DENTAL UNIT

OPERATING INSTRUCTIONS
IMPORTANT This manual provides operating instructions for BELMONT PROGIA. The instructions contained in this booklet should be thoroughly read and understood before operating the chair and unit. File this manual and refer back to it for future maintenance.

TABLE OF CONTENTS

Page 1. OVERALL VIEW AND MAJOR COMPONENTS -------- 1 2. DIMENSIONS AND SPECIFICATIONS 2-1. DIMENSIONS ---------------------------------------------- 2 2-2. SPECIFICATIONS ----------------------------------------- 2 3. OPERATING INSTRUCTIONS 3-1. DOCTOR TABLE SECTION ----------------------------- 3 3-2. CUSPIDOR UNIT SECTION ----------------------------- 6 3-3. FOOT CONTROL SECTION ----------------------------- 8 4. CARE AND MAINTENANCE -------------------------------- 8

1. OVERALL VIEW AND MAJOR COMPONENTS

12 14 2 3 4 13 9 8 7 23 10 6 17 18 21 19 11 20 22 24 1 5 16 15

Fig.1-1 Overall View and Major Components MAJOR COMPONENTS q Cuspidor Unit w Cupfiller Nozzle e Bowl Flush Nozzle r Cuspidor Bowl t Solid Collector y Assistant Arm u Saliva Ejector Handpiece i Assistants Syringe o Vacuum Handpiece !0 Assistant Instrument Holder !1 Foot Controller !2 Light Pole !3 Main Post !4 Dental Light (IO 5000) !5 Horizontal Arm !6 Balance Arm !7 Table Tray !8 Main Control Panel !9 Handle @0 Handpiece Holder @1 Table Housing @2 Waste Receptacle @3 Assistant Side Control Panel @4 Dental Size Film Viewer

-1-

2. DIMENSIONS AND SPECIFICATIONS


2-1. DIMENSIONS -mm650 330 330

67

826

24

67

240
350

90

690

600
730

42.7
89

312

50

650

169

89

90
642 528

226

217

453

CHAIR MOUNT TYPE


493

690

554

880 ~ 1180

371

76
565

140

Net Weight : 26 kg
530 400

CART TYPE Fig.2-1 Dimensions 2-2. SPECIFICATIONS Doctor Side Instrument ---------------- 2 x High Speed Turbine Tubings 1 x Low Speed Air Motor Tubing 1 x 3 Way Syringe Assistant Side Instrument -------------- Saliva Ejector (Central Vacuum Type) Vacuum (Central Vacuum Type) 1 x 3 Way Syringe Bowl Flush ------------------------------- Timer Mode and Continuous Mode Cupfiller ---------------------------------- Timer Net Weight ------------------------------- 90.8 kg (Without Dental Light) -2-

543

90

10 37

65

3. OPERATING INSTRUCTIONS
3-1. DOCTOR TABLE SECTION
Master Switch

CAUTION
Turn off the master switch after daily operation or in long term interval.
ON
I MA I N O

(1) Master Switch (Fig.3-1 & Fig.3-2) Turn on the master switch located on the facing right side under table, the power indicator on the main control panel illuminates in green.
1 10 9 8 7

OFF

Fig.3-1 Master Switch

0 1

LP 2

11 6 5

q w e r t y

Power Indicator Cupfiller Switch Bowl Flush Switch Flush Out Switch Extra Switch Light Pack Switch

u i o !0 !1

Electric Motor Direction Control Switch Electric Motor Maximum Speed Setting Switch Chair Auto Mode Switches Chair Manual Control Switches Safety Lock Indicator

Fig.3-2 Main Control Panel (2) Main Control Panel A. Cupfiller Switch (Fig.3-2) Momentarily press the cupfiller switch ( ), water comes out the cupfiller nozzle for about 2.5 seconds and stops automatically. During cupfiller is working, the indicator illuminates in green. During cupfiller is working, momentarily pressing the cupfiller switch ( ) cancels cupfilling. Note : The cupfiller water volume can be adjusted by the cupfiller flow control knob located in the cuspidor body. Refer to 3-2.(3). B. Bowl Flush Switch (Fig.3-2) Two type of bowl flush modes are available. a. Timer Mode Momentarily press the bowl flush switch ( ), water flushes for about 5 seconds and stops automatically. During flushing water, the indicator illuminates in green. b. Continuous Mode Keep pressing the bowl flush switch( ) for about 3 seconds, the indicator starts blinking and water flushes continuously until the bowl flush switch ( ) is pressed again. Note : During bowl flushing, momentarily pressing the bowl flush switch ( ) cancels bowl flush in any mode. The bowl flush water flow volume can be adjusted by the bowl flush flow control knob located in the cuspidor body. Refer to 3-2.(3) -3-

C. Flush Out Switch (Fig.3-2) Pick up the handpieces from the holder and set them in bowl. Momentarily press the flush out switch ( ), the indicator illuminates in green and water comes out from the handpieces to clean up the handpiece water lines. After flushing about 40 seconds, water stops automatically and the indicator turns out. Note : During flush out is working, momentarily pressing the flush out switch ( flush out immediately. D. Extra Switch (Fig.3-2) The extra switch ( ) can be used for the dental light ON/OFF switch. During the extra switch is ON, the indicator illuminates in green. E. Light Pack Switch (Optional)(Fig.3-2) Pick up the fiber optic handpiece from the holder, momentarily press the light pack switch ( the indicator illuminates in green and the fiber optic power turns on until the light pack switch is pressed again. ) cancels

),

F. Electric Motor Rotation Direction Control Switch (Optional)(Fig.3-2) During picking up the electric motor from the holder, the electric motor rotation direction ) and the rotation direction is indicated can be changed by momentarily pressing this switch ( in the indicator. Indicator in green : Forward Rotation Indicator in amber : Reverse Rotation G. Electric Motor Maximum Speed Setting Switch (Optional)(Fig.3-2) During picking up the electric motor from the holder, the maximum rotation speed of the electric motor can be set with this switch ( ). The maximum speed can be select 3 steps (High, Medium, Low) and it is indicated in the indicator. High (100% rpm. for electric motor) Medium (50% rpm. for electric motor) Low (25% rpm. for electric motor) * Illuminating L.E.D H. Chair Manual Control Switches (Fig.3-2) a. Seat Lifting -------------------------------- Keep pressing ( ) switch until the seat is lifted up to the desired position. b. Seat Lowering ----------------------------- Keep pressing ( ) switch until the seat is lowered to the desired position. c. Backrest Reclining ------------------------ Keep pressing ( ) switch until the backrest is reclined to the desired position. d. Backrest Raising -------------------------- Keep pressing ( ) switch until the backrest is raised up to the desired position. I. Chair Auto Mode Switches (Fig.3-2) a. Preset Operation Momentarily press the preset-1 switch ( 1 ), the chair moves to the preset 1 position and stops automatically. Preset 2 position operated by the preset switch ( 2 ). Note : Preset position adjustment refers to Progia chair Manual.

-4-

b. Auto Return Operation Momentarily press the auto return switch ( 0 ), the chair returns to the initial position (the seat is the lowest and the backrest is the upright position) and stops automatically. c. Last Position Memory Operation Momentarily press the last position memory switch ( LP ) at the treatment position, the backrest raises up to the mouth rinsing position (upright position) and stops automatically. Momentarily press the last position memory switch ( LP ) again, the backrest returns to the previous treatment position and stops automatically. d. Emergency Stop (Safety Stop) During automatic procedure (preset, auto return and last position memory), momentarily pressing any chair control switch will cancel the automatic movement immediately. f. Safety Lock Indicator During the handpiece is running, the safety lock indicator illuminates in red and all chair control switches can not operate.
E

Handpiece Spray Water Control Knobs

MAIN O I

(3) Doctor Table Section Control Knobs A. Handpiece Spray Water Control Knobs (Fig.3-3) The handpiece spray water control knobs are located on the facing right side under the doctor table. Each handpiece spray water control knob is lined up from the facing left side HP1,HP2,HP3,... The handpiece spray water can be controlled independently. B. Doctor's Syringe Flow Control Knobs (Fig.3-3) Doctor's syringe flow control knobs are located on the facing right side under the doctor table. The flow control knobs adjust doctor's syringe air and water flow volume. The yellow capped knob is the air flow control knob, the blue capped knob is the water flow control knob.

Electric Motor Speed Control Knob

Doctor's Syringe Flow Control Knobs Electric Scaler Power Control Knob

Fig.3-3 Doctor Table Section Control Knobs

Increase

Decrease (Stop)

Fig.3-4 Control Knob

Note : Turning a control knob counterclockwise increases flow volume and turning clockwise decreases. (Fig.3-4) C. Electric Scaler Power Control Knob (Optional)(Fig.3-3) Electric scaler power control knob is located on the facing right side under the doctor table. Turning the knob clockwise increases the electric scaler power, turning counterclockwise decreases. D. Electric Motor Speed Control Knob (Optional, Bien Air PLPE843)(Fig.3-3) Electric motor speed control knob is located on the facing right side under the doctor table. Turning the knob clockwise increases the electric scaler power, turning counterclockwise decreases. Electric motor maximum speed setting switch can be set the electric motor speed controlled by this knob. (Refer to 3-1.(2) G.) -5-

(4) Handpiece Pressure Gauge (Fig.3-5) Handpiece drive air pressure gauge is located on the rear side of the doctor table. During a handpiece is working, the handpiece drive air pressure is indicated in the handpiece pressure gauge.

2 1

3 4 5

Handpiece Drive Air Pressure Gauge

Fig.3-5 Handpiece Drive Air Pressure Gauge (5) Balance Arm Air Brake (Fig.3-6) Balance arm air brake is located on the handle. When the master switch is ON, the balance arm is locked. Grasp the handle and press the air brake button to adjust the table height. Release the air brake button at the desired table position, the balance arm is locked. Note : When the master switch is off. The arm lock is released. Do not load over 1.5 kg on the table.

Balance Arm

Air Brake Button

Fig.3-6 Air Brake Button


Bowl Flush Switch Cupfiller Switch

3-2. CUSPIDOR UNIT SECTION (1) Assistant Side Control Panel (Fig.3-7) Cupfiller switch, bowl flush switch and chair auto mode switches are located on the assistant side control panel. Refer to 3-1.(2) A,B,I.

0 1

LP 2

Chair Auto Mode Switches

Fig.3-7 Assistant Sise Control Panel

Dental Light Mode Selection Switch Water Heater Water Heater Switch Indicator
O
MANUAL SENSOR

Service Water Control Knob Service Water Outlet Service Air Outlet

Fig.3-8 Cuspidor Unit Control Panel (2) Cuspidor Unit Control Panel A. Dental Light (IO 5000) Mode Selection Switch (Fig.3-8) Dental light (IO 5000) can be operated (ON/OFF) either by the non-contact switch located on the light head, or by the manual switch. q To operate by the non-contact switch ---- Set the switch lever to Right w To operate by manually -------------------- Set the switch lever Left -ON. Set the switch lever to Centre-OFF. -6-

B. Water Heater Switch (Optional)(Fig.3-8) Water heater switch is located on the cuspidor unit control panel. Turn on the water heater switch, the water heater indicator illuminates in green and cupfiller water will be warmed up. C. Service Water Outlet (Fig.3-8) The service water outlet provides quick-connect water. D. Service Water Outlet Control Knob (Fig.3-8) The water volume of service water outlet can be adjusted by the service water outlet control knob. Turning the knob counterclockwise increases flow volume and turning clockwise decreases. E. Service Air Outlet (Optional)(Fig,3-8) Service air outlet provides quick-connect air.

Lock (3) Cuspidor Unit Body Maintenance A. How to Remove Maintenance Panel (Fig.3-9) panel Assistant's syringe flow control knobs, cupfiller Release flow control knob and bowl flush control knob are located inside the cuspidor body. Remove the maintenance panel for adjustment. Cover Fixing Push the cover fixing buttons (Buttons come out) Button to release and remove the maintenance panel. After adjustment, reset the cuspidor cover and Fig.3-9 Removing Cuspidor Cover push into the fixing buttons to lock the maintenance panel in place.

B. Assistant's Syringe Flow Control Knobs (Fig.3-10) The yellow capped knob is to adjust the assistant's syringe air flow volume, and the blue capped knob is to adjust water flow volume. C. Cupfiller Flow Control Knob (Fig.3-10) Cupfiller flow volume can be controlled by the cupfiller flow control knob. Loosen the lock nut and adjust cupfiller water flow volume by turning the knob. Tighten the lock nut after adjustment. D. Bowl Flush Flow Control Knob (Fig.3-10) Bowl flush flow volume can be controlled by the bowl flush flow control knob. Loosen the lock nut and adjust bowl flush water flow volume by turning the knob. Tighten the lock nut after adjustment. Note : Turning a knob counterclockwise increases flow volume and turning clockwise decreases. -7-

Bowl Flush Flow Control Knob

Cupfiller Flow Control Knob

Water

Air

Lock Nut Assistant's Syringe Flow Control Knobs

Fig.3-10 Flow Control Knobs in Cuspidor Body

(4) Assistant Instrument Holder Picking up an instrument (Saliva ejector or Vacuum handpiece) from the assistant holder starts the instrument working automatically. Returning the instrument to the holder stops automatically.

3-3. FOOT CONTROL SECTION (1) Standard Type Foot Control (Fig.3-11) A. Drive Air Pedal Pick up a handpiece from the handpiece holder and depress the drive air pedal, the handpiece starts running. Depressing degree controls the handpiece rotation speed. B. Coolant Water Switch Coolant water switch allows handpiece coolant water to be turned on or off. C. Chip Blower Pedal By depressing the chip blower pedal, chip blower comes out from handpiece without bur turning. (2) Safety Lock Device (Fig.3-2) While the handpiece is running, all chair electrical movements are locked inactive for safety. It is indicated on the safety lock indicator on the main control panel.

Chip Blow Pedal

OFF ON Coolant Water Switch Drive Air Pedal

Fig.3-11 Standard Type Foot Control

4. CARE AND MAINTENANCE


CAUTION
Turn OFF the master switch after daily operation or in long term interval. Keep the main water valve OFF after daily operation or in long term interval.

Cleaning Unit

CAUTION

Do not drench the unit for cleaning. Do not use polishing powder, solvents, strong disinfectant or hot water for cleaning. After cleaning, wipe with a dry soft cloth and keep the unit dry. Painted, metal and plastic surfaces can be cleaned with weak ethanol. Membrane switches and all tubings can also be cleaned with weak ethanol.

-8-

(1) Cuspidor Bowl (Fig.4-1) Take out the drain cap and the basket strainer in the centre of cuspidor bowl and clean them.
Solid Collector

Drain Cap Basket Strainer

(2) Solid Collector (Fig.4-1) Pull out the solid collector with filter and clean it. After cleaning, set the solid collector firmly.

Cuspidor Bowl

(3) Vacuum Handpiece and Saliva Ejector (Fig.4-2) A. Unscrew and remove the top parts of each handpiece and clean each strainer. B. After daily operation, run two cups of clean water through the handpieces to clean inside.
Close

Fig.4-1 Basket Strainer and Solid Collector

Vacuum Tip Open Close

Saliva Ejector Tip Open

Note : After cleaning the handpieces, apply a white vaseline lightly on the rubber parts (O-Ring) and screw parts for long life.
Vacuum Handpiece

Strainer Saliva Ejector Handpiece

(4) Disinfection A. Vacuum tip, saliva ejector tip and syringe tip are autoclavable. B. All tubings and hoses can be cleaned with weak ethanol.

Fig.4-2 Vacuum Handpiece and Saliva Ejector

Main Water Valve

Air Filter Drain Valve Close

(5) Air Filter Drain Valve (Fig.4-3) Air filter drain valve is located on the left side of utility section. Once a week open the drain valve and drain off water from the air line.

Open

Close

Open

(6) Main Water Valve (Fig.4-3) Main water valve is located on the left side of utility section. Turn off the main water valve after daily operation or in long term interval.

Fig.4-3 Air Filter Drain Valve and Main Water Valve

-9-

NOTE

TAKARA BELMONT CORPORATION


1-1, 2-Chome, Higashi-shinsaibashi,Chuo-ku,Osaka,Japan TEL. : (06) 213-5945 TELEFAX : (06) 212-3680
Printed in Japan 9808 MA

3-3. FOOT CONTROL SECTION (2) Progia Type Foot Control (Fig.3-12) A. Drive Air Pedal Pick up a handpiece from the handpiece holder and depress the drive air pedal downward, the handpiece starts running. Depressing degree controls the handpiece rotation speed. B. Coolant Water Switch Coolant water switch allows handpiece coolant water to be turned on or off. C. Chip Blower Button By depressing the chip blower button, chip blower comes out from handpiece without bur running. D. Electric Motor Rotation Control (Optional) Pick up the electric motor from the handpiece holder and slide the drive air pedal horizontally to right, the electric motor starts running. The rotation speed increases by sliding the drive air pedal to further right. The speed control by the foot control is within the limits of the electric motor speed setting. Refer to 3-2. (2) G.

Chair Manual Control Stick Switch Electric Motor Speed Control Drive Air Control OFF

Drive Air Pedal Chip Blow Button

ON Coolant Water Switch

Fig.3-12 Progia Type Foot Control


Seat Lifting

Backrest Reclining

Backrest Raising

Stick Switch Seat Lowering

Fig.3-13 Chair Manual Control Stick Switch E. Chair Manual Control Stick Switch (Fig.3-12 & Fig.3-13) a. Seat Lifting -------------------- Keep inclining the stick switch to ( ) side until the seat is lifted up to the desired position. b. Seat Lowering ----------------- Keep inclining the stick switch to ( ) side until the seat is lowered to the desired position. c. Backrest Reclining ------------ Keep inclining the stick switch to ( ) side until the backrest is reclined to the desired position. d. Backrest Raising -------------- Keep inclining the stick switch to ( ) side until the backrest is raised to the desired position.