Académique Documents
Professionnel Documents
Culture Documents
LAB INCHARGE:
FACULTY TEAM:
Experiments List:
A. Calibration of Instruments:
1. Calibrate the given Bourdon Gauge using the Mercury Filled Manometer. Also find out the span, error and accuracy of the bourdon gauge. 2. Calibrate and report the accuracy of the given Bourdon Gauge using Dead Weight Tester. 3. Calibrate the Given Resistance Temperature Detector using the mercury filled Thermometer. 4. Calibrate the given Thermocouple using Thermometer.
B. Process Analysis:
5. Find out the time constant of the given Mercury Filled Thermometer and also find the response y(t) of the system when t =, t = 2, t = 3 6. Find out the time constant of the Liquid Level System
C. Control Loops:
7. Report the response and variations in the process variable PV( flow of water)output in i. Proportional Mode (P- mode) by giving the values of gain as 1.00, 0.6 and 1.6 while the output of controller set on 30 % i.e. set point = 30 %. Proportional Integral mode (PI mode) by giving the values of Reset time, as 0.1, 0.15 and 0.5 while the output of controller set on 30 %. Proportional Integral Derivative mode (PID mode) by giving the values of rate minute as 1, 2, 3 while the output of controller set on to 30 %.
ii.
iii.
8. Study the response of the process variable (temperature) in the on-off algorithm. Also plot the graph between time and temperature during heating and cooling.
Table of Contents:
Contents 1. Calibration of Bourdon gauge using Manometer 2. Calibration of Bourdon gauge using dead weight tester 3. Calibration of RTD using thermometer 4. Calibration of thermocouple using thermometer 5. Time constant of thermometer 6. Time constant of liquid level system 7. Flow control loop 8. Temperature control loop Page no. 4 8 12 15 18 22 26 32
Apparatus:
Bourdon gauge Mercury filled thermometer Air compressor
Procedure:
Start the air compressor before performing the experiment. Mercury level in the U-shaped manometer is checked. The Manometer shows the value of the change in pressure by inches of mercury and Bourdon gauge report the value of same pressure in psi or bar. The readings of both the manometer i.e., inches of mercury and Bourdon gauge i.e., psi or bar are noted after the alteration in the pressure by means of controlling valve. The readings of manometer i.e., in inches of mercury are converted into psi by multiplying with an appropriate factor. The graph between manometer readings at X-axis and the Bourdon gauge reading at Y-axis is plotted. A straight line at 45 form origin is drawn. The maximum difference between the actual plotted line and the 45 line is the Span error. Formulae Used: 1. Pressure Reading using Manometer
( Hg air ) gh P = gc
Where Hg is density of Mercury air is density of Air h = difference in height between two limbs of manometer 2. Average Error
n E AvgError = i =1 n
X = % Error =
RESULT:
Objective:
To calibrate and report the accuracy of the given Bourdon gauge, using dead weight tester.
Apparatus:
Bourdon gauge Dead weight tester Weights.
Working Principle:
The working principle of the above depicted Dead weight tester is based on Pascal Law. This law states that if pressure is applied on a fluid at rest, the pressure is equally distributed to all directions, i.e. one to the piston of the dead weight tester and the other to the Bourdon gauge.
Procedure:
1. 2. 3. 4. 5. 6. 7. Initially the gauge which is to be calibrated is connected with the dead weight tester. The lever is moved outward completely After putting a pressure plate of suitable weight upon dead weight tester, the lever is During closure of lever, Bourdon gauge showed increase in pressure. The piston is moved At that point Bourdon gauge reading is noted. Again the same procedure is repeated right from the beginning by placing another pressure plate on the top of already placed pressure plate. The graph between dead weight tester reading on X-axis and Bourdon gauge reading on Y-axis is plotted.
gradually moved outwards. inwards until the scale of Bourdon gauge stops with the jerk.
Formulae used:
8
1. Accuracy:
X = % Error =
Where E = Error for ith observation n = number of observations Observations & Calculations: Pressure applied Sr. No. by weights, P1 kg/cm2 Bourdon gauge Reading P2 kg/cm2 Error, E=P1-P2 kg/cm2
Calibration Curve:
Result:
Results Accuracy Range Average Error
10
To calibrate the given Resistance Temperature Detector using the Mercury Filled Thermometer.
Apparatus:
Resistance Thermometer (Platinum) Beaker Oil bath Thermometer Avometer
Procedure: Initially the Avometer is standardized by joining the two ends of the wires of RTD to Avometer and then wires are short circuited in order to set the pointer at zero. The resistance thermometer is inserted in the oil beaker which already had the mercury filled thermometer. The oil in the beaker is heated and different sets of readings are taken for resistance and the temperature for every five degree centigrade temperature rise. Finally a graph between temperature and the resistance is plotted. The straight line drawn showed the fitness of the resistance thermometer under consideration for the required purpose.
Observations:
12
Sr. No.
Temperature (C)
Resistance (ohm)
Calibration Curve:
Summary Of RTD
Category Temperature sensor 13
Resistance of metal increase with increase Working Principle Material of construction Sensitivity Limits of application in temperature. Platinum Nickel 0.004/C to 0.005/C Up to 650C for Platinum Up to 300C for Nickel High accuracy. Wide Range of application Advantages Good reproducibility Higher signal to noise ratio. Can be used in radiation environment Slower response time Disadvantages Expensive
Apparatus:
Potentiometer Thermocouple Oil bath Mercury filled Thermometer
Procedure:
1. The standardization of potentiometer is done, when the thermocouple is not connected to it. 2. After connecting the thermocouple with potentiometer, the thermocouple and a mercury filled thermometer are inserted into oil beaker. 3. The oil in beaker is heated up to 180 C then for every 5 C drop in temperature a corresponding change in E.M.F. is noted via potentiometer. 4. The change in E.M.F. along with the change in temperature of the system is plotted on a graph.
No. of Obs.
E.M.F. mV
Calibration Curve:
15
Summary of thermocouple
16
Temperature Sensor
Material of construction
Copper-Constantan Platinum,Rhodium-Platinum
Limits of application
-100 to 1100C Low cost with reasonable accuracy. Wide Range of application
Advantages
Disadvantages
Low value of emf is corrupted with noise Temperature is not exactly linear with emf
Type E- Type E (chromelconstantan) Type K- Type K (chromel{90 percent nickel and 10 percent chromium}alumel)
Thermocouple types
Type J- Type J (ironconstantan) Type N- Type N (NicrosilNisil) Type R- Type R thermocouples use a platinumrhodium alloy containing 13% rhodium
17
Apparatus:
Two mercury filled thermometer Oil bath Heating device Stop watch.
Procedure:
1. Initially the room temperature is noted. 2. The mercury filled thermometer is dipped in the oil bath which is placed on the heating arrangement. 3. The whole arrangement is heated till the temperature of the bath is reached to 220C. The attained temperature is maintained. 4. Another thermometer, whose time constant is to determine is dipped in the same oil bath and the rise in temperature after every 5 seconds is noted along with time. 5. The required parameters are calculated and a graph between above mentioned quantities is plotted.
18
19
Response of system at when t=, t=2, t=3. At t/ = 1, Y(t)/A = 0.63 (From graph) Y(t) = 0.63 x A C Results: Results (sec) Y(t) when t/ = 1 Y(t) when t/ = 2 Y(t) when t/ = 3
Time Constant
C C C
Response of System
20
Summary of thermometer
Category Working Principle Material of construction Temperature Measuring Device Expansion of fluid with increase in temperature. Ordinary Soda Lime Fused Quartz Hg is opaque. Hg is good conductor. It does not wet the glass surface Up to 350C for mercury Less than 120C for alcohol Low cost Can be used easily. Advantages Used as STANDARD equipment for calibration of temperature sensors Good accuracy Disadvantages Cannot be used in industry for automatic control.
Advantages of mercury
Limits of application
21
Apparatus:
Storage tank, Tubes (of certain lengths), Beaker, Scale Stop watch.
Procedure:
1. A tube of certain length and diameter was fitted at the bottom of tank. Storage tank was filled up to a certain level (say h 1). A finger was put at the end of the tube so that no water can flow. 2. Then a beaker was placed under the tube and the finger was removed from the lower end of the tube. The water began to flow and at the same time, a stopwatch was operated and the time for which the level of water fell to a certain height in the storage tank (say h2) was noted. The volume of the water that fell into the beaker was measured. 3. The mean of height or level of water was also noted. 4. Diameter of Storage tank was also measured. 5. By drawing a graph of mean level of water (along X axis) vs. the flow rate(along Y-axis), the resistance was noted. 6. The time constant of the system can be noted by the formula. Time Constant = Resistance x Storage Capacitance 7. The experiment was repeated by taking tubes of different length and diameter.
22
Formulae Used:
1. Area of tank A= Where d = diameter of tank 2. Time Constant: =RxA Where A = Area of Tank R= Resistance to flow 3. Resistance (from graph): cm2
Serial # 1 2 3 4
Volume (ml)
23
5 6 7
4 3 2
12
10
) s / m ( Q , e t a r w o l F
4 2 0 2 4 6 8 10 12
24
Results:
Results Dia of tube = constant = d cm Time Constant for L1 = Time Constant for L2 = Time Constant for L3 = 1 2 3
Length of tube = constant = L cm Time Constant for d1 = Time Constant for d2 = Time Constant for d3 = 1 2 3
25
Procedure:
1. The pump should not be energized/ operated until your control setting is complete and whenever you are giving new values/changing values of gain or etc the pump should be switched off. 2. 3. Initially the controller is to be set in Auto mode. Then the set point is brought to 30 % by increasing or decreasing keys without touching any other keys. 4. Then the setup key is to be pressed once to have tuning mode of controller, value of Gain is adjusted by increasing/ decreasing keys and give a specified value i.e. 1.0. 5. 6. Now press function key once and adjust value of Rate minute at 0.00. Again press the function key to adjust value of Reset Rate (of integral mode) to 0.10 or a high value than this. 26
7.
The pump is still not started. When the pump is started immediately one person will record the readings of magnetic flow transmitter and other person simultaneously record the reading (maxima and minima)of the output displayed with reference to the time i.e. readings can be taken after every 5 seconds or any other time which may be suitable. Also observes how quickly the system stabilizes.
8.
When set point is achieved and approximately oscillations are stopped then the pump is to be stopped from the control panel switch. Same procedure is repeated for different gain values (0.6, 1.6) and keeping other parameters constant.
Procedure: Methodology adopted would be the same and the variables will have the following values: Gain = 0.6 for all the three inputs Rate minute = 0.00 for all the three inputs Reset min = 0.1, 0.15 and 0.5
3) PID Mode
Procedure: Methodology adopted would be the same and the variables will have the following values, Gain = 0.6 for all the three inputs Rate minute = 1, 2, 3 Reset min = 0.1 for all the three inputs
27
Block Diagram:
28
Flow Diagram:
1. Storage Tank 2. Pump 3. Control Valve 4. Restriction Valve 5. Bottom valve FT-1: Flow Transmitter FC-1: Flow Controller LC-2: Positioner PI-1: Pressure gauge 29
2) PI mode
Proportional Controller
Set point=30% Plot Graph for all gains by taking time against x-axis and Controller Output across Y axis.
30
31
Gain = 0.6 Tuning Parameters Time (s) 0 5 10 15 20 25 Integral time=0.1 Rate min=1 Controlled Variable (%TO)
32
Procedure:
1. Initially the water supply tube from the cooling element is connected to the main water supply. 2. The outlet tube from the solenoid valve is connected to the drainage. 3. The process container is filled with water so as to cover the cooling element. 4. The electric power supply is connected. 5. The controller is set according to the following procedure. 6. The set point Select key is pressed until Algorithm is displayed. 7. The set point value is set to the required value (according to heating or cooling system) 8. For the above set point, note down reading of temperature against time. 9. Plot a graph b/w temperature (along y-axis and time (along x-axis) both for cooling and heating process.
33
Sr no. 1. 2. 3. 4. 5. 6. 7. 8. 9. 10.
Time (sec) 5 10 15 20 25 30 35 40 45 50
34
Reference:
1. Coughanowr, D. R.:Process Systems Analysis and Control, 2nd ed., McGraw-Hill. Sr no. 1. 2. 3. 4. 5. 6. 7. 8. 9. 2. 3. Time (sec) Temperature (oC)
Fribance, A. E.: Industrial Instrumentation Fundamentals, McGraw-Hill, 1962. Luyben, W. L.: Process Modeling, simulation, and Control for Chemical Engineers, 3rd ed,McGraw-Hill, Inc., 1997.
4.
Seborg, D. E. et al.: Process Dynamics and Control, 2nd ed, John Wiley & Sons, Inc., 1989.
35
5.
Stephanopoulos, G.: Chemical Process Control, An Introduction to Theory and Practice, Prentice-Hall, Inc., 1984.
6.
Bela G Liptac.:,Instrument Engineers Handbook, Process Control and Optimization, Vol 2, 4th ed. Taylor & Francis, 2006.
7.
Smith C.A. and Corripio B.A., Principles and Practice of Automatic Process Control. 3rd Edition, J.W. & Sons, New York, USA. 2006.
8.
Richardson J.F. and Peacock D.G.: Chemical Engineering, Volume 3, 3rd Edition, Butterworth Heinemann, 2010
9. 10.
11.
Marlin, Thomas, Process Control, Designing Processes and Control Systems for Dynamic Performance,3rd Ed., McGraw-Hill, New York, 2000
Websites: 1. 2.
3.
http://www.isa.org/ http://www.manchester.ac.uk
(www.fieldbus.org/endusersupport/enduserpresentations/interkama2004/presentations/chris baltus_ffbenefits.pps) http://www.sensotec.com/intrinsically_safe_rated.asp#ex http://www.omega.com/thermocouples.html
4. 5.
36