.

CHLORIDE ANALYSER Daily: Ensure that: 1. Pressure: 15 psi 2. Flow: 40 ml/min 3. Level of Reference solution in bottle Not to increase 4. Level of Reagent in Bottle: Not to increase Level in Reagent bottle should not increase more than the available level. If it increases than it indicates that there is a leakage in tube inside the bottle. Weekly: 1. All the points of Daily Maintenance. 2. Chloride Analyser Refrigeration (Chiller) unit water level should be maintained. Monthly: 1. All the points of Weekly Maintenance. 2. Calibration of the Instrument (chloride Analyser). Three Monthly Maintenance: 1. All the points of Monthly Maintenance. 2. Change the Reagent bottle if empty. 3. Change the Diffusion tube. 4. Change the Reference Solution bottle if empty. Yearly Maintenance: 1. All the points of 3 monthly Maintenance. 2. Change the Electrode and De-ionization Cartridge. Things to do during Plant Shut Down Because leaving reagent bottle on instrument for unit shutdown of ONE DAY or LONGER will cause Chloride Electrode Failure, diffusion tubing breakage, exhaustion of deionization cartridge and corrosion in cooling coils. 1. Change the Reagent Bottle with the bottle filled with DM Water. 2. Remove both the electrode carefully from the Flow Cell and clean with DM water. Wipe it with tissue paper and let them hang by their connectors. 3. Remove the Reference Solution from the Electrode and rinse clean with DM water. Wipe it with tissue paper, as salt will be formed at the Reference junction. Low-Level Chloride Monitor (Product Number: 1517) Model 1517 Low-Level Chloride Auto Zero Monitor, complete with chloride electrode (100025), reference electrode (100031), with one 2-oz. bottle internal filling solution, deionization cartridge (150013), shorting straps (150030), auxiliary cooling unit (60 Hz), and one instruction manual. Requires reagent (151711), or consumables kit (151750), and dynamic calibrator (150095). Chloride Consumables Kit (Product Number: 151750) Chloride 1 Year Consumables Kit, includes four 1-liter bottles reagent with diffusion tubing (151711), chloride electrode (100025), reference electrode (100031), with one 2-oz. bottle internal filling solution, five 2-oz. bottles reference electrode internal filling solution (150071), deionization cartridge (150013), four sets inlet filter (4 x 151720), two 60 micron filters (181170), four sets O-ring kit (4 x 151735), syringe kit (150096), and one pint chloride standardizing solution (941707). SAMPLE COOLER MAINTENANCE (SWAS)

23

Keep the heat transfer surface clean. Brush the internal surface of shell & inner cylinder by steel wire to remove scale deposits. Check regularly for any clogging of the small bore coil by foreign particles or sediments, specially the bottom portion. Physically inspect the parts as a good engineering practice. During operation, the drain plug provided on the shell can be occasionally opened and cooling water allowed to blow through allowing the sediments to be ejected. Do regularly tighten the flanges bolts to ensure leak tightness. Dont ever shut off the cooling water supply before shutting off the steam sampling line. This may overheat the coil tube. Temperature of Sample line was not being maintained at 175 Deg C as envisaged in system

PROBLEMS ENCOUNTERED 1. Chiller Unit for Sample Cooling: Motor winding got burnt due to voltage fluctuation in customer Power supply. Both Chiller pumps failed 2. DM Water Pump: DM water pump, which is used to pump the DM water sample to SWAS Wet panel, has failed. It was sent to their vendor works for repair 3. Cooling Water Sample Line: This line got choked very frequently due to muddy cooling water. 4. SWAS Panel: SWAS Room layout was changed by vendor at latter date. Due to this the location of wet panel and dry panel has also changed. Sensor cable with 4Cx1.5 sqmm and 6Cx1.5 sqmm in 10 meters length loose pieces. After change in room layout the sensor cable required between wet panel and dry panel were of 20 meter length pieces. New cables were procured. 5. Dissolved Oxygen Analyser (Dearator Do2 Analyser) Of Unit 4 SWAS System: Electrodes silver coating pilled off during calibration in the presence of Vendor representative. 6. Hotwell Conductivity Module Failed During Operation: Card was replaced by vendor. 7. Isolation Valves: As per the specification the Isolation valves, which came, fitted with SWAS wet panel were not of SS316 material. Vendor was informed about the discrepancy and therefore vendor replaced 16 nos. isolation valves per unit.

THE END Prepared by: Vinay Kuhikar, Sr. Manager, BHEL, PSWR