700XA Natural Gas Chromatograph

Product Data Sheet

NGC_PDS_700XA

April 2014

Danalyzer 700XA
Natural Gas Chromatograph
Danalyzer 700XA Gas Chromatographs provide
the most accurate analysis of natural gas
available in a field mounted gas chromatograph
(GC). The 700XA features a design that increases
analytical capability, maximizes ease of use, and
widens the range of analysis options in a GC with
an ambient temperature rating of -20 C to
60 C/ -4 F to 140 F. These enhanced features
make the 700XA ideal for natural gas custody
transfer and applications requiring advanced
analysis such as C9+ (with hydrocarbon dew
point and cricondentherm calculations) and C6+
with hydrogen sulfide (H2S).
Danalyzer 700XA Natural Gas Chromatograph

Features

Unmatched measurement performance


Best C6+ heating value/BTU repeatability available
/ -0.01 % ( / -0.1 BTU/1000 BTU) in temperature controlled
environment
/ 0.015 % ( / -0.15 BTU/1000 BTU) in uncontrolled
environment (-20 to 60 C/-4 to 140 F) with a three minute
cycle time
Best-in-industry C9+ repeatability available
0.0125 % of heating value (0.125 BTU/1000 BTU)
for controlled environment
0.025 % (0.25 BTU/1000 BTU) of heating value for
uncontrolled environment (-20 to 60 C/-4 to 140 F)
with a five minute cycle time
Wide dynamic range from percent to trace level components
down to 2 ppm
Reliable performance over broad ambient temperatures
-40 C to 60 C/-40 F to 140 F

One package for fiscal metering or gas quality


Custody transfer analysis from C6+ to C9+
Contaminant monitoring trace hydrogen sulfide, carbon
dioxide, oxygen, etc.
Combine measurements and reduce analysis cost

C9+ with hydrocarbon dew point


C6+ with H2S (3 to 30 ppm H2S)
C6+ with oxygen
C6+ with helium and hydrogen
C9+ with methanol and water

Reduced installation costs


Standard 24V DC power or optional 120/240V AC power
Integrated controller electronics
Pipe-mount, wall-mount, or floor-mount

Easy to use

Lower operation and maintenance costs

Single bolt analytical valve design for easy overhaul


in the field
Large column canister for both micro-packed and capillary
columns of long lengths
Internal stream selection and valve actuation solenoids that
can be replaced easily in under five minutes
Easy to use MON2020 software for diagnostics
Two portholes in the electronics housing for easy wiring
access

No shelter or instrument air required


Low carrier and power consumption
Longest gas chromatograph valve and column warranties
available in the market today

700XA Natural Gas Chromatograph

April 2014

Applications
Standard Natural Gas Applications

Hydrocarbon Dew Point Monitoring

Emerson Process Management has made popular end-user


energy and gas quality applications standard on all Danalyzer
gas chromatographs. Applications may vary by components
of interest, analysis time, reduced hardware, or improved
precision. For non-standard natural gas applications, Emerson
can custom-engineer the 700XA Gas Chromatograph to fit

The 700XA Gas Chromatograph offers accurate and reliable


hydrocarbon dew point calculations from the extended
C9+ analysis by combining two detectors and a controller
within a single housing reducing complexity, minimizing
maintenance and spare parts requirements, simplifying the
scope of analyzers at the pipeline, and reducing the overall
cost of the analytical solution.

most requirements.

Energy Measurement (to C6+ and C9+)


The 700XA Gas Chromatograph standard applications for
energy measurement include C6+ (three minutes), C7+
(ten minutes), and C9+ (five minutes). The latest GPA 2145/2172,
AGA-8, and ISO 6976 calculations are available and can be
configured to be calculated together.

Standard Measurement Ranges


Methane
Ethane
Propane
N-Butane
Iso-Butane
N-Pentane
Iso-Pentane
Neo-Pentane
Hexane C6+**
Nitrogen
Carbon Dioxide
Hexanes*
Heptanes*
Octanes*
Nonane+*
*

C6+ C9+

65 to 100 mole %
0 to 20 mole %
0 to 10 mole %
0 to 5 mole %
0 to 5 mole %
0 to 1 mole %
0 to 1 mole %
0 to 1 mole %
0 to 0.7 mole %
0 to 20 mole %
0 to 20 mole %
0 to 1 mole %
0 to 1 mole %
0 to 0.5 mole %
0 to 0.5 mole %

C9+ analysis only

** Not included in C9+ analysis results

The 700XA integrates hydrocarbon dew point software into


the gas chromatograph to provide dew point temperatures
for up to four user-entered pressures and the cricondentherm
using the Peng-Robinson or the Redlich-Kwong-Suave
equations of state. Real-time dew point results can be provided
by using analog or Modbus inputs from another device for the
calculation pressures.
The measured C6/C7/C8 and C9+ components allow for
an accurate determination of the hydrocarbon dew point
for pipeline-quality natural gas using reliable and lowmaintenance thermal conductivity detectors (TCDs), avoiding
standalone dew point analyzers or flame ionization detectors
(FIDs), which require additional utility gas requirements.
For heavier gas applications where significant amounts of
components above C10 are expected, an FID can be combined
with a TCD to provide for further extended analysis.

Custom Applications
If Emersons standard applications do not fit your needs, the
700XA Gas Chromatograph can be customized to meet many
measurement requirements. Simply submit a completed
application data sheet found at the end of this data sheet with
your request, or ask our Application Engineers for assistance.

Gas Quality Analysis


Natural gas contaminants, such as hydrogen sulfide and
oxygen, reduce pipeline integrity over time. Most contaminants
can be easily measured in the 700XA Gas Chromatograph
for online quality assurance. Contaminant monitoring can be
combined with energy measurements for complete custody
transfer analysis. To the extent possible, these combined
applications utilize independent gas chromatograph valves,
detectors, and columns for each primary measurement. This
technique offers greater reliability, increased speed, and easier
troubleshooting. This application approach also simplifies field
upgrades and re-applications in the 700XA Gas Chromatograph
by minimizing internal piping changes.

A typical phase curve showing the Cricondentherm and the


hydrocarbon dew point calculated at three pressures for practical
operational use.

www.Danalyzer.com

April 2014

700XA Natural Gas Chromatograph

Superior Performance
The Danalyzer Difference

Thermal Conductivity Detectors

Building off of the proven valves, columns, and detectors of the


Model 500 and Model 700 Gas Chromatographs, the 700XA
Gas Chromatograph analytical oven has been redesigned for
maximum serviceability and expandability. It features a new,
cleaner architecture with fewer cables, making the 700XA
simple to maintain. In addition, the oven features a unique,
pivot-top base that provides maximum accessibility to the
components below.

The thermal conductivity detector (TCD) is the detector of


choice for most applications due to its universal response to
all components of interest in natural gas and light refinery and
hydrocarbon processing gas analysis. The TCD found in the
700XA Gas Chromatograph is able to measure well beyond the
normal ranges seen in other designs and is sensitive enough
to perform many applications with low parts-per-million
measurement requirements. This greatly simplifies the gas
chromatograph design and lowers the cost to the end user

Multiple temperature control zones and up to four six-port


or ten-port valves and two independent detectors provide
extreme application flexibility and range. All components in the
oven are completely accessible and serviceable in the field to
reduce the total cost of ownership over the life of the analyzer.

when a simple and rugged TCD can be used.

Micro Flame Ionization Detector


(FID)
The micro-flame ionization detector, coupled with a
new detector pre-amplifier/electrometer board, permits
measurement of trace hydrocarbons in a variety of samples at
parts-per-billion (ppb) concentrations. The FID is unique in the
industry because of its small size (less than three inches high)
that fits inside the explosion-proof housing of the 700XA
Gas Chromatograph. Typical applications include measuring
trace impurities in gases and light hydrocarbons, as well as
ambient air monitoring.

Micro-packed Columns
Unique analytical assembly design pivots to allow instant access to
all components.

Gas Chromatograph Valves


The 700XA has the capacity to support up to four ten-port or
six-port diaphragm/piston gas chromatograph valves. These
pneumatic valves are guaranteed for the life of the gas
chromatograph and are specified to operate over five million
cycles. The unique, double-diaphragm design removes the need
for springs, o-rings, or lubrication. Valve service is performed by
replacing a cost-effective diaphragm set, which can normally be
completed in less than ten minutes.

Stream Switching Module


The internal sample stream switching module is available in four
or eight-stream versions. For applications with widely varying
stream composition, double-block-and-bleed configuration is

The 700XA Gas Chromatograph offers micro-packed columns


with a superior combination of features found in both capillary
and conventional packed columns speed, sharp peak
resolution, and low carrier gas consumption. In addition, the
unique design provides for greatly extended column life and
the longest warranty available on the market (five years on the
standard C6+ natural gas set). Standard capillary columns may
also be used in 700XA applications as required.

Six-port and ten-port XA valves build on


legendary performance of the
Model 500 six-port valve, with
reduced dead volume
and a single retention
bolt to simplify
maintenance.

optionally available.

www.Danalyzer.com

700XA Natural Gas Chromatograph

Controller Electronics and


Communications

April 2014

Modular Electronics

Data Archiving and Reports

The control electronics, option cards, and field termination


boards are all packaged conveniently in the lower section of the
700XA Gas Chromatograph. Customer-terminated power and
output connections are also made in this lower section of the

Includes expanded reports and data archiving that conforms to


the latest API 21.1 requirements for metering audit purposes
and back up of the primary systems such as flow computers or
SCADA systems. Every analysis is time and date stamped and
archived for retrieval via the MON2020 software.
Security four levels of password-protected security,
configurable to read/write or read-only for third-party
access.
Event logs a continuous record of all operator changes,
with time, date, and password-identified name records
Alarm logs a continuous record of all historical alarms,
time/date stamped with alarm state and description
Maintenance log a ''scratch pad'' for tracking maintenance
or testing performed on the gas chromatograph system
Archiving over 31,700 analysis records (which is over
65 days for a C6+ three minute application), 370 final
calibration records, 370 Validation records are archived
automatically with time and date stamps.
Chromatograms over 1,700 worth of analysis
chromatograms and 370 chromatograms and user selected
''Protected Chromatograms'' that are permanently stored,
including the factory testing chromatograms.
Drawings and Documents user manuals and drawings
in several file formats are stored in the controller
memory for convenient retrieval using the MON2020
software eliminated the risk of manuals and drawings
being misplaced. User generated documents (such as
maintenance checksheets or installation drawings) can also
be uploaded to the controller for later retrieval.

gas chromatograph.

Local Indication and Operation Panel


Analyzer health and valve status can be viewed through the
front cover of the gas chromatograph. The panel displays green
(healthy), yellow (warning), and red (failure) LEDs, along with
LEDs indicating gas chromatograph valve on/off actuations,
power, and CPU health. Each valve can be actuated manually
for simplified troubleshooting and fast system purging after
maintenance.

Touch Key Local Operator Interface


(Optional)
The 700XA local operator interface (LOI) permits maintenance
and operation of a 700XA without a laptop or PC. The LOI is
a high-resolution color display that is touch key infrared
activated and supports all core GC operations.
Features include:
Color LCD with full VGA (640 x 480 pixels) resolution
User adjustable auto-backlighting
Eight infrared-activated touch keys and screen saver
Indicates complete GC status, control, and diagnostics,
including full chromatogram display

Flexible I/O
The 700XA is built with enough I/O to handle most applications,
including:
Five discrete digital outputs
Five discrete digital inputs
Two analog inputs
Six analog outputs
Three Modbus serial ports (RS-232/RS-485/RS-422)
Two Modbus enabled Ethernet ports
(one with DHCP server for local access)
FOUNDATION Fieldbus (optional)
If there is a need for more I/O, the 700XA includes two
expansion slots that use the I/O cards from the ROC-800 family
of flow computers.

Standard reports include:


Average reports hourly, 24-hour, weekly, monthly
and variable averages.

Analysis reports physical property calculations for


component and group analysis and alarms.

Raw data report retention times, peak areas, detector


number, method, integration start/stop, and peak width for
the analysis.

Calibration report raw component data, new response


factors, retention times, and deviation from last calibration

Final calibration report results from the calibration


response factors and retention time adjustments.

Molecular Weight vs. Response Factor Graph the


response factors plotted on a log/log graph as outlined
in the GPA2198-03 Appendix B to confirm the fidelity of the
detector response across components.

www.Danalyzer.com

700XA Natural Gas Chromatograph

April 2014

MON2020 Software

The Danalyzer 700XA Gas Chromatograph is designed to


operate unattended. If adjustments are needed, our exclusive
MON2020 software allows complete control of the XA series of
gas chromatographs either locally or remotely.
From within MON2020, a user can:
JJ Start or stop analysis, calibration, or validation cycles
JJ Generate and save current and historical analysis and
calibration reports
JJ Review and modify analytical settings
JJ Upload and display multiple chromatograms for comparison
JJ Upload and trend any of the measured results
JJ Export data for use in other third-party applications in text,
HTML, or ExcelTM file formats
JJ Check original calibration against last calibration
JJ Perform GC operation checks and modifications
simultaneously
JJ Upload and view manuals and drawings stored in the gas
chromatograph controller
MON2020 is Windows-based software designed to make
analyzer configuration, maintenance, and data collection easy.
With intuitive drop-down menus and fill-in-the-blank tables,
even new users can quickly navigate through the software.
Users of previous-generation MON software will be familiar
with the layout and functionality of the software, and will be
impressed with the additional features that make the software
even easier to use.

With the ability to communicate to the enterprise network or


export to numerous file types, MON2020 is a powerful software
tool that ensures operators, engineers, maintenance personnel,
and management have access to critical data, such as current
and archived chromatograms, alarm history, event logs, and
maintenance logs.
The chromatogram viewer allows you to look and compare
chromatograms stored in a connected GC or offline disk
storage, and the currently running analysis. The small
chromatogram files (less than 100 KB in size) also include
analysis and calculation results, integration and valve timing
settings, retention time settings, and the raw peak data. The
small size and format of the chromatogram files (file extension
.xcgm) facilitate easy sharing of chromatograms over email.
The trend viewer makes it easy to trend multiple variables on
a single chart. To help diagnosing process or analysis issues,
single or multiple points on the trend can be selected and the
corresponding chromatogram(s) opened straight from the
trend viewer screen. The trends can be saved as MON2020 trend
files or exported as text, comma seperated values, or Microsoft
Excel files.
The MON2020 software connects to an XA series GC via
Ethernet (10/100 Mb) directly or over the local or wide area
network. A multiple level username/password security
implementation is used to limit and control access to a gas
chromatograph providing read only or up to four levels of write
enabled access.

Figure 1 - MON2020 Interface


Simple
Drop-down
Menus

Response Factor
Fidelity Chart

Connect to any GC
with a Mouse Click

Fully Detailed
Timed Event Table

Full Featured
Chromatogram
Display

Quickly Add
Chromatograms
to Overlay
Automatic Listing
of Measured
Components

Display Results

Save Chromatograms
to Hard Drive

www.Danalyzer.com

700XA Natural Gas Chromatograph

Integration with ThirdParty Networks

April 2014

FOUNDATION Fieldbus

Whether you want to network gas chromatographs throughout


your network or simply link a single gas chromatograph to a
flow computer, the 700XA can be configured to handle most
any scenario.
Choice of Ethernet, FOUNDATION fieldbus, Modbus Serial,
or 420 mA analog outputs
Uses industry standard non-proprietary communication

protocols that allows connection to other XA series or

legacy gas chromatographs over the same network.
Use MON2020 for diagnotics, configuration, and data

retrieval over Ethernet, Serial, or Modem connections.

Emerson's 1500XA and 700XA Gas Chromatographs are the


first and only gas chromatographs that are certified by the
Fieldbus Foundation. It is quickly becoming the industry
standard, and use of this protocol reduces the amount of
engineering necessary during the installation process, as it
doesnt require the manual point mapping of the Modbus
protocol. It also requires less wiring, fewer junction boxes, cable
trays, and I/O cards, which means a cleaner, simpler, easier to
understand analytical footprint as compared to traditional I/O
installations.

The 700XA Gas Chromatograph supports four types of


communication interfaces
10/100 Mbps Ethernet connectivity
A FOUNDATION fieldbus H1 interface
RS-232, RS-422, and RS485 serial communication links

420 mA analog outputs

The Modbus protocol is widely used today because it is simple


and effective. The 700XA can be configured to use the flow
computer industry standard SIM-2251 Modbus map and be
compatible with communication links from legacy Danalyzer
GC installations, or fully customized Modbus maps can be
configured using either single register per floating point
(ENRON Format) or two registers per floating point format used
in DCS and PLC systems.

Ethernet Connectivity
Two Ethernet interfaces are available on the 700XA Gas
Chromatograph. Each interface can be configured with a static
IP address, subnet mask, and gateway. The RJ45 connector
Ethernet port can be configured to operate as a DHCP host to
simplify local PC access. The Ethernet interfaces on the 700XA
serve two purposes serve MON2020 connections and serve
Modbus TCP requests. The dual Ethernet interfaces can be used
in many ways. Examples:
One to connect to a plant network for GC maintenance
personnel and the other to a control network using
Modbus TCP
One to a broadband cellular wireless gateway for remote
GC access for data collection and maintenance, and the
other for a local laptop connection

OPC
With the optional GC-OPC server, the 700XA can connect
via OPC with fully configurable definition files and remote
operation control capabilities.

Modbus Serial

Three hard-wired serial ports that can be configured as RS-232,


RS-485, or RS-422 links to communicate to host systems using
the Modbus protocol. Additionally, an additional fourth serial
port is configured for RS-232 with a nine-pin D-type connector
and can be used for direct connection to a Daniel ultrasonic
meter or local MON2020 access. The ultrasonic flow-meter
link enables the online calculation of the speed of sound for
continuous validation of the custody measurement. If more
serial links are required, up to two additional serial ports can be
added using the two expansion ports using ROC800 series I/O
cards.

420 mA Analog Outputs


The 700XA Gas Chromatograph supports six isolated 420 mA
analog outputs that can be expanded to 14 analog outputs with
optional expansion cards.

www.Danalyzer.com

April 2014

Custom-Engineered
Solutions

A complete online analytical solution is more than just the


analyzer. Sample conditioning systems to prepare the sample
for analysis, communication links to the plant control computer,
and packaging of the analytical equipment into a cabinet or
shelter all play an important role.
Emerson Process Management has decades of experience
providing complete turnkey solutions ranging from simple
single-analyzer cabinets up to large integrated shelters with
multiple types of analyzers.
The key to successful system integration begins at the proposal
stage where Emerson Process Management develops a
custom-engineered solution. This is followed by experienced
project management during the system fabrication and on to
installation and training once the system is delivered to the
field.

700XA Natural Gas Chromatograph

Environmental Chamber Testing


Every Emerson Gas Chromatograph that leaves our facility
undergoes rigorous testing throughout assembly. The
majority of our systems are put into a 24-hour environmental
chamber test, where they must operate to specification in an
environment where the temperatures cycle between 0 and
130 F (-18 and 54 C) for a minimum of 24 hours.
Our product testing procedures are much stricter than the
industry standard for analytical measurement products. When
you purchase an Emerson Gas Chromatograph, you can be
assured that you are investing in the highest-quality online gas
chromatograph available. As a result of chamber testing, we
can guarantee all gas chromatographs that we ship will operate
to the performance specifications across the stated operating
temperature range.

Custom-Engineered Sample Systems


Any gas chromatograph is only as good as the quality of the
sample it measures. The standard sample system includes
a particulate filter and liquid filter/shut-off for each sample
stream. If required, the sample system can also be custom
engineered for the specific requirements of the application.
Common features include:
Heated and open-panel designs
All components rated for the area classification
Variety of sample probes to extract a reliable and
stable sample from the process

Danalyzer 700XA Gas Chromatograph inside the


environmental testing chamber

www.Danalyzer.com

700XA Natural Gas Chromatograph

April 2014

(This page purposely left blank for your notes)

www.Danalyzer.com

700XA Natural Gas Chromatograph

April 2014

Specifications
Please consult Rosemount Analytical if your requirements are outside the specifications listed below. Improved performance, other
products and material offerings may be available depending on the application.

Construction

Detector: Thermal conductivity detector (TCD), flame ionization


detector (FID), TCD/TCD or TCD/FID dual detector configurations
possible; flame photometric detector (FPD) available.

Hazardous Area Certified for: -20 to 60 C/-4 to 140 F


Reliable performance over broad ambient temperatures:
40 to 60 C/-40 to 140 F

Gating Options: Fixed-time, slope sensing gating of peaks

Enclosure Protection Rating: IP66


Dimensions (without sample system):
Wall-mount: 711 mm H x 445 mm W x 498 mm D
(28'' H x 17.5'' W x 19.6'' D)

Streams: Up to 20 externally controlled streams or up to 8 internal


(includes calibration stream)

Chromatograms stored/archived internally: Stores over 80 days of


analysis report data and up to 2500 individual chromatograms.

Electronics

Pipe-mount: 711 mm H x 445 mm W x 671 mm D

Power:

(28'' H x 17.5'' W x 26.4'' D)

Standard: 24V DC (2130 VDC)

Floor-mount: 1532 mm H x 445 mm W x 612 mm D

Optional: 90264V AC, 4763 Hz

(60.3'' H x 17.5'' W x 24.1'' D)

Typical Power Consumption at 22 C (72 F):


Startup: 105 Watts DC (125 Watts AC)
Steady State: 35 Watts DC (40 Watts AC)

Corrosion Protection:

GC Enclosure Material: Copper free aluminum coated with



industrial grade powder coat suitable for high humidity and salt- Note: Add 15.5 Watts DC (18 Watts AC) for LOI
laden environments.

Communications (Standard)

Process Wetted Materials: Stainless steel. Where the function of


an item excludes the use of stainless steel (e.g. glass rotameter
tubes), materials that are resistant to corrosion are used.

Electronics: All electronic circuit boards are tropicalized with a clear


conformal coating.

Mounting: Floor-standing (standard), wall- or pipe-mount (optional)

Approximate Weight (without sample system): 50 kg (110 lbs.)

Area Safety Certification Options:*


CSA:
USA and Canada
Class I, Division 1, Groups B, C, and D
ATEX / IECEx
Ex II 2G
Ex d IIC Gb T6
(Ta = -20 C to 60 C)

Ethernet: Two ports one RJ-45 and one four-wire


with 10/100 mbps
Analog inputs: Two standard isolated inputs filtered with transient
protection, 420 mA (user scalable and assignable)
Analog outputs: Six self-powered isolated outputs (420 mA)
Digital inputs: Five inputs, user assignable, optically isolated,
rated to 30V DC @ 0.5 A
Digital outputs: Five user-assignable outputs, Form C and
electromechanically isolated, 24V DC
Serial: Three termination blocks, configurable as RS-232, RS-422
or RS-485 and one RS-232 D-sub (9-pin) Modbus/PC Connection

Communications (Optional)
Two expansion slots available for additional communications.
Each slot has the capacity to add one of the following:

Four analog inputs (isolated) card

Four analog outputs (isolated) card

Eight digital inputs (isolated) card

Five digital outputs (isolated) card

One RS-232, RS-422 or RS-485 serial connection card

One modem card, 30019.2k baud

*Stated T-ratings can vary based on applications.

Performance Capabilities

Oven: Airless, maximum 150 C (302 F)


Valves: Six-port and ten-port diaphragm chromatograph valves. Other
Additionally, a FOUNDATION fieldbus module is available.
types of valves, such as liquid injection or rotary valves, may be used
depending on the application
Memory Capacity: 1 GB of flash memory for data storage; 128 MB of
SDRAM system memory with 2 MB static RAM (battery-backed)
Carrier Gas: Application-dependent. Typically zero-grade helium,
nitrogen, or hydrogen
Sample & Calibration Gas Input Pressure Range:
0.20682.0684 bar: 1.0342 bar (recommended) or 15 psig
Carrier Gas Input Pressure Range (recommended):
6.20526.8947 bar (90100 psig)

www.Danalyzer.com

Touch Key Local Operator Interface (Optional)


The 700XA local operator interface (LOI) allows for maintenance and
operation of a 700XA without a laptop or PC. The LOI is a state-of-theart high resolution color display that is touch key infrared activated and
supports all core GC operations.

700XA Natural Gas Chromatograph

April 2014

Recommended Installation
The drawings below represent the minimum recommended installation guidelines for the 700XA Gas Chromatograph.
Please consult Rosemount Analytical for detailed installation recommendation of your application.

Floor Mounted Details

10

www.Danalyzer.com

April 2014

700XA Natural Gas Chromatograph

Recommended Installation
The drawings below represent the minimum recommended installation guidelines for the 700XA Gas Chromatograph.
Please consult Rosemount Analytical for detailed installation recommendation of your application.

Wall and Pole Mounted Details

www.Danalyzer.com

11

700XA Natural Gas Chromatograph

Product Data Sheet

NGC_PDS_700XA

April 2014

www.RosemountAnalytical.com
www.Danalyzer.com
www.analyticexpert.com

www.twitter.com/RAIhome

www.youtube.com/user/RosemountAnalytical

www.facebook.com/EmersonRosemountAnalytical

AMERICAS
Emerson Process Management
Rosemount Analytical
Analytical Center of Excellence
10241 West Little York, Suite 200
Houston, TX 77040 USA
Toll Free 866 422 3683
T +1 713 396 8880 (North America)
T +1 713 396 8759 (Latin America)
F +1 713 466 8175
gc.csc@emerson.com
EUROPE
Emerson Process Management
Bond Street, Dumyat Business Park
Tullibody FK10 2PB UK
T +44 1259 727220
F +44 1259 727727
sales.gcema@emersonprocess.com

MIDDLE EAST AND AFRICA


Emerson Process Management
Emerson FZE
Jebel Ali Free Zone
Dubai, United Arab Emirates, P.O. Box 17033
T +971 4 811 8100
F +971 4 886 5465
gc.csc@emerson.com

2014 Emerson Process Management. All rights reserved.


The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount Analytical
is a mark of one of the Emerson Process Management family of companies. All other marks are
the property of their respective owners.
The contents of this publication are presented for information purposes only, and while effort has
been made to ensure their accuracy, they are not to be construed as warranties or guarantees,
express or implied, regarding the products or services described herein or their use or
applicability. All sales are governed by our terms and conditions, which are available on request.
We reserve the right to modify or improve the designs or specifications of our products at any
time without notice.

ASIA-PACIFIC
Emerson Process Management
Asia Pacific Private Limited
1 Pandan Crescent
Singapore 128461
Republic of Singapore
T +65 6 777 8211
F +65 6 777 0947
analytical@ap.emersonprocess.com

Product Data Sheet

Model 700

DAN-PDS-GC-700

Daniel Danalyzer Model 700


Gas Chromatograph
Danalyzer Model 700 Gas Chromatographs offer the
broadest range of analysis options available today in a fieldmount gas chromatograph (GC). Whether it is heating value
measurement, trace contaminant monitoring, pipeline integrity,
or product quality/process control, the Danalyzer Model 700
is flexible enough to meet your analysis needs. Based on the
same proven technology and software as the Danalyzer
Model 500 series, the Danalyzer Model 700 design offers
superior reliability and precision, lower installation and
operating costs, greater application flexibility, and unmatched
measurement performance.

Features and Benefits


One package for fiscal metering or gas quality
Custody transfer analysis from C6+ to C9+
Contaminant monitoring - trace H2S, CO2, oxygen,
etc.
Combine measurements and reduce analysis cost
Oxygen (0-2%)
Trace H2S (0-50 ppm)
CO2 (as fast as 90 seconds)
Helium/Hydrogen (0-10%)
Hydrocarbon Dew Point (2C)
Reduced installation costs
24 VDC power standard
Integrated controller electronics
Pipe-mount, wall-mount, or floor mount
Lower operation and maintenance costs
No shelter or instrument air required
Low helium and power consumption
Longest gas chromatograph valve and column
warranties available
Visit us online at
www.daniel.com

Unmatched measurement performance


Highest stated precision 0.25 BTU/1000 for
broad ambient temperature C6+ analysis in -18 to
54C (0-130F) environmental conditions
Wide dynamic range from percent to trace level
components
Reliable performance over broad ambient
temperatures -30 C to 60 C (-20-140 F)
Easy to use
Compatible with Danalyzer Model 500 Gas
Chromatographs
MON2000 software setup and diagnostics
Identical setup and operation for all models no
additional training

Product Datasheet
DAN-PDS-GC-700

Hydrocarbon Dew Point Monitoring

APPLICATIONS
Standard Natural Gas Applications

We have made our most popular energy and gas


quality applications standard. Applications may vary
by components of interest, analysis time, reduced
hardware (economics), or improved precision. Choose
the one that is right for you, or we will create one that
is customized for your specific application.
Energy Measurement (to C6+ and C9+)
The Danalyzer Model 700 offers applications for
energy measurement from C6+ hydrocarbon ranges to
C9+ hydrocarbon ranges. Calculations based on GPA
2145 or ISO 6976 standards can be provided.

Standard Measurement Ranges


Methane
65 to 100-mole %
Ethane
0 to 20-mole %
Propane
0 to 10-mole %
N-Butane
0 to 5-mole%
Iso-Butane
0 to 5-mole %
N-Pentane
0 to 1 mole %
Iso-Pentane 0 to 1 mole%
Neo-Pentane 0 to 1 mole%
Hexane+
0 to 0.7-mole%
Nitrogen
0 to 20 mole %
Carbon Dioxide 0 to 20 mole%
Hexanes
0 to 1 mole%
Heptanes
0 to 1 mole%
Octanes
0 to 0.5 Mole%
Nonane+
0 to 0.5 mole%

Gas Quality Analysis

C 6+ C 9+

Natural gas contaminants reduce pipeline integrity


over time. Most contaminants can be easily measured
in the Model 700 for online quality assurance.
Contaminant monitoring can be combined with
energy measurements for complete custody transfer
analysis. To the extent possible, these combined
applications utilize independent gas chromatograph
valves, detectors, and columns, for each primary
measurement. This technique offers greater reliability,
increased speed, and easier troubleshooting. This
application approach also makes field upgrades and
re-applications in the Danalyzer Model 700 easy by
minimizing internal piping changes.

Page 2

Accurate and reliable hydrocarbon dew point


determination can now be made as an option to any
C9+ extended analysis application. The DewCalc
Hydrocarbon Dew Point software is completely integral
to the gas chromatograph and provides dew point
temperatures and cricondentherm for up to four userentered pressures. Alternately, real-time dew point
results can be provided by using analog or Modbus
inputs from another device.
DewCalc software is loaded with features,
including:
Choice of two equations of state, Peng-Robinson or
Soave-Redlich-Kwong
Library of 93 hydrocarbon compounds for detailed
breakdown of C9+ fraction
DewCalc report with analysis results and factors
used

Custom Applications

If our standard applications do not fit your needs, the


Danalyzer Model 700 can be customized to meet
your measurement requirements. Simply submit a
completed application data sheet found at the end of
this data sheet with your request, or ask our Application
Engineers for assistance.

Product Datasheet
DAN-PDS-GC-700

Standard Danalyzer Applications and Repeatability


Application
Code

Components
Measured

Application Name

Analysis
Time

% Repeatability
@ Controlled temperature

% Repeatability Over
0-130F

604

Standard BTU/CV

C1 to C6+, CO2, N2

4 min

0.0125

0.025

6GT

Gas Turbine BTU/CV/


Density

C1 to C6+, CO2, N2

2 min

0.05

0.1

TS2

Gas Quality (H2S)

Trace H2S (3 - 30 ppm)

90 sec

2 ppm

3 ppm

LF8

Landfill Gas BTU/CV

C1, N2, CO2, O2

8 min

0.1

0.2

LF3

Landfill Gas (mole-sieve)

C1, N2, CO2, O2

3 min

0.1

0.2

6PS

Std C6+ with % H2S

C1 to C6+, CO2, N2, H2S


(%)

10 min

0.05

0.1

6PO

Std C6+ with % Oxygen

C1 to C6+, CO2, N2, O2

10 min

0.05

0.1

6HH

Std C6+ with Helium/


Hydrogen

C1 to C6+, CO2, N2, He/


H2

5 min

0.05

0.1

6TS

C6+ with ppm H2S

C1 to C6+, CO2, N2, H2S


(ppm)

7 min

0.0125/2 ppm

0.015/3 ppm

905

Std C9+

C1 to C9+, CO2, N2

5 min

0.025

0.05

HC9

Std C9+ and Hydrocarbon Dewpoint

C1 to C9+, CO2, N2

5 min

0.025

0.05

712

Standard C7+ BTU/CV

C1 to C7+, CO2, N2

10-12 min

0.025

0.05

SUPERIOR PERFORMANCE
The Danalyzer Difference
Modular Analytical Oven
An element that is key to the Danalyzer Model 700 Gas
Chromatographs application flexibility is the modular
analytical oven. Using field-proven Danalyzer valves,
columns, and detectors, the analytical oven has been
redesigned to make it modular and scalable. Six- or
ten-port valves and detectors mount to a triangular
kiosk block which facilitates valve heating and
pneumatic actuation. Three temperature control zones
and up to three valves and two detectors operating
independently provide extreme application flexibility
and range. All components in the oven are completely
accessible and serviceable in the field.

Gas Chromatograph Valves


At the heart of the
Danalyzers superior
reliability are the six- and
ten-port diaphragm/piston
gas chromatograph valves. These pneumatic valves
are guaranteed for the life of the gas chromatograph
and specified to operate over 5 million times between
service. By minimizing internal movement (1/1,000
inch) of the pistons, which never come in contact
with sample, abrasive mechanical wear is virtually
eliminated. This unique double-diaphragm design
removes the need for all springs, o-rings, or lubrication.
Valve service is performed by replacing a cost-effective
diaphragm set, normally completed in less than 10
minutes.
Detectors
The thermal conductivity detector
(TCD) is the detector of choice
for most applications due to
its universal response to all
compounds. The TCD found in
the Danalyzer Model 700 Gas Chromatograph is able
to go well beyond the normal measuring ranges seen
in other designs by being able to do many applications

Page 3

Product Datasheet
DAN-PDS-GC-700
with low parts-per-million measurement requirements.
This greatly simplifies the gas chromatograph design
when a simple and rugged TCD can be used. The
flame ionization detector (FID) for trace hydrocarbon
measurements and the flame photometric detector
(FPD) module for trace sulfur measurements are also
available.
Micro-packed Columns
Danalyzer micro-packed columns
offer a superior combination of
features found in both capillary and
conventional packed columns
speed, sharp peak resolution, and low
carrier gas consumption. In addition,
this unique design provides for greatly extended column
life and the longest warranty available on the market
(five years on the standard C6+ natural gas set).
Stream Switching Module
The Danalyzer Gas
Chromatograph offers an
internal sample stream
switching module. Available
in four- or eight-stream
versions, this module utilizes the same helium-rated
24-volt solenoids used in the analytical oven. It is
temperature-controlled for operation above the sample
dewpoint. For applications with widely varying stream
composition, double block and bleed configuration is
optionally available.

Local Indication and Operation Panel


Analyzer health and valve status can be viewed through
the glass cover of the lower enclosure. The panel
displays green (OK), yellow (warning), and red (failure)
LEDs, along with LEDs for all gas chromatograph valve
actuations, power, and CPU health. Each valve can be
actuated manually for simplified troubleshooting and
fast system purging after maintenance.
Touch Key Local Operator Interface (Optional)
The Model 700 local operator interface (LOI) allows
for maintenance and operation of a Model 700 without
a laptop or PC. The LOI is a state-of-the-art high
resolution color display that is touch key infrared
activated and supports all core GC operations.
Features of the LOI include:
- Color LCD with QVGA (320 x 240 pixels) resolution
- ASCII text and graphics modes
- Auto-backlighting (adjustable)
- 8 infrared-activated touch keys and screen saver
- Eliminates external magnetic pen requirement
- Maintains the Model 700 hazardous area
classifications
- Operates in the same ambient temperatures as the
base Model 700
- Complete GC status, control and diagnostics,
including full chromatogram display

CONTROLLER ELECTRONICS AND


COMMUNICATIONS
Modular Electronics
The controller electronics, option cards, and field
termination boards are packaged conveniently in the
lower enclosure of the Danalyzer Model 700. Separate
power and output connections are made through this
enclosure.
Compatible Electronics and Software
The Danalyzer Model 700 uses the same CPU board
as the Danalyzer Model 500, thus making features,
options, and operation nearly identical. No new
training is required.

Page 4

Model 700 shown with


optional touch-key LCD LOI

Product Datasheet
DAN-PDS-GC-700
Data archiving and reports
The Danalyzer Model 700, with expandable solid-state
memory, virtually eliminates the need for external
data storage. Every analysis is time/date stamped
and archived, for retrieval via the MON2000 software.
Pre-configured reports can be displayed, printed, and/
or stored internally. Results can be trended directly or
exported easily in ASCII format.
Security - three levels of password protected
security configurable to read/write or read only for
third party access
Audit logs - data and event logging fully conforms
to API report 21.1 for metering audit purposes and
backup to primary systems (flow computer, SCADA,
DCS)
Event logs - a continuous record of all operator
changes, with time, date, and operator name
recorded
Alarm logs - a continuous record of all historical
alarms, time/date stamped with alarm state and
description

Standard reports include:


Average reports - hourly, 24-hr., weekly, monthly
and variable averages
Analysis reports - physical property calculations by
component and group for analysis and alarms
Raw data report - retention times, peak areas,
detector number, method, integration start/stop, and
peak width for the analysis
Calibration report - raw component data, new
response factors, retention times, and deviation for
last calibration
Final calibration report - results of final calibration,
response factor and retention time adjustments

ENVIRONMENTAL CHAMBER TESTING

We provide the most thorough gas chromatograph


testing in the world. Prior to shipment, we test
nearly every Danalyzer Model 700 in our walk-in
environmental test chambers, where temperatures
are cycled between 0F and 130F for a minimum
of 24 hours. This is all part of our commitment to
provide analyzers that are capable of providing reliable
measurements in the field.

Maintenance log - a scratch pad for keeping track


of maintenance or testing performed on the gas
chromatograph system

BTU Repeatability In Environmental Chamber


Model 700 C 9 + Application - Pipeline Gas
COLD SOAK

COLD RAMP

HOT SOAK

HOT RAMP

1040

1039

BTU

130oF

1038

1037

70oF

1036

1035

0oF

Stated Repeatability (oven temperature): 0.05% CV


Repeatability as tested: 0.0165%

ELAPSED TIME

17 Jan 04

22:23

21:16

20:08

19:01

17:54

16:47

15:40

14:33

13:25

12:18

11:11

8:57

10:04

7:50

6:43

5:35

4:28

3:21

2:14

1:07

1034

0:00

Archiving - the last 35 days of each analysis and 64


calibrations can be archived automatically by time/
date stamp. Last chromatogram stored for each
sample and calibration stream.

D eg C

Superior measurement stability in extreme climates is


tested for every Model 700 Process Gas
Chromatographs before it ships.

Page 5

Product Datasheet
DAN-PDS-GC-700

MON2000 SOFTWARE
The Danalyzer Model 700 Gas Chromatograph is
designed to operate unattended. If adjustments are
needed, our exclusive MON2000 software allows
complete control of your gas chromatographs either
locally or remotely. From within MON2000, a user can:

MON2000 also has a number of tools built in to help


users manage their analyzers such as:
Automatic recording of alarms in a log file
Event logs that provide a continuous record of all
operator changes with time and user name stored

Review and modify analytical settings on one screen

Maintenance log scratch pad for keeping track of


maintenance or testing done

Upload and display multiple chromatograms on the


screen for comparison

Download and install the latest GPA 2145 and ISO


6976 physical property factors

Upload and trend any of the measured results


Export data for use in other third-party applications
Overlay multiple chromatograms for troubleshooting
and calibration
Check original calibration against last calibration

Data collected from the Danalyzer Model 700 can be


stored and displayed in a wide range of options such as
trend lines on the screen and logs automatically documenting all changes made to the gas chromatograph.
Data can also be exported in formats compatible with
most third-party Windows applications.

The MON2000 software is Windows -based software


designed to make analyzer configuration, maintenance
and data collection easy. With intuitive drop-down
menus and fill-in-the-blank tables, even new users can
quickly navigate through the software. The MON2000
software can display both current and multiple archived
chromatograms on the screen streamlining the time
needed to perform routine analyzer maintenance.

Simple dropdown menus

Powerful Yet Simple MON2000 Software


Summary of application
event status and time

Connect to
any GC with
a mouse click

Automatic listing of
measured
components

Full featured chromatogram


display

Manually adjust
X/Y scale

Page 6

Quickly add
chromatograms
to overlay

Interactive cursur

Display results

Save chromatograms
to hard drive

Product Datasheet
DAN-PDS-GC-700

ANALYZER NETWORKING AND DATA


COMMUNICATION

Secure Modbus Connectivity

For online gas chromatographs, Modbus continues to


be the preferred choice to connect to a gas chromatograph network. Our Modbus design avoids the use of
central interface cards or computers that can act as a
single-point of failure in the Modbus link. Instead, the
flow computer or SCADA system can talk directly to
each gas chromatograph to gather the data needed.
Furthermore, the register and coil addresses can be
easily customized to meet the specific data structure of
Data communication options include simple analog
and discrete signals as well as Modbus serial links. To the flow computer or SCADA system. There is also a
preserve the integrity of the analysis data, all Danalyzer program built into the MON2000 workstation to test the
Model 700 Gas Chromatographs are capable of storing Modbus link if troubleshooting is needed.
up to 35 days of analysis or calibration data in the event
Simplified Network Communications
of loss of communication.
Danalyzer Gas Chromatographs can be configured
in a number of networking and data communication
schemes to meet the communication requirements
of the natural gas industry. Options include modem,
Ethernet, as well as multi-drop RS-485 networks. It
is even possible to set up automatic polling for data
collection if desired.

All Danalyzer Gas Chromatographs are designed to


operate unattended. Occasionally, adjustments to
the analyzer analytical method or a review of possible
alarms may be needed. Using our exclusive MON2000
software loaded on either a PC or laptop running
Windows, youll have complete control of your gas
chromatographs either locally or remotely.

- Multiple serial ports with Ethernet


option can be field configured to provide
redundant data highways.
INTERNET
- RS 232C
- RS 485/422
- ETHERNET

MODEL 700
MON2OOO
WORKSTATION

- RS 232C
- RS 485/422
- ETHERNET

MODEL 700

PRINTER / LOGGER

Networking Flexibility
Whether you want to network gas chromatographs
throughout the plant or simply link a single gas chromatograph to a flow computer system, the Danalyzer
Model 700 can be configured to handle most of the
scenarios:
Choice of Ethernet or RS-485 network
Can use same network to connect Danalyzer Model
700, Model 500, and Model 1000 Gas Chromatographs

OPTIONAL LAPTOP
FOR LOCAL
COMMUNICATION

REDUNDANT
RS 485/422
UP TO
32 GCS

MODEL 700
OPTIONAL LAPTOP
FOR LOCAL
COMMUNICATION

MODEL 700

DCS OPC SERVER


ETHERNET - UNLIMITED NUMBER OF GCS

OPC SERVER

MODEL 500
- RS 232C
- RS 485/422
- ETHERNET

MODEL 700

PQR
PQR

MON2OOO
WORKSTATION

MODEL 500
- RS 232C
- RS 485/422
- ETHERNET

MODEL 1000

MODEL 500

MON2OOO
WORKSTATION

MODEL 1000

MODEL 500

Able to connect multiple PC workstations using


MON2000
Connectivity to flow computer or SCADA systems
using industry standard protocols such as Modbus
and OPC

Page 7

Product Datasheet
DAN-PDS-GC-700

ACCESSORIES

We offer standard and custom gas chromatograph


accessories to meet your installation and application
requirements.

Gas Chromatograph System Enclosures

Complete range of gas chromatograph enclosure


solutions ranging from simple sun-shields and
cabinets to skids systems and shelters
API 14.1 compliant solutions with heating of sample
lines and calibration gas

Standard Auxiliary Equipment

3 - Sided Shelter

Multi-stream Heated Sample


Conditioning System

Sample probes / pressure regulators / filters for a


wide range of sample handling requirements
Carrier gas systems with dual manifold regulators for
uninterrupted operation
Calibration gases and heating blankets
Carrier Manifold

GAS CHROMATOGRAPH ACCESSORIES

Laptop Computer Shelf mounts easily - folds down


on side (PN 2-3-0700-143 )
"Condulet" Junction Box for quick laptop PC (db9
connector) access

Condulet Junction
Box

Sample Probe

Custom-Engineered Sample Systems


Any gas chromatograph is only as good as the quality
of the sample it measures. Every sample system for
Danalyzer Gas Chromatographs is custom-engineered
for the specific requirements of the application.
Common features include:
Heated and open panel designs
All components rated for the area classification
Automatic calibration / validation available as an
option
Variety of sample probes to extract a reliable and
stable sample
Laptop Computer
Shelf

Page 8

Calibration Gas
with Heating Blanket

Product Datasheet
DAN-PDS-GC-700
Danalyzer Model 700 Gas Chromatograph
CSA Certified Danalyzer GC, 6 or 10 port valves, single or dual detectors
Ordering Matrix (must select only 1 option for each column)
77X
C
W
DC
4 SN
N S4 612 1N
S
(example)
Code Approvals
C CSA
A ATEX
Code
W
P
F

Mounting
Wall-mount
Pipe-mount
Floor-mount
Code Power
DC 24 VDC
AC Universal 85 to 240 VAC
Code Digital/Analog I/O
4 5 Digital In; 4 Digital Out; 4 Analog In; 4 Non-isolated Analog Out
8 5 Digital In; 4 Digital Out; 4 Analog In; 8 Analog Out, 4 Non-Isolated/4 Isolated
12 5 Digital In; 4 Digital Out; 4 Analog In; 12 Non-isolated Analog Out

Code Serial Communications/Local Keypad-LCD


SN Standard/no keypad - LCD {(2) RS232, (1) RS422/486 Factory/Field configurable}
EN Extended/no keypad - LCD {(4) RS232, (2) RS422/486 Factory/Field configurable}
Code Internal Modem and/or Ethernet
N No Modem or Ethernet
M Modem only
E Ethernet only
B Modem and Ethernet
X Special (consult factory)
Code Internal Stream Switching
(Total max. # streams = sample + calibration streams)
S4 Standard Up to 4 streams
S8 Standard Up to 8 streams
D4 Double block and bleed - Up to 4 streams
D8 Double block and bleed - Up to 8 streams
ES External stream switching - (consult factory)
X Special (consult factory)

Manuals: (one set included with each system)


Hardware manual - P/N 3-9000-521
Software manual - P/N 3-9000-522
All Danalyzer Model 700 Gas Chromatographs must be
crated for shipment.
* Users with rich natural gas applications may require
heated sample system ovens and sample transport
tubing to insure that the sample is kept above its
sample dew point. Consult the factory to determine if
heated sample system is needed.

Code Oven Assembly/Application


612 BTU/CV C6+ / 12 min.
604 BTU/CD C6+ / 4 min.
6GT BTU/CV C6+ / 2 min./Turbine
TS2 Trace H2S / 3 min.*
LF8 Landfill Gas / 8 min.
LF3 Landfill Gas / 3 min. (mole-sieve)
6PS Std. C6+ with % H2S
6PO Std. C6+ with % Oxygen
6HH Std. C6+ with Helium/Hydrogen
6TS C6+ / 12 min and ppm H2S / 7 min.*
905 Std. C9+ and HC Dew
X Special
Code Sample Conditioning (sample streams only)
1N 1 stream plate
1G 1 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off
2N 2 stream plate
2G 2 stream plate, with Fast Loop/Membrane Filters/Liq. Shut-off
3N 3 stream plate
3G 3 stream plate, with Fast Loop/Membrane Filters/Liq. Shut-off
4N 4 stream plate
4G 4 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off
5N 5 stream plate
5G 5 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off
6N 6 stream plate
6G 6 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off
7N 7 stream plate
7G 7 stream plate, with Fast Loop/Membrane Filter/Liq. Shut-off
O Heated Sample System (Order Separately)
X Special (consult factory)
Code Application Software
S Std. GPA 2172/GPA2145
I ISO 6976;1995
X Special (consult factory)

Page 9

Application
Data Sheet
Product Datasheet
DAN-PDS-GC-700
Every Danalyzer Model 700 Gas Chromatograph is

custom built for the specific application requested. To


request a free quotation, simply fill out the information
below and send it to your world area contact information
located on the back cover of this datasheet.
Name:

Your Reference #:

Title:

Project Name:

Company:

Process Unit:

Address:

Stream Name(s):
Hazardous Area Classification:

Phone #:
Fax #:
Email:

Stream Composition Units

Stream 1
Min

Normal

Stream 2

Max

Measure

Min

Normal

Max

(For more than two streams, make copies of this page)

Stream 1

Stream 2

Stream Temperature
Stream Pressure
Stream Phase (vapor/liquid)
Stream Contaminants

Distance to GC from sample point


Power:

100-240 VAC ( )

24 VDC ( )

Mounting:

Wall ( )

Pipe ( )

Stand ( )

Data communication

Analog output ( )

Modbus ( )

OPC ( )

Analyzer network:

Ethernet ( )

RS-485 Multi-Drop ( )

Other options:

Carrier Gas ( )

Calibration Gas ( )

Special Instructions:

Page 10

Start-up ( )

Printer ( )
Training Class ( )

Modem ( )

Measure

Product Datasheet
DAN-PDS-GC-700
Specifications
Sample Input Pressure Range (recommended): 15-20 psig

Power:
Standard: 24 VDC, (24.5-26.5 VDC)
Optional: 90-264 VAC, 47-63 Hz
80 watts start-up; 33 watts nominal

Carrier Gas Input Pressure Range (recommended): 90-100 psig


Detector: Thermal conductivity detector (TCD), flame ionization detector
(FID), TCD/TCD or TCD/FID dual detector configurations possible; flame
photometric detector (FPD) available (see FPD module data sheet)

Environment: -29 to 60C (-20 to 140F)


Enclosure Protection Rating: IP66 and NEMA 4X

Gating Options: Fixed-time automatic slope sense gating of peaks

Dimensions (without sample system):


Wall-mount: 914mm H x 462mm W x 483mm D
(36 H x 18.2 W x 19 D)
Pipe-mount: 914mm H x 462mm W x 645mm D
(36 H x 18.2 W x 25.4 D)
Floor-mount: 1483mm H x 419mm W x 577mm D
(58.4 H x 16.5 W x 22.7 D)

Streams: Up to eight streams (including calibration stream) internally


controlled; up to 12 streams externally controlled
Chromatograms stored/archived internally: Most recent analysis and/or
calibration per stream
Analog Inputs: Four inputs filtered with transient protection (user scalable and
assignable)

Mounting: Free-standing (standard); wall-mount or pipe-mount (optional)

Analog Outputs: Four non-isolated analog outputs, 4 20 mA (standard); four


additional isolated analog outputs or eight additional non-isolated analog
outputs (optional)

Approximate Weight (without sample system): 80 kg (175 lbs.)


Area Safety Certification Options:
CSA: Class 1, Division 1, Groups B, C, D, T4

USA: Class I, Zone 1, Ex d IIB+H2, T4, Enclosure Type 4

Canada: Class I, Zone 1, Ex d IIB+H2, T4, Enclosure Type 4
ATEX: Ex II 2G Ex d IIC T4 Tamb = 60C
IECEx: Ex d IIC T4 Gb Tamb = 60C

Serial Communication Ports: Three serial ports standard with option for a
total of seven. Depending on the port, choice of RS-232, RS-422, and
RS-485 is available as well as the Modbus protocol
Digital Inputs: Five digital inputs; user assignable
Digital Outputs: One gas chromatograph alarm and four user-assignable
outputs (2 form A, 3 solid-state optically isolated)

Oven: Airless heat sink, maximum 130C (266F)


Valves: 6-port and 10-port diaphragm chromatograph valves. Other types of
valves may be used, such as liquid injection or rotary valves, depending
on the application

Parallel Printer Port: One parallel port available for printed reports
Internal Modem (optional): Field-configurable; 300 to 33.6k baud rate
Ethernet Card (optional): 10 mbps with RJ-45 port

Carrier Gas: Application-dependent. Typically zero-grade helium, nitrogen, or


hydrogen

RAM (DATA) Memory Capacity: 128 Mb RAM (standard)

Outline and Mounting Details


25.4
22.4

4 OR 2 POLE
MOUNT OPTION

15.50

22.7

4X .50 THRU
13.6

26.25
REMOVAL
CLEARANCE
TYPICAL

56.6
1.56

58.4

SAMPLE PLATE

12.0

36.8

33.8
RECOMMENDED
FROM GROUND
OR FLOOR

SAMPLE PLATE
16.75

16.5

22.1

FRONT VIEW FLOOR MOUNT UNIT

SIDE VIEW FLOOR MOUNT UNIT

18.2

FRONT VIEW WALL MOUNT UNIT


POLE MOUNT UNIT

Page 11

Product Datasheet
DAN-PDS-GC-700

Notes:

Emerson Process Management


Daniel Measurement and Control, Inc.
World Area Headquarters
Houston, Texas, USA
T: 713-827-6380, F: 713-827-3865
USA Toll Free 1-866-GC-CENTER
gc.csc@emerson.com
www.daniel.com
Calgary, Alberta, Canada,
T: 403-279-1879, F: 403-236-1337
Singapore - Asia Pacific,
T: +65-6777-8211, F: +65-6770-8001
Stirling, Scotland - Europe, Mid-East, Africa,
T: +44-1259-727-222, F: +44-1786-433-606

2010 Daniel Measurement & Control, Inc. All rights reserved. Unauthorized
duplication in whole or in part is prohibited. Printed in the USA.
Page 12
May 2010

Daniel Measurement and Control, Inc. and Rosemount Analytical are


a wholly owned subsidiaries of Emerson Electric Co., and are divisions
of Emerson Process Management. The Daniel name and logo are
registered trademarks of Daniel Industries, Inc. The Emerson logo is
a registered trademark and service mark of Emerson Electric Co. All
other trademarks are the property of their respective companies. The
contents of this publication are presented for informational purposes
only, and while every effort has been made to ensure their accuracy,
they are not to be construed as warranties or guarantees, express or
implied, regarding the products or services described herein or their
use or applicability. All sales are governed by Daniels terms and
conditions, which are available upon request. We reserve the right to
modify or improve the designs or specifications of such products at
any time. Daniel does not assume responsibility for the selection, use
or maintenance of any product. Responsibility for proper selection,
use and maintenance of any Daniel product remains solely with the
purchaser and end-user. Daniel products are protected by United States
and international patents and patents pending.