Académique Documents
Professionnel Documents
Culture Documents
DOI: 10.1111/ajo.12081
Original Article
Background: Lactate measurements have become increasingly preferred over pH analysis in the evaluation of fetal
acidaemia in labour. In a busy labour ward, often the umbilical cord may be sampled late and as a result yield unreliable
lactate values.
Aim: To investigate the agreement of hand-held device Lactate Pro with a reference method blood gas analyser and
evaluate the stability of umbilical cord lactate values over time.
Methods: Prospective study carried out at elective caesarean section. Sixteen umbilical cords were double clamped
immediately after delivery with paired arterial and venous blood samples collected by an independent researcher, at
varying time intervals, and processed by two Lactate Pro devices and a reference method blood gas analyser.
Results: A signicant difference of 0.41 to 0.10 mmol/L was found when different groups of Lactate Pro devices were
compared with blood gas analyser at lactate values up to 5.70 mmol/L, with average lactate value of 2.45 mmol/L. Over
time, there is progressive rise in lactate samples obtained from the umbilical cord.
Conclusion: Lactate Pro devices have a signicant difference, but when used in clinical practice on cord blood after
delivery, this is unlikely to be meaningful. In intrapartum fetal surveillance, a systematic overestimation might lead to
unnecessary intervention. It is possible to retrospectively predict the likely level of lactate at birth in delayed cord samples.
Key words: agreement, blood gas analyser, comparison, lactate, umbilical cord blood.
Introduction
Fetal lactate analysis is increasingly being used as a tool to
detect metabolic acidaemia and the decision for urgent
delivery in the presence of a nonreassuring cardiotocogram
(CTG) in birth units worldwide. Lactate values have been
found to correlate with, and are comparable to, fetal pH,
base decit and perinatal outcome.1 It has also been shown
that fetal lactate has better predictive value than pH for Apgar
score <4 at ve minutes as well as in relation to moderateto-severe hypoxicischaemic encephalopathy in scalp
samples.2 In addition, pH analysis has been demonstrated
to be associated with more intrapartum sampling failure
(1120%) compared with lactate analysis.3,4 As a result,
lactate measurements have become an attractive alternative
375
T. Y. Su et al.
Protocol
Each umbilical cord was double clamped for the longest
length available immediately after delivery, placed in a
kidney dish and passed to the independent researcher for
sampling of all cords. Outside the operating room,
arterial samples were obtained rst followed by venous
samples using 23 gauge needles into prelabelled
preheparinised syringes kept on ice to be used as
control and then analysed at 5 to 20-minute intervals
up to 60 minutes. At the same time, umbilical cord
arterial and venous samples were collected via 23 gauge
needles, and also analysed at 5 to 20-minute intervals to
60 minutes. The maximum number of samples able to
be obtained per umbilical cord was processed at time
intervals as quickly as allowable by the same blood gas
analyser and Lactate Pro device. The rst samples were
collected from within 1 to 10 minutes of birth.
Statistical analysis
Results were analysed by paired t-tests. Values are
presented as mean differences of the groups compared
and standard deviation (SD), with P-value of <0.05
considered signicant. BlandAltman graphs were plotted
with mean difference of the group on Y-axis against the
mean lactate value on X-axis. Linear mixed-effects model
with SPSS software was used in the second part of the
study with calculation of change from initial values over
time.
As this study conforms to the standards established by
the NHMRC23 for ethical quality review, ethics approval
was not sought.
Results
Part one: agreement study
A total number of 479 lactate values were obtained from
the Lactate Pro devices and blood gas analysers. However,
17 results were excluded as these samples were not
processed by a Lactate Pro device and a blood gas
analyser at the same time to allow valid comparison. This
left 462 lactate values or 231 paired values for analysis in
groups for comparison as listed in Table 1.
A summary of the comparison groups and their mean
differences with limits of agreement is presented in
Table 2.
From Table 2, it can be seen there is a mean difference
of 0.10 mmol/L between the two hand-held Lactate Pro
devices, with a mean lactate value of 2.53 mmol/L. There
is also a consistent difference in LPro1 and ABG1 of
0.41, which is also signicant, whereby the Lactate Pro
device has consistently lower values of lactate compared
with blood gas analyser 1. On the other hand, LPro1
Equipment
2
3
4
5
6
Devices compared
No. of samples
48
62
123
7
65
185
Table 2 Mean differences with limits of agreement of the lactate groups compared
Groups for
comparison
LPro1
LPro1
LPro1
LPro2
LPro2
LPro1
LPro2
ABG1
ABG2
ABG1
ABG2
ABG (1 + 2)
Mean difference
(mmol/L)
0.10
0.41
0.01
0.14
0.01
0.13
SD
P-value
0.26
0.26
0.34
0.20
0.25
0.37
48
62
123
7
65
185
0.010
<0.001
0.747
0.106
0.884
<0.001
Discussion
The Lactate Pro meter is a convenient and versatile tool
currently used by many institutions to assess fetal acidosis
in the presence of a nonreassuring intrapartum CTG.24 In
our study, we have compared the Lactate Pro hand-held
devices with a blood gas analyser (ABL735) to assess the
Limits of
agreement
0.42
0.93
0.67
0.54
0.50
0.87
0.62
0.11
0.69
0.26
0.50
0.61
Mean of the
groups
Mean lactate
(mmol/L)
SD
(LPro1,LPro2)
(LPro1,ABG1)
(LPro1,ABG2)
(LPro2,ABG1)
(LPro2,ABG2)
(LPro1, ABG1 + 2)
2.53
2.47
2.53
2.36
2.31
2.51
0.79
0.64
0.88
0.52
0.80
0.81
377
T. Y. Su et al.
(a)
(b)
(c)
(d)
(e)
Figure 1 (a) The two Lactate Pro devices are compared with each other, with the differences of LPro1 to LPro2 plotted against the
mean lactate values of the two devices. Broad dashed line is the mean difference of 0.10 mmol/L, with ne dashed lines being limits of
agreement from 0.42 to 0.62. (b) Lactate Pro (LPro1) is compared with blood gas analyser (ABG1), with differences plotted against
mean lactate values. Broad dashed line is the mean difference of 0.41 mmol/L, with ne dashed lines being limits of agreement of 0.93
to 0.11. (c) Lactate Pro (LPro1) is compared with blood gas analyser (ABG2), with differences plotted again mean lactate values. Broad
dashed line is the mean difference of 0.01 mmol/L, with ne dashed lines being limits of agreement of 0.67 to 0.69. (d) Lactate Pro
(LPro2) is compared with blood gas analyser (ABG2), with differences plotted again mean lactate values. Broad dashed line is the mean
difference of 0.01 mmol/L, with ne dashed lines being limits of agreement of 0.50 to 0.50. (e) Lactate Pro (LPro1) is compared with
both blood gas analyser results combined (ABG1 + 2), with differences plotted against mean lactate values. Broad dashed line is the mean
difference of 0.13 mmol/L, with ne dashed lines being limits of agreement of 0.87 to 0.61.
378
mmol/L
Type: Cord
10
20
30
40
50
Minutes
60
70
80
mmol/L
Type: Cord
10
20
30
40
50
Minutes
60
70
80
Acknowledgement
We wish to thank Karen Byth statistician at Westmead
Hospital for her invaluable contribution to our data analysis.
379
T. Y. Su et al.
References
1 Westgren M, Divon M, Horal M et al. Routine measurements
of umbilical artery lactate levels in the prediction of perinatal
outcome. Am J Obstet Gynecol 1995; 173: 14161422.
2 Kruger K, Hallberg B, Blennow M et al. Predictive value of
fetal scalp blood lactate concentration and pH as markers of
neurologic disability. Am J Obstet Gynecol 1999; 181: 1072
1078.
3 Tuffnell D, Haw WL, Wilkinson K. How long does a fetal scalp
blood sample take? BJOG 2006; 113: 332334.
4 Westgren M, Kruger K, Ek S et al. Lactate compared with
pH analysis at fetal scalp blood sampling: a prospective
randomised study. Br J Obstet Gynaecol 1998; 105: 2933.
5 Saling E, Schneider D. Biochemical supervision of the foetus
during labour. J Obstet Gynaecol Br Commonw 1967; 74: 799
811.
6 Wiberg-Itzel E, Lipponer C, Norman M et al. Determination
of pH or lactate in fetal scalp blood in management of
intrapartum fetal distress: randomised controlled multicentre
trial. BMJ 2008; 336: 12841287.
7 Saugstad O. Is lactate a reliable indicator of tissue hypoxia in
the neonatal period? Acta Paediatr 2002; 91: 1719.
8 Ridenour R, Gada R, Brost B, Karon B. Comparison and
validation of point of care lactate meters as a replacement for
fetal pH measurement. Clin Biochem 2008; 41: 14611465.
9 Ramanah R, Martin A, Riethmuller D et al. Value of fetal
scalp lactate sampling during labour: a comparative study with
scalp pH. Gynecol Obstet Fertil 2005; 33: 107112.
10 Nordstrom L. Fetal scalp and cord blood lactate. Best Pract
Res Clin Obstet Gynaecol 2004; 18: 467476.
11 Westgren M, Kublickas M, Kruger K. Role of lactate
measurements during labor. Obstet Gynecol Surv 1999; 54: 4348.
12 Pennell C, Tracy M. A new method for rapid measurement of
lactate in fetal and neonatal blood. Aust N Z J Obstet Gynaecol
1999; 39: 227233.
13 Nordstrom L, Chua S, Roy A, Arulkumaran S. Quality
assessment of two lactate test strip methods suitable for
obstetric use. J Perinat Med 1998; 26 (2): 8388.
14 Pyne D, Boston T, Martin D, Logan A. Evaluation of the
lactate pro blood lactate analyser. Eur J Appl Physiol 2000; 82:
112116.
380