Académique Documents
Professionnel Documents
Culture Documents
4
Atomic Absorption Spectrometry
CHEM 137.1 2L
Group # 2
ABSTRACT
I. INTRODUCTION
The standards and samples were prepared prior to analysis. For the trace
analysis of lead, the water sample was prepared by accurately measuring 1.0 L
and around 400 mL of it was transferred to a 600-mL beaker. 5.00 mL
concentrated HNO3 was then added. The beaker was then placed in a hotplate
and the sample was evaporated to almost 100 mL. The remaining amount of
water was then added and concentrated by evaporating the sample until the
remaining amount was around 100 mL again. Another 5.00 mL concentrated
HNO3 was then added and the beaker was then covered with a watchglass. The
sample was concentrated to around 50 mL, cooled and filtered. The
concentration of lead in the water sample was determined using the external
calibration method and standard addition method. For the standard addition
method, 0.5 mL from the 1000 ppm Pb 2+ stock solution was obtained to prepare
50.0 mL of 10 ppm Pb2+ standard solution. Using this 10 ppm Pb 2+ standard
solution, 0, 1.0, 2.0, 4.0, 5.0 mL were pipetted into five different 50-mL
volumetric flasks. 5.0 mL of the water sample was then added into each of the
volumetric flask and was diluted to mark with deionized water.
On the other hand, for the analysis of calcium and potassium, 2.0 g of
Bear Band powdered milk was accurately weighed and transferred to an
evaporating dish. Three trials were made. The milk sample was placed in a hot
plate until almost black. The evaporating dish was then transferred to the
furnace at 550 for 3-4 hours or until a white/gray residue was observed.
The sample was then removed from the furnace and was cooled to room
temperature. 2 mL of concentrated HNO3 was then added followed by 10 mL of
distilled water. The solution was heated to boiling, cooled and transferred
quantitatively while filtering to a 50-mL volumetric flask and was diluted to mark
with distilled water. From the 1000 ppm Ca 2+ and K+ stock solution, 0.5 mL were
obtained from each to prepare 50.0 mL of 50 ppm Ca 2+ and 50 ppm K+ standard
solutions. Using this 50 ppm Ca 2+ and 50 ppm K+ standard solutions, 0, 1.0, 2.0,
3.0, 4.0, 5.0 mL were pipetted into six different 50-mL volumetric flasks. 5.00
mL of 2% La3+ solution was then added to volumetric flasks for calcium analysis
while 5.00 mL of 5% Cs solution was supplemented to volumetric flasks for
potassium analysis. Each volumetric flask was then diluted to mark using %
HNO3. Signals obtained from the working curve were then compared to the
signals obtained from the milk sample.
All prepared standards and samples were read in AAS. Absorption mode
was used for calcium and trace analysis of lead while flame emission mode was
used for potassium analysis. Figure 4.1 shows the instrument used in the
experiment. Blank readings of each metal analyzed were then subtracted to
correct the output absorbance.
The aerosol particle size must be taken into consideration for aerosol
droplets that are too large can have insufficient exposure in the flame leading to
incomplete atomization aerosol droplets that are too small and can be easily lost
through wall collisions. The exposure of the sample in the flame is dependent on
the flow rate of the fuel and oxidizer, thats why optimization should be applied.
The flame also has an effect on the degree of atomization. The frequently used
flames used in analyses are air-acetylene and nitrous oxide-acetylene. The flame
temperature of the gases is 2300 oC and 2800OC, respectively. Nitrous oxideacetylene is recommended in analyses not solely due to higher flame
temperature but also to some excited cyanogen radicals that can reduce
refractory oxides to atoms. These refractory oxides has the ability to endure
flame without atomization and metals such as Al, Ba, Be, Ca, Sc, Si, Ta, Ti, U, V,
W, Zr can be in the form of refractory oxides. The air-acetylene, on the other
hand, is favored to those elements that have low ionization potentials like Li, Na,
K, Rb, and Cs since a flame with higher temperature can ionize the said
elements that can result to the reduction of the amount of the metal.
The ability of this technique to analyze up to 70 elements is due to the
atomizer. The electrons of the atoms in the atomizer can be promoted to higher
orbitals for a short amount of time by absorbing a set quantity of energy (i.e.
light of a given wavelength). This amount of energy (or wavelength) is specific
to a particular electron transition in a particular element, and in general, each
wavelength corresponds to only one element. This gives the technique its
elemental selectivity. And as the quantity of energy (the power) put into the
flame is known, and the quantity remaining at the other side (at the detector)
can be measured, it is possible, from Beer-Lambert law, to calculate how many
of these transitions took place, and thus get a signal that is proportional to the
concentration of the element being measured.
The light source used in AAS is usually a discontinuous source since a
broad spectrum, produced by a continuous source, is not recommended in AAS
analyses. Discontinuous sources (line sources) used today are hollow cathode
lamps (HCL) and electrodeless discharge lamps. HCL is manufactured from the
element of interest and this lamp is filled with an inert under low pressure. The
mechanism of this source is through the production of inert gas discharge that is
responsible for the vaporization or sputtering of the element in the cathode.
These atoms are then excited via collision with the inert gas ions. Electrodeless
discharge lamp, on the other hand, has the same components inside like that of
HCL. The discharge is commenced by supplying electrons with a Tesla coil. This
type of lamp produces higher intensity atomic spectra than the HCL but the
spectral outputs are often unstable.
Trial 1
Abs
Corrected
Trial 2
Corrected
0
2
4
6
8
10
AbsUnknown
Equation of the line
R2
Calculated Cx, ppm
Average Cx, ppm
Theoretical
unknown, ppm
Percent error (%)
Flow rate: 6.3mL/min
0.0002
0.0000
0.0361
0.0359
0.0655
0.0653
0.0992
0.0990
0.1325
0.1323
0.1597
0.1595
0.0764
0.0014
0.0000
0.0357
0.0343
0.0634
0.0620
0.0971
0.0957
0.1296
0.1282
0.1555
0.1541
0.0761
y = 0.016x + 0.002
y = 0.0155x + 0.0015
0.9989
0.9987
4.65
4.812903226
4.731451613
4.5
5.143369176 %
Trial 1
0.18
0.16
f(x) = 0.02x + 0
R = 1
0.14
0.12
0.1
Absorbance 0.08
0.06
0.04
0.02
10
12
[Pb2+], ppm
Fig. 4.5. Graph of Trial 1 for trace analysis of lead using external calibration.
Trial 2
0.18
0.16
f(x) = 0.02x + 0
R = 1
0.14
0.12
0.1
Absorbance 0.08
0.06
0.04
0.02
10
12
[Pb2+], ppm
Fig. 4.6. Graph of Trial 2 for trace analysis of lead using external calibration.
Table 4.2. Data for trace analysis of lead using standard addition
method.
Vol. of 10 ppm Pb2+,
mL
0
1
2
4
5
Trial 1
0.0095
0.0129
0.0164
0.0239
0.0261
Abs
Corrected
0.0000
0.0034
0.0069
0.0144
0.0166
y = 0.0034x + 7E-05
y = 0.0035x - 8E-05
0.9962
0.9968
4.117647059
4.571428571
4.344537815
Trial 2
0.0079
0.0112
0.0145
0.0223
0.0247
Corrected
0.0000
0.0033
0.0066
0.0144
0.0168
Trial 1
0.02
0.02
f(x) = 0x + 0
R = 1
0.01
0.01
0.01
Absorbance 0.01
0.01
0
0
0
Fig. 4.7. Graph of Trial 1 for trace analysis of lead using standard addition
method.
Trial 2
0.02
0.02
f(x) = 0x - 0
R = 1
0.01
0.01
0.01
Absorbance 0.01
0.01
0
0
0
Fig. 4.8. Graph of Trial 2 for trace analysis of lead using standard addition
method.
For calcium, the absorbance of the standards were obtained and plotted
against their respective concentrations generating the standard curve. From this
curve, interpolation was done to determine the concentration of Ca 2+ in the
sample. Data were summarized and presented in the tables and figure on the
next page.
Abs [Ca2+]
0.0019
0.0229
0.0474
0.0850
0.1174
0.1513
0.0000
0.0210
0.0455
0.0831
0.1155
0.1494
y = 0.0305x - 0.0072
0.9912
Absorbance
0
0.0684
0.0569
521.0966514
0.0564
555.2304716
Calcium
0.16
0.14
0.12
0.1
Absorbance 0.08
0.06
0.04
0.02
0
On the other hand, potassium analysis was employed using the Atomic
Emission technique. The emission of the standards were obtained and plotted
against their respective concentrations generating the standard curve. From the
curves, interpolation was done to determine the concentration of K + in the
sample. Data were summarized and presented in the table and figure below.
Table 4.4. Data for K analysis
Vol. of 50 ppm K+,
mL
0
1
2
3
4
5
Equation of the line
R2
Blank
Trial 1 (m = 2.002 g)
Emission [K+]
0.0000
0.0616
0.1060
0.1329
0.1592
0.1954
y = 0.037x + 0.0166
0.9728
Emission
Calculated CK, ppm
0
0
0.1908
235.4054054
Trial 2 (m = 2.023 g)
Trial 3 (m = 2.013 g)
Average CK, ppm
Wavelength K+: 404.0 nm
50
DFtrial for K+: ( 1 )
0.1367
0.1299
162.2972973
153.1081081
183.6036036
Potassium
0.25
0.2
0.15
Emission
0.1
0.05
0
Another factor that caused deviation from the linearity is the presence of
interferences. Interferences can be classified into spectral, chemical, and matrix.
Spectral interferences occur due to other forms of radiation that overlaps the
light source. This is often observed when organic solvents are used and the
presence of magnesium in sodium analysis, copper in iron, and nickel in iron
analyses. Other spectral interferences are due to some effects that result to the
broadening of spectral line. Some of the broadening effects according to
Sevostianova are:
1 Doppler effect due to the different components of velocity in the line
of observation
2 Lorentz effect due to the presence of foreign atoms in the
environment of emitting or absorbing atoms. The magnitude of the
broadening varies with the pressure of the foreign gases and their
physical properties.
3 Quenching effect - In a low-pressure spectral source, quenching
collision can occur in flames as the result of the presence of foreign
gas molecules with vibration levels very close to the excited state of
the resonance line.
4. Self absorption or self-reversal effect - The atoms of the same kind as
that emitting radiation will absorb maximum radiation at the center of the
line than at the wings, resulting in the change of shape of the line as well
as its intensity.
Matrix interferences, on the other hand, occur due to the variation in the
surface tension and the viscosity of the analyte and the standard solution.
Chemical interferences, the most common of all, arise from the ionization of the
analyte and the formation of some compounds that do not dissociate in the
flame like calcium and strontium phospates, which both leads to a reduction in
the signal observed.
Treatment of the chemical interferences can be done by manipulating the
flame conditions and through the application of certain chemical agents. A low
temperature flame can inhibit ionization of the metals and a high temperature
flame can further atomize compounds. The other technique is the application of
releasing agents, suppressor, and protective agents. A releasing agent like La is
a competing ion that reacts with the interfering substance instead of the metal
of interest, hence releasing the analyte from interferences. Addition of
suppressor like Cs is useful to those elements with low ionization energy like Na.
These suppressors, usually with lower ionization energy than the metal of
interest, are ionized instead of the metal of interest. On the other hand,
protective agents prevent interference by preferentially forming stable but
volatile species with the analyte. Application of these techniques leads to a
greater amount of sample detection. The application of either a releasing agent
or suppressor led to an increase in the concentration of the metal which is
consistent with the effect of such agents. These agents act as shields in order to
protect the metal of interest from various reaction conditions such as ionization
and formation of compounds that are not decomposed in flame.
Where:
Cx = concentration of unknown
b = y-intercept (from the graph of absorbance versus volume
of aliquot)
Cs = concentration of standard solution
= 10 ppm
m = slope (from the graph of absorbance versus volume of
aliquot)
Vx = final volume of unknown
= 50 mL or 5x10-3 L
Cx =
Considering Trial 1:
From the standard solutions:
0.0034
slope
interce
7 x 105
pt
Cs b
Vxm
10 ppm x 7 x 105
Cx=
=4.117647059 ppm
0.0034 x 5 x 103 L
Average Cx =
4.117647059+ 4.571428571
=4.344537815 ppm
2
Using the equation of the line obtained from Beers Law plot: A = abc
Plot absorbance (A) vs. concentration (c)
y = mx + b
A = ac + b
Where
c =
A = absorbance
a = absorptivity/slope
b = y-int
c = concentration
A-b
a
Considering Trial 1:
0.1908- 0.0166
c =
=4.708108108 ppm
0.037
Calculation of the concentration of K+ in milk sample
50
= 4.708108108 ppm x DF ( 1 = 235.4054054 ppm
V. CONCLUSIONS
VI. REFERENCES