This document is part of the ASPO Service & Repair set for Dyna and describes the procedure
how to adjust, repair and calibrate the instruments.
Key Words: Pull-off-tester, Repair, Service, Calibration, Upgrade, Manual, ASPO
Document Number :
Document Revision:
Revision Date:
Document State:
Author:
345.13
01
31-Oct-2008
released
Rudolf Bhend
Company:
Proceq SA
Bahnstrasse 23
CH-8603 Schwerzenbach
Switzerland
Classification:
Doc: 345.13
Rev: 01
rudolf.bhend@proceq.com
Page 1/10
Working copy if printed
Table of Contents
1
2
3
3.1
3.2
3.3
3.4
3.5
3.6
3.7
Appendix 1
Appendix 2
Doc:
Rev: 01
Page 2/10
Working copy if printed
Doc: 345.13
Rev: 01
Page 3/10
Working copy if printed
2.
Data of the LEO1 manometer. From left to right: Instrument-ID. SW-version, pressure range, serial
number, status.
Check the pressure range only (100 bar)
2.
The unit set will be displayed, but only after the battery is changed (or disconnected and connected
again.
3.
Switch off time in minutes. The LEO1 will switch off automatically after the set time.
4.
Lock/unlock units
The units for the display of the hydraulic pressure can be locked or unlocked.
5.
Doc: 345.13
Rev: 01
Page 4/10
Working copy if printed
3.
3.1
Prerequisites
For the calibration of the DYNA Z instruments with the LEO1 a calibration device that meets the local
requirements and that can measure with a precision of at least 1% is necessary. For the record of the values
obtained during the calibration the respective Excel-files Certificate Record DYNA Z shall be used (see
Chapter 3.7).
3.2
Preparations
Place DYNA tester to be calibrated on the device and connect. Switch on the display unit of the
calibration device and perform the zero-setting.
Subject unit to full load (refer to the display of the calibration unit) by turning the handle of the pump
clockwise, remain load for 5 minutes, release load (turn handle of pump counter clockwise back to the
stop and turn maximum one turn clockwise). Reinstall the instrument without applying any load.
Insert the battery on the LEO1 and switch off the manometer (press Select twice).
Install the LEO-software according to points 1 to 5 explained in chapter 1(if not done yet). In case the software
is already installed, connect manometer with the RS232-cable to the PC, switch on the LEO1 and start the
software.
Continue with point 6 necessary settings as explained below:
Check pressure range on the display with the one on the manometer (100 bar).
Select under Allgemein (see point 2 in chapter 2) User 1 as unit.
Set 10 for 10 minutes under Ausschalten nach (see point 3 in chapter 2).
Remove all check marks in the fields under Einheiten freischalten(see point 4 in chapter 2), so no any
unit is released. By this the value under user 1 will be shown in the display of the LEO1 and not the
hydraulic pressure of the oil in the DYNA Z- instrument (this is of no interest for us and the client).
Insert calibration values as explained in the following chapters.
3.3
3.3.1
Insert in the first line under Kundeneinheit (costumer unit, see point 5 of chapter 2) the values for kN:
- Set the check mark left to the 1.
- Set the factor 0.12500 for the DYNA Z 6 and DYNA Z 6 S,
respectively 0.26100 for the DYNA Z 16 and DYNA Z 16S.
- Set the letters C_n_u_t for kN)
(The choice of letters and numbers is limited, e. g. no k, no N).
Press the key Schreiben (write).
Doc: 345.13
Rev: 01
4/9
Page 5/10
Working copy if printed
Close the window of the LEO1-software (click on x top right). Open the software again and check if the
correct values are shown in line 1.
If this is not the case repeat the procedure, i.e. set the corrected values, press the key Schreiben (write),
switch off and on again the software and check the values in line 1
In case the correct values are shown:
- Remove RS 232 cable from LEO1
- Switch to the unit Cnut in LEO1(see INFO-sheet addition to operation manual)
- Do not set the manometer to the Peak-mode
- Set the unit kN in the display of the calibration device
3.3.2
3.4
Connect the RS 232 cable to LEO1, switch off the manometer, switch off / on the software and switch on the
manometer
Insert in the second line under Kundeneinheit (costumer unit, see point 5 of chapter 2) the values for lbf:
- Set the check mark left to the 2.
- Calculate the factor for the display of the force in lbf. Multiply the factor obtained for the force in kN by
224.809 (in the record sheet this is done automatically).
- Set the numericals L_b_F_0 (the last one is a zero).
(The choice of numericals is limited, e. g. no l, no f).
For the pull-off testers DYNA Z 6 and DYNA Z 16 on line 3 the settings for the display of the strength on a
2
test disc diameter 50 mm in N/mm and on line 4 in psi must be installed as explained for line 1 and line 2
above.
2
For N/mm on line 3 the letters to be set are n_P_n_n. The factor for the force in kN must be divided by
1.9635 (in the record sheet this is done automatically).
For psi on line 4 the numericals to be set are P_S_J_0 (the last one is a zero). The factor for the force in kN
must be multiplied by 73.8672 (in the record sheet this is done automatically).
Additional units can be set as explained in chapter 3.5.
In case all the values in line 2 to 5 are set continue as follows:
- Press the key Schreiben (write).
- Close the window of the LEO1-software (click on x top right). Open the software again and check if the
correct values are shown in line 2.
Doc: 345.13
Rev: 01
Page 6/10
Working copy if printed
If this is not the case repeat the procedure, i.e. set the corrected values, press the key Schreiben (write),
switch off / on the software and check the values in line 2.
In case the correct values are shown:
- Remove RS 232 cable from LEO1.
- Switch to the unit LbF0 in LEO1 (see INFO-sheet addition to operation manual).
- Set the unit lbf (or Kip) in the display unit of the calibration device.
Check with the calibration device at 20%, 50% and 100% steps of the nominal load if the values of line 2 are
within the tolerances.
In case the values are not within the tolerances or are very close to the upper and/or lower limits:
- Check the factor for kN in the record sheet and if the factors in the software are set according to the one
calculated in the record sheet. Amend values, if necessary.
- Connect the RS 232 cable to LEO1, switch off the manometer, switch off / on the software and switch on
the manometer.
- Repeat procedure starting from line 3 Press the key Schreiben (write) above (except switching to the
unit LbF0 in LEO1, since this is done already).
In case the values are within the tolerances ( 2 % of the actual load) proceed with checking of the values in
lines 3, 4 and 5 as described above for line 2. Start with - Remove RS 232 cable from LEO1.
2
In the calibration device the corresponding units and if necessary, the test areas in mm must be set.
The measured and valid values of the 20%, 50% and 100% steps must be registered in the record (see
chapter 3.7 and Appendix 2).
3.5
On the pull-off testers DYNA Z 6 and DYNA Z 16 line 5 is free for an additional unit, on the extraction testers
DYNA Z 6 S and DYNA Z 1 6 S three additional units can be set in lines 3, 4 and 5 according to clients
request.
E.g. On the pull off tester in line 5 the values for the strength in psi on a test disc of diameter 100 mm may
be set. In this case the factor for the force in kN must be multiplied by 18.4668. For the display of the
strength in psi the numericals P_S_J_1 may be set.
On the extraction tester units and factors for the display of the shell friction of a specific dowel size may
be set.
2
In the calibration device the corresponding units and if necessary, the test areas in mm must be set.
The measured and valid values of the 20%, 50% and 100% steps must be registered in the record (see
chapter 3.7 and Appendix 2).
3.6
All instruments are calibrated in the unit kN. The respective conversion factor and unit are always set in line 1.
The calibration factors in italics are theoretical ones. They are slightly different for each instrument as to verify
during the calibration.
Pull-off Tester
DYNA Z 6
Calibration
Doc: 345.13
Official
Letter
sequence
Extraction Tester
0.125
DYNA Z
16S
DYNA Z 6S
Line 1
Calibration
Factor
DYNA Z 16
Units
Line 1
0.261
kN
Cnut
0.125
0.261
Line 2
Line 2
Rev: 01
Page 7/10
Working copy if printed
Factor
28.10075
Calibration
Factor
58.67515
lbf
LbF0
28.10075
Line 3
Line 3
0.063662
0.132962
N/mm
nPnn
0.0636618
Line 4
multiply with 73.8672*
Factor
9.2334
19.27934
---
50mm*
psi
PSJ0
---
Line 5
Factor
*) Factor in
line 1 to
--Line 5
0.132962
Line 4
Calibration
Calibration
58.67515
4.819835
---
100mm*
psi
PSJ1
---
---
Additional factors and units must be calculated, respectively introduced for other instruments like the DYNA Z
25, Z 50 and Z 100 and/or based on the requirements from the clientele.
3.7
Proceq SA has prepared calibration certificates and records. For each instrument a Excel file exists with one
folder for the record of the calibration values including the calibration factors of different units. The recorded
values are automatically inserted in the folder of the certificate. The files can also be used for the verification of
the values after new units have been set. An example each of the certificate and the record is shown in the
Appendix 1, respectively Appendix 2.
Doc: 345.13
Rev: 01
Page 8/10
Working copy if printed
Appendix 1
DYNA - Gerte
Appareils DYNA
DYNA Instruments
Kalibrier - Zertifikat
Certificat d'talonnage
Calibration Certificate
Prfobjekt
Objet talonner
Test piece
DYNA Z 6 LEO1
6.00
Prfkraft
Force d'essai
Test load
kN
Anzeige:
Affichage:
Display:
Prfeinrichtung
Dispositif d'talonnage
Calibration device
0.01
Serie-Nr.
Srie-No.
Serial No.
1-1152
Manometer Nr.
Manomtre No.
Manometer No.
SN 8486
kN
BM.2.ZK.57
AE.2WM.58
Ablesung am
LEO1
Lecture
LEO1
Displayed on
LEO1
[%]
kN
kN
kN (Cnut)
10
0.60
0.59-0.61
0.59
lbf (LbF0)
N/mm (nPnn)
psi (PSJ0)
90.0
88.2 - 91.8
0.59
89.1
20
1.20
1.18-1.22
1.19
280
274 - 286
30
1.80
1.76-1.84
1.79
278
40
2.40
2.35-2.45
2.39
50
3.00
2.94-3.06
3.00
700
686 - 714
1.50
1.47 -1 .53
225.0
220.5 - 229.5
60
3.60
3.53-3.67
3.59
697
1.49
224.2
70
4.20
4.12-4.28
4.18
80
4.80
4.70-4.90
4.79
0.0
0.0
450.0
441.0 - 459.0
90
5.40
5.29-5.51
5.39
1'400
1'372 -1 '428
100
6.00
5.88-6.12
6.00
1'400
3.00
449.4
0.0
1
0.12560
2
28.23601
3
0.0640
4
9.2777
5
0.00000
*)
Cnut
Schwerzenbach,
Proceq SA
ISO 9001
SQS-Reg. No.
20.07.2005
10956-04
Doc: 345.13
Rev: 01
Page 9/10
Working copy if printed
Appendix 2
Doc: 345.13
Rev: 01
Page 10/10