Académique Documents
Professionnel Documents
Culture Documents
Description
The GSP instrument is a high throughput analyzer intended for quantitative or qualitative measurement of neonatal screening samples on 96-well microplates. Operated by means of a touch-screen, the instrument is fully automated, performing every stage of an assay from retrieval of the sample plate from the stacker to measurement and reporting of results.
System Specifications
Physical dimensions Power requirements Environmental conditions Noise Height Width Depth Weight Power consumption Mains voltage Operating Temperature Temperature range Temperature range < 60 dB 1960 mm (77.2) 1310 mm (51.6) 760 mm (30.0) 610 kg (1345 lb) 1500 VA 100 - 240 V, 50 - 60 Hz 18 - 30 C 18 - 27 C 28 - 30 C Relative humidity Relative humidity 10 80 % 10 65 %
The water used for preparing wash solution and for rinsing has to be deionized. Samples Sample type Loading Bulk Reagents Washing and rinsing Wash solution Automatic dilution of Wash concentrate from a mains water line (deionized) or from an external bottle outside the instrument Sample identification Continuous sample loading Continuous reagent loading Continuous bulk reagent loading Blood disks from dried blood spots punched into 96-well plates Barcode reader or keyboard input
Rinse water Automatic filling from a mains water line (deionized) or from an external bottle outside the instrument Liquid waste Waste capacity Automatic disposal if connected to a waste line OR manual disposal 16 L (15 plates)
Performance Specifications
Capacity Sample capacity Reagent capacity Heated incubator module Shaking module 26 plates (2496 wells) 52 plates (up to 13 reagent cassettes) 12 plates 3 x 6 plates 11 plates before user intervention required No restrictions
Manual water filling or waste emptying Automatic water filling and waste emptying
Low volume reagent pipette High volume reagent pipette Measurement unit Humidity Temperature Analytes
Volume range Precision (CV%) Accuracy Volume range Precision (CV%) Accuracy Measurement modes Signal with 1 nM Eu Plate storage area
5 - 50 l < 1.5 % within 8 % 25 - 200 l < 1.5 % within 4 % TRF, Fl, Abs 1 000 000 cps 17% < 60 % 25 2 C 10 2 C 37 1 C 13 simultaneously
Temperature inside the instrument Reagent storage temperature Heated incubator module Maximum number of different analytes
Reagent storage
Temperature controlled storage for assay reagents Room for 13 reagent cassettes (up to 52 plates) Enables continuous loading
Plate Manipulator
Elevator for transferring plates between different modules of the GSP Forms the body of the instrument
Liquid module
Prepares Wash solution from the Wash concentrate Handles the waste solutions
Measurement module
xy-conveyer to move the plate under the measurement head Enables time resolved fluorescence, prompt fluorescence and absorbance measurements
www.perkinelmer.com
Ordering information
2021-0010 Genetic Screening Processor
The Genetic Screening Processor includes GSP instrument and GSP Workstation.
PerkinElmer, Inc. 940 Winter Street Waltham, MA 02451 USA Phone: (800) 762-4000 or (+1) 203-925-4602 www.perkinelmer.com
PerkinElmer, Inc. Wallac Oy PO Box 10 20101 Turku, Finland Phone: + 358 2 2678 111 Fax: + 358 2 2678 357
For a complete listing of our global offices, visit www.perkinelmer.com/ContactUs 2009-2011 PerkinElmer, Inc. All rights reserved. Wallac is a trademark and GSP and PerkinElmer are registered trademarks of PerkinElmer, Inc. All other trademarks depicted are the property of their respective holders or owners. PerkinElmer reserves the right to change this document at any time and disclaims liability for editorial, pictorial or typographical errors.
Printed in Finland