We were asked to obtain the FFT spectrum of this signal, measure the
frequency and the amplitude of the fundamental frequency and the first
four harmonics.
Firstly, this is the FFT spectrum of the signal:
Second harmonic at 4.96 Khz, third harmonic at 6.96 KHz and fourth
harmonic at 8.96 KHz
Here are the hardcopies that weve taken of the FFT spectrum with the
corresponding measurements.
Evaluation:
1. For frequency domain measurements, the frequency scale needs
to be expanded in order to accurately measure the frequency
components. This could be done with the time base (sec/div)
control. What is the effect of doing this on the measured
bandwidth? Information can be found in reference [6] and [7].
In the oscilloscope, when we change the time base (sec/div) control, we
change the sample interval. The sample interval is related to the sample
rate or frequency by this formula:
fs = record time/time interval = record time/ d.timebase
In order to represent the signal accurately, the sample frequency should
be at least twice the highest frequency. Therefore, the bandwidth and
the time base control is inversely proportional to the bandwidth. The
higher the time base, the lower the bandwidth displayed.
2. Use the hardcopies taken to discuss the effect of changing the duty
cycle on the FFT results.
Here is the frequency spectrum of the same signal, the second one being
at 20 % duty cycle:
As we can see from the comparison of the two graphs, in the 20% duty
cycle some of the frequencies dont have an amplitude, as they tend to
have in the original signal. This is because in the 20% duty cycle, the
signal is active only in 20% of the time.
3. What is the effect of the DC offset in problem 4.3.2.3 if you look at
the hard- copy?
Unfortunately the hard copies we posses do not clearly visualize the 0
Hz frequency, but the effect of the offset is that we have another
component in the FFT spectrum which would represent the DC offset of
the signal at 0 Hz frequency.
1.
500 Hz
980 Hz
2 kHz
3.6 kHz
As mentioned in the pre-lab the frequencies are 500 Hz, 1000 Hz, 2000
Hz and 3600 Hz. We can see that only one of the frequencies slightly
differs by just 20 Hz. But this could be due to the cursor placement in
the oscilloscope. So we can say that there is no actual difference
between the calculated results and measured results in this experiment.