PickPen®

GENERAL DESCRIPTION PickPen, 1 magnet PickPen, 8 magnet


®
The PickPen is designed to transfer magnetic particles when
working with BioControl’s Assurance GDS™ pathogen Eject button
detection kits.
Operating button /
Part No’s: 73038 (8 magnet), 73028 (1 magnet) lever

PROCEDURE
A. Loading Disposable Tips
Carrier sleeves
With magnets withdrawn into the carrier sleeves, position
the PickPen over PickPen tips and push down firmly to seat Magnets
the tips over the carrier sleeves.
B. Collection of Magnetic Particles
(a) Check that the tips are firmly in place by depressing the CLEANING
operating button (PickPen, 8 magnet) or moving the The PickPen should be cleaned and disinfected with an anti-
operating lever down (PickPen, 1 magnet) to extend the DNA agent followed by a 70% alcohol solution after each use.
magnets and stretch the tips. Depressing the operating
button a second time or pushing the operating lever up will To clean and disinfect the device, moisten a paper towel with
retract the magnets. the appropriate solution. Gently wipe device with moistened
(b) With the magnets locked in the extended position, dip the towel. Avoid a introducing any liquid into the openings at the
tips into the suspension containing the magnetic particles. end of the carrier sleeves.
Collect particles by stirring gently for the specified amount Exposure of the magnets or the inside of the carrier sleeves to
of time while continually moving up and down from the moisture may result in corrosion, preventing the device from
surface to the bottom of the well. operating properly. Do not spray any solutions directly on the
C. Transfer of Magnetic Particles PickPen. Do not autoclave.
(a) Once the particles are collected, withdraw the tips from Recommended anti-DNA agents:
the solution. Tap the PickPen tips against inside of sample
wells to remove and residual liquid. The particles will (a) 10-20 % Bleach (0.6 – 1.2% Sodium Hypochlorite) solution
remain attached to the ends of the tips due to the magnetic (b) RNase AWAY® (Molecular BioProducts, Inc)
attraction.
(b) If necessary, transfer particles to a suitable wash solution to (c) DNAZap™ (Ambion®, Inc.)
further remove any residual materials. WARRANTY
(c) Transfer particles to a suitable buffer and depress the
operating button or move the operating lever up to retract BioControl Systems, Inc. (BCS) warrants this product to be free from defects in
materials and workmanship, when stored under labeled conditions and used as
the magnets. Gently stir tips in the buffer solution to intended for 1 year from date of purchase. BCS agrees during the applicable
completely release the particles. warranty period to replace all defective products after return to BCS. BCS shall
not have obligation under this Limited Warranty to make replacements which
D. Ejecting Tips result, in whole or in part, from negligence of the Buyer, or from improper use
After completing transfer of magnetic particles, the tips can of the products, or use of the product in a manner for which it was not
indicated. Buyer shall notify BCS of any products which it believes to be
be ejected into a suitable biohazard container by depressing defective during the warranty period. At BCS option, such products shall be
the eject button. returned to BCS, transportation and insurance prepaid. BCS shall replace any
such product found to be defective, at no charge. Should BCS examination not
STORAGE disclose any defect covered by the foregoing warranty, BCS shall so advise
Buyers and dispose of the product in accordance with Buyer’s instructions.
Exposure to temperatures above 80°C can demagnetize the
magnetic tips. ©BioControl Systems, Inc. 2007. Printed in USA. Assurance GDS is a trademark
of BioControl. U.S. and foreign patents pending. All rights reserved.
The PickPen does not require any calibration and should not be
disassembled.

Part No., 73028, 73038 55057.R002.062008