Académique Documents
Professionnel Documents
Culture Documents
POLAROGRAPHY
5.1. Polarographic methods
In direct current polarography (DCP) a constant potential is applied during the entire
drop-life time. A current-voltage curve is constructed by applying a series of potential steps, each
step being synchronized with the drop fall. In most instruments, however, linearly changing
potential is applied, with a rate slow enough that the change of potential throughout the drop-life
time is about a few millivolts. The current is measured at the end of the drop life.
In normal pulse polarography (NPP) the mercury-drop electrode is held for most of its
duration at a constant potential Ein, at which no electrochemical reaction takes place under given
experimental conditions. The potential of interest Ep is applied in the last stage of the drop life,
for a length of time tp (of the order of few milliseconds). The values of Ein and tp are kept
constant throughout the recording of the polarogram and E p is changed from drop to drop (Fig.51a).
The limiting current in NPP is diffusion controlled. The experimental requirements for
diffusion control are the same as those for DC polarography. Since t p is of the order of
milliseconds, the diffusion layer thickness is very small compared to the radius of the mercury
drop reached at the end of its life. The equations for planar diffusion can be applied with much
better agreement than for DC polarography (for t p = 1 ms, the diffusion layer thickness is about
210-4 cm, while the radius of the mercury drop is about 0.05 cm at t = 5 sec and m = 1 mg/sec).
Furthermore, the area of the drop is virtually constant during the application of the pulse (t is
much larger than tp). The constancy of the area implies also that no correction factor for the
expansion of the diffusion layer is required (the (7/3)1/2 factor). Thus,
(1)
For comparison, the current for DC polarography at time t is
(2)
The current under NP conditions is larger than that under DC conditions. The enhancement of the
current is
(3)
For typical conditions, t = 5 s and tp = 5 ms, this ratio is about 20. The contribution of the
charging current is identical for both NPP and DCP, and thus the detection limit of NPP is lower
than that of DCP by an order of magnitude.
The i/E relationship for Nernstian processes at DCP and NPP is given by
(4)
The detection limit of NPP is about 210-7 M.
Differential pulse polarography (DPP). From analytical point of view, the sensitivity of
DPP is even better than that of NPP. The potential sequence on a single mercury drop and the
potential sequence used for recording an entire differential pulse polarogram are given in Fig.5-1.
The current is sampled twice during a drop life-time: (i) at t 1, just before the pulse, and (ii) at t,
just before the drop fall. The polarogram represents the current difference
as
function of the base potential E b. The curve is peaked shaped. The height of the peak is
proportional to the concentration of the electroactive species.
For Nernstian processes Ox + ne = Red and C* red = 0, the faradaic component of the
current at the peak
(5)
(6)
The peak form of the DP voltammogram is explained below. At sufficiently positive
potentials, Eb - E0 > (120/n) mV, the faradaic currents
also
is zero. At sufficiently negative base potentials, E b - E0 < (120/n) mV, the faradaic
process proceeds at maximum rate and the diffusion-limited current is reached; the current is
independent of potential and
difference
is determined by
the current
peak.
Although (DiF)peak is smaller than the limiting current at NPP, the detection limit at DPP is
lower (about 10-7 M) due to the efficient compensation for the charging current.
Square wave polarography (SWP). High sensitivity and low detection limit are
obtained with this technique. It is similar to DPP, however, the entire potential sequence is
applied during the life-time of a single drop. The voltammogram is obtained in a few seconds,
compared to much longer times with the other techniques.
The potential sequence is applied several seconds after the drop birth, in order to take
advantage of a larger surface area of the mercury drop.
As in the case of DPP, the difference of the current before and after the application of the
pulse is measured. The performance of the method is better for Nernstian processes than for nonNernstian ones. Detection limit of the method is about 10-7 M.
5.2. Handling the dropping-mercury electrode (DME)
The mercury capillary may be easily clogged and care must be exercised to prevent
contamination. The DME must never be allowed to stand in solution when mercury is not
flowing. Before allowing any solution to come in contact with the DME, raise the leveling bulb
and check that mercury drops are formed at the end of the capillary. Allow the mercury flow
during the entire laboratory session. After use, the capillary should be washed thoroughly with
distilled water. The mercury reservoir is then lowered to stop the flow.
Note: BE AWARE! Mercury vapors are poisonous!
Notify the instructor in the event of mercury spill. Mercury should be
cleaned up immediately. Do not throw it down the drain.
Reference electrode
An Ag/AgCl/1 M KCl reference electrode is used. Its potential is -19 mV vs SCE.
Chemicals
1.
2.
3.
4.
5.
6.
7.
8.
1 M KNO3
0.1 M KNO3
5.00 mM ZnCl2
0.2 % Triton X-100
0.01 M TlNO3
0.01 M KIO3
0.1 M NaCl
0.1 mM Tl+, 0.1 mM Cd2+, 0.1 mM Zn2+, 0.1 mM Ni2+, 0.002 % Triton X-100
in 0.1 M NH4OH - 0.1 M NH4Cl buffer
9. 0.5 M acetate buffer (pH 4.6)
10. 0.2 % ascorbic acid (freshly prepared)
(a) to verify that the limiting current of the TI+ - reduction polarographic wave is diffusion
controlled;
(b) to determine the half-wave potential and the number of electrons involved in the reduction.
Prepare 0.5 mM solution of TlNO3 in 0.1 M KNO3. (The Tl+ concentration has been
chosen low enough in order to neglect potential drop in the solution). Run a polarogram in order
to choose a potential in the limiting current region. Record a current/time plot at the chosen
potential. Prove, that the current is totally diffusion controlled, and calculate the diffusion
coefficient, assuming that m = 1 mg/sec.
In order to enable an accurate measurement of the currents along the wave, run an
additional polarogram at a slow scan rate (5 mV/s).
Determine E1/2 of the Tl+ reduction wave and compare with literature.
Determine the number of electrons involved in the reduction process from E vs log[(id i)/i] plot.
Exp. 7. Quantitative determination of Zn2+ in drinking water
Fill the cell with the sample of drinking water and record a polarogram. You may prefer
to use one of the more sensitive polarographic methods (normal pulse, differential pulse or
square wave). Make a rough estimation of the concentration of Zn2+.
For the analytical determination use both the standard addition method and the calibration
curve method. Show the instructor your detailed plan of operation, including the composition of
the solutions. Prepare the standard solutions in 50 ml volumetric flasks. For the calibration curve
method use as the supporting electrolyte the solution of 10 mM NaCl. Compare the results of the
two methods of determination.
Exp. 8. Determination of ascorbic acid (Vitamin C) in citrus juice by the standard-addition
method.
Ascorbic acid yields a well-defined polarographic oxidation wave. The determination can
be carried out directly in freshly prepared diluted juice as well as in conserved citrus juice. The
method of standard additions is used for the quantitative determination of the concentration of
the Vitamin C.
Ascorbic acid
Procedure
1.
Calibration curve
Dehydroascorbic acid
mV/s) and SW (
, tp = 20 ms, pulse amplitude = 20 mV, scan rate = 100 mV/s)
polarograms over the potential range: -150 to +200 mV vs. Ag/AgCl/1 M KCl, with E in = -150
mV. Consult with the instructor about the preferable mode of polarography to be used in further
experiments.
Plot id vs. concentration of ascorbic acid. Is the plot linear and does it pass through the
origin? On the basis of these observations decide if the standard-addition method is applicable.
2.
Determination of ascorbic acid in citrus fruits
Squeeze an orange, grapefruit or lemon until about 10 ml of juice is obtained. Filter the juice
through a porous funnel (pore size about 1 mm).
Prepare four 25 ml volumetric flasks. Add to each, 0.5 ml of 0.5 M acetate buffer, 2.0 ml of
the juice and standard additions of 0, 200, 400 and 600
of 0.2 % ascorbic acid. Dilute to the
mark with distilled water.
Record polarograms under the same conditions as in the calibration step.
Draw the standard additions plot and determine the concentration of the analyte. Report the
concentration of ascorbic acid (Vitamin C) in the original sample (juice) in mol/l and ppm.
3.
Go to Main Page
6. STRIPPING VOLTAMMETRY
(b)
(c)
A short-time electrolysis (30 sec to 5 min) in a stirred solution and at a potential suitable for the
reduction of the ions of interest (E -E1/2 about -200 mV) may result under proper conditions in a
fairly concentrated amalgam. This step is called the electrodeposition step. The concentration of
the metal ion in the film depends on concentration of Mn+ in solution, time of electrolysis and
rate of stirring. Since the electrodeposition is carried out on small electrodes, the amount of
material deposited into it usually does not change significantly the concentration of the metal
ions Mn+ in the solution. Step 1 is only an intermediate step and there is no need to know the
concentration reached in the amalgam. However, it can be estimated.
The enrichment of the metal M in the amalgam in respect to the initial concentration of
M , CMn+, is estimated in stirred solutions, using Nernst simplified model (eq.3 in
section"Coulometric titrations").
n+
For typical laboratory conditions the Nernstian layer thickness in well-stirred solutions is about
20
.
As a result of electrolysis for a time t, the concentration of M in the amalgam, CM(Hg), is
where A and V are the area and the volume of the mercury electrode.
The enrichment of the metal in the amalgam, using D = 10-5 cm2/s, is
As a result of a 100 sec electrolysis, the concentration of the metal in the amalgam will be
~5104 times larger than that of the metal in the solution.
The enrichment factor for a hanging mercury-drop electrode is 1000 times smaller than
that for a thin mercury film. Nevertheless, substantial increase in concentration is achieved also
with this type of electrode for electrolysis time t larger than 100 sec.
The degree of decrease of concentration in solution as result of the electrodeposition step
is calculated for a mercury film electrode for the following conditions: the area of the film, A =
0.2 cm2; the initial concentration of the metal ions in the solution, C Mn+ = 10-8 M; the volume of
the tested solution is 20 ml; the time of deposition is 100 s; the Nernstian layer thickness during
the deposition step is 20
. At the end of the electrolysis step the number of moles of the metal
electrolyzed in the film, CM(Hg) = 10-12 mole (verify this value!). Thus, the decrease of
concentration of Mn+ in solution after the electrolysis is negligible (0.5 %).
Step 2. Rest period
After a predetermined time, the stirring of the solution is turned off. The solution is
allowed to become quiescent and the concentration of the metal in the amalgam - to reach
uniformity. The rest period extends for about 30 sec, during which the applied potential remains
unchanged, thus ensuring that no reoxidation of the metal by traces of oxygen takes place.
During the rest period the electrodeposition current decreases (Explain why).
Step 3. Stripping
After the preconcentration step, the deposited metal M is oxidized ("stripped") from the
mercury electrode back into the solution by oxidation to the ionic form under conditions of
diffusion control, using one of the voltammetric methods:
The anodic diffusion current is used to determine the concentration of the metal in the amalgam,
which is proportional to time of electrolysis, stirring rate and concentration of Mn+ in solution.
Sample preservation
Drinking water samples should be preserved by adding 10 ml of 1 M HNO 3 to each liter
of sample immediately after collection. Acidification to a pH between 1 and 2 is an essential first
step in water sampling: a) it arrests biological growth; b) it affects the release of trace metals
from organic components (humic and fulvic acids) and c) it reduces metal loss caused by
adsorption on the container wall.
Sample pretreatment
The presence of surface-active compounds in the samples consists a major problem in the
use of solid electrodes. Surfactants compete with the analyte on the electrode surface for active
adsorption sites, causing distortion, lower sensitivity and instability of the signal of the analyte.
The hanging mercury-drop electrode, owing to its renewable surface, is inherently less sensitive
than solid electrodes to surfactants.
Severe peak distortions are overcome with different methods: UV irradiation 2 of the
sample, coating the electrode with permselective film or membrane that prevents the surfactant
from reaching the electrode surface3, addition of fumed silica to the analyte 4, wet digestion and
others.
Exp.1. Determination of Pb2+ and Cd2+ in drinking water, using hanging mercury-drop
electrode
Chemicals
Electrodes:
working - HMDE
reference - Silver wire (1.5 mm diameter)
t = 120 s
E = -900 mV
t = 60 s
t=5s
Potential scan = from -900 to -200 mV
Stirring rate
800 rpm
Test of reproducibility. Repeat several times the basic procedure in the same solution. Report the
average and the standard deviation of the slope.
Effect of electrolysis time. Repeat the stripping procedure with different electrolysis times: from
30 to 120 minutes. Note that the signal is linearly changed with the time.
Calibration curve. A calibration curve for the analyzed metal is constructed for two reasons: (a)
to enable rough estimation of the concentration of the metal in the unknowns by comparing peak
heights, and (b) to test if linearity between analytical signal and concentration is obtained in the
concentration range if interest. This is a basic condition for applying the standard-additions
method, used for the analytical determination in this experiment.
Insert into the cell 5.00 ml of the supporting electrolyte. Record a stripping curve of the blank
following the basic procedure. Repeat the experiment with five subsequent additions of 0.05 ml
of the 10-6 M Pb2+ solution. Plot the calibration curve.
Exp. 2. Determination of Pb2+ in drinking water, using the silver rotating-disk electrode
Fig.6-1 The rotating-disk electrode. The electrode and the insulated material are marked by
filled and clear areas, respectively.
Since the flow is laminar, it is possible to calculate rigorously the rate of mass transport.
The rotating-disk electrode was developed following the mathematical solution given by Levich
of the hydrodynamic equations describing the rate of transfer of substance in solution to a
rotating disk surface, in terms of the angular velocity of rotation
(
, N in rps), the
0
diffusion coefficient D, the concentration C of the substance and the kinematic viscosity of
the solution. For the case when the reaction is relatively fast and the current is determined by
mass transport, the corresponding equation for the limiting current density
Levich, is:
, developed by
where is in A/cm2, D in cm2/s, in cm2/s , in rad/s and C0 in mol/cm3. In such a case the
limiting current is independent of potential over a wide range. This range of potential is limited
at one end by the reversible potential and a small overpotential needed to drive even a very fast
reaction to mass-transport limitation and at the other end by another reaction which may take
place, usually the evolution or oxygen or hydrogen in aqueous solutions.
Experimental procedure
The silver rotating-disk electrode4 is well suited for the determination of lead in ppb and
sub-ppb concentrations.
A two-electrode cell configuration is used (cf., Fig.6-2). The working electrode is a silver
rotating disk. A silver wire acts as a combined counter/quasi-reference electrode. The potential of
the silver wire is stabilized by the presence of chloride ions in working solution.
Handling of the silver disk electrode. The silver disk electrode can be used over
hundreds of measurements without any pretreatment, provided the surface-active agents in the
samples are destroyed. The proper functioning of the electrode is judged according to two
criteria:
(a) the reproducibility of the anodic stripping voltammograms;
(b) the degree of linearity between the analytical signal and the concentration.
If the criteria are not fulfilled to our satisfaction, the electrode should be polished with
0.05 mm alumina powder, rinsed thoroughly with distilled water, sonicated for 3 min for removal
of alumina particles and rinsed again with distilled water.
Chemicals
Electrodes:
alumina powder
E = -40 mV
t = 10 s
E = -700 mV
t = 30 s
t=3s
The analytical signal can be quantified by: a) the stripping peak current, b) the stripping
peak area, and c) the average slope at the two inflection points of the peak. Use the slope for
quantitative evaluation of analyte, since this parameter yields the best linear response vs.
concentration.
3. Precision. The precision of an analytical method is the degree of mutual agreement
among results obtained under identical conditions. Statistically, the precision is expressed as the
relative standard deviation of a set of measurements.
Run five subsequent voltammograms and calculate the average and the relative standard
deviation of the slope.
4. Effect of rotation rate. Repeat the stripping procedure for different rotation rates
(2000 - 8000 rpm). Plot the analytical signal as function of the square root of the rate of rotation.
5. Effect of electrolysis time. Repeat the stripping procedure with different electrolysis
times: from 15 to 120 seconds.
6. Background correction. Traces of impurities in the reagents contribute to the
background and should be taken into account. Pipet into the cell 5.00 ml of the combined 10 mM
HNO3and 10 mM NaCl solution and run a stripping voltammogram in the supporting electrolyte
only. Note the value of the analytical signal of lead in the background (the slope) and use it later
to correct the calibration curves.
7. Calibration curve. A calibration curve in the range of 0 to 810-8 M of Pb2+ is
constructed for two reasons: (a) to enable rough estimation of the concentration of Pb 2+ in the
unknowns by comparing peak heights, and (b) to test if linearity between analytical signal and
concentration is obtained in the concentration range if interest. This is a basic condition for
applying the standard-additions method, used for the analytical determination in this experiment.
Insert into the cell 5.00 ml of the combined supporting electrolyte. Record a stripping
curve of the blank by following the basic procedure, but increase the deposition time to 60 s.
Repeat the experiment with five subsequent additions of 0.1 ml of the 1.0010 -6 M Pb2+ solution.
Plot the calibration curve.
For the unknown samples (a) and the blank (b) perform a quantitative determination
based on standard-additions method. Run stripping voltammogram of the sample. Make a rough
estimation of the lead concentration by comparing the analytical signal of the unknown with
those of the calibration curve. If the agreement between the duplicates is satisfactory, perform the
standard-additions procedure on one of the duplicates only. Keep the following rules in respect
with the standard additions: (a) the volume of a single standard addition should be smaller than
5% of the volume of the solution in the cell; (b) the increase in concentration with each addition
should be lower than 40% of the concentration of lead, roughly estimated by consulting the
calibration curve, recorded under identical conditions. Calculate the concentration of lead in the
digested unknown sample and in the digested blank.
Calculate the concentration of lead in the drinking water sample. Report the
concentration of the analyte in mol/l and ppb. Compare your results with the official permissible
levels.
Lead compounds have been and are still used in the manufacture of ceramic glazes and crystal
glasses.
Certain compounds of lead, particularly brightly colored lead oxides, have been used for leaded
glasses and leaded glazes on ceramics for thousands of years. The addition of lead improves the
appearance and cutting properties of crystal glass. Small amounts of lead are also present in
optical glass. The major application of leaded glass is in television screens and computer
monitors, and serves to protect viewers from the harmful X-rays generated by these appliances.
Lead-containing glazes are used in some pottery, tiles and tableware.
Crystal glasses contain up to 30 % lead oxide, PbO. The European definition of lead crystal as
glass that contains a minimum of 24% lead oxide is a feature that once contributed to its
reputation for quality. The purpose of lead oxide is to increase the density of glass, which in turn,
increases the refractive properties. When it is cut and polished, it creates more colors and sparkle
than regular glass. On contact with water or other solutions, the lead is slowly released in trace
amounts, especially in acidic solutions.
Electrodes:
Chemicals
alumina powder
Experimental procedure
The same configuration of the electrochemical cell is used (see Exp. 2).
For the preparation of the sample, heat about 100 ml of deionized water in a glass beaker until
the water boils, add 1 ml of 10 mM HCl (pH 4). Transfer the hot water to a glazed ceramic vessel
or to a crystal glass, wait for about 20 min. Use this solution for determination of the degree of
contamination by Pb2+.
Follow the basic procedure of the anodic stripping (see Exp. 2) for recording voltammograms for
blank and analyte solutions.
Perform a quantitative determination using the standard-addition method. Consult with the
instructor about the details, concerning the volume of the sample and the standard additions.
Estimate the values of the analytical signal from the voltammograms. Plot the analytical signal
corrected for background vs. concentration of the metal in the cell. Report the concentration of
the metal in the sample in mol/l and ppm.
References
1. C. M. G. Van den Berg, Anal. Chim. Acta, 1991, 250, 265.
Recommended Literature
1.
2.
3.
4.
(a)
Accumulation step, in which the metal ion is accumulated on the electrode as an
adsorbed metal complex.
A surface-active complexing agent is added in the solution (at a concentration not greater than 1
mM). The ligand forms an adsorbed layer on the electrode surface (usually on a hanging
mercury-drop electrode) as soon as it is introduced into the solution. The analyte (at nanomolar
or sub-nanomolar concentration) reacts with the adsorbed ligand to form an adsorbed complex.
This process is carried out in stirred solution and under potential control. The analyte reaches the
electrode by convection (as in the preconcentration step of anodic-stripping voltammetry).
(b)
Stripping the adsorbed metal ion, which is realized by electroreduction of the
adsorbed complex.
Square-wave voltammetry is used in this step. The amount of adsorbed metal is determined from
the value of the peak current or from the sum of the slopes at the inflection points of the peak.
The ligands used in AdSV are surface-active compounds or are able to interact
chemically with mercury (the presence of -electrons benefit the adsorption process and the
presence of S donors helps the chemisorption process). The kinetics of the complex formation
must be fast.
In this experiment the ligands are dimethylglyoxime (DMG) and 8-hydroxyquinoline
(oxine). In some applications the use of mixture of ligands enables simultaneous determination
of up to six trace metals.
The detection limit of AdSV depends on the chemistry of the system and can be as low as
10 M.
-12
organic matter in the samples are: addition of fumed silica (adsorbs the surface-active
substances), ultraviolet irradiation, wet and dry ashing.
Cleanliness of glassware
The glassware used is especially cleaned for this highly sensitive analysis. Prior to use
the glassware is immersed for a few hours in a liquid cleaner, followed by immersion in 1:1
aqueous HNO3 and then rinsed with copious amounts of deionized water.
It is recommended to use dedicated vessels for the various analytes and for different
concentrations.
In order to minimize contamination, the amount of glassware and plastic ware should be
reduced as much as possible.
This experiment follows the work of C. M. G. van den Berg et al 1. Van den Berg
demonstrated the advantages of using mixture of ligands in AdSV. With DMG
(dimethylglioxime) as ligand only Ni and Co are determined by adsorptive stripping
voltammetry 2,3. Addition of oxine (8-hydroxyquinoline) enables simultaneous determination of
several metals: Cu, Pb, Cd, Ni, Co and Zn. The concentration range depends on the species
present and is roughly from sub-nanomolar up to about 50 nM. At higher concentrations of the
metals deviation from linearity may be observed. Thus, an excess of one of the analytes may
interfere with the analysis. For this reason, drinking tap water, which normally contains
micromolar concentration of Zn, cannot be analyzed by using mixture of ligands, the content of
Ni and Co in tap water can be determined using DMG only.
(In this experiment the HEPES buffer has been replaced with an ammonia buffer).
The method of standard additions is used for the quantitative determination of the metal
concentrations.
Chemicals
Electrodes:
t = 120 s
E = -200 mV
t = 30 s
Rest step
t = 10 s
technique SWV
E = from -200 to -1400 mV
800 rpm
oxine.
Experimental procedure
) in 10 mM HCl.
Add 100
of each 0.02 ppm solution to the cell containing the blank. Run a stripping
voltammogram. Well defined peaks of Ni and Co should appear.
c) Adding Cu2+, Pb2+, Cd2+ and Zn2+
Following section 1b, add to the cell 100
of a freshly prepared solution containing 1
of
2+
2+
2+
2+
each of Cu , Pb , Cd and Zn in 10 mM HCl. Record a stripping voltammogram. Do you
observe any change as result of the presence of the added metals?
d) Adding oxine
Add to the above cell 50
of 1 mM oxine solution. Record a stripping voltammogram. Note
that the peaks of all the species appear only in the solution containing the mixture of ligands.
If the signal is in the range of the calibration curve, proceed by the standard addition
method.
(b) Always perform first a measurement of the analytical signal in the background. For
obtaining results with reasonable accuracy the signal of the unknown should be at least twice the
value of the background.
4. Applications
4.1. Determination of contamination of water by Ni2+ as a result of contact with stainless steel
Stainless steel being in contact with hot water release traces of Ni. The contamination of Ni is
higher in presence of acids.
This experiment can be performed with any stainless steel equipment: any kind of vessels,
heating elements for preparing water for hot drinks, kettles and pots with stainless steel parts,
etc.
To prepare the samples, heat 100-200 ml of deonized water containing 10 mM NaCl (the
addition of the salt is to simulate drinking water). Leave the hot water in contact with the
stainless steel for about 20-40 min. Use this solution to determine the amount of Ni 2+ leached
from the stainless steel.
Add to the cell 5 ml of deionized water, 1 ml of 0.5 M ammonia buffer, 0.1 ml of 0.01 M DMG.
Record the stripping voltammogram of the background.
Add to the cell 0.05 2.0 ml of the analyte solution and record a voltammogram. The aliquot of
the unknown is determined by the concentration of the analyte: the larger the concentration the
smaller the portion of added sample (cf., note 3.1.). Make at least two standard additions, using
the proper diluted solution of Ni2+ (0.02 ppm or 0.2 ppm) and run the voltammograms again.
Consult with the instructor about the volume of the standard additions.
Estimate the values of the analytical signal from the voltammograms. Plot the analytical signal
corrected for background vs. concentration of the metal in the cell. Report the concentration of
the metal in the sample in mol/l and ppb. What is the pH of the analyte sample?
4.2. Determination of contamination of water by Ni2+ as a result of contact with a nickel strip
Prepare a sample by inserting a Ni strip (total area about 60 cm2) into a 100 ml glass beaker filled
with deionized water. Wait for about 20-30 min. It is possible to use hot water for shortening the
waiting time. Use this solution to determine the amount of Ni2+ leached from the strip.
Add to the cell 5 ml of deionized water, 1 ml of 0.5 M ammonia buffer, 0.1 ml of 0.01 M DMG.
Record the stripping voltammogram of the background.
Add to the cell 0.050-0.100 ml of the analyte solution and record a voltammogram. Make at least
two standard additions, using the 0.2 ppm solution of Ni2+ and run the voltammograms again.
Consult with the instructor about the volume of the standard additions.
Estimate the values of the analytical signal from the voltammograms. Plot the analytical signal
corrected for background vs. concentration of the metal in the cell. Report the concentration of
the metal in the sample in mol/l and ppm.
4.4. Contamination of water by Pb2+ as a result of contact with glazed ceramic vessel
Lead compounds have been and are still used in the manufacture of ceramic glazes and crystal
glasses. Certain compounds of lead, particularly brightly colored lead oxides, have been used for
leaded glasses and leaded glazes on ceramics for thousands of years. The addition of lead
improves the appearance and cutting properties of crystal glass. Small amounts of lead are also
present in optical glass. The major application of leaded glass is in television screens and
computer monitors, and serves to protect viewers from the harmful X-rays generated by these
appliances. Lead-containing glazes are used in some pottery, tiles and tableware. On contact with
water or other solutions, the lead is slowly released in trace amounts, especially in acidic
solutions.
On a hot plate, heat about 100 ml of deionized water in a glass beaker until the water boils, add 1
ml of 10 mM HCl (pH 4). Transfer the hot water to a glazed ceramic vessel, wait for about 20
min. Determine the degree of contamination by Pb2+ in this solution.
Add to the cell 5 ml of deionized water, 1 ml of 0.5 M ammonia buffer, 0.1 ml of 0.01 M DMG
and 50
Add to the cell 0.1 0.2 ml of the analyte solution and record a voltammogram. Make at least
two standard additions, using the 1
solution of Pb2+ and run the voltammograms again.
Consult with the instructor about the volume of the standard additions.
Estimate the values of the analytical signal from the voltammograms. Plot the analytical signal
corrected for background vs. concentration of the metal in the cell. Report the concentration of
the metal in the sample in mol/l and ppm.
On a hot plate, heat about 100 ml of deionized water in a glass beaker until the water boils, add 1
ml of 10 mM HCl (pH 4). Transfer the hot water to a crystal glass, wait for about 20 min. Use
this solution as analyte for the determination of Pb2+.
Add to the cell 5 ml of deionized water, 1 ml of 0.5 M ammonia buffer, 0.1 ml of 0.01 M DMG
and 50
Add to the cell 0.1 0.2 ml of the analyte solution and record a voltammogram. Make at least
two standard additions, using the 1
solution of Pb2+ and run the voltammograms again.
Consult with the instructor about the volume of the standard additions.
Estimate the values of the analytical signal from the voltammograms. Plot the analytical signal
corrected for background vs. concentration of the metals in the cell. Report the concentration of
the metal in the sample in mol/l and ppm.
4.6. Determination of traces of Pb2+ released in water as a result of contact with brass parts
used in water-delivery lines
Brass is an alloy of copper, containing mainly copper and zinc. It is used in connecting parts of
water-delivery lines. Small amounts of lead are sometimes added to improve the mechanical
characteristics of the alloy.
Lead rarely occurs naturally in water. Most lead contamination takes place at some point in the
water delivery system. This occurs as a result of corrosion, the reaction between the water and
lead in parts of the water-delivery system. Components of water-delivery systems which may
contain lead include service connections, pipes, solder and brass fixtures. Lead in drinking water
from plumbing is most often a problem in either very old or very new houses. Many household
faucets, plumbing fittings, check valves and well pumps are manufactured with brass parts.
While brass contains some lead to make casting easier and the machining process more efficient,
the lead content of brass plumbing components is now restricted to a maximum of 8 %. Even at
this low level, however, lead can be leached from new brass faucets and fittings. Eventually, if
the water is not corrosive, hard-water minerals deposit on the interior of the plumbing. These
deposits form a calcium carbonate lining inside pipes and fittings, which protects against lead
contamination. It may take up to five years for an effective calcium carbonate lining to form.
Softening naturally hard water with an ion-exchange water softening unit can either prevent or
dissolve the calcium carbonate scale, eliminating its possible protective effect.
Place in a beaker a brass connector from a water-delivery line and fill with deionized water at
room temperature. Leave over night. Use this solution as analyte for the determination of traces
of Pb2+.
Add to the cell 5 ml of deionized water, 1 ml of 0.5 M ammonia buffer, 0.1 ml of 0.01 M DMG
and 50
Add to the cell 0.1 0.2 ml of the analyte solution and record a voltammogram. Make at least
two standard additions, using the 1
solution of Pb2+ and run the voltammograms again.
Consult with the instructor about the volume of the standard additions.
Estimate the values of the analytical signal from the voltammograms. Plot the analytical signal
corrected for background vs. concentration of the metals in the cell. Report the concentration of
the metal in the sample in mol/l and ppm.
References
1. Carlo Colombo, Constant M. G. van den Berg, Anal. Chim. Acta, 1997, 337, 29-40.
2. Miropi G. Paneli and Anastasios Voulgaropoulos, Electroanalysis, 1993, 5, 355-373.
3. S. B. Adeloju, A. M. Bond and M. H. Briggs, Anal. Chim. Acta, 1984, 164, 181.