Image acquisition for colocalization using optical microscopy

David R. L. Scriven, Ronald M. Lynch and Edwin D. W. Moore


Am J Physiol Cell Physiol 294:C1119-C1122, 2008. First published 19 March 2008; doi: 10.1152/ajpcell.00133.2008 You might find this additional info useful... This article cites 13 articles, 2 of which you can access for free at: http://ajpcell.physiology.org/content/294/5/C1119.full#ref-list-1 This article has been cited by 4 other HighWire-hosted articles: http://ajpcell.physiology.org/content/294/5/C1119#cited-by Updated information and services including high resolution figures, can be found at: http://ajpcell.physiology.org/content/294/5/C1119.full Additional material and information about American Journal of Physiology - Cell Physiology can be found at: http://www.the-aps.org/publications/ajpcell This information is current as of February 4, 2013.
Downloaded from http://ajpcell.physiology.org/ by guest on February 4, 2013

American Journal of Physiology - Cell Physiology is dedicated to innovative approaches to the study of cell and molecular physiology. It is published 12 times a year (monthly) by the American Physiological Society, 9650 Rockville Pike, Bethesda MD 20814-3991. Copyright 2008 the American Physiological Society. ISSN: 0363-6143, ESSN: 1522-1563. Visit our website at http://www.the-aps.org/.

Am J Physiol Cell Physiol 294: C1119C1122, 2008. First published March 19, 2008; doi:10.1152/ajpcell.00133.2008.

Perspectives in Cell Physiology

Image acquisition for colocalization using optical microscopy


David R. L. Scriven,1 Ronald M. Lynch,2 and Edwin D. W. Moore1
Department of Cellular and Physiological Sciences, University of British Columbia, Life Sciences Institute, Vancouver, British Columbia, Canada; and 2Department of Physiology, University of Arizona, Arizona Health Sciences Center, Tucson, Arizona
Submitted 29 February 2008; accepted in nal form 17 March 2008
1

Scriven DR, Lynch RM, Moore ED. Image acquisition for colocalization using optical microscopy. Am J Physiol Cell Physiol 294: C1119C1122, 2008. First published March 19, 2008; doi:10.1152/ajpcell.00133.2008.Colocalization, in which images of two or more uorescent markers are overlaid, and coincidence between the probes is measured or displayed, is a common analytical tool in cell biology. Interpreting the images and the meaning of this identied coincidence is difcult in the absence of basic information about the acquisition parameters. In this commentary, we highlight important factors in the acquisition of images used to demonstrate colocalization, and we discuss the minimum information that authors should include in a manuscript so that a reader can interpret both the uorescent images and any observed colocalization.

THE ANALYSIS AND DISPLAY

of colocalization in optical images obtained through microscopy is now a commonly used experimental tool. In this technique, two or more molecules are labeled with uorescent probes, and the images, acquired at the appropriate wavelengths, are overlaid. The extent of colocalization of the probes is then visualized and often quantied. This general description of a colocalization experiment covers a host of microscopic techniques, labeling methods, and image analysis algorithms that can be used in various combinations. To provide some uniformity in presentation, we are recommending several procedures for collecting colocalization data. We also present guidelines for the information that should be included in the methods section of the manuscript that will allow any reader to interpret the results. We are focusing our discussion on images acquired with wide-eld and confocal microscopes since these are by far the most common approaches for image acquisition. The Nyquist Criteria

When the term colocalization is used in optical microscopy, the objects that are being compared are usually pixels or voxels in digital images (A pixel is the smallest element in a twodimensional digital image and a voxel is its three-dimensional counterpart). When we ask if two molecules are colocalized, we are, in effect, asking Is the voxel illuminated by a uorophore attached to (or part of) molecule A the same voxel as that illuminated by another uorophore attached to molecule B? Thus the size of the voxels is important for both planning and interpreting the experiments. When the image is digitized (voxelized), it has the same effect as placing a grid over the image and replacing the objects at each grid point (voxel) with the total intensity of all of the objects contained in it. As voxel
Address for reprint requests and other correspondence: E. D. W. Moore, Dept. of Cellular and Physiological Sciences, Univ. of British Columbia, Life Sciences Institute, 2350 Health Sciences Mall, Vancouver, British Columbia V6T 1Z3, Canada (e-mail: edmoore@interchange.ubc.ca). http://www.ajpcell.org

size increases, resolution decreases so that separate structures become lumped together in the same voxel. At some point, the voxelized image will no longer accurately represent the original object and artifacts may appear. The voxel size at which an object is accurately represented by an image is given by the Nyquist criteria, which are a function of several variables: the numerical aperture of the objective, the emission wavelength of the uorophore, the type of microscope (wide eld or confocal) and, for depth, the refractive index of the medium. Further details can be found in Castleman (2) and van der Voort and Strasters (13). The formulas and a summary of appropriate values for a conventional microscope, using oil immersion lenses of various numerical apertures and illuminating wavelengths, are given in Table 1. Confocal microscopes, using laser light and a typical pinhole size of 250 nm, have Nyquist values that are similar to those listed (13). Ideally, imaging parameters should be optimized to reach these values to obtain the best resolution for evaluating colocalization. In real systems, noise will increase the theoretical values listed in the table, and values within 10% of those tabulated will still give an acceptable result. Voxels larger than those specied by the Nyquist criteria will under sample the image, create artifacts, and result in false colocalizations. Figure 1 shows a ventricular myocyte collected at voxel sizes satisfying the Nyquist criteria (Fig. 1, A and C) and the same image resampled with voxels four times larger in X and Y and 1.6 times in Z (Fig. 1, B and D). It is clear that not only does a large amount of colocalization appear in the under-sampled image but more importantly, structures appear that are not present in the original (compare Fig. 1, C and D). Collecting images with voxels smaller than the Nyquist values requires longer exposure times to achieve a comparable signal-to-noise ratio, increasing photobleaching and photodamage, while producing little additional information. For proper evaluation of a manuscript, authors should therefore report the objectives numerical aperture, the wavelengths that are being used, the voxel sizes, the refractive index of the medium and the Z spacing between image planes, and show how these values relate to the Nyquist criteria for the experiment. Diffuse Labeling In some cases, one of the uorophores is present in almost every voxel. This can occur when labeling a cytosolic protein or a widely distributed protein like actin. Under these conditions, colocalization by chance has almost 100% probability, making any observed or measured colocalization, not signi-

Downloaded from http://ajpcell.physiology.org/ by guest on February 4, 2013

The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked advertisement in accordance with 18 U.S.C. Section 1734 solely to indicate this fact. C1119

0363-6143/08 $8.00 Copyright 2008 the American Physiological Society

Perspectives in Cell Physiology C1120


IMAGE ACQUISITION FOR COLOCALIZATION

Table 1. Lateral and axial sampling distances determined by the Nyquist criteria
NA Fluorophore Emission Wavelength, nm 1.2 1.3 1.4

DAPI/Alexa 350 Cy2 FITC/Alexa 488 Cy3/Alexa 555 TRITC/Alexa 546 Texas red/Alexa 594 Cy5/Alexa 647

450 505 520 565 575 615 670

97 394 105 426 108 441 118 479 120 487 128 521 140 568

89 316 97 341 100 353 109 383 111 390 118 417 129 455

83 248 90 269 93 278 101 302 103 307 110 329 120 360

The x and y sampling distances are calculated (in nm) using the formula /(4NA) (4) where is the emission wavelength (in nm) and NA is the numerical aperture of lens. The axial (z) sampling distances (in italics) are for an oil (n 1.515) immersion objective and are calculated with the formula, z /[2n (1 cos)], where sin1 (NA/n) and n is the refractive index of the medium (13). DAPI, 6-diamidino-2-phenylindole; TRITC, tetramethylrhodamine isothiocyanate, Cy2, Cy3, Cy5, different uorescent cyanine dyes.

the other. Bleed through of signal from one channel to another can result in an apparent colocalization when none, in fact, exists. For example, even with modern lters, there is significant bleed through from FITC to rhodamine, or from cyan uorescent protein to yellow uorescent protein. Bleed through should be estimated by creating a control with labeling in one channel and none in the other and then recording the signal coming from the unlabeled wavelength(s). The resultant image should be random, having no discernable pattern, and its intensity should be less than the chosen threshold. If, however, there is signicant bleed through, the control images can be used in concert with spectral linear unmixing (8, 14) to separate the two signals. Once corrected, the images can then be used for analysis. Some newer instruments have spectral unmixing built into their software and correct for bleed through at the time of acquisition, although it is important to check how well the signals are separated before accepting the results (15). Colocalization in X-Y versus Z In all optical microscopes, except the 4Pi, the resolving power in Z is only one-third to one-quarter of that in X and Y (9). The consequence is that colocalization is most powerful when performed on structures that run along the XY plane, parallel to the coverslip, whereas it is least accurate for objects that are running parallel to the Z-axis of the microscope. The consequences of this are best seen in a structure that runs both parallel and perpendicular to the XY plane. This is diagrammed in Fig. 2A, which shows a cross-section through a cell in which uorophores of the same color are separated by 200 nm, and the two different uorophores are interspersed at 100-nm intervals. Figure 2B shows the result of superimposing these planesthe pixels that lie in the transverse plane are not colocalized, but those at the edges, lying along the Z-axis appear to be. This effect can be observed in real cells. Figure 2C shows a rat ventricular myocyte labeled with antibodies specic for vinculin (green) and caveolin-3 (red). A 1-m thick section from the cells center is displayed in Fig. 2Ci and demonstrates almost complete colocalization at the cell surface

Downloaded from http://ajpcell.physiology.org/ by guest on February 4, 2013

cant. Experimental approaches other than colocalization by optical microscopy are required to resolve this problem. Deconvolution Wide-eld images contain much out-of-focus light, which will confound colocalization analysis. For this reason, it is mandatory that wide-eld images be deconvolved before colocalization analysis is attempted. Authors should state which algorithm or software package was used. It should be noted that the measurement of colocalization in confocal images is also improved by deconvolution (7, 11). Bleed Through A fundamental assumption made when measuring colocalization is that the signal at each wavelength is independent of

Fig. 1. The effects of pixel size: An image, 2 m thick, of a rat ventricular myocyte labeled with antibodies for vinculin (green) and caveolin-3 (red), with colocalized voxels white. The scale bar is 5 m and the voxel dimensions are: (A) 100 100 250 nm, which satises the Nyquist criteria and (B) 400 400 400 nm, which does not. The colocalization of vinculin with caveolin-3 (calculated as the percentage of voxels labeled for vinculin that also contained caveolin-3) was 19% in A and 61% in B and the colocalization of caveolin-3 with vinculin was 6% in A and 32% in B. (C, D) a blowup of the inserts (white square) in A and B respectively. Note the structural artifacts in the under-sampled image D that do not appear in the original (A, C).

AJP-Cell Physiol VOL

294 MAY 2008

www.ajpcell.org

Perspectives in Cell Physiology


IMAGE ACQUISITION FOR COLOCALIZATION

C1121

running parallel to the Z-axis (arrowheads). Figure 2Cii is an image of the top surface of the same cell running parallel to the XY axis, which shows 5% colocalization (boxed regions). A stereo pair of the complete data stacks (Fig. 2Ciii) shows that colocalization is conned to those regions of the sarcolemma that are oriented largely parallel to the microscopes Z-axis, but these molecules are, in fact, not colocalized. Sometimes the objects to be colocalized have little or no X-Y component (e.g., synapses or gap junctions), and in this case the measured colocalization has less predictive power than it would for objects oriented in the X-Y plane, since the colocal-

ized objects could be separated by 500 nm or more (i.e., 2 the Nyquist limit in Z, see Table 1). Sampling the Cell The great advantage of using optical microscopy to assess colocalization, as opposed to other techniques including electron microscopy, is that images of entire cells or their substructures can be readily collected and analyzed, and regional variations can be visualized. When focusing an analysis on specic regions of a cell, care needs to be taken when deciding

Downloaded from http://ajpcell.physiology.org/ by guest on February 4, 2013

Fig. 2. Edge Effects causing Colocalization. (A and B): A diagram showing how the effect occurs. (A) A cross-section in which uorophores of the same color are separated by 200 nm, and the two different uorophores are interspersed at 100 nm intervals. The grid represents 4 image planes, 250 nm apart in Z and 100 100 nm in X and Y. (B) Projection of the image planes onto a line of voxels; those in which colocalization occurs are colored yellow. (C) A rat ventricular myocyte labeled for vinculin (green) and caveolin-3 (red), colocalized voxels are white. 2 m thick sections from the center of the cell (i) and the cell surface (ii). The arrowheads indicate sarcolemma that is parallel to the optical axis, and the arrows indicate sarcolemma that is perpendicular to the optical axis. The boxed regions, in which only 4% of the vinculin and 5% of the caveolin-3 were coincident, were used to numerically assess colocalization. (iii) Stereo-pair, 6 rotation, of the entire data set, 15 m deep.

AJP-Cell Physiol VOL

294 MAY 2008

www.ajpcell.org

Perspectives in Cell Physiology C1122


IMAGE ACQUISITION FOR COLOCALIZATION Lynch), a Grant-in-Aid from the Heart and Stroke Foundation of British Columbia and the Yukon, and by grants from the National Sciences and Engineering Research Council and the Canadian Institutes of Health Research (to E. D. W. Moore). REFERENCES 1. Anderson JB, Carol AA, Brown VK, Anderson LE. A quantitative method for assessing co-localization in immunolabeled thin section electron micrographs. J Struct Biol 143: 95106, 2003. 2. Castleman KR. Resolution and sampling requirements for digital image processing, analysis, and display. In: Electronic Light Microscopy, edited by Shotton D. New York: Wiley-Liss, 1993, p. 7193. 3. Dan P, Lin E, Huang J, Biln P, Tibbits GF. Three-dimensional distribution of cardiac Na-Ca2 exchanger and ryanodine receptor during development. Biophys J 93: 2504 2518, 2007. 4. Gonzalez RC, Woods RE. Digital Image Processing. New York: Addison-Wesley, 1992. 5. Grifths G, Parton RG, Lucocq J, van Deurs B, Brown D, Slot JW, Geuze HJ. The immunouorescent era of membrane trafc. Trends Cell Biol 3: 214 219, 1993. 6. Jares-Erijman EA, Jovin TM. FRET imaging. Nat Biotechnol 21: 13871395, 2003. 7. Landmann L. Deconvolution improves colocalization analysis of multiple uorochromes in 3D confocal data sets more than ltering techniques. J Microsc 208: 134 147, 2002. 8. Nadrigny F, Rivals I, Hirrlinger PG, Koulakoff A, Personnaz L, Vernet M, Allioux M, Chaumeil M, Ropert N, Giaume C, Kirchhoff F, Oheim M. Detecting uorescent protein expression and co-localisation on single secretory vesicles with linear spectral unmixing. Eur Biophys J 35: 533547, 2006. 9. Schrader M, Bahlmann K, Giese G, Hell SW. 4Pi-confocal imaging in xed biological specimens. Biophys J 75: 1659 1668, 1998. 10. Scriven DR, Klimek A, Asghari P, Bellve K, Moore ED. Caveolin-3 is adjacent to a group of extradyadic ryanodine receptors. Biophys J 89: 18931901, 2005. 11. Sedarat F, Lin E, Moore ED, Tibbits GF. Deconvolution of Confocal Images of Dihydropyridine and Ryanodine Receptors in Developing Cardiomyocytes. J Appl Physiol 97: 1098 1103, 2004. 12. Sekar RB, Periasamy A. Fluorescence resonance energy transfer (FRET) microscopy imaging of live cell protein localizations. J Cell Biol 160: 629 633, 2003. 13. van der Voort HTM, Strasters KC. Restoration of confocal images for quantitative image analysis. J Microsc 178: 165181, 1995. 14. Zimmermann T, Rietdorf J, Pepperkok R. Spectral imaging and its applications in live cell microscopy. FEBS Lett 546: 8792, 2003. 15. Zucker RM, Lerner JM. Wavelength and alignment tests for confocal spectral imaging systems. Microsc Res Tech 68: 307319, 2005.

which part of the cell is sampled, because this can dramatically affect the results. Specically, two recent papers (3, 10) have shown that regional variations in colocalization can occur in the absence of any obvious structural differences. Guidelines for Publication To enable readers to interpret images and to evaluate the meaning of colocalization from a given experiment, manuscripts should include the following information: The pixel sizes, the Z spacing between image planes, the refractive index of the medium, and how this relates to the Nyquist criteria for the experiment (Table 1). The objectives magnication and numerical aperture. For images obtained with a confocal microscope, either the pinhole size or the axial resolution should be reported. The emission wavelengths of the uorophores and the lter sets used. Scalebars should be included in each image group. In addition, the following conditions should be satised. Wide-eld images must be deconvolved, and the methodology reported. Images in which the colocalization is primarily along the Z-axis should, where possible, be examined for edge artifacts. Bleed through should be evaluated and corrected for, if present. If corrections are made, details of the methodology used should be provided. Conclusion Colocalization can be a powerful method for determining whether molecules have the potential to associate or interact, but it cannot conrm that they do. The inherent limitations of the technique mean that conrmation may require electron microscopy (1, 5) or other microscopic techniques such as uorescence resonance energy transfer (6, 12).
GRANTS This work was funded by the American Heart Association GIA 93014700 and National Heart, Lung, and Blood Institute Grant HL-66044 (to R. M.