ENTER PATIENTS DETAILS/INFOS: 1. CONTROL NUMBER (Refer to the last patients record) 2. FIRST NAME (Forename) 3. LAST NAME (Surname) 4. DATE OF BIRTH
o When done press ( ) SIGN.
DOUBLE CHECK THE DETAILS:
When done press ( ) SIGN (Enter).
TESTS SCREEN:
Press the ( PTA ) Pure Tone Audiometry. Instruct the patient: 1. HEADPHONES (Air Conduction) Red Right Blue - Left 2. PATIENTS RESPONSE ( Joystick) 3. Bone Conduction (if NECESSARY)
PURE TONE AUDIOMETRY (PTA): Start with RIGHT Ear
Response Icon
UP
DOWN
LEFT RIGHT Speaker Icon
1. Start with 0.25 kHz & 40 dB. Using the arrows LEFT and Right then UP and DOWN, You can adjust the level of the Kilohertz (kHz) and Decibels (dB). 2. Once ready, advise the patient that you will now start the test. 3. Press the SPEAKER ICON (just below the DOWN arrow), just enough for the patient to hear the beep/tone. 4. Focus on the RESPONSE ICON (found above the UP arrow), if it turns GREEN (the patient responded to the tone), move one level by pressing the UP ARROW until the patient couldnt hear a tone or no response to tone, then PLOT on the level with no response. The move to the next level with the RIGHT ARROW and move downwards using the DOWN ARROW to 30dB - 40dB level. Then repeat the whole process until you reach the level 8k. 5. To save the results/measurement, press the HOME Button, a dialog box will appear, then press YES. 6. You will go back to the TESTS SCREEN.
To test the LEFT Ear 1. Press the ( PTA ) Pure Tone Audiometry. 2. Change the setting by pressing the EAR Icon, to change the color of the icon from RED to BLUE
3. Start with 0.25 kHz & 40 dB. Using the arrows LEFT and Right then UP and DOWN. 4. Once ready, press the SPEAKER ICON (just below the DOWN arrow), just enough for the patient to hear the beep/tone. 5. Focus on the RESPONSE ICON (found above the UP arrow), if it turns GREEN (the patient responded to the tone), move one level by pressing the UP ARROW until the patient couldnt hear a tone or no response to tone, then PLOT on the level with no response. The move to the next level with the RIGHT ARROW and move downwards using the DOWN ARROW to 30dB - 40dB level. Then repeat the whole process until you reach the level 8k. 6. To save the results/measurements, press the HOME Button, a dialog box will appear, then press YES.
TO REVIEW THE PATIENTS RESULTS/MEASUREMENTS (For Plotting)
1. On the TESTS SCREEN, press the BOOK ICON, found at the lower part of the screen, second icon.
2. Youll be able to see the recorded test results/measurements of patients audiology report.
3. Press on the options to preview the results/measurements.
HOW TO PLOT THE RESULTS/MEASUREMENT Plot the results/measurements based on the AUDIOLOGY REPORT FORM. Just simply copy the results/measurements on the AUDIOLOGY FORM, use: o RED Pen for RIGHT EAR Result o BLUE Pen for LEFT EAR result Use : symbol for the result of RIGHT EAR symbol for the result of LEFT EAR EXAMPLE:
RIGHT EAR
LEFT EAR
** Please turn to the next page for the sample of the AUDIOLOGY REPORT FORM.
AUDIOGRAM ANALYSIS o 20 - NORMAL LEVEL o 21 29 may be considered as Normal but it depends on the average results of 500k, 1k & 2k. o To compute the average result of the audiogram: o ADD the results of: 0.5 kHz (500 kHz) 1k 2k o Then DIVIDE the total SUM with three (3). EXAMPLE: Right Ear: 15 +15 + 10 = 40 / 3 = 13.3 Average result: 13 Left Ear: 20 + 20 + 20 = 60 / 3 = 20 Average result: 20
If the result is more than 30 (< 30) suggest for referral to an ENT.
In case there is a SUDDEN DROP at the level of 4k & 8k, But with Normal Average Result, if you were been asked by the physician about it, it is better to advise the physician to refer the client for an ENT Consultation, for further evaluation. Sudden drop at 4K and 8K means the client has been exposed to loud noise or sound. Further ENT Evaluation is recommended. For additional INFOS & CLARIFICATIONS you may call: DR. PIOs Clinic 472-4199 (St. Louise 10am - 12pm) and 227- 2143 (BCDC 3pm 5pm, Mon Fri). By Saturday 10 am 12pm only. Ms. Aiza Marie, Agent Audiometer - +639272652770
IN CONDUCTING BONE CONDUCTION TEST (BC) On the TESTS SCREEN, press PTA.
Press the Third Icon (Middle Icon) and change the icon into a look like LOCK ICON.
Then press on the SETTINGS ICON, next to the LOCK ICON. Change the settings of the TRANSDUCER, by choosing the second icon, the smiley face with an arrow down.
o When done press ( ) SIGN.
Example: RIGHT EAR
LEFT EAR
The process of the Bone Conduction (BC) Test is just the same as the Air Conduction (AC) Test.
PLOTTING THE BONE CONDUCTION RESULTS Plot the results/measurements based on the AUDIOLOGY REPORT FORM. Just simply copy the results/measurements on the AUDIOLOGY FORM, use: o RED Pen for RIGHT EAR Result o BLUE Pen for LEFT EAR result Use : < symbol for the result of RIGHT EAR > symbol for the result of LEFT EAR EXAMPLE: Right Ear
Left Ear
** ALL RESULTS ARE RECORDED ON THE BOOK ICON, THAT CAN BE FOUND AT THE TESTS SCREEN.