Overview
Thechamberisaheavy
duty,clearpolyvinyl
chloride(PVC)enclosure
supportedbyanaluminum
frameandmountedona
plywoodbasecoveredwith
afoampadandheavyvinyl.
Itisequippedwitha
vacuumairlocksystemthat
allowsthetransferof
materialswhilemaintaining
aninternalanaerobic
atmosphere.
Theairlockisconnectedtotwogastankscontaining(i)100%N2and(ii)agasmix
of5%H2,85%N2,10%CO2.Standardsettingsfortheairlocksincludetwo
consecutivecyclesofvacuumandN2purgeforO2displacement,followedbyone
cyclevacuumandgasmixpurgetoequilibratetheatmospherewiththeinsideof
thechamber.
Vinylchambersaccuratelymaintainastrictanaerobic
atmosphereof05partspermillionO2.Internalatmosphereis
keptO2freewithacombinationofH2containinggasmixand
thepalladiumcatalyst.Thecatalystiscomposedofalumina
pelletscoatedwithathinlayerofpalladiumchloride.The
palladiumchlorideadsorbsH2fromthegasmixallowingitto
reactwithanyO2inthechambersatmospheretogenerateH2O,whichisthen
adsorbedbythealumina.Thecatalystiscontainedinametalwiremeshcontainer
(StakPak)thatishousedinafanboxtopromotegascirculation.
TheO2H2gasanalyzerconstantlymonitorsanddisplayslevelsin
thechambersatmosphere.Itiscalibratedtoproducevisibleand
audiblealarmswhenO2levelshaverisenaboveaset
concentrationorH2levelsareoutsideaconcentrationrange.
OperationandMaintenanceoftheGeneralUser
Transferringitemsintoandoutofthechamber:
1. Priortointroducingitemsintheairlock,checkthefollowing:
a. H2 levelsdisplayedonLEDoftheO2 H2 analyzerare~2%orabove
andO2levelsarewithindesiredrange(approachingorat0)
IfH2levelsarebelow2%orO2levelsareabove100ppmsee
GloveBoxMaintenancebelow.
b. Innerdoorofairlockiscompletelyclosedandlocked
2. Loadmaterialsthroughouterdoor
a. Sealedpackagesandtubesshouldbeopenedslightlytopermitgas
exchangeduringairlockcycles
b. Putobjectsinsideairlock,closeandlatchfrontdoor,andpressStart
button.Aftercycling(4exchanges),theairlockwillindicatethatitis
anaerobicandsafetoopentheinnerdooroftheairlock.
3. Transferringmaterialsoutofglovebox
a. Outerdoormustbelockedandairlockchambermustbeanaerobic
priortoopeninginnerdoor.Ifyouareunsureofthestatusofthe
airlock,performexchanges(i.e.,pressstart)tomakeitanaerobic.
b. Innerdoorcanbeopenedandmaterialscanbetransferredinto
airlockcabinet
c. Lockinnerdoortosealairlockcabinet
d. Outerdoorcanbeopenedtoremovematerials
e. ItisrecommendedthatouterdoortoairlockcabinetisLOCKEDAT
ALLTIMEStominimizeriskofO2intrusions
GeneralRulesforGloveBoxoperation:
1. Doors:Neveropenairlocksouterentrydoorbeforecheckingtoseeif
theinnerdoorissecurelyclosedandviceversa
2. Airlock:Alwayscheckthattheanaerobicdisplayisonbeforeopening
innerdoor
3. Sharps:Avoidusingsharpsinoraroundglovebox,ifnecessarykeep
Sharpsboxinchamber(Removeanysharpjewelrypriortouse)
4. Cleaning:useabsorbentbenchpaperinchambertocontainspillsand
debrisandtocleanworkspace(70%ethanolorotherdisinfectantscan
beused)
5. Emergency:Aroleofvinyltapeandscissorsshouldbekeptinornear
thechamberforemergencyrepair
6. Maintenance:
a. RegeneratecatalystStakPakinovenonceaweek
b. Periodicallyreplace~1/4ofthechambersatmospherewithgas
mixtomaintainoptimumH2levels
7. Economy:Minimizethenumberofchamberentriesbycoordinating
transferswithotherlabmembers
AsepticTechniquesinanAnaerobicChamberEnvironment
1. Noflamecanbeusedintheglovebox
a. LackofO2 preventsH2 ignition
b. IncandescentflamingdevicecanbepurchasedfromCoyif
necessary
2. Propercleaningofanaerobicchamberandairlocktoprevent
contamination
a. Priortowork,thechamberworkpad,instrumentationand
(especiallybetweenfingers)canbecleanedwith70%ethanolor
dampdisinfectantwipes.
b. Utilizedisposableitemswhenpossible(i.e.loops,spreaders,etc.)
c. Anyusedutensilsaredisposedofinawastecontainerandshould
beremovedfromtheglovebox,emptiedanddisinfectedafter
eachuse
3. Otherbasicrulesofsteriletechnique
a. Anymaterialshousedingloveboxforlongperiodsoftimeshould
beresterilizedfrequentlytopreventcontamination
b. Gloveboxshouldbecleanedperiodicallywithdisinfectantsaspart
ofitsregularmaintenance
AdditionalNotesforCoyUse
Gasanalyzeralarmgoesoff
1. LowH2 levels
a.ToincreaseH2concentrationinchamber,makesureinnerairlockis
anaerobic,openinnerairlockdoor,andtransferairlocktoManuel
modebypressingEnter/Menubutton,thenStartbutton.
b.Evacuate~1/4ofthechamberatmosphereusingthevacuum(^
button)
i. Deflateglovebox12inchesbelowthetopaluminumframe
ii. Replenishchambersatmospherewithgasmix(Start
button)untilinsidepressureissufficienttoinflatethebag
butnotthearmsleeves
iii.
RepeatasnecessarytoincreasetheH2levelsto23%
iv.
CloseinnerdooroftheairlockandpressEnter/Menu
buttontoexitManuelmodewhenfinished
c. Automaticgasinjectionsystemcanbeprogrammedtoreplacea
setvolumeofthechambersatmospherewithgasmixforafixed
timeperiodandatafixedfrequency
Timeouts
1.
Ifthereisatimeouterror,youcanusuallyidentifytheproblem
basedonwhatpartofthecyclingthetimeoutoccurred.
Iftheerrorhappenedduringavacuumcycle,oneofthedoorsis
probablyopen,socheckthatallouterandinnerdoorsare
completelyclosedandtryagain.
b. Ifthereisanerroronthepurgeorgasmixsteps,therespective
tankislikelyempty.Checkthelevelsindicatedonthetank
regulator,andifatankisclosetoempty,replaceit.
i. Thiswillalsohappensometimeswithanewtankif
someonehasforgottentoopenthetankortheregulator,so
checkthoseproblemsifthetimeoutpersists.
ii. Inmostcases,addressingthesepossibleissuesthen
turningtheCoyoffandbackonwillresolvetheproblem.
a.
Leaks
1.
IfthepressureinonesideoftheCoyisabnormallylow
a. Switchintomanualmodeandincreasepressureusingthegas
mixuntilpositivepressureisreached.
b. Ifthepressurehasdroppedthenextday,theremaybeasmall
holesomewhere,oftennearthegloves.
i. Usetheleakdetectorand/orsoapywatertoidentifythe
hole,thenpatchit.
Contamination
1. Openasolidifiedrichmediuminchambertocheckforpresenceof
contaminatingorganisms
i. Ifcontaminationissuspected,removeallequipmentand
cleanchamberwith50%IPA,12%Clorox
a. Routinelycleantherubbersealairlockdoorstoremovedustand
dirttopreventairleaks
LabJobDutiesforCoy
HeatcatalystplatesinovenunderCoy(150200oC)for>2hours,thenallow
platestocooltoapproximatelyroomtempandexchangewithplates
currentlyinsideCoy(weekly).
2. Checkoxygenlevelstomakesurethattheyareeitherat0ppmorare
steadilydecreasingtothat(daily).
a. Iflevelsaredecreasingtooslowly,exchangecatalystimmediately.
b. Ifoxygenstayshigh,checkforleaksinwallsofCoyanddoors.
3. Checkhydrogenlevelstomakesuretheyaremaintainedaround2.0%.
a. Iflevelsaretoolow,doanexchangetomakethetransferchamber
anaerobic,thenraisehydrogenlevelsbyswitchingCoyintomanual
mode,openingtheinteriordoorconnectedtothatside,and
repeatedlyalternatingcyclesofvacuumandgasmix.
b. Afteryouaredone,returnCoytostandardmodeandensurethat
1.
cyclingprogramisreturnedtoP1usingupanddownarrows(daily).
4. Throwawayleftoverplatesandcultures,especiallythosethatseemtobe
abandoned(crackedplates,driedoutcultures,>1monthold,etc.).
a. Ifyouwanttogivepeoplethechancetocleantheirownthings,give
themafewdaysnoticetomovethingstheywanttokeep,thentoss
anythingthatwasntmoved.Also,occasionallycheckandemptytrash
containers(asnecessary,everycoupleofmonths).
5. Checkhumidityindicatortoensurethathumidityis7080%.Ifnot,make
surethatdehumidifiersarentfull,andiftheyare,emptythem(asneeded,
everycoupleofmonths).
GloveBoxMaintenance
1. Catalyst
a. Palladiumcatalystiscontainedinametalwiremesh(StakPak)
i. Moisturecanberemovedbybakinginadryoven
(125200C)foraminimumof2hrs;regenerateoncea
week
ii. ReplaceStakPakevery1to2years
b. Testtomonitorcatalystperformanceistomeasuretheheat
producedduringthecatalysisreactionbetweenpalladium,H2and
O2
i. Placeatraycontainingthepalladiumcatalystinsidethe
airlock
ii. Placeathermometerindirectcontactwithcatalyst
iii.
Manuallyevacuategasatmospherefromairlock.
Vacuumpressureinsidetheairlockchamberisachieved
whenpressureinsidethechamberreads20in.mercury
units
iv.
Filltheairlockchamberwithgasmix(containingH2)
v. Ifcatalystisworkingcorrectly,thetemperaturewill
increaseatleast10Coveraperiodof1015min
2. Moisture
a. ExcessivemoisturedecreasesH2 levelsinchamberatmosphere
andsaturatesthecatalyst,increasinggasconsumption
b. Minimizehumidity
i. Coysareequippedwithdehumidifier
ii. Regeneratecatalystmoreoftenthanrecommended
schedule
iii.
Incubatesolidmediaplatesinairtightfoodcanisters
containingapetridishwithaluminadesiccantpellets
iv.
AvoidlongtermstorageofplatesandculturesinCoy
c. Ifmoisturedevelopsinthechamberandcondensationbecomes
noticeableonvinyl,removeimmediatelywithabsorbentpaperto
preventfungalgrowth
3. SharpsDisposal
a. Allsharpsshouldbedisposedofapprovedsharpscontainers
b. Operatorshouldremoveanysharpjewelrythatcouldpuncture
glovesand/orsleeves
4. Hygienicmeasures
a. Cleantheoutsidevinylofthegloveboxbeforeeachuse
b. 70%ethanolforworkpadandgloves
c. 50%I.P.A.or12%CloroxforPVC
GloveBoxTroubleshooting
Problem
O2H2analyzeralarm
PossibleCause
LowH2levels
Solution
ReplenishH2levelsinchambers
atmospherewithgasexchange
AirlockDoors
Ensuretheinnerandouterdoors
areproperlyclosedandthereis
nodirtbuilduponseals
Removeanyobjectsblockingthe
catalystand/orfan
Fanboxesareblocked
CatalystStakPak
ReplaceStakPaks(theyare
eithersaturatedwithmoistureor
dontwork)
Punctureinglovebox
Inflategloveboxwithgasmix
untilglovesarefullinflated.
CheckforleakswithH2detector
andrepairpunctures
Faultyregulators
Checkforleakswithsoapy
solution.Redoconnectionsor
replaceregulator
Leaksinconnectionstoairlock
Checkforleakswithasoapy
solutionorgasleakdetectorfor
H2containinggases
RegardlessofthesourceofO2,oncerepaired,purgeofthechambers
atmosphereandreplacewithgasmix.RepeatuntiloptimalH2levelsarereached.
Condensation/DeHumidifier
CatalystStakPak
DesiccantStakPak
ExchangecatalystStakPakwith
rejuvenatedones(theyeitherare
saturatedwithmoistureordont
work)
ReplacedesiccantStakPakwith
Experimentreleasingexcess
moisture
rejuvenatedones(theyareeither
saturatedwithmoistureordont
work)
Emptydehumidifiers
ReplacedesiccantStakPakwith
rejuvenatedonesandaddmore
desiccantStakPaksifneeded
Oncerepaired,purgeofthechambersatmosphereandreplacewithgasmix.
Repeatasnecessary.
Contamination
Improperlysterilizedmaterials
Resterilizeallsterilematerials
houseinchamber
Contaminatedcultures
Discardoldorcontaminated
cultures.Placeplatesinsealed
canisterstoprevent
crosscontamination
Disinfectthechamberandallsurfaceswithadisinfectant(I.P.A.,Clorox).Cleanwith
peraceticacidtosterilizechamber.
Bagpressuretoolowortoohigh Autoinjector
Airlockerror/Timeouterror
Testand/orreprogramthe
autoinjector.Ensurevacuumis
functioningproperly
Leaksinvinylbagorneoprene
rubbergloves
Detectleakswithgasleak
detector.Sealpunctures.
Replaceglovesifnecessary
Vacuum/gasexchangecycles
Usingtheairlock,evacuateor
injectgasasneededtobringthe
bagtotheappropriatepressure
Gastanks
Checkthatthetanksareturned
onandnotempty,andregulators
areonandfunctioningproperly
WorkingPSIshouldbe2025(on
regulator)
Airlockchamber
Ensuretheinnerandouterdoors
areproperlyclosedandthereis
nodirtbuildupontheseals
Vacuum
Ensurethevacuumisfunctioning
properly.Checkconnectionsto
vacuumpump.Replaceorfixas
needed
Onceproblemisdiagnosed,applyvacuumoraddgasasneededtoequalize
pressureintheairlocksothedoorscanbeopened.