0 évaluation0% ont trouvé ce document utile (0 vote)
132 vues30 pages
This document provides a summary of EMC requirements for medical devices, including emissions and immunity standards from the 1st through 4th editions of IEC 60601-1-2. It defines performance criteria for EMC testing and lists the standards, test levels, and applicable dates for electromagnetic emissions, electrostatic discharge, radiated RF immunity, electrical fast transients, surge withstand, conducted RF immunity, magnetic field immunity, dips/interruptions, and other EMC tests. The document is intended for internal use in understanding EMC requirements for medical devices.
This document provides a summary of EMC requirements for medical devices, including emissions and immunity standards from the 1st through 4th editions of IEC 60601-1-2. It defines performance criteria for EMC testing and lists the standards, test levels, and applicable dates for electromagnetic emissions, electrostatic discharge, radiated RF immunity, electrical fast transients, surge withstand, conducted RF immunity, magnetic field immunity, dips/interruptions, and other EMC tests. The document is intended for internal use in understanding EMC requirements for medical devices.
This document provides a summary of EMC requirements for medical devices, including emissions and immunity standards from the 1st through 4th editions of IEC 60601-1-2. It defines performance criteria for EMC testing and lists the standards, test levels, and applicable dates for electromagnetic emissions, electrostatic discharge, radiated RF immunity, electrical fast transients, surge withstand, conducted RF immunity, magnetic field immunity, dips/interruptions, and other EMC tests. The document is intended for internal use in understanding EMC requirements for medical devices.
Prepared By: Victor H. Kee, P.Eng Ultratech Engineering Labs Inc.
EMC REQUIREMENTS SUMMARY For Internal Use Only 2 of 30 26/05/2014 Table of Contents PUBLISHED DATE DEFINITIONS .............................................................................................................................. 3 PERFORMANCE CRITERIA DEFINITIONS ............................................................................................................... 3 EN 60601-1-2:1993 (1 ST Edition) Immunity for Medical Electrical Equipment .................................................................. 4 IEC/EN 60601-1-2:2001 (2 nd Edition) Immunity for Medical Electrical Equipment ............................................................ 5 IEC/EN 60601-1-2:2007 (3 nd Edition) Immunity for Medical Electrical Equipment ............................................................ 6 IEC 60601-1-2:2014 (4 th Edition) Immunity for Medical Electrical Equipment ................................................................... 7 Regulatory Impact Analysis .................................................................................................................................................... 26 Sections 15.21 & 15.105 - Information to User ...................................................................................................................... 29 EN 55011 Emissions Group 1 ................................................................................................................................................. 30
EMC REQUIREMENTS SUMMARY For Internal Use Only 3 of 30 26/05/2014 Published date Definitions doa - latest date of announcement of the European standard at national level
dop - latest date by which the EN has to be implemented at national level by publication of an identical national standard or by endorsement
dow - latest date by which the national standards conflicting with the EN have to be withdrawn
Performance Criteria Definitions Performance Criteria A for phenomena of a continuous nature: The apparatus shall continue to operate as intended. No degradation of performance or loss of function is allowed below a performance level specified by the manufacturer, when the apparatus if used as intended. In some cases the performance level may be replaced by a permissible loss of performance. If the minimum performance level or the permissible performance loss is not specified by the manufacturer then either of these may be derived from the product description and documentation (including leaflets and advertising) and what the user may reasonably expect from the apparatus if used as intended. Performance Criteria B for phenomena of a transient nature: The apparatus shall continue to operate as intended after the test. No degradation of performance or loss of function is allowed below a performance level specified by the manufacturer, when the apparatus if used as intended. In some cases the performance level may be replaced by a permissible loss of performance. During the test, degradation of performance is however allowed. No change of actual operating state or stored data is allowed. If the minimum performance level or the permissible performance loss is not specified by the manufacturer then either of these may be derived from the product description and documentation (including leaflets and advertising) and what the user may reasonably expect from the apparatus if used as intended. Performance Criteria C for power supply failure phenomena: Temporary loss of function is allowed, provided the loss of function is self recoverable or can be restored by the operation of the controls. Functions and or information protected by a battery backup shall not be lost. Performance Criteria D: The apparatus or sub-system shall continue to operate as intended. No degradation of performance or loss of function is allowed other than a failure into a safe-mode. Performance Criteria R for power supply failure phenomena: The equipment shall withstand the test without damage or other disturbances ( such as the corruption of software or misoperation of fault protection facilities) and shall operate properly within the specified limits after transient electromagnetic phenomenon has ceased. (It is not necessary to operate properly while the test condition is present). The exposure may cause the operation of fuses or other specified devices which have to be replaced or reset before normal operation is restored.
EMC REQUIREMENTS SUMMARY For Internal Use Only 4 of 30 26/05/2014 EN 60601-1-2:1993 (1 ST Edition) Immunity for Medical Electrical Equipment
Class A equipment CISPR 11 CISPR 11 N/A N/A Class B equipment CISPR 11 CISPR 11 N/A N/A
Immunity Requirements
Description
Severity Required Performance Criteria
Standard Electrostatic Discharge 4kV contact, 8kV air B IEC 801.2 Radiated RF Immunity 3V/m, 26-1000 MHz, 1kHz 80%AM modulation A IEC 801.3 Electrical Fast Transient 1kV on AC (2kV for permanent installed eqpt.), 500V on I/O and DC lines B IEC 801.4 Surge Withstand 2kV common mode, 1kV differential on AC lines B IEC 801.5 Conducted RF Immunity 3Vrms, 0.15-80 MHz, 1kHz 80% AM modulation (sel freq. 1, 2, 7.1, 13.56, 21, 27.12 & 68 MHz) A IEC 801.6 Magnetic Field Immunity 50 Hz, 3A/m A IEC 801.8 Dips Dropouts Interrupts Voltage variations 30% for 10 mS, 100% for 100 mS >95% for 5 sec 15% from Nominal B C C C IEC 801.11
EMC REQUIREMENTS SUMMARY For Internal Use Only 5 of 30 26/05/2014 IEC/EN 60601-1-2:2001 (2 nd Edition) Immunity for Medical Electrical Equipment (dop) 2002-08-01 (dow) 2004-11-01
Class A equipment CISPR 11 CISPR 11 IEC 1000-3-2 (230VAC, 50Hz) IEC 1000-3-3 (230VAC, 50Hz) Class B equipment CISPR 11 CISPR 11 IEC 1000-3-2 (230VAC, 50Hz) IEC 1000-3-3 (230VAC, 50Hz)
Immunity Requirements
Description
Severity Required Performance Criteria
Standard Electrostatic Discharge 6kV contact, 8kV air A IEC 61000-4-2 Radiated RF Immunity 3V/m, 80-2500MHz, 1kHz/2Hz 80%AM modulation
10V/m, 80-2500MHz, 1kHz/2Hz 80%AM modulation for life support equipment Dwell Time is 3 seconds for 2Hz modulation Dwell Time is 1 second for 1kHz modulation A IEC 61000-4-3 Electrical Fast Transient 2kV on AC and DC 5KHz Repetition 1kV on I/O A IEC 61000-4-4 Surge Withstand 2kV common mode, 1kV differential on AC lines (minimum 5 surges at each phase angle) A IEC 61000-4-5 Conducted RF Immunity 3Vrms, 0.15-80 MHz, 1kHz/2Hz 80% AM modulation
3Vrms, 0.15-80 MHz, 1kHz/2Hz 80% AM modulation for life support equipment
(Calibration required level set to 0-25% of target voltage) Patient coupled leads are tested with current clamp A IEC 61000-4-6 Magnetic Field Immunity 50 and 60 Hz, 3A/m A IEC 61000-4-8 Dips Dips Dropouts Interrupts Dip to 40% for 5 cycles (100 ms) Dip to 70% for 25 cycles (500ms) Dropout to 5% for 10 ms >95% for 5 sec
*A *A *A C IEC 61000-4-11 *Applicable to all life support equipment and any equipment with less than 1kVA input power. Equipment with input power above 1kVA and below 16A/phase need only meet Criteria B. EMC REQUIREMENTS SUMMARY For Internal Use Only 6 of 30 26/05/2014 IEC/EN 60601-1-2:2007 (3 nd Edition) Immunity for Medical Electrical Equipment (dop) 2008-02-01 (dow) 2012-06-01
Class A equipment CISPR 11 CISPR 11 IEC 1000-3-2 (230VAC, 50Hz) IEC 1000-3-3 (230VAC, 50Hz) Class B equipment CISPR 11 CISPR 11 IEC 1000-3-2 (230VAC, 50Hz) IEC 1000-3-3 (230VAC, 50Hz)
Immunity Requirements
Description
Severity Required Performance Criteria
Standard Electrostatic Discharge 6kV contact, 8kV air A IEC 61000-4-2 Radiated RF Immunity 3V/m, 80-2500MHz, 1kHz/2Hz 80%AM modulation
10V/m, 80-2500MHz, 1kHz/2Hz 80%AM modulation for life support equipment Dwell Time is 3 seconds for 2Hz modulation Dwell Time is 1 second for 1kHz modulation A IEC 61000-4-3 Electrical Fast Transient 2kV on AC and DC 5KHz Repetition 1kV on I/O A IEC 61000-4-4 Surge Withstand 2kV common mode, 1kV differential on AC lines (minimum 5 surges at each phase angle) A IEC 61000-4-5 Conducted RF Immunity 3Vrms, 0.15-80 MHz, 1kHz/2Hz 80% AM modulation
3Vrms, 0.15-80 MHz, 1kHz/2Hz 80% AM modulation for life support equipment
(Calibration required level set to 0-25% of target voltage) Patient coupled leads are tested with current clamp A IEC 61000-4-6 Magnetic Field Immunity 50 and 60 Hz, 3A/m A IEC 61000-4-8 Dips Dips Dropouts Interrupts Dip to 40% for 5 cycles (100 ms) Dip to 70% for 25 cycles (500ms) Dropout to 5% for 10 ms >95% for 5 sec
*A *A *A C IEC 61000-4-11 *Applicable to all life support equipment and any equipment with less than 1kVA input power. Equipment with input power above 1kVA and below 16A/phase need only meet Criteria B. EMC REQUIREMENTS SUMMARY For Internal Use Only 7 of 30 26/05/2014 IEC 60601-1-2:2014 (4 th Edition) Immunity for Medical Electrical Equipment (dop) 2015-01-01 (dow) 2017-04-01 Emissions Requirements Conducted Emissions Radiated Emissions Harmonic Emission Flicker
Class A equipment CISPR 11 CISPR 11 IEC 1000-3-2 (230VAC, 50Hz) IEC 1000-3-3 (230VAC, 50Hz) Class B equipment CISPR 11 CISPR 11 IEC 1000-3-2 (230VAC, 50Hz) IEC 1000-3-3 (230VAC, 50Hz)
Immunity Requirements
Description
Severity Required Performance Criteria
Standard Electrostatic Discharge 8kV contact, 15kV air Risk Assessment IEC 61000-4-2 Radiated RF Immunity 3V/m, 80-2700MHz, 1kHz 80%AM modulation Professional Healthcare Facility Environment
10V/m, 80-2700MHz, 1kHz 80%AM modulation Home Healthcare Environment Modulation frequencies and dwell times defined by Risk Assessment and test plan Risk Assessment IEC 61000-4-3 Electrical Fast Transient 2kV on AC and DC 100KHz Repetition 1kV on I/O 100KHz Repetition (patient coupling ports excluded) Risk Assessment IEC 61000-4-4 Surge Withstand 500V,1kV & 2kV common mode, 500V & 1kV differential on AC & DC lines (>3m) 2KV common mode on I/O lines (Outdoor cables) Risk Assessment IEC 61000-4-5 Conducted RF Immunity 3Vrms, 0.15-80 MHz, 1kHz 80% AM modulation 6Vrms in ISM bands (I/O cables <3m excluded) Patient coupled ports tested with current clamp Modulation frequencies defined by Risk Assessment Risk Assessment IEC 61000-4-6 Magnetic Field Immunity 50 and 60 Hz, 30A/m Risk Assessment IEC 61000-4-8 Dips Dips Dropouts
Interrupts Dip to 0% for 1 cycles @ 0phase angle Dip to 70% for 25/30 cycles @ 0 phase angle Dropout to 0% for 0.5 cycles @ 0, 45, 90, 135 , 180, 225, 270 & 315 phase angles 100% for 250/300 cycles Risk Assessment
IEC 61000-4-11
EMC REQUIREMENTS SUMMARY For Internal Use Only 8 of 30 26/05/2014 3 rd EDITION NOTABLE ITEMS TESTING REQUIREMENTS EMISSION ME Equipment and ME Systems are classified as either Class A or Class B and Group 1 or Group 2 as per CISPR 11 (Industrial, Scientific and Medical (ISM) Radio-Frequency Equipment Electromagnetic Disturbance Characteristics Limits and Methods of Measurement). Class A equipment is equipment suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes Class B equipment is equipment suitable for use in all establishments including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes Warning: Class A equipment is intended for use in an industrial environment. In the documentation for the user, a statement shall be incl uded drawing attention to the fact that there may be potential difficulties in ensuring electromagnetic compatibility in other environments, due to conducted as well as radiated disturbances.
Class B equipment is equipment suitable for use in domestic establishments and in establishments directly connected to a low voltage power supply network which supplies buildings used for domestic purposes. Compliance criteria EMC REQUIREMENTS SUMMARY For Internal Use Only 9 of 30 26/05/2014
EMC Risk Analysis and Test Plan
The risk assessment must identify what effected systems, subsystems and components will cause an effect and whether the affected part is "critical" to the function of the product. This risk assessment will form the basis for identification of pass/fail. A test plan drawing up how the equipment is to be setup, monitored and the test points involved is required to be carried out prior to any EMC tests. The essential function of the equipment must be clearly identified detailing the minimum performance criteria acceptable for tests of a continuous nature as well as for tests of a transient nature. The test plan will form the envelope in how the EMC tests are to be performed.
Production firmware/software must be provided for the EMC test portion as it has been shown in the past that software code can be an EMC threat.
Identification and Marking EMC REQUIREMENTS SUMMARY For Internal Use Only 10 of 30 26/05/2014 EMC REQUIREMENTS SUMMARY For Internal Use Only 11 of 30 26/05/2014 Information to User
a) A marking on the outside of the medical device that contains an RF transmitter or that applies RF energy for diagnosis or treatment; b) A marking on the outside of the medical device that uses the connector ESD exemption; c) A marking on the outside of the medical device that must be used in a shielded location; d) A statement that the medical device needs special precaution regarding EMC; e) A statement that the medical device can be affected by portable and mobile RF communications equipment; f) If the ESD exemptions is used: 1) The ESD warning symbol; 2) A warning that pins of the connector should not be touched; 3) An ESD precautionary procedure; 4) A recommendation that all staff be trained in the ESD precautionary procedure; 5) A specification of the minimum contents of the ESD precautionary procedure. g) If claimed, a warning that type A professional equipment may cause interference; h) In the technical description: 1) A list of cables, maximum lengths of cables, transducers and other accessories that may affect compliance; 2) A warning that use of accessories, cables, and transducers other than those specified may result in increased emissions and decreased immunity; 3) Tables regarding the emissions and immunity of the medical device; 4) A warning that the medical device should not be used adjacent to or stacked with other equipment; 5) A list of the products essential performance; 6) Any justifications for lower immunity levels; 7) A table with safety separation distances for portable and mobile RF transmitters. .
EMC REQUIREMENTS SUMMARY For Internal Use Only 12 of 30 26/05/2014 Guidance and Manufacturers Declaration
EMC REQUIREMENTS SUMMARY For Internal Use Only 13 of 30 26/05/2014 EMC REQUIREMENTS SUMMARY For Internal Use Only 14 of 30 26/05/2014
EMC REQUIREMENTS SUMMARY For Internal Use Only 15 of 30 26/05/2014
EMC REQUIREMENTS SUMMARY For Internal Use Only 16 of 30 26/05/2014
EMC REQUIREMENTS SUMMARY For Internal Use Only 17 of 30 26/05/2014
EMC REQUIREMENTS SUMMARY For Internal Use Only 18 of 30 26/05/2014 EMC REQUIREMENTS SUMMARY For Internal Use Only 19 of 30 26/05/2014 CISPR11 or EN 55011 Conduced and Radiated Emissions
All medical electrical equipment and/or systems7, must be correctly classified as defined in CISPR 11 as Group 1 or Group 2; Class A or Class B. This must be done in accordance with CISPR 11, based on the equipments intended use and intended site of use. The intended use is specified by the manufacturer, not by a testing laboratory.
CCC.3 Division into classes
EQUIPMENT and SYSTEMS predominantly intended for use in domestic environments and connected to the PUBLIC MAINS NETWORK (e.g. home care EQUIPMENT and EQUIPMENT for doctors offices in residential areas) should meet the requirements for CISPR 11 Class B.
EQUIPMENT and SYSTEMS that are intended to be connected (e.g. in hospitals) to dedicated supply systems (normally fed by separation transformers) should meet the requirements for either CISPR 11 Class A or Class B.
EQUIPMENT and SYSTEMS that are specified for use only in a shielded location may be classified based on compliance of the system formed by the EQUIPMENT or SYSTEM together with the specified type of shielded location, i.e. with the assumption that the EQUIPMENT or SYSTEM has been installed in a shielded location meeting the EQUIPMENT or SYSTEM manufacturers specifications for minimum RF shielding effectiveness and minimum RF filter attenuation. If classification is made on this basis, 6.8.3.201 c) 2) of this standard requires a statement of this fact to appear in Table 201, as well as a recommendation to verify the actual shielding effectiveness and filter attenuation of the shielded location.
Tests on medical equipment shall include all modes of operation including DC operations, AC 120V, AC 230V, AC 120V with DC Backup, AC 230V with DC backup, Standby mode, Blower On mode, Dialup and communications modeetc.
EMC REQUIREMENTS SUMMARY For Internal Use Only 20 of 30 26/05/2014 EN 61000-3-2/3 Harmonics Emissions and Voltage Flicker
Harmonics emissions and voltage flicker is required to be performed.
EN 61000-4-2 Electrostatic Discharge
All ports which may be exposed during normal operation must have detachable cover to protect the interface pins from ESD damage. If a protective cover is not provided, the port pins become a valid test point for ESD. The other option is to provide adequate labeling to exempt the port from ESD testing.
EN 61000-4-3 Radiated Immunity
The range is extended to 2500MHz up from 1000MHz. Modulation frequency is set to 2Hz Amplitude Modulation with 80% depth for Respironics Equipment. Dwell time at each frequency interval is 3 seconds minimum
EN 61000-4-4 Electrical Fast Transients EMC REQUIREMENTS SUMMARY For Internal Use Only 21 of 30 26/05/2014
For PATIENT coupling points that do not have a conductive contact, the PATIENT coupling point shall be terminated with the artificial hand and RC element specified in CISPR 16-1. The metal foil of the artificial hand shall be sized and placed to simulate the approximate area and location of PATIENT coupling in NORMAL USE. The metal foil of the artificial hand shall be connected to terminal M of the RC element and the other terminal of the RC element shall be connected to the ground reference plane.
For PATIENT coupling points that have conductive contact to the PATIENT, terminal M of the RC element (see CISPR 16-1) shall be connected directly to the conductive PATIENT connection, and the other terminal of the RC element shall be connected to the ground reference plane. If normal operation of the EQUIPMENT or SYSTEM cannot be verified with terminal M connected to the coupling point, an insulating material with a maximum thickness of 5 mm may be applied between the metal foil of the artificial hand (see CISPR 16-1) and the PATIENT coupling point. In this case, the metal foil of the artificial hand shall be sized and placed to simulate the approximate area and location of PATIENT coupling in NORMAL USE, and terminal M of the RC element shall be connected to the metal foil but not to the PATIENT coupling point. The other terminal of the RC element shall be connected to the ground reference plane in all cases.
EN 61000-4-5 Surge Withstand
The test methods and equipment specified by IEC 61000-4-5 for the combination wave test shall apply, with the following modifications:
1) Only power lines and a.c. inputs to a.c.-to-d.c. converters and battery chargers are tested; however, all EQUIPMENT and SYSTEM cables shall be attached during the test.
2) Five surges at each voltage level and polarity shall be applied to each power line at each of the following a.c. voltage waveform angles: 0 or 180, 90 and 270. NOTE While testing at both 0 and 180 is allowed, testing at only one of these two phase angles, in addition to 90 and 270, is required.
*3) EQUIPMENT and SYSTEMS that do not have a surge protection device in the primary power circuit may be tested only at 2 kV line(s) to ground and 1 kV line(s) to line(s). However, in case of dispute, the EQUIPMENT or SYSTEM shall comply at all the IMMUNITY TEST LEVELS specified in 36.202.5 a).
*4) Class II EQUIPMENT and SYSTEMS without any grounded interconnections are exempt from line(s) to ground testing.
EMC REQUIREMENTS SUMMARY For Internal Use Only 22 of 30 26/05/2014 5) For INTERNALLY POWERED EQUIPMENT and SYSTEMS without the option of a.c. or d.c. power inputs, this test does not apply.
6) For EQUIPMENT and SYSTEMS that have, for power input, multiple voltage settings or autoranging voltage capability, the test shall be performed at the minimum and maximum RATED input voltages. The test may be performed with the EQUIPMENT or SYSTEM powered at any one of its NOMINAL power frequencies.
7) For EQUIPMENT and SYSTEMS with internal battery backup, it shall be verified that the EQUIPMENT or SYSTEM continues operation from mains power after the tests specified in this subclause.
EN 61000-4-6 Conducted RF Immunity
Modulation frequency is set to 2Hz Amplitude Modulation with 80% depth for Medical Equipment related to biological functions else it is 1kHz, AM with 80% depth.. Dwell time at each frequency interval is 3 seconds minimum
EN 61000-4-8 Magnetic Field Immunity
EMC REQUIREMENTS SUMMARY For Internal Use Only 23 of 30 26/05/2014 EN 61000-4-11 Voltage Dips
EMC REQUIREMENTS SUMMARY For Internal Use Only 24 of 30 26/05/2014 4 th EDITION NOTABLE ITEMS
1. Only Voltage dips, short interruptions and voltage variations need to be tested at more than one line voltage. All other EMC tests to be tested only at one line voltage/frequency. 3rd Edition required max and min line voltage tests on EFT, Surge, Dips/Dropouts and Magnetic Field Immunity. 2. Both standby and operating modes need to be tested for emissions and the modes for maximum emissions should be used. For immunity, it is suggested that standby mode be included especially where medical equipment is in standby mode for long periods of time. 3. The environment the medical equipment is to be operated in dictates the Emissions standard and class to be used. For portable devices apply CISPR 11 Class B, CISPR 25/ISO 7637-2 and RTCA-DO-160/ISO 7137/EUROCAE ED-14 with harmonics/flicker. Interesting to note that for professional healthcare facility environment, harmonics/flicker is referred to as a Basic EMC standard although it appears in the Official Journal as a Product Standard. 4. Specific procedures are provided on when EUT damage as a result of EMC tests occur. If the test is repeated three times and the equipment still meets basic safety and essential performance, it is deemed to pass that test. This provides more wiggle room when an ancillary function is affected but the essential performance remains unaffected. 5. Specific medical environments are now clearly defined as; Emergency Medical Services Environment, Home Healthcare Environment, Professional Healthcare Facility Environment and Special Environments. Professional healthcare environment is considered a controlled EM environment. MRI, x-ray and other medical equipment requiring shielded rooms have special label and user information detailing the use in special environments. X-ray generators operating in intermittent mode may have the quasi-peak limits relaxed by 20dB for discontinuous conducted and radiated emissions. 6. ESD Levels for contact discharge has been increased from 6kV to 8kV and air discharge has been increased from 8kV to 15kV. This is major item change that will have huge consequences as current medical equipment will be affected by the much higher ESD test levels to be met. 7. Radiated Immunity upper frequency has been extended to 2.7GHz from 2.5GHz. More importantly, all home healthcare environment equipment must meet 10V/m which will also apply to all portable equipment. The 4th edition outlines risk assessment to also take into account fields as high as 28V/m when portable wireless devices are used within 30cm of the medical device. This could mean additional test levels and modulations.
8. Magnetic Field Immunity is increased from 3A/m to 30A/m. 9. Electrical Fast Transient levels stay the same but the newer 100kHz repetition rate is now mandatory over the 5kHz rate allowed before. Patient coupled ports are excluded. 10. Surge levels stay the same but previous lower level surge is now mandated so surge test is carried out at 500V, 1000V for differential mode and 500V, 1000V and 2000V for common mode on AC and DC ports. Additional 2kV common mode surge has been added for ports providing outdoor I/O cables.
EMC REQUIREMENTS SUMMARY For Internal Use Only 25 of 30 26/05/2014 11. Conducted RF immunity levels have been raised to 6Vrms in the ISM & Amateur radio bands and 3Vrms in non ISM & Amateur radio bands. ISM & Amateur radio bands are;
I/O cables shorter than 3m are excluded. Patient coupled cables are tested with current clamp and tubes with liquids are also tested as patient coupled cables. 12. Medical equipment capable of being powered by vehicles battery must meet the ISO 7637-2 automotive transient pulses. 13. The artificial hand is required to be applied onto the medical device. The artificial hand is a requirement for CISPR 11 conducted emissions and according to CISPR 16-1-2, for conducted RF immunity. Under the 4th edition, the 220pf/510 circuit must be used to terminate patient coupled points that have conductive contact to the patient. 14. A Test plan with 27 items is now specified to be completed before EMC tests are performed. 15. Patient coupled conducted emissions is recommended to be measured and limited to 24dBuA from 1-30 MHz. This is informative but may be included in the test plan. 16. Performance degradation is now defined in the risk assessment and test plan and guidelines are provided as example cases in Annex H.
EMC REQUIREMENTS SUMMARY For Internal Use Only 26 of 30 26/05/2014 Regulatory Impact Analysis The 4th edition IEC 60601-1-2:2014 represents a huge step in re-defining the electromagnetic performance for medical devices and raises the immunity levels required for medical devices to operate safely and within its essential performance requirements. Many existing medical products will not meet the 4th edition requirements and a gap analysis will reveal that incrementally, the following minimum additional tests need to be conducted from the 3rd edition standard;
Electrostatic Discharge 8kV contact, 15kV air Radiated RF Immunity 3V/m, 2500-2700MHz, 1kHz 80%AM modulation Professional Healthcare Facility Environment 10V/m, 80-2700MHz, 1kHz 80%AM modulation Home Healthcare Environment Modulation frequencies and dwell times defined by Risk Assessment and test plan Electrical Fast Transient 2kV on AC and DC, 100KHz Repetition rate 1kV on I/O 100KHz Repetition rate (patient coupling ports excluded) Surge Withstand 2KV common mode on I/O lines (Outdoor cables) Conducted RF Immunity 6Vrms in ISM & Amateur Radio bands (I/O cables < 3m excluded) Patient coupled ports tested with current clamp Magnetic Field Immunity 50 and 60 Hz, 30A/m Dips Dips Dropouts Interrupts Dip to 0% for 1 cycles @ 0phase angle Dip to 70% for 25/30 cycles @ 0 phase angle Dropout to 0% for 0.5 cycles @ 0, 45, 90, 135 , 180, 225, 270 & 315 phase angles 100% for 250/300 cycles Gap Analysis of Test Requirements
A gap analysis of the incrementals will reveal that only the 3rd edition surge tests may be not be required to be re- tested. This being the case, it may be more expeditious to just test everything to the 4th edition and the latest requirements than to provide the gap analysis and technical rational linking 3rd edition reports/test plans and risk analysis.
There is usually a 3 year transition period provided before an adopted standard become mandatory. Manufactures are advised to start addressing the elevated EMC levels that will impact the current and future medical products as design changes will probably be required to address any EMC deficiencies detected. New products currently in the pipes may need to have their design specifications revised to account for the increased EMC immunity the device will be eventually required to meet.
EMC REQUIREMENTS SUMMARY For Internal Use Only 27 of 30 26/05/2014 FCC part 15 Subpart B Emissions
Exemption for Medical Devices
The following devices are subject only to the general conditions of operation in Secs. 15.5 and 15.29 and are exempt from the specific technical standards and other requirements contained in this part. The operator of the exempted device shall be required to stop operating the device upon a finding by the Commission or its representative that the device is causing harmful interference. Operation shall not resume until the condition causing the harmful interference has been corrected. Although not mandatory, it is strongly recommended that the manufacturer of an exempted device endeavor to have the device meet the specific technical standards in this part.
(e) Specialized medical digital devices (generally used at the direction of or under the supervision of a licensed health care practitioner) whether used in a patient's home or a health care facility. Non-specialized medical devices, i.e., devices marketed through retail channels for use by the general public, are not exempted. This exemption also does not apply to digital devices used for record keeping or any purpose not directly connected with medical treatment.
Conducted Emissions Frequency (MHz) QP Class A (dBV) AVG Class A (dBV) QP Class B (dBV) AVG Class B (dBV) 0.15 to 0.5 79 66 66 to 56* 56 to 46* 0.5 to 5 73 60 56 46 5 to 30 73 60 60 50 * linearly decreasing with the log of the frequency Notes. 1. The tighter limit applies at the band edges. 2. Both limits must be met. If the quasi-peak value meets the average limit then the limits are deemed to have been met. 3. The line voltage used must be 120VAC for FCC compliance. 4. The limit applies to any equipment which is connected to the AC mains network regardless of whether there is power present or if the unit is in battery operation. EMC REQUIREMENTS SUMMARY For Internal Use Only 28 of 30 26/05/2014 Radiated Emission @ 3m Frequency (MHz) QP Class A Limit (dBV/m) QP Class B Limit (dBV/m) 30 to 88 49.54 40.00 88 to 216 53.98 43.52 216 to 960 56.90 46.02 Above 960 60.00 53.98
Radiated Emission @ 10m Frequency (MHz) QP Class A Limit (dBV/m) QP Class B Limit (dBV/m) 30 to 88 39.08 29.54 88 to 216 43.52 33.06 216 to 960 46.44 35.56 Above 960 49.54 43.52
Notes. 1. The preferred distance for Class B is 3 meters. The preferred distance for Class A is 10 meters. 2. The tighter limit applies at the band edges. 3. The measurement resolution bandwidth is 120kHz. For 30-1000MHz and 1MHz above 1GHz. 4. Above 1GHz, an average detector is used. FCC Labeling Requirements for Computing Devices
EMC REQUIREMENTS SUMMARY For Internal Use Only 29 of 30 26/05/2014 Sections 15.21 & 15.105 - Information to User The users manual or instruction manual for an intentional or unintentional radiator shall caution the user that changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
(a) For a Class B digital device or peripheral, the instructions furnished the user shall include the following or similar statement, placed in a prominent location in the text of the manual.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference and (2) this device must accept any interference received, including interference that may cause undesired operation.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital devices, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one of more of the following measures:
Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help.
Warning: Changes or modifications not expressly approved by <manufacturer> could void the users authority to operate the equipment.
EMC REQUIREMENTS SUMMARY For Internal Use Only 30 of 30 26/05/2014 EN 55011 Emissions Group 1
Conducted Emissions Frequency (MHz) QP Class A (dBV) AVG Class A (dBV) QP Class B (dBV) AVG Class B (dBV) 0.15 to 0.5 79 66 66 to 56* 56 to 46* 0.5 to 5 73 60 56 46 5 to 30 73 60 60 50 * linearly decreasing with the log of the frequency Notes 5. The tighter limit applies at the band edges. 6. Both limits must be met. If the quasi-peak value meets the average limit then the limits are deemed to have been met. 7. The line voltage used must be 230VAC for European compliance. 8. The limit applies to any equipment which is connected to the AC mains network regardless of whether there is power present or if the unit is in battery operation. Radiated Emission @ 3m Frequency (MHz) QP Class A Limit (dBV/m) QP Class B Limit (dBV/m) 30 to 230 50 40.5 230 to 1000 57 47.5
Radiated Emission @ 10m Frequency (MHz) QP Class A Limit (dBV/m) QP Class B Limit (dBV/m) 30 to 230 39.5 30 230 to 1000 46.5 37
Radiated Emission @ 30m Frequency (MHz) QP Class A Limit (dBV/m) QP Class B Limit (dBV/m) 30 to 230 30 20.5 230 to 1000 37 27.5 Notes. 1. The tighter limit applies at the band edges. 2. The preferred distance for Class A is 30 meters and for Class B is 10 meters.
Medical Electrical Equipment - Part 2-41: Particular Requirements For Basic Safety and Essential Performance of Surgical Luminaires and Luminaires For Diagnosis