Vous êtes sur la page 1sur 5

AMSL/CHD/2011-12

November 10, 2011

PROPOSAL II
(Digital Radiography System 800mA 03 Nos. Sl. No. 17, Department of Radiology)

The Head (P & CD),


HLL Lifecare Limited,
For and on behalf of Ministry of Health & Family Welfare,
Procurement and Consultancy Division,
B -14A, Sector 62,
NOIDA 201 301 (U. P.).
Ref.: - Your Tender Enquiry No. HLL/PCD/PMSSY/23/11-12 dated 11.07.2011, Corrigendum No. 1 dated
24.08.2011, Corrigendum No. 2, Corrigendum No. 3 & Corrigendum No. 4 for the Supply of Medical
Equipments at Medical Colleges under PMSSY Scheme, due on 17.11.2011 at 2:00 p.m. (IST), Techno
Commercial Tender to be opened on 17.11.2011 at 2:30 p.m. (IST).

MARS 80
(Digital Radiography System with U Arm Stand and Flat Panel based Detector)
TECHNICAL COMPLIANCE
Sr.
No.

SPECIFICATIONS ASKED IN
THE TENDER
Direct Digital X-ray unit with one
flat panel for complete range of
general radiography. It should be
possible to do all the general
radiological imaging like chest,
abdomen, musculoskeletal, joints,
PNS, Skull and Spine etc. with or
without movement of table/ Table
- tube assembly. It would have
following specifications:

1.

Generator

a)
b)

Microprocessor controlled high


frequency generator
Output 64 kW or more

c)

KVP range 40 KV to 150 KV.

d)

Should have automatic exposure


control with ionization chamber.

ALLENGERS SPECIFICATION

Page No.

Offered model MARS 80 is a Solid-State Flat


Panel based Digital Radiography X-ray unit
with a flat panel detector suitable for complete
range of general radiography. Unit is capable
to do all the general radiological imaging like
chest, abdomen, musculoskeletal, joints, PNS,
Skull and Spine etc. with or without
movement of table/ table-tube assembly.
Our offered unit has following features: -

Microprocessor controlled high frequency XRay generator will be provided.


Power output: 80KW
(Better Specs)
KV range: 40 KV to 150 KV.
Automatic exposure control device with
ionization chamber enables auto selection of
Radiographic factors, saves time, eliminates
retake, increases diagnostic capability and
lowers radiation dose is provided.

Page 1 of 5

2.

Table

a)

It should have a movable carbon


fibre or equivalent radiology
table with effective and fool
proof braking system
It should be possible to do all the
general radiological imaging like
chest and abdomen in supine and
erect position, all musculoskeletal
and joints imaging, PNS, Skull
and spine in AP, PA and lateral
views etc. with or without
movement of table in supine erect
or sitting position.
Cross lateral imaging for chest
should be possible.

b)

c)
3.

a)

The detectors should be able to do


all
routine
Radiography
examination. It should have
following Features:
Amorphous Silicon technology.

b)

Cesium iodide scintillation

c)

X-Ray sensing surface (Detector


Size) 42cm x 42 cm (16.8" x
16.8") or more.
Image Matrix 2500 x 2500 Pixel
or more.
Pixel size 160 micron or less

e)
f)
g)
h)
a)

Offered unit is capable to do all the general


radiological imaging like chest and abdomen
in
supine
and
erect
position,
all
musculoskeletal and joints imaging, PNS,
Skull and spine in AP, PA and lateral views etc.
with or without movement of table in supine
erect or sitting position.
Cross lateral imaging for chest is possible.

Detector system:

d)

Mobile table having table top of carbon fibre


material with effective and foolproof braking
system is provided.

The detector is able to do all routine


radiography examination.
It has following features:
Flat panel detector is based on Amorphous
Silicon (a-Si) technology.
Scintillator material: Cesium Iodide

X-ray sensing surface: 43cm x 43cm (17 x


17)
(Better Specs)
Image matrix: 3072 x 3072 pixels
(Better Specs)
Pixel Size: 143m.
(Better Specs)
Very high DQE (Digital Quantum DQE (Digital Quantum Efficiency): > 70%
Efficiency)
(Better Specs)
Spatial resolution 3 LP / mm. or Spatial resolution: 3.5 line pairs/mm
better.
Continuously adjustable source to Continuously adjustable source to image
image distance (SID) of 100 cm - distance (SID): 100cm to 180cm.
180 cm.
The tube assembly should follow U-Arm stand is provided.
the detector system for auto X-Ray tube and detector is fitted on U-Arm
anatomical positioning. It is stand due to which X-Ray tube is auto aligned
clarified that the bidders may with flat panel detector, at all times during
quote U arm type/C arm application.
type/ceiling suspended or floor
mounted models as long as other
parameters of specifications are
met.
Page 2 of 5

4.

Rotating Anode X-ray Tube

Rotating Anode X-ray Tube of


minimum 30 / 70 KW capacity,
Dual Focus:0. 6mm/1.2 mm, 9000
RPM. Anode heat storage capacity
minimum 300 KHU or more.

One No. Dual focus rotating anode X-Ray


tube with following focal spots is provided:

Operating console software should


be integrated with X-ray control
and should have facility to show
all X-ray parameters/does for
examination done.
Operating Console should have
facility for patient ID entry,
Viewing,
Processing
and
Documentation of images.
Should have monochrome medical
grade LCD Display monitor of 18"
size or more with antireflective
front screen and matrix size 1024
x 1024.
Diagnostic image should be
available 5 second after exposure.
Specify the time.
The next exposure should be
possible while processing is in
progress on the operating system.
Image storage capacity 30,000

Yes Provided.

Small focus: 0.6mm with power output


40KW.
Large focus: 1.2mm with power output
100KW.
Anode heat storage capacity: 300KHU.
Anode rotation speed: 10,000 rpm
(Better Specs)
Operating Console (To be placed in console room)

5.
a)

b)

c)

d)
e)
f)

Operating Console has facility for patient ID


entry,
Viewing,
Processing
and
Documentation of images.
One no. 19 monochrome medical grade LCD
monitor with antireflective front screen and of
matrix size 1280 x 1024 will be provided.
(Better Specs)
Diagnostic image is available in 3 seconds
after exposure.
(Better Specs)
The next exposure is possible while processing
is in progress on the operating system.

g)

Hard disc with image storage capacity of


30,000 images will be provided.
System should have inbuilt System has inbuilt CD/DVD writer with
CD/DVD writer with facility to facility to write images.
write images.

6.

Connectivity to Camera

Connectivity to Camera - The


bidder to provide connectivity to
dry chemistry camera already
available in the department.
Camera with 500 dpi or more to
be quoted separately as optional.

Offered unit can be connected to any existing


dry chemistry camera available in the
department.
1 No. Dry chemistry camera of resolution
508PPI is quoted separately as optional in
price bid.

Page 3 of 5

7.

DICOM compatibility.

8.

It should be DICOM compatible


to receive, send, print images. It
should be ready for connectivity
for any network computer.

Unit is DICOM compatible to receive, send


and print images.
Facility of connectivity to any network
computer is available.

Accessories

One No. Voltage Stabilizer of 100 KVA with


spikes protectors for whole system is
provided.
(Better Specs)

The price of following optional items is


quoted separately in price bid.
Acceptable.

a)

The price of following optional


items to be quoted separately.
The final decision to purchase
these items will be taken at the
time of placing orders.
Image stitching software

Image stitching software

b)

Wireless microphone

One No. Wireless microphone

c)

Lead glass of size 80 x 100 cms

One No. Lead glass of size 80 x 100 cm

d)

500 CDS and 250 DVDs (Blank).

500 Nos. Blank CDs and 250 Nos. DVDs.

e)

There should be vertical stand


with
bucky
for
chest
examination (Price to be
quoted separately)

One No. Vertical Bucky Stand along with


oscillating grid of ratio 8:1, 85 lines / inch is
provided. Chest stand is provided to take
chest radiographs.
(Price quoted separately in price bid).

10.

Installation:

9.

Voltage stabilizer 75 KVA with


spikes protectors for the whole
system.
Optional Items:

11.

12.

The finished room will be provide


to the bidder
The bidder has to do all necessary
civil, electrical and mechanical
work for installation of machine.

Acceptable
Acceptable

Service Centre:
The company should have an
established Registered service
centre with address and phone
numbers in Delhi.

Address of our registered office at Delhi is enclosed.

Spares availability :
Company should give undertaking Undertaking regarding the spares availability for next ten
regarding the spares availability years is enclosed.
for next ten years.

Page 4 of 5

13.

Product datasheet:

All specifications to be provided


with original product data sheet.
All technical specifications should
be supported with original data
sheet highlighting the page
number in the compliance sheet.
Photocopy/computer print will not
be accepted.

Original product datasheet enclosed.

Turnkey:
X-Ray Installation 800mA

Acceptable as per details in tender documents quoted


separately in price bid.

Being an ISO-9001: 2008 and ISO 13485:2003 Company with its Products CE
Certified, Allengers ensure International Quality Standards for its Manufacture, Design, Supply,
Installation & Service of X-Ray Machine.

Radiation Protection: Our Unit is approved by AERB (Atomic Energy Regulatory


Board) for Radiation Safety.

Thanking you,
Yours faithfully,
For ALLENGERS MEDICAL SYSTEMS LTD.
(AUTHORISED SIGNATORY)

Page 5 of 5

Vous aimerez peut-être aussi