Vous êtes sur la page 1sur 7

Europace Advance Access published October 30, 2014

CLINICAL RESEARCH

Europace
doi:10.1093/europace/euu259

Right ventricular septal pacing as alternative for


failed left ventricular lead implantation in cardiac
resynchronization therapy candidates
M. Hafez A. Alhous 1*, Gary R. Small 1, Andrew Hannah 1, Graham S. Hillis 2,
Michael Frenneaux 1, and Paul A. Broadhurst1
1

Department of Cardiology, Aberdeen Royal Infirmary/University of Aberdeen, Aberdeen AB25 2ZN, UK; and 2The George Institute for Global Health, University of Sydney, Australia

Received 14 April 2014; accepted after revision 25 August 2014

Aims

----------------------------------------------------------------------------------------------------------------------------------------------------------Keywords

Cardiac resynchronization therapy (CRT) Right ventricular outflow tract pacing Heart failure Left ventricular
dyssynchrony Left ventricular systolic function

Introduction
Cardiac resynchronization therapy (CRT) is an effective treatment in
patients with refractory heart failure due to severe left ventricular
(LV) systolic impairment in the presence of left bundle branch
block (LBBB).1 It improves the quality of life and prolongs survival.2 5 Cardiac resynchronization therapy is, therefore, recommended in international guidelines for appropriate patients with
heart failure6,7 and used extensively in clinical practice.8 Despite

the benefits of CRT it can pose technical challenges, in particular,


with placement of the LV pacing lead.9 Implantation of the LV lead
through the coronary sinus (CS) may be impossible in 510% of
cases due to variations in cardiac venous anatomy.9,10 Other technical reasons for LV lead failure include: lead dislodgement, phrenic
nerve stimulation, and lead positioning at a scarred site.11 16
When percutaneous LV lead implantation is not possible an alternative approach is to perform invasive surgical LV lead implantation;
this appears to offer functional benefits similar to those of standard

* Corresponding author. Tel: +44 1224 550761; fax: +44 122 550692. E-mail address: h.alhous@abdn.ac.uk
Published on behalf of the European Society of Cardiology. All rights reserved. & The Author 2014. For permissions please email: journals.permissions@oup.com.

Downloaded from by guest on December 13, 2014

To compare the effects on left ventricular (LV) function of right ventricular (RV) septal pacing vs. cardiac resynchronization therapy (CRT) in patients with an indication for the latter. Cardiac resynchronization therapy is an effective therapy in
patients with drug-refractory heart failure. Despite advances in implantation techniques, LV lead placement can be impossible in up to 10% of cases. We, therefore, assessed the effects of RV septal pacing from mid septum (RVmIVS)
and outflow tract (RVOT) on cardiac performance, in comparison with CRT.
.....................................................................................................................................................................................
Methods
Twenty-two patients scheduled for CRT underwent dual-chamber temporary pacing. The ventricular lead was placed at
the RV apex (RVA), RVmIVS, and RVOT in random order. Comprehensive echocardiography was performed in a baseline
and results
AAI mode and then at each RV position in dual chamber pacemaker function (D pacing, D sensing, D dual responses)
mode and repeated on the next day following CRT implantation. Right ventricular apex pacing did not change any of
the assessed echocardiography parameters. Both RVmIVS and RVOT pacing increased LV ejection fraction (EF):
29 + 7% at baseline vs. 32 + 6% (P 0.02) and 32 + 5% (P 0.04) with RVmIVS and RVOT pacing, respectively. Similarly, the dyssynchrony index (Ts-SD) decreased: 50 + 19 ms at baseline vs. 39 + 17 ms (P 0.04) and 37 + 17 ms
(P 0.006) with RVmIVS and RVOT pacing, respectively. Cardiac resynchronization therapy further improved LVEF
and Ts-SD to 36 + 7% and 34 + 15 ms, respectively, however, only LVEF was significantly higher compared with
RVmIVS and RVOT pacing (P 0.03 and P 0.01 respectively). There were no significant differences in either LVEF
or Ts-SD between RVmIVS and RVOT.
.....................................................................................................................................................................................
Conclusion
Right ventricular septal pacing from mid septum or RVOT pacing improves LVEF and LV synchrony in CRT candidates.
Further improvement in LVEF was achieved by CRT, which remains the gold standard therapy in these patients.
However, RV septal pacing is worthy of further study as an alternative strategy when LV lead implantation fails.

Page 2 of 7

Whats new?
The effects of right ventricular (RV) septal pacing from RV mid
septum and RV outflow tract on LV function and dyssynchrony
in patients with indication for CRT (cardiac resynchronization
therapy).
Right ventricular septal pacing as an alternative to failed left
ventricular lead in patients with indication for CRT.

Methods
Study population
Patients who were scheduled to undergo conventional CRT implantation
were recruited to the study. Patients with a current atrial tachyarrhythmia
were excluded. The temporary pacing protocol was performed 4
weeks prior to the date of CRT implantation. The study was approved
by the Grampian Research Ethics Committee and informed consent
was obtained.

Pacing procedure
Dual-chamber temporary pacing
Temporary pacing was performed in a sterile manner via the right femoral
vein using two intra-cardiac electrical catheters (TORQRTM , Medtronic).
Under fluoroscopic guidance one catheter was placed in the high right
atrium; the second catheter was then placed in the RV. Temporary
atrial pacing (AAI) was established at 10 beats per minute (b.p.m.)
above the baseline sinus rate to ensure constant pacing throughout
the study.29 Temporary dual chamber pacemaker function (D pacing, D
sensing, D dual responses) (DDD) pacing was subsequently commenced
at the established atrial rate. The pacing rate was kept the same throughout the study at all RV sites. Atrio-ventricular (AV) delay was optimized
according to the trans-mitral pulse-wave Doppler echocardiographic
method.31 The ventricular pacing catheter was placed in random order
at three RV pacing sites: RV apex (RVA), RVmIVS and RVOT. Ventricular
pacing lead positions were confirmed by both fluoroscopy in both left anterior oblique and right anterior oblique views (to ensure the septum, not
the free wall was being paced) and electrocardiogram (ECG) characteristics.32,33 Pacing was carried out at twice diastolic threshold. Direct
His bundle capture was avoided. A full study protocol (detailed

echocardiography and ECG) was undertaken with AAI pacing (baseline)


and repeated at each RV pacing sites.

Cardiac resynchronization therapy device implantation


Cardiac resynchronization therapy device implantation was performed in
a standardized fashion.34 Venous access was obtained via the left axilliary
vein using ultrasound (SonoSite-iLook, Providian Medical Equipment
LLC). The RV pacing or defibrillation lead was positioned at the RV
apex in all patients. The LV lead was implanted in one of the tributaries
of the CS in all patients. The proximal CS was intubated with a guiding
sheath and the pacing lead was advanced to a second- or third-order
branch of the CS to pace the LV lateral wall. Finally, a permanent
passive fixation atrial pacing lead (CapSurew SP Novus, model 5592,
MedtronicTM ) was positioned at the right atrial appendage in all patients.
All pacing lead checks were performed using a Medtronic programmer
(The Medtronic CareLinkw Programmer, MedtronicTM ). A CRT generator (MedtronicTM ) was used in all study patients. Detailed echocardiography and ECG were carried out the following day, using the same atrial
pacing rate as in the RV studies.

Electrocardiography
A 12-lead surface ECG was performed prior to the study, during AAI
pacing, at each RV pacing site studied, and post CRT. QRS duration was
calculated using the first to the last sharp vector crossing the isoelectric
line in all leads. The mean of these values was used for statistical analysis.

Echocardiography
A detailed trans-thoracic echocardiogram was performed during AAI
pacing, during DDD pacing at each RV pacing site, and on the day following CRT device implantation. Subjects were paced for 10 min at each RV
pacing site prior to echocardiographic assessment. Images were obtained
using a 3.5 MHz phased-array transducer and a Vivid 7 echocardiography
machine (General Electric Healthcare). For all echocardiography data, at
least three consecutive paced cardiac cycles were stored in cine-loop
format for off-line analysis (EchoPAC, General Electric Healthcare). All
echocardiographic variables were measured in triplicate from three
different paced beats and the average value was calculated and used for
statistical analysis.
Left ventricular ejection fraction was measured using Simpsons
bi-plane method and cardiac output calculated.35 Left ventricular diastolic function was assessed from the mitral inflow in the apical four-chamber
view using pulse-wave Doppler.36 Systolic and diastolic function was also
evaluated using colour tissue Doppler imaging (TDI) to measure mitral
annular and myocardial velocities.37
Intra-ventricular dyssynchrony (LV dyssynchrony): TDI was used to
assess longitudinal LV dyssynchrony. Myocardial velocity curves were
constructed from the three standard TDI apical views and the peak sustained myocardial systolic velocity (during the ejection phase) (Sm) for
each of the 12-LV segments was identified. The time-to-peak Sm (Ts)
was measured with reference to the onset of the QRS complex.31,38
From these measurements, a variety of previously described dyssynchrony parameters were derived. These included the difference in Ts
between the basal septum and the lateral wall (septal-to-lateral delay),
where a difference of 65 ms is a widely used indicator of LV dyssynchrony, and the standard deviation for Ts among all 12 basal and
mid-LV segments (the dyssynchrony index), where a cut-off value of
32.6 ms is suggested.38,39
Inter-ventricular dyssynchrony: The time from the Q-wave on the
ECG to semi-lunar valve opening was evaluated using pulsed-wave
Doppler of the LV outflow tract in the apical five-chamber view and
the RVOT in the short axis parasternal view. An inter-ventricular

Downloaded from by guest on December 13, 2014

CRT;17,18 however, these benefits come at the expense of perioperative complications associated with the use of general anaesthesia in
patients with severe LV impairment.19 21 A more simple alternative
strategy to failed LV lead implantation could be achieved by percutaneous placement of the pacing lead at a site close to the LV such as the
RV outflow tract or RV mid septum. Right ventricular outflow tract
lead placement is a relatively simple procedure that is often used in
permanent pacing for bradyarrhythmias and can be carried out at
no additional risk.22 27 Furthermore, in patients who have bradycardia indications for pacing with concomitant heart failure, early studies
reported an improvement in LV ejection fraction (LVEF) and
New York Heart Association (NYHA) class with alternative RV
site pacing.28 30 In CRT candidates, however, it remains undetermined whether alternative RV site pacing may improve LV function.
The aim of the current study is to compare the effects of pacing
from the RV apex, RV mid septum (RVmIVS) and RVOT on LV synchrony and function with standard CRT.

M. H. A. Alhous et al.

Page 3 of 7

RV septal pacing as alternative for failed LV lead implantation

mechanical delay (IVMD) 40 ms was considered indicative of interventricular dyssynchrony.31,38

Statistical analysis
Continuous data are expressed as mean and standard deviation (SD)
where normally distributed and median with interquartile range (IQR)
were skewed. Non-parametric related data were compared using the
Wilcoxon signed-rank test. Analysis of variance also was conducted to
determine which of the paired comparisons differ significantly using
post-hoc multiple comparisons for observed means. Inter- and
intra-observer variability were determined using the Bland Altman
methods.40 Comparisons were made between each of the four pacing
sites and modes: AAI (baseline), RVmIVS, RVOT, and CRT.
Significance was established at a P-value of ,0.05. However, for
repeated comparisons a Bonferroni correction to control for a Type I
error was applied. All statistical calculations are undertaken using SPSS
version 20.

A total of 22 patients were recruited into the study. The study was
completed in all the patients. The pacing rate needed to achieve

Patients
(Total n 5 22)

................................................................................
Age (years), mean + SD

70 + 11

Male, n (%)
NYHA class

13 (59)
18 (82)

NYHA class IV, n (%)


De novo CRT/upgrade from PPM

4 (18)
20/2

Known coronary heart disease, n (%)

14 (64)

Mitral regurgitation (moderate severe), n (%)


Previous paroxysmal atrial fibrillation, n (%)

8 (36)
5 (23)

Hypertension, n (%)

8 (36)

Diabetes mellitus, n (%)


Peripheral vascular disease, n (%)

7 (32)
2 (9)

Renal failure, n (%)

5 (23)

Anaemia, n (%)
Hypothyroidism, n (%)

4 (18)
4 (18)

Chronic obstructive lung disease, n (%)

3 (14)

Heart rate (HR) b.p.m., mean + SD


Blood pressure (BP) (pre-pacing), mmHg

65 + 13

Systolic BP (SBP), mean + SD

124 + 21

Diastolic BP (DBP), mean + SD


Pulse pressure (PP), mean + SD

69 + 11
55 + 18

QRS duration (ms), mean + SD

166 + 26

Axis (8), mean + SD


Heart rhythm before pacing:

14 + 68

Sinus rhythm, n (%)

13 (59)

First degree AV block, n (%)


Second degree AV block, n (%)

8 (36)
1 (5)

SD, standard deviation; n, number of patients.

Left ventricular ejection fraction was between 36 and 44% in three


patients at baseline (AAI pacing), and 35% in the remaining patients.
Left ventricular ejection fraction was unchanged by RVA pacing compared with baseline. Pacing from the RVmIVS or RVOT significantly
increased LVEF compared with baseline and compared with RVA
pacing. There was no significant difference in the effect between
the two RV septal pacing sites on LVEF. A further significant
incremental increase in LVEF was achieved by CRT (Table 2 and
Figure 1).

Pacing from any site in the RV did not change any of the LV diastolic
function parameters when compared with baseline (Table 2). There
were no significant differences in diastolic measurements between
CRT and baseline or between CRT and either RV septal pacing site.

Effects of different pacing sites on


intra-ventricular dyssynchrony
Ts-SD was abnormal in 18 (82%) patients at baseline (Table 3). Right
ventricular apex pacing did not significantly change Ts-SD compared
with baseline. Pacing the RVmIVS or RVOT significantly decreased
Ts-SD (Table 3 and Figure 2a). Cardiac resynchronization therapy
also significantly reduced intra-ventricular dyssynchrony vs. baseline,
but there were no significant differences in Ts-SD between either of
the RV septal pacing sites or between CRT and either RV septal
pacing site.

Effect of different pacing sites on


inter-ventricular mechanical delay
At baseline 13 (59%) patients had inter-ventricular dyssynchrony, as
assessed by IVMD (Table 3). Inter-ventricular mechanical delay was
not changed by RVA pacing. Pacing from both RVmIVS and RVOT
resulted in a significant decrease in IVMD compared with baseline.
Similarly, CRT decreased IVMD compared with baseline. There
was no significant difference in IVMD between CRT and either of
the RV septal pacing sites, or between the two RV septal pacing
sites (Figure 2b).

Effects of different pacing sites on QRS


duration
All patients had LBBB with QRS duration (QRSd) .120 ms (mean
166, SD 26 ms) at baseline, and it remained .120 ms in all patients
at each of the pacing sites including CRT. QRS duration increased
with RV pacing from any site compared with baseline. QRS duration
was significantly narrower with any of the RV septal pacing sites compared with RVA, but there was no significant difference in QRSd
between the two RV septal pacing sites. Cardiac resynchronization
therapy resulted in a narrower QRSd compared with baseline or
with any of the RV pacing sites (Table 4 and Figure 3).

Downloaded from by guest on December 13, 2014

Table 1 Patient demographics

NYHA class III, n (%)

Effects of different pacing sites on left


ventricular systolic function

Effects of different pacing sites on left


ventricular diastolic function

Results

Characteristics

constant pacing during the study ranged between 70 and


100 b.p.m. Optimal AVD ranged between 70 and 160 ms (mean
111, SD 27). Patients demographics are shown in Table 1.

45 (23 69)

P = 0.008
P = 0.04
P = 0.02
P = 0.32

50

40
LVEF (%)

AAI, single atrial pacing; RVA, RV apex; RVmIVS, RV mid inter-ventricular septum; RVOT, RV outflow tract; CRT, cardiac resynchronization therapy; LVEF, LV ejection fraction; LVEDV, LV end diastolic volume; LVESV, LV end systolic volume; CO,
cardiac output; Sm, myocardial systolic velocity; E, E wave velocity; A wave velocity; DT, E wave deceleration time; Em, myocardial early diastolic velocity; cm/s, centimetre per second; ms, millisecond; SD, standard deviation.

0.44

0.43
0.75
0.31
0.05
0.60
0.89
0.74
31 (28 47)
36 (24 47)
32 (23 52)

0.73

E/Em (ratio), median (IQR)

40 (26 65)

0.30

0.44
0.23

0.30
0.20

0.72
0.51

0.51
0.21

0.26
0.88

0.64
250 + 73

1.1 + 0.7
1.0 + 0.5

184 + 53
188 + 50

1.1 + 0.5
0.9 + 0.4

DT (ms), mean + SD

184 + 65

1.1 + 0.7

179 + 67

E/A (ratio), mean + SD

2.6 + 0.7
Sm (cm/s), mean + SD

LV diastolic function

0.66

0.71
0.94
0.23
0.14
0.90
0.81
0.16
2.6 + 0.8
2.5 + 0.6
2.5 + 0.9

0.76

4.9 + 1.4

2.3 + 0.8

0.02
0.03
0.01
0.76
0.78
0.78
5.6 + 1.8
4.9 + 1.6
4.9 + 1.4

0.81

CO (l), mean + SD

4.8 + 1.4

0.26
0.12
0.12
0.43
0.56
0.37
132 + 53
142 + 53
142 + 51
150 + 40
146 + 41
LVESV (mL), mean + SD

0.83
0.01

0.57
0.67

0.03
0.008

0.61
0.65

0.04
0.02

0.74
0.86

0.32
36 + 7

204 + 66
207 + 68

32 + 5
32 + 6

206 + 62
207 + 47

28 + 7
29 + 7

205 + 52
LVEDV (mL), mean + SD

LVEF (%), mean + SD

.............................................................................................................................................................................................................................................

CRT
RVOT
RVmIVS

LV systolic function

RVmIVS vs. RVOT

RVmIVS vs. RVOT


RVOT vs. CRT

RV septal pacing vs. CRT

RVmIVS vs. CRT


CRT vs. AAI
RVOT vs. AAI

Pacing sites vs. baseline

RVA
AAI

RVA vs. AAI

Pacing modes
Baseline

RVmIVS vs. AAI

...................................................... ....................................................................... .......................................

P-values
Pacing sites

.................................................................... .........................................................................................................................................

M. H. A. Alhous et al.

30

mean

20

10
AAI

RVA

RVmIVS RVOT

CRT

Figure 1 Left ventricular ejection fraction at different pacing


sites.

Discussion
In patients with standard indications for CRT, pacing from the
RVmIVS or RVOT significantly increased LVEF and reduced the
degree of intra- and inter-ventricular dyssynchrony compared with
baseline, with no significant difference between these two RV
pacing sites. However, CRT resulted in a further significant improvement in LVEF.

Prior studies
Previous studies have reported conflicting results with regard to the
effects of RV septal pacing in heart failure patients. Some groups29,30
have found an improvement in LV systolic function and heart failure
class with RV septal pacing but others41 43 report no improvement
in LVEF. However, these early studies were performed on heterogeneous groups of patients, many of them with atrial fibrillation and with
differing degrees of LV systolic dysfunction. More importantly, the
patients in these studies did not fulfil the standard criteria for CRT,
whereas in the current study accepted indications for CRT were
strictly adhered to.6,7 More recently, Vlay et al. 44 documented the
clinical response of 22 patients who met the standard criteria for
CRT but in whom standard LV pacing was not possible and RVOT
pacing was used instead. All patients had improved NYHA functional
class, reduced levels of cyanosis, dyspnoea, and improved exercise
tolerance.
The mechanism whereby RV septal pacing improves the synchrony
and force of LV contraction in CRT candidates may be explained by
the earlier activation of the proximal LV through the thinner part of
the upper septum: similar to the mechanism whereby RV septal
pacing is thought to reduce dyssynchrony when compared with
RVA pacing in patients requiring a permanent pacemaker for bradyarrhythmia.22
The current study demonstrates that pacing from either RV septal
site results in less dyssynchrony than RVA pacing and a higher LVEF. It
also suggests that pacing from either septal site has a similar impact on
traditional echocardiographic indices of cardiac performance or
more novel parameters, including Ts-SD and IVMD. Moreover,
there was no significant difference in QRSd between RVmIVS and
RVOT pacing sites. These findings may offer practical flexibility as

Downloaded from by guest on December 13, 2014

Table 2 Left ventricular systolic and diastolic function: changes in LVEF with different RV pacing sites and CRT

Page 4 of 7

Pacing sites

P values

Baseline

Pacing modes

Pacing sites vs. baseline

AAI

RVA

...............................................................
.................................................
RVmIVS

RVOT

CRT

...............................................................................................................................................
..........................................................................

RV septal pacing vs. CRT

RVmIVS vs. RVOT

........................................

RVA vs. AAI

RVmIVS vs. AAI

RVOT vs. AAI

CRT vs. AAI

RVmIVS vs. CRT

RVOT vs. CRT

RVmIVS vs. RVOT

.............................................................................................................................................................................................................................................
Intra-ventricular dyssynchrony
Ts-SD (ms), mean + SD

50 + 19

43 + 14

39 + 17

37 + 17

34 + 15

0.12

0.04

0.006

0.002

0.17

0.17

0.91

Ts-12 (ms), mean + SD

155 + 58

133 + 50

118 + 51

110 + 44

104 + 41

0.14

0.002

0.01

0.002

0.20

0.35

0.57

SLD (ms), mean + SD

56 + 45

56 + 37

42 + 40

27 + 23

41 + 33

1.00

0.46

0.01

0.03

0.94

0.15

0.38

35 + 19

27 + 21

27 + 21

20 + 16

0.09

0.003

0.01

0.002

0.38

0.23

0.47

Inter-ventricular dyssynchrony
IVMD (ms), mean + SD

48 + 29

AAI, single atrial pacing; RV, right ventricle; RVA, RV apex; RVmIVS, RV mid inter-ventricular septum; RVOT, RV outflow tract; CRT, cardiac resynchronization therapy; Ts-SD, dyssynchrony index; SLD, septal to lateral delay; Ts-12, maximum delay
between 12 LV segments (6 basal & 6 mid); IVMD, inter-ventricular mechanical delay; SD, standard deviation; ms, millisecond; n: number of patients.

(a)

(b)

Ts-SD (ms)

100

80

60

40

20

100

80

60

40

20

pacing sites.

P = 0.04
P = 0.006

CRT

CRT

Page 5 of 7

P = 0.002

P = 0.002

RVmIVS RVOT

P = 0.01

RVmIVS RVOT

P = 0.003

RVA

P = 0.09

RVA

P = 0.12

AAI

AAI

Figure 2 (a) Ts-SD at different pacing sites. (b) IVMD at different

Downloaded from by guest on December 13, 2014

to where to place the pacing lead when a clinical decision is made to


pace the RV septum in this group of patients.
Importantly, although we found no significant difference in intraand inter-ventricular dyssynchrony between CRT and RV septal
pacing sites, CRT resulted in a further incremental improvement in
LVEF compared with RV septal pacing from either site. Furthermore,
CRT resulted in a narrower QRSd compared with RV septal pacing.
Atrio-ventricular delay optimization may also have contributed to
the improvement in LV function and LV dyssynchrony observed
during RV septal pacing. However, the lack of improvement in
these parameters during RVA pacing (compared with baseline AAI
measurements), despite AV delay optimization, suggests that the
site of RV pacing site is the predominant determinant.

Limitations of the study

The effects of different pacing sites were assessed by echocardiography and this may not reflect clinical outcome. This was an acute
study and a longer term study would be useful to demonstrate a sustained response. The number of patients in this study was small and
larger studies to confirm our findings and explore clinical end
points should be considered.

IVMD (ms)

RV septal pacing as alternative for failed LV lead implantation

Table 3 Intra- & Inter-ventricular Dyssynchrony: changes in Ts-SD and IVMD with different RV pacing sites and CRT

Page 6 of 7

Mean DRSd (ms)

0.59
,0.001

AAI, atrial pacing; RV, right ventricle; RVA, RV apex; RVmIVS, RV mid inter-ventricular septum; RVOT, RV outflow tract; CRT, cardiac resynchronization therapy; QRSd, QRS duration.

,0.001
0.003
0.004
0.02
,0.001
144 + 20
178 + 19
180 + 22
196 + 25
166 + 26

.............................................................................................................................................................................................................................................

160

110

60

QRSd (ms),
mean + SD

AAI

RVA

RVmIVS

RVOT

CRT

RVA vs.
AAI

RVmIVS vs.
AAI

RVOT vs.
AAI

CRT vs.
AAI

RVmIVS vs.
CRT

RVOT vs.
CRT

RVmIVS vs.
RVOT
RVmIVS vs.
RVOT
RV septal pacing vs. CRT
Pacing sites vs. baseline

...................................................... ........................................................................ .......................................

Baseline

Pacing modes

P values

.................................................................... ...........................................................................................................................................

Pacing sites

210

AAI

RVA RVmIVS RVOT


Pacing site/Pacing mode

CRT

Figure 3 QRS duration at different pacing sites.

Conclusions
Cardiac resynchronization therapy is superior to either RV septal
pacing site and remains the gold standard. Nevertheless, these data
indicate that RV septal pacing from either site may be an option for
those patients in whom LV pacing is not possible for technical
reasons and is worthy of further study.
Conflict of interest: none declared.

References
1. Abraham WT, Hayes DL. Cardiac resynchronization therapy for heart failure. Circulation 2003;108:2596 603.
2. Cazeau S, Leclercq C, Lavergne T, Walker S, Varma C, Linde C et al. Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction
delay. N Engl J Med 2001;344:873 80.
3. Cleland JG, Daubert JC, Erdmann E, Freemantle N, Gras D, Kappenberger L et al. The
effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl
J Med 2005;352:1539 49.
4. Bristow MR, Saxon LA, Boehmer J, Krueger S, Kass DA, De MT et al.
Cardiac-resynchronization therapy with or without an implantable defibrillator in
advanced chronic heart failure. N Engl J Med 2004;350:2140 50.
5. Abraham WT, Fisher WG, Smith AL, Delurgio DB, Leon AR, Loh E et al. Cardiac
resynchronization in chronic heart failure. N Engl J Med 2002;346:1845 53.
6. Brignole M, Auricchio A, Baron-Esquivias G, Bordachar P, Boriani G, Breithardt OA
et al. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy:
the task force on cardiac pacing and resynchronization therapy of the European
Society of Cardiology (ESC). Developed in collaboration with the European
Heart Rhythm Association (EHRA). Europace 2013;15:1070 118.
7. Epstein AE, DiMarco JP, Ellenbogen KA, Estes NA III, Freedman RA, Gettes LS et al.
ACC/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on practice guidelines (writing committee to revise the ACC/
AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and
antiarrhythmia devices) developed in collaboration with the American Association
for Thoracic Surgery and Society of Thoracic Surgeons. J Am Coll Cardiol 2008;51:
e1 62.
8. van Veldhuisen DJ, Maass AH, Priori SG, Stolt P, van Gelder IC, Dickstein K et al. Implementation of device therapy (cardiac resynchronization therapy and implantable
cardioverter defibrillator) for patients with heart failure in Europe: changes from
2004 to 2008. Eur J Heart Fail 2009;11:1143 51.
9. Bisch L, Da CA, Dauphinot V, Romeyer-Bouchard C, Khris L, Mbaye A et al. Predictive factors of difficult implantation procedure in cardiac resynchronization therapy.
Europace 2010;12:1141 8.

Downloaded from by guest on December 13, 2014

Table 4 Changes in QRSd with different RV pacing sites and CRT

M. H. A. Alhous et al.

RV septal pacing as alternative for failed LV lead implantation

27. Lewicka-Nowak E, Dabrowska-Kugacka A, Tybura S, Krzyminska-Stasiuk E,


Wilczek R, Staniewicz J et al. Right ventricular apex versus right ventricular
outflow tract pacing: prospective, randomised, long-term clinical and echocardiographic evaluation. Kardiol Pol 2006;64:1082 91.
28. Cowell R, Morris-Thurgood J, Ilsley C, Paul V. Septal short atrioventricular delay
pacing: additional hemodynamic improvements in heart failure. Pacing Clin Electrophysiol 1994;17:1980 3.
29. Lieberman R, Padeletti L, Schreuder J, Jackson K, Michelucci A, Colella A et al. Ventricular pacing lead location alters systemic hemodynamics and left ventricular function in patients with and without reduced ejection fraction. J Am Coll Cardiol 2006;48:
1634 41.
30. Muto C, Ottaviano L, Canciello M, Carreras G, Calvanese R, Ascione L et al. Effect of
pacing the right ventricular mid-septum tract in patients with permanent atrial fibrillation and low ejection fraction. J Cardiovasc Electrophysiol 2007;18:1032 6.
31. Gorcsan J III, Abraham T, Agler DA, Bax JJ, Derumeaux G, Grimm RA et al. Echocardiography for cardiac resynchronization therapy: recommendations for performance and reportinga report from the American Society of Echocardiography
Dyssynchrony Writing Group endorsed by the Heart Rhythm Society. J Am Soc Echocardiogr 2008;21:191213.
32. Lieberman R, Grenz D, Mond HG, Gammage MD. Selective site pacing: defining and
reaching the selected site. Pacing Clin Electrophysiol 2004;27:8836.
33. Mond HG, Hillock RJ, Stevenson IH, McGavigan AD. The right ventricular outflow
tract: the road to septal pacing. Pacing Clin Electrophysiol 2007;30:482 91.
34. Morgan JM, Delgado V. Lead positioning for cardiac resynchronization therapy: techniques and priorities. Europace 2009;11:v228.
35. Schiller NB, Shah PM, Crawford M, DeMaria A, Devereux R, Feigenbaum H et al.
Recommendations for quantitation of the left ventricle by two-dimensional echocardiography. American Society of Echocardiography Committee on Standards,
Subcommittee on Quantitation of Two-Dimensional Echocardiograms. J Am Soc
Echocardiogr 1989;2:35867.
36. Appleton CP, Jensen JL, Hatle LK, Oh JK. Doppler evaluation of left and right ventricular diastolic function: a technical guide for obtaining optimal flow velocity
recordings. J Am Soc Echocardiogr 1997;10:271 92.
37. Yu CM, Sanderson JE, Marwick TH, Oh JK. Tissue Doppler imaging a new prognosticator for cardiovascular diseases. J Am Coll Cardiol 2007;49:1903 14.
38. Ypenburg C, Westenberg JJ, Bleeker GB, VAN de Veire N, Marsan NA,
Henneman MM et al. Noninvasive imaging in cardiac resynchronization therapy
part 1: selection of patients. Pacing Clin Electrophysiol 2008;31:1475 99.
39. Yu CM, Fung WH, Lin H, Zhang Q, Sanderson JE, Lau CP. Predictors of left ventricular reverse remodeling after cardiac resynchronization therapy for heart failure secondary to idiopathic dilated or ischemic cardiomyopathy. Am J Cardiol 2003;91:
684 8.
40. Inoue K, Okayama H, Nishimura K, Ogimoto A, Ohtsuka T, Saito M et al. Right ventricular pacing from the septum avoids the acute exacerbation in left ventricular dyssynchrony and torsional behavior seen with pacing from the apex. J Am Soc
Echocardiogr 2009.
41. Buckingham TA, Candinas R, Attenhofer C, Van HH, Hug R, Hess O et al. Systolic and
diastolic function with alternate and combined site pacing in the right ventricle. Pacing
Clin Electrophysiol 1998;21:1077 84.
42. Victor F, Mabo P, Mansour H, Pavin D, Kabalu G, de PC et al. A randomized comparison of permanent septal versus apical right ventricular pacing: short-term results.
J Cardiovasc Electrophysiol 2006;17:238 42.
43. Stambler BS, Ellenbogen K, Zhang X, Porter TR, Xie F, Malik R et al. Right ventricular
outflow versus apical pacing in pacemaker patients with congestive heart failure and
atrial fibrillation. J Cardiovasc Electrophysiol 2003;14:1180 6.
44. Vlay SC. Alternate site biventricular pacing: Bi-V in the RVis there a role? Pacing Clin
Electrophysiol 2004;27:5679.

Downloaded from by guest on December 13, 2014

10. Gras D, Leclercq C, Tang AS, Bucknall C, Luttikhuis HO, Kirstein-Pedersen A.


Cardiac resynchronization therapy in advanced heart failure the multicenter
InSync clinical study. Eur J Heart Fail 2002;4:31120.
11. van Rees JB, de Bie MK, Thijssen J, Borleffs CJ, Schalij MJ, van EL. Implantation-related
complications of implantable cardioverter-defibrillators and cardiac resynchronization therapy devices: a systematic review of randomized clinical trials. J Am Coll Cardiol
2011;58:995 1000.
12. Leon AR, Abraham WT, Curtis AB, Daubert JP, Fisher WG, Gurley J et al. Safety of
transvenous cardiac resynchronization system implantation in patients with chronic
heart failure: combined results of over 2000 patients from a multicenter study
program. J Am Coll Cardiol 2005;46:2348 56.
13. Biffi M, Exner DV, Crossley GH, Ramza B, Coutu B, Tomassoni G et al. Occurrence of
phrenic nerve stimulation in cardiac resynchronization therapy patients: the role of
left ventricular lead type and placement site. Europace 2013;15:77 82.
14. Yu CM, Wing-Hong FJ, Zhang Q, Sanderson JE. Understanding nonresponders of
cardiac resynchronization therapycurrent and future perspectives. J Cardiovasc
Electrophysiol 2005;16:1117 24.
15. Bleeker GB, Schalij MJ, Van Der Wall EE, Bax JJ. Postero-lateral scar tissue resulting in
non-response to cardiac resynchronization therapy. J Cardiovasc Electrophysiol 2006;
17:899 901.
16. Delgado V, van Bommel RJ, Bertini M, Borleffs CJ, Marsan NA, Arnold CT et al. Relative merits of left ventricular dyssynchrony, left ventricular lead position, and myocardial scar to predict long-term survival of ischemic heart failure patients
undergoing cardiac resynchronization therapy. Circulation 2011;123:70 8.
17. Ailawadi G, Lapar DJ, Swenson BR, Maxwell CD, Girotti ME, Bergin JD et al. Surgically
placed left ventricular leads provide similar outcomes to percutaneous leads in
patients with failed coronary sinus lead placement. Heart Rhythm 2010;7:61925.
18. Mair H, Sachweh J, Meuris B, Nollert G, Schmoeckel M, Schuetz A et al. Surgical epicardial left ventricular lead versus coronary sinus lead placement in biventricular
pacing. Eur J Cardiothorac Surg 2005;27:235 42.
19. Leclercq F, Hager FX, Macia JC, Mariottini CJ, Pasquie JL, Grolleau R. Left ventricular
lead insertion using a modified transseptal catheterization technique: a totally endocardial approach for permanent biventricular pacing in end-stage heart failure. Pacing
Clin Electrophysiol 1999;22:1570 5.
20. Garrigue S, Jais P, Espil G, Labeque JN, Hocini M, Shah DC et al. Comparison of
chronic biventricular pacing between epicardial and endocardial left ventricular
stimulation using Doppler tissue imaging in patients with heart failure. Am J Cardiol
2001;88:858 62.
21. Jais P, Douard H, Shah DC, Barold S, Barat JL, Clementy J. Endocardial biventricular
pacing. Pacing Clin Electrophysiol 1998;21:2128 31.
22. Alhous MH, Small GR, Hannah A, Hillis GS, Broadhurst P. Impact of temporary right
ventricular pacing from different sites on echocardiographic indices of cardiac function. Europace 2011;13:1738 46.
23. Wilkoff BL, Cook JR, Epstein AE, Greene HL, Hallstrom AP, Hsia H et al. Dualchamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial. JAMA
2002;288:3115 23.
24. Zhang XH, Chen H, Siu CW, Yiu KH, Chan WS, Lee KL et al. New-onset heart failure
after permanent right ventricular apical pacing in patients with acquired high-grade
atrioventricular block and normal left ventricular function. J Cardiovasc Electrophysiol
2008;19:136 41.
25. Sweeney MO, Hellkamp AS, Ellenbogen KA, Lamas GA. Reduced ejection fraction,
sudden cardiac death, and heart failure death in the mode selection trial (MOST):
implications for device selection in elderly patients with sinus node disease.
J Cardiovasc Electrophysiol 2008;19:1160 6.
26. Tse HF, Yu C, Wong KK, Tsang V, Leung YL, Ho WY et al. Functional abnormalities in
patients with permanent right ventricular pacing: the effect of sites of electrical
stimulation. J Am Coll Cardiol 2002;40:1451 8.

Page 7 of 7

Vous aimerez peut-être aussi