Vous êtes sur la page 1sur 54

Advanced

Ad
d Topic:
T i Tips
Ti
and Tricks of LC User
Maintenance

Edward Kim
Application Engineer
February 19
19, 2010

Agilent 1200 Series LC Agilent Restricted


Page 1

2/18/2010

mA
U

Troubleshooting in HPLC
20
00
15
00
10
5
00
0
0
0

Page 2

1
0

1
5
Time
(min)

2
0

2
5

Agilent 1100/1200 Series System


Operating Principles

Pumping principle
Sampler principles
TCC principle
Detector principle

Agilent 1200 Series LC Agilent Restricted


Page 3

2/18/2010

Fast Serviceability Still Unmatched


Savings from E-PAC Design:
f60% reduction in mechanical
parts
f90% reduction in screw joints
f70% reduction in assembly time
f90% reduction in disassembly
time
E-PAC: Expanded Polypropylene foam

All pure materials for low cost recycling

Agilent 1200 Series LC Agilent Restricted


Page 4

2/18/2010

Agilent 1100/1200 Module


Serial Number Format
10 characters:

02274
DE

916

Where made?

Date code

JP = Japan
DE = Germany
US = USA
CN = China
FR = France

IIndividual
di id l module
d l counter
t
(how many have been made?)

(16st week of 1999)

Each module has it's own module counter.


The date code marks a major change in production
production.
The module counter starts at about 100 for each type of module
and continues in order regardless of changes to the date code.

Individual module counter higher 55000 refers to 1200 Series


lower 55000 to 1100Series.
Agilent 1200 Series LC Agilent Restricted
Page 5

2/18/2010

Degassers - Overview
G1322A Vacuum Degasser (aka Standard Degasser/Online Degasser)
G1379A Micro Vacuum Degasser 1100 Series (obsolete)
G1379B Micro Vacuum Degasser 1200 Series

Agilent 1200 Series LC Agilent Restricted


Page 6

2/18/2010

Standard degasser operating principle

Agilent 1200 Series LC Agilent Restricted


Page 7

2/18/2010

Avoid growth of algae

Agilent 1200 Series LC Agilent Restricted


Page 8

2/18/2010

Comparison of degassers
G1322A:
better for applications using methanol
recommended for use with refractive index detector (RID)
better for high flow rates (>5 ml/min)
G1379A/B
better in all other cases

Agilent 1200 Series LC Agilent Restricted


Page 9

2/18/2010

Maintenance
Cleaning Flush the solvent tubes with
isopropanol.
When using buffers, flush with water.
May
M h
have tto exchange
h
th
the vacuum pump,
sensor, solenoid valve, or chamber.
Replace solvent inlet frits
Solvent

Pump
10

Pumps - Overview

G1310A IIsocratic
ti P
Pump: one solvent
l
t ((might
i ht b
be mixture)
i t )
G1311A Quaternary Pump: four solvents, low pressure mixing
G1312A Binary Pump: two (four) solvents, high pressure mixing
[G1312B: binary pump SL (RRLC)
Special pumps for other purposes: cap, nano, prep systems]

Agilent 1200 Series LC Agilent Restricted


Page 11

2/18/2010

Isocratic pump

Agilent 1200 Series LC Agilent Restricted


Page 12

2/18/2010

Quaternary/binary pump

Quaternary Pump

Binary Pump

high pressure
mixing

multi-channel
gradient valve (MCGV)
l
low
pressure mixing
i i
Agilent 1200 Series LC Agilent Restricted
Page 13

2/18/2010

Pump options
Solvent Selection Valve (P/N G1312-60000)
G1312 60000)
Normal Phase Seals

(P/N 0905-1420)

Seal Wash

(P/N 01018
01018-68722)
68722)

Active Seal Wash

(P/N G1311-68711)

Agilent 1200 Series LC Agilent Restricted


Page 14

2/18/2010

Seal wash option


Seal Wash, active/passive

Agilent 1200 Series LC Agilent Restricted


Page 15

2/18/2010

System Pressure Test

Pressure tightness up to 400 bar of the system (this is no pump test!)


Ti ht system
Tighten
t
with
ith blank
bl k nut,
t preferably
f bl att TCC
Use isopropanole as solvent
Pressure increased to 390 bar, decrease lower than 2 bar/minute

Agilent 1200 Series LC Agilent Restricted


Page 16

2/18/2010

Leak test for quaternary pump

Pressure tightness of the quaternary pump


Tighten the pump outlet with a blank nut
Use isopropanole as solvent, flush system
y
prior to test (remove bubbles)
Steps:
Ramp 1: both plungers deliver for two strokes (1-2-1-2)
Plateau 1: plunger 2 with low flow (2 l/min)
Ramp 2: Plungers 2 and 1, high flow (153 l/min)
Plateau 2: Plunger 1 delivers with low flow (2 l/min)
Ramp 3, 4: Plungers 2-1-2-1 at high flow, up to 390 bar
tightness of pump? degassing efficiency?

2
1

problems at plateau 1
check OBV
OBV, plunger/seal 2
2,
purge valve, blanking nut
problems at plateau 2
check AIV, plunger/seal 1
also affected by chamber 2
Agilent 1200 Series LC Agilent Restricted

Page 17

2/18/2010

G1312A Leak Test

IF A2 & B2 failed:
everything after the both OBV could be
leaking
g including
g blank nut and p
purge
g valve
Agilent 1200 Series LC Agilent Restricted
Page 18

2/18/2010

Pump parameters
flow: volume delivered per time
time, range depends on pump type
composition: percentage for each solvent (bin./quat. pump)
stroke:
t k the
th volume
l
iin a range off 20 ll tto 100 ll off compressed
d
solvent delivered by one movement of piston 1
p
mixing
g but decreases lifetime of seals
a lower stroke improves
and valves
stroke
100 l

auto
settable stroke

auto: minimum possible


stroke for a chosen flow
((recommended setting)
g)

20 l

0.1 ml/min

5 ml/min

flow
Agilent 1200 Series LC Agilent Restricted

Page 19

2/18/2010

Pump parameters: compressibility


Compressibility: relative volume change per pressure change
= V/(Vp), e.g. water 46 10-6 bar, acetonitrile 115 10-6 bar,
isopropanole 100 10-6 bar, see manual/on-line help
is corrected by initial pump speed for avoiding pressure
fluctuations (pressure ripple should be below 1%)

Agilent 1200 Series LC Agilent Restricted


Page 20

2/18/2010

Using the pressure plot

piston 2 delivers

pressure (ripple)

piston 1 delivers

piston 2 delivers

pressure drop
p
p

Agilent 1200 Series LC Agilent Restricted


Page 21

2/18/2010

Optimization of compressibility settings

compressibility
ibilit setting
tti ttoo hi
high
h

compressibility setting too low

compressibility changes during gradients = changing composition


prior to mixing for quaternary pumps
Agilent 1200 Series LC Agilent Restricted
Page 22

2/18/2010

Terms for volumes in HPLC


Delay volume: volume from the first point of mixing to the head
of the column =>
> delay increases time needed for gradient to
arrive at column => elution is delayed
Dispersion volume: volume between the point of injection and
end of flow cell. Mixing dispersion causes peak broadening
The volume between point of injection and head of the column
contributes to both the delay and dispersion volume
volume, ii.e.
e peaks
come later and are broader
=> use short connections and small diameters (columns,
(
,
capillaries)

Agilent 1200 Series LC Agilent Restricted


Page 23

2/18/2010

Delay and dispersion volumes


for binary and quaternary pump

Quaternary pump: 800 l to 1100 l


Binary pump:

180 l to 480 l

Largest contribution by damper (1l/bar)

Agilent 1200 Series LC Agilent Restricted


Page 24

2/18/2010

Autosamplers - overview
G1313A Autosampler (obsolete)
G1329A Thermostated Standard Autosampler (1100/1200)
[G1329B: ALS SL (RRLC) high pressure variant]

G1367A Well-Plate
Well Plate Autosampler (1100 Series)
G1367B High-Performance Autosampler (1200 Series),HiP ALS
[1367C: HiP ALS SL (RRLC)]
[1367D HiP ALS SL+ (RRLC,
[1367D:
(RRLC llow carry over als)]
l )]

G1330A/B Autosampler Thermostat


[special samplers for other applications: micro autosampler
autosampler, micro wellplate sampler
sampler, dual-loop
autosampler, prep autosampler CTC]

Agilent 1200 Series LC Agilent Restricted


Page 25

2/18/2010

Standard ALS Principle

Mainpass
Pump

to Column

Draw Sample
(Bypass)

to Column

Inject
Pump

Pump

to Column

Agilent 1200 Series LC Agilent Restricted


Page 26

2/18/2010

HiP ALS Principle

Agilent 1200 Series LC Agilent Restricted


Page 27

2/18/2010

ALS (Std/HiP) Options


Extended Volume Metering Device 900l
MultiDraw Kit 400l & 1400l & (5000l prep)
Red Needle Seat Capillary 0,12mm ID

Agilent 1200 Series LC Agilent Restricted


Page 28

2/18/2010

Problems with Reproducibility Peak Areas

Peak Areas not


Reproducible

With peak height


Rotor seal cross-port
cross port leak or injection valve not tight
Piston seal of metering unit leaking
Needle partially blocked
With retention time
Variable pump flow rate
Other
Capillary
p
y from injector
j
to detector not tight
g
Detector equilibration problems

Page 29

Seal
Pump

Column
Rotor Seal

Thermostatted Column Compartment


- overview
Th
Thermostatted
t tt d Column
C l
C
Compartment
t
t G1316A TCC
[G1316B: TCC SL (RRLC)]
[G1316C: TCC SL+ (Method Devellopment Solution)]

Agilent 1200 Series LC Agilent Restricted


Page 30

2/18/2010

TCC principle
G1316A

Agilent 1200 Series LC Agilent Restricted


Page 31

2/18/2010

Agilent 1100/1200 Series Thermostatted


Column Compartment
G1316A

Options:
Column Switching Valve 6 or 10 Port
Specs:
Specs
10 degrees below ambient to 80 deg C
up to 80 deg C: max flow rate 5ml/min

Agilent 1200 Series LC Agilent Restricted


Page 32

2/18/2010

Agilent Capillary Color Code

C l
Color

id
i.d.

Red

0.12 mm

Green

0.17 mm

Blue

0.25 mm

Note: All fittings in the 1100


HPLC are Swagelok compatible,
EXCEPT the ports on model
77251 Rheodyne valve.

33

Fittings
g

Swagelok

Waters

0 130 iin.
0.130

0 090 iin.
0.090

Rheodyne

Parker

0.170 in.

0.090 in.

Valco

Uptight

0.080 in.

0.090 in.

Troubleshooting LC Fittings Part II,


II
J.W. Dolan and P. Upchurch, LC/GC Magazine 6:788 (1988)
34

Detectors
G1314A Variable Wavelength Detector (1100 Series) VWD obs.
G1314B Variable Wavelength Detector (1200 Series) VWD
[G1314C Variable Wavelength Detector SL (1200 Series) VWD SL]
[G1314D/E Variable Wavelength Detector SL+ (1200 Series) VWD SL+]

G1315B Diode-Array Detector (1100/1200 Series) DAD


[G1315C Diode-Array Detector SL (1200 Series) DAD SL
G1315D Diode
Diode-Array
Array Detector (1100/1200 Series) DAD SL]

G1365B Multiple-Wavelength Detector (1100/1200) MWD


[[G1365C/D Multiple-Wavelength
p
g Detector SL ((1200 Series)) MWD SL
G1321A Fluorescence Detector (1100/1200 Series) FLD
G1362A Refractive Index Detector (1100/1200 Series) RID
G4218A Evaporative Light Scattering Detector (1200 Series) ELSD]
Agilent 1200 Series LC Agilent Restricted
Page 35

2/18/2010

Lamp Matrix
Detector

New

Original

DAD

G1315A/B/C

MWD

G1365A/B/C

DAD SL

G1315D/E

MWD SL

G1365D/E

1290 Infinity

G4212A (New)

VWD

Secondary

VIS Lamp

2140-0813

5181-1530
(2140-0820)

G1103-60001

2140-0820

5181-1530
(2140-0813)

G1103-60001

5190-0917

NA

G1103-60001

G1314A/B/C

G1314-60100

2140-0813

G1103-60001

VWD SL+

G1314D/E

G1314-60101

2140-0820

G1103-60001

UV VIS

8453A

2140-0605

2140-0813

G1103-60001

CE

G1600A

2140-0585

NA

NA

CE

G7100A (New)

5190-0917

NA

NA

New SL and SL+ detectors shipped with RFID lamp


SL detectors can use lamps without RFID chip (Backwards compatible)
When using a secondary lamp in a VWD, the setting in the detector needs to be changed
The UV-VIS system can run with a long life lamp. The CE system can not.
Lamps with RFID chip can be used with old detectors, but no functionality of the chip.

Page 36

Group/Presentation Title
Agilent Restricted
Month ##, 200X

VWD principle

I0

I0
Agilent 1200 Series LC Agilent Restricted
Page 37

2/18/2010

VWD flow cells

Typical Column
Length

Typical Peak
Width

<= 5 cm

0.025

10 cm

0.05

20 cm

0.1

>= 40 cm

0.2

Recommended Flow Cell


Micro Flow Cell
Semimicro Flow
Cell

High-Pressure
Flow Cell

Standard Flow
Cell

For Pressure
Above 100 bar

Typical Flow Rate

0.05-0.2 mL/min

0.2-0.4 mL/min

0.4-0.8 mL/min

1-2 mL/min

Internal Column Diameter

1.0 mm

2.1 mm

3.0 mm

4.6 mm

0.05-5 mL/min

Agilent 1200 Series LC Agilent Restricted


Page 38

2/18/2010

Baseline stabilization

VWD baseline shows a slight


g drift for the first 60 minutes:

Agilent 1200 Series LC Agilent Restricted


Page 39

2/18/2010

VWD diagnostic tests

Agilent 1200 Series LC Agilent Restricted


Page 40

2/18/2010

Wavelength calibration
Or: how to calibrate the rotation angle of the grating
0

1
1
656 nm

=>
>T
Two positions
iti
needed:
d d zero order
d and
d fifirstt order
d 656 nm emission
i i liline
of deuterium lamp
required after repair, part replacements or for PM
the 656 nm position is checked each time the lamp is switched on
Agilent 1200 Series LC Agilent Restricted
Page 41

2/18/2010

VWD: Holmium oxide test

Similar to wavelength calibration, done for 3 absorption lines of holmium filter


Verifies success of wavelength calibration, used in OQ/PV and
after flow cell maintenance/repair
Agilent 1200 Series LC Agilent Restricted
Page 42

2/18/2010

VWD intensity test


Measures intensity for complete wavelength
range and for a flow cell filled with water.

lamp performance
cleanliness of flow cell
Agilent 1200 Series LC Agilent Restricted
Page 43

2/18/2010

Dark Current Test


Measures the detector noise (electronics)
while the lamp is off (dark current)
=> checks integrity of sample and reference photodiode, sample and
reference VWA boards and VWM ((main)) board

Digital/Analog Converter (DAC) Test

Converts a test voltage of 10 V to an analog signal, which should have a


noise below 3 V, otherwise D/A converter on VWD main board is defect
(requires board replacement)
Agilent 1200 Series LC Agilent Restricted
Page 44

2/18/2010

VWD Grating Motor Test


Checks the motor position steps for grating
movement. To be done if calibration or holmium filter tests fail.
Caused by defective grating motor, grating position sensor of
VWD main board

VWD Filter Motor Test


Checks
Ch
k the
h motor off the
h second
d order
d cut-off
ff filter.
fil
movement. To be done if holmium filter tests fails.
Caused by defective filter motor
motor, filter position sensor of VWD
main board

Agilent 1200 Series LC Agilent Restricted


Page 45

2/18/2010

VWD scan
During a method execution, a sample spectrum can be
acquired Therefore two scans are needed: a solvent/blank scan
acquired.
and a sample scan. The online spectrum is displayed as the
difference of both scans. The flow needs to be stopped
(flow = 0) during the sample scan.

Agilent 1200 Series LC Agilent Restricted


Page 46

2/18/2010

Flow Cells
Standard and Micro

Maximum pressure 40 Bar


Materials: SST, quartz, PTFE
Standard cell: 10mm, 14ul
Micro cell: 5mm,, 1ul
Gasket #2 has different
aperture

High Pressure

Maximum pressure 400 Bar


Materials: SST, quartz, Kapton
10mm, 14ul

High Pressure Cell

Standard and Micro Cell

Group/Presentation Title
Agilent Restricted
Month ##, 200X

Diode Array Detector (DAD) principle

I0

Measurement of I0?
Agilent 1200 Series LC Agilent Restricted
Page 48

2/18/2010

DAD settings: slit width

smaller slits increase the spectral resolution and

noise due to lower intensity


Agilent 1200 Series LC Agilent Restricted

Page 49

2/18/2010

DAD settings: peak width/


response time/time resolution

smaller peak widths


increase time resolution
improves separation of peaks
increase noise
increase amount of data

Agilent 1200 Series LC Agilent Restricted


Page 50

2/18/2010

DAD diagnostic tests


filter
filt ttestt
intensity test
dark-current
dark current test
slit test: checks intensity change when changing slit widths
holmium oxide test
spectral flatness test: maximum noise in spectrum < 2 mAU
ASTM noise test: noise over 20 minutes without flow cell < 0.02 mAU after
1 hour of equilibration
cell test: ratio of intensity in full range with and without cell
wavelength verification and re-calibration based on and emission lines
of deuterium lamp
test chromatogram
Agilent 1200 Series LC Agilent Restricted
Page 51

2/18/2010

DAD intensity test

Agilent 1200 Series LC Agilent Restricted


Page 52

2/18/2010

DAD flow cells

Typical Column
Length

Typical Peak
Width

<= 5 cm

0 025
0.025

10 cm

0.05

20 cm

0.1

>= 40 cm

0.2

Recommended Flow Cell


500 nl Flow Cell
Semimicro Flow
Cell

High-Pressure
Flow Cell

Standard Flow
Cell

Typical Flow Rate

0.01-0.2 mL/min

0.2-0.4 mL/min

0.4-0.8 mL/min

1-2 mL/min

Internal Column Diameter

0.3-1
0.3
1 mm

2.1 mm

3.0 mm

4.6 mm

0.05-5 mL/min

Agilent 1200 Series LC Agilent Restricted


Page 53

2/18/2010

Flow Cells
Standard and Semi-micro
Maximum pressure 120 Bar
M t i l SST,
Materials:
SST quartz,Teflon
t T fl
Apertures of front and rear
gaskets have different diameters
Standard cell uses cell body for
additional
dditi
l heat
h t exchange
h
Semi-micro cell: 6mm, 5uL
Standard: 10mm, 13uL

Micro
Maximum pressure 400 Bar
M t i l SST,
Materials:
SST quartz,
t Kapton
K t
6mm, 1.7ul

Mi
Micro
C
Cellll
Different apertures

Standard and Semi


Semi-Micro
Micro Cell
Group/Presentation Title
Agilent Restricted
Month ##, 200X