Académique Documents
Professionnel Documents
Culture Documents
The HPLC (LC-20A) has been designed to provide high accuracy and high sensitivity
analysis. System configurations along with description of the operation of the various
components are provided below. Each component of the system is controlled locally for
stable analysis.
The low pressure gradient unit mixes upto 4 mobile phases that have been degassed by the
degasser
**
The mixer enhances the mixing efficiency of the mobile phases. This item is required
for low or high pressure gradient system.
**
The autosampler automatically injects the samples into the flow lines. By adding a
rack changes it is possible to automatically change autosampler racks.
Steps
Mobile phase is drawn out of the reservoir bottle and pumps through the tubing
by the pump.
The degasses removes dissolved air from the mobile phase preventing the air
bubbles and consequent rise drift or other baseline irregularities caused by
dissolved air.
Pump sends the mobile phase through injector, column and detector and finally
into the waste container.
The detector detects the components eluted from the column and sends the
signal data to a chromato pacer computer.
Reservoir bottle
Degasser
Low Pressure Gradient Unit
Pump Unit
Mixer
Injector
Auto Samples
Column
Detector
Water Container
Column Oven:
It maintains the temperature of the Liquid Chromatography System column and flow
lines at a content temperature, in order to separate performance. The oven is an aircirculated thermostat chamber for HPLC. Column oven regulates temperatures with a heater
while AC model equipped with cooling mechanisms that enables it to regulate the oven and
flow lines to temperatures below room temperature. The air circulation ratio of the fan
automatically matches the set temperature to keep temperature constant. At high
temperatures, the air circulation ratio is increased to keep the temperature constant at
temperature close to the ambient temperature, the air circulation ration is decreased to
suppress heat generation due to air circulation. The fan speed is controlled so as to enable
stable temperature regulation.
Autosampler:
The applicable pressure in 20 MPa-35MPa
The injection capacity ranges from 0.1l to 2000 l
There re two modes for samples injection
Standard injection mode that facilitates condition setting. This
is the simplest mode. In this mode, specify the sample vial
number, number of injections, injection volume and analysis
time in sample table setting screen. Set needle stroke, rinse
volume, rinsing speed, sampling speed in parameter setting
Bracket sequence analysis mode that insertion analysis can
be performed in a certain cycle. This mode can be used to
inject a predetermined number of samples repeatedly at
periodic intervals.
Preparing Samples
Completely dissolve the sample in a solvent with an identical or similar composition
to mobile phase.
Filter the sample to eliminate undissolved particles to avoid clogging the needle.
Dilute highly viscous sample before use as they as not aspirated.
Carefully load the sample into a vial by injecting.
Slide the sample rack. The sample rack check into place when it is inserted correctly.
(If the sample rack is not inserted correctly, the sampling needles may stick the
wrong position and may be bent.
Close the door.
The sample is injected along with the mobile phase into the flow
lines passes through the high pressure valve and finally into column
where analysis begins. Purge of measurement flow line (Rinse
solution aspiration). The low pressure value rotates 300 counter
clockwise. Plunger of the measuring pump has returned to original
position, then the low pressure value rotates 600 clockswise the
plunger operates to aspirate rinse solution.
Rinse Solution Dischargement
The low pressure valve rotates 600 counter clockwise measuring
pump plunger operates to discharge rinse solution to rinse port.
175 vials
105 vials
70 vials
50 vials
Sample Vials
4ml sample vial
4ml cap
borosil glass
1.5 ml
B Glass
1.1 ml
B Glass
1 ml sample vial
1 ml
Polyethylene
1ml
Cap
Polyethylene
--
--
Both glass and plastic (polypropylene) sample vials are used. In case
of glass vials, the ionic substances, acids or bases may be adsorbed onto glass surface.
Thus yield is poor and reliability of the analysis is lost. To avoid the problem of adsorbtion
use:
10-100 mM aqueous Perchloric acid solution or mixture of Perchloric acid + Organic
solvent (acetonitrite, methanol, ethanol).
Organic solvent solution of 10mM trifluoroacetic acid (TFA).
Use Polyethlylene vials.
Some times in plastic sample vials, the hydrophobic properties of substances cause
surface adsorbtion results into poor results and reliability. Use mixture of water or buffering
solution with organic solvent (20:50 v/v).
Degasser
Degasser is a device removes
dissolved gases from liquids using
a
special
membrane.
degasser
synthetic
By
to
polymer
connecting
solvent
the
delivery
2.
Bubbling helium gas in a mobile phase in the reservoir and driving out the
dissolved air.
Water/methanol
1.5 ml/min
Water/acetonitrite
1.7 ml/min
1.
2.
3.
The signal is processed using the right and left portions of the photodiode as
individual elements. Data processing is performed as a two partitioned
photodiode divided into right and left. When refractive index in the cell varies, the
balance of the incident light intensity into right and left parts of the photodiode
changes. This change is converted into refractive index and can be recorded.
Signal processing is performed using upper and lower parts of the photodiode
elements. The refractive index is one-twentieth of mode A.
Ide
ntifi
cati
Use of retention time Rt along with absorbance spectra provides greater certainty in
anlayte identification.
Check of impurities (Eluted peaks can be checked to determine if they are generated
form single component or mixed with impurities.
Temperature of Flowcell:
If ambient temperature is 20-300C, set cell temperature to 400C
If ambient temperature is 30-350C, set cell temperature to 500
The fraction collector is a device to collect the fraction of a sample for further analysis
using either MS.NMR. X-rays and crystallation graphic analysis.
For 4 ml volume 144 fractions and 64 fractions are used for large fractionation.
The delay time from the detector to fractionation head is automatically calculated. It is
not necessary to calculate
it every time the flow rate is
changed.
10
When exceeding the volume of the vial, the collector automatically moves to the next
vial before it is overflown. So there is no worry for the loss of precious fraction.
Collection Method
There are two methods:
Unnecessary liquid in analysis if flowed in the drain tubing so that test tubes are not
wasted.
Liquid from the previous fractionation remains between inside of the valve and nozzle
tip 50 l. Therefore, a small amount of contamination may occur. Time programme is
set.
Contamination does not occur since liquid is collected into test tubes from the tube
connected to the detector or cell without passing through the flow line selector.
Liquid may be dispersed in moving from one test tube to another when the flow rate
is high.
11
12