Académique Documents
Professionnel Documents
Culture Documents
AMVn
Automated Microviscometer
'ALILEO 'ALILEI
-EASURE
WHAT IS MEASURABLE
AND MAKE MEASURABLE
THAT WHICH IS NOT
Instruction Manual
AMVn
Automated Microviscometer
Firmware Version: 1.7x
Anton Paar GmbH assumes no liability for technical or printing errors or omissions in this document.
Nor is any liability assumed for damages resulting from information contained in the document.
Anton Paar GmbH reserves the right to content changes. This also extends to changes to delivery
volumes or any features of delivered parts.
All rights reserved (including translation). This document, or any part of it, may not be reproduced,
changed, copied, or distributed by means of electronic systems in any form (print, photocopy, microfilm
or any other process) without prior written permission by Anton Paar GmbH.
Trademarks, registered trademarks, trade names, etc. may be used in this manual without being marked
as such. They are the property of their respective owner.
Published by Anton Paar
Printed: Anton Paar, Austria
Copyright 2008 Anton Paar GmbH, Graz, Austria
Address:
Date:
July 2008
Contents
1 About the Instruction Manual ......................................................................................................... 8
2 Safety Instructions........................................................................................................................... 9
2.1
2.2
Liability ..................................................................................................................... 9
2.1.2
2.1.3
2.1.4
Disposal.................................................................................................................. 10
2.1.5
Returns................................................................................................................... 10
2.1.6
3 AMVn - An Overview...................................................................................................................... 12
4 Unpacking and Checking the Supplied Items ............................................................................. 13
4.1
4.2
4.2.2
4.2.3
4.2.2.1
4.2.2.2
5 Functional Components................................................................................................................ 17
5.1
Front of AMVn........................................................................................................................ 17
5.2
5.3
5.4
6.1
6.2
6.2.2
6.2.3
6.2.4
6.2.5
6.2.6
Measurement................................................................................................................................. 23
7.1
Setting up AMVn.................................................................................................................... 23
7.2
B80IB01
7.2.1
7.2.2
7.2.3
7.3
7.4
7.5
8.2
8.3
8.1.2
8.1.2.2
8.1.3
8.1.4
8.1.5
Software Update..................................................................................................... 31
8.1.6
Method ................................................................................................................... 32
8.2.2
Measuring System.................................................................................................. 34
9.2
Menu Calibration.................................................................................................................... 36
9.3
10 Maintenance ................................................................................................................................... 38
10.1 Cleaning the Measuring System............................................................................................ 38
10.2 Cleaning the Filling Adapter................................................................................................... 38
10.3 Cleaning AMVn...................................................................................................................... 39
10.4 Wetted Parts of AMVn ........................................................................................................... 39
10.5 Replacing O-rings in the Filling Adapter ................................................................................ 40
10.5.1
10.5.2
10.5.3
Viscosity..................................................................................................................... 42
A.1 Viscosity Standard Fluids .................................................................................... 42
A.2 Calibration Constant K1 ...................................................................................... 43
A.3 Determining the Viscosity.................................................................................... 43
Appendix B:
Appendix C:
B80IB01
B80IB01
B80IB01
2 Safety Instructions
2 Safety Instructions
2.1
2.1.1
Liability
The instruction manual at hand does not claim to address all safety issues
associated with the use of the instrument and samples. It is your
responsibility to establish health and safety practices and determine the
applicability of regulatory limitations.
Only use AMVn for the purpose described in this instruction manual. Anton
Paar GmbH is not liable for damages caused by incorrect use of AMVn.
2.1.2
Do not use any accessories or wearing parts other than those supplied or
approved by Anton Paar GmbH.
Make sure all operators are trained to use the instrument safely and correctly
before starting any applicable operations.
Check AMVn for chemical resistance to the samples and cleaning agents.
2.1.3
B80IB01
The results delivered by AMVn not only depend on the correct function of the
instrument, but also on various other factors. We therefore recommend to
have the results checked (e.g. plausibility tested) by skilled personnel before
consequential actions are taken based on the results.
2 Safety Instructions
2.1.4
2.1.5
Returns
For repairs send the cleaned AMVn instrument to your Anton Paar
representative. Only return the instrument together with the filled out
"Maintenance/Error Report" and the "Standard Maintenance Contract". Find
the applicable forms on the Anton Paar homepage (www.anton-paar.com).
2.1.6
10
Disposal
Observe and adhere to your national safety regulations for handling the
measured samples (e.g. use of safety goggles, gloves, respiratory protection
etc.).
Only store the minimum required amount of sample, cleaning agents and
other inflammable materials near the AMVn.
Make sure that the setup location is sufficiently ventilated. The environment
of AMVn must be kept free of inflammable gases and vapors.
Connect AMVn to the mains via a safety switch located at a safe distance
from the instrument. In an emergency, turn off the power using this switch
instead of the power switch on AMVn.
B80IB01
2 Safety Instructions
2.2
B80IB01
11
3 AMVn - An Overview
3 AMVn - An Overview
Anton Paar GmbH thanks you for buying AMVn, the global vanguard measuring
instrument for the automatic, high-precision viscosity determination of low viscous
substances. It provides you with measurement solutions of utmost precision in a
very broad viscosity and temperature measurement range.
The measurement principle of AMVn is the rolling ball/falling ball principle in
accordance with DIN 53015 and ISO 12058. It consists of the measurement of a
balls rolling time in a diagonally mounted glass capillary filled with sample. The
shear rate is influenced by changing the inclination angle of the capillary.
In this way, investigations are implemented on whether the measured samples
are Newtonian or non-Newtonian (pseudoplastic or dilatant) fluids.
A built-in Pt100 temperature sensor is used for precise temperature
measurements and temperature regulation.
Choose various measurement methods and systems that accord with your
measurement applications.
You can print and display measurement results with sample names and
measurement system data. Only the most recent measurement result is saved in
the AMVn instrument.
In the case of multiple measurement results, it is recommendable to immediately
print or note the data.
With the optional VisioLab Windows Software, all results can be saved and
evaluated in Excel.
The optionally delivered, fully automatic sample changers of Anton Paar GmbH
automatically fill the AMVn capillaries with samples. Between measurements the
capillaries are automatically cleaned with two cleaning agents, and consecutively
dried in air.
Our trained Anton Paar GmbH personnel always welcome your requests and will
gladly provide you with additional offers.
12
B80IB01
AMVn was tested and packed carefully before shipment. However, damage may
occur during transport.
Keep the packaging material (box, foam piece, transport protection) for
possible returns and further questions from the transport- and insurance
company.
Compare the supplied parts to those noted in the table to check the delivery
for completion.
If a part is damaged, contact the transport company and your Anton Paar
representative.
4.2
The following tables list all supplied items contained in your order. Every
instrument is delivered as a unit.
4.2.1
Symbol
B80IB01
Supplied Parts
Pcs.
Article Description
Mat. No.
1569
Instruction manual
German or English
Mains cable:
Euro- or
GB- or
US- or
PRC
52112
61865
52656
27011
Temperature tube
22109
Protective cap
22252
see
Chapter
4.2.2.
13
4.2.2
A measuring system set is delivered in a box, including all necessary parts for
a measurement with AMVn. Every measuring system set includes the measuring
system that accords with your order.
The measuring system is a combination of a capillary and a diametrically
matching ball.
The identification is determined by the internal diameter of the capillary:
e.g. the measuring system (set) 1.6 which is comprised of:
a capillary (diameter 1.6 mm) and balls (diameter 1.5 mm).
Find a listing of all measuring system sets and all parts contained in the box in the
tables below. You can order all registered parts either in a set (Chapter 4.2.2.1)
or separately, in reference to the material number listed on the right (Chapter
4.2.2.2).
4.2.2.1
Symbol
14
Article Description
Mat. No.
67603
63875
63877
63878
Measuring
system
set
uncalibrated
Capillary diameter: 1.6 mm
Volume:
400 L
Ball diameter:
1.5 mm
Measuring range: 0.3-10 mPa.s
1.6
19570
B80IB01
4.2.2.2
Pcs.
Article Description
Calibrated capillary
with data medium and calibration
certificate
Capillary 1.6 mm calibrated or
Capillary 1.8 mm calibrated or
Capillary 3.0 mm calibrated or
Capillary 4.0 mm calibrated
74273
74275
74276
74277
Uncalibrated capillary
Capillary 1.6 mm uncalibrated
67605
Balls
d = 1.5 mm for 1.6
d = 1.5 mm for 1.8
d = 2.5 mm for 3.0
d = 3.0 mm for 4.0
73109
63869
63870
63871
300
300
500
500
B80IB01
mm capillaries or
mm capillaries or
mm capillaries or
mm capillaries
Mat. No.
Plastic tweezers
63132
Syringe, 5 mL
63130
Funnel
for 1.5 mm balls or
for 2.5 mm and 3.0 mm balls
20743
20742
63390
12760
Replacement-O-rings Viton
(Chapter 10.5)
O-ring viton 6.4x1.9 mm FPM 75
(for filling adapter with engraved
marking ring)
oder
O-ring viton 6.5x1.5 mm FPM 80
(for filling adapter without engraved
marking ring)
23726
65555
O-ring tool
24943
63388
63865
15
4.2.3
Symbol
Optional Accessories
Pcs.
Article Description
69079
69578
Ball dispenser
for 1.5 mm balls or
for 2.5 mm and 3.0 mm balls
63604
63605
10
Mat. No.
70116
20659
20660
20661
Keyboard
German
English
77143
77144
Printer
9423
Printer paper
59149
8139
Uncalibrated capillary:
1.6 mm
1.8 mm
3.0 mm
4.0 mm
67605
63180
63181
63182
Disposable syringes:
1 mL Luer
2 mL Luer
5 mL Luer
68015
6771
6772
5034
300
500
500
1
1
100
100
1
16
B80IB01
5 Functional Components
5 Functional Components
5.1
Front of AMVn
A
B
C
D
E
B
D
5.2
A
B
C
D
A
E
B F
C
D G
E H
F
G
H
J I
K
L
I
J
K
L
Fig. 5. 2
Information:
The interfaces COM 1, COM 3, PC and S-Bus are only necessary if the
VisioLab Windows software is used.
Density meters are connected to COM 1.
Sample changers are either connected to COM 3 (SPV and SP3-V) or to
the S-BUS interface (SP-1m and SP-3m).
B80IB01
17
5 Functional Components
5.3
Measuring System
A
B
C
D
E
F
G
D
C
B
F
G
A
Fig. 5. 3
5.4
Time
actual SETTINGS
Sample
Method
Meas. System
Instr. Settings
10:48:12
Name
Remark
Temp.
Dens.
Menu
:
:
:
:
Sample 1
Testlauf 1
20.00C
0.9982g/cm3
Eval
NewRun
ESC
Back
Edit
Cont
.....
.....
.....
.....
......
.....
.....
.....
Keyboard use:
The keyboards function keys (F1 F5) accord with the function keys of AMVn.
Use these to choose the menu and/or input field you wish to access. An external
keyboard enables a much faster and easier input.
18
B80IB01
6.1
Activating AMVn
1. Set up AMVn on the lab desk and level it off with the adjustable feet and the
built-in bubble level. Make sure all four feet touch the lab desk.
Caution:
Before activating the AMVn instrument make sure that the correct line voltage
is applied (85 V to 264 V, 50 Hz to 60 Hz). If large voltage fluctuations are to
be expected, we recommend using a constant-voltage source (UPS).
2. Connect the power inlet of the AMVn instrument to the mains using the
delivered power cord.
3. Activate AMVn with the power switch located on its rear side. The green
display light signals that the power is on.
Warning:
After its activation, AMVn independently runs through the coil- and zero
position initialization. In the course of this the capillary block first moves to its
final position before it remains in its home position leaning backwards at
20 . The capillary block should not be touched during this movement, in order
to avoid possible clamping or displacement of the zero position. The capillary
block should only be moved by hand when the power of the instrument is
switched off.
4. After the successful initialization you will see the following reading on the
display:
actual SETTINGS
Sample
Method
Meas. System
Instr. Settings
Menu
10:48:12
Name
Remark
Temp.
Dens.
:
:
:
:
Sample 1
Testlauf 1
20.00C
0.9982g/cm3
Eval
NewRun
"actual SETTINGS" displays the settings of the last measurement. Browse with
F3 or the down button to check the following:
Sample:
sample name, remark, temperature, density
Method:
the last method used
Meas. system: the measuring system used for the last measurement
Instr. Settings: instrument constant K2, temperature equilibration mode
F4 <Eval> displays/prints the results of the last measurement.
B80IB01
19
6.2
Before starting a measurement, fill the capillary and mount the measuring system
in the capillary block.
6.2.1
1. Hold the capillary (B) vertically and place the funnel (A) on
the capillary opening (Fig. 6. 1).
2. Use the delivered plastic tweezers to introduce a ball into the
funnel.
(B)
Fig. 6. 1
Information:
Never touch the balls with your fingers, as these need to be completely
clean to function correctly. Only use clean capillaries and clean or new
balls.
You can also use an optionally available ball dispenser to introduce the
ball.
6.2.2
Fig. 6. 2
20
B80IB01
6.2.3
Warning:
When handling caustic fluids, acids, easily inflammable fluids etc., use the
required protective clothing (protective gloves, safety goggles).
6.2.4
Always mount the temperature tube and the protective cap (Chapter 6.2.6) before
you start a measurement.
1
2
Fig. 6. 4
... Temperature
tube
... Protective cap
Screw the temperature tube on to the rear side of the capillary block. For
functionality reasons it should always stay attached to the capillary block and
should only be unscrewed for cleaning.
B80IB01
21
6.2.5
Before mounting the filled capillary, you need to activate AMVn and wait until the
self-test is concluded (Chapter 6.1).
Caution:
Make sure that the self-test is concluded and no measurement program has
started, so that the capillary block does not start any surprising movements.
Do not push too hard when mounting the capillary, because the glass could
break. Do not use excessive force!
1. Carefully insert the capillary into the capillary block, starting from the front.
2. Turn the capillary so that the glass knob easily slips into the corresponding
recess of the capillary block opening.
As soon as the glass knob has slid into the opening, completely insert the capillary
into the capillary block until the red silicone O-ring of the adapter is no longer
visible and you meet a resistance.
6.2.6
1. After mounting a capillary, attach the protective cap (2, Fig. 6. 4) in a straight
position in the front of the capillary block.
2. Fix the protective cap by turning it clockwise.
Leave the protective cap on the capillary block during the measurement and only
remove it once you change the capillary.
Hot surface:
When attaching or removing the temperature tube and the protective cap at
high measuring temperatures you might be operating in the direct vicinity of
hot surfaces. Therefore only touch the capillary block and the capillary with
adequate protective gloves (Chapter 2.2) or wait until the capillary block has
cooled down.
Warning:
If the samples in the capillary are heated up to measuring temperatures up to
135 C, the thermal expansion of the fluid in the glass capillary may result in
the unintentional opening of the capillarys sealing caps. It is imperative to use
the temperature tube and the protective cap.
22
B80IB01
7 Measurement
Measurement
7.1
Setting up AMVn
MEASURE PREPARE
Sample Name
Remark
Density
Temperature
11:18:26
Sample 1
Back
Scroll through all items with the F3 <down button> and make sure all settings
always listed on the right are correct before starting the measurement:
F5 <Enter>:
F2, F3, F4:
F1 <Back>:
to change a setting
enter values
to return to "Actual SETTINGS" menu
Information:
A new method or measuring system can only be entered in the submenu
"Setup Measurement" in "MENU".
Individual settings of the temperature equilibration mode are only possible
in the submenu "Setup Instrument" - "Temperature equilibration" .
Keep in mind that only measuring systems that apply to the chosen
method are displayed.
B80IB01
23
7 Measurement
7.2
*** Start *** starts the measurement, for this option press F5 <ENTER>.
MEASURE PREPARE
11:19:23
Temp. Equi.
Method
Measuring System
*** Start ***
Back
Every measuring procedure consists of three consecutive steps:
1. Thermostatting of capillary block
2. Sample thermostatting
3. Sample measurement
7.2.1
If the set temperature ("set T:") does not accord with the actual temperature ("act
T:"), the instrument holds up the temperature equilibration of the sample until the
display reading "!Temp. diff >!" disappears. This occurs at differences smaller
than 0.05 C.
7.2.2
Sample Thermostatting
The sample thermostatting takes place immediately after the thermostatting of the
capillary block. You can find settings for this in Chapter 8.1.2.
RUN - TempEqui
Time (s)
to Start: 2.65
Runtime : 11.34
to End : 0.63
var Coef: min.rep = 4
Skip
F2 <Show>:
F1 <Skip>:
Show
*auto
set
act
set
act
10:55:02
T
:
T
:
ang.:
ang.:
rep :
20.00C
20.00C
70
70.0
2/4
Strtd
Every start and stop signal is accompanied by a beeping sound. After the second
measurement the actual and desired deviation ("var Coef:") is displayed. After a
successful temperature equilibration, the measuring instrument starts the actual
measurement.
24
B80IB01
7 Measurement
7.2.3
The display shows the actual measuring angle, the current number of actual
repetition, and, from the third measurement onwards, the determined viscosity.
RUN
11:25:12
Time (s)
to Start: 2.65
Runtime : 11.34
to End : 0.63
Visc.
: 1.002 mPa.s
Stop
F2 <Show>:
F1 <Stop>:
Esc:
Enter:
set
act
set
act
T
:
T
:
ang.:
ang.:
rep :
Show
20.00C
20.00C
70
70.0
Strtd
Caution:
Choose the angle of your measurement in such a way that the ball rolling
times do not drop below 10 sec. Otherwise turbulence is caused in the
capillary and a seemingly higher viscosity is read.
In this case choose another measuring method (Chapter 8.2.1).
Information:
Always enter an even number of repetitions, because viscosity is always
determined from two repetitions.
For the viscosity calculation a maximum of 20 repetition periods is used. If you
want more than 20 repetitions, you have the further option of connecting
AMVn to a PC. The AMVn Visiolab Windows software allows a nearly
unlimited amount of repetitions, which is only one of its many benefits.
7.3
EVALUATION
Sample:Sample1
dyn. Visco
kin. Visco
Infos
Rep.times
ESC
B80IB01
11:23:56
Testlauf1
Mean
: 1.0019 mPa.s
Time
: 11.35 s
St.Dev :0.00006 mPa.s
Var.Co : 0.01 %
25
7 Measurement
7.4
----- A M V n Microviscometer
SNo: 409247
----------
-----
SW_Ver: V1.70
EVALUATION
-----------
Sample: Water
Remark:
Method: Standard 1.6
Angle:
70 deg
Repetitions: 4
CapSys: 1.6/1.500 mm
Density :
0.9982 g/cm3
act. Temp:
20.00 C
MeasSystem: 15844194-70A
K1
: 0.984e-02
dyn. Viscosity: 1.0020 mPa.s
kin. Viscosity: 1.0038 mm2/s
mean Time:
15.27 s
Var. Coeff.:
0.00 %
Started on 2010-03-10 at 08:38
26
B80IB01
7 Measurement
7.5
B80IB01
27
8.1
In "Setup Instrument" you can view or change all settings concerning AMVn.
With F5 <Enter> you can access each of the following submenus, with F1 <Back
or <ESC> you return to the menu.
"Date" Date
"Time" Time
"Zero adjust" Zero adjustment
"Temperature equilibration" Temperature equilibration
"Instrument Constant K2" Instrument constant: preset
"Default Settings" Default settings
"Configuration Printer" Printer configuration
"Info" Information on Firmware Version, serial number of AMVn and your
Anton Paar GmbH contact address
"Service Utilities" password-protected service menu for authorized
service personnel of Anton Paar GmbH
"Software Update" Update instrument software
8.1.1
Change date and time with the arrow keys and confirm the entry with <Enter>.
SETUP Date
Date :
10:49:37
2000-09-23 Saturday
YYYY-MM-DD Day
ESC
28
B80IB01
8.1.2
Temperature Equilibration
You can choose between two different modes: constant oder auto mode. The
active mode is highlit and marked with an asterisk (*).
F2, F3:
F5 <Enter>:
F4 <Edit>:
8.1.2.1
Choose "constant mode" if you know how long your sample will take to reach
measuring temperature.
F2, F3, F4 arrow keys: adjust waiting periods until measurement start
F5 <Enter>:
accept changes
F1 <ESC>:
discard changes
waiting time:
ESC
8.1.2.2
10:55:24
60s
Edit
ESC
angle:
var.coeff:
rep. to fit:
max.rep.:
:
:
:
:
70
0.1%
4
20
Edit
measurement angle
variation coefficient to reach temperature equilibration
number of repetitions to calculate the variation coefficient
max. number of repetitions to reach temperature equilibration
F5 <Enter>:
F2, F3, F4 arrow keys:
F5 <Enter>:
F1 <ESC>:
B80IB01
angle
var. coeff.
rep. to fit
max. rep.
10:55:31
29
8.1.3
Default Settings
In the submenu "Default Settings" you can reset user-defined settings to factory
default values.
Prnt
F2 <Prnt>:
F3 <Meth>:
F4 <Meth>:
8.1.4
10:55:56
Back
1-2
Meth
Meas
Printer Configuration
In the submenu "Configuration Printer" you can change your printer settings.
:
:
:
11:48:42
COM2
OFF
1.0 s
ESC
Press F4
F5 <Enter>:
F1 <ESC>:
confirm changes
discard changes
Delays are necessary for printers without a receive buffer. In case characters are
missing in your printout, increase the delay times.
The parameters of the printer interface on the AMVn instrument are pre-adjusted
and cannot be changed:
baudrate 9600, 7 data bits, even parity, 1 stop bit, no handshake
30
B80IB01
8.1.5
Software Update
In the submenu "Software Update" you can install new instrument software
versions. Further information will be included with your update.
Whenever applicable, contact your local Anton Paar representative for possible
firmware updates.
8.1.6
AMVn is already delivered pre-adjusted. In case you observe evident run time
differences between even and uneven repetitions, you can compensate these
with the zero point adjustment.
Before starting a zero point adjustment, make sure that the instrument is placed
on a level, stable surface. If necessary, level AMVn using the adjustable feet and
the built-in bubble level.
Follow the submenu instructions on the screen.
Step:
Step:
Step:
Step:
Back
F3 <Meas>:
F5 <Edit>:
F5
F3
F4
F3
Show
to
to
to
to
1-2
10:55:02
edit method
start Run
adjust 0-Position
control adjustment
Meas
Zero
Edit
starts the zero point adjustment with the method "Zero Adjust"
changes the zero point adjustment method
After the selected angle is measured, AMVn automatically raises the angle by 1
degree and starts a new run. This second run substantially raises the accuracy of
the zero point adjustment.
B80IB01
31
8.2
SETUP Measurement
10:59:35
Method
Measuring System
Measuring System: Calibrate
Menu
With F4 <Print> you can print the content of the highlit menu.
8.2.1
Method
A method involves the measuring angle, the number of repetitions, and the
measuring system used for a measurement.
In the menu "Method" you can create, revise or delete a method.
Information:
When using more than one method it is advisable to assign slightly different
names. This might prove helpful later in matching them to the applications.
32
B80IB01
SETUP: Measure-Method
->
2/05
Name
Angle
Rep.
CapSys
:
:
:
:
Standard 1.6
70
04
1.6/1.500 mm
Back
F2, F3 arrow keys:
e.g. 2/05:
F1 <Back>:
F5 <Enter>:
F5 <Edit>:
F3 <Delete>:
F1 <Back>:
F4 <Enter>:
11:01:37
New
display all stored methods
the second of five stored methods is displayed
back to the menu
choose the displayed method to then press:
to edit or
to delete or
to return
enter a new method:
.......
ESC
F5 <Enter>:
11:00:44
Save
choose a parameter
The option F5 <Overw.> (Overwrite) is only available after you edited a previously
existing method.
B80IB01
33
8.2.2
Measuring System
SETUP: Measure-MeasSystem
->
1/05
11:04:31
Cap
:11118183-70A
K1
:1.28800e-02
CapSys
:1.6/1.500
Ball Dens:7.8500e+00
Back
1999-07-22
14:41:00
New
1/05:
F4 <Enter>:
F2, F3 arrow keys:
F5 <Enter>:
F5 <Edit>:
F3 <Delete>:
F1 <Back>:
SerNo
Angle
K1
CapSys
ESC
11:04:46
.......
Save
F5 <Enter>:
choose a parameter
arrow keys or external keyboard: change parameters:
SerNo:
8 digit serial number (engraved in the capillary)
Angle:
measurement angle
K1:
calibration constant K1
CapSys:
measuring system (capillary-/ball diameter)
B-Dens:
density of the ball
Date:
date of calibration
Time:
time of calibration
F4 <Save>:
F3 <Insert>:
F5 <Overw.>:
F1 <Back>:
The option F5 <Overw.> (Overwrite) is only available after you edited a previously
existing measuring system.
34
B80IB01
8.3
In the third submenu of "MENU" "Print Configuration", you can print the
configuration of AMVn.
The printer has to be attached to the selected connection
(default selection: COM 2).
PRINT: Configuration
11:12:02
Printing on COM2
Please wait
ESC
F5 <Enter>:
<ESC>:
You can find the settings for the printer interface in Chapter 8.1.4.
Example of a printout
(numbers and values are different on every instrument):
AMVn Microviscometer
SNo: 409247
SW_Ver: V1.70
B80IB01
35
9.1
There are calibrated and uncalibrated measuring system sets (Chapter 4.2.2) and
calibrated and uncalibrated capillaries:
When using a new or alternative ball set the calibration constant only needs to
be redefined in case the measuring deviation is above tolerance limit (0.3 %) and/
or the ball set features a different batch number. Read the batch number on the
sticker attached to your ball set.
When calibrating a measuring system, choose the viscosity standard in such a
way that it falls within the viscosity range of the measurement at the usual
measuring temperature (Chapter A.1).
9.2
Menu Calibration
SETUP: Measure-MeasSystem-Calib
->
1/05
Back
F2, F3:
F4 <New>:
F5 <Enter>:
1/05:
36
Cap
:11118183-70A
K1
:1.28800e-02
CapSys
:1.6/1.500
Ball Dens:7.8500e+00
11:04:31
1999-07-22
14:41:00
New
browse all stored measuring systems and check all values
calibrate a new measuring system
choose the measuring system you wish to recalibrate
the first of five stored measuring systems is displayed
B80IB01
9.3
1. Press F4 <New>.
2. Enter all values:
SerNo
Angle
K1
B-Dens
ESC
SerNo:
Angle:
CapSys:
B-Dens:
S-Name:
S-Rem:
S-Visc:
S-Dens:
Temp:
Rep:
11:06:39
.......
Cont
8-digit serial number (engraved in the capillary)
measurement angle
measuring system (capillary-/ball diameter)
density of the ball (e.g. steel ball = 7.85 g/cm3)
name of calibration medium
remark on calibration medium
dyn. viscosity of calibration medium at the chosen temperature
density of calibration medium at the chosen temperature
temperature for the calibration medium values
number of repetitions for the calibration
F5 <Enter>:
F2, F3, F4 arrow keys:
F4 <Cont>:
F3 <Insert>:
F5 <Overw.>:
F1 <ESC>:
edit a value
make entries
continue
insert measuring system
overwrite an existing measuring system
return to previous screen
3. Fill the capillary with the suitable viscosity standard fluid (Chapter A.1).
4. Press F5 <Start> to start the calibration.
The calibration procedure runs automatically.
5. After measurement conclusion press:
F2, F3:
view the detailed evaluations
F5 <Print>:
print the result
F4 <Save>:
save to use the calibration constant in measurements.
Information:
If you do not save the calibration constant K1, you lose all calibration results.
B80IB01
37
10 Maintenance
10 Maintenance
Clean AMVn and its accessories regularly.
We recommend regular maintenance of your measuring instrument by authorized
Anton Paar GmbH personnel.
38
B80IB01
10 Maintenance
Use a soft cloth and warm water to clean the casing and display.
Check the capillary block for pollution and clean it regularly, especially at the
thermostat and protective cap fixations.
Never remove the warning signs on the capillary block. If these come loose,
they need to be attached again.
B80IB01
Material
Capillary
Borosilicate glass
Ball
Filling adapter
PCTFE
O-ring
Viton
O-ring
Kalrez
LUER cap
PCTFE
LUER plug
PCTFE
39
10 Maintenance
10.5.1
Silicone
Fig. 10. 5
The O-rings on the inside are made of viton and have the dimension
6.5 x 1.5 mm FPM 80. They have contact to the sample and have a sealing
function.
10.5.2
Engraved
marking ring
Silicon e
Fig. 10. 6
The filling adapter with engraved marking ring contains O-rings of the dimension
6.4 x 1.9 mm FPM 75. In this filling adapter with engraved marking ring O-rings
made of viton and made of Kalrez can be used.
Kalrez-O-rings should be used if viton is not resistant against the measured
sample. You can optionally order it (Chapter 4.2.3).
10.5.3
Caution:
The filling adapter without engraved marking ring and the filling adapter with
engraved marking ring are NOT compatible.
The O-rings have another dimension (see Chapter 4.2.2 and Chapter 4.2.3).
The red, outer O-ring is made of silicone and does not usually need to be
exchanged. The inner O-rings are sealing rings. If the O-rings need to be
replaced, first remove them from the adapter using the O-ring tool. Then carefully
insert the new O-rings.
40
B80IB01
11 Technical Data
Time measuring range:
Resolution:
Accuracy:
0 - 250 s (999 s *)
0.001 sec
<0.002 sec
Viscosity range:
Repeatability:
Reproducibility:
Temperature range:
Resolution:
Accuracy:
+5 C to +135 C
0.01 C
<0.05 C
Mains connection:
15 kg (150 N)
Ambient temperature:
15 to 35 C
Air humidity:
41
*)
**)
***)
typical value for the range, where the calibration constant was determined
B80IB01
Appendix A: Viscosity
Appendix A: Viscosity
For a precise viscosity determination, a suitable calibration standard fluid is
required for the calibration of the capillaries. A calibration constant needs to be
determined and/or entered for every capillary (Chapter 9).
Information:
The calibration procedure has to be repeated periodically, depending on the
specific accuracy requirements of your laboratory.
A.1
Due to the different measuring ranges of the capillaries, you need different
viscosity standard fluids. You can order these viscosity standard fluids at Anton
Paar GmbH.
The table below shows general values for the different measuring systems.
The best results are achieved if you calibrate a measuring system at the usual
measuring temperature and in its viscosity range.
(e.g. if you measure approx. 50 mPa.s with a 3.0 measuring system at 100 C,
use e.g. the viscosity standard fluid DKD 10000. At 100 C this standard has
approx. 57 mPa.s).
For measuring aqueous samples we recommend bidistilled water as a viscosity
standard fluid.
Viscosity value for calibration [mPa.s]
20 C
40 C
100 C
Capillary [mm]
Capillary [mm]
Capillary [mm]
1.002
0.9 / 1.6
0.6529
0.9 / 1.6
---
33
1,8 / 3,0
15
1.8
100
3.0 / 4.0
Mat. No.
Bidistilled water
26750
3.1
1.6 / 1.8
12935
38
1.8 / 3.0
6.2
1.6 / 1.8
9754
1100
4.0 / 4.0 WK/KO*
330
4.0
32
1.8 / 3.0 / 4.0
12936
7900
4.0 WK/KO*
1500
4.0 / 4.0 WK/KO*
57
1.8 / 3.0 / 4.0
DKD 10000AW
301364
B80IB01
42
Appendix A: Viscosity
A.2
Calibration Constant K1
For every measuring angle you plan to use, first determine the calibration
constant K1.
N
K 1 = ----------------------------------t1(K N )
K1
N
t1
K
N
Information:
The calibration constant is different at every angle and with every
capillary.
A calibration datasheet with pre-determined calibration constants for the
stated angles is enclosed with every calibrated capillary.
If a ball is exchanged, K1 only needs to be readjusted if the measuring
deviation is above the tolerance limit (0.3 %). Readjust the constant if you
use balls with different batch numbers. The batch number is written on the
ball vessel.
The viscosity of the standard fluid should fall within the viscosity range of
the measured samples.
A.3
The dynamic viscosity of the sample is calculated from the calibration constant
K1 of the measuring system, the rolling time t1 and the density difference
between ball and sample.
= K1 * (K - P) * t1
K1
K
P
t1
...
...
...
...
...
The serial number and the ball density of the measuring system are known (steel
ball: K = 7.85 g/cm3, tungsten carbide-cobalt ball: K = 15 g/cm3). The density of
the sample, often a known number, must also be entered.
If there is no information on the density of the sample to be measured, you can
order a manual density meter DMA 35N from Anton Paar GmbH, which measures
the density with an accuracy of +/- 0.001 g/cm3.
Caution:
Choose the angle of your measurement in such a way that the ball rolling
times do not drop below 10 sec. Otherwise turbulence is caused in the
capillary and a seemingly higher viscosity is read.
43
B80IB01
A measuring range extension through an increase in run time from 250 to 999 sec
quadruples the maximum measuring range.
The run time increase can only be defined by means of a "remote" command from
the PC.
Proceed as follows:
1. Connect AMVn and PC with the interface cable.
2. Switch on your AMVn and the PC.
3. Start the Windows Terminal Program
PC port: 9600, 8 databit, no parity, 1 stop bit, no handshake.
4. The following commands, which need to be typed using capital and small
letters, define the max. runtime to 999 sec:
<REM
<STATUS 00000>
<setrlim = 999>
Runtime-Limit = 999
<LOC
OK
5. End the Terminal Program.
6. Check with the printer in menu Print Configuration.
7. Check the print:
AMVn Microviscometer
SNo: 80074021
SW_Ver: V1.70
B80IB01
44
B.2
45
B80IB01
B80IB01
F1
F1
Menu
F5
F1
Meas. System:
Instr. Settings:
F5
Date
Time
Zero adjust
Temperature equilibration
Instrument Constant K2
Default Settings
Configuration Printer
Info
Service Utilities
Software Update
Name
Angle
Repetitions
Capillary System
Method:
F2
Name
Remark
Temperature
Density
Sample:
Setup Instrument
F2 F3
actual SETTINGS
Initialization
Start
F5
F1
46
47
F5
F1
F4
F1
New Run
Evaluation
F5
F1
F5
F1
F2 F3
F2 F3
F5
Sample Name
Remark
Density
Temperature
Temp. Equi
Method
Measuring System
Start
Angle / Method
Repetitions
Temperature
Density
Mean
Time
St.Dev.
Var. Co.
Kin. Visco.:
1/2
2/2
F1
F1
Mean
Time
St.Dev.
Var. Co.
Infos:
F1
Measuring System
Method
Dyn. Visco.:
Print Configuration
F2 F3
Setup Measurement
F1
F2
F1
Start Time
Capillary SerNo.
K1
Capillary System
Stop measurement
Show runtimes
Setup: Measure-MeasSystem-Calib
Setup: Measure-Method
Print Methods
RUN
F5
F4
F1
F4
F5
F5
B80IB01