Users Manual
UW-48240
1/72
Contents
Safety Precautions
3
1.Before Using the Product
4
1.1. Model Codes
4
1.2. Checking Each Unit Type and Its Accessories 5
1.3. Check Before Use 5
2. About Assembly
6
2.1. Assembly Method 6
3. Block Diagram
7
4. Mounting the Product 8
4.1. Mounting the Product to the Panel 8
5. Terminal Description and Connection Method
9
5.1. Connecting Power 9
5.2. Connecting Input Signal
9
5.3. Connecting External Controls
13
5.4. Connecting Comparison Output
13
5.5. Connecting Analog Output 14
5.6. Connecting Serial Communication 14
6. Components and Their Functions
15
6.1. Multi Display
15
6.2. Single Display
16
7. Parameter Setup
17
7.1.Parameter Group 17
7.2.LED Display Method
17
7.3.Each Unit and Unit Number
17
7.4. Data lists and Default Settings
18
7.5. Differences Between Display Units 19
7.6.Operation Flow
20
7.7. Condition Data
21
7.8. Comparator Data 44
7.9. Scaling Data
46
7.10. Linearization Data
51
7.11. Calibration Data 52
8. Other Functions
54
8.1. The Display Shift Function 54
8.2. Monitor Mode
54
9. Control Functions
55
9.1. About Control Function
55
9.2. Hold Function
55
9.3. Digital Zero Function
55
9.4. Peak Hold Function
56
10. Output Functions
56
10.1. Comparison Output Function
56
10.2. Analog Output Function 56
10.3. RS-485 Interface Function
56
10.4. RS-232C Interface Function
56
11. Error Messages
57
12. Specifications and External Dimensions
58
12.1. Input Specifications
58
12.2. Common Specifications
62
12.3. Output Specifications
63
12.4. External Dimensions
64
13.Detailed Communications Specifications
65
13.1. Terminal Assignments and Connection Method
13.2. Communication Function Parameters
66
13.3. RS-485 Transmission/Reception Formats
66
13.4. Communication Commands
67
14. Warranty and After-service
72
14.1. Warranty
72
14.2. After-service
72
2/72
65
Safety Precautions
!
Caution
3/72
Model Codes
The model lineup of the A5000 series is shown below. Check that the model code and specifications
of your product match those you specified when ordering.
4/72
1.2
Unit Type
Power supply units
YA5100
Description
AC power unit
YA5200
DC power unit
Display units
YA5010-XX
Accessories
Case x 1, front panel x 1
Case-fastening hardware x 2,
Case-fastening hardware fixing screw x 2
Rear panel x 1, Rear panel-fixing screw x 4
Type label x 1, Connector label x 1, Blind seal x 1
2-pin connector x 1, 4-pin connector x 1, Operation
manual for the main unit (this manual) x 1
Case x 1, Front panel x 1
Case-fastening hardware x 2,
Case-fastening hardware fixing screw x 2
Rear panel x 1, Rear panel-fixing screw x 4
Type label x 1, Connector label x 1, Blind seal x 1
2-pin connector x 1, 4-pin connector x 1, Operation
manual for the main unit (this manual) x 1
Front sheet (without judgment monitor) x 1
Front sheet (with judgment monitor) x 1
Front sheet x 1
YA5020-XX
Output units
YA5001-XX
YA5002-XX
Multi-display unit
YA5003-XX
RS-232C unit
YA5004-XX
RS-485 unit
YA5005-XX
YA5006-XX
YA5007-XX
Input units
YA5001-01
YA5001-02
YA5001-03
YA5001-04
YA5001-05
YA5001-06
YA5001-07
YA5001-08
YA5001-09
YA5001-10
YA5001-11
YA5001-12
YA5001-13
YA5001-14
YA5001-15
YA5001-16
YA5001-17
YA5001-18
1.3
About Assembly
The A5000 series panel meter allows the customer to change over (add) a function by replacing the
corresponding unit with the desired one (or installing it).
2.1
Assembly Method
When replacing the unit, follow the figure below to assemble it into the meter.
To take each unit from the case, remove the hooks of the front panel from the hook-fastening holes
at the front of the side-faces of the case. (If an output unit has been installed, unscrew the output
unit-fixing screws [number 21 in the figure below] and then take this step.)
* Before replacing the unit, always disconnect the A5000 series panel meter from the power
supply.
No.
1
2
3
4
5
6
7
8
9
10
11
12
Description
Front sheet (for multi-display)
Front sheet (for single display)
Front panel
Display unit (for multi-display)
Display unit (for single display)
Power supply unit
Output unit
Printed board guides
Input unit
Case
Type label
Connector label
No.
13
14
15
16
17
18
19
20
21
22
23
Description
Case-fastening hardware
Case-fastening hardware fixing screw
Rear panel-fixing screw
Power-connecting connector
Control line-connecting connector
Input-connecting connector (with a terminal cover for 26 ranges)
Comparison output-connecting connector
Analog output-connecting connector
Output unit-fixing screw
Blind seal
Rear panel
6/72
7/72
POWER
POWER
HOLD
DZ
P/H
COM
LO
HI
Power unit
Power
circuit
Input
circuit
Input unit
A/D
Output unit
Power
circuit
Photo
coupler
Photo
coupler
microcomputer
9999
RS-232C driver
or
RS-485 driver
Analog
output
circuit
Relay
Relay
Relay
Display unit
RXD(+)
TXD(-)
SG
V-OUT
A-OUT
COM
LO-b
LO-c
LO-a
GO-c
GO-c
HI-b
HI-c
HI-a
Note that an input units connector and input circuit vary depending on the input unit to be installed.
3
Block Diagram
Caution
(1) The recommended panel thickness is 0.8 to 5mm. The tightening torque should be approximately
0.39 to 0.49Nm(4 to 5kgf cm).
.
(2) Do not install the instrument in locations exposed to direct sunligth, an environment where the
ambient temperature goes beyond the range of 0 to 50 or the ambient humidity goes beyond the
range of 35 to 85? , or places where dew condensation is formed due to rapid temperature changes.
(3) Do not install the unit where it is exposed to dust, particles, chemicals harmful to electric components,
corrosive gases, etc.
(4) When this unit is installed inside other equipment, pay attention to the heat radiation and keep the
heat inside the equiqment 50 or below.
(5) Do not install the unit where it is exposed to excessive vibration or shock.
(6) Install the unit horizontally; otherwise, ventilation will be adversely affected and may result in
deterioration.
8/72
Serial
communication
12 13 14 15 16 17 18 19
Analog output
24 25 26
20 22
21 23
10 11
External
control
Power
connector
* The overall length of each wire should be 30m or less with the exception of the power cable. If the
wiring exceeds 30m, the EN/IEC standards will not be met.
5.1
Connecting Power
10
No.
Name
10
POWER
11
11
Description
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
5.2
No.
Name
HI
NC
LO
Description
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
9/72
5.2.2
No.
Name
Description
12
13
14
15
LO
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
5.2.3
No.
Name
23
24
25
Description
Positive input terminal for range 23
(9.999 mA)
Positive input terminal for range 24
(99.99 mA)
Positive input terminal for range 25
(999.9 mA)
4
LO
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
5.2.4
No.
Name
Description
11-12
13
LO
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
5.2.5
No.
Name
Description
14
15
LO
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
10/72
5.2.6
No.
Name
Description
23
24
25
4
LO
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
5.2.7
No.
Name
HI
Input terminal
LO
Input terminal
Description
5.8 mm
or less
5.8 mm
or less
* For crimp-on terminals, use the type shown in the figure above.
5.2.8
No.
Name
HI
LO
+S
-S
LO
Description
To change the measurement method, change the ST1 socket position of the
resistance-measuring unit. For removal of the unit, see 2.1, Assembly Method.
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
11/72
5.2.9
5.2.10
No.
Name
NC
Description
Positive terminal for thermocouple
Do not connect this terminal.
Negative terminal for thermocouple
No.
Name
Description
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
5.2.11
No.
Name
Description
HI
LO
+15V
0V
COM
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
5.2.12
No.
Name
Description
HI
Input terminal
NC
LO
Input terminal
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
12/72
5.2.13
No.
Name
Description
+SIG
-SIG
+EXC
-EXC
COM
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
5.2.14
No.
Name
Description
V-IN
A-IN
LO
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
5.3
No.
Name
HOLD
DZ
PH
COM
Description
Control for hold function. Enabled when
short-circuited or at the same potential as
COM.
Control for digital zero function. Enabled
when short-circuited or at the same
potential as COM.
Control for peak hold function. Enabled
when short-circuited or at the same
potential as COM.
Common for all external control terminals.
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
5.4
12 13 14 15 16 17 18 19
No.
Name
Description
12
LO-b
13
LO-c
14
LO-a
15
GO-c
16
GO-a
17
HI-b
18
HI-c
19
HI-a
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
13/72
5.5
24 25 26
No.
Name
24
COM
Description
Common terminal for analog output
25
26
V-OUT
* Wires should be single wires of AWG28 to 12, or stranded wires of AWG30 to 12.
5.6
No.
Name
20
RXD(+)
21
Description
RS-232C: transmission;
RS-485: Non-reverse output
RS-232C: reception;
TXD(-)
RS-485: Reverse output
22
NC
23
SG
14/72
Multi Display
(2)
(4)
(1)
(5)
(3)
(6)
No.
Name
(2)
Judgment
indicators
GO
LO
ME
PH
Function
(3)
indicators
DZ
RE
(4) Sub-display 1
(5) Sub-display 2
(7)
(8)
(9)
Main Functions
During measurement
During parameter setup
IIndicates information on the
Indicates the measured value.
parameter to be set.
Indicates the result of judgment and
turns on if the measured value > HI
judgment value.
Indicates the result of judgment and
turns on if LO judgment value the
measured value HI judgment value.
Indicates the result of judgment and
turns on if the measured value < LO
judgment value.
Turns on if
digital zero backupis on.
Turns on if
peak hold/valley hold/
peak - valley holdis on.
Turns on if
digital zerois on.
Turns on if remote control is being
performed through RS-232C or RS485 interface.
Indicates the HI side judgment value.
Indicates the item in the maximum/
minimum/(maximum-minimum)/input
value monitoring mode.
Indicates the LO side judgment value. Indicates the item to be set.
Indicates information on the item in
the maximum/minimum/(maximumminimum)/input value monitoring mode.
Pressing the Enter and Mode keys
Returns to the measurement
together changes to the parameter
mode.
setting mode.
Pressing the Enter and Increment keys
together changes to the maximum/
minimum/(maximum-minimum)/input
value monitoring mode.
Switches from the maximum/
minimum/(maximum-maximum/
minimum/(maximum-minimum)/input
value monitoring mode to the
comparative judgment reading mode.
Pressing the Mode and Enter keys
together changes to the parameter
Selects the item to be set.
setting mode.
Pressing the Mode and Shift keys
together changes to the shift function
setup mode.
Pressing the Mode and Incremental
keys together turns on/off the Digital
zeroindicator.
Pressing the Shift and Enter keys
together changes to the parameter
Changes the digit to be set.
checking mode. (Comparator data can
be set.)
Pressing the Shift and Mode keys
together changes to the shift function
setup mode.
Selects from items in the maximum/
minimum/(maximum-minimum)/input
value monitoring mode. (Hold down the
key for about one second.)
Pressing the Increment and Mode keys Changes the value or content
together turns on/off the Digital
of a selected digit.
zeroindicator.
(Increments the value)
Pressing the Increment and Enter keys
changes to the maximum/minimum/
(maximum-minimum)/input value
monitoring mode.
Resets the maximum/minimum/
(maximum-minimum)/input value
monitoring mode. (Hold down the key
for about one second.)
15/72
6.2
Single Display
(2)
(6)
No.
Name
HI
(2)
Judgment
indicators
GO
LO
PH
DZ
Function
(3)
indicators
ME
RE
(1)
(7)
(8)
(3)
(9)
Main Functions
During measurement
During parameter setup
IIndicates information on the
Indicates the measured value.
parameter to be set.
Indicates information on the item in
the maximum/minimum/(maximumminimum)/input value monitoring mode.
Indicates the result of judgment and
turns on if the measured value > HI
judgment value.
Indicates the result of judgment and
turns on if LO judgment value the
measured value HI judgment value.
Indicates the result of judgment and
turns on if the measured value < LO
judgment value.
Turns on if peak hold/valley hold/
peak - valley holdis on.
Turns on if digital zerois on.
Flashes when linearization data output
values are set.
Turns on if digital zero backupis on.
Turns on if remote control is being
performed through RS-232C or RS485 interface.
Flashes when linearization data input
values are set.
Pressing the Mode and Enter keys
Returns to the measurement
together changes to the parameter
mode.
setting mode.
Pressing the Enter and Increment keys
together changes to the maximum/
minimum/(maximum-minimum)/input
value monitoring mode.
Switches from the maximum/
minimum/(maximum-maximum/
minimum/(maximum-minimum)/input
value monitoring mode to the
comparative judgment reading mode.
Pressing the Mode and Enter keys
Selects the item to be set.
together changes to the parameter
setting mode.
Pressing the Mode and Shift keys
together changes to the shift function
setup mode.
Pressing the Mode and Incremental
keys together turns on/off the
Digital
zeroindicator.
Pressing the Shift and Enter keys
together changes to the parameter
Changes the digit to be set.
checking mode. (Comparator data can
be set.)
Pressing the Shift and Mode keys
together changes to the shift function
setup mode.
Holding down the Shift key for about
one second moves to the HI judgment
value indicator.
Selects from items in the maximum/
minimum/(maximum-minimum)/input
value monitoring mode. (Hold down the
key for about one second.)
Pressing the Increment and Mode keys Changes the value or content
together turns on/off the
Digital
of a selected digit.
zeroindicator.
(Increments the value)
Holding down the Increment key for
about one second moves to the LO
judgment value indicator.
Pressing the Increment and Enter keys
changes to the maximum/minimum/
(maximum-minimum)/input value
monitoring mode.
Resets the maximum/minimum/
(maximum-minimum)/input value
monitoring mode. (Hold down the key
for about one second.)
16/72
Parameter Setup
7.1
Parameter Group
The A5000 series' parameters are broadly classified into groups depending on their use and
operational system. The groups of parameters are as shown in the table below.
Group name
Condition data
Comparator data
Scaling data
Linearize data
Calibration data
7.2
7.3
Description
A group of the parameters concerning basic operations (such as
measurement range power supply frequency, and sampling
speed) and the operations of special functions and optional
functions.
A group of parameters concerning comparison operations such
as HI/LO comparison judgment value and hysteresis.
A droup of parameters for setting the correlations between input
signals and a displays, the correlations between displays and
analog outputs etc.
A group of parameters concerning the linearize (linearity
correction) function.
A group of parameters concerning sensor calibrations when
strain gauge input units are mounted.
unit
Description
DC voltage measuring unit (range 11)
DC voltage measuring unit (range 12 to 15)
DC current measuring unit (range 23 to 25)
AC voltage measuring unit (Average rms range 11 to 13)
AC voltage measuring unit (Average rms range 14,15)
AC voltage measuring unit (True rms range 11 to 13)
AC voltage measuring unit (True rms range 14,15)
AC current measuring unit (Average rms range 23 to 25)
AC current measuring unit (Average rms range 26)
AC current measuring unit (True rms range 23 to 25)
AC current measuring unit (True rms range 26)
Resistance measuring unit
Temperature measuring unit (TC)
Temperature measuring unit (RTD)
Frequency measuring unit (Open collector,logic and magnet)
Frequency measuring unit (50 to 500Vrms)
Strain gauge input unit
Process input unit (4 to 20mA, 1 to 5V)
17/72
Output unit
NO.
Description
0
None
1
Comparison output
2
Analog output
3
RS-232C
4
RS-485
5
Comparison output and analog output
6
Comparison output, analog output and RS-232C
7
Comparison output, analog output and RS-485
7.4
Default Equipped
value
as
PH
*1
1
OFF
1
OFF
C
9600
7
E
2
CR.LF
00
OFF
OFF
CLR
OC
00
01
10
OFF
OFF
ON
01
02
03
04
05
06
07
11
15
25
13
15
13
15
25
26
11
12
13
14
15 16 17
18
25
26
14
B JPB 14 14
2A
*1
*1
OFS
*1
OIN
*1
1000
500
0
0
FIN
9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999
9999 9999 9999 9999 9999 9999 9999 9999 5000 9999 5000 9999
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
PS
Pre-scaling value setup
1
PPR
Frequency division setup
1
DLHI
Digital limiter HI value setup
9999
Linearization data
*5
Calibration data
ZERO Zero input value *6
0
SPIN
Span input value *6
2000
Shift data
SHF
Shift data setup
0
*1
Each value in the lower part of a cell in the columns on the right is the default value.
*2
Tracking zero width setup parameter is not indicated if the tracking time is set to OFF(0).
*3
5000 for 1 V range and 2000 for 2 A range
*4
1000 for 1 V range and 400 for 2 A range
*5
Linearization data are not set up for the default values.
*6
This value is not indicated if calibration is done using an actual load.
The shaded parts show the parameters that must be set for each unit.
18/72
9999
*3
*4
7.5
Setup contents
Parameter name
19/72
7.6
Operation Flow
The basic operation flow applied in setting each parameter of the A5000 series panel meter is as
shown below. The individual setting method and description of each parameter are separately
described later.
Calibration data
Measurement
Condition data
Comparator data
Scaling data
Linearization data
Pressing the ENTER key saves the data and returns to the measurement mode.
(Data are backed up with EEPROM even when the power is turned off.)
* The figure above pertains to "the multi-display unit," but the same operation flow applies to the
single-display unit.
* The figure above pertains to "the multi-display unit," but the same operation flow applies to the
single-display unit.
20/72
7.7
Condition Data
7.7.1
Setting the Peak Hold Type
This subsection lets you make settings relating to the A5000 series peak holding function. The
peak holding function holds the maximum value (peak hold), the minimum value (valley hold),
or the maximum and minimum values (peak and valley hold) by performing control through an
external control terminal, and it produces each output with respect to the relevant value.
List of the display and description (The * denotes the default.)
Display
Description
PH
Peak hold
*
VH
Valley hold
PVH
Peak valley hold
* The figure below shows how to set the peak hold type to valley hold.
Multi-display unit
or
21/72
7.7.2
Setting the Measurement Ranges
This subsection lets you make settings relating to the A5000 series measurement ranges. The
setting of this parameter differs depending on the input unit installed.
List of the display and description (The *denotes the default.)
Input unit No.01:DC voltage measuring unit
Display
Description
11
Range 11 (99.99mV)
Input unit
Display
*
12
13
14
15
Input unit
Display
23
24
25
Input unit
Input unit
Display
11
12
13
No.04:AC voltage measuring unit (AVG) Input unit No.05:AC voltage measuring unit (AVG)
No.06:AC voltage measuring unit (RMS) Input unit No.07:AC voltage measuring unit (RMS)
Description
Description
Display
Range 11 (99.99mV)
14
Range 14 (99.99V)
Range 12 (999.9mV)
15
Range 15 (600V)
*
Range 13 (9.999V)
*
Input unit
Input unit
Display
23
24
25
No.08:AC current measuring unit (AVG) Input unit No.09:AC current measuring unit (AVG)
No.10:AC current measuring unit (RMS) Input unit No.11:AC current measuring unit (RMS)
Description
Description
Display
Range 23 (9.999mA)
26
Range 26 (5A)
*
Range 24 (99.99mA)
Range 25 (999.9mA)
*
Input unit
Display
12
13
14
15
Input unit No.13:Temperature measuring unit (TC) Input unit No.14:Temperature measuring unit (RTD)
Display
Description
Display
Description
KA -50.0 to 199.9
PA -100.0 to 199.9
KB -50 to 1200
PB -100 to 600
J
-50 to 1000
JPA -100.0 to 199.9
T
-50 to 400
JPB -100 to 600
*
S
0 to 1700
R
-10 to 1700
B
100 to 1800
*
Input unit No.15:Frequency measuring unit
Input unit No.16:Frequency measuring unit
Display
Description
11 0.1 to 200Hz
12 1 to 2000Hz
13 0.01 to 20kHz
14 0.1 to 200kHz
* For letters of a message actually displayed, check them by referring to 7.2. LED Display Method.
22/72
The figure below shows how to set the measurement range of the DC voltage-measuring unit (12 to
15 ranges) to range 13.
1 sec.
or
* For units equipped with multiple measurement ranges, setting up (changing) a measurement
range also causes the terminal to which the relevant input signal is connected to be changed. (For
the terminal to be connected, see Chapter 5. Terminal Description and Connection Method.)
23/72
7.7.3
Setting the Number of Averaging Times
This subsection lets you make settings relating to the number of averaging times (sampling
rate) of the A5000 series. Setting of this parameter differs depending on the input unit installed.
List of the display and description (The * denotes the default.)
For input units other than the frequency measuring units
Display
Description
1
Single averaging (sampling rate12.5 times/s)
*
2
Double averaging (sampling rate6.25 times/s)
4
Four-time averaging (sampling rateapprox.3.1 times/s)
8
Eight-time averaging (sampling rateapprox.1.6 times/s)
Display
10
20
40
80
10-time
20-time
40-time
80-time
averaging
averaging
averaging
averaging
Description
(sampling rate1.25 times/s)
(sampling rateapprox. 0.6 times/s)
(sampling rateapprox. 0.3 times/s)
(sampling rateapprox. 0.1 times/s)
The figure below shows how to set the number of averaging times (sampling rate) of the
temperature-measuring unit (thermocouple) to 8 times (approx. 1.6 times/s).
1 sec.
or
* The sampling rate of the A5000 series power meter is controlled by the simple number of
samplings of the basic sampling rate (12.5 times/s).
* For cases where the change of a signal under measurement is slow, such as temperature
measurement, or cases where the instrument is used in an environment subject to significant
noise effects, reducing the number of averaging times (increasing the sampling rate)
inadvertently may cause the display to flutter.
* For the frequency-measuring units, the number-of-averaging-times parameter will not be
displayed.
24/72
7.7.4
Setting the Number of Moving Averaging Times
This subsection deals with settings relating to the number of moving averaging times of the
A5000 series. Unlike simple averaging, the moving averaging function offers filtering effects
without reducing the sampling rate.
List of the display and description (The * denotes the default.)
Display
Description
Display
Description
OFF
Without moving averaging
*
8
Eight-time moving averaging
2
Double moving averaging
16
16-time moving averaging
4
Four-time moving averaging
32
32-time moving averaging
The figure below shows how to set the number of moving averaging times to four times.
1 sec.
or
* Increasing the number of moving averaging times allows the filtering effects to improve. However,
this causes the response to a change of a transient input signal to be accordingly slow. Use this
function at an appropriate moving averaging count in consideration of the signal to be measured
and the number of averaging times (sampling rate) noted in the previous subsection.
25/72
7.7.5
Setting Step Wide
This subsection deals with settings relating to the A5000 series step wise function. The step
wide function forces the resolution of the least significant digit to change in order to prevent
display drift, etc.
List of the display and description (The * denotes the default.)
Display
Description
1
Least significant digit's resolution 1
2
Least significant digit's resolution 2
5
Least significant digit's resolution 5
0
Least significant digit's resolution 0 (resolution 1/10)
The figure below shows how to set the step wide function to "5."
1 sec
or
26/72
7.7.6
Setting up Display Blank
This subsection lets you make settings relating to the A5000 series display blank function. The
display blank function allows you to adjust the brightness of the display.
List of the display and description (The * denotes the default.)
Display
Description
OFF
Display blank function OFF
B-3
Display blank level 3 (slightly dark)
B-2
Display blank level 2 (dark)
B-1
Display blank level 1 (very dark)
ON
Display blank function ON (extinguished)
The figure below shows how to set the display blank function to display blank level 1.
1 sec.
or
* If the display blank function is activated, the main monitor and sub-monitor (only for the
multi-display type) are completely extinguished. (When setting a parameter, the display blank
function is OFF and the monitor(s) is lit.)
27/72
7.7.7
Setting the Baud Rate
This subsection lets you make settings relating to the baud rate of the A5000 series
communication functions. This parameter is displayed only when an output unit with
communication functions has been installed.
List of the display and description (The * denotes the default.)
Display
Description
9600
9600bps
*
4800
4800bps
2400
2400bps
38438400bps
19219200bps
The figure below shows how to set the baud rate to 4800 bps.
1 sec.
or
28/72
7.7.8
Setting the Data Length
This subsection lets you make settings relating to the data length of the A5000 series
communication functions. This parameter is displayed only when an output unit with
communication functions has been installed.
List of the display and description (The * denotes the default.)
Display
Description
7
7bits
*
8
8bits
The figure below shows how to set the data length to 8 bits.
1 sec.
or
29/72
7.7.9
Setting the Parity Bit
This subsection makes settings relating to the parity bit of the A5000 series communication
functions. This parameter is displayed only when an output unit with communication functions
has been installed.
List of the display and description (The * denotes the default.)
Display
Description
E
Even parity
*
O
Odd parity
N
No parity
The figure below shows how to set the parity bit to odd parity.
1 sec.
or
30/72
7.7.10
Setting the Stop Bit
This subsection lets you make settings relating to the stop bit of the A5000 series
communication functions. This parameter is indicated only when an output unit with
communication functions has been installed.
List of the display and description (The * denotes the default.)
Display
Description
2
2bits
*
1
1bit
The figure below shows how to set the stop bit to 1 bit.
1 sec.
or
31/72
7.7.11
Setting the Delimiter
In this subsection, you can make settings relating to the delimiter of the A5000 series
communication functions. This parameter is indicated only when an output unit with
communication functions has been installed.
List of the display and description (The * denotes the default.)
Display
Description
CR.LF CR and
*
CR.
CR.
1 sec.
or
32/72
7.7.12
Setting the Equipment ID
This subsection lets you make settings relating to the equipment ID of the A5000 series RS-485
function. This parameter is displayed only when an output unit with an RS-485 interface
function has been installed.
The figure below shows how to set the equipment ID to "10."
1 sec.
&
or
33/72
7.7.13
Setting the Analog Output Type
This subsection lets you make settings relating to the output type of the A5000 series analog
output function. This parameter is displayed only when an output unit with an analog output
function has been installed.
List of the display and description (The * denotes the default.)
Display
Description
OFF
Without analog output
*
0-1
0 to 1V output
0-10
0 to 10V output
1-5
1 to 5V output
4-20
4 to 20mA output
The figure below shows how to set the analog output type to 4 to 20 mA.
1 sec.
or
* Setting up (changing) the analog output type also causes the terminal to which connections are to
be made to change (for the terminal to which connections are to be made, see Chapter 5. Terminal
Description and Connection Method).
34/72
7.7.14
Setting Digital Zero Backup
This subsection deals with settings relating to the digital zero backup function of the A5000
series. The digital zero backup function writes a digital zero value to EEPROM (memory) at the
instant the digital zero terminal is turned ON. This stored value takes effect when the
instrument is started with the digital zero terminal turned ON.
The assured number of writes to the EEPROM is 100,000.
List of the display and description (The * denotes the default.)
Display
Description
OFF
Digital zero backup OFF
*
ON
Digital zero backup ON
The figure below shows how to set the digital zero backup function to ON.
1 sec.
or
35/72
7.7.15
Setting Linearize Function
In this subsection, you can make settings relating to the A5000 series linearize function. The
linearize function corrects the linear relation between input and display at any point to change
the gradient.
List of the display and description (The * denotes the default.)
Display
Description
OFF
Linearize OFF
*
ON
Linearize ON
The figure below shows how to set the linearize function to ON.
1 sec.
or
* This parameter sets whether to use the linearize function. For the specific method of setting up the
linearize function, see 7.10. Linearize Data.
36/72
7.7.16
Setting Input Switching
This subsection lets you make settings relating to the input type of the A5000 series
frequency-measuring unit. The input type function allows you to select the input type (open
collector, logic, or magnet) of the frequency-measuring unit. This parameter is displayed only
when the frequency-measuring unit (open collector, logic, or magnet) has been installed.
List of the display and description (The * denotes the default.)
Display
Description
O.C
Open collector input
*
LGC
Logic input
MAG
Magnet input
The figure below shows how to set input switching to logic input.
1 sec.
or
37/72
7.7.17
Setting Tracking Zero Time
This subsection lets you make settings relating to correction time of the A5000 series tracking
zero function. The correction time is set within the range of the number of sampling, 0 to 99. If
the correction time is set to "0", the tracking function is turned OFF.
The figure below shows how to set tracking zero time to "10."
* The default value is "0."
1 sec.
&
or
38/72
7.7.18
Setting Tracking Zero Width
This subsection lets you make settings relating to a correction width of the A5000 series
tracking zero function. This parameter appears only when the setting of tracking zero time has
been setup. The setting range is from 1 to 99 digits.
The figure below shows how to set the tracking zero width to 10 digits.
* The default value is "1."
1 sec.
&
or
39/72
7.7.19
Setting Sensor Power
In this subsection, you can make settings relating to the sensor power supply (supply voltage to
the sensor) of the A5000 series strain gauge input unit. This parameter appears only when the
strain gauge input unit has been installed in the meter.
List of the display and description (The * denotes the default.)
Display
Description
10
10V (30mA)
*
5
5V (15mA)
1 sec.
or
* The sensor power-connecting terminal is the same for either 10V or 5V. Check the maximum
application voltage of the load cell, etc. first and then set the sensor power to the correct voltage.
40/72
7.7.20
Setting Power-ON Delay Time
This subsection deals with settings relating to the A5000 series power-ON delay function. The
power-ON delay function stops operation for a given time when the power is turned ON (in that
case, all the displays become "----").
List of the display and description (The * denotes the default.)
Display
Description
Display
OFF
Power-ON delay function OFF
*
16
1
----displayed for approx. 1 second
17
2
----displayed for approx. 2 seconds
18
3
----displayed for approx. 3 seconds
19
4
----displayed for approx. 4 seconds
20
5
----displayed for approx. 5 seconds
21
6
----displayed for approx. 6 seconds
22
7
----displayed for approx. 7 seconds
23
8
----displayed for approx. 8 seconds
24
9
----displayed for approx. 9 second
25
10
----displayed for approx. 10 seconds
26
11
----displayed for approx. 11 seconds
27
12
----displayed for approx. 12 seconds
28
13
----displayed for approx. 13 seconds
29
14
----displayed for approx. 14 seconds
30
15
----displayed for approx. 15 seconds
Description
----displayed for approx. 16 seconds
----displayed for approx. 17 seconds
----displayed for approx. 18 seconds
----displayed for approx. 19 seconds
----displayed for approx. 20 seconds
----displayed for approx. 21 seconds
----displayed for approx. 22 seconds
----displayed for approx. 23 seconds
----displayed for approx. 24 seconds
----displayed for approx. 25 seconds
----displayed for approx. 26 seconds
----displayed for approx. 27 seconds
----displayed for approx. 28 seconds
----displayed for approx. 29 seconds
----displayed for approx. 30 seconds
The figure below shows how to set the power-ON delay time to approx. 10 seconds.
1 sec.
or
* If the power-ON delay time is set up, the operation is activated in the order of segment check
immediately after power ON, i.e., elapsing of delayed time, unit number display, and then
measurement operation.
41/72
7.7.21
Setting Protect Function
This subsection lets you make settings relating to the A5000 series protect function. The protect
function prevents changes being made to all of the parameters with the exception of condition
data.
List of the display and description (The * denotes the default.)
Display
Description
OFF
Protect OFF
*
ON
Protect ON
The figure below shows how to set the protect function to ON.
1 sec.
or
42/72
7.7.22
Setting Unit Number Display
This subsection lets you make settings relating to the A5000 series unit number display
function. The unit number display function displays the numbers of the installed units when
the power is turned ON.
List of the display and description (The * denotes the default.)
Display
Description
ON
Unit number display ON
*
OFF
Unit number display OFF
The figure below shows how to set the unit number display to OFF.
1 sec.
or
43/72
:
:
:
:
900
0
300
0
Indicated value
HI side judgment value
1000
900
500
300
LO side judgment value
0
Judgment
7.8.2
GO
HI
GO
LO
GO
:
:
:
:
900
200
300
150
Indicated value
HI side judgment value
1000
900
500
150 LO side hysteresis range
300
Judgment
GO
HI
G
O
LO
44/72
GO
7.8.3
Setting Method
The method of setting comparator data is common to all comparator data.
The display of each parameter, and the list of the description
Display
Description
Default
S-HI
HI side judgment value
1000
S-LO
LO side judgment value
500
H-HI
HI side hysteresis value
0
H-LO
LO side hysteresis value
0
The figure below shows how to set the HI judgment value to "900."
Multi display unit
&
or
* The setup conditions are HI side judgment value LO side judgment value, HI side judgment
value LO side judgment value + LO side hysteresis, and LO side judgment value HI side
judgment value-HI side hysteresis. If these conditions are not satisfied, an error indication
appears and the display returns to the HI side judgment value setup.
45/72
F S C
F I N
OFS
O I N
DLHI
DLLO
D E P
:
:
:
:
:
:
:
8000
8000
9999
0
0
9999
-9999
The place of 102 is lit.
Input
0
10V
7.9.1.2
Scaling Setting Example 2
Of the DC current-measuring units ranges (12 to 15 ranges), range 13 is used.
Display
5000
F S C
F I N
OFS
O I N
DLHI
DLLO
D E P
:
:
:
:
:
:
:
5000
9999
0
0
3000
-2000
The place of 103 is lit.
3000
DLHI
Input
0
-2000
10V
DLLO
* If a digital limiter is set up, a value exceeding the range of DLHI to DLLO is not displayed and is
held at DLHI (or DLLO). (Note that if an input signal exceeds the range, a range excess is
displayed.)
46/72
7.9.1.3
Scaling Setting Example 3
Of the DC voltage-measuring units ranges (12 to 15 ranges), range 13 is used.
Display
6000
F S C
F I N
OFS
O I N
DLHI
DLLO
D E P
:
:
:
:
:
:
:
6000
7800
1000
600
9999
-9999
The place of 102 is lit.
1000
Input
0.6V
7.8V
7.9.1.4
Scaling Setting Example 4
Of the process-measuring units ranges, range 2A is used.
Display
4000
3000
2000
F S C
F I N
OFS
O I N
DLHI
DLLO
D E P
:
:
:
:
:
:
:
-2000
20.00
2000
4.00
9999
-9999
The place of 101 is lit.
1000
20mA
Input
4mA
-1000
-2000
47/72
7.9.1.5
Scaling Setting Example 5
Of the DC voltage-measuring units ranges (12 to 15 ranges), range 13 is used.
Display
9999
Input voltage : 0 to 5V
D i s p l a y : 0 to 9999
F S C
F I N
OFS
O I N
DLHI
DLLO
D E P
:
:
:
:
:
:
:
9999
5000
0
0
9999
-9999
None
Input
0
5V
* In this example, the resolution of the least significant digit is (5V / 9999) = approx. 0.5mV,
exceeding the maximum resolution (1mV) of range 13. Thus, the indication of the least
significant digit will be removed (however, it can be set).
7.9.2
Display Scaling of the Frequency-measuring Units
When a frequency-measuring unit has been installed, the panel meter achieves display scaling
using the PS and PPR parameters (parameters FSC, FIN, OFS, and OIN are not indicated).
Determining the revolution speed (rpm) using the rotary encoder set to 30 pulses per minute:
(1)Determine the measurement range by calculating the maximum frequency.
The figure below shows an example where the revolution rises to a maximum speed of
about 100 rpm.
30
100
60
50
Number of pulses per second
Revolution speed per second
Number of pulses per revolution at the
rotary encoder
(2)Since the number of pulses determined in (1) is 50 per second (50 Hz), set the range to
range 11(for how to set the range, see the section on setting condition data).
(3)The display unit shows 500 if 50 Hz pulse input is measured under range 11 (when PS=1
and PPR=1 by default). Therefore, the parameters should be set as PS=2 and PPR=1 so that
the decimal point is positioned in the 101 digit(100.0 is is indicated 50 Hz input).
7.9.2.1
Example of Setting Pre-scale/Frequency Division
* The pre-scale/frequency division settable range is 0 < PS 5.000 or 1 PPR 100. (If PS 5 or
PPR 100 is set, the panel meter causes PS = 5.000 or PPR = 100 to be set).
* In this example, the least significant digit resolution is 0.05 Hz, exceeding the maximum
resolution (0.1 Hz) of range 11. Thus, the indication of the least significant digit is removed.
48/72
7.9.3
Analog Output Scaling
Analog output scaling is set up using the AOHI and AOLO parameters. Set the value indicated
when 1 V (10 V, 5 V, or 20 mA) is output to AOHI and the value indicated when 0 V (1 V or 4 mA)
is output to AOLO.
* The analog signal out of the setting range cannot be accurately output.
These parameters are displayed only when an output unit with an analog output function has
been installed.
7.9.3.1
Scaling Setting Example 1
Use the analog output type in 4 to 20mA.
Output
20mA
D i s p l a y : 0.0 to 500.0
O u t p u t : 4 to 20mA
A O H I : 5000
AOLO : 0
4mA
Display
0
5000
7.9.3.2
Scaling Setting Example 2
Use the analog output type in 0 to 10V.
Output
10V
D i s p l a y : 200.0 to -200.0
O u t p u t : 0 to 10V
A O H I : -2000
AOLO : 2000
Display
-2000
49/72
2000
7.9.4
Setting Method
The method of setting up scaling data is common to all scaling data.
Display
FSC
FIN
OFS
OIN
PS
PPR
Description
Full-scale indicated value
Full-scale input value
Offset indicated value
Offset input value
Pre-scale value
Frequency division value
Default
9999
9999
0
0
1.000
1
Display
DLHI
DLLO
AOHI
AOLO
DEP
Description
Digital limit HI set value
Digital limit LO set value
Analog output HI set value
Analog output LO set value
Decimal-point position
Default
9999
-9999
9999
0
None
The figure below shows how to set the full scale value to 8000.
Multi display unit
&
or
50/72
&
&
&
51/72
&
52/72
7.11.2
&
&
53/72
Other Functions
8.1 The Display Shift Function
The display shift function arbitrarily shifts the display only, without changing the gradient of the
input signal. Press the Shift key for approximately 2 seconds with the Mode key held down to enter
the setting conditions. Then use the Shift key and Increment key to set up any shift.
The figure below shows how to set up shifting of the indicated value by -15 digits.
Press the Shift key with the Mode key held down
during measurement operation.
&
54/72
Control Functions
9.1 About Control Function
The A5000 series panel meter has the hold, digital zero, and peak hold control functions.
5V
Input circuit
GND/LO
5V
10k
HOLD/
DZ/PH
COM
CPU
Indicated value
3
2
1
Display0
-1
-2
-3
Start of
Five samplings
tracking zero
function (Tracking zero time)
Setting example Tracking zero time setting 5 (correction at every five sampling)
Tracking zero width setting 3 (correction value
3 digits)
55/72
9.4
10 Output Functions
10.1 Comparison Output Function
The A5000 series of unit meters is designed so that the two judgment values HI and LO can be set
for the measured (indicated) value to provide the results of judgment as relay contact output.
(This function is effective when the meter is equipped with a comparison output unit.) For details
on the contact ratings and other specifications, refer to the section Output Specifications.
10.2 Analog Output Function
The A5000 series of unit meters can output an analog signal for an indicated value (when the
meter is equipped with an analog output unit). There are four output ranges, 0 to 1 V/0 to 10 V/1
to 5 V/4 to 20mA, from which a selection can be made using the condition data. In addition, the
analog output of the A5000 series allows for arbitrary output scaling. This scaling can be achieved
by setting the indication value for an output of the maximum scale value (20mA for 4-20mA output
range) in the AOHI parameter of the scaling data.
10.3 RS-485 Interface Function
The A5000 series can be equipped with an RS-485 interface (when the meter is provided with an
RS-485 unit). For details on the RS-485 function, see the 13. Detailed Communications
Specifications.
10.4 RS-232C Interface Function
The A5000 series can be equipped with an RS-232C interface (when the meter is provided with an
RS-232C unit). For details on the RS-232C function, see 13. Detailed Communications
Specifications.
56/72
11 Error Messages
The table below describes the inspection points, remedies, etc. taken if abnormal displays or erroneous
operating conditions occur during use of the A5000 series panel meter.
Didplay
Error description
Remedy
One of segments a
Note
If any of indications 1 to 6 appears frequently, the panel meter can be assumed to be
being affected by noise, etc. Take appropriate measures against noise.
57/72
Measurement
range
11
99.99mV
Highest
Input
Maximum
Accuracy
resolution impedance permissible input
Offset : 9999
About
10V
100V
(0.1% of FS)
Full scale : 0 to 9999
100M
Indication
12.1.2
Measurement
range
12
999.9mV
13
9.999V
14
99.99V
15
600.0V
Highest
Input
Maximum
Accuracy
resolution impedance permissible input
About
100V
100V
100M
About
1mV
250V
(0.1% of FS)
Offset : 9999
1M
About
Full scale : 0 to 9999
10mV
250V
10M
About
100mV
600V
(0.15% of FS)
10M
Indication
12.1.3
Measurement
range
23
9.999mA
24
99.99mA
25
999.9mA
Highest
Input
Maximum
Accuracy
resolution impedance permissible input
About
1A
100mA
10
(0.2% of FS)
Offset : 9999
About
10A
500mA
Full scale : 0 to 9999
1
About
100A
3A
(0.3% of FS)
0.1
Indication
12.1.4
Measurement
Highest
Input
Maximum
Indication
Accuracy
range
resolution impedance permissible input
99.99mV
10V
Offset : 9999
1M
100V
(0.2% of FS
999.9mV
100V
Full scale : 0 to 9999
or more
+10digit)
9.999V
1mV
250V
58/72
12.1.5
Measurement
range
14
99.99V
15
600.0V
Highest
Input
Maximum
resolution impedance permissible input
Indication
Offset : 9999
Full scale : 0 to 9999
10mV
100mV
1M
or more
250V
600V
Accuracy
(0.2% of FS
+10digit)
(0.3% of FS
+10digit)
12.1.6
Measurement
Highest
Input
Maximum
Indication
Accuracy
range
resolution impedance permissible input
99.99mV
10V
Offset : 9999
1M
(0.2% of FS
100V
999.9mV
100V
Full scale : 0 to 9999
or more
+20digit)
9.999V
1mV
250V
12.1.7
Measurement
range
14
99.99V
15
600.0V
Highest
Input
Maximum
resolution impedance permissible input
Indication
Offset : 9999
Full scale : 0 to 9999
10mV
100mV
1M
or more
250V
600V
Accuracy
(0.2% of FS
+20digit)
(0.3% of FS
+20digit)
12.1.8
Measurement
range
23
9.999mA
24
99.99mA
25
999.9mA
Highest
Input
Maximum
Accuracy
resolution impedance permissible input
About
1A
100mA
10
Offset : 9999
About
(0.5% of FS
10A
500mA
Full scale : 0 to 9999
1
+10digit)
About
100A
3A
0.1
Indication
59/72
12.1.9
Measurement
range
Indication
Highest
Input
Maximum
resolution impedance permissible input
26
5A
Offset : 9999
Full scale : 0 to 9999
1mA
(CT)
Accuracy
(0.5% of FS
+10digit)
8A
12.1.10
Measurement
range
23
9.999mA
24
99.99mA
25
999.9mA
Highest
Input
Maximum
Accuracy
resolution impedance permissible input
About
1A
100mA
10
Offset : 9999
About
(0.5% of FS
10A
500mA
Full scale : 0 to 9999
1
+20digit)
About
100A
3A
0.1
Indication
12.1.11
Measurement
range
Indication
Highest
Input
Maximum
resolution impedance permissible input
26
5A
Offset : 9999
Full scale : 0 to 9999
1mA
(CT)
8A
Accuracy
(0.5% of FS
+20digit)
12.1.12
Measurement
Highest
Indication
Circuit current
range
resolution
99.99
10m
About 5mA
999.9
Offset : 9999
100m About 500A
9.999k
Full scale : 0 to 9999
1
About 50A
99.99k
10
About 5A
Accuracy
(0.2% of FS)
60/72
12.1.13
Input
sensor
K
K
J
T
S
R
B
Highest
resolution
0.1
Indication
-50.0 to 199.9
-50 to 1200
-50 to 1000
-50 to 400
0 to 1700
-10 to 1700
100 to 1800
Accuracy
(0.5% of FS)
(0.2% of FS)
(0.6% of FS)
(0.4% of FS)
Note : The accuracy of range B is applicable
to temperatures of 500 or more.
12.1.14
Input sensor
PT100
JPt100
PT100
JPt100
Indication
Highest resolution
Accuracy
-100.0 to 199.9
0.1
(0.15% of FS)
-100 to 600
(0.3% of FS)
12.1.15
Measurement
range
0.1 to 200Hz
1 to 2000Hz
0.01 to 20kHz
0.1 to 200kHz
Input type
Open collector
Logic
Magnet
Accuracy
(0.2% of FS)
Maximumpermissible input
30V
15V
Duty ratio : 50
12.1.16
Measurement
range
0.1 to 200Hz
1 to 2000Hz
0.01 to 20kHz
0.1 to 200kHz
Input type
Voltage
Duty ratio : 50
61/72
Accuracy
(0.2% of FS)
12.1.17
12.1.18
Measurement
range
1V
1 to 5V
2A
4 to 20mA
Indication
Offset : 9999
Full scale : 0 to 9999
Input
Maximum
impedance permissible input
About
100V
100M
About
100mA
10
Accuracy
(0.2% of FS)
7-segment LED display (character height : 14.2mm on main display and 8mm on
sub-display)
Indication range
Over-range alarm
Decimal point
Zero indication
External control
External dimensions
Weight
: 450g
Withstand voltage
(AC power supply)
: 2000 V AC for 1 min. between power terminals and input terminal, and
between power terminals and each output terminal
Withstand voltage : 500 V DC for 1 min. between power terminals and input terminal, and
(DC power supply)
between power terminals and each output terminal
Withstand voltage
(common)
: 500 V DC for 1 min. between input terminal and each output terminal,
and between analog output terminal and communication terminals
2000 VAC for 1 min between case and each output terminal (common
to both AC and DC supply)
Insulation resistance
Conformity standard
: EN61000-6-2,EN50081-2,IEC1010-1
It applies only to the product with which CE mark is indicated
on the label.
Grounding environment
Altitude
Fuse
: 2000m max
: K19372 1.0A (DC power supply)
62/72
: -9999 to +9999
: Can be set in the range of 1 to 999 digits for each judgment value
Operating speed
Output method
: Relay contact output (Make and break contacts for HI and LO and make
contacts for GO)
: 240V AC, 8A (resistive load) and 30V DC, 8A (resistive load)
Output rating
Mechanical life
Electric life
12.3.2
(Resistance load)
Analog Output
Output type
Load resistance
0 to 1V
10k or more
0 to 10V
10k or more
1 to 5V
10k or more
4 to 20mA
550 or less
Accuracy
Ripple
50mVp-p
(0.5% of FS)
25mVp-p
Note: The ripple ratings for the 4 to 20mA output are when the load resistance of 250 and
the output current of 20mA are applied.
Conversion system
Resoluton
Scaling
: Equivalent to 13 bits
: Digital scaling
Response speed
12.3.3
: PWM conversion
Communicating function
Synchronization
Communication method
Transmission rate
Number of start bits
Data length
Error detection
RS-232C
RS-485
(Compatible with EIA RS-232C)
(Compatible with EIA RS-485)
Start-stop
Full-duplex
Two-wire half-duplex (polling/selecting)
2400,4800,9600,19200,38400 bps
1 bit
7 bits / 8 bits
Even parity, odd parity, or no parity
Block check character (BCC) checksum
1 bit / 2 bits
ASCII
Non-peocedural
TXD,RXD,SG
Non-inverting (+), inverting (-)
1
31 for meters
15m
Delimiter
63/72
64/72
Terminal Assignments
21 TXD(-)
22 NC
20 RXD(+)
23 SG
Terminal
No.
20
21
22
23
Description
Name
RS-232C/RS-485 connector
13.1.2
13
25
20 21 22 23
14
20 21 22 23
9-pin D-sub
(male)
25-pin D-sub
(female)
Pin 2: TXD
Pin 3: RXD
Pin 4: RTS
Pin 5: CTS
Pin 7: SG
Pin 23: SG
Pin 2: RXD
Pin 3: TXD
Pin 7: RTS
Pin 8: CTS
Pin 5: SG
Pin 23: SG
* The above-illustrated connection of the CTS and RTS terminals on the host side is only a
typical example for hardware control. Consult your system designer for further details on how
to cope with the terminals.
65/72
13.1.3
Example of RS-485 Connection
If the panelmeter is positioned to be an end station as the result of an RS-485 connection, set
the terminator to ON by using the selector socket in the A5000 output unit.
+
RS-485/
RS-232C
converter
SG
+
+
Note: If the panelmeter is an
end station, set the terminator to
ON.
Side-mounted
terminal block
Selector socket
ON OFF
Communication
connector
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
0 1 CR LF
Note: Set a two-digit number as the device ID (00 is void).
Char. Length
ENQ
EOTCR LF
Note: Communication is still possible when another device ID is specified without releasing
the communication link.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
0 1 CR LF
(Normal response)
Note: The response time is 40 ms maximum.
ACK
66/72
Char. Length
13.3.2
Hexadecimal
02H
03H
04H
05H
06H
Name
Start of Text
End of Text
End of Transmission
Enquiry
Acknowledge
Description
Marks the starting point of text.
Marks the ending point of text.
Marks the end of transmission.
Denotes an enquiry.
Denotes an affirmative reply.
13.3.3
BCC Checksum
As a means of error detection, a block check character (BCC) checksum is added to the RS-485
communication function of the A5000 panelmeter. See the following illustrations for details on
the transmission and reception formats.
Example of BCC Checksum for Transmission
STX
ETX
Character
D
S
P
Hexadecimal 02H 44H 53H 50H 03H
[1]
[2]
[3]
CR
LF
41H
45H
0DH
0AH
[4]
ETX
CR
LF
20H
20H
20H
35H
30H
30H
30H
20H
48H
49H
03H
39H
44H
0DH
0AH
[1]
[2]
[3]
[4]
[5]
[6]
[7]
[8]
[9]
[10]
[11]
1
D
2
S
3 4 5 6
P CR LF
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
Char. Length
S CR LF
M CR LF
H CR LF
A CR LF
S T H
S CR LF
(Sets the hold function to OFF.)
S T H
H CR LF
(Sets the hold function to ON.)
10
I CR LF
11
CR LF
10
I CR LF
11
CR LF
5 0 0 0
CR
(+ reading, w/o decimal point)
- 5 0 0 0
CR
(- reading, w/ decimal point)
< =
9 8 0 0
CR
(+ overrange reading, w/o decimal point)
< = - 9 8 0 . 0
CR
(- overrange reading, w/ decimal point)
All responses are the fixed length of 12 characters.
H I
(In HI, a judgment result is.)
G O
(In GO, a judgment result is.)
L O
(In LO, a judgment result is.)
All responses are the fixed length of 15 characters.
10
10
LF
12
LF
12
LF
12
LF
12
CR LF
15
CR LF
15
CR LF
15
S T A R T
CR LF
(Response with the status of the hold function being set to OFF by remote
control)
H O L D
CR LF
(Response with the status of the hold function being set to ON by remote control)
S T A R T
CR LF
(Response with the OFF status of the hold function)
H O L D
CR LF
(Response with the ON status of the hold function)
CR LF
CR LF
67/72
CR LF
Char. Length
T CR LF
M CR LF
11 12 13 14 15 16 17 18 19 20
CR LF
Trigger input
1 2 3 4 5 6 7 8 9 10
5 0 0 0
H I
(+ reading, w/o decimal point)
- 5 0 0 0
H I
(- reading, w/o decimal point)
5 0 0 . 0
H
(+ reading, w/ decimal point)
< =
9 8 0 0
H I
(+ overrange reading, w/o decimal point)
< = - 9 8 0 . 0
H
(- overrange reading, w/ decimal point)
P H
5 0 0 0
H I
(peak hold reading, w/o decimal point)
5 0 0 . 0
H I CR LF
(+ reading, w/ decimal point)
Note: The response is the same as that of gDSP h.
Y
CR LF
11
Function
Digital zero remote control response
1
D
2
Z
3 4 5 6
R CR LF
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
Char. Length
A CR LF
D Z R
O N CR LF
(Sets the digital zero function to ON.)
D Z R
O F F CR LF
(Sets the digital zero function to OFF.)
D Z R
1 0 0 0 CR LF
(Sets the digital zero function to ON at 1000.)
CR LF
CR LF
CR LF
CR LF
Y CR LF
CR LF
CR LF
CR LF
CR LF
CR LF
H CR LF
D CR LF
E
R
P
E
M CR LF
A CR LF
68/72
P V H
O F F CR LF
(Response with the OFF status of the peak hold function)
P V H
O N CR LF
(Response with the ON status of the peak hold function)
10
10
CR LF
CR LF
CR LF
CR LF
CR LF
P H
5 0 0 . 0 CR LF
(Response with the peak hold value)
V H
- 1 0 0 . 0 CR LF
(Response with the valley hold value)
P V
6 0 0 . 0 CR LF
(Response with the peak/valley hold values)
12
12
12
CR LF
CR LF
CR LF
M C L
M A CR LF
(Clears the MAX value.)
M C L
M I CR LF
(Clears the MIN value.)
M C L
M M CR LF
(Clears the (MAX-MIN) value.)
10
3
3
X CR LF
P V H
P H - O F F CR LF
(Response with the status of the peak hold function being set to OFF by
remote control)
P V H
V H - O F F CR LF
(Response with the status of the valley hold function being set to OFF by
remote control)
P V H
P V - O F F CR LF
(Response with the status of the peak/valley hold function being set to OFF by
remote control)
P V H
P H - O N CR LF
(Response with the status of the peak hold function being set to ON by
remote control)
P V H
V H - O N CR LF
(Response with the status of the valley hold function being set to ON by
remote control)
P V H
P V - O N CR LF
(Response with the status of the peak/valley hold function being set to ON by
remote control)
P C L
P H CR LF
(Clears the peak hold value.)
P C L
V H CR LF
(Clears the valley hold value.)
P C L
P V CR LF
(Clears the peak/valley hold values.)
M CR LF
P V H
O N CR LF
(Sets the peak hold function to ON.)
P V H
O F F CR LF
(Sets the peak hold function to OFF.)
P V H
P H CR LF
(Sets to the peak hold.)
P V H
V H CR LF
(Sets to the valley hold.)
P V H
P V CR LF
(Sets to the peak/valley hold.)
A CR LF
R L Y
O F F CR LF
(Response with the status of the comparison output being set to OFF by
remote control)
R L Y
H I CR LF
(Response with the status of the comparison output being set to ON for HI by
remote control)
R L Y
G O CR LF
(Response with the status of the comparison output being set to ON for GO by
remote control)
R L Y
L O CR LF
(Response with the status of the comparison output being set to ON for LO by
remote control)
M CR LF
R L Y
H I CR LF
(Sets to ON for HI.)
R L Y
G O CR LF
(Sets to ON for GO.)
R L Y
L O CR LF
(Sets to ON for LO.)
R L Y
O F F CR LF
(Sets all comparison outputs to OFF.)
D Z R
O F F CR LF
(Response with the OFF status of the digital zero function)
D Z R
O N CR LF
(Response with the ON status of the digital zero function)
Char. Length
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
D Z R
O F F CR LF
(Response with the status of the digital zero function being set to OFF by
remote control)
D Z R
1 0 0 0 CR LF
(Response with the status of the digital zero function being set to ON by
remote control)
Note: The main unit responds with the reading when the digital zero function is
set to ON.
Y E S
CR LF
N O
?
CR LF
(Response with the status of every function not being remote-controlled)
S T H CR LF
(Response with the status of the hold function being remote-controlled)
P V H CR LF
(Response with the status of the peak hold function being remote-controlled)
D Z R CR LF
(Response with the status of the digital zero function being remote-controlled)
R L Y CR LF
(Response with the status of the comparison output function being
remote-controlled)
Note: The main unit responds with the statuses of functions by separating
them with delimiters if multiple functions are being remote-controlled.
M A X
5 0 0 . 0 CR LF
(Response with a MAX value)
M I N
- 1 0 0 . 0 CR LF
(Response with a MIN value)
M - M
6 0 0 . 0 CR LF
(Response with a (MAX-MIN) value)
Note: The main unit responds with these values at one time by separating them
with delimiters.
5
5
3
3
3
3
9
9
9
CR LF
CR LF
CR LF
Function
Range response (except for
thermometers)
Range setting
Device ID response
Device ID setting
1
R
2
N
3 4 5 6
G CR LF
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
G CR LF
R N G
1 2 CR LF
(Sets to range 12.)
R N G
2 4 CR LF
(Sets to range 24.)
R N G
2 A CR LF
(Sets to range 2A.)
R N G
K A CR LF
(Sets to range KA.)
R N G
T CR LF
(Sets to range T.)
R N G
J P b CR LF
(Sets to range JPb.)
Note: The command varies depending on the input unit or the range of interest.
A
G CR LF
V CR LF
W D CR LF
S W D
1 CR LF
(Sets the step width to 1.)
S W D
1 0 CR LF
(Sets step width to 10.)
R
- CR LF
A D R
0 1 CR LF
(Sets the device ID to 01.)
A D R
9 9 CR LF
(Sets the device ID to 99.)
O
6
6
5
7
P CR LF
8
8
8
5
Y E S
CR LF
Note: The main unit switches to the range immediately after the setting.
Y E S
CR LF
Note: The main unit switches to the range immediately after the setting.
Y E S
CR LF
Note: The main unit switches to the range immediately after the setting.
Y E S
CR LF
Note: The main unit switches to the range immediately after the setting.
Y E S
CR LF
Note: The main unit switches to the range immediately after the setting.
Y E S
CR LF
Note: The main unit switches to the range immediately after the setting.
N O
?
CR LF
(Response when a nonexistent range is set.)
A V G
1 CR LF
(Response with the status of the averaging frequency being once.)
A V G
8 0 CR LF
(Response with the status of the averaging frequency being 80 times.)
CR LF
M A V
O F F CR LF
(Response with the status of moving average calculation being set to OFF.)
M A V
O N = 4 CR LF
(Response with the status of the moving average calculation frequency
being 4 times.)
M A V
O N = 1 6 CR LF
(Response with the status of the moving average calculation frequency
being 16 times.)
2
1
1
1
1
1
2
2
3
3
6
5
5
CR LF
CR LF
Y E S
CR LF
Note: The main unit switches to the OFF status immediately after the setting.
S . W D
1 CR LF
(Response with the status of the step width being 1.)
S . W D
1 0 CR LF
(Response with the status of the step width being 10.)
CR LF
7
5
Y E S
CR LF
Note: The main unit switches to the setpoint immediately after the setting.
1 9 2 0 0 - 7 - E - 2 - C R / L F CR LF
(Response with the status of the baud rate being 19200 bps, data length being 7
bits, parity being even, number of stop bits being 2, and delimiter being CR+LF.)
9 6 0 0 - 8 - N - 1 - C R CR LF
(Response with the status of the baud rate being 9600 bps, data length being 8
bits, parity being none, number of stop bits being 1, and delimiter being CR.)
17
21
17
3
CR LF
Y E S
CR LF
Note: The main unit switches to the setpoints immediately after the setting.
0 1 CR LF
(Response with the status of the device ID being 01)
9 9 CR LF
(Response with the status of the device ID being 99)
13
3
2
2
CR LF
CR LF
A . O U T
O F F CR LF
(Response with the status of the analog output type being OFF.)
0 - 1 CR LF
(Response with the status of the analog output type being 0-1 V.)
0 - 1 0 CR LF
(Response with the status of the analog output type being 0-10 V.)
1 - 5 CR LF
(Response with the status of the analog output type being 1-5 V.)
0 - 2 0 CR LF
(Response with the status of the analog output type being 0-20 mA.)
4 - 2 0 CR LF
(Response with the status of the analog output type being 4-20 mA.)
N O
?
CR LF
(Response with the status of the analog output unit being not installed.)
9
3
4
3
4
4
5
CR LF
CR LF
CR LF
CR LF
CR LF
CR LF
N O
?
CR LF
(Response with the status of the analog output unit being not installed.)
69/72
Char. Length
K A CR LF
(Response with the status of range KA being selected)
K b CR LF
(Response with the status of range KB being selected)
J CR LF
(Response with the status of range J being selected)
T CR LF
(Response with the status of range T being selected)
R CR LF
(Response with the status of range R being selected)
S CR LF
(Response with the status of range S being selected)
B CR LF
(Response with the status of range B being selected)
P A CR LF
(Response with the status of range PA being selected)
P b CR LF
(Response with the status of range Pb being selected)
J P A CR LF
(Response with the status of range JPA being selected)
J P b CR LF
(Response with the status of range JPb being selected)
Y E S
CR LF
Note: The main unit switches to the frequency immediately after the setting.
A O P
O F F CR LF
(Sets the analog output type to OFF.)
A O P
0 - 1 CR LF
(Sets the analog output type to 0-1 V.)
A O P
0 - 1 0 CR LF
(Sets the analog output type to 0-10 V.)
A O P
1 - 5 CR LF
(Sets the analog output type to 1-5 V.)
A O P
0 - 2 0 CR LF
(Sets the analog output type to 0-20 mA.)
A O P
4 - 2 0 CR LF
(Sets the analog output type to 4-20 mA.)
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
R A N G E
1 2 CR LF
(Response with the status of range 12 being selected)
R A N G E
2 4 CR LF
(Response with the status of range 24 being selected)
R A N G E
2 A CR LF
(Response with the status of range 2A being selected)
N O
?
CR LF
(Response with the status of no range being selected)
Note: The response varies depending on the input unit or the status of range
selection (8 characters max.).
R S 1 9 2 0 0 - 7 - E - 2 - C R / L F CR LF
(Sets the baud rate to 19200 bps, data length to 7 bits, parity to even, number of stop bits to
2, and delimiter to CR+LF.)
R S 9 6 0 0 - 7 - E - 1 - C R CR LF
(Sets the baud rate to 9600 bps, data length to 8 bits, parity to none, number of stop bits to 1,
A D R CR LF
M A V
4 CR LF
(Sets the moving average calculation frequency to 4 times.)
M A V
1 6 CR LF
(Sets the moving average calculation frequency to 16 times.)
M A V
0 CR LF
(Cancels moving average calculation (OFF).)
S
A V G
1 CR LF
(Sets the averaging frequency to once.)
A V G
8 0 CR LF
(Sets the averaging frequency to 80 times.)
M
Char. Length
Function
Digital zero backup status response
Protection response
Protection setting
1
B
2
D
3 4 5 6
Z CR LF
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
B D Z
O N CR LF
(Sets digital zero backup to ON.)
B D Z
O F F CR LF
(Sets digital zero backup to OFF.)
S
V CR LF
L CR LF
R CR LF
N CR LF
O CR LF
CR LF
CR LF
CR LF
O CR LF
C O M CR LF
Note: The reading of the main unit changes to COM the moment it receives a COM command.
N CR LF
3
3
3
5
CR LF
CR LF
CR LF
N O
?
CR LF
(Response with the status of the frequency measurement unit not being
installed.)
O N
T = 1
W = 1 CR LF
(Response with the status of the tracking zero time being 1 and tracking zero
width being 1.)
O N
T = 9 9
W = 9 9 CR LF
(Response with the status of the tracking zero time being 99 and tracking zero
width being 99.)
T R K
O F F CR LF
(Response with the status of the tracking zero function being set to OFF.)
10
12
CR LF
CR LF
CR LF
CR LF
CR LF
S N S R
1 0 CR LF
(Response with the status of sensor power supply being set to 10 V.)
S N S R
5 CR LF
(Response with the status of sensor power supply being set to 5 V.)
7
6
CR LF
CR LF
P O N
O F F CR LF
(Response with the status of the power-on delay function being set to OFF.)
P O N
1 CR LF
(Response with the status of the power-on delay function being set to 1 sec.)
P O N
3 0 CR LF
(Response with the status of the power-on delay function being set to 30 sec.)
7
5
6
CR LF
CR LF
CR LF
P R O
O F F CR LF
(Response with the status of the protection function being set to OFF.)
P R O
O N CR LF
(Response with the status of the protection function being set to ON.)
CR LF
CR LF
K E Y
O F F CR LF
(Sets the key operation prohibition function to OFF.)
K E Y
O N CR LF
(Sets the key operation prohibition function to ON.)
O . C CR LF
(Response with the status of the input selection option being open collector.)
L O G CR LF
(Response with the status of the input selection option being logic.)
M A G CR LF
(Response with the status of the input selection option being magnet.)
N O
?
CR LF
(Response with the status of the frequency measurement unit not being
installed.)
Y CR LF
N O
?
CR LF
(Response with the status of digital zero backup being OFF.)
P R O
O F F CR LF
(Sets the protection function to OFF.)
P R O
O N CR LF
(Sets the protection function to ON.)
U
P O N
1 CR LF
(Sets the power-on delay function to 1 sec.)
P O N
3 0 CR LF
(Sets the power-on delay function to 30 sec.)
P O N
0 CR LF
(Sets the power-on delay function to OFF.)
S N S R
1 0 CR LF
(Sets the sensor power supply to 10 V.)
S N S R
5 CR LF
(Sets the sensor power supply to 5 V.)
T R K
T = 1 CR LF
(Sets the tracking zero time to 1.)
T R K
T = 9 9 CR LF
(Sets the tracking zero time to 99.)
T R K
W = 1 CR LF
(Sets the tracking zero width to 1 digit.)
T R K
W = 9 9 CR LF
(Sets the tracking zero width to 99 digits.)
T R K
T = 0 CR LF
(Sets the tracking zero function to OFF.)
Char. Length
K CR LF
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
B D Z
O N CR LF
(Response with the status of digital zero backup being ON.)
B D Z
O F F CR LF
(Response with the status of digital zero backup being OFF.)
I S E L
O . C CR LF
(Sets the input selection option to open collector.)
I S E L
L O G CR LF
(Sets the input selection option to logic.)
I S E L
M A G CR LF
(Sets the input selection option to magnet.)
Char. Length
7
6
I
- 1 7 , O - 6 CR LF
(Response with the status of the input unit being strain gauge measurement
and the output unit being a combination of comparison output, analog output and
RS-232C.)
I
- 0 1 , O - 3 CR LF
(Response with the status of the input unit being DC voltage measurement (11
ranges) and the output unit being RS-232C.)
K E Y
O F F CR LF
(Response with the status of the key operation prohibition function being set to
OFF.)
K E Y
O N CR LF
(Response with the status of the key operation prohibition function being set to
ON.)
CR LF
CR LF
S - H I
1 0 0 0 CR LF
(Response with the HI-side judgment value.)
S - L O
5 0 0 CR LF
(Response with the LO-side judgment value.)
H - H I
0 CR LF
(Response with the HI-side hysteresis.)
H - L O
0 CR LF
(Response with the LO-side hysteresis.)
Y E S
CR LF
10
3
1
N CR LF
N CR LF
10
10
10
R CR LF
Note: The main unit returns to measurement operation upon an R command
(or returns to the HI-side judgment value if an N command is sent.)
C O M CR LF
Note: The reading of the main unit changes to COM the moment it receives a COM command.
8 0 0 0 CR LF
(Sets the HI-side judgment value to 8000.)
N CR LF
0 CR LF
10
0 CR LF
10
0 CR LF
10
4 0 0 0 CR LF
(Sets the LO-side judgment value to 4000.)
R CR LF
Note: If an R command is sent after setting required data, the main unit saves data provided
up to that moment and then returns to measurement operation.
0 CR LF
10
70/72
4
CR LF
E r r o r
CR LF
(Response when a value that does not meet the setting conditions is input.)
5
6
Function
Scaling data response
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
M E T CR LF
Note: The reading of the main unit changes to MET the moment it receives an MET command.
N CR LF
N CR LF
N CR LF
N CR LF
N CR LF
N CR LF
N CR LF
M E T CR LF
Note: The reading of the main unit changes to MET the moment it receives an MET command.
8 0 0 0 CR LF
(Sets the full-scale reading to 8000.)
N CR LF
2 0 CR LF
(Sets the offset reading to 20.)
R CR LF
Note: If an R command is sent after setting required data, the main unit saves data provided
up to that moment and then returns to measurement operation.
N CR LF
L I N
O F F CR LF
(Sets the linearization function to an OFF status.)
L I N
O N CR LF
(Sets the linearization function to an ON status.)
L I N
C L R CR LF
(Sets the linearization function to a cleared status.)
L N D
0 1 CR LF
Note: A reading can be made from any of the data items 01 to 16.
Note: The reading of the main unit changes to LINE the moment it receives an LND XX
command.
N CR LF
10
10
10
10
10
10
10
6
5
9 CR LF
10
0 CR LF
10
9 CR LF
10
0 CR LF
10
0 CR LF
10
CR LF
E r r o r
CR LF
(Response when a value outside the setpoint range is input.)
L I N
O F F CR LF
(Response with the status of the linearization function being set to OFF.)
L I N
O N CR LF
(Response with the status of the linearization function being set to ON.)
L I N
C L R CR LF
(Response with the status of the linearization function being cleared.)
7
6
7
CR LF
CR LF
CR LF
N O
?
CR LF
(Response with the status of the linearization function being cleared.)
Note: Since the linearization data are all cleared when the linearization function is
cleared, the main unit does not accept either a LIN ON or LIN OFF command.
(Set the linearization function status after setting the linearization data again.)
L N O
0 0 CR LF
(Response with the status of the linearization function being cleared.)
L N O
0 2 CR LF
(Response with the status of the number of linearization correction data items
being 02.)
L N O
1 6 CR LF
(Response with the status of the number of linearization correction data items
being 16.)
6
6
CR LF
CR LF
E r r o r
CR LF
(Response when linearization data is not correctly set.)
Note: Set the number of linearization correction data items after setting
linearization data.)
L N D
0 1
I =
0 CR LF
(Response with the input value of linearization data N-01.)
14
L N D
0 1
O =
0 CR
(Response with the output value of linearization data N-01.)
L N D
0 2
I =
0 CR
(Response with the input value of linearization data N-02.)
L N D
0 2
O =
0 CR
(Response with the output value of linearization data N-02.)
L N D
0 3
I =
0 CR
LF
14
LF
14
LF
14
LF
14
L N D
1 6
I =
0 CR LF
(Response with the input value of linearization data N-16.)
L N D
1 6
O =
0 CR LF
(Response with the output value of linearization data N-16.)
Y E S
CR LF
14
N CR LF
N CR LF
N CR LF
N CR LF
N CR LF
R CR LF
Note: The main unit returns to measurement operation upon an R command (or returns to the
response with the input value of linearization data N-01 if an N command is sent.)
L N D
0 1 CR LF
Note: Setting can be made from any of the data items 01 to 16.
Note: The reading of the main unit changes to LINE the moment it receives an LND XX
command.
- 1 0 0 0 CR LF
(Sets the input value of linearization data N-01 to -1000.)
N CR LF
- 9 0 0 CR LF
(Sets the output value of linearization data N-01 to -900.)
N CR LF
- 5 0 0 CR LF
(Sets the input value of linearization data N-02 to -500.)
N CR LF
- 6 0 0 CR LF
(Sets the output value of linearization data N-01 to -600.)
R CR LF
Note: If an R command is sent after setting required data, the main unit saves data provided
up to that point and then returns to measurement operation.
71/72
10
L N O
0 2 CR LF
(Sets the number of linearization correction data items to 02.)
L N O
1 6 CR LF
(Sets the number of linearization correction data items to 16.)
Char. Length
O CR LF
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
F S C
9 9 9 9 CR LF
(Response with a full-scale reading.)
F I N
9 9 9 9 CR LF
(Response with a full-scale input value.)
O F S
0 CR LF
(Response with an offset reading.)
O I N
0 CR LF
(Response with an offset input value.)
D L H I
9 9 9 9 CR LF
(Response with the HI value of the digital limiter.)
D L L O
- 9 9 9 9 CR LF
(Response with the LO value of the digital limiter.)
A O H I
9 9 9 9 CR LF
(Response with the HI reading of the analog output.)
A O L O
0 CR LF
(Response with the LO reading of the analog output.)
D E P
4 CR LF
(Response with the position of the decimal point.)
Y E S
CR LF
1
1
N CR LF
N CR LF
R CR LF
Note: The main unit returns to measurement operation upon an R command (or returns to the
response with the full-scale reading if an N command is sent.)
Scaling data setting
Note: This example shows a case
where FSC is set to 8000 and OFS
to 20.
Char. Length
14
0 CR LF
14
0 CR LF
14
0 CR LF
14
0 CR LF
14
0 CR LF
14
0 CR LF
14
0 CR LF
14
0 CR LF
14
CR LF
E r r o r
CR LF
(Response when a value outside the setpoint range is input.)
5
6
Function
Calibration data response
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23
C A L 2 CR LF
Note: The reading of the main unit changes to CAL2 the moment it receives a CAL2
command.
J CR LF
N CR LF
1 CR LF
N CR LF
(Executes zero calibration.)
Note: The main unit moves to the setting of a span reading if a J command is sent.)
5 0 0 0 CR LF
(Sets the span reading to 5000.)
N CR LF
(Executes span calibration.)
R CR LF
Note: If an R command is sent during setting, the main unit saves data provided up to that
moment and then returns to measurement operation (span reading is not saved, however).
Calibration data setting
(Equivalent calibration)
R CR LF
Note: The main unit returns to measurement operation upon an R command (or returns to the
zero point input value response if an N command is sent.)
Calibration data setting
(Calibration with actual load)
Char. Length
2 CR LF
0 0 0 4 CR LF
(Sets the zero input value to 0.004 mV/V.)
N CR LF
(Executes zero calibration.)
5 0 0 0 CR LF
(Sets the span reading to 5000.)
N CR LF
(Executes span calibration.)
V CR LF
O CR LF
0 CR LF
4
10
10
Y E S
CR LF
(Response when span calibration is executed correctly.)
Note: The main unit automatically returns to measurement operation when span
calibration is executed correctly.)
E r
r o r
X CR LF
S P A N
9 0 0 0 CR LF
(Response when an input signal that is outside the zero adjustment range or
contains a gain error is applied.)
Note: The main unit returns to the setting of a span reading.
Note: Check the applied input signal level or span reading and set the calibration
data again.
Y E S
CR LF
7
10
V CR LF
15
V CR LF
15
S P I N
2 . 0 0 0 m V / V CR LF
(Response when zero calibration is executed correctly and a transition is made to
span input value setting.)
E r
r o r
X CR LF
Z E R O
0 . 0 0 0 m V / V CR LF
(Response when a value outside the zero adjustment range is set.)
Note: Resend a value within the zero adjustment range (0.3 mV/V).
2
V CR LF
15
15
0 CR LF
10
0 CR LF
10
Y E S
CR LF
(Response when span calibration is executed correctly.)
Note: The main unit automatically returns to measurement operation when span
calibration is executed correctly.)
E r
r o r
X CR LF
S P I N
2 . 0 0 0 m V / V CR LF
(Response when a value that contains a gain error or is outside the range is set.)
Note: The main unit returns to span input value setting.
Y E S
CR LF
(Normal response)
N O
?
CR LF
(Response to, e.g., undefined commands.)
E r
r o r
CR LF
(Response to out-of-range data or data that do not meet setting conditions.)
E R R O R
A CR LF
(Response when there is any communication parameter failure.)
Note: Under normal conditions, this response is made only once.
Note: ERROR B to ERROR F are available as the response commands for
acknowledgment.
Note: If a response to the ERROR X command is received, execute the same
processing once again. If the panelmeter still does not recover, turn off the
panelmeter once and then turn it on again.
15
10
Z E R O CR LF
S P A N
9 0 0 0 CR LF
(Response when zero calibration is executed correctly and a transition is made to
span calibration.)
E r
r o r
X CR LF
(Response when an input signal outside the zero adjustment range is applied.)
Note: Apply an input signal within the zero adjustment range (0.3 mV/V and resend the N command.
S
15
Common response
S P I N
2 . 0 0 0 m V
(Response with a span input value.)
S P A N
9 0 0 0 CR LF
(Response with a span reading.)
Y E S
CR LF
Char. Length
4
1
1 5 0 2 CR LF
(Sets the span input value to 1.502 mV/V.)
N CR LF
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
Z E R O
0 . 0 0 0 m V / V CR LF
(Response with a zero input value.)
D A T A
L O S T
C O
(Response in case of condition data failure.)
Note: Set the condition data again.
D A T A
L O S T
C O
(Response in case of comparator data failure.)
Note: Set the comparator data again.
D A T A
L O S T
M E
(Response in case of scaling data failure.)
Note: Set the scaling data again.
D CR LF
15
7
15
5
5
6
7
14
M CR LF
13
T CR LF
13
Homepage http://www.asahikeiki.co.jp
72/72