1.X / 2.X to CT/i Upgrade Manual

Copyrighted 1998 by General Electric Company

Operating Documentation

GE MEDICAL SYSTEMS
REV 5

WARNING

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

THIS SERVICE MANUAL IS


AVAILABLE IN ENGLISH ONLY.

D IF A CUSTOMERS SERVICE PROVIDER REQUIRES A LANGUAGE


OTHER THAN ENGLISH, IT IS THE CUSTOMERS RESPONSIBILITY
TO PROVIDE TRANSLATION SERVICES.
D DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS
SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
D FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO
THE SERVICE PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC
SHOCK, MECHANICAL OR OTHER HAZARDS.

AVERTISSEMENT

CE MANUEL DE
MAINTENANCE NEST
DISPONIBLE QUEN ANGLAIS.

D SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS


UNE AUTRE LANGUE QUE LANGLAIS, CEST AU CLIENT QUIL
INCOMBE DE LE FAIRE TRADUIRE.
D NE PAS TENTER DINTERVENTION SUR LES QUIPEMENTS
TANT QUE LE MANUEL SERVICE NA PAS T CONSULT ET
COMPRIS.
D LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRANER
CHEZ LE TECHNICIEN, LOPRATEUR OU LE PATIENT DES
BLESSURES DUES DES DANGERS LECTRIQUES, MCANIQUES
OU AUTRES.

WARNUNG

DIESES KUNDENDIENSTHANDBUCH
EXISTIERT NUR IN ENGLISCHER
SPRACHE.

D FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE


BENTIGT, IST ES AUFGABE DES KUNDEN FR EINE
ENTSPRECHENDE BERSETZUNG ZU SORGEN.
D VERSUCHEN SIE NICHT, DAS GERT ZU REPARIEREN, BEVOR
DIESES KUNDENDIENSTHANDBUCH NICHT ZU RATE GEZOGEN
UND VERSTANDEN WURDE.
D WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU
VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES
BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE
SCHLGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.

AVISO

ESTE MANUAL DE SERVICIO


SLO EXISTE EN INGLS.

D SI ALGN PROVEEDOR DE SERVICIOS AJENO A GEMS


SOLICITA UN IDIOMA QUE NO SEA EL INGLS, ES
RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE
TRADUCCIN.
D NO SE DEBER DAR SERVICIO TCNICO AL EQUIPO, SIN
HABER CONSULTADO Y COMPRENDIDO ESTE MANUAL DE
SERVICIO.
D LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR
LUGAR A QUE EL PROVEEDOR DE SERVICIOS, EL OPERADOR O EL
PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS
ELCTRICAS, MECNICAS O DE OTRA NATURALEZA.

ATENO

ESTE MANUAL DE ASSISTNCIA


TCNICA S SE ENCONTRA
DISPONVEL EM INGLS.

D SE QUALQUER OUTRO SERVIO DE ASSISTNCIA TCNICA,


QUE NO A GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA,
DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIOS
DE TRADUO.
D NO TENTE REPARAR O EQUIPAMENTO SEM TER
CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSISTNCIA
TCNICA.
D O NO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A
SEGURANA DO TCNICO, OPERADOR OU PACIENTE DEVIDO A
CHOQUES ELTRICOS, MECNICOS OU OUTROS.

AVVERTENZA

IL PRESENTE MANUALE DI
MANUTENZIONE DISPONIBILE
SOLTANTO IN INGLESE.

D SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS


RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE
TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE.
D SI PROCEDA ALLA MANUTENZIONE DELLAPPARECCHIATURA
SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE
COMPRESO IL CONTENUTO.
D NON TENERE CONTO DELLA PRESENTE AVVERTENZA
POTREBBE FAR COMPIERE OPERAZIONI DA CUI DERIVINO
LESIONI ALLADDETTO ALLA MANUTENZIONE,
ALLUTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

IMPORTANT! . . . X-RAY PROTECTION


X-ray equipment if not properly used may cause
injury. Accordingly, the instructions herein
contained should be thoroughly read and
understood by everyone who will use the equipment
before you attempt to place this equipment in
operation. The General Electric Company, Medical
Systems Group, will be glad to assist and cooperate
in placing this equipment in use.
Although this apparatus incorporates a high degree of
protection against x-radiation other than the useful beam,
no practical design of equipment can provide complete
protection. Nor can any practical design compel the
operator to take adequate precautions to prevent the
possibility of any persons carelessly exposing themselves
or others to radiation.

the recommendations of the National Council on


Radiation Protection and Measurements as published in
NCRP Reports available from NCRP Publications, 7910
Woodmont Avenue, Room 1016, Bethesda, Maryland
20814, and of the International Commission on Radiation
Protection, and take adequate steps to protect against
injury.
The equipment is sold with the understanding that the
General Electric Company, Medical Systems Group, its
agents, and representatives have no responsibility for
injury or damage which may result from improper use of
the equipment.
Various protective material and devices are available. It is
urged that such materials or devices be used.

It is important that everyone having anything to do with


x-radiation be properly trained and fully acquainted with

CERTIFIED ELECTRICAL CONTRACTOR


STATEMENT
All electrical installations that are preliminary to positioning of
the equipment at the site prepared for the equipment shall be
performed by licensed electrical contractors. In addition,
electrical feeds into the Power Distribution Unit shall be
performed by licensed electrical contractors. Other
connections between pieces of electrical equipment,
calibrations, and testing shall be performed by qualified GE
Medical personnel. The products involved (and the
accompanying electrical installations) are highly
sophisticated, and special engineering competence is

required. In performing all electrical work on these products,


GE will use its own specially trained field engineers. All of
GEs electrical work on these products will comply with the
requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified
personnel (i.e., GEs field engineers, personnel of third-party
service companies with equivalent training, or licensed
electricians) to perform electrical servicing on the
equipment.

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of
delivery. If damage is apparent, have notation damage in
shipment written on all copies of the freight or express
bill before delivery is accepted or signed for by a
General Electric representative or a hospital receiving
agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in
any event, within 14 days after receipt, and the contents
and containers held for inspection by the carrier. A
transportation company will not pay a claim for damage if
an inspection is not requested within this 14 day period.

ii

Call Traffic and Transportation, Milwaukee, WI (414)


8273449 / 8*2853449 immediately after damage is
found. At this time be ready to supply name of carrier,
delivery date, consignee name, freight or express bill
number, item damaged and extent of damage.
Complete instructions regarding claim procedure are
found in Section S of the Policy & Procedure Bulletins.

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

OMISSIONS & ERRORS


Customers, please contact your GE Sales or Service
represenatives.

GE personnel, please use the GEMS CQA Process to


report all omissions, errors, and defects in this
documentation.

iii

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

NOTES

iv

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Chapter 1

CT/i Upgrade Manual

Terminology
Internet Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Boot Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating System Level (Irix) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Applications Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Upgrade20
Upgrade72
Upgrade72
Upgrade72

Upgrade1

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Table Of Contents
Section

Description

Section 1
Work Flow Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page

Upgrade11

Section 2
Before you Start the Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . Upgrade13
2.1
2.2
2.3
2.4
2.5
2.6
2.7
2.8
2.9
2.10
2.11

Collect RP system data . . . . . . . . . . . . . . . . . . . . . . . . . . . .


Write Down System Internet Addresses . . . . . . . . . . . . . .
Record ISO Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Record Collimator Data . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Record N Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Write Down Tube Usage Infromation . . . . . . . . . . . . . . . . .
Write Down Window / Level Hardkey Values . . . . . . . . . .
Record Reconfig Information . . . . . . . . . . . . . . . . . . . . . . .
IP Addresses in CT/i . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
IP Addresses in RP System . . . . . . . . . . . . . . . . . . . . . . . .
Check CT/i Modem is Available . . . . . . . . . . . . . . . . . . . . .

Upgrade13
Upgrade13
Upgrade14
Upgrade14
Upgrade15
Upgrade15
Upgrade16
Upgrade17
Upgrade19
Upgrade20
Upgrade20

Section 3
Equipment Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Upgrade21
3.1
3.2
3.3
3.4
3.4.1
3.4.2
3.4.3
3.4.4

REMOVE SCU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
REMOVE GENESIS CONSOLE . . . . . . . . . . . . . . . . . . . .
Remove CT/i Console from Skid . . . . . . . . . . . . . . . . . . . .
Open Console and Attach Keyboard Shelf . . . . . . . . . . . .
Open Console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Keyboard Table Top Bracket Attachment and Console
Placement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Keyboard Table Top Attachment . . . . . . . . . . . . . . . . . . . . .
Seismic Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Upgrade21
Upgrade22
Upgrade23
Upgrade24
Upgrade24
Upgrade25
Upgrade26
Upgrade28

Section 4
System Power and Ground Guidelines . . . . . . . . . . . . . . . . . . Upgrade29
4.1
4.2
4.3

Upgrade2

Electrical Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cable Color Identifiers . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Upgrade29
Upgrade30
Upgrade30

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Section

Description

Page

Section 5
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Upgrade31
5.1
5.2

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
How to Use this Section . . . . . . . . . . . . . . . . . . . . . . . . . . . .
INDIGO COMPUTER CONSOLE

Upgrade31
Upgrade31

5.3
Power Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.4
Incoming Power Detail w/Global or Compact PDU . . . . .
5.5
Incoming Power Detail w/GE PDU . . . . . . . . . . . . . . . . . . .
5.6
Computer Carrier Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.7
Specialix Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.8
Back Bulkhead Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.9
Display Monitor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.10
Rx Monitor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
OCTANE COMPUTER CONSOLE

Upgrade32
Upgrade33
Upgrade34
Upgrade35
Upgrade36
Upgrade37
Upgrade38
Upgrade39

5.11
5.12
5.13
5.14
5.15
5.16
5.17
5.18
5.19
5.20
5.21
5.22
5.23
5.24
5.25
5.26

Octane Rear Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


Remove Shipping Strap and Bracket . . . . . . . . . . . . . . . . .
Modem Installation 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Modem Installation 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DASM Installation 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DASM Installtion 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DASM Installation 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DASM Isntallation 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Incoming Power Detail w/Global or Compact PDU . . . . .
Incoming Power Detail w/GE PDU . . . . . . . . . . . . . . . . . . .
Computer Carrier Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Back Bulkhead Detail 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Back Bulkhead Detail 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Display Monitor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Rx Monitor Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Upgrade40
Upgrade41
Upgrade42
Upgrade43
Upgrade44
Upgrade45
Upgrade46
Upgrade47
Upgrade48
Upgrade49
Upgrade50
Upgrade51
Upgrade52
Upgrade53
Upgrade54
Upgrade55

Keyboard / Trackball Detail . . . . . . . . . . . . . . . . . . . . . . . . .


Gantry/Console Cable Connections . . . . . . . . . . . . . . . . . .
EStop/Intercom/Scan Control . . . . . . . . . . . . . . . . . . . . . .
Autovoice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SCOM/FEP Fiber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
System Ground Connections . . . . . . . . . . . . . . . . . . . . . . .

Upgrade56
Upgrade57
Upgrade57
Upgrade58
Upgrade61
Upgrade63

5.27
5.28
5.28.1
5.28.3
5.28.5
5.29

Upgrade3

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Section

Description

Page

Section 6
COMPUTER INTEGRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Upgrade72
DEFINITIONS
6.1
6.2
6.3
6.4
6.5
6.6
6.7
6.9
6.8
6.7
6.11
6.14
6.16
6.18
6.19

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tools Needed: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Requirements/Assumptions: . . . . . . . . . . . . . . . . . . . . . . . .
Service Key Install . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power Up RCU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Keyboard Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Default Passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Touch Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DASM Checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ReConfig . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Install Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Verify Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
KVTest and ZAlign Baselines . . . . . . . . . . . . . . . . . . . . . .
Monitor Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Upgrade72
Upgrade72
Upgrade72
Upgrade73
Upgrade73
Upgrade74
Upgrade74
Upgrade75
Upgrade76
Upgrade75
Upgrade75
Upgrade77
Upgrade79
Upgrade81
Upgrade83
Upgrade84

Section 7
Table / Gantry Characterization . . . . . . . . . . . . . . . . . . . . . . . . . Upgrade88
7.1
7.2
7.3
7.4
7.5

Collimator Characterization . . . . . . . . . . . . . . . . . . . . . . . . .
Tilt Characterization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Elevation Characterization . . . . . . . . . . . . . . . . . . . . . . . . . .
Cradle Characterization . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Interference Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Upgrade88
Upgrade91
Upgrade92
Upgrade94
Upgrade96

Section 8
MX200 Tube Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Upgrade98
Section 9
Tube Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Upgrade99
9.1
9.2

Upgrade4

New Tube Seed Shift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


Auto mA Cal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Upgrade99
Upgrade101

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Section

Description

Page

Section 10
System Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Upgrade102
10.1
10.2
10.3
10.4
10.5

QCal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Crosstalk Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Alpha Vector Cal Generation . . . . . . . . . . . . . . . . . . . . . . . .
Hot ISO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Air/Phantom Calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . .

Upgrade103
Upgrade105
Upgrade106
Upgrade107
Upgrade108

Section 11
Packing Material For Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Upgrade110
11.1
11.2
11.3
11.3.1
11.3.2

Console Packing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SCU Packing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Label Return Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Return Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Return Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Upgrade110
Upgrade110
Upgrade111
Upgrade112
Upgrade114

Upgrade5

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

List Of Illustrations
Illustration

Illustration 11
Upgrade Flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Illustration 12
CT/i Network Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Illustration 13
RP Network Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Illustration 14
Console Stabilizer Pad Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Illustration 15
Console Front Cover Latches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Illustration 16
Console Top and Keyboard Shelf Mounting . . . . . . . . . . . . . . . . . . . . . .
Illustration 18
CT/i Console Power Wiring Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Illustration 111
CT/i Console Computer Carrier Wiring Detail . . . . . . . . . . . . . . . . . . . .
Illustration 112
CT/i Console Specialx Wiring Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Illustration 113
CT/i Console Back Bulkhead Wiring Detail . . . . . . . . . . . . . . . . . . . . . .
Illustration 132
CT/i Console Keyboard / Trackball Wiring Detail . . . . . . . . . . . . . . . . . .
Illustration 114
CT/i Console Display Monitor Wiring Detail . . . . . . . . . . . . . . . . . . . . . .
Illustration 115
CT/i Console Rx Monitor Wiring Detail . . . . . . . . . . . . . . . . . . . . . . . . . .
Illustration 133
Gantry A1A5 Panel Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Illustration 134
CT/i Gantry Interconnect Cable Connections . . . . . . . . . . . . . . . . . . . . .
Illustration 136
System Ground Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Upgrade6

Page

Upgrade11
Upgrade19
Upgrade20
Upgrade23
Upgrade24
Upgrade25
Upgrade32
Upgrade35
Upgrade36
Upgrade37
Upgrade56
Upgrade38
Upgrade39
Upgrade60
Upgrade60
Upgrade63

GE MEDICAL SYSTEMS
REV 5

Section

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Description

Illustration 137
Non EMC HiSpeed CT/i Interconnect Diagram . . . . . . . . . . . . . . . . . . .
Illustration 138
EMC HiSpeed CT/i Interconnect Diagram . . . . . . . . . . . . . . . . . . . . . . .
Illustration A1
Elevation Distance Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Illustration A2
Cradle Characterization Pin Locations . . . . . . . . . . . . . . . . . . . . . . . . . .

Page

Upgrade66
Upgrade67
Upgrade93
Upgrade95

Upgrade7

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

List Of Tables
Table

Table 11 System Interconnect Drawing Matrix . . . . . . . . . . . . . . . . . . . . .


Table 12 CT/i Ground Wire Sizes and Connections . . . . . . . . . . . . . . . . .

Upgrade8

Page

Upgrade31
Upgrade64

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

REVISION HISTORY
REV

DATE

REASON FOR CHANGE

A . . . . . . . . . . . . . . . . . . 24 July, 1996 . . . . . . . Initial Draft of document.


0 . . . . . . . . . . . . . . . . . . 18 Sept, 1996 . . . . . . . Initial Release.
1 . . . . . . . . . . . . . . . . . . 18 Oct, 1996 . . . . . . . . Pilot Feedback.
2 . . . . . . . . . . . . . . . . . . 02 June, 1997 . . . . . . 3.6 Software Release
3 . . . . . . . . . . . . . . . . . . 27 Oct 1997 . . . . . . . . 4.0 Software Release
4 . . . . . . . . . . . . . . . . . . 18 Mar 1998 . . . . . . . . Octane Clinical Release
5 . . . . . . . . . . . . . . . . . . 31 Dec 1998 . . . . . . . . Changes for 5.3 M4 Octane, 4.1 M4 Indigo

Upgrade9

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

NOTES

Upgrade10

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

SECTION 1
WORK FLOW CHART
Illustration 11
UPGRADE FLOWCHART

Get RP System Info:


ISO, IP Addresses,
Tube Slice, etc...

SECTION 2.0

Disconnect &
Remove GENESIS
Console

Disconnect SCU
cables and remove
SCU

Remove CT/i
Console from skid

SECTION 3.2

SECTION 3.1

SECTION 3.3

Unplug GENESIS
Console Power Cord
at PDU. Tape
Console end or
remove cable.

Move SCU cables to


CT/i Console location

Move Console
to/near final position

Disconnect SCU
ground connection at
Table Junction
Ground Bus. Tape
both ends or remove.

Install Console
Stabilizer Leg
Leveling Pads before
lowering console to
floor.

Lower console to
floor and remove
shipping dollies

Remove console
front cover and cover
on front of scan
chassis
Attach keyboard shelf
to console sides

Attach keyboard shelf


to shelf supports

Place keyboard
assembly on the
keyboard shelf

Route and terminate


monitor video &
power cables in
console base.
Illustration 18 &
Illustration 111
Position both
monitors on console
top

Terminate cables at
CT/i console

SECTION 5

SECTION 3.3

Power On

SECTION 6
Computer Integration

SECTION 7
Table / Gantry
Characterization

SECTION 8
Option If performing a MX200
tube upgrade do so at this point

Yes
Does this upgrade
include a MX200
tube upgrade?

SECTION 9

No

Verify Images

Calibrate Tube

SECTION 10
Calibrate System

Upgrade11

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

NOTES

Upgrade12

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

SECTION 2
BEFORE YOU START THE UPGRADE
This section describes things that should be completed before
disconnecting or removing any of the HSA RP 1x/2.x system pieces.

2.1

Save RP System State Information


This section describes the procedure to share system specific files from the RP CT
Scanner to a CT/i CT scanner.
In this document output from the system is shown in this font/style: (ex.

login:),,

site specific output is shown in this font/style: (ex. <suiteID>),


and commands to be typed by the user are shown in this font/style: (ex.
/usr/g).

2.2

cd

Write Down System Internet Addresses


(1) Login as root on the RP OC.
<suitId>_OC0 Login:
Password:

root<enter>

<Root Password>

<suite_id>_OC0

1: ypcat hosts|grep <suite_id>

This represents an example output of the primary IP addresses of interest.


192.20.20.2

<suiteID>_SBC0

192.20.20.1

<suiteID>_OC0 suti_oc0

<suiteID>_sbc0

The following IP address is associated with the optional second ethernet adapter
board that provides access to networks outside of the CT suite.
3.7.52.101

<suiteID>_OC0gate suti_oc0gate

Record number <suiteID>_SBC0:

________________________

Record number <suiteID>_OC0:

________________________

Record number <suiteID>_OC0gate: ________________________

Upgrade13

GE MEDICAL SYSTEMS
REV 5

2.3

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Record iso Values


From a Unix Shell
Type:

su

Type:

grep

<enter>

Log in as root.

isoSmallSpot

/usr/g/g_config/ScanHardware.cfg

Record number:
Type:

grep

isoLargeSpot

________________________
/usr/g/g_config/ScanHardware.cfg

Record number:

2.4

________________________

Record Collimator Data


Collimator Data
Closed
1mm
3mm
5mm
7mm
10mm
CRC

Upgrade14

<enter>

<enter>

GE MEDICAL SYSTEMS
REV 5

2.5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Record N numbers
Small

Large

80 kV
100 kV
120 kV
140 kV

2.6

Write Down Tube Usage Information


The RP system must be up and running for this procedure.
(1) Open a Unix Shell
<suite_id>_OC0

1: tubeUsage

Record current tube slice count:


Patient:

________________________

NonPatient:

________________________

Transfer the current tube slice count into your site records. The slice count for CT/i
will start with zero after the software is reconfigured.

Upgrade15

GE MEDICAL SYSTEMS
REV 5

2.7

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Write Down Window / Level Hardkey Information


Window / Level Hard Key Values
Anatomy
Head
Mediastinum
Spine
Abdomen
Lung
Vertebrae

Upgrade16

Window

Level

GE MEDICAL SYSTEMS
REV 5

2.8

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Record Recofig Information


(1) Shutdown Applications.
(2) Open a Unix Shell.
(3) As genesis on the RP OC.
<suitId>_OC0

Login: sd i<enter>

(4) Wait for Applications to shutdown.


(5) Login as root on the RP OC.
<suitId>_OC0

Login: root<enter>

Password: <Root Password>


<suite_id>_OC0

2: reconfig

(6) Record the following information from the reconfig screens.

Suite ID
Hostname
Hosptial Name
Service ID
Starting Patient Exam
Starting Optical Disk
Timezone
Doctors Title
Tube Housing PN
Tube Housing SN
Tube Insert PN
Tube Insert SN

Upgrade17

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Camera Type
DAS Part Number
DAS Type
ACRNEMA Address
Suite ID of PAN Host
Hostname of PAN Host
Subnet Netmask
Gateway Netmask
(7) Go back to the first screen and exit without changes.

Upgrade18

GE MEDICAL SYSTEMS
REV 5

2.9

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

IP Addresses in CT/i

Illustration 12
CT/I NETWORK DEVICES

InSite
address

SBC0
address

SBC Control LAN


address

Modem

DLAN
Bit3
GIO64

du0

vd0
External
Network

SGI IP22

Bit3
VME
vd0

To Gantry STC

ei0
SBC

VME
Bus

ec0

du0 MODEM DEVICE MNEMONIC

gateway
address

OBC0
address

ec0 IP22 ETHERNET TRANSCEIVER DEVICE MNEMONIC


vd0 BIT3 DEVICE MNEMONIC
ei0 SBC ETHERNET TRANSCEIVER DEVICE MNEMONIC

Upgrade19

GE MEDICAL SYSTEMS
REV 5

2.10

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

IP Addresses in RP System

Illustration 13
RP NETWORK DEVICES

SBC Control LAN


address

OC0
address

Modem

ei0

ei0

Motorola
Hawk
Ethernet

External
Network

hk0

SBC
SBC0
address

DLAN

To Gantry STC

hk0
CLA

VME
Bus

ei0 MOTOROLA CPU ETHERNET TRANSCEIVER DEVICE MNEMONIC


hk0 HAWK ETHERNET TRANSCEIVER DEVICE MNEMONIC

gateway
address

2.11

ei0 SBC ETHERNET TRANSCEIVER DEVICE MNEMONIC


hk0 CLA (HAWK) ETHERNET TRANSCEIVER DEVICE MNEMONIC

Check CT/i Modem is Available


Insure that a unique Internet Address is available for the CT/i InSite Modem.
HiSpeed CT/i has made a change to the way the OnLine Centers communicate with
the scanner. In the past the protocols used to communicate with the scanner where
based upon standard/special serial protocols and special security access controls.
HiSpeed CT/i makes use of a Point to Point Protocol (PPP) to communicate with the
OnLine Centers. The PPP allows standard TCP/IP connectivity tools to be used as if
the modem connection where part of a TCP/IP based network. In addition, multiple
levels of access security are used to insure that unauthorized users cannot get into
the system.
For PPP to work correctly for the InSite connection, a unique IP address must be
assigned to either the CT/i modem or to the SGI computer gateway.

Upgrade20

Note 1

If the Host Computer (GENESIS) had a second ethernet board to connect as a


gateway to the facilities network and had unique IP addresses assigned by the
facilities Network Administrator, then that number may be used for the SGI Host
Computer IP Gateway Address and may also be used for the Modem IP Address.

Note 2

If the Host Computer (GENESIS) does not have a Unique IP address assigned by
the facilities Network Administrator, then a unique number for the InSite Modem can
be obtained from the OnLine Center during the InSite Connection CheckOut.

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

SECTION 3
EQUIPMENT REMOVAL
This section tells you how to remove the Console and the SCU Cabinet.

3.1

REMOVE SCU
(1) Disconnect these cables from the SCU and protect from damage. These cables
will be moved and reconnected to the HiSpeed CT/i Computer/Console:
>

Note 3

Note 4

Disconnect SCU Power Cable (2118139 or 46278722G1) at SCU.

In case the SCU power cord was cut to length and reterminated for the RP system,
a new Power Cord 2118139 is included in the Upgrade kit of parts.
>

Disconnect Scan Abort Cable (2118148 or 46297565G1) at SCU.

>

Disconnect Control LAN Cable (2118150 or 46296454G2) at SCU

>

Disconnect DAS Data Fiber (21178482) at FEP Board in SCU.

>

Disconnect Network cable connection if present

>

Disconnect Camera Video and Pixel CLock Cables

The HiSpeed CT/i System requires a Camera Host Control Interface for filming
functions.
>

Disconnect Camera Control Cable

(2) Disconnect remaining SCU cables and discard.


(3) Disconnect the SCU Ground wire at the Table/Gantry Junction Ground Bar. Tape
both ends or remove ground wire from conduit or duct.
(4) Move SCU cabinet out of the work area.

Upgrade21

GE MEDICAL SYSTEMS
REV 5

3.2

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

REMOVE GENESIS CONSOLE


(1) Disconnect the cables from the GENESIS Operators Console.
DANGER

FAILURE TO DISCONNECT THE GENESIS CONSOLE POWER CORD AT THE


PDU COULD RESULT IN A SHOCK WHEN THE SYSTEM IS TURNED BACK ON.
(2) Disconnect the GENESIS Console power cable from the PDU.
(3) Tape up the Console end of the old power cord or remove the cord from the
conduit or duct.
(4) Remove the trim covers over the Console Stabilizer Leg hardware on one side of
the console and remove the hardware.

DANGER

REMOVAL OF THE HARDWARE THAT ATTACHES THE GENESIS CONSOLE TO


THE FLOOR OR RISERS IN THE NEXT STEP MAKES THE GENESIS CONSOLE
LESS STABLE AND EASIER TO KNOCK OVER.
WE RECOMMENDED THAT YOU ATTACH THE CONSOLE SHIPPING DOLLIES
TO THE CONSOLE BEFORE YOU REMOVE ANY OF THE CONSOLE
ANCHORING HARDWARE.
(5) Remove GENESIS Console Floor anchors if the console is attached to the floor
or remove anchors that attach the console to the risers.
(6) Move the GENESIS Console to a location that will allow access to the hardware
that attaches the stabilizer legs.
(7) Attach the GENESIS Console Shipping Dolly and Fischpaper provided in the
Upgrade Kit to the side of the console that the stabilizer leg was just removed
from.
(8) Remove the trim covers over the Console Stabilizer Leg hardware on the other
side of the console and remove the hardware.
(9) Attach the GENESIS Console Shipping Dolly and Fischpaper provided in the
Upgrade Kit to the side of the console that the stabilizer leg was just removed
from.
(10)Move the GENESIS Console out of the work area.

Upgrade22

GE MEDICAL SYSTEMS
REV 5

3.3

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Remove CT/i Console from Skid


(1) Raise the CT/i Console off of the shipping skid using the adjustable dolly casters.

Raise the console enough to clear the wood blocking attached to the shipping
skid.

Roll the console away from the shipping skid.


(2) Move the console to very near its final position. Do not lower the console to the
floor yet!

Note 5

Even though the CT/i Console has internal casters, use the Shipping Dollies to move
the console any significant distance.
The larger diameter casters on the shipping dollies make it easier to move over door
and elevator thresholds and cause less shock to the console.
(3) Install the Console Stabilizer Leg Leveling Pads before lowering the Console to
the floor. See Illustration 14.

Illustration 14
CONSOLE STABILIZER PAD INSTALLATION

Screw the pads all the way into the Stabilizer Legs.
After the console has been placed in its final position lower the pads so that the
console does not rock forward.
(4) Lower the Console to the floor and remove the shipping dollies.

Upgrade23

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

3.4

Open and Assemble Console

3.4.1

Open Console
(1) Front Cover Removal
>

Unlatch the two fasteners at the top corners of the console front cover, rotate
counterclockwise to the 5 oclock position. See Illustration 15.

Illustration 15
CONSOLE FRONT COVER LATCHES

Rotate
CounterClockwise to Unlatch

Unlatched

Latched

(2) Once the front cover is unlatched slide the top edge forward 1/2 inch then lift the
cover up and away from the console.

Upgrade24

GE MEDICAL SYSTEMS
REV 5

3.4.2

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Adjust Table Top Height and Console Placement


The Console should have arrived with the Monitor top mounted, starting with the
fourth bolt from the bottom. The Keyboard Table Top should be attached starting
with the third hole from the bottom. This arrangement will provide for optimum
operator comfort.

Note 6

Your site may require the console to be adjusted from this position.
The important factor in setting the height is operator visibility of the patient being
scanned. If you need to adjust from this starting position keep the relationship of
mounting the Keyboard Table Top one hole lower then the Console Table Top, see
Illustration 16 for clarification.

Illustration 16
CONSOLE TOP AND KEYBOARD SHELF MOUNTING

Wall
Minimum

Note 7

START WITH THE THIRD HOLE FROM THE BOTTOM.


SPEC 8 +/ 1 FOOT POUND
START WITH THE FOURTH HOLE FROM THE BOTTOM.
THIS WILL LEAVE 2 1/2 HOLES SHOWING.
SPEC 8 +/ 1 FOOT POUND

When placing the console in its final position keep a 2 minimum distance from the
wall to maintain console airflow.

Upgrade25

GE MEDICAL SYSTEMS
REV 5

3.4.3

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Attach Keyboard Table Top Select Illustration that matches your system.

Illustration 17
ATTACH KEYBOARD TO TABLE TOP

Use spring loaded handles to attach


keyboard top to support arms.
You may want to delay installing the top
until the Octane computer is wired in Section 2.

Upgrade26

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

NOTES

Upgrade27

GE MEDICAL SYSTEMS
REV 5

3.4.4

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Seismic Mounting
See PreInstallation Manual 2152916100.

Upgrade28

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

SECTION 4
SYSTEM POWER AND GROUND GUIDELINES
This section provides reference information used to insure a good
power installation for the system.
At this point, you MUST physically inspect the System Power and Ground
Connections per the following:

The system should have a dedicated ground. The ground wire should originate at
the supplying transformer and shall have no secondary ground connected to it
until the ground bar in our PDU.

The system ground wire should be copper and sized equal to or larger than the 3
phase supply wires, with a minimum size of AWG 1/0. The ground wire should
take the same physical path (conduit) as the supply wires.

The ground wire impedance from the system disconnect, including the ground
rod, shall not have an impedance greater than 2 ohms to earth.

The neutral from the 480 volt system should never be brought into the PDU.
3 phase wires, power cables, neutrals, and grounds should be routed in the
same conduit or raceway duct. These cables must not be routed in the same
conduit or raceway duct as system control cables.
CAUTION

Check all the ground connections on the system for tightness and proper
termination. Check ground connection at the supply transformer. Check
ground from transformer to its earth connection.

Note 8

If you suspect that a system performance problem is related to, or caused by, a
power/grounding issue, call:

GEMS Americas
Power/Grounding Hotline
18003217937
Press 1 for OLC
Press 7 for Power and Ground

GEMS Europe
Tel: (33 1) 39 20 00 07

GEMS Asia
Taiwan:
Australia:
SMS:
YMS:

(886 2) 713 4150


(61 2) 318 9100
(82 2) 553 9737
(81 426) 56 0033

(ask for technical support)

4.1

Electrical Inspection
Arrange for electrical code inspection, if required.

Upgrade29

GE MEDICAL SYSTEMS
REV 5

4.2

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Identification

CT2 Gantry
CT1 Patient Table
PM Power Distribution Unit
OC Operators Console
WL XRay on Warning Light

All system cables are identified using the system component designators (see
above). Each end of a system cable has a label and may have a color near the
connector (see below). This indicates the component and the jack identifier of the
component.

4.3

Cable Color Identifiers


The ends of the cables may be marked with a color, either red, orange, blue, green
or brown, to help with the cable installation. The chart indicates which color is
associated with which sub component.

Upgrade30

Subcomponent

Color

Gantry

Blue

Table

Yellow

PDU

Red

Console

Green

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

SECTION 5
WIRING
5.1

Introduction
This upgrade manual supports 2 Console and 4 PDU types:
CONSOLE:

INDIGO
OCTANE

PDU:

5.2

21364002
2169455

46296220G1
46327106G1
2113764 EMC/NONEMC
2133533 EMC/NONEMC

How to Use this Section


This section contains detailed drawings of the individual cable connections and system
interconnect drawings.
Use Sections 5.3 through 5.10 for a Console with an Indigo Computer.
Use Sections 5.19 through 5.26 for a Console with an Octane Computer.
To determine which system interconnect drawing to use consult Table 11, SYSTEM
INTERCONNECT DRAWING MATRIX

Table 11
SYSTEM INTERCONNECT DRAWING MATRIX

PDU Type

Console Type
Indigo PN21384002

Octane PN2169455

46327106G1

Illustration 139 page Upgrade68

Illustration 142 page Upgrade71

46296220G1

Illustration 139 page Upgrade68

Illustration 142 page Upgrade71

2113764 EMC

Illustration 138 page Upgrade67

Illustration 141 page Upgrade70

2133533 EMC

Illustration 138 page Upgrade67

Illustration 141 page Upgrade70

2113764 NON EMC

Illustration 137 page Upgrade66

Illustration 140 page Upgrade69

2133533 NON EMC

Illustration 137 page Upgrade66

Illustration 140 page Upgrade69

Upgrade31

GE MEDICAL SYSTEMS
REV 5

5.3

Power Detail

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Console PN
21384002

Illustration 18 CT/I CONSOLE POWER WIRING DETAIL

Front View
P# DEVICE

P#

DEVICE

P1
P2
P3
P4
P5
P6
P7
P8

P9
P10
P11
P12
P13
P14
P15
P16

VIEB 
DASM 
IMAGE PRINTER 
VCR 
REMOTE MONITOR 
SPARE
SPARE
COMPUTER CARRIER 

RX MONITOR
EMC CHASSIS 
KIEB 
MODEM 
VIDEO SPLITTER 
PRINTER 
DISPLAY MONITOR
SPARE

P1

P3

P5

P7

P2

P4

P6

P8

P9

P11

P13

P15

P10

P12

P14

P16

Notes:
 Shown for reference, cable already
installed.
 Option: Shown for reference, cable
may or may not be installed on this
system.

Upgrade32

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

5.4

Console PN
21384002

Incoming Power Detail For a System with Global or Compact PDU


If you have a PDU PN 46329106G1 or 46296220G1 see Section 5.5.
Illustration 19 CT/I CONSOLE INCOMING POWER WIRING DETAIL

Front View

P1

P3

P5

P7

P2

P4

P6

P8

P9

P11

P13

P15

P10

P12

P14

P16

12

Console Power 2118139


From : PDU A4J5

Indicates Cable Run Number

Upgrade33

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

5.5

Console PN
21384002

Incoming Power Detail


For a System with a PDU PN 46329106G1 or 46296220G1
If you have a Global or Compact PDU see Section 5.4
Illustration 110 CT/I CONSOLE INCOMING POWER WIRING DETAIL

Front View

You will need this Section 5.5 if your RP Console was


powered from a wall outlet and not a power cord coming
from the PDU. See Illustration 139 cable run 12 for more
information. In this situation two options exist for cable run
12.
(1) You can reuse the power cord ( 46278722G1 or
46327187G1 ) which connected the SCU to the wall power
for cable run 12.
or
(2) If your site requires a more pliable power cord. You can
rework the power cord 2118139 that came with the kit. To
perform the rework follow the steps below.
(a) Cut cable 2118139 to the desired length measuring from
the female connector end.
(b) Remove the male plug from the power cord which was
used on the RP Console that was just removed.
P1

P3

P5

P2

P4

P6

P9

P11

P13

P10

P12

P14

(c) Install the male plug on cable 2118139.


see detail (a) for plug wiring.
detail a

Blue
Brown Z

X blank

12
Y
Console Power 2118139 or 46278722G1
or 46327187G1
From : RP Console Wall Outlet

Indicates Cable Run Number

Upgrade34

Blac
k

Center
Yellow/Green

GE MEDICAL SYSTEMS
REV 5

5.6

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Computer Carrier Detail

Console PN
21384002

Illustration 111 CT/I CONSOLE COMPUTER CARRIER WIRING DETAIL

Front View

Rx Video
2142221
From : Rx Monitor

Display Video
2142221 Color or
2143804 B/W
From : Display Monitor

J39

J40

J46
J45

J44

110 VAC 1

J35

Audio In 1

J37

J47

Insite Modem 1
Security Key 1

Audio Out 1

SCSI 1 1

BIT3 1
VCR Video 2
Specialix 1

Camera SCSI 0 1

J33

J43

Inter Suite 1
Line Printer 1
Inter Suite 1

J41
J42

5
Mouse:
Route through
console top using
hole provided.
Keyboard 1

Indicates Cable Run Number


Notes:
1
Shown for reference, cable already
installed.
2
Option: Shown for reference, cable
may or may not be installed on this
system.
Upgrade35

GE MEDICAL SYSTEMS
REV 5

5.7

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Specialix Detail

Console PN
21384002

Illustration 112 CT/I CONSOLE SPECIALIX WIRING DETAIL

Front View

Touch 2139581 
From : Rx Monitor

Indicates Cable Run Number


2










  

Indicates Cable Run Number

Upgrade36




Notes:
 Shown for reference, cable already
installed.
 Option: Shown for reference, cable
may or may not be installed on this
system.

GE MEDICAL SYSTEMS
REV 5

5.8

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Back Bulkhead Detail

Console PN
21384002

Illustration 113 CT/I CONSOLE BACK BULKHEAD WIRING DETAIL

Rear View
Indicates Cable Run Number

Keyboard 2139579
From : Keyboard

J19

EStop

2139578 *

Intercom
J20

Scan Control

10

DAS Data 21178482


From : Gantry A1A5J5
or SCOM

Existing cables disconnected


from RP Console

A10J8 EStop

A10J9 Intercom

A10J4 Scan Control

*Alternate method for cable run 8 is


to run a completely new cable PN
2139577 from J20 to A1A5J1

LAN 2118150 or 46296454G2


From : Gantry A1A5J3

J17

11
J18

Scan Abort 2118148


From : Gantry A1A5J4
or
Scan Abort 46297565G1
From : PDU

Upgrade37

GE MEDICAL SYSTEMS
REV 5

5.9

Display Monitor Detail

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Console PN
21384002

Illustration 114 CT/I CONSOLE DISPLAY MONITOR WIRING DETAIL

Rear View
Display Monitor

Rx Monitor

DISPLAY MONITOR
COLOR OPTION

RGB

110VAC
From: P7

2142221 Video
From: Computer Carrier
Right Hand Video Plug

Indicates Cable Run Number

Upgrade38

GE MEDICAL SYSTEMS
REV 5

5.10

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Rx Monitor Detail
Note 9

Console PN
21384002

The Rx monitor is color and may have a touch option.

Illustration 115 CT/I CONSOLE RX MONITOR WIRING DETAIL

Rear View
Display Monitor

Rx Monitor

RX MONITOR

RGB

110VAC
From: P1

2139581 Touch 1
From: Specialix #2

1
2142221 Video
From: Computer Carrier
Left Hand Video Plug

Notes:
1
Option: Shown for reference, cable
may or may not be installed on this
system.

Indicates Cable Run Number

Upgrade39

GE MEDICAL SYSTEMS
REV 5

5.11

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Octane Rear Access

Console PN
2169455

To install cables at the rear of the Octane in the following subsections it will be
necessary to gain access to the rear of the Octane computer.
(1) Remove the screws at locations A and B.
(2) Slide the drawer forward.
(3) Remove the velcro strap and metal bracket, see Illustration 117, page
Upgrade41.

Note 10

To prevent overheating of the Octane computer you must remove the velcro strap
before turning on power.

Note 11

Once cable installation is complete return to step (4)


(4) Reverse the procedure to secure the Octane back into the console. Be aware of
tight clearance points while sliding the tray back in.

Location C, between the Octane and SCSI connector.


Illustration 116 CT/I CONSOLE

Front View Console


C

OCTANE

Upgrade40

GE MEDICAL SYSTEMS
REV 5

5.12

Remove Shipping Strap and Bracket

Illustration 117 STRAP DETAIL

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Console PN
2169455

WARNING: Strap must be removed before power is turned on.


The strap is blocking cooling air vents.

Remove strap before turning on power.


The strap is blocking cooling air vents.
Store strap for future use.

Remove and store bracket.

Many items removed for clarity

Upgrade41

GE MEDICAL SYSTEMS
REV 5

5.13

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Modem Installation (Hardware Option)


How to Install

Console PN
2169455

(1) Some Consoles will come with the modem already installed.
(1) Refer to Illustration 118 and Illustration 119.
(2) Feed cable 2206138, modem power and phone line through the shelf provided for the modem
from the back of the console to the front.
(3) Connect cable 2206138, power and phone line to the modem.
(4) Place the modem in the console on the shelf provided, see Illustration 118.
(5) Plug the modem power cord/pack into P12, see Illustration 115.
Illustration 118 MODEM DETAIL 1

Place Modem Here

Modem Cables Shown;


Serial
Line In
Power

MODEM

Upgrade42

GE MEDICAL SYSTEMS
REV 5

5.14

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Modem Installation (Hardware Option)

Console PN
2169455

(6) Connect the other end of cable 2206138 to serial port 1 of the Octane.
Illustration 119 MODEM DETAIL 2

OCTANE
MODEM

2206183

Connect to Serial Port 1

Many items removed for clarity

Upgrade43

GE MEDICAL SYSTEMS
REV 5

5.15

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Dasm Installation (Hardware Option)


How to Install

(1) Refer to Illustration 120.


(2) Remove screws (items C) and cover (item A) and set aside.
(3) Verify/Remove third jumper pin from ribbon cable (view B).

Note 12

Following this step will ensure the DASM is set to SCSI id 1.

(4) Replace cover and secure with screws.


Illustration 120 DASM DETAIL 1

BOTTOM
VIEW
DASM

Upgrade44

Console PN
2169455

GE MEDICAL SYSTEMS
REV 5

5.16

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Dasm Installation (Hardware Option)


(5) Refer to Illustration 121, Analog DASM only.

Console PN
2169455

(6) Attach cable 46296980G1 to front of DASM.


Illustration 121 DASM DETAIL 2

Analog DASM

Many items removed for clarity

Upgrade45

GE MEDICAL SYSTEMS
REV 5

5.17

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Dasm Installation (Hardware Option)

(7) Refer to Illustration 122, disconnect the Central Data Box end of SCSI cable A.
Illustration 122 DASM DETAIL 3

Rear View of Octane and Central Data Box


SCSI Cable Configuration without DASM

OCTANE
CENTRAL
DATA
BOX

Many items removed for clarity

Upgrade46

Console PN
2169455

GE MEDICAL SYSTEMS
REV 5

5.18

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Dasm Installation (Hardware Option)


(8) Refer to Illustration 123.

Console PN
2169455

(9) Install SCSI 3 to 1 Adapter to the back of the DASM.


(10)Attach the free end of cable A to adapter installed in step (9).
(11)Attach cable B between the Central Data Box and DASM as shown.
(12)Plug DASM power cord into J10, see Section 5.10.
(13)Reattach remaining cable(s) from camera to DASM.
Illustration 123 DASM DETAIL 4

Place the DASM inside the console next


to the Octane. No
mounting is provided.

Use SCSI 3 to
1 Adapter for
cable coming
from Octane

CENTRAL
DATA
BOX

OCTANE
DASM

Many items removed for clarity

Upgrade47

GE MEDICAL SYSTEMS
REV 5

5.19

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Console Power

Console PN
2169455

Illustration 124 CT/I CONSOLE RX WIRING DETAIL

Front View

J# Device

J#

Device

J1 Rx Monitor

J9

Vieb 

J2 Powered Transceiver 

J10

DASM 

J3 Kieb 

J11

Image Printer 

J4 VCR 

J12

Modem 

J5 Printer 

J13

Remote Monitor 

J6 Video Splitter 

J14

Spare

J7 Display Monitor

J15

Central Data Box 

J8 Octane 

J16

EMC Chassis 

J1

J3

J5

J7

J9

J11

J13

J15

J2

J4

J6

J8

J10

J12

J14

J16

Notes:
 Shown for reference, cable already installed.
 Option: Shown for reference, system may or may not have this cable

Upgrade48

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

5.20

Console PN
2169455

Incoming Power Detail For a System with Global or Compact PDU


If you have a PDU PN 46329106G1 or 46296220G1 see Section 5.21
Illustration 125 CT/I CONSOLE INCOMING POWER WIRING DETAIL

Front View

J9

J11

J13

J15

J10

J12

J14

J16

12
Console Power 2118139
From: PDU A4J5

Indicates Cable Run Number

Upgrade49

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

5.21

Console PN
2169455

Incoming Power Detail


For a System with a PDU PN 46329106G1 or 46296220G1
If you have a Global or Compact PDU see Section 5.20
Illustration 126 CT/I CONSOLE INCOMING POWER WIRING DETAIL

You will need this Section 5.21 if your RP Console was


powered from a wall outlet and not a power cord coming
from the PDU. See Illustration 139 cable run 12 for more
information. In this situation two options exist for cable run
12.

Front View

(1) You can reuse the power cord ( 46278722G1 or


46327187G1 ) which connected the SCU to the wall power
for cable run 12.
or
(2) If your site requires a more pliable power cord. You can
rework the power cord 2118139 that came with the kit. To
perform the rework follow the steps below.

J9

J11

J13

(a) Cut cable 2118139 to the desired length measuring from


the female connector end.

J15

(b) Remove the male plug from the power cord which was
used on the RP Console that was just removed.
J10

J12

J14

(c) Install the male plug on cable 2118139.


see detail (a) for plug wiring.

J16

detail a

Blue

12
Brown Z
Console Power:
2118139
or 46278722G1
or 46327187G1
From : RP Console Wall Outlet

Indicates Cable Run Number

Upgrade50

X blank
Y
Black

Center
Yellow/Green

GE MEDICAL SYSTEMS
REV 5

5.22

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Computer Carrier

Illustration 127 COMPUTER CARRIER DETAIL

Note:

Front View

Console PN
2169455

Cable runs 1 and 3 may already be


connected at the Octane end. The
monitor end of these cables can be
found at the rear cable exit of the
console. The monitor power cords
will also be plugged into the power
panel and be attached to the video
cables.

Rear View of Octane


Headphone
Cable to
Bulkhead

Headphone

Microphone
Cables to
Bulkhead

L
IN
OUT

Coaxial

SCSI Cable to
Central Data
Box or DASM

IN
Optical
OUT

3D

PCI BIT 3

Cable to Network
Transceiver
Mouse: Route
through top using
hole provided

Display Video
2142221 Color
From: Display
Monitor

Keyboard
Security Cable

1
Keyboard Mouse

Rx Video
2142221
From: Rx
Monitor

3D

110 VAC

Cable to modem
Notes:
Shown for reference, cable already installed.
Indicates Cable Run Number

Upgrade51

GE MEDICAL SYSTEMS
REV 5

5.23

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Back Bulkhead 1

Illustration 128

Console PN
2169455

Front View

Rear View

Routed from front


of FEP board.

LAN 2118150
From: Gantry A1A5J3
DAS Data: 21178482
From: Gantry A1A5J5

Indicates Cable Run Number

Keyboard 2139579
From: Keyboard

Upgrade52

10

11

Scan Abort 2118148


From: Gantry A1A5J4

See Illustration 129 on page


Upgrade53 for cable run 8.

GE MEDICAL SYSTEMS
REV 5

5.24

Back Bulkhead 2

Illustration 129

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Console PN
2169455

Rear View

2139579 *
Intercom

Scan Control

*Alternate method for cable run 8 is


to run a completely new cable PN
2139577 from J20 to A1A5J1
Indicates Cable Run Number

EStop

Existing cables
disconnected from
RP Console

A10J8 EStop

A10J9 Intercom

A10J4 Scan Control

Upgrade53

GE MEDICAL SYSTEMS
REV 5

5.25

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Display Monitor Color Option

Illustration 130

Rear View

Display Monitor

Rx Monitor

DISPLAY MONITOR
COLOR OPTION

RGB

110VAC
From: P7

2142221 Video
From: Computer Carrier Top Video Plug
Indicates Cable Run Number

Upgrade54

Console PN
2169455

GE MEDICAL SYSTEMS
REV 5

5.26

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Rx Display Monitor

Illustration 131

Console PN
2169455

Rear View

Display Monitor

Rx Monitor

RX MONITOR

RGB

110VAC
From: P1

1
2142221 Video
From: Computer Carrier
Bottom Video Plug
Indicates Cable Run Number

Upgrade55

GE MEDICAL SYSTEMS
REV 5

5.27

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Keyboard / Trackball Detail


Connect the Trackball and Keyboard cables as shown. Take note that the angled
end of the keyboard cable connects to the keyboard not the straight end.
Do not connect the mouse to the keyboard. Route the mouse cable through the
console top and connect it directly to the computer carrier. The mouse is cable run
5.

Illustration 132 CT/I CONSOLE KEYBOARD / TRACKBALL WIRING DETAIL

Trackball

4
Keyboard Cable

Indicates Cable Run Number

Upgrade56

GE MEDICAL SYSTEMS
REV 5

5.28

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Gantry/Console Cable Connections


Select the console type that applies: INDIGO, Section 5.28.1 or
OCTANE, Section 5.28.2.

5.28.1

EStop/Intercom/Scan Control

Console PN
21384002

(1) At the console; cables A10J4 (Scan Control), A10J8 (EStop) and A10J9
(Intercom) have been disconnected from the RP Console. Use adapter cable
2139578 at the console end of the cable to make the connection to CT/i Console
Rear Bulkhead J20. See Illustration 113 and Illustration 139.

Note 13

Alternate Method for cable run 8.


If desired, the Console Gantry Signal Cable may be removed from A1A5 J1 and
replaced with Upgrade Kit cable 2139577 between Gantry A1A5 J1 and CT/i Console
Rear Bulkhead J20.
(1) Connect new cable from upgrade kit 2139577; from Console J20 to Gantry A1A5
J1. See Illustration 113.

5.28.2

EStop/Intercom/Scan Control

Console PN
2169455

(1) At the console; cables A10J4 (Scan Control), A10J8 (EStop) and A10J9
(Intercom) have been disconnected from the RP Console. Use adapter cable
2139578 at the console end of the cable to make the connection to CT/i Console
Rear Bulkhead J20. See Illustration 129 and Illustration 142.

Note 14

Alternate Method for cable run 8.


If desired, the Console Gantry Signal Cable may be removed from A1A5 J1 and
replaced with Upgrade Kit cable 2139577 between Gantry A1A5 J1 and CT/i Console
Rear Bulkhead J20.
(1) Connect new cable from upgrade kit 2139577; from Console J20 to Gantry A1A5
J1. See Illustration 129.

Upgrade57

GE MEDICAL SYSTEMS
REV 5

5.28.3

Autovoice

Note 15

Upgrade58

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Console PN
21384002

This cable is not needed with software revision 4.1 or greater. The LAN Cable now
carries this function.

GE MEDICAL SYSTEMS
REV 5

5.28.4

Autovoice

Note 16

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Console PN
2169455

This cable is not needed with software revision 5.3 or greater. The LAN Cable now
carries this function.

Upgrade59

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Illustration 133 GANTRY A1A5 PANEL LOCATION

A1 A5 panel
Detail A

Gantry Front View

Illustration 134 CT/I GANTRY INTERCONNECT CABLE CONNECTIONS

J3

J1
J7

8
2139577

10
2118150

J2

15

J4
J5

2118146

11
(J6 does not exist
on an EMC gantry)

J6
Used for VX only
(not used on RP or IG)

EMC gantry has 4position TS1.


(Terminal Strip 1.)

Upgrade60

9
21178482

A1A5 Panel
Detail A

CAUTION

**

2118148

Indicates Cable Run Number

** Use extreme caution when installing the fiber optic DAS data cable. (Part
Number 21178482). It is fragile and should not be stepped on, kinked or bent
sharply.

GE MEDICAL SYSTEMS
REV 5

5.28.5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

SCOM/FEP Fiber
Look at the SCU and determine if your system is currently using a FEP Fiber link to
the gantry.
If it is this Section 5.28.5 is complete, continue with Section 5.29.
If not continue with the following steps 5.28.5.1 and 5.28.5.2.

5.28.5.1

SCOM Board

The SCOM board is a static sensitive device, be sure to use ESD procedures. Use wrist strap
provided.
Remove the cover to the STC and determine if the SCOM board has a FEP Fiber
connection.
If it does leave the board installed and continue with step 5.28.5.2.
If it does not replace the board using the following steps.
(1) Disconnect the coax cables attached to the Axial board in the STC.
(2) Remove the SCOM Board
(3) Verify the dip switch on the SCOM board (S1) has switch 1 on and switches 2, 3
and 4 are off.
(4) Install the SCOM board in the Gantry STC. The SCOM board is installed in the
A5 slot (far right).
(5) Install the ground lead from the SCOM board to the STC chassis.
(6) Connect the coaxial cables attached to the Axial board in the STC.

5.28.5.2

FEP Fiber Cable


(1) Route the fiber cable from the SCU to the gantry and tyrap as shown in
Illustration 135. Verify that the fiber cable does not interfere with the gantry
shrouds when the gantry is tilted.
(2) Connect the fiber cable to the fiber connector on the SCOM board in the STC.
Connect the black ground wire (46327241G1) from the SCOM board to the
STC power supply mounting plate as shown on Illustration 135 detail A.
(3) Install the new STC cover provided (2101947). Extra cover screws and washers
are included, if needed.

Upgrade61

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Illustration 135 FEP FIBER CABLE ROUTING

SCOM

46327241G1
SCOM Lead Data
(See Detail A).

Attach SCOM
ground wire to
lower screw.
Detail A

21178482 Fiber Optic Cable


TyRap cable to existing harness and
to the outside of the flexible trough

Upgrade62

GE MEDICAL SYSTEMS
REV 5

5.29

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

System Ground Connections

Illustration 136 SYSTEM GROUND WIRING


PDU

A3 EQUIPMENT
GROUND BUS

A4 GROUND BUS

GANTRY

# 1/0

VAULT
GROUND

# 2*

CT2

#4ROTATING PART
#4

CT1 RACEWAY
GROUND BUS

#1/0*

TABLE/GANTRY
RACEWAY

SEE TABLE BELOW


FOR CONNECTIONS
TABLE
CT 1
# 2*

# 2*

20

E
T
C

The 1 inch braid is


attached to the table
with two metal
plates,
46297427P1 .
Position the braid
under each plate
and screw down the
plate with the
two bolts.

DETAIL

CONSOLE
(OC)

* use ground wire supplied with system for these grounds

Upgrade63

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Table 12
CT/I GROUND WIRE SIZES AND CONNECTIONS

AWG

CONNECTION

P.M.

Upgrade64

INSTALLED
Connection To

Connection To

#1/0**

EARTH GROUND

PDU

#2

OC

CT1 RACEWAY #10

#2

CT2

CT1 RACEWAY #9

#2

CT1

CT1 RACEWAY #8

#1/0

PM

CT1 RACEWAY #11

GND Braid (2)

CT1

CT2

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

NOTES

Upgrade65

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Illustration 137
NON EMC HISPEED CT/I INTERCONNECT DIAGRAM

USE ON PDU 2133533, 2113764


How to use this page: Check box when cable end is installed.

TABLE
CT1
(YELLOW)

A NEW 2118139
CABLE IS
INCLUDED IN THE
UPGRADE KIT.

GANTRY 
CT2
(BLUE)

PDU
PM
(RED)

Indicates Cable Run Number

CONSOLE

12

Power Panel

HATCHED CABLES ARE


MOVED FROM THE SCU
CABINET TO THE NEW
CT/I CONSOLE

2118139 Console Power

Back Bulkhead

Back Bulkhead

11

2118148 Scan Abort

10

2118150 LAN

21178482 DAS Data 

A1A5
J4

2139577 Intercom, Scan, EStop Adapter

Back Bulkhead J20

NEW CABLE FROM


UPGRADE KIT

q HVDC Power q

46297080G1 (HVDC)

Use Existing cable with adapter 2139577 or run


new cable 2139577 Intercom, Scan, EStop q

A1A5
J1

Axial Motor q
Power 46327155G1

A1A5
J2

LAN

Tilt Limit Switch

2139581 Touch

A1TS1
1, 2, 3

A4 J4

q 120VDC Power q
FIXED

Tilt Motor Power

Signal

2119603

FIXED

A2A7J5

A4 J2

16
FIXED

A2A7J3

Specialx #2

Signal
2118146

FIXED

q
A2A7J2

RX
MONITOR

A1 TS11
TS12
TS13

15

A2A7J1
2142221 Video

A2 TS11
TS21
GND

14

A1TS1
9, 10, 11

TABLE
3

A4 J5

DISPLAY
MONITOR

MOUSE
5

13

A1TS1
5, 6, 8

A1A5
J3
A1A5
J5

DAS Data Fiber Connected to FEP

PDU

GANTRY

COMPUTER
CARRIER

Console PN
21384002

17

Back
Bulkhead

2142221 Video

A2A7J7

A4 J3

2118141 Table Power

Footswitch

Footswitch

A2A7J8

A2A7J6

KEYBOARD

Raceway
FootFIXED switches

FIXED

4
2139579 Keyboard q

NON EMC ONLY

TrackBall 
6

Upgrade66

NOTES:
NonEMC gantry has a 12position TS1.
2
Use extreme caution when installing the fiber optic
das data cable, part #21178482. It is fragile and
should not be stepped on, kinked or bent sharply.
3
A variety of Monitors are available as options, see
pages 7, 1112 for instructions.
4
Option.
1

GE MEDICAL SYSTEMS
REV 5

Illustration 138
EMC HISPEED CT/I INTERCONNECT DIAGRAM

USE ON PDU 2133533, 2113764


How to use this page: Check box when cable end is installed.

TABLE
CT1
(YELLOW)

HATCHED CABLES ARE


MOVED FROM THE SCU
CABINET TO THE NEW
CT/I CONSOLE

12

Power Panel

Console PN
21384002

A NEW 2118139
CABLE IS
INCLUDED IN THE
UPGRADE KIT.

GANTRY 
CT2
(BLUE)

2118139 Console Power

2118148 Scan Abort

10

Back Bulkhead

2118150 LAN

DAS Data Fiber Connected to FEP


Back Bulkhead

2139577 Intercom, Scan, EStop Adapter

21178482 DAS Data 

NEW CABLE FROM


UPGRADE KIT

Computer Carrier

MOUSE

Use Existing cable with adapter 2139577 or run


new cable 2139577 Intercom, Scan, EStop q

2118143(550)

A1A5
J5

A1TS1
3, 4

Axial Motor
Power 2118145

2
Specialx #2

Back
Bulkhead

2139581 Touch

RX
MONITOR

A2A7J5

q
A2A7J7

A1A5
J2

LAN

Tilt Limit Switch

FIXED

2142221 Video

Signal

2118141 Table Power

FIXED

A2A7J6
KEYBOARD
4

Tilt Motor Power

EMC ONLY

Footswitch

Footswitch

Signal
2118146

A4
J2

A4
J4

Line Filter
q 120VDC Power q
Top: Blue
2118142
Middle: Green/Yellow
Bottom: Brown

FIXED

16

17
A4 J3
Raceway

TrackBall 

*
* with
* Shields grounded

FIXED

A2A7J2

A2A7J8

2139579 Keyboard q

A1 TS11
TS12

15

A2A7J3

clamp at PDU cabinet and


gantry base

A2A7J1
2142221 Video

14

A1A5
J1

TABLE

A2 TS11
TS21

q HVDC Power q

A1A5
J3

A4 J5

13

A1TS1
1, 2

A1A5
J4

DISPLAY
MONITOR

PDU

GANTRY

11

Back Bulkhead

PDU
PM
(RED)

Indicates Cable Run Number

CONSOLE

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Footswitches
FIXED
FIXED
NOTES:
1
EMC gantry has a 4position TS1.
2
Use extreme caution when installing the fiber optic
das data cable, part #21178482. It is fragile and
should not be stepped on, kinked or bent sharply.
3
A variety of Monitors are available as options, see
pages 7,1112 for instructions.
4
Option.

Upgrade67

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

USE ON PDU 46296220G1. 46327106G1


Illustration 139
HISPEED CT/I INTERCONNECT DIAGRAM

CONSOLE

Back Bulkhead
J17
DLAN

DAS Fiber Optic


Connect to front of FEP

Console PN
21384002

CABLE MOVES FROM


SRU TO CT/I CONSOLE

LAN

46296454G5

A1A5
J6

NEW CABLES FOR


UPGRADE

DAS DATA Fiber Optic Cable

SCOM in
STC

Back Bulkhead
J20
Intercom Audio / EStop /
Scan Control

Adapter Cable

SCAN CONTROL/AUTOVOICE
A10J4

A1A5
J8

46296986G1

2139578

INTERCOM A10J9

A1A5
J5

46297264G1

ETHERNET

(CT2)

10

462118482

XCVR

GANTRY

Indicates Cable Run Number

CABLES
MOVE FROM
GENESIS
CONSOLE TO
CT/I CONSOLE

PDU
(PM)

CONSOLE EMERGENCY STOP


A10J8

A3A3J15

46297089G1

Back Bulkhead EXP INTERLOCK


J18
Exp Ilock / Scan Abort 46297565G1
Power Panel
J17

SRU POWER

A3A3J19

CABLES MOVE FROM SRU TO CT/I CONSOLE

46327187G1 or 46278722G1 or 2139577

SGI Computer Carrier


J41

11

CONSOLE
WALL
OUTLET

12

Mouse

5
Back Bulkhead
J9

SGI Computer Carrier


Slot 5

SGI Computer Carrier


Slot 2
SpecialX MTA
SP2

2139579
Keyboard

2142221
Video
2142221
Video

Keyboard
4

Display
Monitor

Rx
Monitor

2139581
Touch Option
2

Upgrade68

Trackball

HATCHED
CABLES ARE
MOVED FROM
THEIR ORIGINAL
LOCATIONS AT
THE SRU TO THE
CT/I CONSOLE

Indicates Cable Run Number

HATCHED
CABLES ARE
MOVED FROM
THEIR ORIGINAL
LOCATIONS AT
THE GENESIS
CONSOLE TO
THE CT/I
CONSOLE

A3K3, TS5

SHADED
CABLES ARE
NEW FROM
UPGRADE KIT.

GE MEDICAL SYSTEMS
REV 5
Illustration 140
NON EMC HISPEED CT/I INTERCONNECT DIAGRAM

USE ON PDU 2133533, 2113764


How to use this page: Check box when cable end is installed.

TABLE
CT1
(YELLOW)

12

Power Panel

Console PN
2169455

A NEW 2118139
CABLE IS
INCLUDED IN THE
UPGRADE KIT.

GANTRY 
CT2
(BLUE)

HATCHED CABLES ARE


MOVED FROM THE SCU
CABINET TO THE NEW
CT/I CONSOLE

2118139 Console Power

11

2118148 Scan Abort

10

2118150 LAN

21178482 DAS Data 

A1A5
J4

Back Bulkhead J20

2139577 Intercom, Scan, EStop Adapter

NEW CABLE FROM


UPGRADE KIT

13

A1TS1
5, 6, 8

q HVDC Power q

A1A5
J3

46297080G1 (HVDC)

Use Existing cable with adapter 2139577 or run


new cable 2139577 Intercom, Scan, EStop q

14

A1TS1
9, 10, 11

A1A5
J1

MOUSE

DISPLAY
MONITOR

Axial Motor q
Power 46327155G1

LAN

Tilt Limit Switch

A2A7J2
2142221 Video

A4 J2

16
FIXED

A1TS1
1, 2, 3

A4 J4

q 120VDC Power q

A2A7J3
RX
MONITOR

Signal
2118146

FIXED

A2A7J1

15

TABLE

A1 TS11
TS12
TS13

A1A5
J2
COMPUTER
CARRIER

A2 TS11
TS21
GND

A1A5
J5

DAS Data Fiber Connected to FEP

GANTRY

Back Bulkhead

PDU

A4 J5

Back Bulkhead

PDU
PM
(RED)

Indicates Cable Run Number

CONSOLE

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

FIXED

Tilt Motor Power

Signal

2119603

FIXED

A2A7J5

1
17

Back
Bulkhead

2142221 Video

A2A7J7

A4 J3

2118141 Table Power

Footswitch

Footswitch

A2A7J8

A2A7J6

KEYBOARD

Raceway
FootFIXED switches

FIXED

4
2139579 Keyboard q

NON EMC ONLY

TrackBall 
6

NOTES:
NonEMC gantry has a 12position TS1.
2
Use extreme caution when installing the fiber optic
das data cable, part #21178482. It is fragile and
should not be stepped on, kinked or bent sharply.
3
A variety of Monitors are available as options, see
pages 7, 1112 for instructions.
4
Option.
1

Upgrade69

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Illustration 141
EMC HISPEED CT/I INTERCONNECT DIAGRAM

USE ON PDU 2133533, 2113764

How to use this page: Check box when cable end is installed.

TABLE
CT1
(YELLOW)

HATCHED CABLES ARE


MOVED FROM THE SCU
CABINET TO THE NEW
CT/I CONSOLE

12

Power Panel

A NEW 2118139
CABLE IS
INCLUDED IN THE
UPGRADE KIT.

GANTRY 
CT2
(BLUE)

2118139 Console Power

2118148 Scan Abort

10

Back Bulkhead

2118150 LAN

DAS Data Fiber Connected to FEP


Back Bulkhead

2139577 Intercom, Scan, EStop Adapter

PDU

A4 J5

GANTRY

11

Back Bulkhead

PDU
PM
(RED)

Indicates Cable Run Number

CONSOLE

Console PN
2169455

21178482 DAS Data 

NEW CABLES FROM


UPGRADE KIT

Use Existing cable with adapter 2139577 or run


new cable 2139577 Intercom, Scan, EStop q

13

A1TS1
1, 2

A1A5
J4

A2 TS11
TS21

q HVDC Power q

A1A5
J3

2118143(550)

A1A5
J5

A1TS1
3, 4

14

A1A5
J1

Axial Motor
Power 2118145

A1 TS11
TS12

*
* with
* Shields grounded

clamp at PDU cabinet and


gantry base
Octane Computer

MOUSE

DISPLAY
MONITOR

15
TABLE
A2A7J1

FIXED

LAN

Tilt Limit Switch

A2A7J2
2142221 Video

FIXED

A2A7J3
RX
MONITOR

1
Back
Bulkhead

A2A7J5
A2A7J7

2142221 Video

A2A7J6
KEYBOARD
4

EMC ONLY

Upgrade70

Signal

2118141 Table Power

FIXED

TrackBall 

Tilt Motor Power

Footswitch

Footswitch

Signal
2118146

A4
J2

A4
J4

Line Filter
q 120VDC Power q
Top: Blue
2118142
Middle: Green/Yellow
Bottom: Brown

FIXED

16

17
A4 J3
Raceway

A2A7J8

2139579 Keyboard q

A1A5
J2

Footswitches
FIXED
FIXED
NOTES:
1
EMC gantry has a 4position TS1.
2
Use extreme caution when installing the fiber optic
das data cable, part #21178482. It is fragile and
should not be stepped on, kinked or bent sharply.
3
A variety of Monitors are available as options, see
pages 7,1112 for instructions.
4
Option.

GE MEDICAL SYSTEMS
REV 5

USE ON PDU 46296220G1. 46327106G1


Illustration 142
HISPEED CT/I INTERCONNECT DIAGRAM

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100
Console PN
2169455
GANTRY

CONSOLE

Back Bulkhead
J17
DLAN

DAS Fiber Optic


Connect to front of FEP

CABLE MOVES FROM


SRU TO CT/I CONSOLE

(CT2)

Indicates Cable Run Number

LAN

46296454G5

A1A5
J6

10

NEW CABLES FOR


UPGRADE

DAS DATA Fiber Optic Cable


462118482

SCOM in
STC

Back Bulkhead
J20
Intercom Audio / EStop /
Scan Control

Adapter Cable

SCAN CONTROL/AUTOVOICE
A10J4

A1A5
J8

46296986G1

2139578

INTERCOM A10J9

A1A5
J5

46297264G1

CABLES
MOVE FROM
GENESIS
CONSOLE TO
CT/I CONSOLE

PDU
(PM)

CONSOLE EMERGENCY STOP


A10J8

A3A3J15

46297089G1

Back Bulkhead EXP INTERLOCK


J18
Exp Ilock / Scan Abort 46297565G1
Power Panel
J17

SRU POWER

A3A3J19

CABLES MOVE FROM SRU TO CT/I CONSOLE

46327187G1 or 46278722G1 or 2139577

SGI Computer Carrier


J41

11

CONSOLE
WALL
OUTLET

12

Mouse

5
Back Bulkhead
J9

SGI Octane Computer


Top Video Head

SGI Octane Computer


Bottom Video Head

2139579
Keyboard

2142221
Video
2142221
Video

Keyboard
4

Display
Monitor

Rx
Monitor

Trackball

HATCHED
CABLES ARE
MOVED FROM
THEIR ORIGINAL
LOCATIONS AT
THE SRU TO THE
CT/I CONSOLE

Indicates Cable Run Number

HATCHED
CABLES ARE
MOVED FROM
THEIR ORIGINAL
LOCATIONS AT
THE GENESIS
CONSOLE TO
THE CT/I
CONSOLE

A3K3, TS5

SHADED
CABLES ARE
NEW FROM
UPGRADE KIT.

Upgrade71

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

SECTION 6
COMPUTER INTEGRATION

DEFINITIONS
Boot Level
This is the software level where no software is running other than what can be run out of
CPU firmware. On GENESIS this was often referred to as Prom Monitor or Boot
Prompt.
Operating System Level (Irix)
This is the software level in between Boot Level and Applications Level. In GENESIS,
this was often referred to as the Unix or Operating System level. The HiSpeed CT/i
system will normally start and login as user ctuser leaving the User Interface ready for
selection of Irix Commands or startup of the Scanner Applications Software.
Applications Level
This is the software level where the Scanner specific software has been initialized and
the system could be used to: scan, archive, display, film, network, etc...

6.1

Introduction
This section describes the reconfig, options installation and touch monitor calibration for
a HiSpeed CT/i System which has had the operating system, platform and diagnostic
software previously loaded.

6.2

Upgrade72

Tools Needed:
GE Service Key (J25)
Option MOD (optional)

GE MEDICAL SYSTEMS
REV 5

6.3

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Requirements/Assumptions:
(1) All stations in a suite must have the same SUITE ID.
(2) The name of all hosts in the suite and their respective Internet/Ethernet numbers.
(3) The first time a reconfig is executed on the system the following information is
needed: Internet (IP) addresses. If the system is to be connected to a network
contact your system administrator to obtain the IP address for the computers in the
suite. If the system is standalone (not connected to any network) the default
internet numbers can be used.
CAUTION

6.4

This procedure instructs the operator to REGEN DATABASE


when executing a reconfig. This is required during the initial
installation or Upgrade of a System because the Databases are
not compatible due to the load/duplication processes used by
Manufacturing to create these disks. During an initial Installation
or Upgrade it is assumed that disks do not contain any
Calibration or Image Data. The reconfig process, as described
in this procedure, will destroy any Image or Calibration Data that
exists on the System. If a reconfig must be executed to change
Site Specific parameters it is strongly recommended that all Info,
Image and Calibration Data be saved on Tape to avoid possible
loss.

Install Service Key


(1) Install the service key into J25, located on the front of the console.

Illustration 143
SERVICE KEY LOCATION

Service Key
(top connector)

Upgrade73

GE MEDICAL SYSTEMS
REV 5

6.5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Power Up RCU
(1) Power up the RCU.
(2) When the window appears Starting up the system . . .
Select

6.6

Stop for Maintenance

Keyboard Configuration
The CT/i System can be used with 1 of 4 keyboards;

o English US (default)
o German DE
o Swedish SE
o French FR

The computer leaves the factory configured for the English US keyboard. If your system
is using that keyboard then continue with section 6.6.2.

6.6.1

NonEnglish Keyboard Configuration


A screen will pop up with several icons:

Start System
Install System Software
Run Diagnostics
Recover System
Enter Command Monitor
Select Keyboard Layout

(1) Use the mouse and click on Select Keyboard Layout


(2) Select the appropriate keyboard type. The currently supported keyboards are:

o English US (default)
o German DE
o Swedish SE
o French FR

(3) Select Apply to reconfigure.

Upgrade74

GE MEDICAL SYSTEMS
REV 5

6.6.2

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Start System
(1) Select Start System.
(2) Stop the CT/i AutoStart by selecting

Cancel

when the pink attention

window appears.

6.7

Passwords
All forward production HISPEED CT/i systems have the following default passwords:
root: #bigguy
ctuser: 4$apps
insite: 2getin
service: 4rhelp
diag: <enter>
These defaults were changed because of customer and field requests for tighter security,
especially those operating on networks. We suggest that the default passwords be changed
ONLY by customer request. Contact your support center with the new passwords.

6.8

DASM Checkout (Hardware Option)


Note 17

If a DASM was installed in Section 5, perform the following steps.


(1) Open a UNIX shell as ctuser.
(2) Type/enter

Note 18

scsistat

Observe the following output. Some of the output has been removed to make it generic.
The important parts of the output are; Device 1 1 Disk, this shows that the DASM was
successfully installed as scsi device 1. You should see 3 devices after the DASM, these
are your central data, MOD and CDROM. If these 3 devices are listed it means that the
installation of the DASM did not disrupt the SCSI 1 bus.

Device 0 1
Device 1 1
Device 1 3
Device 1 4
Device 1 6

Disk
Disk
............. DASM
Optical
Comm
CDROM

Upgrade75

GE MEDICAL SYSTEMS
REV 5

6.9

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Touch Calibration (Hardware Option)


If your system has the touch option it will be necessary to calibrate this feature. Follow the
instructions below for touch calibration.
(1) In a Unix Shell window , become root: su
password)
(2) Type : cd

/usr/g/xtouch

(3) Type : setup


Note 19

<Enter>

(then enter the root

<Enter>

<Enter>

The Serial Device Name field should contain: /dev/ttya2


(4) Click on software calibrate
(5) Touch on lower left corner
(6) Touch on upper right corner
(7) Click on save settings
(8) Click on exit

Note 20

These settings will not take effect until the next time ctuser logs in.
(9) In the Unix Shell type: exit

6.10

<Enter> to log out as root.

Restore iso Spot Values


Restore the HSA RP iso Large and Small Spot values to the CT/i System that was saved
in section 2.3 on page Upgrade14 of this direction.
(1) Open a UNIX Shell on the OC.
(2) Type: jot /usr/g/config/ScanHardware.cfg
(a) change isoSmallSpot line to the number recorded in Section 2.3.
(b) To find the value move the mouse to the jot top menu bar to Select pull
down a menu and select Search Forward . Enter isoSmallSpot into the
search filed box and push enter. The isoSmallSpot filed will now be
highlighted. Use the backspace key to remove the default value.
(c) change isoLargeSpot line to the number recorded in Section 2.3.
(d) Use the left mouse button select File Save at the top of the window.
(e) Use the left mouse button select File Exit at the top of the window.
(3) Type: cd /usr/g/config
Copy the ScanHardware.cfg file to the SBC.
(4) Type: rcp ScanHardware.cfg
<suiteID>_sbc0:/usr/g/g_config/scanhardware.cfg

Upgrade76

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

6.11

Reconfig

6.11.1

Rconfig the OC
Note 21

When you reconfigure the OC, you MUST reconfig the SBC. The document collector
box that arrived with your system, contains an OC/SBC System Reconfiguration
procedure, PN 2168070100 for your reference.

Applications must not be up.


(1) Type su

<enter> to change to root

(2) Type cd /usr/g/scripts


(3) Type reconfig.
(4) Select config <enter> to open the reconfig UIF.
(5) Select Regenerate Database? |Yes| on the first configuration screen, to erase the
former calibration information.
(6) Complete the remaining configuration screen data field selections. Refer to the
OC/SBC System Reconfiguration procedure, PN 2168070100 (located in the
document collector box that arrived with your system), for illustrations of the
Configuration screens, and explanations of the data fields.
(7) When you finish all the configuration screen data parameters, select |Accept| at the
top of the screen.
(8) Select |Yes| when the screen prompts you to reboot now.

6.11.2

Rconfig the SBC


Note 22

The OC reconfig program cannot change the timezone on the SBC. Before running
reconfig on the SBC perform the timezone command below.
(1) Open a UNIX shell.
(2) Type/enter cu sbc
(3) Login as root
(4) Type/enter

Note 23

reconfig

Wait for the SBC reconfig to complete. The software automatically reboots the sbc when
it completes the SBC reconfig.
(5) Exit the cu session type ~

i.e.: to exit the cu session press the following keys: <tilde><dot><Enter>

Upgrade77

GE MEDICAL SYSTEMS
REV 5

6.12

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Set Date and Time


(1) Open a UNIX Shell on the OC.
(2) Type/enter su root to change to root.
(3) Type/enter setdate

to set the date and time.

(4) Type exit <enter>

6.13

Applications StartUp
(1) In the Console window type: st

Upgrade78

<Enter>

GE MEDICAL SYSTEMS
REV 5

6.14

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Install Option(s)
Note 24

This system may have a MOD which contains options that the customer has
purchased. If this system has such an MOD install it at this time.
(1) Select |Service|

Service
(2) Select |SYSTEM INTEGRATION|

SYSTEM
INTEGRATION

(3) Select |COMPUTER INTEGRATION|

COMPUTER

INTEGRATION

TABLE / GANTRY INTEGRATION


GENERATOR

CALIBRATION

DAS INTEGRATION
ALIGNMENTS
SYSTEM TESTS

(4) Select |START|

. . . START. . .

Upgrade79

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100
(5) Select |Install Options|

Computer Integration

New Tube Config


Restore System State
Install Options
Verify Options
___
Computer Diags
Display Diags
Operator I/O Diags
RP Diags
Dismiss

(6) Insert the options MOD into the top MOD drive.
(7) Select the options in the lefthand column to install the corresponding software.
(8) Select |install|

6.15

Restore Window / Level Hardkey Values


Restore the HSA RP System Information saved in section 2.1 on page Upgrade13 of this
direction.
The CT/i system must be up and running the CT applications. Connect and startup the CT/i
system as normal before starting this procedure.
Restore the Window / Level values in Image Works.
(1) Select the Image Works desktop.
(2) Select Viewer.
(3) Select User Prefs on the Viewer.
(4) Fill out the Window /Level table for the CT values from the table you saved in
Section 2.7.
(5) Click on Save Defaults.

Upgrade80

GE MEDICAL SYSTEMS
REV 5

6.16

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

New Tube Config


(1) Select |New Tube Config|

Computer Integration
New Tube Config
Restore System State
Install Options
Verify Options

Computer Diags
Display Diags
Operator I/O Diags
RP Diags
Dismiss

(2) Answer the questions.


Note 25

You must perform a system shutdown and startup for the tube usage count to be correct.
Contiune with steps (3) thru (5) to perform this.
(3) Select |Utilities|

UTILITIES

Upgrade81

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100
(4) Select |Applications Shutdown|

.
.
.
.
.
.
.
.
.
.
Applications Shutdown

(5) After Applications have been shutdown, in the Console shell type: st <Enter>

6.17

Verify Options
If options were installed perform this section.
(1) Select |Verify Options|

Computer Integration
New Tube Config
Restore System State
Install Options
Verify Options

Computer Diags
Display Diags
Operator I/O Diags
RP Diags
Dismiss

(2) A shelltool will open: Verify that the options loaded in the previous section are
listed.

Upgrade82

GE MEDICAL SYSTEMS
REV 5

6.18

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

KVTest and ZAlign Baselines


KVTest baseline and ZAlign baseline are not saved from the RP System. KVTest baseline
and ZAlign baseline will need to be reestablished prior to the next tube change.

Upgrade83

GE MEDICAL SYSTEMS
REV 5

6.19

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Monitor Setup
This chapter describes the steps required to set up the two Display Monitors. The flowchart
in Illustration 144 provides an overview of the Monitor Setup procedure. Use the detailed
procedures in the following sections to set up the monitors.

Illustration 144
SETUP FLOWCHART

Start

Factory
PreSet

Set
Color

Adjust Brightness
and Contrast

Adjust H/V Size,


Center and Tilt

Adjust Convergence
Pincushion

End

Upgrade84

Select the 6500K color temperature

GE MEDICAL SYSTEMS
REV 5

6.19.1

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Monitor Manual Use


The following steps are the proper order to follow for adjustment. You will need to reference the
monitor manual that arrived in the monitor box for specific control procedure.

6.19.2

Factory PreSet
(1) Reference the monitor manual and reset all adjustments to factory preset levels.

6.19.3

Center the Raster


(1) Reference the monitor manual and center the raster.

The monitors normally have a blank area to the left and right of the display,
to maintain the correct aspect ratio.

6.19.4

Set Monitor Color Temperature


(1) Reference the monitor manual and select the 6500K color temperature.

6.19.5

Monitor Brightness and Contrast Setup


Color monitors have a lower light output than black and white monitors. Take care when you adjust
the monitor brightness & contrast on CT/i systems.
If the operators tell you they think the image on the monitor looks softer than the image on the film,
and they like the film, adjust the monitor Brightness and Contrast to compensate for its lower light
output. Use the following procedure to adjust the monitor Brightness and Contrast, until the operator
sees the anatomical structure (window width) at the expected window levels.
(1) Adjust the ambient room light to the level used by the operator during image
review.
(2) Display the Image Works desktop.
(3) Display the SMPTE pattern:

Use the browser to select Exam 1000, which contains the SMPTE pattern,
and enlarge the image to full screen display.
or
(a) Select the Service Desktop
(b) Select |Trouble Shooting|
(c) Select |Display Processing|
(d) Select |Install SMPTE|
(4) Increase the Contrast to maximum.
(5) Increase the Brightness to maximum.
(6) Decrease the Brightness, until the raster just fades into, and matches, the
monitor screen background.

Upgrade85

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

6.19.6

Adjust the Raster Horizontal and Vertical Size and Tilt.


(1) Reference the monitor manual and adjust the horizontal, vertical raster size and
tilt.

6.19.7

Adjust Pincushion
(1) Reference the monitor manual and adjust the pincushion.

6.19.8

Adjust Convergence
(1) Display the Service Desktop Manager.
(2) Select |Operator I/O|
(3) Select |Host Diags|
(4) Select the |Monitor| icon.
(5) Select |Convergence| from the monitor menu, to display the convergence
pattern on the exam Rx monitor.
Note 26

Swap video cables to adjust convergence on the display monitor.


(6) Reference the monitor manual and adjust the convergence.
(7) If you swapped cables to display the convergence pattern on the Display monitor,
restore the cables to the original configuration.

Upgrade86

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100
NOTES

Upgrade87

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

SECTION 7
TABLE GANTRY CHARACTERIZATION

7.1

Collimator Characterization
(1) Select SERVICE

Service

(2) Select SYSTEM INTEGRATION

SYSTEM
INTEGRATION

(3) Select TABLE / GANTRY INTEGRATION

COMPUTER

INTEGRATION

TABLE / GANTRY INTEGRATION


GENERATOR

CALIBRATION

DAS INTEGRATION
ALIGNMENTS
SYSTEM TESTS

Upgrade88

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100
(4) Select START

. . . START. . .

(5) Select Characterization

Table/Gantry Installation
HV DC Enable Switch
Characterization

Power Supply Checks


Collimator
Rotor Verification

Dismiss

Upgrade89

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100
(6) Select Collimator Data
File

Help

Mechanical Characterization Tool


Hardware
Reset
Characterize
Tilt
Table
Elevation
Cradle

Data View
Collimator
Data
Charact
Limits
View
Values
View
Log

(7) Perform

enter the characterization values that were recorded in SECTION 2.4.


Note 27

Use the <tab> key to move to the next data input field for collimator characterization
values.
Select . . . . . Continue

Note 28

Upgrade90

Note message which says that characterization was completed successfully (if you
entered new values).

GE MEDICAL SYSTEMS
REV 5

7.2

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Tilt Characterization
(1) Select

Tilt
Help

File
Mechanical Characterization Tool
Hardware
Reset
Characterize
Tilt
Table
Elevation
Cradle

Data View
Collimator
Data
Charact
Limits
View
Values
View
Log

(2) Perform Tilt Characterization per the instructions given on the screen.

Upgrade91

GE MEDICAL SYSTEMS
REV 5

7.3

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Elevation Characterization
(1) Remove upper right and lower right table covers.
Small Lshaped measurement block is stored on top of the bigger measurement block. Remove
small Lshaped measurement block from its storage position and install it in its characterization
position at the right rear of the table. Move measurement plate on right side of table to out
position. Tighten mounting screws for both plates.
(2) Select Table

Elevation
File

Help

Mechanical Characterization Tool


Hardware
Reset
Characterize
Tilt
Table
Elevation
Cradle

Data View
Collimator
Data
Charact
Limits
View
Values
View
Log

(3) Perform Elevation Characterization per the instructions given on the screen.
Note 29

When measuring elevation distance, measure as vertically as possible. Measure


between the bottom surface of the appropriate measurement block and the center of
the rear leg upper pivot pin, per Illustration 72 below.
This measurement is critical. Elevation characterization will fail if this measurement is
incorrect.

Reinstall measurement plates to original positions.

Upgrade92

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Illustration A1
ELEVATION DISTANCE MEASUREMENT

Rear leg upper pivot pin


Measure
This
Distance
Measurement from bottom of rear measurement block for lower
position
measurement

Gantry

etc board

Top of
Weldment
Plate

Weldment Plate

LOWER POSITION DISTANCE = 14.7 INCHES(373.4MM)


CT38885A

Rear leg upper pivot pin

Measure
This
Distance

Gantry

etc board

Measurement from bottom of right side measurement block


for upper position measurement

Weldment Plate

Top of
Weldment
Plate

UPPER POSITION DISTANCE = 36.44 INCHES(925.6MM)


CT38886A

Upgrade93

GE MEDICAL SYSTEMS
REV 5

7.4

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Cradle Characterization
(4) Select Cradle
Help

File
Mechanical Characterization Tool
Hardware
Reset
Characterize
Tilt
Table
Elevation
Cradle

Data View
Collimator
Data
Charact
Limits
View
Values
View
Log

(5) Perform Cradle CharacterizatIon per the instructions given on the screen.
IMPORTANT CRADLE CHARACTERIZATION INFORMATION
Note 30

The following changes apply to the onscreen directions given for the
alignment light and endofcradle position characterization steps, for tables
equipped with the METALFREE cradle only.

In the alignment light procedure (3rd and 4th screens), use the BLACK
DOT on the cradle rightside bumper, instead of the black stripe, for both
steps. Note that the table height must be at least 40mm from isocenter in
order for the black dot to reach the external scan plane alignment light
before the cradle reaches its home position.

In the endofcradle position procedure (5th screen), use the same


BLACK DOT on the cradle rightside bumper, instead of the end of the
cradle.

Upgrade94

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Illustration A2
CRADLE CHARACTERIZATION PIN LOCATIONS

Cradle Characterization Pin


Storage Location
(Under Front Cover)

Cradle Characterization Pin


Rear Characterization Location

Cradle Characterization Pin


Front Characterization Location

etc board

CT38886A

Gantry

Upgrade95

GE MEDICAL SYSTEMS
REV 5

7.5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Interference Test
At this point you must do a hardware reset to download the new characterization values
to the ETC, OBC, and STC before doing the table/gantry interference test on this page.
(1) Select Hardware Reset
File

Help

Mechanical Characterization Tool


Hardware
Reset
Characterize
Tilt
Table
Elevation
Cradle

Data View
Collimator
Data
Charact
Limits
View
Values
View
Log

Note 31

If the following elevation and/or tilt display readings are incorrect, DO NOT adjust
limit switches. Recharacterize elevation and/or tilt.

Note 32

When discussing gantry tilt on this page, (minus sign) refers to having the top of
the gantry tilted away from the table. + (plus sign) refers to having the top of the
gantry tilted toward the table.

Note 33

Pressing the fast pushbutton only will cause the elevation limits for the present
elevation/tilt combination to be displayed.
(2) Move table to the first position, down and out:
(a) Move the cradle to the home position.
(b) Using the Down pushbutton, lower the table to the minimum height.
(c) Verify that the elevation display reads 577.5 +/ 3 mm.
(3) Move Elevation to maximum:
(a) Using the Up pushbutton, raise the table to the maximum height.

Upgrade96

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100
(b) Verify that the elevation display reads 25.5 +/ 3 mm.
(4) Check Elevation minimum clearance with zero tilt:
(a) Advance the cradle from the home position about 36 inches.
(b) Tilt the gantry to 23 degrees.
(c) Use the Down pushbutton to lower the table until it stops automatically.
(d) Verify that the elevation display reads 115.5 +/ 3 mm.
(e) Verify that minimum clearance between table and gantry covers is no less
than 1/2 inch.
(5) Check Elevation clearance with gantry tilted minus 20:
(a) Tilt the gantry to 20 degrees.
(b) Use the Down pushbutton to lower the table until it stops automatically.
(c) Verify that the elevation display reads 209.5 +/ 3 mm.
(d) Verify that minimum clearance between table and gantry covers is no less
than 1/2 inch.
(6) Check Tilt Interference with table raised:
(a) Using the Up pushbutton, raise the table until the elevation display reads
between 119.5mm and 209.5mm.
(b) Tilt the gantry to 10 degrees.
(c) Tilt the gantry back and verify that it stops at 22.0 +/ 0 degrees.
(d) Tilt the gantry forward and verify that it stops at +18.5 +/ 0 degrees.
(7) Verify +/ 30 degree tilt:
(a) Return the cradle to home position.
(b) Verify that the gantry will now tilt +/ 30.0 degrees.
(c) Leave gantry tilted at 30 degrees.
(8) Check Cradle movement below 209mm:
(a) Raise the table enough to allow cradle movement.
(b) Advance the cradle from the home position about 12 inches.
(c) Using the Home Button, return the cradle to the home position, tilt the
gantry to 0 degrees and lower the table to minimum height.
(d) Verify that no cradle movement is possible below 209.5 +/ 3mm.
(e) Verify that the cradle Lock Indicator on the gantry display is lit.
(9) Select File, Exit
(10)Select Dismiss
(11)Select Cleanup

Upgrade97

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

SECTION 8
MX200 TUBE UPGRADE
If your system will be upgrading to the MX200 tube do so at this point in this upgrade
procedure. The MX200 upgrade procedure will complete any remaining steps for the CT/i
Upgrade. You will not be returning to complete any remaining sections of this manual.

Upgrade98

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

SECTION 9
TUBE CALIBRATION

9.1

New Tube Seed Shift


Use the following sequence to run a partial Install New Tube, and set the calseed values on a new
system.

(1) Select |Generator Calibration|

COMPUTER

INTEGRATION

TABLE / GANTRY INTEGRATION


GENERATOR

CALIBRATION

DAS INTEGRATION
ALIGNMENTS
SYSTEM TESTS

(2) Select |Start|

START

Upgrade99

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100
(3) Select |New Tube Seed Shift|

Generator Installation and Verification


mA Meter Check
kV Test
Auto mA Cal
HHS Scans
HHS Report
Ground Leakage Test
Heat Soak & Seasoning

New Tube Seed Shift


Bleeder Setup (DDC)
Rise and Fall Time (DDC)
Internal Scan Time (DDC)
Heat Soak & Seasoning (Diag)
Dismiss

(4) The screen prompts you to confirm the displayed tube type.
(5) Verify the installed and configured tube types match.
(6) Press (Start Scan) when it illuminates, to automatically run the filament current seed
shift program.
(7) ABORT the program after the filament current seed shift scans, and before ductility
warmup

Upgrade100

GE MEDICAL SYSTEMS
REV 5

9.2

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Auto mA Cal
If you have just run a Heat Soak and Seasoning, wait 30 minutes before you run the Auto mA
Cal.
(1) Select |Auto mA Cal|

Generator Installation and Verification


mA Meter Check
kV Test
Auto mA Cal
HHS Scans
HHS Report
Ground Leakage Test
Heat Soak & Seasoning

New Tube Seed Shift


Bleeder Setup (DDC)
Rise and Fall Time (DDC)
Internal Scan Time (DDC)
Heat Soak & Seasoning (Diag)
Dismiss

(2) Press (Start Scan) when it flashes.


(3) Complete the scans.

Upgrade101

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

SECTION 10
SYSTEM CALIBRATION
All Calibrations plus nnumber correction must be performed to insure optimum image quality.

The QCal procedure must be run before calibration and scanning can begin.
Performing a QCal destroys all old phantom CALS. The following warmups, etc. should
be strictly observed:

The DAS should have power turned ON for at least 2 hours before calibrating.
A very cold tube is required. There should be no exposures for at least one and
one half hours before QCal is started (this also means no tube warmup).
Tests that do not interfere with tube cooling will be completed during this wait.

Verify that the mylar window is centered on the patient alignment lights. If the
mylar window is not properly centered, one of the metal bands of the mylar
window assembly could be shading the detector, causing QCal to fail.

Upgrade102

GE MEDICAL SYSTEMS
REV 5

10.1

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

QCal
(1) Select SERVICE

Service
(2) Select SYSTEM INTEGRATION

SYSTEM
INTEGRATION

(3) Select SYSTEM TESTS

COMPUTER

INTEGRATION

TABLE / GANTRY INTEGRATION


GENERATOR

CALIBRATION

DAS INTEGRATION
ALIGNMENTS
SYSTEM TESTS

(4) Select START

. . . START. . .

Upgrade103

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100
(5) Select Qcal
A Shelltool will open.
System Tests
FastCal
Perpendicular Alignment
Tomo Plane Indication
Exposure Time Accuracy
System Scanning Tests
Set / Adjust N#s
Image Analysis
Phantom Self Calibration
Image Series

Qcal
xtalk
alpha vector
Hot ISO
Air cal
P cal
System Reli Scanning
Save System State
Heat Soak & Seasoning
Streak Test
Teflon Pin Test

(6) Select Cal Types in order


1 Cold Qcal
2 Heat Qcal
3 Hot Qcal
Note 34

Upgrade104

40 scans
30 scans
40 scans

If the time since the last QCal was done is greater than 96 hours, QCals may be
redone to give a higher degree of confidence that the image series will pass. However,
this is not a requirement.

GE MEDICAL SYSTEMS
REV 5

10.2

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Crosstalk Calibration
(1) Select xtalk
System Tests
FastCal
PerpendicularAlignment
Tomo Plane Indication
Exposure Time Accuracy
System Scanning Tests
Set / Adjust N#s
Image Analysis
Phantom Self Calibration
Image Series

Qcal
xtalk
alpha vector
Hot ISO
Air cal
P cal
System Reli Scanning
Save System State
Heat Soak & Seasoning
Streak Test
Teflon Pin Test

A Shelltool will open.


(2) Select Cal Types in order
4 Xtalk Air

4 scans

Be sure the table is referenced so the table height is reported on the gantry. Position the table
as low as possible. Adjust the phantom holder so it is as high as possible with respect to
the table. Place the Crosstalk phantom on the phantom holder and center it using the
alignment lights. Raise the table 180 mm from the centered position. Now execute the
Crosstalk phantom scans:
Select Cal Type
5 Xtalk Phantom

8 scans

Upgrade105

GE MEDICAL SYSTEMS
REV 5

10.3

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Alpha Vector Cal Generation


Note 35

If AutomA calibration has not been run recently ( <4hours), do so now.


(1) Select alpha vector
System Tests
FastCal
PerpendicularAlignment
Tomo Plane Indication
Exposure Time Accuracy
System Scanning Tests
Set / Adjust N#s
Image Analysis
Phantom Self Calibration
Image Series

Qcal
xtalk
alpha vector
Hot ISO
Air cal
P cal
System Reli Scanning
Save System State
Heat Soak & Seasoning
Streak Test
Teflon Pin Test

A Shelltool will open.


(2) Select Cal Types in order
6 Afterglow

Upgrade106

2 scans

GE MEDICAL SYSTEMS
REV 5

10.4

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Hot ISO Cal Generation


(1) Select Hot ISO
System Tests
FastCal
PerpendicularAlignment
Tomo Plane Indication
Exposure Time Accuracy
System Scanning Tests
Set / Adjust N#s
Image Analysis
Phantom Self Calibration
Image Series

Qcal
xtalk
alpha vector
Hot ISO
Air cal
P cal
System Reli Scanning
Save System State
Heat Soak & Seasoning
Streak Test
Teflon Pin Test

(2) Perform the scans

Upgrade107

GE MEDICAL SYSTEMS
REV 5

10.5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Perform Air and Phantom Calibrations

(1) Select
Scanner Utilities

Detailed Calibration
(2) Select

(3) Refer to Illustration 145. Select the following screen parameters:


(a) Calibration Type: Air, Small SFOV, Large FOV
(b) KV: Select all the kV settings.
(c) Aperture: Scaled.
Illustration 145
DETAILED CALIBRATION SCREEN

Detailed Calibration

Calibration
Type

Air

Small
SFOV

Large
SFOV

KV

80KV

100KV

120KV

Aperture

Scaled

1mm

3mm

Adjust CT
Number...

140KV

5mm

(4) Run the selected Calibrations.


(5) Perform N number correction.

Upgrade108

7mm

10mm

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100
NOTES

Upgrade109

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

SECTION 11
PACKING MATERIAL FOR RETURN
This section describes the packing for shipment of the GENESIS Console and the
SCU cabinet.

11.1

Console Packing
The console should already be on the shipping dollies with the fischpaper supplied in the
Upgrade Kit between the dollies and the console legs. Verify this now.
(1) Raise the dollies enough to move the console over the 48 x 40 wooden skid that the
CT/i console was delivered on.
(2) Ensure that no part of the console is extending out over the edges of the skid. THis
includes the ground stud on the back of the console.
(3) Using the dollies as a template, drill 4 holes in the skid, and bolt the console to the skid
using the Lag Screws 5/16 x 1 1/2 and washers provided in the Upgrade kit return
materials.
(4) Drill 2 holes in the console lumber brace and bolt it between the dollies using the
provided hardware.
(5) Place a piece of KimPack across the top of the console to protect the bezel.
(6) Place cardboard sheets (1 set 46303734P1) across the top and front of the console.
Bend and tape in place.
(7) Place a plastic bag over the entire console.

11.2

SCU Packing
(1) Install the shipping brace diagonally across the back of the SCU cabinet.
(2) Attach the two smaller yellow brackets to the large yellow bracket loosely with the
provided hardware.
(3) Position the braces diagonally across the back of the scu cabinet between the cabinet
vertical rails and install the hardware to hold it in position.
(4) Tighten the brace hardware to keep the cabinet from twisting during shipment.
(5) Attach the cabinet keys to the cabinet using tape.
(6) Place the SCU cabinet on to the skid (LD00506) provided in the Upgrade Kit Return
Materials.
(7) Bolt the SCU in place on the skid using the L brackets and hardware provided in the
Upgrade Kit Return Materials.
(8) Cover the SCU with a plastic bag provided in the Upgrade Kit Return Materials.

Upgrade110

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

(9) Place cardboard sheets (one set 46166887P1) on the sides of the cabinet and bend
tham into place around the front and rear of the cabinet.
(10) Tape the cardboard to the Cabinet.

11.3

Label Return Material


(1) Fill in the information on page Upgrade112 and Upgrade114 and attach one each to
the console and SCU.

Upgrade111

GE MEDICAL SYSTEMS
REV 5

11.3.1

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Return Information
Call: GEMS Recycling Center
4147476997
Arrange for pickup of the GENESIS Console, SCU Cabinet, and the CT/i Shipping Dollies.
Fill in the requested information on these sheets and attach to the outside of the GENESIS
Console and SCU Cabinet.

HiSpeed CT/i Equipment


Removal Information
(This page is not a substitute for equipment deinstallation cards)

Attention: GEMS Recycling Center


MAXINE or CHRISTINE
FDO:
Site Name:
Removal Date:
Field Engineer:
SCU Serial Number:
Console Serial Number:

Upgrade112

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

NOTES

Upgrade113

GE MEDICAL SYSTEMS
REV 5

11.3.2

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

Return Information
Call: GEMS Recycling Center
4147476997
Arrange for pickup of the GENESIS Console, SCU Cabinet, and the CT/i Shipping Dollies.
Fill in the requested information on these sheets and attach to the outside of the GENESIS
Console and SCU Cabinet.

HiSpeed CT/i Equipment


Removal Information
(This page is not a substitute for equipment deinstallation cards)

Attention: GEMS Recycling Center


MAXINE or CHRISTINE
FDO:
Site Name:
Removal Date:
Field Engineer:
SCU Serial Number:
Console Serial Number:

Upgrade114

GE MEDICAL SYSTEMS
REV 5

1.X / 2.X TO CT/I UPGRADE MANUAL


DIRECTION 2159823100

NOTES