Académique Documents
Professionnel Documents
Culture Documents
Experimentalevaluationoftheabsorptionspectrumofanumberofsamplesbased
onlampillumination.
(BenGurionUniversitygroup).
Summary
Though an experimental system is in place to measure the absorption spectrum in the range of 380 to
1700 nm based on direct solar radiation (described in our first-year report), we decided to build a
second system based on a Xenon arc discharge lamp whose output is stable and not dependent on
weather conditions. The system uses a heated sample holder which eventually will allow
measurements at temperatures of up to 700C. The sample holder can accommodate both the smaller
(30x30mm) and the 60x60mm samples.
We evaluated this system and performed detailed measurements on the first set of samples we
received during October 2011 from Camille.
The system is not perfect as we will detail below, but first results appear indicative of the capability
of the system and is hopefully useful in the further development of the anti-reflective coating.
New Experimental Setup
Previous experiments were performed using the direct sun directed into the lab via a heliostat
(reported in our first-year report). A small waviness of the (nominally flat) heliostat caused
significant variations in the irradiation of the sample and influenced the measurements. As a result it
was decided to build an optical system for measuring the absorption spectrum based on a Xe arc
discharge lamp (Fig 1).
The new experimental setup was also designed for measuring the spectral absorptivity at high
temperatures. The source of radiation (Fig.1a) is a Xe lamp (300W nominal power), which reaches
stability within about 15 minutes of operation and remains stable with a drift of 0.5% and
fluctuations of 1% (manufacturers data). We added two flat mirrors to increase incident light on
the sample. The distance between the light source and the target was set to be 60 cm, a compromise
between sufficient light at the sample (the samples are highly absorbing!) and light uniformity at the
sample. The light is not completely collimated but subtends an angle of about 4. We measured the
light intensity as a function of position on the active area of the sample (5cm across) and found it to
be within ~0.5% of peak radiation.
The sample holder can be heated via a custom made heating element whose outside dimensions are
100x100x50 mm (Figs. 1b and 1c). During pre-experiments we successfully heated a sample (60x60
mm) up to 1000K. The temperature was measured by means of a portable thermocouple-integrated
high performance handheld Infrared thermometer (OMEGA OSXL689).
A reasonable uniform temperature distribution on the sample with minor temperature gradients
towards the edges was observed (approximately 5C between center and edge at 320C). As a
reference, we use a gray sample (Fig. 1d) from Labsphere with an NIST-traceable calibrated
absorption spectrum (Fig 2).
(a)
(b)
(c)(d)
Figure 1. (a) left: side view of system, right: view of the lamp and mirrors from direction of the
sample. (b) The heated sample holder and optical fiber with collimating optics (c) Electric heater
with the sample removed, (d) reference sample from Labsphere with an absorption coefficient
0.96 (Lambertian reflector) in the visible and NIR part of spectrum.
reflectance
0.1
0.09
0.08
0.07
0.06
0.05
0.04
0.03
0.02
0.01
0
400
reflectance
600
800
1000
1200
1400
1600
1800
wavelenght (nm)
Figure 2. Reflectance spectrum of the Labsphere gray spectralon reference sample, as supplied by the
manufacturer, NIST traceable calibration.
Fig.3. Comparing the reflectance spectrum of the reference sample to itself at different angles s. In
other words, if the sample was truly Lambertian, we should get unity for the entire spectrum.
B. Experiments on four samples.
A series of experiments was performed on four samples (MS-1.1, MS-1.3, MS-1.5 and TS1). Their
dimensions were 30x 30 mm for the first three (MS-1.1, MS-1.3, MS-1.5 with 1, 2 and 3 layers of
coating respectively) and 60 x 60 mm for the TS1. Two series of experiments were performed: in the
first we kept the samples at ambient temperature and in the second the samples were heated to
around T=600 K. In all cases we measured the reflected spectral intensity r =f(, s, s, T) of
reflected light by means of two spectrometers, one in the visible range (380-900nm) and a second in
the NIR (900-1700nm) part of the spectrum. Both spectrometers are by Control Development, the
visible is the PDA-512, and the near infrared is NIR128L-1.7TS. Two angles, which characterize the
incident direction of light were kept constant for all experiments, namely, i=0, and i= (Figure
3). The solid angle subtended by the receiver was also fixed (determined by the fiber aperture and
the collimating optics). The angles, which define the direction from which the reflected light are
measured, are polar angle s and the azimuth angle s (Fig. 3). In our preliminary measurements we
varied only the polar angle s (15, 30,45 and 60).
Figure 3. Definitions of incidence angles and angles of reflection. I is incident radiation on the
sample surface. The ZB axis is normal to the sample surface, and the XB axis lies in the plane
defined by ZB and I. The incident direction is given by (i,i), where i = by definition. The
direction of the reflected radiation is given by (s, s). is the solid-angle subtended by the receiver.
(a)
(b)
(c)
(d)
(a)
(b)
(c)
(d)
Future work
1. To measure at more broad range of angles to be able to measure, with higher accuracy, the total
hemispherical absorbance.
2. Perform experiments with more powerful power supply to reach higher temperature and
increasing the incident intensity, in order to minimize the noise.
3.
Increasing the light input in order to improve the signal to noise ratio.