Level measurement

in liquids
Ultrasonic
VEGASON61
VEGASON62
VEGASON63
Product Information
Contents
2 Ultrasonic Level measurement in liquids
Contents
1 Description of the measuring principle. . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Type overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Mounting information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4 Electrical connection
4.1 General requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.2 Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.3 Connection cable and installation . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.4 Cable screening and grounding. . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
4.5 Wiring plans. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
5 Adjustment
5.1 Adjustment, general . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
5.2 Compatibility acc. to NAMUR NE 53 . . . . . . . . . . . . . . . . . . . . . . . . 16
5.3 Adjustment with the indicating and adjustment module PLICSCOM . . 16
5.4 Adjustment with PACTware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
7 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
8 Product code. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Take note of safety
instructions for Ex
application
Please note the Ex specic safety information which you will nd on our homepage
www.vega.com\services\downloads and which come with the appropriate instru-
ment. In hazardous areas you should take note of the appropriate regulations, con-
formity and type approval certicates of the sensors and power supply units. The
sensors must only be operatedonintrinsically safe circuits. The permissible electrical
values are stated in the certicate.
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
1 Description of the measuring principle
Measuring principle
Short ultrasonic pulses in the range of 35 kHz to 70 kHz are emittedby the transducer
to the product surface, reected there and received by the transducer. The pulses
travel at the speed of sound - the elapsed time from emission to reception of the
signals depends on the level in the vessel.
The latest microcomputer technology and the proven ECHOFOX software select the
level echofromamonganynumber of falseechoesandcalculatetheexact distanceto
the product surface. An integrated temperature sensor detects the temperature in the
vessel and compensates the inuence of temperature on the signal running time.
By simply entering the vessel dimensions, a level-proportional signal is generated
from the distance. It is not necessary to ll the vessel for adjustment.
Wide application range
VEGASON 61, 62 and 63 ultrasonic sensors are especially suitable for level measu-
rement of liquids, but are also good for solids. The instruments dier in the measuring
range, the transducer version and the process tting. Through dierent, adapted
emitting frequencies, levels in a measuring range of 5 15 m can be measured.
Resistant materials for transducers and process ttings also allow applications in
corrosive products (depending on the model). A practical mounting strap (option)
enables easy orientation of VEGASON 63.
Unaected by product properties
Fluctuations in product composition or even complete product changes do not inu-
ence the measuring result. A fresh adjustment is not necessary.
Service and maintenance friendly
Thanks tothenon-contact measuringprinciple, VEGASON61, 62and63sensors are
particularly easy to service and maintain.
1.1 Application examples
Open basins
Fig. 1: Level measurement in an open basin with VEGASON 61
Atypical applicationfor VEGASON61 sensors is level measurement onopen basins.
Products such as rain water or sewage water, i.e. with impurities. Here is where the
advantagesof non-contact measurement withVEGASONcomeintotheir own: simple
Description of the measuring principle
Ultrasonic Level measurement in liquids 3
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
andmaintenance-free. Thedegreeof pollutionof thewater or anaccumulationof mud
in the basin is not important, because VEGASON measures the surface.
Sludge container
Fig. 2: Level measurement in a container with VEGASON 63
Insewagetreatment plants, theaccumulatedsludgeis dewateredandtransportedvia
conveyor belts to containers. The VEGASON 63 sensor measures the lling of the
container. An empty container can thus be readied in good time before the max. level
is reached. Thanks to the metal transducer diaphragm, measurement functionality is
also ensured even under conditions of uctuating temperatures and steam genera-
tion.
Description of the measuring principle
4 Ultrasonic Level measurement in liquids
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
2 Type overview
VEGASON 61 VEGASON 62
Applications: liquids and solids in virtually all
industries, particularly in water and
waste water management
as VEGASON 61
Measuring range: Liquids: 0.25 5 m (0.8 16.4 ft)
Solids: 0.25 2 m (0.8 6.6 ft)
Liquids: 0.4 8 m (1.3 26.2 ft)
Solids: 0.4 3.5 m (1.3 11.5 ft)
Process tting: G1A of PVDF G2A of PVDF
Process temperature: -40 +80C (-40 +176F) -40 +80C (-40 +176F)
Process pressure: -0.2 2 bar (-20 200 kPa) -0.2 2 bar (-20 200 kPa)
(-2.9 29 psi) (-2.9 29 psi)
VEGASON 63
Applications: liquids andsolidsinvirtuallyall industries
Measuring range: liquids: 0.6 15 m (2 49.2 ft)
solids: 0.6 7 m (2 23 ft)
Process tting: compression ange or mounting strap
Process temperature: -40 +80C (-40 +176F)
Process pressure: -0.2 1.0 bar (-20 100 kPa)
(-2.9 14.5 psi)
Type overview
Ultrasonic Level measurement in liquids 5
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Indicating and
adjustment module
PLICSCOM
Housing
Plastic Stainless steel Aluminium Aluminium
(double cham-
ber)
Electronics
4 20 mA/
HART
4 20 mA/
HART four-
wire
Probus PA Foundation
Fieldbus
Sensors
Transducer
1"
Transducer 2" Transducer 4"
Approvals
Gas explosion
protection
Type overview
6 Ultrasonic Level measurement in liquids
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
3 Mounting information
The reference plane for the measurement is the lower edge of the transducer. All
statements concerning the measuring range as well as the internal signal processing
refer to this.
With all instruments, a min. distance fromthe lower edge of the ange - the so-called
dead zone, in which measurement is not possible - must be maintained. The exact
value of the dead zone, depending on the instrument version, is stated in the Techni-
cal data.
1
2
Fig. 3: Min. distance to the max. level
1 Dead zone
2 Reference plane for the measurement
Note:
If the product reaches the transducer, buildup can formon it over a period of time and
later cause measurement errors.
1 3 2
100%
0%
Fig. 4: Measuring range and max. measuring distance
1 full
2 empty (max. measuring distance)
3 max. measuring range
Gauge pressure in the vessel does not inuence VEGASON. Low pressure or va-
cuum, however damp the ultrasonic pulses. This inuences the measuring result,
particularly if the level is very low. With pressures under -0.2 bar (-20 kPa) use a
dierent measuring principle, e.g. radar or guided microwave.
When mounting VEGASON, keep a distance of at least 200 mm to the vessel wall. If
thesensor is installedinthecenter of dishedor spherical vessel tops, multiple echoes
can arise. These can, however, be faded out by an appropriate adjustment.
Mounting information
Ultrasonic Level measurement in liquids 7
Measuring range
Pressure/Vacuum
Installation position
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
If you cannot keep this distance, a false echo storage should be carried out during
setup. This applies particularly if buildup on the vessel wall is expected. In this case,
we recommend repeating the false echo storage later on with existing buildup.
1
2
> 200 mm
Fig. 5: Mounting on dished vessel tops
1 Reference plane
2 Vessel center or symmetry axis
In vessels with conical bottom it can be advantageous to mount the sensor in the
center of the vessel, as measurement is then possible down to the lowest point of the
vessel bottom.
Fig. 6: Vessel with conical bottom
Themountingsocket shouldpreferably be dimensionedtoallowthelower edge of the
transducer to protrude at least 10 mm out of the socket.
Mounting information
8 Ultrasonic Level measurement in liquids
Socket
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
c
a
.

1
0

m
m
Fig. 7: Recommended socket mounting
If the reective properties of the medium are good, you can mount VEGASON on
sockets longer than the transducer. You will nd recommended values of the socket
heights inthefollowingillustrations. Thesocket endshouldbesmoothandburr-free, if
possible also rounded. A false echo storage is recommended.
>

1
0

m
m
d
h
d h
100 mm/4"
150 mm/6"
300 mm/12"
400 mm/16"
Fig. 8: Deviating socket dimensions with VEGASON 61
>

1
0

m
m
d
h
d h
100 mm/4"
150 mm/6"
200 mm/8"
300 mm/12"
400 mm/16"
500 mm/20"
Fig. 9: Deviating socket dimensions with VEGASON 62
Mounting information
Ultrasonic Level measurement in liquids 9
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
d
4
5

h
m
a
x
.
d
100 mm/4"
150 mm/6"
200 mm/8"
250 mm/10"
300 mm
400 mm
500 mm
600 mm
h
max.
Fig. 10: Deviating socket dimensions with VEGASON 63
With liquids, align the sensor as close to vertical as possible to achieve optimum
measuring results.
Fig. 11: Orientation in liquids
To reduce the min. distance to the medium, you can also mount VEGASON with a
beam deector of corrosion-resistant material. By doing this, it is possible to ll the
vessel nearlytomaximum. Suchanarrangement is suitableprimarily for openvessels
such as e.g. overow basins.
~200
~
4
0
0
x
4
0
0
4
5

Fig. 12: Beam deector


Theultrasonic sensor shouldbe installedat a locationwherenoinstallations cross the
ultrasonic beam.
Vessel installations such as, for example, ladders, limit switches, heating spirals,
struts, etc. can cause false echoes superimposed on the wanted echo. Make sure
when planning your measuring location that the ultrasonic signals have "free access"
to the measured product.
If there are existing vessel installations, a false echo storage should be carried out
during setup.
Mounting information
10 Ultrasonic Level measurement in liquids
Sensor orientation
Vessel installations
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
If large vessel installations such as struts or supports cause false echoes, these can
be attenuated through supplementary measures. Small, inclined sheet metal or plas-
tic baes above the installations scatter the ultrasonic signals and avoid direct false
echoes.
Fig. 13: Cover smooth proles with deectors
Theinstruments must not bemountedinor abovethellingstream. Makesurethat the
product surface and not the inowing material is detected.
Fig. 14: Inowing liquid
Through the action of lling, stirring and other processes in the vessel, dense foams
which considerably damp the emitted signals may form on the product surface.
If foams are causing measurement errors, the sensor should be used in a standpipe
or, alternatively, the more suitable VEGAFLEX guided microwave sensors should be
used.
Guided microwaves are not inuenced by foam generation and are particularly sui-
table for such applications.
If there are strong air currents in the vessel, e.g. due to strong winds in outdoor
installations, or because of air turbulence, you should mount VEGASON in a stand-
pipe or use a dierent measuring principle, e.g. radar or guided radar (TDR).
By using a standpipe (surge or bypass tube), the inuence of vessel installations,
foam generation and turbulence is excluded.
Mounting information
Ultrasonic Level measurement in liquids 11
Inowing material
Foam
Air ow
Standpipe measurement
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
max.
min.
1
Fig. 15: Standpipe in a tank
1 Vent hole 5 10 mm
Astandpipemeasurement withVEGASONis possiblewithtubediameters larger than
40 mm (VEGASON 61) or 50 mm (VEGASON 62).
Avoid large gaps and thick welding joints when connecting the tubes. A false echo
storage should always be carried out.
Measurement in a standpipe is not recommended for very adhesive products.
Mounting information
12 Ultrasonic Level measurement in liquids
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
4 Electrical connection
4.1 General requirements
The power supply range can dier depending on the instrument version. The exact
range is stated in the Technical data.
Take note of country-specic installation standards (e.g. the VDE regulations in Ger-
many) as well as prevailing safety regulations and accident prevention rules.
In hazardous areas you should take note of the appropriate regulations, conformity
and type approval certicates of the sensors and power supply units.
4.2 Supply voltage
Power supply andcurrent signal arecarriedover thesametwo-wireconnectioncable.
The requirements on thepower supply are statedin theTechnical data of this Product
Information manual.
The VEGA power supply units VEGATRENN 149AEx, VEGASTAB 690, VEGADIS
371 as well as the VEGAMET signal conditioning instruments are suitable for power
supply. When one of these instruments is used, a reliable separation of the supply
circuit from the mains circuits acc. to DIN VDE 0106 part 101 is ensured for VEGA-
SON.
Power supply and current output are carried on two separate connection cables.
The standard version can be operated with an earth-connected current output, the
Exd version must be operated with a oating output.
The instrument is designed in protection class I. To maintain this protection class, it is
absolutely necessary that the ground conductor be connected to the internal ground
conductor terminal.
The power supply is provided via a Probus DP/PA segment coupler or a VEGALOG
571 EP input card.
Power supply via the H1 Fieldbus cable.
4.3 Connection cable and installation
The sensors are connected with standard cable without screen. An outer cable dia-
meter of 5 9 mm ensures the seal eect of the cable entry.
As an option, VEGASON are also available with standard plug connectors (see
Technical data).
If strongelectromagnetic interference is expected, screenedcableshouldbe usedfor
the signal lines.
The installation must be carried out acc. to the appropriate bus specication. VEGA-
SON is connected respectively with screened cable acc. to the bus specication.
Power supply anddigital bus signal are transmitted via the same two-wire connection
cable. Make sure that the bus is terminated via appropriate terminating resistors.
In Ex applications, the corresponding installation regulations must be noted for the
connection cable.
Electrical connection
Ultrasonic Level measurement in liquids 13
General
4 20 mA/HART two-wire
4 20 mA/HART four-wire
Probus PA
Foundation Fieldbus
Probus PA, Foundation
Fieldbus
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
4.4 Cable screening and grounding
If screened cable is required, the cable screen must be connected on both ends to
ground potential. If potential equalisation currents are expected, the connection on
the evaluation side must be made via a ceramic capacitor (e.g. 1 nF, 1500 V).
In systems with potential equalisation, connect the cable screen at the power supply
unit, in the connection box and at the sensor directly to ground potential.
In systems without potential equalisation, connect the cable screen directly to ground
potential only at the power supply unit and at the sensor - do not connect to ground
potential in the connection box or the T-distributor.
4.5 Wiring plans
I
2
C
Display
1
1 2 5 6 7 8
Fig. 16: Connection HART two-wire, Probus PA, FoundationFieldbus
1 Power supply and signal output
I
2
C
1
1 2
Fig. 17: Connection HART two-wire, Probus PA, FoundationFieldbus
1 Power supply and signal output
4 ... 20 mA
PE
/ L
/ N
L1 N GND
1 2
1 2 3 4
4...20mA IS
Fig. 18: Connection 4 20 mA/HART four-wire
1 Supply voltage
2 Signal output
Electrical connection
14 Ultrasonic Level measurement in liquids
Probus PA, Foundation
Fieldbus
Single chamber housing
Double chamber hou-
sing two-wire
Double chamber hou-
sing 4 20 mA/HART
four-wire
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
+
-
1
2
Fig. 19: Wire assignment, connection cable
1 br (+) and bl () for power supply or to the processing system
2 Screen
Electrical connection
Ultrasonic Level measurement in liquids 15
Housing IP 66/IP 68,
1 bar
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Adjustment
16 Ultrasonic Level measurement in liquids
5 Adjustment
5.1 Adjustment, general
Per theelectronics version, VEGASONcanbe adjustedwiththefollowingadjustment
media:
l the indicating and adjustment module PLICSCOM
l an adjustment software acc. to FDT/DTMstandard, e.g. PACTware and PC
l a HART handheld (4 20 mA/HART)
l the adjustment programAMS or PDM
l a conguration tool (Foundation Fieldbus)
The entered parameters are generally saved in VEGASON, optionally also in
PLICSCOMor in the respective adjustment program.
5.2 Compatibility acc. to NAMUR NE 53
VEGASON meet NAMUR recommendation NE 53. VEGA instruments are generally
upward and downward compatible:
l sensor software for DTM-VEGASON HART, PA or FF
l DTMVEGASON for adjustment software PACTware
l adjustment module PLICSCOMfor sensor software
The parameter adjustment of the basic sensor functions is independent of the soft-
ware version. The range of available functions depends on the respective software
version of the individual components.
5.3 Adjustment with the indicating and adjustment module
PLICSCOM
PLICSCOMis apluggableindicationandadjustment modulefor plics

sensors. It can
be placed infour dierent positions on theinstrument (each displacedby 90). Indica-
tion and adjustment are made via four keys and a clear, graphic-capable dot matrix
indication. The adjustment menu with language selection is clearly structured and
enables easy setup. After setup, PLICSCOMserves as indicatinginstrument: through
the screwed cover with glass insert, measured values can be read directly in the
requested unit and presentation.
1.1
2
3
1
Fig. 20: Indicating and adjustment elements
1 LC display
2 Indication of the menu item number
3 Adjustment keys
Setup and indication
PLICSCOM adjustment
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Adjustment
Ultrasonic Level measurement in liquids 17
l [OK] key:
- move to the menu overview
- conrm selected menu
- edit parameter
- save value
l [>] key to select:
- menu change
- list entry
- editing position
l [+] key:
- modify value of a parameter
l [ESC] key:
- interrupt input
- jump to the next higher menu
5.4 Adjustment with PACTware
Independent of the respective signal output, whether 4 20 mA/HART, Probus PA
or Foundation Fieldbus, VEGASON sensors can be adjusted directly on the instru-
ment via PACTware. To adjust with PACTware, an instrument driver for the
particular VEGASON model is required. All currently available VEGA DTMs are pro-
vided in a DTM Collection with the current PACTware version on CD. They are
available fromthe responsible VEGAagency for a token fee. The basic version of this
DTMCollection incl. PACTwareis available as a free-of-charge download fromthe
Internet.
To use the entire range of functions of the DTM incl. project documentation, a DTM
licence is required for the particular instrument family, e.g.VEGASON . This licence
can be bought from the responsible VEGA agency.
~
=
Power supply
VEGACONNECT 3
PACTware /
TM
>PA<
2
3
1
Fig. 21: PC connected directly to the sensor
1 RS232 connection
2 VEGASON
3 IC adapter cable for VEGACONNECT 3
To adjust with PACTware, a VEGACONNECT 3 with IC adapter cable (art. no.
2.27323) as well as a power supply unit is necessary in addition to the PC and the
suitable VEGA-DTM.
Key functions
PACTware/DTM
Connecting the PC di-
rectly to the sensor
Connecting the PC to
the signal cable
(4 20 mA/HART)
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Adjustment
18 Ultrasonic Level measurement in liquids
2
3
1
4
~
=
Power supply
VEGACONNECT 3
PACTware /
TM
Fig. 22: Connecting the PC to the signal cable
1 RS232 connection
2 VEGASON
3 HART adapter cable for VEGACONNECT 3
4 HART resistance 250 Ohm
To adjust with PACTware, a VEGACONNECT 3 with HART adapter cable (art. no.
2.25397) as well as a power supply unit and a HART resistor with approx. 250 Ohmis
required in addition to the PC and suitable VEGA DTM.
Note:
With power supply units with integrated HART resistor (inner resistor approx.
250 Ohm) there is no additional external resistor necessary. This applies e.g. for
VEGA instruments VEGATRENN 149A, VEGADIS 371, VEGAMET 381. Also stan-
dard Ex separators are normally equipped with a sucient current limitation resistor.
In these cases, VEGACONNECT3 can connectedin parallel to the 4 20 mAcable.
2
3
1
4
5
Power supply
VEGACONNECT 3
PACTware /
TM
Fig. 23: Connecting the PC to the signal cable
1 RS232 connection
2 VEGASON
3 HART adapter cable for VEGACONNECT 3
4 HART resistance 250 Ohm
5 4 20 mA processing
To adjust with PACTware, a VEGACONNECT 3 with HART adapter cable (art. no.
2.25397) and a HART resistor with approx. 250 Ohmis required in addition to the PC
and the suitable VEGA-DTM.
Connecting the PC to
the signal cable
(4 20 mA/HART four-
wire)
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Technical data
Ultrasonic Level measurement in liquids 19
6 Technical data
General data
Materials, wetted parts
- Process tting PVDF
- Transducer PVDF
- Seal transducer/process tting EPDM
Materials, wetted parts
- Mounting strap 1.4301
- Process tting UP
- transducer diaphragm 1.4571
- Seal transducer/process tting EPDM
Materials, non-wetted parts
- compression ange (VEGASON 63) PPH, 1.4435
- Housing plastic PBT (Polyester), Alu-die casting
powder-coated, 316L (1.4435)
- Seal ring between housing and hou-
sing cover
NBR (stainless steel housing), silicone
(Alu/plastic housing)
- Inspection window in housing cover
for PLICSCOM
Polycarbonate
- Ground terminal 316Ti/316L (1.4571/1.4435)
Weight
- VEGASON 61 and 62 1.8 4.0 kg (4.0 8.8 lbs), depending
on process tting and housing
- VEGASON 63 2.7 5.7 kg (6 12.6 lbs), depending
on process tting and housing
Output variable
Output signal 4 20 mA/HART
Resolution 1.6 A
Fault signal current output unchanged; 20.5 mA;
22 mA; <3.6 mA (adjustable)
Current limitation 22 mA
Load see load diagram in Power supply
Integration time (63 % of the input va-
riable)
0 999 s, adjustable
Fullled NAMUR recommendation NE 43
Output signal digital output signal, format acc. to
IEEE-754
Sensor address 126 (default setting)
Current value constantly 10 mA; 0.5 mA
Integration time (63 % of the input va-
riable)
0 999 s, adjustable
Rise time 500 ms (ti: 0 s, 0 100 %)
Output
- Signal digital output signal, Foundation Field-
bus protocol
- Physical layer acc. to IEC 61158-2
VEGASON 61 and 62
VEGASON 63
4 20 mA/HART
Probus PA
Foundation Fieldbus
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Technical data
20 Ultrasonic Level measurement in liquids
Channel Numbers
- Channel 1 Primary Value
- Channel 2 Secondary Value 1
- Channel 3 Secondary Value 2
Transmission rate 31.25 Kbit/s
Current value 10 mA, 0.5 mA
Integration time (63 % of the input va-
riable)
0 999 s, adjustable
Rise time 500 ms (ti: 0 s, 0 100 %)
Input variable
Parameter distance between lower edge of the
transducer and product surface
Dead zone
- VEGASON 61 0.25 m (0.8 ft)
- VEGASON 62 0.4 m (1.3 ft)
- VEGASON 63 0.6 m (2 ft)
Measuring range
- VEGASON 61 up to 5 m(16.4 ft) liquid/up to 2 m(6.6 ft)
solid
- VEGASON 62 up to 8 m (26.2 ft) liquid/up to 3.5 m
(11.5 ft) solid
- VEGASON 63 upto15 m(49.2 ft) liquid/upto7 m(23 ft)
solid
Accuracy (similar to DIN EN 60770-1)
Reference conditions acc. to
DIN EN 61298-1
- Temperature 18 30C (64 86F)
- Relative humidity 45 75 %
- Atmospheric pressure 860 1060 mbar (86 106 kPa/
12.5 15.4 psi)
Characteristic curve deviation and measurement characteristics
1)
Average temperature coecient of the
zero signal (temperature error)
0.06 %/10 K
Resolution, general max. 1 mm
Ultrasonic frequency
- VEGASON 61 70 kHz
- VEGASON 62 54 kHz
- VEGASON 63 35 kHz
Interval >2 s (dependent on the parameter ad-
justment)
1)
Relating to the nominal range, incl. hysteresis and repeatability, determined acc. to the limit point
method.
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Technical data
Ultrasonic Level measurement in liquids 21
Beam angle
- VEGASON 61 and 62 5.5
- VEGASON 63 3
Adjustment time
2)
>3 s (dependent on the parameter ad-
justment)
Accuracy better than 0.2 % or 4 mm (see dia-
gram)
5 m 4 m 3 m 2 m 1 m
10 mm
4 mm
-4 mm
-10 mm
Fig. 24: Accuracy diagramVEGASON 61
Accuracy better than 0.2 % or 4 mm (see dia-
gram)
8 m 4 m 3 m 2 m 7 m 6 m 5 m 1 m
10 mm
16 mm
4 mm
-4 mm
-10 mm
-16 mm
Fig. 25: Accuracy diagramVEGASON 62
Accuracy better than 0.2 % or 6 mm (see dia-
gram)
2)
Time to output the correct level (with max. 10 % deviation) after a sudden level change.
VEGASON 61
VEGASON 62
VEGASON 63
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Technical data
22 Ultrasonic Level measurement in liquids
15 m 8 m 6 m 4 m 12 m 10 m 3 m
10 mm
20 mm
30 mm
6 mm
-6 mm
-10 mm
-20 mm
-30 mm
Fig. 26: Accuracy diagramVEGASON 63
Ambient conditions
Ambient, storage and transport tem-
perature
- without indicating and adjustment
module
-40 +80C (-40 +176F)
- the indicating/adjustment module -20 +70C (-4 +158F)
- Version IP 66/IP 68 1 bar with con-
nection cable PE
-20 +60C (-4 +140F)
Process conditions
Vessel pressure
- VEGASON 61, 62 -20 200 kPa (-0.2 2 bar)
- VEGASON 63 with compression
ange
-20 100 kPa (-0.2 1 bar)
- VEGASON 63 with mounting strap 0 kPa (0 bar), because no sealing pos-
sibility
Process temperature (transducer tem-
perature)
-40 +80C (-40 +176F)
Vibration resistance mechanical vibrations with 4 g and
5 100 Hz
3)
Electromechanical data version IP 66/IP 67 and IP 66/IP 68, 0.2 bar
Cable entry/plug
4)
- Single chamber housing l 1x cable entry M20x1.5 (cable-
5 9 mm), 1x blind stopper
M20x1.5
or:
l 1x closing cap NPT, 1x blind
stopper NPT
3)
Tested acc. to the regulations of German Lloyd, GL directive 2
4)
Depending on the version M12x1, acc. to DIN 43650, Harting, Amphenol-Tuchel, 7/8" FF; note plug
protection.
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Technical data
Ultrasonic Level measurement in liquids 23
or:
l 1x plug depending on the version, 1x
blind stopper M20x1.5
- Double chamber housing l 1x cable entry M20x1.5 (cable-
5 9 mm), 1x blind stopper
M20x1.5, plug M12x1 for VEGADIS
61 (option)
or:
l 1x closing cap NPT, 1x blind
stopper NPT, plug M12x1 for VE-
GADIS 61 (option)
or:
l 1x plug (depending on the version),
1x blind stopper M20x1.5, plug
M12x1 for VEGADIS 61 (option)
Spring-loaded terminals for wire cross sections up to 2.5 mm
Electromechanical data version IP 66/IP 68, 1 bar
Cable entry
- Single chamber housing l 1x IP 68 cable entry M20x1.5; 1x
blind stopper M20x1.5
or:
l 1x closing cap NPT, 1x blind
stopper NPT
- Double chamber housing l 1x IP 68 cable entry M20x1.5; 1x
blind stopper M20x1.5; plug M12x1
for VEGADIS 61 (option)
or:
l 1x closing cap NPT, 1x blind
stopper NPT, plug M12x1 for VE-
GADIS 61 (option)
Connection cable
- Conguration four cores, one suspension cable, one
breather capillary, screen braiding, foil,
mantle
- wire cross section 0.5 mm
- wire resistance <0.036 Ohm/m
- tensile load >1.200 N (270 pounds force)
- Standard length 5 m (16.4 ft)
- max. length 1000 m (3280 ft)
- Min. bending radius 25 mm (with 25C/77F)
- Diameter approx. 8 mm
- Colour standard PE Black
- Colour standard PUR Blue
- Colour Ex version Blue
Indicating and adjustment module
Power supply and data transmission through sensor via gold-plated sliding
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Technical data
24 Ultrasonic Level measurement in liquids
contacts (IC bus)
Indication LC display in full dot matrix
Adjustment elements 4 keys
Protection
- unassembled IP 20
- mountedinto the sensor without cover IP 40
Materials
- Housing ABS
- Inspection window Polyester foil
Supply voltage
Power supply two-wire instrument
- non-Ex instrument 14 36 V DC
- EEx ia instrument 14 30 V DC
Permissible residual ripple
- <100 Hz U
ss
<1 V
- 100 Hz 10 kHz U
ss
<10 mV
Load see diagram
1000
750
500
250
14 18 16 20 22 24 26 28 30 32 34 36

V
4
1
2
3
Fig. 27: Voltage diagram
1 HART load
2 Voltage limit EEx ia instrument
3 Voltage limit non-Ex instrument
4 Supply voltage
Supply voltage four-wire instrument 2072VDC, 20253VAC, 50/60Hz
Power consumption four-wire instru-
ment
max. 4 VA; max. 2.1 W
Supply voltage
- non-Ex instrument 9 32 V DC
- EEx ia instrument 9 24 V DC
Power supply by/max. number of sen-
sors
- DP/PA segment coupler max. 32 (max. 10 with Ex)
- VEGALOG 571 EP card max. 15 (max. 10 with Ex)
Supply voltage
- non-Ex instrument 9 32 V DC
- EEx ia instrument 9 24 V DC
4 20 mA/HART
Probus PA
Foundation Fieldbus
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Technical data
Ultrasonic Level measurement in liquids 25
Power supply by/max. number of sen-
sors
- H1 Fieldbus cable/Power supply max. 32 (max. 10 with Ex)
Electrical protective measures
Protection
- Plastic housing IP 66/IP 67
- Double chamber Alu-housing, four-
wire instruments
IP 66/IP 67
- Alu and stainless steel housing, two-
wire instruments
IP 66/IP 68 (0.2 bar)
5)
- Alu and stainless steel housing optio-
nal, two-wire instruments
IP 66/IP 68 (1 bar)
Overvoltage category III
Protection class
- two-wire, Probus PA, Foundation
Fieldbus
II
- four-wire I
Approvals VEGASON 61 and 62
6)
ATEX ATEX II 1G, 1/2G, 2G EEx ia IIC T6
FM FMCl.I-III, Div 1 (IS)
Ship approvals GL, LRS, ABS, CCS, RINA, DNV
Approvals VEGASON 63
7)
Ship approvals GL, LRS, ABS, CCS, RINA, DNV
CE conformity
EMC (89/336/EWG) Emission EN 61326: 1997 (class A),
susceptibility EN61326: 1997/A1: 1998
NSR (73/23/EWG) EN 61010-1: 2001
Environmental instructions
VEGA environment management sys-
tem
8)
certied acc. to DIN EN ISO 14001
5)
Requirement to maintain the protection is the suitable cable.
6)
Deviating data with Ex applications: see separate safety instructions.
7)
Deviating data with Ex applications: see separate safety instructions.
8)
You will nd detailed information under www.vega.com. 2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Dimensions
26 Ultrasonic Level measurement in liquids
7 Dimensions
Housing in protection IP 66/IP 67 and IP 66/IP 68, 0.2 bar
1
1
2
m
m

(
4
1
3
/
3
2
"
)
1
1
7
m
m

(
4

3
9
1
1
4
m
m

(
4

3
1
/
6
4
"
)
1
2
0
m
m

(
4
2
3
/
3
2
"
)
~ 69mm
(2 23/32")
77mm
(3 1/32")
~ 69mm
(2 23/32")
~ 116mm (4 9/16")
~ 87mm (3 27/64")
77mm
84mm (3 5/16")
84mm
/16")
20x1,5
NPT NPT
M20x1,5/
NPT 1 2 3 4
Fig. 28: Housing versions in protection IP 66/IP 67 and IP 66/IP 68, 0.2 bar (with integrated indicating and
adjustment module the housing height increases by 9 mm/0.35 in)
1 Plastic housing
2 Stainless steel housing
3 Aluminium double chamber housing
4 Aluminium housing
Housing in protection IP 66/IP 68, 1 bar
1
1
7
m
m

(
4

3
9
/
6
4
"
)
1
1
4
m
m

(
4

3
1
/
6
4
"
)
1
2
0
m
m

(
4

2
3
/
3
2
"
)
~ 103mm
(4 1/16")
~ 150mm (5 29/32")
~ 105mm (4 9/64")
77mm
(3 1/32")
84mm (3 5/16")
84mm
(3 5/16")
M20x1,5 M20x1,5 M20x1,5
M20x1,5/
NPT
M16x1,5
1 2 3
Fig. 29: Housingversions inprotectionIP66/IP68, 1 bar (withintegratedindicatingandadjustment module
the housing height increases by 9 mm/0.35 in)
1 Stainless steel housing
2 Aluminium double chamber housing
3 Aluminium housing
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Dimensions
Ultrasonic Level measurement in liquids 27
VEGASON 61
1 2
1
5
5
m
m

(
6

7
/
6
4
"
)
2
2
m
m

(
5
5
/
6
4
"
)
6
6
m
m
(
2

1
9
/
3
2
"
)
39mm
(1 17/32")
74mm
(2 58/64")
60mm
(2 23/64")
G1A /
1"NPT
Fig. 30: VEGASON 61
1 Dead band: 0.25 m (0.8 ft)
2 Meas. range: in liquids up to 5 m (16.4 ft), in solids up to 2 m (6.6 ft)
VEGASON 62
1 2
1
5
3
m
m

(
6

1
/
3
2
"
)
2
0
m
m

(
2
5
/
3
2
"
)
6
3
m
m
(
2

3
1
/
6
4
"
)
50mm
(1 31/32")
74mm
(2 58/64")
60mm
(2 23/64")
G2A /
2"NPT
Fig. 31: VEGASON 62
1 Dead zone: 0.4 m (1.3 ft)
2 Meas. range: in liquids up to 8 m (26.2 ft), in solids up to 3,5 m (11.5 ft)
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Dimensions
28 Ultrasonic Level measurement in liquids
VEGASON 63
3 4
1
1
8
m
m

(
4

4
1
/
6
4
"
)
1
9
5
m
m

(
7

4
3
/
6
4
"
)
148mm (5 53/64")
130mm (5 1/8")
158mm (6 7/32")
118mm (4 41/64")
M8x12
1
2 DN100/ANSI4"
Fig. 32: VEGASON 63
1 Mounting strap
2 Compression ange
3 Dead zone: 0.6 m (2 ft)
4 Meas. range: in liquids up to 15 m (49.2 ft), in solids up to 7 m (23 ft)
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Product code
Ultrasonic Level measurement in liquids 29
8 Product code
VEGASON 61
Approval
XX without
XM Ship approval
CX ATEX II 1G, 1/2G, 2G EEx ia IIC T6
CM ATEX II 1G, 1/2G, 2G EEx ia IIC T6 + Ship approval
Version / Process temperature
A Standard / -40...80C
Process connection / Material
G Thread G1A / PVDF
N Thread 1"NPT / PVDF
Electronics
H Two wire 4...20mA/HART
V Four wire 4...20mA/HART
1)
P Profibus PA
F Foundation Fieldbus
Housing / Protection
K Plastic / IP66/IP67
A Aluminium / IP66/IP68 (0.2 bar)
D Aluminium double chamber / IP66/IP68 (0.2 bar)
V Stainless steel 316L / IP66/IP68 (0.2bar)
Cable entry / Plug connection
M M20x1.5 / without
N NPT / without
Indicating/adjustment module (PLICSCOM)
X without
A top mounted
SN61.
1)
Only in conjunction with Housing / Protection "D"
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Product code
30 Ultrasonic Level measurement in liquids
VEGASON 62
Approval
XX without
XM Ship approval
CX ATEX II 1G, 1/2G, 2G EEx ia IIC T6
CM ATEX II 1G, 1/2G, 2G EEx ia IIC T6 + Ship approval
Version / Process temperature
A Standard / -40...80C
Process connection / Material
G Thread G2A / PVDF
N Thread 2"NPT / PVDF
Electronics
H Two wire 4...20mA/HART
V Four wire 4...20mA/HART
1)
P Profibus PA
F Foundation Fieldbus
Housing / Protection
K Plastic / IP66/IP67
A Aluminium / IP66/IP68 (0.2 bar)
D Aluminium double chamber / IP66/IP68 (0.2 bar)
V Stainless steel 316L / IP66/IP68 (0.2bar)
Cable entry / Plug connection
M M20x1.5 / without
N NPT / without
Indicating/adjustment module (PLICSCOM)
X without
A top mounted
SN62.
1)
Only in conjunction with Housing / Protection "D"
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
Ultrasonic Level measurement in liquids 31
VEGASON 63
Approval
XX without
XM Ship approval
Version / Process temperature
A Standard / -40...80C
Process connection / Material
X without compression flange
C Mounting loop / 1.4301
B Compression flange DN100PN16 / 316L
A Compression flange DN100PN16 / PPH
E Compression flange ANSI 4" 150psi / 316L
D Compression flange ANSI 4" 150psi / PPH
Electronics
H Two wire 4...20mA/HART
V Four wire 4...20mA/HART
1)
P Profibus PA
F Foundation Fieldbus
Housing / Protection
K Plastic / IP66/IP67
A Aluminium / IP66/IP68 (0.2 bar)
D Aluminium double chamber / IP66/IP68 (0.2 bar)
V Stainless steel 316L / IP66/IP68 (0.2bar)
Cable entry / Plug connection
M M20x1.5 / without
N NPT / without
Indicating/adjustment module (PLICSCOM)
X without
A top mounted
SN63.
1)
Only in conjunction with Housing / Protection "D"
2
9
0
2
3
-
E
N
-
0
5
0
8
1
0
VEGA Grieshaber KG
Am Hohenstein 113
77761 Schiltach
Germany
Phone +49 7836 50-0
Fax +49 7836 50-201
E-Mail: info@de.vega.com
www.vega.com
You can nd at www.vega.com
downloads of the following
l operating instructions manuals
l menu schematics
l software
l certicates
l approvals
and much, much more
29023-EN-050810