also twisted the two wiresin the stem to cancel the net flux passing between th
em.
The loop is sensitive to both electric and magnetic fields, but we are interested
only in the magnetic field response. A
balun is used to remove the signal caused by
the changing electric field.Following Roleson(l), we used #24 enameled wire
bifilar wound seven times around a small ferrite core.
Use
The shielded loop antenna is simple to use. Drive the coil to be tested with a
cw or pulse signal at the frequency of interest. Attach the shielded loop antenn
a to a
high impedance input of anoscilloscope and make measurements of the relativ
e
field strength and direction by comparing the magnitudes of the rf that is bein
g
detected by the probe. With our particular birdcage coil, wesee 20 volts peakto- peak into the scope for an effective B
field of 0.175 gauss which gives rise to a
proton /2 pulse of about 350 s. With this high sensitivity, it is easy to test a
coil which isexcited by a
weak cw source. This is convenient for us because the high
power transmitter of the NMR apparatus is often not available when the coil is
being developed. A
loop to be used withhigh power pulses can be smaller and thus
have better spatial resolution.
One test of how well this loop works for measuring the rf magnetic field is to
rotate it in the rf coil to be tested. A maximum in the magnitude of the rf
detected corresponds to an orientation withthe plane of the probe cutting the
most field lines, whereas a
minimum corresponds to an orientation with the loop cutting
no field lines. If the loop-coil system is working well, this minimumwill be a
zero, but if the electric field is not cancelled fully, there will be a
residual signal at minimum.
1. Scott Roleson, "Evaluate EMI Reduction Schemes With Shielded-Loop Ant
ennas, EDN, 203-207; May 17, 1984.