Verification
How to verify that the accuracy is within 5V?
STCP-195 (Rev 0)
BBC-665 (Rev 0)
Copy right @ 2010 All rights reserved National Metrology Centre CDC-987 (Rev 0)
STCP-195 (Rev 0)
Method
Copy right @ 2010 All rights reserved National Metrology Centre CDC-987 (Rev 0)
Calibration Result
Error at 1.0000mV is -2.0V having expanded uncertainty of 1.5V with confidence level of approximately 95% at k=2
Verification
Specification of voltmeter=5V Absolute value of error + expanded uncertainty=|2.0V|+1.5V=3.5V It is smaller than the specification and therefore, the meter conforms to specification
12 Copy right @ 2010 All rights reserved National Metrology Centre
Verification
Error (V) Upper Limit: 5V
0 -2 Uncertainty bar
Conclusion
Identify test specification Select measuring instrument with accuracy 10 times or at least 4 times better than the test specification Select calibration laboratory with CMC 10 times or at least 4 times smaller than the specification Expanded uncertainty shall be provided with the calibration result Absolute value of error + expanded uncertainty should be < specification to verify conformity The same method applies for the conformity assessment of product. Uncertainty must be known too!
14 Copy right @ 2010 All rights reserved National Metrology Centre