Académique Documents
Professionnel Documents
Culture Documents
William McMullen
Vice President
BIOPAC Systems, Inc.
Lesson 7
ECG & PULSE
Mechanical Action of the Heart
Peripheral Pressure Pulse
Plethysmography
Page 2
I. INTRODUCTION
The primary purpose of the heart is to pump blood throughout the body. To pump blood, the heart has a
rhythmical sequence of both electrical and mechanical events, the cardiac cycle. The electrical activity,
recorded as an electrocardiogram (ECG), initiates the mechanical activity of the heart (contraction and
relaxation of atria and ventricles). When the heart chambers contract, they pump blood to the next section of
the cardiovascular system. This lesson will focus on the actions of the left ventricle, which pumps blood to
the systemic circulatory system, producing a pulse.
During the cardiac cycle the electrical activity of the ventricles, as represented by the QRS complex of the
ECG, precedes the mechanical event of ventricular muscle contraction (ventricular systole). Within the
range of normal resting heart rates, systole begins at the time of the R wave peak and ends at the end of the T
wave. The T wave, which represents repolarization of the ventricles, occurs during the time the ventricles are
in systole. Ventricular diastole, a period of relaxation of ventricular muscles, begins at the end of systole
and lasts until the next R wave peak. Since each cardiac cycle contains one period of ventricular systole
immediately followed by one period of ventricular diastole, the duration of one cardiac cycle, or heartbeat,
can be measured as the time between successive R waves. (Fig 7.1) In the ECG cycle, electrical activity
precedes and initiates mechanical activity.
Fig. 7.1
Contraction of the ventricles (ventricular systole) pushes a volume of blood (stroke volume) into arteries.
From the left ventricle, the blood goes into the aorta and throughout the rest of the body. Each section of
blood bumps the downstream, neighboring section of blood to facilitate blood flow. The aorta and other
arteries have muscular walls, which allow the arterial walls to expand slightly to receive the volume of blood
during systole and then, elastic recoil of the arteries helps to continue pushing the blood through the rest of
the system. The arterial pressure throughout the cardiac cycle is the main force for blood flow.
The pumping action of the ventricles also initiates a pressure wave that is transmitted via the arterial walls.
The pressure increases with systole and decreases with diastole. The stiffness of the vessel walls helps
transmit the pressure wave. The stiffer the walls, the faster the transmission of the pressure wave, but the
more work is required by the heart to move the same blood volume.
When the pressure wave is transmitted to the periphery, e.g., fingertip, there is a pulse of increased blood
volume. The tissues and organs change in volume as blood vessels dilate or constrict and as pulses of blood
pass through the blood vessels during each cardiac cycle. Changes in blood volume of organs may be
brought about by the autonomic nervous system acting on the cardiovascular system, environmental factors
(such as temperature), metabolic activity of an organ, and a variety of other variables.
For example, temperature regulation involves controlling blood flow to the skin; when heat needs to be
conserved, blood flow to the skin is minimized and when excess heat is being generated, the opposite occurs.
Page 3
The actual blood flow is slower than the transmission of the pressure wave. The aorta has the fastest blood
flow in the body at approximately 40-50 cm/sec (approximately 1 mile per hour) whereas the speed of the
pressure wave can be much faster.
The travelling speed of the pressure wave from the heart to the periphery can be affected by many
interrelated factors, including the hearts ability to contract strongly, blood pressure, the relative elasticity of
the arteries, and the diameters of systemic arteries and arterioles. These factors change in response to body
positions, sympathetic nervous system input, emotions, etc. For example, the travelling speed of the pressure
wave has been shown to correlate with sympathetic influence and systolic blood pressure.
The study of blood volume changes within an organ by using volume displacement techniques is known as
plethysmography. In this lesson, you will simultaneously record the ECG and subsequent pulse. One type
of transducer used in plethysmography operates by converting light energy to electrical energy and thus is
called a photoelectric transducer. The photoelectric transducer works by shining a beam of light through the
skin and measuring the amount of light that is reflected. Blood absorbs light in a manner proportional to
blood volume. The greater the blood volume, the greater the light absorption, and vice-versa. The
photoelectric transducer converts the reflected light into electrical signals, which can then be processed and
displayed by the recorder.
III. MATERIALS
Page 4
For further explanation, use the online support options under the Help Menu.
A. SET UP
FAST TRACK Set Up
1. Turn the computer ON.
Fig. 7.2
4. Turn the MP35/30 Data Acquisition
Unit ON.
5. Place three electrodes on the Subject
(Fig. 7.3).
one on right forearm
(just above wrist)
Fig. 7.3
Set Up continues
Page 5
right forearm
WHITE lead
right leg
BLACK lead
(ground)
left leg
RED lead
Fig. 7.4
Each of the pinch connectors on the end of the electrode cable
needs to be attached to a specific electrode. The electrode cables
are each a different color. Follow Fig. 7.4 to ensure that you
connect each cable to the proper electrode.
The pinch connectors work like a small clothespin, but will only
latch onto the nipple of the electrode from one side of the
connector.
When the electrode cable is connected properly, the LEAD II
electrode configuration will be established.
7. Clean the window of the sensor.
Velcro strap
wraps around
finger
Set Up continues
Page 6
Position the electrode cables such that they are not pulling on
the electrodes or the transducer.
Connect the electrode cable clip (where the cable meets the
three individual colored wires) to a convenient location (can be
on the Subjects clothes). This will relieve cable strain.
The Subject should not be in contact with nearby metal objects
(faucets, pipes, etc.), and should remove any wrist or ankle
bracelets.
END OF SET UP
B.
Page 7
CALIBRATION
The Calibration procedure establishes the hardwares internal parameters (such as gain, offset, and scaling)
and is critical for optimum performance. Pay close attention to the Calibration procedure.
FAST TRACK Calibration
1. Double check the electrodes, and make
sure the Subject is relaxed.
2. Click Calibrate.
Fig. 7.6
There should be a greatly reduced ECG waveform with a
relatively flat baseline in the upper band. There should be
wavelike forms in the pulse band. If your data resembles
Fig. 7.6, proceed to the Data Recording section.
If the data shows any large spikes, jitter, or large baseline drifts,
or if you have a flat line instead of a clear signal on the pulse
channel, then you should redo the calibration by clicking Redo
Calibration and repeating the entire calibration sequence.
END OF CALIBRATION
Page 8
C.
Hints for obtaining optimal data:
To minimize muscle (EMG) corruption of the ECG signal and
baseline drift:
a) The Subject should keep still during all of the recording
segments because the recording from the pulse
transducer is sensitive to motion and the ECG recording
is sensitive to EMG artifact.
b) The Subject should be in a relaxed state for each
recording segment.
c) Initially, the Subjects forearms should be supported on
armrests.
d) After the Subject has repositioned his/her forearm,
check to make sure that the cable is not pulling on the
pulse transducer.
e) The recording should be suspended before the Subject
prepares for the next recording segment.
f) Make sure electrodes do not peel up.
Recording continues
Page 9
Segment 1
2. Click Record.
When you click Record, the recording will begin and an append
marker labeled Seated and relaxed will automatically be
inserted.
4. Click Suspend.
The recording should halt, giving you time to review the data
and prepare for the next recording segment.
If all went well, your data should look similar to Fig. 7.7 and you
can proceed to Step 6.
If correct, go to Step 6.
Recording continues
Page 10
9. Click Suspend.
The recording should halt, giving you time to review the data
and prepare for the next recording segment.
If all went well, your data should look similar to Fig. 7.8 and you
can proceed to Step 11.
Recording continues
Page 11
Segment 3
11. Subject remains seated, raises the
right hand (with transducer
attached), extends the arm above the
head, and holds that position.
If all went well, your data should look similar to the Fig. 7.9 and
you can proceed to Step 16.
END OF RECORDING
Remove the electrode cable pinch connectors, and peel off the
electrodes. Throw out the electrodes (BIOPAC electrodes are not
reusable). Wash the electrode gel residue from the skin, using
soap and water. The electrodes may leave a slight ring on the
skin for a few hours. This is normal, and does not indicate that
anything is wrong.
Page 12
V. DATA ANALYSIS
FAST TRACK Data Analysis
1. Enter the Review Saved Data mode.
Note Channel Number (CH)
designation:
Channel
Displays
CH 1
ECG
CH 40
Pulse
Fig 7.10
2. Setup your display window for optimal
viewing of the entire recording.
T (time interval)
CH 1
BPM (rate)
CH 1
p-p
CH 40
p-p
Be sure to zoom in far enough so that you can easily measure the
intervals between peaks, approximately 4 cardiac cycles.
Page 13
A
6. Repeat the above measurements for
each of the data segments.
A
Fig. 7.11
7. Using the I-Beam cursor, select the
area between two successive pulse
peaks (one cardiac cycle).
A
8. Repeat the above measurements for
each of the data segments.
Fig. 7.12
A
9. Select individual pulse peaks for each
segment and determine their
amplitudes.
B
Fig. 7.13
10. Using the I-beam cursor, select the
interval between the R-wave and pulse
peak.
C
11. Save or print the data file.
You may save the data to a drive, save notes that are in the
journal, or print the data file.
END OF LESSON 7
Page 14
Page 15
Lesson 7
DATA REPORT
Students Name:
Lab Section:
Date:
Height
Age
Weight
Gender:
Male
or
Female
Table 7.1
Condition
Arm Relaxed
Segment 1
Temp. Change
Segment 2
Arm Up
Segment 3
Measurement
R-R Interval
Heart Rate
Pulse Interval
Pulse Rate
R-R Interval
Heart Rate
Pulse Interval
Pulse Rate
R-R Interval
Heart Rate
Pulse Interval
Pulse Rate
Channel
CH 1
T
BPM CH 1
CH 1
T
BPM CH 1
CH 1
T
BPM CH 1
CH 1
T
BPM CH 1
CH 1
T
BPM CH 1
CH 1
T
BPM CH 1
Cycle I Cycle 2
Cycle 3
Mean
Page 16
Table 7.2
Measurement
Arm Resting
Temperatur
e
Arm Up
Segment 1
Segment 2
Segment 3
QRS Amplitude
CH1 p-p
Relative Pulse Amplitude (mV)
CH 40 p-p
II.
Questions
D. Referring to data in table 7.1, are the values of heart rate and pulse rate similar for each condition?
Yes / No
Explain why the values might differ or be similar.
E. Referring to Table 7.2 data, how much did the amplitude of the QRS complex change between
conditions?
Extreme temp Arm Resting?
__________ mV
Page 17
F. Referring to Table 7.2 data, how much did the pulse amplitude change between arm positions?
Extreme temp Arm Resting?
__________ mV
H. Describe one mechanism that causes changes in blood volume to your fingertip.
I.
Referring to data from section C of this report, how would you explain the difference in speed, if
any?
J.
Which components of the cardiac cycle (atrial systole and diastole, ventricular systole and diastole)
are discernible in the pulse tracing?
K. Would you expect the calculated pulse wave velocities of other students to be very close if not the
same as yours? Why or why not?
L. Explain any amplitude or frequency changes that occurred with arm position.