Vous êtes sur la page 1sur 29

HydroCOM

Service & Configuration Software 4.1


User Manual
Version 06/A

makes compressor control


and monitoring
an easy job

Contents
HSS english User Manual 06A.doc / KLS

2012-03-29

Introduction..........................................................................................................3

Installation............................................................................................................4

2.1

Minimum System Requirements ................................................................................................4

2.2

Set Up ...........................................................................................................................................4

The HSS User interface .......................................................................................5

3.1
Menu Bar ......................................................................................................................................5
3.1.1
File..........................................................................................................................................5
3.1.2
Settings...................................................................................................................................6
3.1.3
History ....................................................................................................................................7
3.1.4
Help ........................................................................................................................................7
3.2

Status Bar.....................................................................................................................................7

3.3

Task Bar........................................................................................................................................7

Analysis................................................................................................................8

4.1

Task Bar........................................................................................................................................8

4.2

Going On Line ..............................................................................................................................9

4.3

Get Parameters Upload Configuration from CIU...................................................................9

4.4

View CIU .......................................................................................................................................9

4.5

View Actuator............................................................................................................................ 10

4.6

View FTIM .................................................................................................................................. 12

4.7

View FTIM Detail ....................................................................................................................... 13

4.8

Import CSV-File in MS-Excel ................................................................................................... 15

4.9

Print............................................................................................................................................ 17

4.10

History-Function ................................................................................................................... 18

Service................................................................................................................22

5.1
Sensor calibration .................................................................................................................... 22
5.1.1
Hydraulic pressure .............................................................................................................. 22
5.1.2
Rod-drop.............................................................................................................................. 25
5.1.3
Indicator pressure................................................................................................................ 26
5.2

Set Parameters Download Configuration to CIU................................................................ 27

5.3

CIU Time and Operating Hours ............................................................................................... 28

HSS english User Manual 06A.doc / KLS

2012-03-29

Introduction
The HydroCOM Service & Configuration Software, HSS, is used for data and
status display, service and configuration of following HOERBIGER products:

HydroCOM
RecipCOM (On-line compressor condition monitoring)
ROD-Position Analyzer

The software has three access levels:

Analysis
Service
Configuration

The Analysis level is used for data and status display. At this level user failures do
never cause system failures.
Service is only applicable for those users being in charge of service and
maintenance. In addition to Analysis the necessary functions are available in order
to calibrate sensors and to download the configuration file to the CIU.
Warning!
If the wrong file is downloaded to the CIU the compressor may trip and the
equipment be damaged.
The Configuration level is dedicated for HOERBIGER Service Engineers or
advanced users only. In addition to Service and Analysis you have access to
configure the system.
Warning!
Wrong configuration may cause system failure, compressor trip and even
equipment damages.
Passwords are necessary for logon to Service and Configuration.
The manual is built up in that way, that it is not necessary for all users to work
through the complete document. In the below diagram you find the manual
sections that are applicable for the access level you have.
Section
Analysis-User
Service-User
Configuration-User

Appendix

The menus and windows described herein might deviate in details from the
supplied software.

HSS english User Manual 06A.doc / KLS

2012-03-29

Installation

2.1

Minimum System Requirements


The computer where HSS is going to be installed shall meet or exceed the
following requirements:

2.2

Operation system

Windows 98
Windows NT 4.0 (> Service Pack 4)
Windows 2000
Windows XP

Processor:

Pentium ( 233 MHz)

Interfaces

RS232

RAM

64 MB (128 MB recommended)

Free HD capacity:

4 MB

Screen:

800 x 600 pixels

Set Up
1)

Create a new directory e.g. c:\programs\HSS

2)

Insert the HOERBIGER CD-ROM into your CD-ROM drive

3)

Start the Windows Explorer

4)

Copy all files from the CD to c:\programs\HSS

5)

The complete HSS comprises of the following files:


HSS.DE
HSS.EN
HSS.exe
HSS.ini
HSSPrtUt.DE
HSSPrtUt.EN
HSSPrtUt.exe

6)

Mark up the file HSS.exe; click the right mouse button and select send to
desktop.

7)

Double click the HSS icon on your desktop to start the software.

HSS english User Manual 06A.doc / KLS

2012-03-29

The HSS User interface

3.1

Menu Bar
The screen is structured in the following way:
Menu bar

Worksheet

Task bar
Status bar

The menu bar contains the menus:


File
Open
Save (as)
New
Exit
Settings
Unit System
Connection
Language
Change Password
History
View
Save as *.txt
Help
About

3.1.1 File
Open
The files generated with HSS have the extension *.su4. These file do not only
contain configuration data, but also calibration data, HydroCOM status and
measurement values at the time when the file was saved.
Use Open for following tasks:
Rework existing configuration file
View configuration
View HydroCOM status and measurements being off-line; HydroCOM status
and measurements are shown from that time when the file was saved.
HSS english User Manual 06A.doc / KLS

2012-03-29

Save (as)
Use Save (as) for following tasks:
Saving of the current HydroCOM status and measurements.
Saving of configuration
Saving of calibration data
Note:
After sensor calibration the file must be saved. Otherwise calibration must
be repeated in case hardware components are exchanged.

New
Use New to start a new configuration.

Exit
Use Exit to exit HSS.

3.1.2 Settings
Unit System
Select the unit system before you start configuration. SI, German, English and
American units are available. The unit system is used for assigning measuring
ranges to the sensors during the configuration process. It is not possible to use the
unit system for changing the units for data display. Once units are configured they
can not be varied for data display.
Note:
If the Unit System is changed for an existing configuration all units are
deleted and have to be defined again. If RecipCOM is connected to the CIU
the units must not be changed.

Connection
Use Connection to assign the COM port you are using for communicating with the
CIU.
Available ports: COM 1 to 8
Communication between your PC/Laptop and the CIU is done via RS232. Each
RS232 interface has a COM port number defined by your operating system. In
most cases it is COM 1. The COM port number can be seen in the Windows
operating system. To check the number go to:
Control panel / system / hardware / device manager / connections

Language
You can choose between German and English. If you change the language exit the
HSS and start it again.

Change Password
Service and Configuration menu are password protected. Use Change Password if
you want to change from the default password to your own.

HSS english User Manual 06A.doc / KLS

2012-03-29

3.1.3 History
View
View of error and warning messages with a detailed date and time indication.

Save as *.txt
The error and warning messages can also be saved as *.txt file.

3.1.4 Help
About
Here you find the software version number (e.g. 4.20)

3.2

Status Bar
This bar has five status output fields.
Most left:
2nd from left:
3rd from left:
2nd from right:
Most right:

3.3

Filename or New
Current access level (analysis, service or
configuration)
Communication with CIU: on- /offline
Communication status: red flashing heart if you are
on-line with CIU
Description of item where mouse cursor is
positioned

Task Bar
The task bar is described in the following manual sections:

Analysis, 4
Service, 5
Configuration, 6 (Appendix)

HSS english User Manual 06A.doc / KLS

2012-03-29

Analysis

4.1

Task Bar
If you start the HSS you have already access to the Analysis level and the below
screen appears.

The task bar contains the buttons:

MainMenu
View CIU
View Actuator
View FTIM
View FTIM Detail
On/Offline
Print

HSS english User Manual 06A.doc / KLS

2012-03-29

4.2

4.3

Going On Line
1)

Select the correct COM port number and language in the menu Settings. If you
change the language restart the HSS.

2)

Click On-/Offline. If you are correctly connected the red heart is flashing.

Get Parameters Upload Configuration from CIU


Click Get Parameters. The HSS is loading the configuration and data from the CIU
to your PC.

4.4

View CIU
Click View CIU and following window opens (example):

HSS english User Manual 06A.doc / KLS

2012-03-29

The SIM status display shows the error codes. Click the buttons to scroll
through the list. The most recent error is on top of the list. The most previous error
is on the bottom of the list.
Click on the Erase button to clear the error list. Errors that are still active can not
be cleared.
The load signal is shown in the load signal output field.
If you move the mouse cursor on the error code the error clear text is displayed in
the status line of CIU window.
The GIM status display shows the status of the LEDs on the GIM hardware.
In addition operating hours, hours of HydroCOM control and speed are shown.
The IPS status display shows the status of the LED on the IPS hardwarePlease refer to the CIU operating instructions for detailed error/warning description
and troubleshooting.
Click on Simulation to start the HydroCOM electronic test run for all actuators. If
Simulation is on the HydroCOM operation is simulated at min. speed.
If you click the buttons SIM the following window opens

4.5

View Actuator
The same window (example) opens if you click View Actuator:

This window contains on the left the SIM status display, load signal and baud rate.
Actuator status, actuator temperature and hydraulic pressure (option) are
displayed for each actuator.
Click Clear Error to clear errors.

HSS english User Manual 06A.doc / KLS

10

2012-03-29

Click Simulation to start and stop the HydroCOM electronic test run. When
Simulation is on the button turns red.

Click the buttons on/off in the line Single Simul. in order to start the test run for
individual actuators. This test run mode is used to test the correct address of each
actuator. If Single Simulation is on the button turns to ON. If Single Simulation is off
the button turns to off.

Click on tabs SIM1, SIM2,..... to switch to actuators on other SIMs.

HSS english User Manual 06A.doc / KLS

11

2012-03-29

4.6

View FTIM
Click View FTIM to display TIM measurement or FTIM average values.
The TIM window (example) contains on the left the SIM status display, load signal
and baud rate.
TIM, sensor status and measurement values are shown for each TIM.

The Fast-TIM window (example) contains on the left the SIM status display, load
signal and baud rate.
Fast-TIM, sensor status and measurement values are shown for each Fast-TIM. If
high frequency measurement is enabled, the indicated value is the average value
of one revolution

Click on tabs SIM1, SIM2,..... to view TIMs/Fast-TIMs on other SIMs.

HSS english User Manual 06A.doc / KLS

12

2012-03-29

4.7

View FTIM Detail


In order of which set of parameters is configured (see also chapter 6.4.3
Configuration of new RPA functions) following values are displayed:
RPA
Clearance
Run-Out
Average @ Sensor
Trigger

Default
MIN
MAX
Average
Trigger

To view a complete set of measurements (profile) for each channel it is necessary


to configure Profile Enable.
Following window (example) opens:

If a channel is not configured on Profile Enable only one value is shown in the
column headed with Single measurement.
Click the button under View Profile to display the corresponding high frequency
measurements over one revolution.

HSS english User Manual 06A.doc / KLS

13

2012-03-29

Example:
Indicator pressure Fast-TIM 2.9; channel 1;
Tag Name: P-Cyl1-CE

Below the x-axis the Load Signal of the relevant stage and the time stamp of the
sampling are displayed.
Click the button Copy to clipboard to export the graph into a word processing
program.
Click the button Load Profile to view a new set of measurements.
Click the button Export to CSV-File to save the displayed profile to a MicrosoftExcel compatible CSV-File. How to get this CSV-File in MS-Excel and make a
diagram is described in the next chapter.
Click the checkbox Automatic CSV export after Load Profile to save the profile
(which can be refreshed by clicking Load Profile) additionally to the CSV-File. With
this function it is possible to save several profiles in one CSV-file.
The file name for the CSV-File is the same like the Tag-Name of the FTIM-channel.
So in our example this would be P-Cyl1-CE. This file is saved direct in the folder
Exports in the folder of the Hoerbiger Service Software HSS (Version 4.20 or
higher).
Zoom Graph
Fix the left window corner with a left mouse click and keep the button pressed.
Drag the button to fix the right corner and the window zooms in.

HSS english User Manual 06A.doc / KLS

14

2012-03-29

4.8

Import CSV-File in MS-Excel


If a profile is as saved as a CSV-File (described in chapter View F-Tim detail), then
this file is saved direct in the folder Exports (this folder is generated automatically)
in the folder of the Hoerbiger Service Software HSS (Version 4.20 or higher).
Following instruction shows the import in MS-Excel:
Step 1: Open File; File-Type: Text files

In row 1 the degree crank angle are displayed, at which the measurement is done
(250 times every 1,44 CA). This row should be configured as the X-Axis.
Starting with row 2 are the measurement (in bar absolute).
Column A to C can be used for labelling the legend.

HSS english User Manual 06A.doc / KLS

15

2012-03-29

Step 2: As diagram-type use the following

Step 3: Change scaling of x-axis

HSS english User Manual 06A.doc / KLS

16

2012-03-29

Example of a diagram in Microsoft-Excel


Profile Pressure Trend

9
19.07.2005 13:12:14 59,80%

19.07.2005 13:13:22 49,90%

Pressure [bar a]

7
6
5
4
3
2
1
0
0

90

180

270

360

Crank Angle []

4.9

Print
Click Print to set printer settings and print the configuration data.

HSS english User Manual 06A.doc / KLS

17

2012-03-29

4.10

History-Function
Click History in the menu bar to View or Save as *.txt File of error and warning
messages and the enable signal with a detailed date and time stamp.

There is also a time measurement in milliseconds called count, which gives the
possibility to have a chronological list of the warning and error messages.

This History file can also be saved as a *.txt file, which then can be inserted in
e.g. a Microsoft-Excel File.
Set Set the error or the warning
Clr Reset (Clear) the error or the warning

Following instruction shows an import of this *.txt file in Microsoft Excel.


HSS english User Manual 06A.doc / KLS

18

2012-03-29

Step 1: Open Import text file dialog

Step 2: Select *.txt - file

HSS english User Manual 06A.doc / KLS

19

2012-03-29

Step 3: Start Import with row 4

Step 4: Define column and press Finished

HSS english User Manual 06A.doc / KLS

20

2012-03-29

Step 5: Press OK in dialog box

Step 6: Example of an imported *txt file

HSS english User Manual 06A.doc / KLS

21

2012-03-29

Service

5.1

Sensor calibration
Warning!
Save configuration file after calibration in order to save the calibration
parameters. Otherwise a new calibration is required if components are
changed.

5.1.1 Hydraulic pressure


5.1.1.1 Actuator-4 operated as ACT4
Single point calibration is required for actuators of generation 4
1)

Go on-line as per 4.2

2)

Click Edit Actuator and following window opens

3)

Check if the threshold is set. If the threshold has not been set, define the
threshold and click Set Parameters to download to the CIU. Setting of the
threshold is only possible in the Configuration level.

4)

Click the button ... in the line Calibr. (bar) Offset right to. 0 @ 0. Following
window opens:

5)

Switch on the hydraulic unit.

HSS english User Manual 06A.doc / KLS

22

2012-03-29

6)

A minimum of 40bar actual pressure is required for calibration. Therefore


increase the pressure of the hydraulic unit at least to 40bar.

7)

Enter the actual pressure (reading from the pressure gage at the hydraulic
unit).

8)

Check the box use Online values

9)

Click the button Calibrate

10) Click the button OK


11) Now in the line Calibr. (bar) Offset following value is shown (example):
-6 @ 56. This means: The sensor was calibrated at a raw value of 56. The
offset is 6 that results into 50 bar actual pressure

HSS english User Manual 06A.doc / KLS

23

2012-03-29

5.1.1.2 Actuator-2
A two point calibration is required for actuators of generation 2.
1)

Go on-line as per 4.2

2)

Click Edit Actuator and following window opens (example)

3)

Check if the threshold is set. If the threshold has not been set, define the
threshold and click Set Parameters to download to the CIU. Threshold
setting is only possible in the Configuration level.

4)

Click the button ... in the line Calibr. (bar) Offset right to. 0 @ 182. Following
window opens (example):

5)

Switch off the hydraulic unit or open bypass valve

6)

Calibration for Low pressure:


Enter 0 bar for the actual pressure und click the Low button

HSS english User Manual 06A.doc / KLS

24

2012-03-29

7)

Switch on the hydraulic unit and adjust the pressure reducing valve for the
correct operating pressure

8)

Calibration for High pressure:


Enter the actual pressure (reading from the pressure gage at the hydraulic
unit) and click the High button.

9)

Click the OK button

10) Now in the line Calibr. (bar) Offset following value is shown (example):
0/45. This means: The sensor was calibrated at 0 and 45 bar

5.1.1.3 Actuator-4 operated as ACT2


Proceed as for Actuator-2. Please refer to 5.1.1.2.

5.1.2 Rod-drop
The calibration and initialisation of Rod-Drop Sensors is described in the separate
RPA calibration manual.

HSS english User Manual 06A.doc / KLS

25

2012-03-29

5.1.3 Indicator pressure


Indicator pressure sensors are already calibrated when delivered. Offset calibration
is not required for commissioning new sensors. However, for compensating
possible long term drift an offset calibration might be necessary later.
1)

Go on-line as described in 4.2

2)

Click Edit FTIM and following window opens (example)

3)

Click Calibr. Offset and following window opens (example)

The window contains


Check box for each measuring channel number: Select the channel you are
going to calibrate by checking the box.
TagName
Transmitter type
Reading
Unit
Current Adjustment
Input filed for correction value
Check box to freeze reading
Buttons for relative or absolute offset calibration
It is possible to use absolute or relative correction values.
Relative correction values:
Click Relative and the value from the correction value input field is added or
subtracted from the reading

HSS english User Manual 06A.doc / KLS

26

2012-03-29

Absolute correction values:


Click Absolute and the value from the correction value input field is taken as the
absolute value for calibration.

5.2

4)

Select the channel you are going to calibrate by checking the relevant box.

5)

Check the box Freeze Reading if the compressor is running or the readings
fluctuate

6)

Enter the correction value in the input sheet (example)

7)

Click the button Relative or Absolute, depending on the calibration you want to
apply. Following window opens, if Relative is selected (example).

8)

Click OK

Set Parameters Download Configuration to CIU


1)

Go on-line as described in 4.2.

2)

There must be no yellow background for an input field, before clicking the
button Set Parameters.

3)

Click the button Set Parameters;

4)

Following message appears:

If you click Yes, the configuration is going to be downloaded to the CIU and
overrides any configuration that is already on the CIU.
Warning!
If an input field is yellow before Set Parameters, this parameter is not
transferred to the CIU.

HSS english User Manual 06A.doc / KLS

27

2012-03-29

5)

5.3

If you click Yes following screen is shown; Watch the status bar right to Action.

CIU Time and Operating Hours


1)

Go on-line

2)

Click View CIU

3)

Click the button right to Hours of Operation

4)

The following window appears (example):

HSS english User Manual 06A.doc / KLS

28

2012-03-29

5)

Decide if you want to set Hours of Operation, Hours of Control or CIU-Time.


Only the inputs where the boxes are checked are considered
Example Hours of Operation:

Check the box CIU-Time;


Enter Time and Date;
Click Set button;
Example Hours of Operation / Hours of Control:

Check the boxes Hours of Operation and Hours of Control;


Enter the hours;
Click Set button;

HSS english User Manual 06A.doc / KLS

29

2012-03-29

Vous aimerez peut-être aussi