Académique Documents
Professionnel Documents
Culture Documents
Calibrator check
Author
Dr.-Ing. Kay Rethmeier | kay.rethmeier@omicron.at
Date
Sep, 2008
Related OMICRON Product
MPD 600
Application Area
Calibration
Version
v2.0
Document ID
ANP_10032_ENU
OMICRON
Page 1 of 7
Content
1
Please use this note only in combination with the related product manual which contains several important safety
instructions. The user is responsible for every application that makes use of an OMICRON product.
OMICRON electronics GmbH including all international branch offices is henceforth referred to as OMICRON.
OMICRON 2010. All rights reserved. This application note is a publication of OMICRON.
All rights including translation reserved. Reproduction of any kind, for example, photocopying, microfilming, optical
character recognition and/or storage in electronic data processing systems, requires the explicit consent of OMICRON.
Reprinting, wholly or in part, is not permitted.
The product information, specifications, and technical data embodied in this application note represent the technical
status at the time of writing and are subject to change without prior notice.
We have done our best to ensure that the information given in this application note is useful, accurate and entirely
reliable. However, OMICRON does not assume responsibility for any inaccuracies which may be present.
OMICRON translates this application note from the source language English into a number of other languages. Any
translation of this document is done for local requirements, and in the event of a dispute between the English and a nonEnglish version, the English version of this note shall govern.
OMICRON 2013
Page 2 of 7
Short Cut
1.
2.
3.
4.
5.
6.
OMICRON 2013
Page 3 of 7
As a next step select the Q-Card and set the Division Factor in the charge calibration menu
manually to 1 (fig. 2).
Set the Measuring Center Frequency to 1 MHz and the Bandwidth to 300 kHz after selecting
the Q-Card.
OMICRON 2013
Page 4 of 7
Select the V-Card and set the voltage trigger to Line. Six PD events for 50 Hz (see black Arrows in fig. 3) will
be visible in the main measuring window; the QIEC reading should fit to the selected output charge of the
calibrator (see red mark in fig. 3).
Use the Time-Window to visualize the pulse (blue mark in fig. 3). The pulse-width and the
amplitude will depend on the selected charge level, polarity and on the type of calibrator
(CAL542 A-B-C-D). The oscilloscope settings can be changed in the PD-Card.
The frequency spectrum of the calibration pulse should be flat and linear up to a certain
cut-off frequency. Choose the FFT-Window and compare it with the curve exemplarily shown in
figure 4.
Please remember that any calibrator will only inject the displayed charge level if the FFT curve
is horizontal from nearly 0 MHz up to the selected measuring center frequency. Regarding to
IEC 60270:2000 (Chapter 6.1 figure 5) the upper cut-off frequency of your calibrator
has to be higher than the selected bandwidth of the PD measuring system.
OMICRON 2013
Page 5 of 7
Figure 5 shows the FFT-curve of a calibrator NOT DESIGNED TO BE USED over the whole
frequency range
Due to "Einbrche" notches in the frequency spectrum at 7 MHz the calibrator cannot inject
the displayed charge level at these frequencies. THE CALIBRATION WILL BE UNVALID in this
case; the PD measurements will led to FALSE RESULTS. This calibrator can only be used in a
frequency range of 0 MHz up to 2 MHz, which may be sufficient for many cases.
Calibrator OK?
As a High Quality Supplier of High-End PD Measuring Technique OMICRON expects that
Your CAL 542 will pass this system check at any time. In the unlikely event of greater derivations,
or if you are even uncertain about some results do not hesitate to contact OMICRON Technical
Support. We will be pleased to help you.
Even if this Self-Test procedure does NOT substitute the IEC recommended performance
checks it provides you quick and easy information about the status of your calibrator.
To gain reliable PD measuring results we recommend to invest these few minutes to check
your calibrator before every PD measurement.
OMICRON 2013
Page 6 of 7
Americas
OMICRON electronics Corp. USA
12 Greenway Plaza, Suite 1510
Houston, TX 77046, USA
Phone: +1 713 830-4660
+1 800-OMICRON
Fax: +1 713 830-4661
info@omicronusa.com
Asia-Pacific
OMICRON electronics Asia Limited
Suite 2006, 20/F, Tower 2
The Gateway, Harbour City
Kowloon, Hong Kong S.A.R.
Phone: +852 2634 0377
Fax: +852 2634 0390
info@asia.omicron.at
www.omicron.at www.omicronusa.com