Académique Documents
Professionnel Documents
Culture Documents
Project
Shell Project No: HP-3000-QAT
Project Document Number: T- 4.250.915
2MH127
Tag/Item Number(s):
N/A
Unit(s):
ASU
PCWBS:
LINDE ENGINEERING
Contractor Project No :
Project Rev.: B
Page 2 of 14
5887
09.11.08
FB
EO
AG
RFC
09.10.08
FB
EO
AG
IFR
02.12.07
BR
UA
MC /ZG
ICC
MS
MC
ZG
IFR
MS
MC
ZG
IIR
GAMA
Rev.
No.
Project
Rev
No.
Date
Description
Status
Document Title:
Contractor
LOC470
Gama Qatar Co. W. L. L.
GAM
PRO
QAC
GTL
028
Org.
Code
Doc.
Type
Disc.
Code
Geogr.
Area
Seq.
No
Rev.
No
Vendor Code
Rev No: B
Q A T A R
Doc. Title
GAM-PRO-QAC-GTL-028 Rev. E
PROCEDURE FOR CALIBRATION
Date
09.11.2008
TABLE OF CONTENTS
1.
SCOPE ________________________________________________________________________ 5
2.
PURPOSE _____________________________________________________________________ 5
3.
REFERENCES _________________________________________________________________ 5
4.
DEFINITIONS _________________________________________________________________ 5
5.
RESPONSIBILITIES ____________________________________________________________ 6
6.
PROCEDURE __________________________________________________________________ 6
6.1.
GENERAL REQUIREMENTS_______________________________________________________6
6.2.
6.3.
6.3.1.
6.3.2.
7.
6.4.
6.5.
6.6.
ATTACHMENTS ________________________________________________________________ 9
T-4.250.915 Rev. B
Page 4 of 14
Q A T A R
Doc. Title
GAM-PRO-QAC-GTL-028 Rev. E
PROCEDURE FOR CALIBRATION
Date
09.11.2008
1. SCOPE
This procedure describes the criteria and methods that shall be applied by GAMA in ensuring
the accuracy of the measuring and test instruments that shall be used in the erection activities
of the QSGTL Pearl Plant C2 Air Separation Units.
2. PURPOSE
The purpose of this Quality Control Procedure is to set a guideline of the testing requirements
as well as the responsibilities of implementing this procedure.
3. REFERENCES
T.4.096.674
PQP-08.F01
PQP-08.F02
4. DEFINITIONS
Company
Contractor
: Linde Engineering AG
Sub Contractor
: Gama Qatar
Project
ITP
QCP
QA/QC
ITL
MRR
HSE
T-4.250.915 Rev. B
Page 5 of 14
Q A T A R
Doc. Title
GAM-PRO-QAC-GTL-028 Rev. E
PROCEDURE FOR CALIBRATION
Date
09.11.2008
5. RESPONSIBILITIES
The QA/QC Manager is responsible for monitoring and implementing the following activities:
Establishing the calibration schedule for all equipment included in the system.
Keeping a record of all calibrated instruments.
Ensuring that instruments are calibrated at all times during the work.
Storage of calibration instruments, which are defined as secondary "MASTER".
Master instruments are generally stored in a locked-up cabinet under the control of the QA/QC
Manager. Particularly, over-sized instruments shall be kept at adequate areas and/or in proper
containers.
The above mentioned activities shall be carried on by the QA/QC Department together with
the department that uses the instrument.
6. PROCEDURE
6.1.
GENERAL REQUIREMENTS
Instruments and/or devices having appropriate calibration certificates can be utilised within the
expiry period. After the expiry dates, the calibration of these instruments and/or devices should
be calibrated.
The present procedure is applicable to instruments that require to be calibrated periodically,
such as pressure gages, hardness testers, paint thickness gages, optical measurement
devices, etc.
Calibration intervals for these instruments and/or devices are given in Attachment 1 according
to T.4.096.674 Calibration of Testing and Measuring Tools and Equipment.
6.2.
IDENTIFICATION OF INSTRUMENTS
This serial number will be placed on the instrument by such means that shall provide
durable identification. All inspection, measuring and test instruments with missing or
illegible serial numbers shall be re-calibrated prior to their use.
b)
All inspection, measuring and test instruments that are employed for the inspection or
acceptance of the construction work shall be marked with utilising stickers or tags, which
shall enable immediate visual confirmation of its calibration status.
T-4.250.915 Rev. B
Page 6 of 14
Q A T A R
6.3.
Doc. Title
GAM-PRO-QAC-GTL-028 Rev. E
PROCEDURE FOR CALIBRATION
Date
09.11.2008
CALIBRATION OF INSTRUMENTS
6.3.1.
Periodic Calibration
This calibration will be carried out before the defined expiry dates by the QA/QC
personnel or by a calibration service or agency under the control of the QA/QC
Department.
6.3.2.
Re-calibration
The instruments may be re-calibrated at any time before the expiry date of the calibration
if there are doubts as to the accuracy of the equipment.
6.4.
PRECISION OF CALIBRATION
The precision of the calibration shall be defined by the QA/QC Department in accordance with
the class of precision stated by the manufacturer of the instrument.
6.5.
METHOD OF CALIBRATION
The calibration of the devices shall be carried out according to the manufacturer instructions.
In case of the calibration procedure and/or instruction of the device manufacturer is not
available, then one of the following procedures shall apply:
a)
b)
c)
T-4.250.915 Rev. B
Page 7 of 14
Q A T A R
6.6.
Doc. Title
GAM-PRO-QAC-GTL-028 Rev. E
PROCEDURE FOR CALIBRATION
Date
09.11.2008
DOCUMENTATION OF CALIBRATION
A "control list" will be compiled for the measuring and test equipment and shall give the
relevant data.
The calibration label (Attachment 2) of the calibrated equipment shall indicate the acceptance
of the calibration. When this labelling is neither practical nor considered essential for the
control purposes, other procedures shall be established to ensure the conformity with this
requirement. Any other procedure shall be in compliance with the requirements of related
specifications and subject to approval.
Prior to use of any inspection, measuring and test instrument for the construction work, the
responsible engineer shall assure that the effective calibration period of the device has not yet
expired and the instrument has not been damaged to a visible degree.
T-4.250.915 Rev. B
Page 8 of 14
Q A T A R
Doc. Title
GAM-PRO-QAC-GTL-028 Rev. E
PROCEDURE FOR CALIBRATION
Date
09.11.2008
7. ATTACHMENTS
Attachment 1 Calibration Intervals
Attachment 2 Calibration Labels
T-4.250.915 Rev. B
Page 9 of 14
Q A T A R
T-4.250.915 Rev. B
Doc. Title
GAM-PRO-QAC-GTL-028 Rev. E
PROCEDURE FOR CALIBRATION
Date
09.11.2008
Page 10 of 14
Q A T A R
T-4.250.915 Rev. B
Doc. Title
GAM-PRO-QAC-GTL-028 Rev. E
PROCEDURE FOR CALIBRATION
Date
09.11.2008
Page 11 of 14
Q A T A R
Doc. Title
GAM-PRO-QAC-GTL-028 Rev. E
PROCEDURE FOR CALIBRATION
Date
09.11.2008
Calibration/
Functional
Test
Calibration
Period
(Months)
CAL
6 months
CAL
1 week
CAL
1 month
CAL
12 months and
each testing
CAL
12 month
CAL
2 years
CAL
12months
CAL
12 months
CAL
12 months
CAL
12 months
Equipment
Calibration
of
survey
instruments
Calibration of Relief valve
Calibration of MT yoke
Calibration
of
optical
measuring devices
Calibration of
ultrasonic
thickness measuring device
Calibration
of
surface
thermometer
Calibration of PMI machine
T-4.250.915 Rev. B
Remarks
Page 12 of 14
Q A T A R
Doc. Title
GAM-PRO-QAC-GTL-028 Rev. E
PROCEDURE FOR CALIBRATION
Date
09.11.2008
ATTACHMENT 2
CALIBRATION LABELS
CALIBRATED
INS
__________
Date :
___/___/___
__________
Sign
__________
S.NO
Due
CALIBRATION"
T-4.250.915 Rev. B
Page 13 of 14
Q A T A R
Doc. Title
GAM-PRO-QAC-GTL-028 Rev. E
PROCEDURE FOR CALIBRATION
Date
09.11.2008
HOLD
T-4.250.915 Rev. B
Page 14 of 14