Vous êtes sur la page 1sur 56

BC-2800 HematoIogy AnaIyzer

7.5' color LCD19 parameters and 3 histograms


3-part diIIerential oI WBC
automatically cleaning sample probe
Automatic diluting, lyzing, mixing, rinsing and clog-clearing
Count speed: 30tests/hr
3 counting modeall parametersWBC/HGBRBC/PLT
Automatic Ilushing, zapping
remain reagents alarm
expired date alarm
build-in recorder
bar code scanner and keyboard option
Up to 10000 sample results (including histograms) can be stored
BC-2800 Specification
BC-2800 InstaIIation---Environment
3;7432039
The environment should be as Iree as possible Irom dust, mechanical vibrations,
loud
noises and electrical interIerence.
Avoid proximity to brush-type motors, Ilickering Iluorescent lights and electrical
contacts that regularly turn on and oII.
Avoid placing the instrument in direct sun or in Iront oI a source oI heat.
BC-2800 InstaIIation---Space
$5,.0
ComIortable working height
At least 40cm on each side is the preIerred access to perIorm service procedures.
At least 10cm behind Ior cabling and ventilation
There is enough space Ior reagents.
BC-2800 InstaIIation---Power
!407
Aproperly grounded socket supplying 220 or 110 volts
The Iluctuation should be in the range oI 10.
The power Irequency should be 50/60Hz.
Power consumption is less than 180VA.
$elect some samples in the sample tabular review menu, and then
press |F5| to access 'extension Iunction menu.
|PRINT| print the current displayed content
|HELP| help inIormation
|MENU| exit menu
Extension Iunction: 'repetition and 'trend graph
'trend graph Iunction only open to administrator
Extension function in tabuIar review menu
Password: 2826
Procedure: press F5 then press F1
Trend graph analysis mean: it is to monitor one patient`s illness trend.
Trend graph
Value oI lyse: 0.5mL(whole blood) 0.36mL(pre-diluted)
Blood needed:
whole blood: 13L
pre-diluted: 20L
Count time less than 2 min / test
Aperture size 80mdiameter70mmdepth
Ratio oI whole blood Capillary
WBC/HGB 1:308 1:400
RBC/PLT 1:44862 1:44402
Volume oI volumetric testing: 300L `2times
BC-2800 sampIe and dispense
Range of Iinear and dispIay
=r=mfr l+u=r l+>,l=,
WBC 0 - 99.9 `10
9
/L 0.0-499.9 `10
9
/L
RBC 0 -9.99`10
12
/L 0.0-9.99 `10
12
/L
HGB 0 - 300g/L 0- 300g/L
MCV ------ 0-250 fL
PLT 0 - 999`10
9
/L 0-2999 `10
9
/L
100 L
Mean
Diff
CJ
=r=mfr linear
4,f+f+u\)
WBC
4.0-7.5`10
9
/L 5 3.0
7.6-15.0`10
9
/L 5 2.0
RBC
3.0-4.5`10
12
/L 5 2.0
4.5-6.5`10
12
/ L 5 1.5
HGB
110 180g/L 5 1.5
MCV
70.0 100.0fL 5 1.0
PLT
100-200 `10
9
/L 5 5.0
201 - 500`10
9
/L 5 4.0
Repetition
AC Input
AC 220V30V 50/60Hz1Hz
Consumption
150 VA
Fuse
250 V 2A
Power SuppIy
Without any operation in 10 minutes, the system will enter the
saver program.
LCD display turns oII
Power indicator is still light.
Any key exists the saver program.
Screen Saver
Power indicator
$tart key
7.5 color LCD
Keypad
Record
$ample probe
Front PaneI
Power switch and Iuse
Equipotential terminal
Diluent coonector
Rinse connector
Lyse connector
Waste connector
Rear paneI
Keyboard
Power or Iloppy
Transition
Printer or upgrade interIace
$canner interIace
Left side
DispIay feature
Title Count Mode System
Status
System time
Window
Error Messages Reagent Status
MENU On-line Help
1~99testsremain reagents and remain waste container
are enough Ior 99tests
20 testany one oI the remain reagents or remain
waste container is not enough Ior 1 test. The warning message
will be show in error area.
Actual testsshow the actual tests number
Remain reagents
P$/2!'1!/
P$/2!'!!!'|Ctrl A|
Keypad
Menu tree
Count
Revi ew
Cal i br at i on Manual
Aut o
Qual i t y cont r ol
Passwor d
Sel f - t est
Log
St at us
Ser vi ce Mai nt enance
Conf i g
Pr i nt
Shut down
Cont r ol s
X- B anal ysi s
Set t i ngs
Set up
Aut o - f r esh bl ood
DiIuent Rinse Lyse
E-Z
cIeanser
NormaI start-up
46.5mI
cIean & once background test)
16.5mI 0.5mI ---
75mI
cIean & twice background test
26mI 1mI ---
AII parameters
(whoIe bIood)
28.5mI 9.5mI 0.5mI ---
AII parameters
(pre-diIuted)
30.5mI
9.5mI 0.36mI ---
NormaI shut down 13.5mI 5.5 mI --- 1mI
Reagents consumption
6D l=>f >DKf du >f=fK>
rdKr
\rm=l >DKfdu l=u fK>+uz =ud >=lzrKud f>f+uz
^>urm=l >DKfdu
l+rfl, >DKfdu
4>f mfr>. Dl >f=r. l=u
fK>+uz =ud >=lzrKud f>f
=l=z r m,f, fK>+uz
r+m fK>+uz. l=u fK>+uz >vr=l
f+m>. >=lzrKud f>f.
1. $witch on BC-2800and the power indicator will be light.
2. The Initializing procedure will continue about 80s.
3. $elI test procedure.
4 The system will enter 'count menu. II there are errors during startup, the
error message will be displayed on leIt bottom oI screen.
Power on and shut down
Background test
=r=mfr ^,f=>l l+m+f
WBC
A 0.3 L 10
9
/ L
RBC
A 0.03 L 10
12
/ L
HGB
A 1 g / L
HCT
A 0.5 %
PLT
A 10 L 10
9
/ L
1. Only ID mode
2. All inIormation
3. Batch inIormation mode
SampIe information input
1. Press [MODE] to select the prediluted mode (any mode preceded by PB).
2. Press [DLUENT] to enter the Add Diluent screen
3. Prepare a clean sample cup under the sample probe and press [START]
to dispense 1.6ml diluent to the cup.
4. Take out the sample cup after the sample
probe rose up and press [ENTER] to exit the
Add Diluent screen.
5. Collect 20L capillary blood specimen and
immediately add it to the sample cup and
shake to mix them.
Preparation of pre-diIuted bIood
|F1| is to select the setting items.
[SETUP]-[Settings]
Reagent
VaIid voIume
Diluent 0-30.0L
Rinse 0-30.0L
Lyse 0-999mL
Waste 0-99.9L
[SETUP]-[Settings]--Reagents Remain
[SETUP]-[Settings]-Print & Communication
[SETUP]-[Settings]-Print & Communication
Printing
Device
Format1 Format2 Format3 Format4
Recorder Parameters +
histograms
Parameters Parameters+
histograms
Parameters
Printer Whole page with
histograms
Whole page
without
histograms
Whole page with
histograms
Whole page
without
histograms
[SETUP]-[Settings]-Date & time
[Setup]-[Settings]---[Gain]
[Setup]-[Settings]---[Gain]--WBC
W too small WBC gain that should be increased
W too large WBC gain that should be decreased.
[Setup]-[Settings]---[Gain]--RBC
W When the difference between the actual MCV analysis result and the
expected result exceeds 6%, need to change the RBC gain.
W For exampIe, assuming the expected MCV result is 90.0fL, while the
actual analysis result is 82.0fL, the gain factor as below:
098 . 1
0 . 82
0 . 90

.9:alMCJ
J Expe.9edMC
[Setup]-[Settings]---[Gain]--HGB
W adjust the HGB gain to change the HGB blank voltage, which
usually should be set between 4.3 ~ 4.5V.
[Setup]-[Settings]--count
[Setup]-[Settings]--Reference Range
=f+uf rK, x ^z
ur=l
\f >,++d \f >,++d
\f >,++d 1` ,=r>
H=l =dKlf> H=u 1` ,=r>
lm=l =dKlf> "m=u 1` ,=r>
d+=fr+
,=f+uf>
\f >,++d - 1 mufD =ud <- 1` ,=r>
\u=f> \f >,++d <1 mufD
[Setup]-[Settings]--other
[Service]-[Maintenance]--Prime
[Service]-[Maintenance]-Zap & FIush
[Service]-[Maintenance]-Probe cIeanser cIean
W f your analyzer is to run 24 hours a day, you should perform this
procedure every 3 days.
W f you follow the shutdown procedure to turn off your analyzer
everyday, you should perform this procedure every week.
W To make sure this analyzer functions normally, every time the
accumulated analyzed samples reach 200, a dialog box will pop
up to remind the user to perform the Probe cleanser cleaning
procedure.
'Lyse test is to check the injected volume Ior lyse is correct or not.
When shall we use this Iunction?
When the WBC result or its histogram is abnormal, we should run this
Iunction.
$crew
oII
these
two
screws
[Service]-[Maintenance]-Lyse test
Remove the
Count bath
shield cover
The
injected
lyse must
over the
second
mark
2ml lyse will be dispensed to the count bath.
[Service]-[Maintenance]-Lyse test
[Service]-[Maintenance]-Drain
W Drain bath
Empty the count bath, and prime it again
W Drain Tubing
Empty all tubing, then turn off the
equipment
[Service]-[Maintenance]-CIean
W Clean bath
Use diluent to clean count bath
W Clean wipe block
After being used for a long time, the bottom of the probe wipe
block may be contaminated by blood.
When the accumulated analyzed samples reach 4,000, a dialog box
will
pop up to remind the user to clean the wipe block
[Service]-[Maintenance]-Prepare to ship
W Empty the tubing
W Wash the tubing
W Empty the tubing
W Turn off the equipment
[Service]-[System test]
The available selI-test items are divided into Iour groups, the
1ubing, Machine, Jalve and Circuit.
[Service]-[seIf-test]
[Service]-[seIf-test]-[Test tubing]
Count Time WBCS& Count Time RBCS
t measures the duration of a WBC or RBC count
Aperture(v)
t measures the voltage (v) over the aperture.
Vacuum
t checks whether the vacuum system functions normally.
Pressure
t checks whether the system flushes the aperture at a normal
pressure.
FiIter
t checks whether the filter functions normally.
[Service]-[seIf-test]-[Machine]
[Service]-[seIf-test]-[VaIve]
It is to monitor that WBC, RBC and PLT signals can be properly converted
into digital signals or not.
[Service]-[seIf-test]-[Circuit]
The recorded events are divided into Iour groups, ll, Para set, Error and Other
(including setting discriminators, system selI-test and updating system soItware)
This analyzer can
save maximum 100
log Iiles and once
the maximum
number has been
reached, the newest
log will
automatically cover
the oldest one.
[Service]-[Log]
Every screen displays 13 items
It only shows BC-2800 current conIiguration, no change.
[Service]-[Configuration]
The printing tasks are queued in this screen
[Service]-[Print management]
[CaIibration]---[ManuaI]
Password: 2826
Valid range:
75---125
[CaIibration]---[Auto]
Password: 2826
Valid range:
75---125
[CaIibration]---[Fresh bIood]
Password: 2826
Valid range:
75---125

Vous aimerez peut-être aussi