Académique Documents
Professionnel Documents
Culture Documents
AEROSOL DEVICES
Two types of aerosol devices are used for aerosol drug
delivery to ventilator-dependent patients: (1) nebulizers and (2)
pressurized metered dose inhalers (pMDIs). Types of nebulizers used for the treatment of intubated and mechanically
ventilated patients are jet nebulizers (JNs), ultrasonic nebulizers (UNs), and vibrating mesh nebulizers (VMNs). Figure 1
shows illustrations of aerosol devices used for the treatment of
ventilator-dependent patients.
Nebulizers
JNs have been commonly used for aerosol drug delivery
during mechanical ventilation since the advent of modern
mechanical ventilation in the mid-1950s. These relatively
inexpensive nebulizers are operated using 2 to 10 L/min of
compressed gas and cause entrainment often resulting in the
injection of additional gas into the ventilator circuit. This can
lead to an increase in volumes and pressures delivered during
mechanical ventilation, especially in smaller patients, if the
ventilator does not compensate for nebulizer gas flow entering
the circuit. Therefore, clinicians may be well advised to adjust
ventilator parameters during aerosol administration and return
the ventilator to pretreatment levels after the completion of
therapy. During operation of JNs, larger particles tend to
deposit on the baffles and walls of the nebulizer. Then, they
return to the reservoir to be renebulized causing an evaporative
effect that increases concentration of medication over time,
especially with continuous nebulization. Contamination is a
risk associated with the use of JNs during mechanical ventilation because the JN reservoir is in a dependent position to
and in direct contact with the ventilator circuit. This allows
fluids, secretions, and pathogens in condensate that forms in
the circuit to drain into the reservoir of the nebulizer where it
contaminates the medication, resulting in the production of
contaminated aerosol which is then delivered to the patient.
Unfortunately these least expensive aerosol generators
tend to be the least efficient for aerosol delivery during
mechanical ventilation. As JNs have large residual volumes of
up to 2 mL, a substantial portion of drugs placed in the nebulizer stays in the reservoir and cannot be delivered to ventilator-dependent patients. Consequently, with typical dose
volumes of 3 mL, the efficiency of JNs is much less than other
aerosol device options, which have lower residual drug
remaining in the nebulizer at the conclusion of dosing.
Whereas aerosol delivery of JNs ranges from 2% to 5.4%
during mechanical ventilation, delivery efficiency of VMNs,
UNs, and pMDIs are 8.7% to 33.3%, 8.7% to 17% and 2.5% to
www.clinpulm.com |
79
Aeron
eb
FIGURE 1. Variety of aerosol devices used for drug delivery to ventilator-dependent patients. From left to right: jet nebulizers, vibrating
mesh nebulizers, ultrasonic nebulizers, and pressurized metered dose inhalers.
80 | www.clinpulm.com
Copyright
DEVICE SELECTION
Previous evidence confirms similar therapeutic effects of
nebulizers and pMDIs with in-line adapter/spacers in ventilator-dependent patients when the drug is available in both
formats.1015 Such studies are largely based on bronchodilators. However, not all aerosol devices and interfaces provide
the same level of efficiency under all conditions. So the key
question is which device to use with which medication for any
specific ventilator circuit.
Drug Availability
The variety of drugs available with pMDIs is largely limited to bronchodilators and corticosteroids. In addition, the dose
volume of pMDIs (30 to 100 mL/actuation) is much smaller than
the 2 to 6 mL delivered with nebulizers. Nebulizers present the
opportunity for delivery of larger payloads of drugs to the lungs
with a broader range of inhaled medication including mucolytics,
mucokinetics, antivirals, antibiotics, diuretics, prostanoids, and
others. Consequently, the selection of aerosol devices for aerosol
administration during mechanical ventilation rests with the needs
of the patient, the intent of the clinician, and the desired therapeutic outcomes. In addition, if the drug is not available in a
pMDI, the nebulizer becomes the default aerosol generator
option. The need for efficient aerosol delivery increases with the
cost of any specific medication.
Drug Costs/
Dose
5d
VMN
5 d JN
125
89.70
10
16
0.40
0.25
1250
900
100
160
8
5
5000
3600
400
640
32
20
www.clinpulm.com |
81
Factors
Factors
Effects
Recommendations
Mode of
ventilation
The efficiency of
Prefer volume-controlled
nebulizer is lower in
ventilation when using
pressure-controlled
a nebulizer, if possible
ventilation than in
Use spontaneous modes,
volume-controlled
if tolerated by patients
ventilation
Spontaneous modes such
as CPAP increase
aerosol drug delivery
up to 23%
Tidal volume Low-tidal volume
Set tidal volume greater
decreases aerosol drug
than the volume of
delivery to the lower
tubing and ETT
respiratory tract
between the nebulizer
and the patient. For
adults it should be set to
>500 mL, if appropriate
for the patient
Inspiratory
Inspiratory time and duty Increase inspiratory time
time and
cycle are directly
while minimizing the
related to aerosol
duty cycle
degree of intrinsic
delivery during
PEEP. Duty cycle of
mechanical ventilation.
Z0.3 is recommended
As inspiratory time or
duty cycle increases,
more aerosol will be
delivered to the patient
Flow
Higher inspiratory flow
Use an inspiratory flow
increases turbulence,
of 30 to 50 L/min, if
transitional flow, and
tolerated, during
inertial impaction,
nebulization
thereby reducing
aerosol delivery
Bias flow
Bias flow dilutes aerosol Use lower bias flow
(breathgenerated proximal to
r2 L/min with
triggering
the patient but low
nebulizers when
mechanism)
flows may improve
possible
aerosol from VMNs and Synchronize actuations of
UNs placed proximal to pMDIs with inspiration
the patient
Inspiratory
Square waveform delivers Choose sinusoidal or
waveform
less aerosol than
decreasing ramp
decreasing ramp or
waveforms
sinusoidal waveforms
Reproduced with permission from Ari and Fink.16
DELIVERY TECHNIQUE
Recommendations
Place the largest diameter
ETT that the patient can
safely tolerate. Keep
ETT clear of secretions
Remove the HME before
aerosol drug
administration and place
it back into the circuit
after the treatment
Do not place nebulizers
between the HME and
patient airway
Do not turn off or remove
the heated humidifier
from the circuit.
Increase the dose, as
needed for clinical
response
82 | www.clinpulm.com
Effects
Copyright
TABLE 4. Device-related Factors, Their Effects, and Recommendations to Optimize Bronchodilator Delivery in Mechanically Ventilated
Adults
Factors
Effects
Position of aerosol
device in circuit
Nebulizer-related
factors
pMDI-related factors
Recommendations
inspiratory nebulization was less efficient than either continuous or intermittent expiratory aerosol generation. Intermittent inspiratory mode did not increase lung delivery but did
dramatically increased treatment time.27,28 Ventilators with
integrated nebulizer controls often compensate for the added
flow to the nebulizer to maintain tidal volume, inspiratory
pressures, and minute ventilation during aerosol delivery,
reducing the need to adjust parameters or alarms, during and
after nebulization.
It is important to prime the pMDI before first use and
when the canister has not been used for >24 hours to avoid
separation of drug from the propellants. Similarly, the pMDI
should be shaken before first actuation for each treatment.
Otherwise, this simple mistake in delivery technique will lead
to a decrease in aerosol deposition to ventilator-dependent
patients.29 Synchronizing pMDI with the beginning of inspiration, using a chamber spacer, and placing the spacer 6 inches
from the Y adapter in the ventilator circuit increases aerosol
drug delivery during mechanical ventilation.17,30,31
Previous studies show that mistakes in delivery technique
influence aerosol drug delivery in ventilated patients.22,3240 These
mistakes in technique are specific to the delivery device used
during treatment. For instance, when nebulizers are used for aerosol delivery during mechanical ventilation, clinicians may set gas
www.clinpulm.com |
83
JN:
Correctly assemble the nebulizer.
Fill the nebulizer with volume on drug
label. You may increase volume to 4-6 mL
to increase delivery.
Place the JN in the inspiratory line near
the Y adapter or proximal to the
ventilator.
VMN:
Correctly assemble the nebulizer.
Pour recommended drug volume into
the nebulizer.
Place the VMN in the inspiratory line
near the Y adapter or proximal to the
ventilator.
pMDI:
Shake and prime the pMDI.
Warm the pMDI canister to hand or
body temperature.
Place the pMDI spacer/adapter in the
inspiratory line near the Y adapter.
Correctly assemble the pMDI canister
with spacer/adapter.
JN:
Connect the nebulizer to a gas source.
If available, use gas source on the
ventilator in order to synchronize
nebulization with inspiration
Otherwise set gas flow to the JN at 6 to 8
l/min (or flow recommended on label).
Adjust ventilator volume or pressure limit
to compensate for added flow.
VMN:
Connect the nebulizer to a power source.
Turn on the power.
pMDI:
Coordinate actuations with beginning of
inspiration.
Do not remove or shake the pMDI
between actuations.
Wait 15 seconds between actuations.
FIGURE 2. An algorithm describing steps for optimal drug administration technique by each aerosol generator. Reproduced with
permission from Ari and Fink.16
flow improperly. Not shaking pMDI before treatment, not synchronizing pMDI actuations with the beginning of inspiration, or
not using the right type of spacer with pMDI will significantly
reduce aerosol drug delivery to ventilator-dependent patients. In
addition, improper positioning of the aerosol device in the ventilator circuit will influence efficiency of treatment during mechanical ventilation.33,34
84 | www.clinpulm.com
Copyright
18.
REFERENCES
23.
Copyright
13.
14.
15.
16.
17.
19.
20.
21.
22.
24.
25.
26.
27.
28.
29.
30.
31.
32.
33.
34.
www.clinpulm.com |
85
35.
36.
37.
38.
39.
86 | www.clinpulm.com
Copyright